Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS TEST REPORT No.: 25041C/2007-EMC Test laboratory: Testing Laboratory "CERTIS" ANO NTCCE "ISEP" Address: Bolshaya Puchkarskaya str., 21, St. Petersburg, Russian Federation phone: +7 (812) 232-7352 fax: +7 (812) 232-7352 Compiled by: A. Lipunov Sign: Date: 2007.05.11 Approved by: G. Zargarianz Sign: Date: 2007.05.14 Description THE COMBINED MEASURING DEVICE Trademark: TKA® Model and/or type reference: TKA-PKM "KEEPER" Manufactured by: TKA Scientific Instruments Serial-No: 40 464п Date of receipt of test item: 2007.04.25 rated voltage: 9 VDC (6LF22) rated current: - rated power: - Transportable equipment Permanently installed equipment Accompanying units and accessories: Client: TKA Scientific Instruments Address: Shkapina str. b6, lit.A, St. Petersburg, 198095, Russian Federation Contact person: Tomskiy Konstantin – General director Phone / fax: 007 (812) 274-7443; 710-74-77 Report-No.: 25041C/2007-EMС 1 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS TEST SPECIFICATION Standard Title Electrical equipment for measurement, control and laboratory use – EMC requirements Electromagnetic compatibility. Limits for harmonic current emissions. Electromagnetic compatibility. Limitation of voltage fluctuations and flicker in low voltage supply systems for equipment with rated of 16A or less. EN 61326:2002 + Corrig.1 EN 61000-3-2:2000 EN 61000-3-3:1995 A1:2001 METHODS AND PROCEDURES The standards listed on the previous page refer to the following tests: Basic Standard EN 55022:1998+A1:2000+A2:2003 EN 61000-3-2:2000 EN 61000-3-3:1995 + A1:2001 EN 61000-4-2:95 + A1:98 + A2:2001 EN 61000-4-3:96 + A1:98 + A2:2001 EN 61000-4-4:95 + A1:2001 + A2:2001 EN 61000-4-5:95 + A1:2001 EN 61000-4-6:96 + A1:2001 EN 61000-4-8:93 + A1:2001 EN 61000-4-11:94 + A1:2001 Report-No.: 25041C/2007-EMС Description Information technology equipment- Radio disturbance characteristics Limits and methods of measurement. Conducted Emissions. Radiated Emissions. Electromagnetic compatibility. Limits for harmonic current emissions. Electromagnetic compatibility. Limitation of voltage fluctuations and flicker in low voltage supply systems for equipment with rated of 16A or less. Testing and measurement techniques. Electrostatic discharge immunity test. Testing and measurement techniques. Radiated, radio frequency, electromagnetic field immunity test. Testing and measurement techniques. Electrical fast transient/burst immunity test. Testing and measurement techniques. Surge immunity test. Testing and measurement techniques. Immunity to conducted disturbances, induced by radio frequency fields. Testing and measurement techniques. Power frequency magnetic field immunity test. Testing and measurement techniques. Voltage dips, short interruptions and voltage variations immunity tests. 2 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS Notes and remarks: This TRF contains 15 pages, numbered 1 to 15 This TRF is a part of documentation with identification PERSONNEL - INITIALS USED THROUGHOUT THIS TRF initials used full name (block capitals; in this TRF clearly readable) A.L. Andrey Lipunov signature Released: 2007.05.10 Report-No.: 25041C/2007-EMС 3 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS CORRESPONDENCE ON REQUIREMENTS OF STANDARDS EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements Class A Class B Ö STANDARD FOR LIMITS AND METHODS OF MEASUREMENT Test Verdict P EN 55022:1998+A1:2000 Emission requirements – IT equipment P EN 61000-3-2:2000 Harmonic Current Emission N EN 61000-3-3:1995 + A1:2001 Voltage Fluctuations N EN 61000-4-2:95 + A1:98 + A2:2001 Electrostatic Discharge Requirements P EN 61000-4-3:96 + A1:98 Radiated Radio-frequency EM Field P EN 61000-4-4:95 + A1:2001 + A2:2001 Electrical Fast Transient/burst Requirements N EN 61000-4-5:95 + A1:2001 Surge Immunity Test N EN 61000-4-6:96 + A1:2001 Conducted Disturbances inducted by RF Fields N EN 61000-4-8:93 + A1:2001 Power frequency magnetic field immunity test P EN 61000-4-11:94 + A1:2001 Voltage dips, Interruptions and Variations N SUMMARY OF TESTING: Test Case Verdict : Test case does not apply to the test object......................: N (ot applicable) Test item does meet the requirements.............................: P (as) Test item does not meet the requirements.......................: F (ail) See page 3 “notes and remarks” : (*) Not tested in this Test report. See “notes and remarks” : (#) TEST MODE Mode 1 Measurement of the temperature Report-No.: 25041C/2007-EMС 4 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS TEST EQUIPMENT LIST: Test Equip ID 1. 2. 3. 4. 5. 6. 7. Test Equipment Spectrum Analyzing Hewlett Packard Test Receiver Rohde@Schwarz Biconical Antenna Log-Periodic Dipole Antenna Signal generator MARCONI Signal generator Amplifier RF 0,15-250 MHz Model Serial No. Cal Date NEXT Cal Date E7401A US39150127 31.10.2006 31.10.2007 ESVD 839904/014 10.10.2006 10.10.2007 HK 116 HL 223 840807/02 840578/11 23.09.2006 23.09.2006 23.09.2007 23.09.2007 T2018 118461/036 11.10.2006 11.10.2007 Г4-176 РУ3-33 02128 007505 10.10.2006 17.10.2006 10.10.2007 17.10.2007 1014 21.09.2006 21.09.2007 10.10.2006 10.10.2007 8. Amplifier RF OPHIR 20-1000 MHz 5125F 9. Field Probe HFV-Z 10. Power and Harmonics Analyser with Flicker Meter HA1600 211377 07.12.2006 07.12.2007 Simulator of electrostatic discharge Simulator fast transient (burst) Simulator surges ЭСР-8000К/ЭСР15000 024 10.10.2006 10.10.2007 ИПП-4000 41 29.03.2006 29.03.2007 ИИП-4000 37 25.06.2006 25.06.2007 Simulator dynamic changes pressure network ИПНП-8 35 10.10.2006 10.10.2007 ИМППЧ-1000 9 18.08.2006 18.08.2007 УСР - 30.06.2006 30.06.2007 17. 18. Simulator power magnetic fields Complete set of devices of a communication - outcome Voltmeter Equivalent network В3-57 NNB-111 8663 1383524 10.10.2006 10.10.2006 10.10.2007 10.10.2007 19. Aspiration psychrometer M-34 14785 19.10.2006 19.10.2007 11. 12. 13. 14. 15. 16. 20. TEST SETUP Temperature Relative Humidity Atmospheric pressure Report-No.: 25041C/2007-EMС 25 ± 10 °C 21 - 25 °C 45 – 80 % 84.0 – 106.7 kPa 48 - 54 % 752-765 mmHg 5 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS CONTENTS TEST SPECIFICATION 2 METHODS AND PROCEDURES 2 TEST RESULTS ON CONFORMITY REQUIREMENTS EMC EN 55022. 7 RADIATED EMISSION LIMIT EUT CLASS B 7 TEST RESULT OF RADIATED EMISSION 7 IMMUNITY TEST 10 PERFOMENCE CRITERIA DESCRIPTION 10 ELECTROSTATIC DISCHARGE (ESD) IMMUNITY TEST 11 RADIATED ELECTROMAGNETIC FIELD IMMUNITY TEST (EM) 13 MAGNETIC FILDS EN 61000-4-8 14 EUT Photo 15 Report-No.: 25041C/2007-EMС 6 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS TEST RESULTS ON CONFORMITY REQUIREMENTS EMC EN 55022. RADIATED EMISSION LIMIT EUT CLASS B Tab 1. Band of frequencies, MHz Radiated Emission, dB(mV/m) 30 – 230 30 230 – 1000 37 Note: – On border frequency by rate is smaller importance of tension of field. Test Procedure The EUT and its simulators are placed on a turn table which is 0.8 meter above ground. The turn table can rotate 360 degrees to determine the position of the maximum emission level. The EUT was positioned such that the distance from antenna to the EUT was 10 meters. The antenna can move up and down between 1 meter and 4 meters to find out the maximum emission level. Both horizontal and vertical polarization of the antenna are set on measurement. In order to find the maximum emission, all of the interface cables must be manipulated on radiated measurement. Radiated emissions were invested over the frequency range from 30MHz to1GHz using a receiver bandwidth of 120 kHz. Radiated was performed at an antenna to EUT distance of 10 meters. TEST RESULT OF RADIATED EMISSION Signal Number 1 Frequency (MHz) 77.249650 Report-No.: 25041C/2007-EMС QP (dBuV/m) 14.2 Limit 1 (dBuV/m) 30.0 QP Delta L1 (dB) -15.8 Polarization Hor. 7 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS TEST RESULT OF RADIATED EMISSION Signal Number 1 2 Frequency (MHz) 111.648040 214.727380 Report-No.: 25041C/2007-EMС QP (dBuV/m) 5.0 7.5 Limit 1 (dBuV/m) 30.0 30.0 QP Delta L1 (dB) -25.0 -22.5 Polarization Ver. Ver. 8 ( 15 ) Test report EN 61326:2002 + Corrig.1 TL CERTIS Electrical equipment for measurement, control and laboratory use – EMC requirements TEST RESULT OF RADIATED EMISSION Signal Number [[ No signal list data available ]] Ö Report-No.: 25041C/2007-EMС Frequency (MHz) 300-1000 PASS QP (dBuV/m) - Limit 1 (dBuV/m) 37 QP Delta L1 (dB) - FAIL 9 ( 15 ) Test report EN 61326:2002 + Corrig.1 TL CERTIS Electrical equipment for measurement, control and laboratory use – EMC requirements IMMUNITY TEST PERFOMENCE CRITERIA DESCRIPTION Performance criterion A: During testing, normal performance within the specification limit. Performance criterion B: During testing, temporary degradation, or loss of function or performance which is self-recovering. Performance criterion C: During testing, temporary degradation, or loss of function or performance which requires operator intervention or system reset occurs. Performance criterion D: Degradation or loss of function which is not recoverable due to damage to equipment, components, software, or to loss of data. Essential Continuous Continuous operation unmonitored monitored (functional safety) operation operation A B B C A A A B A B B B A B B C A A A C A B C C ESD EN 61000-4-2 EM EN 61000-4-3 Burst EN 61000-4-4 Surge EN 61000-4-5 Conducted RF EN 61000-4-6 Voltage interrupts EN 61000-4-11 Report-No.: 25041C/2007-EMС Non-continuous operation 10 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS ELECTROSTATIC DISCHARGE (ESD) IMMUNITY TEST Tab 2. Specifications Test method Exposure on Discharge voltage Charge capacitance Discharge impedance Performance criteria: (see generic standards) air contact Exposure during test EN 61326 EN 61000-4-2 Enclosure, Control Unit ±4 kV ±4 kV 150 pF 330 W Essential operation B Test Procedure Direct application of discharges to the EUT: Contact discharge was applied only to conductive surfaces of the EUT. Air discharges were applied only to non-conductive surfaces of the EUT. During the test, it was performed with single discharges. For the single discharge time between successive single discharges will be keep longer 1 second. It was at least ten single discharges with positive and negative at the same selected point. Indirect application of discharges to the EUT: Vertical Coupling Plane (VCP): The coupling plane, of dimensions 0.5m x 0.5m, is placed parallel to, and positioned at a distance 0.1m from, the EUT, with the Discharge Electrode touching the coupling plane. The four faces of the EUT will be performed with electrostatic discharge. It was at least ten single discharges with positive and negative at the same selected point. Horizontal Coupling Plane (HCP): The coupling plane is placed under to the EUT. The generator shall be positioned vertically at a distance of 0.1m from the EUT, with the Discharge Electrode touching the coupling plane. The four faces of the EUT will be performed with electrostatic discharge. It was at least ten single discharges with positive and negative at the same selected point. Report-No.: 25041C/2007-EMС 11 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS Tab 3. Test Result of Air Discharge Test Result of Contact Discharge Criterion of quality of action Test Point Possible Result Block of the processing the signal B А Measuring head B А Ö Report-No.: 25041C/2007-EMС PASS Test Point Criterion of quality of action Possible Result Horizontal (At Front) B А Horizontal (At Left) B А Horizontal (At Right) B А Horizontal (At Rear) B А Vertical (At Front) B А Vertical (At Left) B А Vertical (At Right) B А Vertical (At Rear) B А Measuring head (screws of the fastening) B А FAIL 12 ( 15 ) Test report EN 61326:2002 + Corrig.1 TL CERTIS Electrical equipment for measurement, control and laboratory use – EMC requirements RADIATED ELECTROMAGNETIC FIELD IMMUNITY TEST (EM) Tab 4. Exposure during test EN 61326 EN 61000-4-3 Enclosure 3 V/m 26-1000 MHz 80%, 1 kHz Essential operation A Specifications Test method Exposure on Field strength Frequency range Modulation Performance criteria: (see generic standards) Test Procedure The EUT and load, which are placed on a table that is 0.8 meter above ground, are placed with one coincident with the calibration plane such that the distance from antenna to the EUT was 3 meters. Both horizontal and vertical polarization of the antenna and four sides of the EUT are set on measurement. All the scanning conditions are as follows: Condition of Test Remarks 1. Field Strength 3 V/m Level 2 2. Radiated Signal AM 80% Modulated with 1kHz 3. Scanning Frequency 80MHz - 1000MHz 4 Dwell Time 3 Seconds 5. Frequency step size Δ f : 1% 6. The rate of Swept of Frequency 1.5 x 10-3 decades/s Tab 5. Degrees Possible Result Susceptible (YES or NO) Horizontal 0 A А NO Horizontal 90 A А NO Horizontal 180 A А NO Horizontal 270 A А NO Vertical 0 A А NO Vertical 90 A А NO Vertical 180 A А NO Vertical 270 A А NO Polarisation Ö Report-No.: 25041C/2007-EMС PASS FAIL 13 ( 15 ) Test report EN 61326:2002 + Corrig.1 TL CERTIS Electrical equipment for measurement, control and laboratory use – EMC requirements MAGNETIC FILDS EN 61000-4-8 Tab 6. Field strength Frequency range Ports cases, Display information 3 A/m 50 Hz Performance criteria: (see generic standards) Essential operation A Exposure on Enclosure 3 A/m 50 Hz No visual effects Test Procedure The EUT and its load are placed on a table which is 0.8 meter above a metal ground plane measured at least 1m*1m min. The test magnetic field shall be placed at central of the induction coil. The test magnetic Field shall be applied 10 minutes by the immersion method to the EUT. And the induction coil shall be rotated by 90° in order to expose the EUT to the test field with different orientation (X, Y, Z Orientations). Tab 7. Criterion of quality of action Possible Result Susceptible (YES or NO) A А NO Ports cases Ö Report-No.: 25041C/2007-EMС PASS FAIL 14 ( 15 ) Test report EN 61326:2002 + Corrig.1 Electrical equipment for measurement, control and laboratory use – EMC requirements TL CERTIS EUT Photo Photo 1 Photo 2 Report-No.: 25041C/2007-EMС 15 ( 15 )