LC-120-L Low Voltage Controller
120 VAC 50/60 HZ Low Voltage Controller
For Lighting Fixture Applications
Low Voltage Controllers enable users to safely operate the highvoltage lighting found in hospital and nursing home rooms with
their pillow speaker or bed communication side rail. They’re tested
and approved for use with all Nurse Call systems that use a switch
with normally-open momentary dry contacts for auxiliaries.
. For use with fluorescent, LED, and other lamp types
. Can be used to operate high-voltage circuits such as lights,
electric doors, and drapes
. Can be controlled from a pillow speaker, bed side rail, or wall
switch with normally-open momentary dry contacts
. Controls two loads independently, or two loads sequentially
. Safe for use near medical equipment that is sensitive to
electromagnetic noise
. UL/C-UL Listed to UL 508
. Compliant with FCC, Title 47 CFR 15 – Class B
. RoHS compliant
Input voltage
120 V AC, 6 A, Gen Use
1/8 hp (2 FLA), 120 V AC, hp
1 A, 120 VAC, Tungsten
1 A, 120 VAC, Standard Ballast
Output voltage
120 VAC, 60 Hz
Output devices
Two 6 amp relays driven by low voltage
microcontroller ESD protected.
Lamp
T5 or T8 lamp of 54 watts or less, one lamp per load
Switching circuits
operation voltage
5 VDC @ 0.5 mA
Power Supply
Classification
Isolation from line voltage via a Class II transformer
Operating Temp. Range
40º to 105º F (5º to 40º C)
External wires
White/Black: 16 AWG stranded
All other wires: 18 AWG stranded
Mounting
Mounts next to a lighting ballast in
lighting applications
Housing
.058" Galvanized Zinc-Plated Steel
business is built on, and it’s the reason that hospitals across the
Dimensions
3.76" (L) x 1.8" (W) x 1.38" (H)
country trust Curbell for a variety of products and solutions for
Warranty
Three Years
the patient room, including pillow speakers and other nurse call
Regulatory Listings and
Compliance
• UL/C-UL Listed to UL 508
• Compliant with FCC, Title 47 CFR 15 – Class B
• RoHS compliant
devices, bed replacement parts, fall management devices, and
About Curbell
Curbell has been integrating patient room communication
products since 1960. In that time, we’ve become a trusted partner
for every major nurse call system manufacturer and an industry
leader in patient room connectivity. Our experience is what our
patient monitoring products.
©2013 Curbell Medical Products, Inc.
MAP1274B
www.curbellmedical.com • 1-800-235-7500
Product Design and Application
Figure 1: For one SPST momentary switch, sequential
The LC-120-L Low Voltage Controller is designed to mount inside a
lighting fixture. All connections should be made within the lighting
fixture and capped with wire nuts. A qualified electrician should
perform the installation.
READ
LIGHT
L1
ROOM
LIGHT
L2
GRN
BLK
WHT
(INDIRECT LIGHTING)
BLK
(DIRECT LIGHTING)
WHT
GRN
BLK
WHT
GRN
PINK
YEL
BLK
AC LINE
VOLTAGE
50/60 HZ
HIGH VOLTAGE SIDE
S1
PILLOW SPEAKER OR
BED SIDE RAIL SWITCH
Figure 2: For one or two individually controlled SPST
momentary switches, combination
READ
LIGHT
L1
ROOM
LIGHT
L2
BLK
Momentary Switch States - Sequential Controller
(INDIRECT LIGHTING)
GRN
(DIRECT LIGHTING)
WHT
On the other side of the control housing is the high voltage side,
which has two groups of wires. One group is the AC voltage in. It
consists of a black wire (hot), and a white wire (neutral). The other
group of wires is the AC voltage out to the load (lights). It consists
of a yellow wire (hot, load 1), a pink wire (hot, load 2), and a green
wire (earth ground). These connections go directly to the room
lights, or other devices being controlled, and earth ground.
TO
REMOTE
WALL
SWITCH
BLK
The three wires on the low voltage side of the unit are designated
as red, blue, and brown. When wiring as a single switch, sequential
controller, you must connect the red and blue wires together to one
side of the switch. The brown wire is the other side of the switch.
This wiring can be connected to a receptacle, or directly to a pillow
speaker if it is hard wired to the nurse call system (Figure 1). When
wiring as a two switch combination control, the brown wire is
common for the two switches. Blue is S1 and red is S2 (Figure 2).
LOW VOLTAGE SIDE
BRN
Wiring
RED
BLUE
LOW VOLTAGE CONTROLLER
WHT
Switches will reset to off position if power is interrupted more than
one second.
3rd Switch
On
On
4th Switch
Off
Off
BLK
On
WHT
Off
Off
AC LINE
VOLTAGE
50/60 HZ
HIGH VOLTAGE SIDE
LOW VOLTAGE CONTROLLER
BRN
On
2nd Switch
GRN
1st Switch
PINK
Off
BLUE
Off
After power up
YEL
Load 2
BLK
RED
Load 1
GRN
LOW VOLTAGE SIDE
READ LIGHT
(DIRECT LIGHTING)
S1
S2
COMMON
ROOM LIGHT
(INDIRECT LIGHTING)
TO
REMOTE
WALL
SWITCH
PILLOW SPEAKER OR
BED SIDE RAIL SWITCH
These illustrations are not intended for installation purposes.