1 Cavity Ionization Chambers: • Ion chambers in general vary a great deal: Volumes vary from ~0.005 cc to ~800 cc. Currents vary from ~1 pA to ~1 µA. • Thimble-Type Chambers: a fundamental tool for medical dosimetry physics. • Cylindrical or spherical cavities in which an electric field is applied between a conductor coated on the surface of the wall and a collector electrode along the center of the cavity: High voltage (~300v) power supply Collection volume V A Wall Electrometer Guard electrode Insulator • The voltage on the inner electrodes is high! It usually uses a triax cable to include a guard. Often the wall is C-552 or graphite. • The current collected is very small (calculation later). • The guard electrode grounds the current that leaks between the wall and collector electrodes. The leak mainly occurs along the surface of the insulator separating the electrodes but the cable joining the ion chamber to the electrometer can leak as well – especially if the triax cable is kinked and bent a lot which always seems to happen in a busy hospital. • The guard defines the collection volume of the ion chamber. • The chamber wall, source of charged particles, is often (not always) made thick enough to establish CPE or TCPE that is fully characteristic of the photon interactions taking place in the wall material or thin enough to not perturb the fluence of charged particles if charged particle dosimetry is of interest. Lecture 25 MP 501 Kissick 2016 2 • Condenser-Type Chambers: • A condenser chamber is not connected to an electrometer. Instead, a high voltage is placed between two electrodes as shown below, from Attix, page 310: • The condenser chamber can be thought of as a cavity in parallel with a capacitor. Switch Collection volume V • Ccapacitor Cchamber The combined capacitance of the chamber and the capacitor is C = Ccap + Cchamb . And the initial charging voltage is P1. • After irradiation, one gets the remaining voltage, P2. • The charge collected from the chamber, ∆Q , is given by ∆Q = Q1 − Q2 = C ( P1 − P2 ) Lecture 25 MP 501 Kissick 2016 3 • Since, the voltage across the chamber varies during the irradiation, there must be recombination of some ions before they can reach the electrodes. This means that some ions produced will not be collected – a correction for this is discussed later. -- The recombination depends on voltage, so other chambers make efforts to keep the voltage constant. • Parallel Plate Chambers: • The voltage across the plates is kept constant. • One or both of the plates is thin enough, and conducting, to allow minimal attenuation or scattering of incident electrons or low energy photons. • The plate separation can be very small and also can be variable (That is what an extrapolation chamber is). In this way, one can get close to a surface dose. Collection volume A V Collector electrode • Guard electrode If the collector electrode is too thick, then extra electrons can be knocked out by photons, primary charged particles, or δ-rays. All of which will lead to an increase in positive charge. If charged positive originally, then it will appear as though fewer ions were created. If originally charged negative, then, it will appear as though more ions are created. primary e- Collector electrode primary γ e- e- Lecture 25 MP 501 Kissick 2016 4 • This ‘polarity’ effect can be corrected for by taking measurements with both polarities, and averaging the results. • All chambers should have triaxial cables to guard against current leakage so that it is not accidentally included in the measurement. • Charge and Current Measurements: • Important – the charge collected and current measured are very small ! • Example: Dose to air in the cavity is 1 Gy and the volume of that air at 22oC, and 1 atm is 1 cc, then the amount of charge of one sign is found by the following: Dair = Therefore, Q= Q M W e air Dair M D ρ V = air air (W / e) air (W / e) air or, Q= • (1J / kg)(1.29 ×10 −3 g / cc)(1cc)(10 −3 kg / g) = 3.80 ×10 −8 C (33.97J / C) This is small. In radiotherapy, a dose like this can take only 30s of continuous beam on time, and so the current would be: i= ∆Q 3.80 x10 −8 C = = 1.27nA ∆t 30 s • A typical ammeter cannot measure this small a current, and is also the reason why you should treat your triax cable with care! A typical DMM can sense ~2 mA. A high end one can sense ~20 µA with 107 Ω input impedance. • An electrometer is used. It is a very high impendence voltmeter that can be used to measure current or voltage on a calibrated capacitor on which this collected charge is accumulated. An electrometer typically has 1014 Ω input impedance. They can cost between 2 to 10 thousand dollars. Lecture 25 MP 501 Kissick 2016 5 • Charge Measurement Specifics: • Modern electrometers use operational-amplifiers (op-amps) to amplify the voltage. The simplest possible configuration for an electrometer would be the following: op-amp + Pi V P0=GPi Where G is the gain of the op-amp. The above does not show the feedback on the op-amp or the ion chamber’s and its cable’s inherent capacitance, Ci. • Consider this circuit (not the most typical!): (ion chamber extra ground not shown) For charge measurements* Ri Ion chamber C - + Q or dQ/dt - * Replace with “ • S Ci Pi R (megaohm) + G V P0 “ for current measurements The negative charge Q flows from the ion chamber and responds to a negative potential, Pi. The op-amp produces an output voltage, P0 given by: P0 = GPi Lecture 25 MP 501 Kissick 2016 6 • The total potential across the capacitor, C, is P0 − (− Pi ) , and it holds a charge, C ( P0 + Pi ) . This capacitor is in parallel with the inherent capacitance, Ci (the chamber and the cable). Therefore the ion chamber charge is as follows: Q = C (P0 + Pi ) + Ci Pi = CP0 + (C + Ci )Pi = C (GPi ) + CPi + Ci Pi We can assume a large gain approximation: CG >> Ci and G +1 ≈ G : Q = CGPi = CP0 (see Attix, page 322 for typical values of G and Ci) • Smaller capacitors, C, allow for smaller charges to be measured: the output voltage, P0, is inversely proportional to C: P0 ≅ Q C • Current Measurement Specifics: • One needs to replace the capacitor, C, above with a very large, megaohm resistor, R. • The current, I, that passes through both R and the inherent Ri is as follows, by using P0 = GPi : i ( R + Ri ) = P0 + Pi = (G + 1) Pi We can assume small inherent resistance and large gain approximations: R >> Ri and G +1 ≈ G : i≅ • GPi P0 = R R Therefore, large resistors, R, allow for smaller currents to be measured. Lecture 25 MP 501 Kissick 2016 7 • A typical Triax mode circuit that you should memorize! Note that the circuit inside the dotted box floats at high voltage – careful when you open the device! • There are other ways to connect things together, such this coax mode circuit with an external high voltage supply: Coax Mode (external HV supply): + C R (megaohm) Ion chamber -dQ/dt +12v + + - -12v Vout + ground common ground • Note that with the resistor for feedback to the inverting input makes the op amp circuit look like a high pass filter (see appendix for more info). Op amps act to force the inverting input (-) have equal voltage to the non-inverting input (+). • Next lecture starts to explore the dynamics of the gas and ionization inside the chamber. Lecture 25 MP 501 Kissick 2016