MPU 12kW
23
High stability Magnet Power Supply
Class 10ppm, 1ppm
Presentation
True Linear bipolar power supply, on all the range (no switching unit) .
The power is provided by a bank of Power Operational Amplifiers.
The basic architecture of performances is in a frame of some ppm before any processing and
corrections :
The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C.
Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm
load resistors.
The zero Volt and the Voltage references are 2ppm/°C class (Analog device).
In normal mode, the setting to a PID is issued from a couple of DAC 20 bits
(4 DAC’s of 20 bits).
The measurements are done by 3 ADC’s of 32bits.
Data management provided by embedded PC X86 (PC104).
The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode
(mini pad) or externally.
General
Main
max voltage Output (V)
max current Output (A)
Max power Output
Polarity
Class
Current set/reading resolution
Field set/reading resolution (Option)
Speed Control Ramping
Stability/AC power source (240Vac,
400Vac, 50Hz)+/-10%
Stability/Variations of 90% load
Stability long term (after 5 minutes
ON)
Low frequency noise (0.01-1Hz)
Noise on 3 milliohms
Stability in Temperature
Regulation
Zero crossing
30 V
400 A
12 kVA
Bipolar 4 quadrants
10ppm, 1ppm
1mA
0.1 G, 0.1mT
mA/s
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
(1.10-6 + 1.10-6.In)
<8 mV rms (150Hz)
<1 mV rms (30Hz-10kHz
±(1.10-6 + 1.10-6 In) / °C
Current, Voltage, Field
Perfect, no switch
Securities
Securities
Heating element
External urgency
Default Indication
Digital or/and analog Emergency
Action
Temperature & High
Inductance Load
Temperature switch
3 interlocks
LED on front panel
Mass
Dimensions
Main Power bus
Temperature of operations
Temperature of storage
Cooling
Up to 400 kg
16U Baie, 600mm
400Vac 32 A
-30°C to +55°C
-55°C to +70°C
Water
Interface
Display
Analog Control
Control by external PC
Control in Local mode
2 OLED Display 128x64
+/- 10Vdc
RS232, TCP/IP, MODBUS
Mini keyboard
Control and Reading
Regulation
Current Regulation
Field Regulation (Option)
Main Current setting 100%
Fine Current setting 1%
Measure
Current/Voltage/Field
Current, Field
Analogic PID & Digital PID
Digital PID
1 DAC 20 bits by Polarity
1 DAC 20 bits by Polarity
3 x ADC 32 bits
Ramp to Zero
CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE
Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net .
APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618
1/5
CAYLAR
08/07/2014 Version 1.3
www.caylar.net
MPU 4.2kW
23
High stability Magnet Power Supply
Class 10ppm, 1ppm
Presentation
True Linear bipolar power supply, on all the range (no switching unit)
The power is provided by a bank of Power Operational Amplifiers.
The basic architecture of performances is in a frame of some ppm before any processing and
corrections :
The Power Operational Amplifier gain is fixed by the ratio of two resistors 10ppm/C.
Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm, using 0.2ppm
load resistors.
The zero Volt and the Voltage references are 2ppm/°C class (Analog device).
In normal mode, the setting to a PID is issued from a couple of DAC 20 bits
(4 DAC’s of 20 bits).
The measurements are done by 3 ADC’s of 32bits.
Data management provided by embedded PC X86 (PC104).
The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode
(mini pad) or externally.
General
Main
max voltage Output (V)
max current Output (A)
Max power Output
Polarity
Class
Current set/reading resolution
Field set/reading resolution (Option)
Speed Control Ramping
Stability/AC power source (240Vac,
400Vac, 50Hz)+/-10%
Stability/Variations of 90% load
Stability long term (after 5 minutes
ON)
Low frequency noise (0.01-1Hz)
Noise on 3 milliohms
Stability in Temperature
Regulation
30 V
240A
4.2 kVA
Bipolar
10ppm, 1ppm
1mA
0.1 G, 0.1mT
mA/s
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
(1.10-6 + 1.10-6.In)
<8 mV rms (150Hz)
<1 mV rms (30Hz-10kHz
±(1.10-6 + 1.10-6 In) par
°C
Current, Field
Securities
Securities
Heating element
External urgency
Default Indication
Digital or analog Emergency Action
Temperature & High
Inductance Load
Temperature switch
3 interlocks
LED on front panel
Ramp to Zero
Mass
Dimensions
Main Power bus
Temperature of operations
Temperature of storage
Cooling
Up to 200 kg
9U Baie, 600mm
400Vac 7 A
-30°C to +55°C
-55°C to +70°C
Water
Interface
Display
Analog Control
Control by external PC
Control in Local mode
2 OLED Display 128x64
+/- 10Vdc
RS232, TCP/IP, MODBUS
Mini keyboard
Control and Reading
Regulation
Current Regulation
Field Regulation (Option)
Main Current setting 100%
Fine Current setting 1%
Measure
Current/Voltage/Field
Current, Field
Analogic PID & Digital PID
Digital PID
1 DAC 20 bits by Polarity
1 DAC 20 bits by Polarity
3 x ADC 32 bits
CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE
Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net .
APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618
2/5
CAYLAR
08/07/2014 Version 1.3
www.caylar.net
MPU 1kW
23
High stability Magnet Power Supply
Class 10ppm, 1ppm
Presentation
True Linear bipolar power supply, on all the range (no switching unit)
The power is provided by a bank of Power Operational Amplifiers.
The basic architecture of performances is in a frame of some ppm before any
processing and corrections :
The Power Operational Amplifier gain is fixed by the ratio of two resistors
10ppm/C.
Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity 3ppm,
using 0.2ppm load resistors.
The zero Volt and the Voltage references are 2ppm/°C class (Analog device).
In normal mode, the setting to a PID is issued from a couple of DAC 20 bits
(4 DAC’s of 20 bits).
The measurements are done by 3 ADC’s of 32bits.
Data management provided by embedded PC X86 (PC104).
The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally.
General
Main
max voltage Output (V)
max current Output (A)
Max power Output
Polarity
Class
Current set/reading resolution
Field set/reading resolution (Option)
Speed Control Ramping
Stability/AC power source (240Vac,
400Vac, 50Hz)+/-10%
Stability/Variations of 90% load
Stability long term (after 5 minutes
ON)
Low frequency noise (0.01-1Hz)
Noise on 3 milliohms
Stability in Temperature
Regulation
30 V
180 A
1.0 kVA
Bipolar
10ppm, 1ppm
1mA
0.1 G, 0.1mT
mA/s
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
(1.10-6 + 1.10-6.In)
<8 mV rms (150Hz)
<1 mV rms (30Hz-10kHz
±(1.10-6 + 1.10-6 In) par
°C
Current, Field
Securities
Securities
Heating element
External urgency
Default Indication
Digital or analog Emergency Action
Temperature & High
Inductance Load
Temperature switch
3 interlocks
LED on front panel
Ramp to Zero
Mass
Dimensions
Main Power bus
Temperature of operations
Temperature of storage
Cooling
Up to 60 kg
19’’ Racks
220Vac 10 A
-30°C to +55°C
-55°C to +70°C
Water
Interface
Display
Analog Control
Control by external PC
Control in Local mode
2 OLED Display 128x64
+/- 10Vdc
RS232, TCP/IP, MODBUS
Mini keyboard
Control and Reading
Regulation
Current Regulation
Field Regulation (Option)
Main Current setting 100%
Fine Current setting 1%
Measure
Current/Voltage/Field
Current, Field
Analogic PID & Digital PID
Digital PID
1 DAC 20 bits by Polarity
1 DAC 20 bits by Polarity
3 x ADC 32 bits
CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE
Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net .
APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618
3/5
CAYLAR
08/07/2014 Version 1.3
www.caylar.net
MPU Supra
23
High stability Supra Magnet Power Supply
Class 10ppm, 1ppm
Presentation
True Linear bipolar power supply, on all the range (no switching unit)
The power is provided by a bank of Power Operational Amplifiers.
The basic architecture of performances is in a frame of some ppm before any
processing and corrections :
The Power Operational Amplifier gain is fixed by the ratio of two resistors
10ppm/C.
Range 0-400 Amperes measured by a DCCT 400 A, class 0.5ppm, linearity
3ppm, using 0.2ppm load resistors.
The zero Volt and the Voltage references are 2ppm/°C class (Analog device).
In normal mode, the setting to a PID is issued from a couple of DAC 20 bits
(4 DAC’s of 20 bits).
The measurements are done by 3 ADC’s of 32bits.
Data management provided by embedded PC X86 (PC104).
The controls can be done through a computer (RS232, Ethernet,MODBUS), or in local mode (mini pad) or externally.
General
Main
max voltage Output (V)
max current Output (A)
Max power Output
Polarity
Class
Current set/reading resolution
Field set/reading resolution (Option)
Speed Control Ramping
Stability/AC power source (240Vac,
400Vac, 50Hz)+/-10%
Stability/Variations of 90% load
Stability long term (after 5 minutes
ON)
Low frequency noise (0.01-1Hz)
Noise on 3 milliohms
Stability in Temperature
Regulation
5V
150 A
500 VA
Bipolar
10ppm, 1ppm
1mA
0.1 G, 0.1mT
mA/s
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
± (1.10-6 + 1.10-6 In)
(1.10-6 + 1.10-6.In)
<8 mV rms (150Hz)
<1 mV rms (30Hz-10kHz
±(1.10-6 + 1.10-6 In) par
°C
Current, Field
Securities
Securities
Heating element
External urgency
Default Indication
Digital or analog Emergency Action
Temperature & High
Inductance Load
Temperature switch
3 interlocks
LED on front panel
Ramp to Zero
Mass
Dimensions
Main Power bus
Temperature of operations
Temperature of storage
Cooling
40 kg
19’’ rack
220Vac 7 A
-30°C to +55°C
-55°C to +70°C
Water
Interface
Display
Analog Control
Control by external PC
Control in Local mode
2 OLED Display 128x64
+/- 10Vdc
RS232, TCP/IP, MODBUS
Mini keyboard
Control and Reading
Regulation
Current Regulation
Field Regulation (Option)
Main Current setting 100%
Fine Current setting 1%
Measure
Current/Voltage/Field
Current, Field
Analogic PID & Digital PID
Digital PID
1 DAC 20 bits by Polarity
1 DAC 20 bits by Polarity
3 x ADC 32 bits
CAYLAR sas 14 Avenue du Quebec - BP 711 91140 Villebon-sur Yvette- FRANCE
Tel.: 33 1 6929 9162 - Fax .: 33 1 6929 9560 - Mail : caylar@caylar.net - www.caylar.net .
APE2651B – Siret 441 581 618 00028 - TVA n° : FR 554 415 81618
4/5
Download