RWG 5020 Ex, Ex potentiometer Technical data Electronic position transmitter/potentiometer RWG 5020 Ex On the basis of the actual potentiometer value, the RWG generates a current signal for signalling the valve position. Data 2-wire system Output current IA 4 – 20 mA 1) V 10 – 28.5 V DC Supply voltage U 1) K < 200 mA Short-circuit current I Power P1) < 0.9 W (UV – 10 V)/20 mA Max. load RB Impact of power supply ≤ 0.1 % Load influence ≤ 0.1 % (RB = 0...600 Ω) Temperature impact < 0.1 ‰/K Transmitter potentiometer 5 kΩ 3) Ambient temperature – 60 °C /–40 °C to +60 °C2) Explosion protection II2G Ex ib IIC T4 EC type test certificate PTB 03 ATEX 2176 Notes on table 1) Power supply Power supply via external, intrinsically safe (Ex ia or Ex ib) power supply unit with type test certificate meeting the technical requirements of RWG 5020 Ex. 2) Ambient temperature Depending on temperature range of the actuator: Refer to name plate 3) For heater in switch compartment Wiring 2-wire system RWG 5020 Connector XK for Customer connection 21 22 23 24 4 – 20 mA 89 3: CC ,P$ T = -40..60 °C BK YE RD 0344 Riester GmbH & Co.KG D-79379 Muellheim RWG 5020.2 Ex II2G Ex ib IIC T4 PTB 03 ATEX 2176 Ser.Nr.: 1 2 Setting: 3 4 5 Refer to the operation instructions relating to actuator. We reserve the right to alter data according to improvements made. Previous documents become invalid with the issue of this document. Y007.184/003/en Issue 1.15 Page 1/2 RWG 5020 Ex, Ex potentiometer Technical data, Ex potentiometer Potentiometer Travel sensor for recording the valve position. Data Wire potentiometer Independent linearity ≤ 1 % Widerstand (Option)1) 0.1 kΩ 0.2 kΩ 0.5 kΩ 1 kΩ 5 kΩ Resistance tolerance +/– 5 % Rated power 1.2 W – 30 °C3)/–20 °C to +60 °C2) Ambient temperature Enclosure protection according to IEC 60529 IP 54 Explosion protection II2G Ex db IIC Gb T4 I M2 Ex db Mb EC type test certificate PTB 03 ATEX 1025 U Notes on table 1) Resistance (option) Further variants on request 2) Ambient temperature Depending on temperature range of the actuator: Refer to name plate 3) For heater in switch compartment The position of the valve can be transmitted as a continuous signal by a potentiometer. The potentiometer is installed in the control unit of the actuator. We recommend: Using the potentiometer as voltage divider. Depending on the supply voltage, suitable series resistors are to be provided. Please observe max. rated power. We reserve the right to alter data according to improvements made. Previous documents become invalid with the issue of this document. Y007.184/003/en Issue 1.15 Page 2/2