The product is the special episcopic LED ring illuminator for the “MULTIZOOM AZ100” microscope. The
illuminator features white LEDs arranged as a ring. It is attached on the tip of the objective for illuminating a
specimen. The product can be attached to an objective of 0.5x or 1x magnification and is recommended for
low reflectance specimens.
Configuration
4
2
6
0
C
O
N
O
R
8
T
1
L
0
O
N
O
F
F
Illuminator
Adapter for the
objective of
1x magnification
Power cord *
Power supply
* The power cord is changed depending on
the local power supply specifications.
Directions for use
Attachable objectives
The product can be attached to the following objectives:
AZ-Plan Apo 0.5x,
AZ-Plan Apo 1x
* The product cannot be attached to the objective of 2x, 4x, or 5x magnification (AZ-Plan Fluor 2x, AZ-Plan
Apo 4x, AZ-Plan Fluor 5x).
Attaching an adapter (for the objective of 1x magnification)
To use the illuminator with the objective of 1x magnification,
attach the special adapter provided with the illuminator.
1. Fully loosen the thumbscrew (clamp screw) on the side
of the illuminator.
Adapter
Fixing screw
for the adapter
(side)
Screw hole
Illuminator
Thumbscrew
(clamp screw)
2. Align the screw hole of the adapter with the clamp
screw of the illuminator and insert the adapter into the
illuminator.
3. Tighten the fixing screw for the adapter on the opposite
side of the thumbscrew (clamp screw) by using a
hexagonal driver provided with the AZ100 Zoom Body.
1
AZ-LED
LED Ring Illuminator
Instructions
Directions for use (continued)
Attaching the illuminator
Attach the illuminator to the objective. For the objective of 0.5x,
the illuminator is attached directly. For the objective of 1x
magnification, use the adapter. (See the item above.)
1. Fully loosen the thumbscrew (clamp screw) on the side
of the illuminator.
2. Be careful not to damage the objective. Attach the
illuminator to the tip of the objective so that the LEDs
are facing downward.
3. Fully raise the illuminator and fix it with the
thumbscrew (clamp screw).
The tip of the thumbscrew catches the groove of the
objective and the illuminator is fixed.
Groove
Thumbscrew
(clamp screw)
Illuminator
(with adapter)
Connecting the power supply
Connect the power supply to the illuminator.
1. Check the position and orientation of the connector on
the side of the illuminator.
2. Connect the cord (DC power cord) fixed to the power
supply.
Draw the cord for your convenient route.
Connecting the power cord
2. Connect the power cord to the AC inlet on the power
supply.
Power switch
Connector
6
8
4
C
O
N
T
R
O
L
10
0
N
2
O
F
F
O
1. Check that the power switch of the power supply is
turned off.
DC power cord
Power supply
Connecting
the power cord
3. Connect another end to the wall outlet.
Illumination control
To turn on the illumination, turn on the power switch on the
power supply. To turn off the illumination, turn off the power
switch on the power supply.
4
6
2
Brightness control
8
0 CONTROL10
Brightness control
ON
To adjust the brightness of the illumination, rotate the
brightness control on the power supply (0 to 100%).
OFF
Power supply
2
Power switch
AZ-LED
LED Ring Illuminator
Instructions
Specifications
Model name
AZ-LED
Light source
White LED (useful life is more than 20,000 hours)
Brightness control range
0% to 100% (brightness)
Packaged parts
Illuminator, adapter, power supply, and power cord
Dimensions (excluding
protrusions)
Illuminator:
Power supply:
ø100 x 26 mm (H)
60.6 mm (W) x 107 mm (D) x 32 mm (H)
Weight
Illuminator:
Power supply:
160 g
150 g
Output ratings (power supply)
21 VDC and 200 mA
Input ratings (power supply)
100 to 240 VAC ±10% and 50/60 Hz, 10 VA
Electric shock protection class
Class 2
Power cord
For Japan:
Use the special power cord, AZ-LED-PCD.
For countries excluding Japan where the power supply is 100 to 120 VAC:
Use the special power cord, AZ-LED-PCU.
For countries where the power supply is 220 to 240 VAC:
Use the special power cord, AZ-LED-PCE.
Operating conditions
Temperature:
Humidity:
Pollution
Installation category:
Altitude:
Indoor
0 to +40°C
85% relative humidity maximum (no condensation)
degree: degree 2
category II
2000 m maximum
use only
Storage conditions
Temperature:
Humidity:
0 to +60°C
90% relative humidity maximum (no condensation)
Safety standards compliance
• CE marking
• This product meets EU Low Voltage Directive requirements.
• This product meets EU EMC Directive requirements. (EN61326)
• Laser safety
This product complies with IEC / EN60825-1 +A2: 2001 Class 1 requirements. (Class 1 LED Product)
• This is the UL-listed product.
• This product meets FCC Part 15B Class A requirements.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial
environment.
This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the
interference at his own expense.
• This Class A digital apparatus complies with Canadian ICES-003.
Cet appareil numérique de la classe A est conforme à la norme NMB-003 du Canada. (ICES-003 Class A)
• This product complies with Australian EMC. (AS/NZS C1SPR11 Group1 Class B)
3
AZ-LED
LED Ring Illuminator
Instructions
Caution
Although Nikon products are designed to provide the utmost safety
during use, incorrect usage or failure to follow the safety instructions
provided may cause personal injury or property damage. Damage or
injury that may occur due to mishandling is unwarranted. To ensure
correct usage, read the instruction manual carefully and thoroughly
before using the product. Do not discard the manual; keep it handy for
easy reference.
Safety instructions within this manual are described with the symbols
on the right to highlight their importance. For your safety, always follow
the instructions accompanied by these symbols.
Symbol
Meaning
WARNING
Disregarding instructions
accompanying this symbol may
lead to serious injury or death.
CAUTION
Disregarding instructions
accompanying this symbol may
lead to injury or property damage.
WARNING
Use of controls or adjustments or performance of procedures other than those specified in this manual may result in
hazardous radiation exposure.
1.
Intended application of the product: This product must be used only for microscopic observation. Do not use this
product for other purpose.
2.
Do not disassemble: Disassembling this product may result in electric shocks or other hazards. Malfunctions and
damage due to disassembling are unwarranted. Never attempt to disassemble any part other than the parts described in
the instructions. If you experience problems with this product or the microscope system, contact your nearest Nikon
representative.
3.
Read the instructions carefully: To ensure safety, carefully read this manual and the manuals provided with any other
equipment used with this product. In particular, observe all warnings and cautions of each manual.
4.
Cautions about the power cord: Make sure to use the specified power cord for the AZ-LED. Using a wrong power cord
may result in malfunctions or fire. For specifications of the power cord, refer to “Specifications.” To prevent an electric
shock, turn off the power switch of the product whenever connecting or disconnecting the power cord.
CAUTION
1.
Handle with care: The product and the microscope system are precision optical instruments. Handle the product
with care to avoid shock or impact. In particular, objectives may loose accuracy when exposed to even a weak
physical shock.
2.
Do not wet the product or allow foreign matter to get inside: Never allow water to come into contact with the
product because that may cause an electric shock or malfunction. Similarly, if foreign matter gets inside, an electric
shock or malfunction may occur. If water or foreign matter enters the product, do not use the product but contact your
nearest Nikon representative.
3. Turn off the power before assembling the product or connecting or disconnecting the cable: To prevent an
electric shock and fire, turn off the power switch and disconnect the power cord from the wall outlet before assembling
the product.
4.
Installation location: Being a precision optical instrument, the microscope may get damaged or loose accuracy if it
is used or stored under unsuitable conditions. When selecting the installation location, note the following:
• Avoid a brightly lit location, such as exposed to direct sunlight or directly under a room light. The image
quality deteriorates if there is excessive ambient light.
• Always install the microscope with a surrounding clear area of 10 cm or more.
• Choose a location that is free from dust or dirt.
• Choose a flat surface with little vibration.
• Choose a sturdy desk or table that is able to bear the weight of the instrument.
• Do not install the microscope in a hot or humid location.
• Select a layout that allows easy removal of the power cord from the microscope’s AC inlet in the event of an
emergency.
• For details about the operating environment and storage environment, see “Specifications.”
5.
Weak electromagnetic waves: The product emits weak electromagnetic waves. Do not install the product near