CAYLAR 08/07/2014 Version 1.3 www.caylar.net MPU 12kW 23 High stability Magnet Power Supply Class 10ppm, 1ppm Presentation True Linear bipolar power supply, on all the range (no switching unit) . The power is provided by a bank of Power Operational Amplifiers. The basic architecture of performances is in a frame of some ppm before any processing and corrections : The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C. Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm load resistors. The zero Volt and the Voltage references are 2ppm/°C class (Analog device). In normal mode, the setting to a PID is issued from a couple of DAC 20 bits (4 DAC’s of 20 bits). The measurements are done by 3 ADC’s of 32bits. Data management provided by embedded PC X86 (PC104). The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally. General Main max voltage Output (V) max current Output (A) Max power Output Polarity Class Current set/reading resolution Field set/reading resolution (Option) Speed Control Ramping Stability/AC power source (240Vac, 400Vac, 50Hz)+/-10% Stability/Variations of 90% load Stability long term (after 5 minutes ON) Low frequency noise (0.01-1Hz) Noise on 3 milliohms Stability in Temperature Regulation Zero crossing 30 V 400 A 12 kVA Bipolar 4 quadrants 10ppm, 1ppm 1mA 0.1 G, 0.1mT mA/s ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) (1.10-6 + 1.10-6.In) <8 mV rms (150Hz) <1 mV rms (30Hz-10kHz ±(1.10-6 + 1.10-6 In) / °C Current, Voltage, Field Perfect, no switch Securities Securities Heating element External urgency Default Indication Digital or/and analog Emergency Action Temperature & High Inductance Load Temperature switch 3 interlocks LED on front panel Mass Dimensions Main Power bus Temperature of operations Temperature of storage Cooling Up to 400 kg 16U Baie, 600mm 400Vac 32 A -30°C to +55°C -55°C to +70°C Water Interface Display Analog Control Control by external PC Control in Local mode 2 OLED Display 128x64 +/- 10Vdc RS232, TCP/IP, MODBUS Mini keyboard Control and Reading Regulation Current Regulation Field Regulation (Option) Main Current setting 100% Fine Current setting 1% Measure Current/Voltage/Field Current, Field Analogic PID & Digital PID Digital PID 1 DAC 20 bits by Polarity 1 DAC 20 bits by Polarity 3 x ADC 32 bits Ramp to Zero CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net . APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618 1/5 CAYLAR 08/07/2014 Version 1.3 www.caylar.net MPU 4.2kW 23 High stability Magnet Power Supply Class 10ppm, 1ppm Presentation True Linear bipolar power supply, on all the range (no switching unit) The power is provided by a bank of Power Operational Amplifiers. The basic architecture of performances is in a frame of some ppm before any processing and corrections : The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C. Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm load resistors. The zero Volt and the Voltage references are 2ppm/°C class (Analog device). In normal mode, the setting to a PID is issued from a couple of DAC 20 bits (4 DAC’s of 20 bits). The measurements are done by 3 ADC’s of 32bits. Data management provided by embedded PC X86 (PC104). The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally. General Main max voltage Output (V) max current Output (A) Max power Output Polarity Class Current set/reading resolution Field set/reading resolution (Option) Speed Control Ramping Stability/AC power source (240Vac, 400Vac, 50Hz)+/-10% Stability/Variations of 90% load Stability long term (after 5 minutes ON) Low frequency noise (0.01-1Hz) Noise on 3 milliohms Stability in Temperature Regulation 30 V 240A 4.2 kVA Bipolar 10ppm, 1ppm 1mA 0.1 G, 0.1mT mA/s ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) (1.10-6 + 1.10-6.In) <8 mV rms (150Hz) <1 mV rms (30Hz-10kHz ±(1.10-6 + 1.10-6 In) par °C Current, Field Securities Securities Heating element External urgency Default Indication Digital or analog Emergency Action Temperature & High Inductance Load Temperature switch 3 interlocks LED on front panel Ramp to Zero Mass Dimensions Main Power bus Temperature of operations Temperature of storage Cooling Up to 200 kg 9U Baie, 600mm 400Vac 7 A -30°C to +55°C -55°C to +70°C Water Interface Display Analog Control Control by external PC Control in Local mode 2 OLED Display 128x64 +/- 10Vdc RS232, TCP/IP, MODBUS Mini keyboard Control and Reading Regulation Current Regulation Field Regulation (Option) Main Current setting 100% Fine Current setting 1% Measure Current/Voltage/Field Current, Field Analogic PID & Digital PID Digital PID 1 DAC 20 bits by Polarity 1 DAC 20 bits by Polarity 3 x ADC 32 bits CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net . APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618 2/5 CAYLAR 08/07/2014 Version 1.3 www.caylar.net MPU 1kW 23 High stability Magnet Power Supply Class 10ppm, 1ppm Presentation True Linear bipolar power supply, on all the range (no switching unit) The power is provided by a bank of Power Operational Amplifiers. The basic architecture of performances is in a frame of some ppm before any processing and corrections : The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C. Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm load resistors. The zero Volt and the Voltage references are 2ppm/°C class (Analog device). In normal mode, the setting to a PID is issued from a couple of DAC 20 bits (4 DAC’s of 20 bits). The measurements are done by 3 ADC’s of 32bits. Data management provided by embedded PC X86 (PC104). The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally. General Main max voltage Output (V) max current Output (A) Max power Output Polarity Class Current set/reading resolution Field set/reading resolution (Option) Speed Control Ramping Stability/AC power source (240Vac, 400Vac, 50Hz)+/-10% Stability/Variations of 90% load Stability long term (after 5 minutes ON) Low frequency noise (0.01-1Hz) Noise on 3 milliohms Stability in Temperature Regulation 30 V 180 A 1.0 kVA Bipolar 10ppm, 1ppm 1mA 0.1 G, 0.1mT mA/s ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) (1.10-6 + 1.10-6.In) <8 mV rms (150Hz) <1 mV rms (30Hz-10kHz ±(1.10-6 + 1.10-6 In) par °C Current, Field Securities Securities Heating element External urgency Default Indication Digital or analog Emergency Action Temperature & High Inductance Load Temperature switch 3 interlocks LED on front panel Ramp to Zero Mass Dimensions Main Power bus Temperature of operations Temperature of storage Cooling Up to 60 kg 19’’ Racks 220Vac 10 A -30°C to +55°C -55°C to +70°C Water Interface Display Analog Control Control by external PC Control in Local mode 2 OLED Display 128x64 +/- 10Vdc RS232, TCP/IP, MODBUS Mini keyboard Control and Reading Regulation Current Regulation Field Regulation (Option) Main Current setting 100% Fine Current setting 1% Measure Current/Voltage/Field Current, Field Analogic PID & Digital PID Digital PID 1 DAC 20 bits by Polarity 1 DAC 20 bits by Polarity 3 x ADC 32 bits CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net . APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618 3/5 CAYLAR 08/07/2014 Version 1.3 www.caylar.net MPU Supra 23 High stability Supra Magnet Power Supply Class 10ppm, 1ppm Presentation True Linear bipolar power supply, on all the range (no switching unit) The power is provided by a bank of Power Operational Amplifiers. The basic architecture of performances is in a frame of some ppm before any processing and corrections : The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C. Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm load resistors. The zero Volt and the Voltage references are 2ppm/°C class (Analog device). In normal mode, the setting to a PID is issued from a couple of DAC 20 bits (4 DAC’s of 20 bits). The measurements are done by 3 ADC’s of 32bits. Data management provided by embedded PC X86 (PC104). The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally. General Main max voltage Output (V) max current Output (A) Max power Output Polarity Class Current set/reading resolution Field set/reading resolution (Option) Speed Control Ramping Stability/AC power source (240Vac, 400Vac, 50Hz)+/-10% Stability/Variations of 90% load Stability long term (after 5 minutes ON) Low frequency noise (0.01-1Hz) Noise on 3 milliohms Stability in Temperature Regulation 5V 150 A 500 VA Bipolar 10ppm, 1ppm 1mA 0.1 G, 0.1mT mA/s ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) ± (1.10-6 + 1.10-6 In) (1.10-6 + 1.10-6.In) <8 mV rms (150Hz) <1 mV rms (30Hz-10kHz ±(1.10-6 + 1.10-6 In) par °C Current, Field Securities Securities Heating element External urgency Default Indication Digital or analog Emergency Action Temperature & High Inductance Load Temperature switch 3 interlocks LED on front panel Ramp to Zero Mass Dimensions Main Power bus Temperature of operations Temperature of storage Cooling 40 kg 19’’ rack 220Vac 7 A -30°C to +55°C -55°C to +70°C Water Interface Display Analog Control Control by external PC Control in Local mode 2 OLED Display 128x64 +/- 10Vdc RS232, TCP/IP, MODBUS Mini keyboard Control and Reading Regulation Current Regulation Field Regulation (Option) Main Current setting 100% Fine Current setting 1% Measure Current/Voltage/Field Current, Field Analogic PID & Digital PID Digital PID 1 DAC 20 bits by Polarity 1 DAC 20 bits by Polarity 3 x ADC 32 bits CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net . APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618 4/5