Specifications for the NI PXI-2566 16-Channel SPDT Relay Module This document lists specifications for the NI PXI-2566 general purpose relay module. All specifications are subject to change without notice. Visit ni.com/manuals for the most current specifications. Configuration ......................................... 16-channel SPDT, non-latching Input Characteristics Maximum switching voltage Channel-to-channel ......................... 150 VDC, 125 VAC, CAT I Channel-to-ground .......................... 150 VDC, 125 VAC, CAT I This module is rated for Measurement Category I and intended to carry signal voltages no greater than 150 V. This module can withstand up to 800 V impulse voltage. Do not use this module for connection to signals or for measurements within Categories II, III or IV. Do not connect to MAINs supply circuits (e.g., wall outlets) of 115 or 230 VAC. Refer to the NI Switches Getting Started Guide for more information on measurement categories. Caution When hazardous voltages (>42.4 Vpk/60 VDC) are present on any relay terminal, safety low-voltage (<42.4 Vpk/60 VDC) cannot be connected to any other relay terminal. Maximum switching current .................. 2 ADC, 2 AAC (per channel) Simultaneous channels at maximum switching current (≤25 °C) .................... 16 Maximum carry current ......................... 5 ADC, 5 AAC (per channel) Simultaneous channels at maximum carry current (≤25 °C)............................ 9 National Instruments™, NI™, and ni.com™ are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents. ni.com © 2003 National Instruments Corp. All rights reserved. April 2003 323468A-01 Module Load Derating at >25 °C Load derating is dependent on the ambient temperature and the sum of the current squared of each channel simultaneously carrying a signal. The result must fall within in the shaded region of Figure 1. The following examples represent this calculation. Example 1: 5 channels carry 4 A while 10 channels carry 2 A (5 × 42) + (10 × 22) = 120 A Example 1 can be used at ambient temperatures between 0 °C and 50 °C. Example 2: 6 channels carry 5 A while 10 channels carry 2 A (6 × 52) + (10 × 22) = 190 A Example 2 can be used at ambient temperatures between 0 °C and 41 °C. Power Load (Amps Squared x Channels) 250 200 150 100 50 0 0 10 20 30 40 50 60 Ambient Temperature (°C) Figure 1. Module Load Derating Specifications for the NI PXI-2566 2 ni.com Maximum switching power ................... 60 W, 62.5 VA (DC to 60 Hz) (per channel) DC path resistance Initial ............................................... <0.1 Ω End of life ....................................... ≥1.0 Ω DC path resistance typically remains low for the life of the relay. At the end of relay life, the path resistance rises rapidly above 1 Ω. Load ratings apply to relays used within the specification before the end of relay life. Thermal EMF ......................................... <9 µV (typical at 23 °C) Minimum switching capacity................. 10 µA at 10 mV DC Bandwidth (–3 dB)................................. ≥70 MHz –0.0 Channel 3 –0.5 Loss (dB) –1.0 –1.5 70 MHz –2.0 Channel 12 –2.5 –3.0 –3.5 –4.0 0 50 100 Frequency (MHz) 150 Figure 2. Insertion Loss Measured in a 50 Ω System (Worst Case, Based on 30 Engineering DUTs) © National Instruments Corporation 3 Specifications for the NI PXI-2566 Crosstalk (typical at 23 °C) Channel-to-channel 10 kHz ...................................... ≤ –75 dB 100 kHz .................................... ≤ –65 dB 1 MHz....................................... ≤ –45 dB 10 MHz..................................... ≤ –25 dB Isolation (typical at 23 °C) Open channel 100 kHz .................................... ≥65 dB 1 MHz....................................... ≥45 dB 10 MHz..................................... ≥25 dB Dynamic Characteristics Maximum cycle speed ............................115 cycles/s Relay operate time Typical.............................................2 ms Maximum ........................................4.4 ms Expected relay life Mechanical ......................................100,000,000 cycles Electrical 30 VDC, 1 ADC resistive.........500,000 cycles 30 VDC, 2 ADC resistive.........100,000 cycles 125 VAC, 0.2 AAC resistive....300,000 cycles 125 VAC, 0.5 AAC resistive....100,000 cycles Specifications for the NI PXI-2566 4 ni.com Trigger Characteristics Input trigger Sources............................................ PXI trigger lines 0–7, Front panel Minimum pulse width ..................... 150 ns Front panel/terminal block input voltage Minimum ................................. –0.5 V VL maximum........................... +0.7 V VH minimum ........................... +2.0 V Nominal ................................... +3.3 V Maximum................................. +5.5 V Output trigger Destinations .................................... PXI trigger lines 0–7, Front panel Pulse width...................................... Programmable (1 µs to 62 µs) Front panel nominal voltage ........... 3.3 V TTL, 8 mA Physical Characteristics Relay type .............................................. Electromechanical, non-latching Relay contact material............................ Gold clad silver alloy I/O connectors ........................................ Two 62-pin DSubs PXI power requirement .......................... 4.5 W at 5 V 2.5 W at 3.3 V Dimensions (W × H × D) ....................... 3 cm × 10 cm × 16 cm (0.8 in. × 3.9 in. × 6.3 in.) Weight .................................................... 250 g (9 oz) © National Instruments Corporation 5 Specifications for the NI PXI-2566 Environment Operating temperature ............................0 °C to 50 °C Storage temperature ................................–20 °C to 70 °C Relative humidity ...................................5% to 85% noncondensing Pollution Degree .....................................2 Approved at altitudes up to 2,000 m Accessories Visit ni.com for more information about the following accessories. Table 1. Accessories Available for the NI PXI-2566 Accessory Note Part Number NI TB-2666 terminal block 778717-66 Backshell and connector kit 778720-01 The front connector is keyed and has pins removed for safety isolation. Glossary channel a single SPDT (form C) relay. Each channel has three terminals—common (COM), normally closed (NC), normally open (NO). cycle actuate a SPDT relay twice, leaving it in its original state operate actuate a SPDT relay once, leaving it in the opposite state Specifications for the NI PXI-2566 6 ni.com K8 K0 NC0 NO0 COM0 NC8 NO8 COM1 NC9 NO9 COM2 NC10 NO10 COM3 NC11 NO11 COM4 NC12 NO12 COM5 NC13 NO13 COM6 NC14 NO14 COM7 NC15 NO15 COM8 K9 K1 NC1 NO1 COM9 K10 K2 NC2 NO2 COM10 K11 K3 NC3 NO3 COM11 K12 K4 NC4 NO4 COM12 K13 K5 NC5 NO5 COM13 K14 K6 NC6 NO6 COM14 K15 K7 NC7 NO7 COM15 Figure 3. NI PXI-2566 Power-On State COM0 NO0 COM1 NO1 COM2 NO2 COM3 NO3 COM4 NO4 COM5 NO5 COM6 NO6 COM7 NO7 COM8 NO8 COM9 NO9 COM10 NO10 COM11 NO11 COM12 NO12 COM13 NO13 COM14 NO14 COM15 NO15 TB_EECLK TB_EEOUT TB_EEEN +5V TB_INTERLOCK TRIG IN 22 43 23 44 24 45 25 46 26 47 27 48 28 49 29 50 30 51 31 52 32 53 33 54 34 55 35 56 36 57 37 58 60 40 61 41 62 42 1 2 3 NC0 NC1 NC2 4 NC3 5 NC4 6 NC5 7 NC6 8 NC7 9 NC8 10 NC9 11 NC10 12 NC11 13 NC12 14 NC13 15 NC14 16 19 20 21 NC15 TB_EEMOSI GND Low Voltage TRIG OUT Figure 4. NI PXI-2566 Front Connector © National Instruments Corporation 7 Specifications for the NI PXI-2566 Compliance and Certifications Safety This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: Note • IEC 61010-1, EN 61010-1 • UL 3111-1, UL 61010B-1 • CAN/CSA C22.2 No. 1010.1 For UL and other safety certifications, refer to the product label or to ni.com. Electromagnetic Compatibility Emissions................................................EN 55011 Class A at 10 m FCC Part 15A above 1 GHz Immunity ................................................EN 61326:1997 + A2:2001, Table 1 EMC/EMI ...............................................CE, C-Tick and FCC Part 15 (Class A) Compliant Note For EMC compliance, you must operate this device with shielded cabling. CE Compliance This product meets the essential requirements of applicable European Directives, as amended for CE Marking, as follows: Low-Voltage Directive (safety)..............73/23/EEC Electromagnetic Compatibility Directive (EMC) .....................................89/336/EEC Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory compliance information. To obtain the DoC for this product, click Declarations of Conformity Information at ni.com/hardref.nsf/. Note *323468A-01* 323468A-01 Apr03