Model CYJ Dual Bellows Differential Pressure Liquid Level Gauge Operation Instruction SAMEEE CORPORATION Ltd. Tel : +82-31-492-1691 Web : www.sameee.net Operation Instruction 1. Summarize Model CYJ Dual Bellows Differential Pressure Liquid Level Gauge is mainly used in cryogenic liquid containers (vessel or tank)liquid level height measurement, such as liquid oxygen, liquid nitrogen, liquid argon, liquid carbon dioxide and liquefied natural gas, and can also be used for gas or liquid the instantaneous flow of tests. 2. Structure and Principle The center plate is the key element connected between low pressure bellows assembly and high pressure bellows assembly. The bellows assembly includes a pair or two pairs of spring element. The center axis is connected with high pressure and low pressure bellows. The center axis orthogonally the torque tube within the center plate component, and the whole cavity filled with fluid. The shaft of torque tube through the linkage system is connected with clockworke. The gauge pointer is installed on the ciockworke and shows the dial readings. The out sides of high and low pressure bellows assembly are enclosed within high and low chambers separately. Differential pressure gauge sensing part is based on the displacement balance principle. When the working liquid (or gas) pass through the throttling device producing the differential pressure comefrom high pressure valve and low pressure valve enter the sensing part of the high-pressure chamber and low-pressure chamber, due to the pressure differential drive the bellows spring to produce displacement, lateral movement driven by the central shaft of torque tube assembly can rotate the mandrel, the mandrel of rotation torque tube rotate 0~9°; amplified through clockwork led gauge pointer rotate 0~270°, the differential pressure measured by sensing part shows in mm water column(mm WC) and kilopascals ( kPa ) . The temperature compensation device is connected to the outer end of the high-pressure bellows, so that within the operating temperature range, the device makes differential pressure gauge readings minimal drifting. The differential pressure gauge only as a level indicator, can not serve as the basis for metering and charging. 3. Technical specification 3.1 Accuracy class: 2.5 3.2 Safe working pressure:3.3MPa (≤4MPa) 3.3 Measuring range: 0~15kPa(0~1500 ㎜ WC) 0~25kPa(0~2500 ㎜ WC) 0~50kPa(0~5000 ㎜ WC) 0~100kPa(0~10000 ㎜ WC) 0~150kPa(0~15000 ㎜ WC) 0~250kPa(0~25000 ㎜ WC) 0~20kPa(0~2000 ㎜ WC) 0~30kPa(0~3000 ㎜ WC) 0~75kPa(0~7500 ㎜ WC) 0~125kPa(0~12500 ㎜ WC) 0~200kPa(0~20000 ㎜ WC) 3.4 Work environment (1)Temperature: -40℃~+70℃ (2)Relative humidity: ≤95% (3)Atmospheric pressure: 86KPa~106KPa 3.5 Dimensions: φ190×135 in mm. 3.6 Weight: about 3.2kg 4. Installation and Use 4.1 Gently shake the pressure gauge checking whether the pointer is aligned with the zero position before installation, The connection of up screw and down screw must be perpendicular to the ground when installed. The point should be adjust if it is not aligned with the zero mark, (See section 6 specific methods) 4.2 Installation drawing Figure 1 1. Low pressure valve 2. High pressure valve 4. Differential pressure gauge 3. Balancing valve 5. Tanks There are two kinds of the liquid level meter installation method, according to the needs of the user to choose one of these methods. One is to use the four M6 screw hole (see figure 2), back of the dial gauge, drilling 4 diameter of 7 holes on the mounting plate ,using M6 screw fixation. Another is use of differential pressure level meter with mounting bracket (see figure 2), 4 of diameter 9 hole, 70 x70 hole ditance , using M8 bolts installation fixed to the bracket of tank. 4.3 Making sure the surface to be perpendicular to the ground, high and low pressure pipe joints must not be reversed, pipe fitting thread size is M18X1.5, be careful when installing, do not crash and banging. 4.4 When use to prevent damage to the differential pressure gauge, strict compliance with the following steps: When the high and low pressure valves and balancing valves are closed, open the table should be open balancing valve, then slowly open the high pressure valve, low pressure valve, and finally close the balancing valve. Then the reading is the liquid leave of the tank. IF remove the gauge the first to open balancing valve, then close the high pressure valve, low pressure valve, and finally close the balancing valve. 5. Maintenance and repair Regularly check high and low pressure valves and joints for leaks. If you need remove and repair, must notify the manufacturer or by a professional repair. 6. Fault and Approach Fault 1. When the different pressure gauge pointer is stationary, the pointer offset 1-3 grid out of zero. 2. Storage tank liquid level changes, pointer does not move 3. Close balancing valves, differential pressure gauge is in working condition, the pointer stops at full scale or down to zero 4. pointer shaking. Solution (1) Open the cover; (2) Hold the pointer ends with one’s fingers. (3) With a T-shaped screwdriver to gently rotating pointer core; (4) Loosen the finger; (5) Repeat the above steps until the pointer back to zero. (Note: Pointer back to zero can still maintain the accuracy of 2.5; gauge should be installed perpendicular to the ground.) Return the instrument manufacturer checks. Check the pipeline of tank without clogging; checking the switching action of low pressure valve, high pressure valve, balance valve meets the practice. Check the high pressure valve, low pressure valve, balance valve connection between whether there is leakage. 7. Acceptance and Storage 7.1 After receiving the product packing, user should first check and accept accessories according to the packing list. 7.2 Remove the gauge check if any question with assembly and accessories, notify the company immediately if you have any questions. 7.3 the level gauge should be stored in 0 ~ 40 ℃ environment temperature and indoor relative humidity is no more than 80%, no corrosive gases in the surrounding air. Avoid tension to spring to avoid deformation. If the storage period last more than half a year, it must be check before use again.