RMO-TW WINDING OHMMETERS Manual 1.800.561.8187 www. .com information@itm.com RMO-TW Contents Contents 1. Introduction ............................................................................................................................................................. 3 1.1 Safety Instructions ............................................................................................................................................ 3 1.1.1 Safety Terms and Symbols ....................................................................................................................... 3 1.1.2 Terms of Use ............................................................................................................................................ 3 1.1.3 Orderly Practices and Procedures ............................................................................................................ 4 1.1.4 Device Maintenance ................................................................................................................................. 4 1.1.5 Operator Qualifications ............................................................................................................................. 4 1.1.6 Safe Operating Procedures ...................................................................................................................... 4 1.2 Power Supply .................................................................................................................................................... 5 1.3 Measurement Category .................................................................................................................................... 5 1.4 Intended Use ..................................................................................................................................................... 6 2 Description ............................................................................................................................................................... 7 2.1 Front and Rear Panel Components .................................................................................................................. 7 3 Testing.................................................................................................................................................................... 10 3.1 Connecting the RMO-TW to a Test Object ..................................................................................................... 10 3.2 The Current Test ............................................................................................................................................. 11 3.2.1 Setting the Measurement Parameters .................................................................................................... 11 3.2.2 The Measurement Process ..................................................................................................................... 11 3.2.3 Viewing the Test Results ........................................................................................................................ 13 3.3 The Tap Changer Test .................................................................................................................................... 13 3.3.1 Setting the Measurement Parameters .................................................................................................... 14 3.3.2 Tap Changer Test with the RMO-TW ..................................................................................................... 15 3.4 Measurement Parameters .............................................................................................................................. 17 4 Error Messages ...................................................................................................................................................... 18 4.1 Error Message "Open Connection" ................................................................................................................. 18 4.2 Error Messages "Connection CH1-2-3" and "Connection CH1" ..................................................................... 18 4.3 Error Message "Polarity" ................................................................................................................................. 19 4.4 Error Messages "Change Current" and “Overcurrent” .................................................................................... 19 4.5 Error Message "Overheat" .............................................................................................................................. 19 4.6 Error Message "Mains Voltage < 90V" ........................................................................................................... 20 4.7 Error Messages "Error Printer" and "Check Paper" ........................................................................................ 20 5 Troubleshooting ..................................................................................................................................................... 21 5.1 Accuracy Check .............................................................................................................................................. 21 5.2 Open Circuit Voltage Check ........................................................................................................................... 21 5.3 Test Currents Check ....................................................................................................................................... 22 5.4 Voltage Measurement Check ......................................................................................................................... 23 5.5 TestCom Diagnostic Software ........................................................................................................................ 23 6 Customer Service ................................................................................................................................................... 24 7. Packing the Instrument for Shipment ..................................................................................................................... 24 8 Technical Data ....................................................................................................................................................... 25 8.1 Mains Power Supply ....................................................................................................................................... 25 8.2 Output Data ..................................................................................................................................................... 25 8.3 Measurement .................................................................................................................................................. 25 8.4 Environmental Conditions ............................................................................................................................... 25 8.5 Dimensions and Weight .................................................................................................................................. 26 8.6 Applicable Standards ...................................................................................................................................... 26 9 Instrument & Accessories ...................................................................................................................................... 27 Manufacturer Contact Information .................................................................................................................................. 28 Manual Version: M-RXX0TW-201-U-EN This Manual refers to the firmware version 11.00 Refers to: RMO10TW, RMO20TW, RMO30TW and RMO50TW models M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 2 information@itm.com RMO-TW Introduction 1. Introduction The purpose of this Manual is to provide helpful instructions on how to use the RMO-TW instruments safely, properly and efficiently. The following instructions will help the user avoid unsafe situations, reduce maintenance costs and will ensure the reliability and durability of the RMO-TW instruments. The RMO-TW must be used in observance of all existing safety requirements and regulations based on national/local standards for accident prevention and environmental protection. In addition, the relevant international standards are listed in paragraph 8.6 of the “Technical data” section of this document. 1.1 Safety Instructions Safety is the responsibility of the user. Before operating the RMO-TW, please read the following safety instructions carefully. It is not recommended that the RMO-TW is used (or even turned on) without careful observation of the instructions listed in this Manual. The RMO-TW should only be operated by trained and authorized personnel. 1.1.1 Safety Terms and Symbols Terms in this Manual. The following terms may appear in this manual: WARNING: Warning statements identify conditions or practices that could result in injury or loss of life. CAUTION: Caution statements identify conditions or practices that could result in damage to this product or to other property. Terms on the Device. The following warning terms used in this document may appear on the device: WARNING: indicates that potential hazard may occur. CAUTION: indicates that potential damage may occur to the instrument or to the test object connected to the instrument. Symbols on the Product. The following symbols may appear on the device: Refer to Manual Protective Earth Terminal 1.1.2 Terms of Use • • • • The RMO-TW shall be used only if it is in good technical condition. Its use shall be in accordance with local safety and industrial regulations. Adequate precautions must be taken to avoid any risks related to high voltages associated with this equipment and nearby objects. The RMO-TW shall be used only for the application purposes specified in the ”Intended Use” section of this Manual. The manufacturer and distributors are not liable for any damage resulting from wrong usage. The user bears responsibility in case of not following the instructions defined in this document. Do not remove the protective casing of the RMO-TW. All service and maintenance work must be performed by qualified personnel only. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 3 information@itm.com RMO-TW Introduction 1.1.3 Orderly Practices and Procedures • • • • • The Manual shall always be available on the site where the RMO-TW is used. Before using the RMO-TW, all personnel (even personnel who only occasionally, or less frequently, work with the RMO-TW) assigned to operate the RMO-TW should read this Manual. Do not make any modifications, extensions, or adaptations to the RMO-TW. Use the RMO-TW only with the original accessories provided by its manufacturer. Use the RMO-TW and its original accessories for the device's intended use only. 1.1.4 Device Maintenance This device should be kept clean in order to prevent excessive cases of dust or other contaminants affecting its operation. It should be cleaned with water/isopropyl alcohol after any dirt/contaminants are noticed on its surfaces. 1.1.5 Operator Qualifications • • Testing with the RMO-TW should only be carried out by authorized and qualified personnel. Personnel receiving training, instruction or education on the RMO-TW should remain under the constant supervision of an experienced operator while working with the test set and the test object. 1.1.6 Safe Operating Procedures • • • • • • • • • • • • Hazardous voltages of up to 400 V can occur inside the RMO-TW. Therefore, it is not permitted to remove the protective casing of the RMO-TW. Hazardous voltages exist on the terminals of the RMO-TW when the red LED is lit. Never assume that the connections are safe even if this LED is off. Switch off and unplug the RMO-TW before touching the connections, particularly if a fault is suspected. Before putting the RMO-TW into operation, check the test set for any visible damage. Cables between the RMO-TW and other equipment shall be connected and disconnected from the RMOTW only while the instrument is switched off. Do not operate the RMO-TW under wet or moist conditions (condensation). Do not operate the RMO-TW if explosive gas or vapors are present. Only the external devices which meet the requirements for SELV equipment according to EN 60950 or IEC 60950 should be connected to the RMO-TW through the serial interface. When setting up the RMO-TW, make sure that the air slots of the test set remain unobstructed Removing the RMO-TW protective casing by unauthorized persons will void the warranty. Any work inside the instrument without prior authorization by DV Power will also void the warranty. If the RMO-TW seems to be malfunctioning, please contact the DV Power Support Team (refer to the "Manufacturer Contact Information” section) after previously checking the “Error Messages” section. Prior to connecting the RMO-TW, ensure that the transformer (object) to be tested is completely deenergized and isolated from both the line and the load. Every terminal should be checked and verified before connecting the RMO-TW. Ground connections may be left in place. Do not use the RMO-TW without the extra protective ground cables supplied with the RMO-TW. It must never be operated in a non-grounded configuration as this may result in an electric shock to the user or M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 4 information@itm.com RMO-TW Introduction damage to the RMO-TW. Always establish this connection as the first step before establishing any other connections, and remove this connection last. Wherever possible, the outer casing of the transformer under test should also be connected to a safety earth to prevent the risk of shock. Where this cannot be achieved, adequate precautions should be taken to prevent access to the transformer (e.g. barriers). • Never open the measuring circuit while the test is in progress as dangerous high voltage may occur. Never change the De-Energized Tap Changer (DETC) position while the test is in progress. This can cause excessive voltage arcs within the transformer and on external terminals. 1.2 • Power Supply Supply the RMO-TW only from a power outlet which is equipped with protective ground. • Besides supplying the RMO-TW from phase-neutral (L1-N, A-N), it may also be supplied from phase to phase (e.g., L1-L2; A-B). However, the voltage must not exceed 264 V AC. Please refer to the “Technical Data” section. The RMO-TW should be positioned in such a way that it is possible to safely disconnect it from the power supply at any moment. WARNING / AVERTISSEMENT This is a class A product. In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures. Il s'agit d'un produit de classe A. Dans un environnement domestique, ce produit peut provoquer des interferences radio, auquel cas l'utilisateur peut être amené à prendre des measures adéquates. 1.3 Measurement Category The RMO-TW is intended to be used for measurements in Measurement Category I (CAT I) for voltages up to 300V. The device is also designed to withstand occasional transient overvoltage of up to 1000Vpk. WARNING / AVERTISSEMENT This equipment is classified as measurement category I and must not be used within measurement category II, III and IV. Cet équipement est classé dans la I catégorie de mesure, et ne doit pas être utilisé pendant les catégories de mesure II, III et IV. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 5 information@itm.com RMO-TW 1.4 Introduction Intended Use The Winding Ohmmeter RMO-TW is a stand-alone or PC-controlled digital instrument designed for detailed analysis of on-load tap changers and for measuring resistances of inductive test objects used in the electric power industry or similar industries. It is employed for resistance measurement during the manufacturing, commissioning and testing of: transformers, on-load tap changers, de-energized tap changers, generators and electrical motors, high current busbar joints, cable splices. The RMO-TW is also used to create the automatic resistance/temperature measurement for the Heat Run test together with the DV-Win software. Any uses of the RMO-TW other than the ones mentioned above are considered improper and will invalidate warranty claims. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 6 information@itm.com RMO-TW Description 2 Description 2.1 Front and Rear Panel Components Figure 2-1: Front panel of the instrument 1. Operator Control Display Displays the settings during the device test mode selection as well as the measured values during and after a test operation. Keyboard Used to control the device. • UP/DOWN buttons are used to navigate and set the test parameters. • ENTER confirms the defined test parameters, language, time and date, and also navigates through menus. During the test it is used to switch to the measurement of the next phase. • STOP stops the test, interrupts the alarm buzzer and to return to the previous menu. • Ω records the resistance value currently displayed on the screen in the RMO-TW instrument internal memory • START starts the test. • SET scrolls between the Current, TapC and Memory menus. Green LED • Lights continuously when the RMO-TW is turned on. • Flashes when the test can be started. • Flashes alternately with the red LED during the Current and TapC test. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 7 information@itm.com RMO-TW Description Red LED • • • • Lights continuously in case of operational error. Lights continuously during a discharging process. Flashes alternately with the green LED during the Current and TapC test. Flashes shortly during a resistance measurement in the TapC test. 2. USB PC Interface The USB PC interface enables controlling the instrument from a PC and downloading previously recorded results from the instrument’s internal memory. A USB cable is provided as an accessory. 3. Discharge LED The red LED remains lit until the accumulated energy of the measured object is reduced to a harmless level. CAUTION While the discharge LED is lit, the measurement cables must not be disconnected. CAUTION While disconnecting cables, do not bridge the circuit with your body parts! In case of shorting a winding for the DRM test, it should be noted that a very small current may remain flowing through the shorted winding even when the discharging process has finished. For that reason a person should use only a single hand when disconnecting the shorting cable, to avoid the other hand creating a return path if touching a transformer. For maximum safety insulating gloves should also be used. 4. Voltage sense cables terminals Terminals for connecting up to two pairs of voltage sense cables. 5. Current cables terminals Terminals for connecting “+ / red” and “- / black” test current cables. Figure 2-2: Rear panel of the instrument 6. Mains Power Connector Connects the RMO-TW to the mains power supply with a power cord. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 8 information@itm.com RMO-TW Description 7. Power Switch Power switch – Double pole switch I - In this position, the RMO-TW is connected to the mains power supply. 0 - In this position, the RMO-TW is separated from the mains power supply. 8. Protective Earth Connector For protection against parasitic currents or voltages, the RMO-TW earth/ground connector should always be connected to protective ground (PE). Only the originally provided cable should be used for this purpose. For safety reasons, this connection should be established as the first one before any other connections and removed last. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 9 information@itm.com RMO-TW Testing 3 Testing 3.1 Connecting the RMO-TW to a Test Object Before the RMO-TW has been connected to a test object (e.g. a power transformer), the following steps have to be taken: • The test object needs to be disconnected from its circuit in accordance to the national safety regulations and properly grounded to protective earth. • The transformer needs to be completely de-energized. • The RMO-TW itself needs to be properly grounded. To do so, the grounding screw should be connected to protective earth using the original grounding cable. • One end of the measured windings needs to be properly grounded. The only exception is in the case of the YN connection being tested and neutral and earth bonds are mutually connected. In this case, one side of the measurement circuit should not be earthed because it is not necessary and because it will incorrectly affect the RMO-TW reading. Note: Always connect measuring cables to the RMO-TW instrument first and then to the test object. When disconnecting, always disconnect the cables from the test object first and then from the RMO-TW. The grounding wire PE should be disconnected last. Not following these instructions may cause life-threatening situations. Figure 3-1: Connecting a test object to the RMO-TW With the RMO-TW turned off, the RMO-TW needs to be connected to the test object in such a way that the measuring cables from the "Voltage Sense" sockets are attached as close as possible to the winding and in between the current feeding cables. That way, the resistance of cables and clamps is almost completely excluded from the resistance measurement. When connecting the measuring cables, attention should be paid to polarity. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 10 information@itm.com RMO-TW Testing To maximize the accuracy and measurement repeatability, all clamps need to have a good connection to the test object. CAUTION For safety reasons, a connection of the device should be always established with the test object in the way described in this Manual. The measuring cables should not be connected after the start of the test. The leads should not be disconnected before the discharge LED and the alarm buzzer are off. The RMO-TW instrument can perform the following tests: - Current test - Tap Changer test. 3.2 The Current Test 3.2.1 Setting the Measurement Parameters After the initial display, the RMO-TW automatically changes to the Current menu and the green LED lights up. Test current Figure 3-2: The Current menu Menu name Test duration time Memory Time and Date 3.2.2 The Measurement Process To start the measuring sequence, the following values need to be selected: Test current. The following values can be chosen: 5 mA, 10 mA, 50 mA, 100 mA, 500 mA, 1 A, 2 A, 5 A, 10 A, 15 A(*), 20 A(*), 25 A(**), 30 A(**), 40 A(***) and 50 A(***). * available with the RMO20TW, RMO30TW and RMO50TW models ** available with the RMO30TW and RMO50TW models *** available with the RMO50TW model Memory position. This parameter sets the memory position in which the first measurement results will be saved. Once the test is completed (either due to time expiration or when pressing STOP), the last measured result will remain registered in the RMO-TW and the memory position is automatically switched over to the next memory position. The result can be recalled later by selecting that position number. Positions 000 - 499 can be selected. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 11 information@itm.com RMO-TW Testing Test duration time. The following values can be selected: 30 s, 60 s, 90 s, 2 min, 3 min, 4 min, 5 min, 6 min, 7 min, 8 min, 9 min, 10 min, 11 min, 12 min, 13 min, 14 min, 15 min, 30 min, 60 min, 90 min, unlimited. The available values are selected by pressing UP/DOWN. The next parameter is selected by pressing RIGHT. Once the information has been confirmed by pressing ENTER, the Ready screen will be displayed. The Ready state displays the selected test current and its defined duration. If one of these values needs to be changed, press STOP to return to the Current menu. Figure 3-3: The Ready screen The flashing green LED indicates that the RMO-TW is now ready to start the test. Pressing START will run the test. At the start of the test, a cable connection check is done. If an open circuit is detected, the alarm will be activated and the appropriate error message will be shown on the display. During the test, both the green and the red LEDs flash alternately. When the current is stabilized, the RMO-TW starts the measurement of the test object resistance. Figure 3-4: The Charging screen The following values are displayed during the test: test current, currently measured resistance values, time remaining until the end of the test procedure and the active memory position. Two resistance values can be measured simultaneously. During the test, it is necessary to wait for the test current stabilization in order to obtain reliable results. The current stabilization can be tracked by observing the Delta R value on the right side of the screen, or by observing the resistance values themselves. The Delta R information provided is the deviation of the resistance Δ which can be used to track the current stabilization process. The closer the Δ is to 0%, the results are more stable. These values are automatically refreshed every second. Note: The Current menu is not intended for on-load tap changer testing and the tap changer should not be operated while testing in this mode. The TapC menu should always be used for tap changer analysis instead. Figure 3-5: The Current menu during the test Pressing Ω will save the measured resistance results in the internal memory of the RMO-TW, and the memory position is automatically switched over to the next memory position. These results can be recalled later by selecting that position number. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 12 information@itm.com RMO-TW Testing The test can be completed in two ways: - By pressing STOP, - Automatically, after the selected time has expired. When the test is finished either because the selected time has expired or the STOP button has been pressed, the RMO-TW switches to the Results menu and starts the discharging process. During the discharging, the discharge LED lights and the alarm beeps. Discharge is completed when the discharge LED and the buzzer are OFF. Also, during that period the “Discharging” message is shown on the display. During the discharging process it is not possible to start another measurement. When the test is stopped, the RMO-TW immediately stops producing output voltage. However, because of the test object inductivity, the current still flows. Because of the enormous amount of energy that can be stored in a magnetic field, precautions should be taken before disconnecting the test leads from the tested transformer. Never remove the leads during the testing process and always allow enough time for the tested transformer to completely discharge. Large transformers can require several minutes to discharge. CAUTION DO NOT DISCONNECT LEADS BEFORE THE “DISCHARGING” MESSAGE DISAPPEARS FROM THE DISPLAY AND THE DISCHARGE LED IS OFF. ATTENTION NE PAS DÉBRANCHER LES CONDUCTEURS AVANT QUE LE MESSAGE ''DÉCHARGER/DISCHARGING'' EST DISPARU DE L'ÉCRAN ET QUE LA DIODE DE DÉCHARGE EST ÉTEINTE. 3.2.3 Viewing the Test Results Once the test is finished, the RMO-TW automatically changes to the Results menu to display the test results. Figure 3-6: The Results menu showing the test results 3.3 The Tap Changer Test The winding resistance of individual taps of a power transformer’s tap changer can be measured by using the Tap Changer (TapC) mode of the RMO-TW. There is no need to charge and discharge the transformer between tap position changes. The inputs of the instrument are protected during the switching process in order to avoid damage to the instrument in case of a faulty tap changer. The RMO-TW injects a constant current into the power transformer. The winding resistance is calculated from this current value and the measured voltage value. An interruption caused by a defective tap changer results in comparatively high measured values of the current ripple. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 13 information@itm.com RMO-TW Testing Figure 3-7: Connecting a test object to the RMO-TW 3.3.1 Setting the Measurement Parameters Turn on the RMO-TW with the power switch on the back of the test set. The display shows the Current menu and the green LED lights up. Press SET to go to the TapC menu. Test current Figure 3-8: The TapC menu Menu name Memory position Active channel Time and Date In the TapC menu, the following values need to be defined: • Test current. The following values can be selected: 1 A, 2 A, 5 A, 10 A, 15 A(*), 20 A(*). * available with the RMO20TW, RMO30TW and RMO50TW models • Memory position. This parameter sets the memory position in which the first measurement results will be saved. Once the test is completed, the last measured result will remain saved in the RMO-TW, and the memory position is automatically switched over to the next memory position. The result can be recalled later by selecting that position number. Positions 000 - 499 can be selected. The available values are selected by pressing UP/DOWN. The next parameter is selected by pressing RIGHT. Once the information has been confirmed by pressing ENTER, the Ready screen will be displayed. Figure 3-9: The Ready screen The Ready state displays the selected test current and memory location. If one of these values has to be changed, press STOP to return to the TapC menu. Note: Only CH1 (the upper channel) is active during the TapC measurement. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 14 information@itm.com RMO-TW Testing 3.3.2 Tap Changer Test with the RMO-TW The flashing green LED indicates that the RMO-TW is now ready to start the test. Press START to run a test. At the start of the test, a cable connection check is done. If an open circuit is detected, the alarm is activated and the error message is shown on the display During the period of current stabilization, both the green and the red LEDs flash alternately and the message “PLEASE WAIT CHARGING” is shown on the display. The following values are displayed during this part of the test: the test current, the effective value of the measured resistance, and the active memory position. The current stabilization can be tracked by observing the Delta R value on the right side of the screen, or by observing the resistance values themselves. These values are automatically refreshed every second. Figure 3-10: The TapC menu when the test current is stabilized At the moment the resistance result is stable, the resistance value of the on-load tap changer starting position can be saved by pressing Ω. The memory position is automatically switched over to the next one and the device is ready to record the current ripple. While the resistance value is being saved, the red LED flashes shortly and the beeper sounds in higher speed intervals. Figure 3-11: The TapC menu, “Change tap position” message The RMO-TW detects the change of the test current during the tap position change and measures the difference between the starting current and the lowest value it reaches during the transition. This value is called the transition ripple, and it is indicated as a percentage of the starting current value. Figure 3-12: The TapC menu, memory location is increased by one The display shows the test current value in the new tap position (here 9.82 A), and the value of the current ripple which occurred during the change from the previous position to the new one (here 4.7 %). The current stabilization can be followed on the display. After it has been stabilized, the RMO-TW shows the resistance M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 15 information@itm.com RMO-TW Testing value in the new tap position (here 576.4 mΩ). Pressing Ω will save the resistance and the ripple in the new tap changer position. This procedure is repeated until the ripples and resistance values are obtained for all the tap changer positions. The tap changer test is completed by pressing STOP. The transformer is then automatically discharged. During the current discharging process, the discharge LED lights and the alarms beep. Discharge is completed when the discharge LED and the buzzer are OFF. It is not possible to start another measurement during this process. CAUTION DO NOT DISCONNECT LEADS BEFORE THE “DISCHARGING” MESSAGE DISAPPEARS FROM THE DISPLAY AND THE DISCHARGE LED AND BUZZER ARE OFF. Figure 3-13: The TapC menu during discharging A properly working tap changer differs from a malfunctioning one by the magnitude of the current ripple value. An open circuit during the transition will result in a much higher current ripple value (99 % or 100 %) than a properly functioning tap changer. Note: The current ripple value depends on the construction of the tap changer. It is recommended to compare current ripple values to the values of other tap transitions in the same measurement. It is also recommended to compare the current ripple results with the reference results from a previous tap changer test on the same transformer. The figure below shows an indication of possible tap changer discontinuity during the tap position change from one tap position to the next one. Figure 3-14: The TapC menu showing a much higher current ripple value If the current circuit is interrupted during the tap position change, the „Discontinuity“ message will be shown on the display, the red LED will light and the buzzer will beep continuously. The discharging process will be initiated. Figure 3-15: The "Discontinuity" message M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 16 information@itm.com RMO-TW 3.4 Testing Measurement Parameters The highest resolution of the reading is always displayed. The resistance range selection is performed automatically and therefore manual setting by the operator is not necessary. Table 3.1: Maximum resistances for each test current Test Current Max Test Voltage RMO10TW Max Test Voltage RMO20TW Max Test Voltage RMO30TW Max Test Voltage RMO50TW Max Resistance RMO10TW Max Resistance RMO20TW Max Resistance RMO30TW Max Resistance RMO50TW 5 mA 10 mA 50 mA 100 mA 500 mA 1A 2A 5A 10 A 15 A 20 A 25 A 30 A 40 A 50 A 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V - 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 37,5 V - 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 42.5 V 34 V 28.33 V - 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 50 V 40 V 33.3 V 25 V 20 V 10 kΩ 5 kΩ 1 kΩ 500 Ω 100 Ω 50 Ω 25 Ω 10 Ω 5Ω - 10 kΩ 5 kΩ 1 kΩ 500 Ω 100 Ω 50 Ω 25 Ω 10 Ω 5Ω 3.33 Ω 1.87 Ω - 10 kΩ 5 kΩ 1 kΩ 500 Ω 100 Ω 50 Ω 25 Ω 10 Ω 5Ω 3.33 Ω 2.125 Ω 1.36 Ω 944 mΩ - 10 kΩ 5 kΩ 1 kΩ 500 Ω 100 Ω 50 Ω 25 Ω 10 Ω 5Ω 3.33 Ω 2.5 Ω 1.6 Ω 1.11 Ω 625 mΩ 400 mΩ Table 3.2: Resolution and full scale for each measuring range Full scale Resolution 0.1 μΩ – 999.9 μΩ 0.1 μΩ 1.000 mΩ – 9.999 mΩ 1 μΩ 10.00 mΩ – 99.99 mΩ 10 μΩ 100.0 mΩ – 999.9 mΩ 0.1 mΩ 1.000 Ω – 9.999 Ω 1 mΩ 10,00 Ω – 99.99 Ω 10 mΩ 100.0 Ω – 999.9 Ω 0.1 Ω 1.000 kΩ – 9.999 kΩ 1Ω Typical accuracy of the RMO-TW instruments is ±(0.1% rdg + 0.1% F.S) for 0.1 μΩ – 1.999 kΩ range, and ±(0.2% rdg + 0.1% F.S) for 2.000 kΩ – 9.999 kΩ range. These accuracies are valid if the maximum possible current is used. As high current as possible should always be selected, but to avoid the risk of heating the winding, it is recommended not to exceed from 10 % to 15 % of the transformer nominal current. It is important to make a good connection between the clamps and the connecting points on the test object to achieve stable and accurate test results. The guaranteed accuracy of the RMO-TW instruments is ± (0.2 % rdg + 0.2 % F.S). M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 17 information@itm.com RMO-TW Error Messages 4 Error Messages Any operational error is indicated by a red LED and accompanied by an alarm. Furthermore, the display indicates an error status message. To stop the alarm buzzer and return to the test menu, the status message should be removed from the display by pressing STOP. 4.1 Error Message "Open Connection" The Error message "Open Connection" is displayed if the current path is interrupted. Figure 4-1: An example of interruption of the current path – a current cable disconnection Figure 4-2: Error message "Open Connection" 4.2 Error Messages "Connection CH1-2-3" and "Connection CH1" The Error message "Connection CH1-2-3" is displayed if channel 1and channel 2 are all disconnected. This message is displayed if either + or – voltage sense cable or both are disconnected either on the device side or on the side of the test object. Figure 4-3: Error message "Connection" The Error message "Connection CH1" is displayed only if CH1 is not connected during the tap changer testing. This message is displayed if either + or – voltage sense cable or both are disconnected either on the device side or the side of the test object. These messages can be displayed only at the beginning of the test. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 18 information@itm.com RMO-TW 4.3 Error Messages Error Message "Polarity" The Error message "Polarity" is displayed if the voltage sense cables are connected with opposite polarity either on the side of a device or the side of a test object, and the measured voltage signal is higher than 60 mV. Unlike other error messages, the test will not be interrupted. A “Polarity” warning will be displayed as shown on the figure below. Figure 4-4: Error message "Polarity" Note: Since 60 mV is the signal threshold, the display will show the result with lower accuracy for lower voltage signal values. If some of the cables are disconnected after starting the test, the error message will not show up and the received results will be erroneous. 4.4 Error Messages "Change Current" and “Overcurrent” This message is displayed if the voltage at the test object is too high during the test. In this case the test current should be reduced and the test repeated. Figure 4-5: Error message "Change Current" The error message “Overcurrent” is displayed if the test voltage is lower than 10 V per channel but the test current is still too high for that resistance. An user should re-start the test with lower test current selected. Figure 4-6: Error message "Overcurrent" 4.5 Error Message "Overheat" This message is displayed when the operating temperature of the RMO-TW rises too high. In this case, it is necessary to wait for the test set to cool down and repeat the test. Figure 4-7: Error message "Overheat" M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 19 information@itm.com RMO-TW 4.6 Error Messages Error Message "Mains Voltage < 90V" This message is displayed when the RMO-TW mains voltage is below 90 V. Figure 4-8: Error message "Mains Voltage < 90V" The RMO-TW should be restarted to reset the status message. 4.7 Error Messages "Error Printer" and "Check Paper" The message "Error printer" is displayed if a hardware error of the printer is detected. The position of the black shaft should be checked. If it is in proper position and the error keeps occurring repeatedly, please contact DV Power Support Team (please refer to the ”Manufacturer Contact Information” section of this Manual). Figure 4-9: Error messages "Error printer" and "Check paper" The message "Check paper" is displayed if the printer is out of paper. The paper roll should be aligned with the paper intake so that there is no horizontal offset, and the central axis of the paper roll must be parallel to the printer as shown in the Figure 4-10. Also, attention should be paid when inserting the paper into the printer. It should enter the paper intake from the lower side of a paper roll. If the paper is not properly aligned, the black shaft should be pulled up to loosen the paper (see the Figure 410). After that, the paper should be aligned and the black shaft pushed down into a proper position. Figure 4-10: Paper loading M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 20 information@itm.com RMO-TW Troubleshooting 5 Troubleshooting If suspected the device is presenting inaccurate results, the following tests should be performed: 5.1 Accuracy Check 1. Connect the test shunt R1 (1 Ω) and the Voltage Sense cables (VS1 input) as shown in the Figure 5-1. 2. Set the first test current from the Table 5.1 and start the test. Read the resistance value measured with the Winding Ohmmeter and compare it to the expected value written in the Table 5.1. Repeat this step for all test currents from the Table 5.1. 3. Connect the test shunt R2 (1 mΩ) and Voltage Sense cables (VS1 input) as shown in the Figure 5-2. 4. Set the first test current listed in Table 5.2 and start the test. Read the resistance value measured with the Winding Ohmmeter and compare it with the expected value written in the Table 5.1. Repeat this step for all test currents from the Table 5.2. 5. Connect the Voltage Sense cables to the VS2 input and repeat steps 1-4. 6. If measured results differ from expected values, please proceed to the next step – determining the root cause of the inaccurate measurements. Figure 5-1 – Connection diagram I [mA] 10 100 1000 Figure 5-2 – Connection diagram R1 [Ω] 0.989 – 1.011 0.989 – 1.011 0.989 – 1.011 I [A] 2 5 10 Table 5.1 – Expected results 5.2 R2 [mΩ] 0.989 – 1.011 0.989 – 1.011 0.989 – 1.011 Table 5.2 – Expected results Open Circuit Voltage Check 1. Short circuit the Voltage Sense input (VS1) and connect a DC voltmeter directly between current terminals of the Winding Ohmmeter, as shown in the Figure 5.3. 2. Set the test current of 10 A and start the test. Check if the voltage measured with the voltmeter is in range 55 – 65 VDC. 3. Check if any resistance measurement value is shown on the device display and if the discharge LED is ON. 4. After 120 seconds the device should display the Open connection error. 5. If the measured voltage is not in the range 55 – 65 VDC, please contact the DV Power Support Team. Otherwise, please proceed to the next step. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 21 information@itm.com RMO-TW Troubleshooting Note: Discharge LED should not light during this test. Expected voltage: 55 VDC – 65 VDC Figure 5.3 – Open circuit voltage check 5.3 Test Currents Check 1. Short circuit Voltage Sense input (VS1) of the device and connect a digital voltmeter in parallel to the R1 (1 Ω) test shunt as shown in the Figure 5.4. 2. Set the first test current listed in the Table 5.3 and start the test. Note the voltage V1 obtained with the voltmeter and compare it with expected one: 𝑉1 = 𝐼 ∙ 𝑅1 . Repeat this step for all test current listed in the Table 5.3. 3. Short circuit Voltage Sense input (VS1) of the device and connect a digital voltmeter parallel to the R2 (1 mΩ) test shunt as shown in the Figure 5.5. 4. Set the first test current listed in the table 5.4 and start the test. Note the voltage V2 obtained with the voltmeter and compare it to the expected one: 𝑉2 = 𝐼 ∙ 𝑅2 . Repeat this step for all test current listed in Table 5.4. 5. If the measured voltages do not differ significantly from the expected ones, the device is generating the current correctly. Please proceed to the next step. Figure 5.4 – Connection diagram I [mA] 10 100 1000 Figure 5.5 – Connection diagram V1[mV] 9.89 – 10.11 98.9 – 101.1 989 – 1011 I [A] 2 5 10 Table 5.3 – Expected results M-RXX0TW-201-U-EN 1.800.561.8187 www. R2[mV] 1.978 – 2.022 4.945 – 5.055 9.89 – 10.01 Table 5.4 – Expected results .com 22 information@itm.com RMO-TW 5.4 Troubleshooting Voltage Measurement Check 1. Connect the test shunt R1 (1 Ω), Voltage Sense cables (VS1 input) and voltmeter as shown in the Figure 5.6. 2. Set the first test current listed in Table 5.5 and start the test. Note the resistance measured with the device (R1meas), the voltage V1 measured with the voltmeter, and calculate the resistance: 𝑅1𝑐𝑎𝑙 = 𝑉1 . 𝐼 Repeat this step for all test currents listed in Table 4.1. 3. Connect the test shunt R2 (1 mΩ), Voltage Sense cables (VS1 input) and voltmeter as shown in the Figure 5.7. 4. Set the first test current listed in Table 5.6 and start the test. Note the resistance measured with the device (R2meas), the voltage V2 measured with the voltmeter, and calculate the resistance: 𝑅2𝑐𝑎𝑙 = 𝑉2 . 𝐼 Repeat this step for all test currents listed in Table 5.6. 5. Connect the Voltage Sense cables to the VS2 input and repeat steps 1-4. 6. If calculated resistances do not differ significantly from the resistances measured with the RMO-TW device, the device is measuring correctly. Figure 5.6 – Connection diagram I [mA] 10 100 1000 R1meas[Ω] 0.989 – 1.011 0.989 – 1.011 0.989 – 1.011 Figure 5.7 – Connection diagram V1[mV] 9.89 – 10.11 98.9 – 101.1 989 – 1011 I [A] 2 5 10 Table 5.5 – Expected results 5.5 R2meas[mΩ] 0.989 – 1.011 0.989 – 1.011 0.989 – 1.011 V2[mV] 1.978 – 2.022 4.945 – 5.055 9.89 – 10.01 Table 5.6 – Expected results TestCom Diagnostic Software In case the TestCom diagnostic software is not provided, please contact the DV Power Support Team to request the latest version. 1. Connect the device to the computer and turn it ON. 2. Run the TestCom.exe file. 3. Click Connect to establish communication between the device and the computer. 4. Click Start and a log file will be generated in the same folder where TestCom.exe is located. 5. Send the log file to DV Power Support Team for analysis. NOTE: If experiencing problems with running TestCom.exe, please install the Patch.exe, which is included in the TestCom archive. After the installation, it should be possible to run TestCom.exe. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 23 information@itm.com RMO-TW Customer Service, Packing the Instrument for Shipment 6 Customer Service Before calling or sending an e-mail to Customer Service for assistance, please perform the following steps: 1. Check all cable connections. 2. Try the test on another instrument, if available. 3. Perform the troubleshoot procedure 4. Have the following information available: • Instrument serial numbers, hardware configuration, and software revision • Exact description of the problem, including the test object information, error messages and the sequence of events before it appeared • List of solutions that have been tried The Customer Support Department can be reached at: Local support (Sweden): +46 8 731 78 24 International support: +46 70 0925 000 E-mail: support@dv-power.com Note: Email communication is preferred for support issues, since the case is then documented and traceable. Also, the time zone problems and issues with occupied telephones do not occur. 7. Packing the Instrument for Shipment If you need to send the instrument to DV Power for servicing, please contact the DV Power Customer Service for return instructions at: Local support (Sweden): +46 8 731 78 24 International support: +46 70 0925 000 E-mail: support@dv-power.com Note: DV Power is not responsible for shipping damage. Please protect each instrument from shipping and handling hazards carefully. Please ensure protective covers are securely in place. Instruments should be sent to DV Power freight pre-paid, unless other arrangements have been authorized in advance by DV Power Customer Service. To prepare an instrument for shipment, please follow these instructions: 1. Disconnect and remove all external cables. Do not include manuals and cables unless recommended by DV Power Customer Service. 2. Reuse the original packing material if it is available. If it is not, pack the instrument for shipment according to the instructions for fragile electronic equipment. It is recommended use two-wall minimum corrugated cardboard box with a minimum 5 cm (2 inch) thick poly foam padding, or a wooden crate with minimum of 5 cm (2 inch) thick poly foam padding all around. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 24 information@itm.com RMO-TW Technical Data, Instrument & Accessories, Manufacturer Contact Information 8 Technical Data 8.1 Mains Power Supply - Connection: - Voltage: - Input power according to EN60320-1/C320 90 V – 264 V AC, 50/60 Hz, Single phase 750 VA (RMO10TW model), 1500 VA (RMO20TW model), 2250 VA (RMO30TW and RMO50TW models) 15 A / 250 V, type F, but not user replaceable - Fuse 8.2 Output Data - Test current RMO10TW RMO20TW RMO30TW RMO50TW 5 mA – 10 A DC 5 mA – 20 A DC 5 mA – 30 A DC 5 mA – 50 A DC - Output power 8.3 Instrument RMO10TW RMO20TW RMO30TW RMO50TW Output power 600 W 750 W 850 W 1 000 W Measurement - Measuring range / Resolution 0.1 µΩ - 999.9 µΩ 1.000 mΩ - 9.999 mΩ 10.00 mΩ - 99.99 mΩ 100.0 mΩ - 999.9 mΩ 1.000 Ω - 9.999 Ω 10,00 Ω - 99.99 Ω 100.0 Ω - 999.9 Ω 1.000 kΩ - 9.999 kΩ 0.1 µΩ 1 µΩ 10 µΩ 0.1 mΩ 1 mΩ 10 mΩ 0.1 Ω 1Ω - Typical Accuracy: ± (0.1 % rdg + 0.1 % FS) for 0,1 μΩ – 1,999 kΩ range ± (0.1 % rdg + 0.2 % FS) for 2.000 kΩ – 9,999 kΩ range - Guaranteed Accuracy ± (0.2 % rdg + 0.2 % FS) 8.4 Environmental Conditions o o o o - Operating temperature -10 C - +55 C / 14 F - +131 F - Storage temperature - Humidity range -25 C - +70 C / -13 F - +158 F 5 - 95%, non-condensing M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 25 information@itm.com RMO-TW 8.5 Dimensions and Weight Dimensions without handle (cm) Dimensions without handle (in) Weight (kg) Weight (lbs) 8.6 Technical Data, Instrument & Accessories, Manufacturer Contact Information RMO10TW RMO20TW RMO30TW RMO50TW 19.8 x 25.5 x 38 19.8 x 25.5 x 38 19.8 x 25.5 x 38 19.8 x 25.5 x 38 7.8 x 10 x 15 7.8 x 10 x 15 7.8 x 10 x 15 7.8 x 10 x 15 7.5 16.5 7.5 16.5 8.5 18.7 8.5 18.7 Applicable Standards - Installation/overvoltage - Pollution - Safety category II degree 2 UL 61010-1 CAN/CSA-C22.2 No. 61010-1, 2nd edition, including Amendment 1 Directive 2004/108/EC (CE Conform) Standard EN 61326-1 - EMC All specifications herein are valid at the ambient temperature of + 23 °C ± 5°C and by using the recommended accessories. Specifications are subject to change without notice. M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 26 information@itm.com RMO-TW Technical Data, Instrument & Accessories, Manufacturer Contact Information 9 Instrument & Accessories Instrument with included accessories Article No Winding Ohmmeters RMO-TW RMOXX0TW-N00 - DV-Win PC software including USB cable - Mains power cable - Ground (PE) cable Recommended accessories Article No 2 Current cables 2 x 10 m 10 mm and Sense cables 2 x 10 m with TTA clamps * 2 Current cables 2 x 10 m 2.5 mm and Sense cables 2 x 10 m with TTA clamps ** Sense cables 2 x 10 m with TTA clamps 2 Current connection cable 5 m 6 mm with TTA clamps * 2 Current connection cable 5 m 10 mm with TTA clamps ** Device bag Cable bag CS-10-10LMWC CS-10-02BPWC S2-10-02BPWC CX-05-062XWC CX-05-102XWC DEVIC-BAG-00 CABLE-BAG-00 * recommended for the RMO10TW and RMO20TW models ** recommended for the RMO30TW and RMO50TW models Optional accessories Article No Test shunt 1 mΩ (150 A / 150 mV) 2 Current cables 2 x 15 m 10 mm and Sense cables 2 x 10 m with TTA clamps 2 Current cables 2 x 20 m 16 mm and Sense cables 2 x 10 m with TTA clamps 2 Current cables 2 x 10 m 10 mm with TTA clamps 2 Current cables 2 x 15 m 10 mm with TTA clamps 2 Current cables 2 x 20 m 16 mm with TTA clamps Sense cables 2 x 15 m with TTA clamps Sense cables 2 x 20 m with TTA clamps 2 Current connection cable 12 m 16 mm Thermal printer 80 mm (built-in) Cable plastic case Cable plastic case with wheels Transport case Bluetooth communication module SHUNT-150-MK CS-15-16LMWC CS-20-16LMWC C2-10-16LMWC C2-15-16LMWC C2-20-16LMWC S2-15-02BPWC S2-20-02BPWC CX-12-162XWC PRINT-080-00 CABLE-CAS-01 CABLE-CAS-W1 HARD-CASE-RT BLUETOOTH-00 M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 27 information@itm.com RMO-TW Technical Data, Instrument & Accessories, Manufacturer Contact Information IBEKO Power AB 2014 This Manual is a publication of IBEKO Power AB, 181 50 Lidingö, Sweden. These documents are protected by Swedish Copyright law and international contracts as intellectual property of the IBEKO Power AB. The documents contain confidential information of IBEKO Power AB which is protected by patent, copyright, trademarks or otherwise as inventions, trademarks or creations of IBEKO Power AB. The reproduction, duplication, transmission or use of these documents or its contents is not permitted without express prior written consent of the IBEKO Power AB. IBEKO Power AB shall not be liable for any incidental or consequential damages resulting from the performance or use of this document or its product. This document has undergone extensive technical approval before being released. IBEKO Power AB reviews this document at regular intervals, and includes appropriate amendments in subsequent issues. While every effort has been made to keep the information herein as accurate and up to date as possible, IBEKO Power AB assumes no responsibility for errors or omissions or for damages resulting from the use of the information contained herein. IBEKO Power AB cannot take over liability resulting in any way from the use of this document or parts thereof. The product information, pictures, drawings and all technical data contained within this manual are not contractually binding and IBEKO Power AB reserves the right to make modifications at any time to the technology and/or configuration without prior notice. Insofar as any information, software or documentation is made available, any liability for defects as to quality or title of the information, software and documentation especially in relation to the correctness or absence of defects or the absence of claims or third party rights or in relation to completeness and/or fitness for purpose are excluded except for cases involving willful misconduct or fraud. In case of a disagreement between the translation and the original English version of this Manual, the original English version will prevail. Manufacturer Contact Information IBEKO Power AB Box 1346, 181 25 Lidingö, Sweden Fax: + 46 8 731 77 99 E-Mail: support@dv-power.com Website: http://www.dv-power.com M-RXX0TW-201-U-EN 1.800.561.8187 www. .com 28 information@itm.com