H100 COND Transmitter Specification Data Sheet

H100 COND Transmitter
Specification Data Sheet
Hamilton H100 COND Transmitter combines ease of use and reliability. It has
been designed for universal process application including use in pharmaceutical,
chemical, food & beverage industries as well as water / waste water treatment.
The self-explaining user interface ensures comfortable and intuitive handling.
Hamilton H100 COND Transmitter provides continuous sensor monitoring and
preventive maintenance indication for maximal reliability.
H100 COND Specifications
COND input
Input for 2-electrodes or 4-electrodes
Effective ranges
Conductivity 0.2 • c ... 1000 mS • c
Measuring ranges
Conductivity
0.000 ... 9.999 μS/cm
00.00 ... 99.99 μS/cm
000.0 ... 999.9 μS/cm
0000 ... 9999 μS/cm
0.000 ... 9.999 mS/cm
00.00 ... 99.99 mS/cm
000.0 ... 999.9 mS/cm
0.000 ... 9.999 S/cm
00.00 ... 99.99 S/cm
Resistivity
00.00 ... 99.99 MOhm • cm
Concentration
0.00 ... 9.99 % by wt
Salinity
0.0 ... 45.0 ‰ (0 ... 35 °C)
Response tine (T 90 )
< 1 s (Sensocheck off)
< 3 s (Sensocheck on)
Meas. error
< 1 % meas. val. + 0.4 μS • c
1, 2, 3)
Concentration Determination
01
NaCI
Modes
02
HCI
03
NaOH
04
H 2SO 4
05
HNO 3
*)
Sensor standardization
Operating modes
- Input of cell constant with simultaneous display of conductivity and temperature
- Input of conductivity of calibration solution with simultaneous display of cell constant and
aitemperature
- Product calibration
- Temperature probe adjustment
Permissible cell constant
H100 DO Specifications (continued)
Sensor monitoring
Sensocheck
Polarization detection and monitoring of cable capacitance
®
Provides information on the sensor
condition (Sensocheck ®)
Sensor monitor
Direct display of measured values from sensor for validation (resistance / temperature)
Temperature input
*)
Pt 100 / Pt 1000 / NTC 30 kOhm
2-wire connection, adjustable
Measuring range
Pt 100 / Pt 1000
-20.0 ... +200.0 °C (-4 ... +392 °F)
NTC 30 kOhm
Resolution
i
-20.0 ... +150.0 °C (-4 ... +302 °F)
0.1 °C / 0.1 °F
Meas. error
< 0.5 (< 1K for Pt 100; < 1K for NTC >100 °C)
1, 2, 3)
Temperature compensation
(ref. temperature 25 °C)
i
i
Output 1
Overrange
0/4 … 20 mA, max. 10 V, floating (galvanically connected to output 2)
22 mA in the case of error messages
*)
Output filter
Low-pass, filter time constant: 0 … 120 s
*)
Minumum spam
Process variable
Temperature
Measurement error
Start/end of scale
1)
< 0.3 % current value + 0.05 mA
*)
Permissible span
Alarm contact
Relay contact, floating
Display
approx. 5 VA, 2.5 W
Overvoltage category II, protection class II
Ambient temperature
–20 … +55 °C
Transport/Storage temp
–20 … +70 °C
Relative humidity
EMC
Emitted interference
Class B (residential area),
Immunity to interference
Industry
Enclosure
Molded enclosure made from PBT (polybutylene terephtlate)
pipe mounting 243082 (
Ingress protection
IP 65 / NEMA 4X
*) User-defined; 1) To IEC 746 Part 1, at nominal operating conditions; 2) ±1 count; 3) Plus sensor error
© 2013 Hamilton Bonaduz AG. All rights reserved.
Calimatic ® and Sensocheck® are registed trademarks by Knick Elektronische Messgeräte GmbH & Co. KG
P/N: 695155/00 — 11/2013
Hamilton Americas & Pacific Rim
Web: www.hamiltoncompany.com
800-648-5950
Europe: +41-81-660-60-60
USA:
4970 Energy Way
Reno, Nevada 89502 USA
Tel: +1-775-858-3000
Fax: +1-775-856-7259
sales@hamiltoncompany.com
Hamilton Europe, Asia, & Africa
Via Crusch 8
CH-7402 Bonaduz, Switzerland
Tel: +41-81-660-60-60
Fax: +41-81-660-60-70
contact@hamilton.ch
To find a representative in your area, please visit hamiltoncompany.com/contacts.