Diagnostic Ability of the Daniel Four Path Ultrasonic Flow Meter page 1 DANIEL MEASUREMENT AND CONTROL WHITE PAPERS Diagnostic Ability of the Daniel Four Path Ultrasonic Flow Meter www.daniel.com 1 INTRODUCTION 3 VELOCITY PROFILE The primary function of the ultrasonic meter is to measure the The four chordal velocities give an indication of the velocity actual volume flow rate. The process involves measuring four profile in the meter, established by the flow through the upstream velocities on chords located in four different radial positions and pipe work. Three helpful ratios can be defined as: Asymmetry = in two different vertical planes. The eight transducers are fired (VA + VB)/(VC + VD), Cross flow = (VA + VC)/(VB + VD) and Swirl about 50 times per second and the transit times to traverse each chord in both directions are measured. This vast array of data = (VB+ VC)/(VA + VD). The asymmetry compares the flow in the top half of the pipe with can be processed to yield useful diagnostic information, which that in the bottom half; in good condition it should be close to 1. forms the subject of this paper and shows that the four-path The cross flow compares the chords in one plane with those in ultrasonic meter does much more than just measure the flow the other plane at right angles: in good condition it should be rate. It has sufficient diagnostic ability to confirm the authenticity close to 1. The swirl compares the inner chords to the outer of the measurement, and develop the source for conditional chords and it is an indicator of swirl due to both the different based maintenance and re-calibration. radial locations and planes. In good condition the swirl should be close to 1.042/0.89 = 1.17 [1]. 2 METER GEOMETRY AND FUNCTION Fig, 1a. End View Fig. 1b. Top View There are two equations for the transit time with (t1) and against (t2) the flow, which can be solved for the chord velocity (Vi) Where C = speed of sound L = distance between transducers X = axial distance in the flow R = meter radius The average velocity (V) and actual volume flow rate (Q) are found from: W = the weighting factor DANIEL MEASUREMENT AND CONTROL WHITE PAPERS Examples of analytical flow profiles with swirl are shown in Fig. 2, and how the swirl ratio can estimate the magnitude of the swirl angle in Fig. 3. Fig. 2 VELOCITY PROFILE WITH SWIRL Fig. 3 SWIRL RATIO (Vb+Vc)/(Va+Vd) Actual measured ratios are shown in Fig. 4 for a 20” meter. Fig. 3 SWIRL RATIO (Vb+Vc)/(Va+Vd) page 2 Diagnostic Ability of the Daniel Four Path Ultrasonic Flow Meter page 3 Because of tolerances in the manufacture of the ultrasonic is higher on the outer chord, nearer the pipe wall where there is meter and the difficulty of achieving ideal fully developed flow, more shear. the actual ratios are never exactly at their theoretical values, and give the meter a unique fingerprint. However they are If flow pulsations or fluctuations are present, this turbulence usually within ±2%, which is sufficient to say that the flow profile parameter will increase above the 2% to 5% value associated is near perfect, which is not surprising since Fig.4 comes from with good flow. a lab calibration. In general, four paths are not sufficient to resolve any arbitrary 3-dimensional flow field containing asymmetry, swirl and cross flow. However, fiscal flow measurement practice attempts to establish good flow conditions, which can certainly be verified by these ratios. Never the less, if the ratios differ significantly from their ideal value they can give a reasonable indication of the type of disturbance, especially if only one of the ratios has 4 SPEED OF SOUND The speed of sound can be evaluated from the same two transit times for each chord: The average speed of sound C is just the average of the chords C = (CA +CB + CC +CD)/4 changed significantly. The speed of sound can be used as a diagnostic in several 3.1 Turbulence, pulsation and fluctuation 4.1 Compare a measured and calculated value In discussing the velocity profile we have been considering the The speed of sound can be calculated from independent average velocity obtained from a batch of transit times from all eight transducers. The velocity depends upon the transit time difference Δt = (t2 – t1). Typically a batch of 20 Δt values is used to determine an average Δt and hence the average velocity. The batch of 20 also allows the calculation of a standard deviation σΔt, and then the ratio σΔt/Δt is a measure of turbulence or ways: measurements of gas composition, pressure and temperature, using methods described in AGA 10 [2]. This is an independent check on the transit time measurement and is important because the ultrasonic meter basically measures transit times. An example is shown in Fig. 6 for a 4” meter. Note that 4” is velocity fluctuation. the smallest ultrasonic meter typically used and the small length Some typical values are given in Fig. 5, for the inner and outer determine the speed of sound. chords of a meter during calibration in good flow. The turbulence varies from 2 to 5% of the velocity. The inner chord-C has an average turbulence of 3%, while the outer chord-D has an average turbulence of 4%. This is in keeping with typical point values in pipe flow of 6% at z/R = 0.309 and 9% at z/R = 0.809, and the general concept that the turbulence Fig. 5 TURBULENCE % = 100* STD-DEV Dt / Dt and short transit time make it the most difficult size to accurately Despite the small size the agreement between measured and calculate speed of sound is good to within ± 0.02%. This calibration at SwRI re-circulates the same gas in a closed loop; hence there is no change in gas composition for the results in Fig. 6. Fig. 6 4” METER SOS DEVIATION FROM AGA 10 DANIEL MEASUREMENT AND CONTROL WHITE PAPERS page 4 Fig 9. SOS LONG - SOS SHORT % 4.2 Compare the 4 chordal values with one another At a reasonable gas velocity (say > 5 ft/s), the gas composition and temperature should be uniform over the pipe cross-section and all chords should read the same speed of sound. See Fig. 7 for an example from a 12” meter. If all four values agree, it is more likely that they are all correct than all in error by exactly the same amount. Fig. 7 CHORDAL SPEED OF SOUND FT/S = 2.7R giving a length ratio = 2.7/1.7 = 1.6. If the speed of sound is the same on chords of different length, this indicates that both the transit times and delay times (time not spent in the gas) are correct. The normal way of determining the delay time is with a gas of known speed of sound, typically nitrogen. Another way is to use two different lengths in the same, but unknown, speed of The values of speed of sound on all four chords agree to within sound. Thus the different chord lengths allow a dynamic method ± 1 ft/s This is not quite as good a test as the one in section 4.1, of checking the delay times. but it has the advantage of not requiring any additional external inputs (pressure, temperature & composition). 4.3 Use the difference between the chordal and average value Fig. 9 shows very small variation of 0.002% SOS or about 20 ns in transit time, which is close to the stability limit of the metering system. Because of manufacturing tolerances the four chordal values Note: three examples have been shown where the SOS of speed of sound will never be exactly equal. These small variations are 0.1%, 0.02% and 0.002%. A good value for the differences can be used as another fingerprint of the meter. average field application would be 0.05%. See Fig.8 as an example Fig. 8 SOS DIFFERENCE FROM AVERAGE % 4.5 Correction for stratification An example is given for a velocity of 2 ft/s in a 12” pipe with the gas temperature at 70o F and the ambient temperature at 90o F, in a long un-insulated pipe, shown in Figs 10 & 11. Fig 10. TEMPERATURE [SOS] GRADIENT 4.4 Check the agreement between chords of different length Fig.9 shows an example for a 12” meter under stable zero flow conditions. The 4-path Daniel meter has paths of different length. The outer chords (A & D) have L = 1.7R and the inner chords (B&C) have L The 4 chords are horizontal and spaced vertically, so they see the temperature stratification: low density gas, with high temperature and higher speed of sound floats on top of the Diagnostic Ability of the Daniel Four Path Ultrasonic Flow Meter page 5 higher density gas, with lower temperature and lower speed of • An upstream flow control valve being adjusted sound. In this case a SOS gradient of 8 ft/s corresponds to the • An upstream branching flow of varying proportion to temperature gradient of 6oF. This temperature difference also the metered flow drives a natural convection current: gas heated by the pipe wall • A flow conditioner trapping debris rises to the top of the pipe, while cold gas in the center falls to • Erosion, corrosion or deposition changing the the bottom of the pipe. This convection current, which is normal to the axial flow, pushes the maximum axial velocity down from upstream pipe roughness the center of the pipe, and the effect is seen in the velocity 5.1 Pipe roughness profile shown in Fig. 11. The meter can detect an increase in pipe roughness by its effect on the velocity profile [3]. In general the velocity profile depends non- upon both roughness and Reynolds Number, but at the high Re representative gas temperature measurement. A reasonable No associated with natural gas transport the roughness effects approach to the correct gas temperature is a Flow Weighted dominate. The profile can be described by the ratio (VB+ VC)/(VA Mean Temperature (FWMT) based on using the same velocity + VD), but here we are looking for change from a flatter profile to flow weighting factors Wi for the chordal temperatures. The result a more pointed profile (Fig.12) with time, and not swirl. in this case is FWMT = 73.5o F seen in Fig.10. Unfortunately The ratio (VB+ VC)/(VA + VD) can be linked to pipe roughness (Fig. the high pipe temperature and the highest gas temperature 13) in terms of relative roughness k/D, where k = the roughness influence the thermo-well located on top of the pipe, and reads height and D = the pipe diameter. Another consequence of stratification is a poor a temperature = 77.8o F. This error of 4.3o F in 530 o R will give a 0.8% error in the corrected volume flow. The SOS gradient can Again the idea is that fixed piping gives a fixed velocity profile, so be used to recognize and correct the error; this is unique to the a changing profile with time is an indication of pipe roughness. If chordal ultrasonic meter, while non-chordal ultrasonic meters the velocity profile does not change with time, it is an indication and other flow meters would suffer from stratification without that surface roughness has probably remained unchanged. knowing. Fig.12 VELOCITY PROFILE - ROUGH Fig11. VELOCITY PROFILE WITH CONVECTION 5. TRENDING All the diagnostics discussed above can be trended to see if anything is changing with time, by recording the data at suitable intervals. The time interval depends on the specific application, but it can be extended if nothing is seen to change. The velocity profile is determined by the pipe and fittings upstream of the meter. The pipe and fittings are fixed, so one would normally not expected the velocity profile to change. However there are some exceptions where the velocity profile would change, due to: DANIEL MEASUREMENT AND CONTROL WHITE PAPERS Fig. 13 VELOCITY RATIO (B + C) / (A + D) - ROUGHNESS page 6 useful diagnostic. As part of the signal detection many checks are made: • The noise, found upstream of the signal • The ratio of signal to noise • The standard deviation of the transit times • The signal quality, a measure of how quickly the signal rises • A comparison between the transmitted and received signals • A comparison between the upstream and downstream received signals 5.2 Speed of Sound The SOS will change with gas composition, temperature and pressure, and can be a useful check on plant operation. In particular, if nitrogen purging is required for safety, the SOS can detect the nitrogen to natural gas change over. The speed of sound depends only on the distance between the transducers and the transit time. If it is not changing with time, either internally between the chords or externally compared to independent calculations, it is a very good sign that the meter is functioning correctly, and the other inputs (P, T & gas composition) needed for the calculation must be correct. If it does change it could be due to deposits on the transducer face or transducer damage. Deposits on the transducer would probably be accompanied by deposit on the pipe wall, which would also show in the velocity profile. The SOS and velocity profile can certainly confirm that the meter is functioning correctly, but if they reveal a problem then a deeper level of diagnostics is required. 6 DIAGNOSTIC LEVEL The discussion so far has made use of the four chord velocities and the four chord values of SOS, which are directly available from the meter output. There is a deeper level of diagnostic associated with the digital signal processing. 6.1 Digital Signal Processing The detection of the received ultrasonic signal must find a consistent reliable zero crossing to use for the transit time measurement. The transmitted signal strength is limited by intrinsic safety considerations, so the received signal amplitude is a function of the acoustic impedance of the gas, the meter size and the gas velocity. An automatic gain control (AGC) is applied to the received signal to always achieve the same amplitude, to simplify detection. The value of the gain is a measure of the health of the transducer or attenuation in the path, and is a • A check for peak switching • The tracking of a consistent zero crossing If a signal does not pass all these tests it is not used for a transit time measurement and an alarm is given that is decoded to explain why the signal was rejected. This is a very useful diagnostic. The number of signals used in a batch is reported as Performance. A 100% performance is a sign that the meter is working well, and is the normally expected performance up to the full rated capacity of the meter. Anything less than 100% is another useful diagnostic. An example of the power of this deeper diagnostics is the ability to detect wet gas [4]. The presence of liquid drops dispersed in the gas affect the SOS, gain, standard deviation, signal to noise, signal quality and performance. Deposits on the pipe wall change the velocity profile and deposits on the transducer may change the SOS, but they would also affect the gain and signal to noise. 6.2 Multiple diagnostic parameters The velocity profile gives a fingerprint, detects asymmetry, swirl and cross flow, while the standard deviation of the velocity indicates turbulence (5 parameters). The SOS gives another fingerprint. It can be compared with calculations from gas composition, pressure & temperature, and gives four values from two different chord lengths, that can detect stratification & convection currents, and can correct temperature measurements (6 parameters). The DSP and AGC add a further 9 parameters for each of the eight transducers. The waveform and spectra (frequency content) can be displayed, another 2 parameters. This gives a total of 22 potentially useful diagnostic parameters. If all 22 parameters are normal, then there is no doubt that the meter is working correctly. If the meter fails, there is sufficient information to diagnose and fix the problem. If there are problems with the flow metering system, and one has sufficient confidence Diagnostic Ability of the Daniel Four Path Ultrasonic Flow Meter page 7 that the meter is working correctly, it is then possible to look To change this situation, a manufacturer could offer a “health for other system problems. Typical system problems are swirl, check” service where say once a month he had remote access turbulence, pulsations, fluctuations and noise. to all the diagnostic data to give the meter a clean bill of health, or raise warning signs. This should be mutually beneficial, leading If a transducer is short circuited to the body by liquid or debris, to conditional based maintenance and recalibration instead of noise through the body will show on all transducers, while fixed routine maintenance and mandatory recalibration intervals. external noise will be much more on the transducers facing the noise source. The meter will indicate where to look for the noise source, upstream or downstream. The waveform and spectra will also give information on the noise as well as the signal. Gain can be useful, but remember it is dependent on the gas Klaus J. Zanker, Daniel Industries, USA 7 REFERENCES 1. Klaus J. Zanker, INSTALLATION EFFECTS ON SINGLE- pressure, or more precisely the gas acoustic impedance = SOS AND MULTI-PATH ULTRASONIC METERS, Flomeko 2000 * density. If gain increase with time (for the same pressure) there 2. AGA Report No 10, SPEED OF SOUND IN NATURAL GAS can be two causes: transducer damage or obstruction of the AND OTHER RELATED HYDROCARBON GASES, July acoustic path. Further diagnostic can follow from a series of 2002 questions: 3. Klaus J. Zanker, THE EFFECTS OF REYNOLDS NUMBER, • Are one or more transducers involved? WALL ROUGHNESS, AND PROFILE ASYMMETRY ON • Are one or more chords involved? SINGLE- AND MULTI- PATH ULTRASONIC METERS, • Is it the bottom D-Chord, most likely to accumulate NSFMW, Oct 1999 debris and liquid? 4. Klaus J. Zanker & Gregor J. Brown, THE PERFORMANCE • Has the velocity profile changed? OF A MULTI-PATH ULTRASONIC METER WITH WET • Has the SOS profile changed? GAS, NSFMW Oct 2000. • Has the signal quality changed? In general, if any diagnostic parameter is suspect, and one can postulates a cause, it then becomes possible to find other parameters to try to confirm or refute the postulate. 7 CONCLUSIONS A major advantage of the ultrasonic meter is that it produces an abundance of diagnostic data. If all the diagnostic parameters are normal, one can have complete confidence that the meter is working correctly. This confidence is very important, because one can then look for problems in other parts of the metering system. If a meter fails, the diagnostics quickly reveals the problem and solution. The range of decline between perfect performance and complete failure is more difficult to quantify. Trending the diagnostic parameters will certainly show if changes are occurring, but knowing how much change is tolerable, before intervention is necessary, is at present very much a judgment call. As a meter manufacturer, we basically only see the extremes of new meter calibrations and field failures. We do not have access to long-term normal operational data in typical fiscal applications. Emerson Process Management Daniel Measurement and Control, Inc. www.daniel.com North America / Latin America: Headquarters USA - Houston, Texas T +1.713.467.6000 F +1.713.827.3880 USA Toll Free 1.888.FLOW.001 Europe: Stirling, Scotland, UK T +44.1786.433400 F +44.1786.433401 Middle East, Africa: Dubai, UAE T +971.4.811.8100 F +971.4.886.5465 Asia Pacific: Singapore T +65.6777.8211 F +65.6777.0947 / 0743 Daniel Measurement and Control, Inc. is a wholly owned subsidiary of Emerson Electric Co., and a division of Emerson Process Management. The Daniel name and logo are registered trademarks of Daniel Industries, Inc. The Emerson logo is a registered trademark and service mark of Emerson Electric Co. All other trademarks are the property of their respective companies. The contents of this publication are presented for informational purposes only, and while every effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding the products or services described herein or their use or applicability. All sales are governed by Daniel’s terms and conditions, which are available upon request. We reserve the right to modify or improve the designs or specifications of such products at any time. Daniel does not assume responsibility for the selection, use or maintenance of any product. Responsibility for proper selection, use and maintenance of any Daniel product remains solely with the purchaser and end-user. ©2010 Daniel Measurement and Control, Inc. All Rights Reserved. Unauthorized duplication in whole or in part is prohibited. Printed in the USA. DAN-TECHNOLOGIES-DIAGNOSTIC-ABILITY-OF-THE-DANIEL-FOUR-PATH-ULTRASONIC-FLOW-METER