Bihar Veterinary College, Patna List of equipments along with

advertisement
 Bihar Veterinary College, Patna
List of equipments along with specifications and quantity
as per VCI norms
Sl. No.
Name of Articles
1
Refrigerator
Specifications





Vertical type,
310 litre cap.,
double door,
Frost free with power cut evercool and new smart inverter
compressor,
with automatic voltage stabilizer
2
Quartz Double
distillation unit (Vertical)
 - Borosil make
 Output cap.- 1.5 litre/hr
230-250 volt single phase 3 kw
3
Electronic balance
Single pan,
Capacity:
(1) 1mg-100g
(2) 50 g – 5000 g
4
Muffle furnace














5
6
7
8
Suction Pump
Digestion set
Micro Kjeldhal
apparatus (Manual)
Soxhlet apparatus set








Inner dimension: 18"X9"X9"
Outer dimension: 28"X20"X20"
Outside Body made of G.I. Sheet powder coated
Inner muffle of Ceramic with high grade ceramic instillation
Door insulation of HFK briks
Heating element are made from kanthal A1 Wire
A Uniform distribution through all 4 sides with kentahl wire
Max. Working temp upto 1160 C
Equipped with digital temp. Controller cum indicator fitted on top
side of the unit with indicating lamps & switches
A special type of thermocouple fuse for safety to avoid over
heating
Power supply 230 Volts, 50 H2
Digital Timer
Solid state fuses for extra safety
PID Controller with printer interface
Max. Vacuum: 650mmHg
Max. Flow Rate: 20L/min (100V,110V/60Hz), 17L/min (220V/50Hz)
Motor Rotation: 1700 RPM (110V,110V/60Hz), 1450RPM (220V/50Hz)
Horse Power: 1/8 HP
Pole: 4P
Gross Wt: 11.24lbs/5.1kg
Port Thread: 5/16"
Noise Level: ~50dB
Complete electrical setup with auto cut thermostat
Std. glass make with complete accessories
Borosil glass make with oilflask, condenser, soxhlet with complete
1 9
Near Infra Red
Spectroscopy (NIRS)























10
Flame photometer
11
Spectronic 20























Weight: 31 lbs/14.06 kg
Dimensions: (Height) 13.93 in/35.4cm, (Width) 15.52 in/39.42 cm,
(Depth) 14.25 in/36.19 cm.
Power Requirements: 120/220 VAC (50-60Hz)
Voltage Range: 90 to 260 VAC (50 - 60 Hz).
Sample results Storage: 3000 Readings
Caliberation Storage: Flash Memory
Display: Touch Screen Display
Optical Filters: 4 to 10 ( 10 Filters Standard)
Serial Computer Interface: RS-232-C and USB
Printer: Optional
Sample Integration Technique: Rotating cup
Filter Band pass: Approximately 10nm
Light Source: Long life Tungsten Halogen
ETL Mark: UL61010-1
Humidity: 5% RH to 95% RH to 27.7° C noncondensing decreasing
linearly to 60% RH at 35° C decreasing linearly to 50% at 40° C.
Altitude: Up to 2000 meters
Operating Temperature: 10 - 40° C (machine accuracy specification
10° C - 35° C).
Storage Temperature: -20° C to +60° C
Calibration Storage: Flash memory - 500 CALIBERATIONS
Typical Analysis Time: 10 seconds
Detector: Thermoelectrically cooled lead sulfide (Pbs)
Reference Measurement: Through entire optical system, against
ceramic disk
Gain Setting: Automatic for reference gain, calibration file controls
gain or sample
Diagnostics: Diagnostic screens for troubleshooting indoor use only
Calibration curve fit routines first order and second order
Data displayed in concentration units i.e. ppm & meq
Automatic filter selection
Auto calibration feature
Printer attachment
PC connecting compatibility
Sample data storage more than 500 results
Sensitivity: General Mode: Na:0.5 ppm, K:0.5 ppm, Li:0.5 ppm, Ca:15 ppm
Filters: Na & K Supplied (Ca & Li are optional at extra cost)
Resolution: 0.1 ppm/meq
Display: 20 x 4 Alphanumeric backlit LCD display
Flame System: LPG & dry oil free air
Detector: Silicon Photodiode
Calibration: Upto 5-Point Calibration with curve fitting software
Gas Control:Adjustable with knobs
Ignition System: Auto Ignition
Automiser: Axial flow type
Power: 230 V + 10% AC, 50 Hz
With computer and complete accessories
Bandwidth: 20 nm
Wavelength range: 340 to 950 nm
Wavelength accuracy :±2.5 nm near 525 nm
Photometric range: Transmittance: 0 to 100.0% T
2 




12
Bomb Calorimeter
13
Gas liquid
Chromatography unit
14
Fluorescent microscope











Absorbance: 0 to 1.95 A
Concentration: 0 to 1999 C
Detector: Solid-state silicon
Output : 0 to 0.1 VDC; RS-232
Stray light : <0.5%T when appropriate stray light filter is used
Source lamp: Tungsten
Display : 3-1/3-digit LED
With power backup stabilizer
Automatic
Adiabatic type
Gallenkamp 301 series
With computer and complete accessories
Automatic power stabilizer
NUCON 5700 Gas Chromatograph make
With Glass Lined Injector,
Quartz Jet TCD & FID detector and Glass Column,
Methaniser,
Head Space Attachment and Manual / Auto Gas Sampling Valve
With all accessories and power backup
Frame : Rigid and vibrations free frame
Illuminator : Built-in-Koehler illumination for transmitted light 12V100W
halogen bulb (pre cantered)with motion sensor detects when an operator leaves
and automatically
turns off the light for long life of bulbs, light preset switch, built in filters
Observation tube : Wide field trinocular observation head with field no. 22-25
mm
Eyepieces : Wide field eyepieces with field No. 22-25 mm
Stage : Rectangular Mechanical Stage, Ceramic-coated coaxial stage
Focus : Vertical stage movement 25mm stage stroke with coarse adjustment
stopper. Torque
adjustment for coarse adjustment knobs. Stage mounting position variable, high
sensitive fine focusing knob with minimum adjustment graduation should be1µm.
Objectives : Plan Achromat Objectives 4X/0.1
Fluorescence/Semi Apochromat Phase Contrast Objectives 10X/0.3
Fluorescence /Semi Apochromat objective 20X/0.5, (spring)
Fluorescence /Universal Plan Semi Apochromat Phase Contrast Objective
40X/0.75, (spring)
Universal Plan Super Apochromat Objective 60X/1.35, (spring, oil)
Universal Plan Super Apochromat Objective 100X/1.4, (spring, oil)
Fluorescence Unit : Fluorescence with alignment free 130W mercury/LED
illumination .Fluorescence axis should have six to eight cube filter turret with
Zero Pixel Shift technology. Single narrow band Filters for UV, Blue & Green
range excitation and one triple band filter for UV/Blue/Green excitation Cooled
Digital Monochrome
Camera : Monochrome camera, 1.4 or more Mega pixel for High Resolution
Images,
Progressive scan interline CCD, Pixel size 6.45um x 6.45um,1280 - 1392 x 1020 1040 pixels,10fps frame rate, Mono, 12bit, includes camera head,1394 Fire
wire digital interface, power supply. High-Speed Readout < Previewing &
focusing in real time 10fps full resolution@12bits(165fps maximum with
binning and ROI functions), Tight synchronization with flash lamps automated
filters, shutter & microscope stages. Linear Full Well 18,000e-(22,000e-with 2x2
binning), Dark current 1.5e /p/s .Minimizing thermal noise during low light, long
exposure imaging, 2/3 inch optical format. 2/3 inch C mount adapter.
System should be capable of handling bright field and fluorescence object
Magnification changer: Having option of 1X, 1.25X, 1.6X and 2X.
3 15
Serologic water bath
16
pH meter digital
17
Hot air oven
18
Incubator
19
Micropipette
20
ELISA reader
Double walled construction, Inner made of SS 304Q mirror finish Exterior made
of GI sheets/Stainless Steel, Heating is by specially designed water heaters/,
Temperature controlled by thermostat Temperature range: 37C - 60C, Control
Accuracy: +_2C
Bath size 18"X10"X6"
Power supply 230 V .A.C 50 HZ Single phase
pH Range :0-14pH
Resolution : 0.01 pH
Temperature Range : 0.0 to 100ºC (Manual Compensation)
Display : 31/2 Digit LED Display
Power Supply : 230VAC ±10%, 50 Hz
Calibration check facility & Calibration Error indication
 Ovens must be triple walled, Inner chamber made of either M.S. or
S.S.304, Gap between the two walled should be filled with ceramic
insulation to avoid heat of loses with insulation of 150 mm
 Heating element are made of special type air heater of S.S. ‘U’ Type.
Should provide with three heating elements on three sides of the
equipment for uniform temperature on all shelves.
 Should be provided with air circulating fan. Should provide with a
variable microprocessor based digital temperature
controller with digital display and thermometer should be provided separate.
 Should have a minimum chamber size of (L*B*H) 24”x24”x36” with 2
stainless steel trays with holes.
 Should provide with air ventilations.
 Should be operated on 230V, 50Hz single phase AC supply and having
temperature ranging between 50-200ºC



Should be operated on 230V,50Hz single phase AC supply, and having
temperature ranging from ambient to 60ºC
Should be double walled with stainless steel inner chamber having a
minimum of two inner stainless steel shelves with holes and powder
coated outer surface.
 Inner chamber should be fabricated with ribs for adjusting shelves to
convenient height.
 Should have a minimum of chamber size of (L*B*H) of
450*450*450mm.
 Should be provided with three side heating elements.
 Should have air circulating fan (Which can be turn ON/OFF on demand)
for uniform temperature on all shelves.
 Should have double door with acrylic transparent door.
 Should provide with a microprocessor based digital temperature
controller with digital display.
 Should have synthetic rubber gasket at the door.
Variable volume 2-20 ul (Inaccuracy(µL) ±0,20 & Imprecision (µL) 0,08), 550ul(Inaccuracy(µL) ±0,30 & Imprecision (µL) 0,15), 10-100ul (Inaccuracy(µL)
±0,80 & Imprecision (µL) 020)
 Should have 96 wells and reading capability of 1 to 96 wells
individually.
 Should have a linear measurement range of 0 to 3.000 Abs.
 Should have wavelength range from 400 to 750nm.
 Should have a photometric accuracy of ±3% or better.
 Should have a resolution of 0.001Abs.
 Should have variable speed plate shaking capability.
 Should have easy access 8 position filter wheel
 Machine should be supplied with 4 standard filters.
 Should have automatic filter selection.
 Should have automatic calibration before each reading.
4 



21
Thermocycler
22
Gel documentation
system
Should have at least 6 second reading speed.
Should have facility for storage of calibration curves.
Capable of doing multi standard tests and controls.
Should have different types of blanking facility like air wise and well
wise.
 Should be capable of reading U, V and flat type wells
 Should be capable of reading 8 or 12 well strip plates.
 Should use halogen light source and two spare bulbs should be provided.
 Should have internal thermal printer and 5 rolls of thermal should be
supplied alongwith the unit.
 Should have external printer connectivity option.
 Should work with input 200 to 240Vac 50 Hz supply.
Technical Specifications for Thermal Cycler
General Specifications:
 Gradient thermal cycler with peltier heating and cooling system.
 Sample block should have 96x 0.2 ml or 96x 0.2 ml well enabled to run
fast chemistry.

Block should have option of dual block of 2 X 48 X .2 ml and
independently control for both blocks or the block should be able to run
PCR for six different annealing temperatures in the same run if it has six
zones.

Maximum ramp rate should be 5˚C/sec and average ramp rate of 3.3˚C/
sec.

Temperature control mode should be calculated and block independent.
 Display: 8 inch LCD display and touch screen
Reaction module specification:
• Temperature range should be 0-100˚C.
• Temperature accuracy of ± 0.2˚Cof programmed target 0-99.0˚C.
• Average temperature rate should have 3.3˚C/ sec.
• Temperature uniformity should be ± 0.4 ˚C to 5.0˚C well to well with in 10-20
sec of arrival at 90˚C and have 6 thermo electrical modules.
• Should have gradient range of 30-100˚C and have dynamic ramping for
gradient.
• Temperature differential range should be 1-24˚C.
• Memory should be 800-1,000 typical programs onboard, unlimited with USB
flash drive expansion.
• Security features should be password protected folders, optional log in, and
secured mode of high regulated environment.
• Instant incubation
• The system may support PCR volumes ranging from 10-30ul
• Automatic options for Step based graphical and text based programming.
• The software should have exportable run logs and system error logs.
• Input power range should be 400W and frequency 50-60, single phase.
• Online 5KVA UPS
• Warranty/Guaranty minimum three years with first one year free spare parts
replacement options
The system should be image fluorescent DNA, RNA, and protein gels,
colorimetric gels and blots and colony arrays.
• High resolution CCD camera: 1.4 mega pixel resolution with 1,360 x1,024 pixel
array.
• Data acquisitions: 12 bit and 4,096 gray level and or Pixel size: 7.4µm x 7.4µm.
• Light tight cabinet to provide increase sensitivity to minimize background noise.
• Motorized control for zoom with numerical feedback and software acquisition
preset integrated into an intuitive and easy to use interface along gel alignment
templates, aperture & lens with broad range filter.
• Gel alignment templates matched to agarose or protein gel tray and ready gel.
• Filter for Fluorescent dyes such as EtBr/SYBR Green/SYBR Safe/Texas
Red/SYPRO orange, coomassie blue, silver stain.
5 • Filter size having 25X26 cm area. Trans illumination area with White light
converter Screen for viewing protein gels or similar features.
• Fire wire connectivity with fast data transfer.
• Minimum 3 position filter slider with amber filter.
• Software for imaging and analyzing 1-D electrophoretic gels, dot blots and
colony counts.
Software should be able to do:
• Quantitate and analyze a variety of data.
• Rapid molecular weight determination with choice of multiple regression
models.
• Band/lane matching analysis with comparative dandogram creation
• Background subtraction correction of gradient gels
• Accurate concentration analysis using sophisticated volume tools, volume box,
volume circle , volume contour, or free hand drawing.
• Local background subtraction for individual bands.
• Colony counting that discriminates colonies and plaques.
• Array tools to analyze and quantitate dot blots, slot blots, and medium density
arrays.
• Annotation tools to add text and lines.
• 3 D viewer for critical analysis of closely spaced bands
• Molecular weight determination
• Volume overlays
• Text and line overlays
• Multiple illumination sources for imaging opaque samples
• Dynamic range should be more than 3 orders of magnitude.
• A Branded work station- desktop computer, 3GB Ram, 160 GB HDD, CD –
DVD Combo Drive, with 17” TFT monitor, windows Compatible for running the
1- D soft ware with suitable LaserJet printer.
23
Refrigerated
micro
centrifuge machine
24
25
Mini submarine gel
electrophoresis unit
Mini dual vertical
electrophoresis unit
26
vortex shaker
27
Orbital Shaker
1. Max RPM/RCF- 17000/27237g with microtube rotor for 1.5
2. Temp. range- -20ºC to +40 ºC with fast cool button
3. Pulse or continuous time control
4. <56DB noice level with imbalance cut off
5. LCD display
6. safety lid lock
lid drop protection with CE mark
unit includes buffer chamber, safety lid with cables, UV transparent tray and one
each of 1.5 thick 8 and 15 well comb
unit include buffer chamber, safety lid with cables, two aluminum notch plate, 10
glass plates, 2 10 well 1 mm thick combs, 4 spacers of 1mm thickness and dual
gel caster
 Heavy metal base with rubber feet
 Choice of continuous and touch mode
 variable speed control with maximum speed 3000 rpm
 one micro tube insert
A. General : Bench top microprocessor controlled Refrigerated Incubator Shaker
B. Temperature Range: Temperature control range from 7°C above ambient to 60°C control
accuracy ±0.1°C.
 Should be able to maintain 37°C during summer months when ambient
temperature is 40-45°C.
C. Shaking Features:
Electronically controlled shaking mechanism provides quiet operation.
Brushless motor is preferable to avoid costly maintenance and noise less.

Type of shaking: Orbital. But having any additional feature is
advantageous.

Speed: 30-300 rpm with control accuracy of ±2rpm

It should have acceleration/deceleration circuit prevent sudden
6 28
Deep Freezer
vertical
with 1KVA voltage
stabilizer
29
Laminar Flow Work
station (Horizontal),
with 2 KVA voltage
stabilizer
starts/stops to eliminate flask closure wetting which would otherwise interfere
with oxygen transfer.

It should have timer of up to 99.9 Hrs.
D. Display
 Large easy to read display screen clearly indicates speed, running time,
alarm condition and temperature character.
 Display is clearly visible from all angles across the room, and in the
dark.
 E Safety Features:
 i There should be over temperature protection.
 ii Preferably there should be a option to mute the audible alarms.
 iii The system should have RS-232 port for Data Logging.
 iv The system should have facility to calibrate the temperature and speed
via keypad.
 v Electronic and mechanical components should be enclosed, protected
from accidental spills.
 vi It should have automatic restart after power interruption with non
volatile memory.
 vii It should have open lid cut out switch stop agitation when chamber is
open.
F. Essential Accessories
 It should come with Universal Platform having capacity to holds
assortment of various size of flask sizes upto 250 ml to 1000 ml. System
should be supplied with 250 ml clamps (5 Nos.), 500 ml clamps (05
Nos.), 1000 ml (02 Nos.)
 (ii) It should be supplied with compatible stabilizer/servo for smooth
operation
 Digital temperature control and LED Door display and systems monitoring
and reporting technology.
 Minimum attainable cabinet temperature: -20°C with an increment of 1oC
 Capacity: Minimum 350 ltrs
 Min 3- 4 multiple freezer compartments with individual insulated doors
 Insulated inner doors
 Epoxy covered SS metallic external case.
 Strong, molded, chemically resistant abs interior.
 Vacuum Insulation Technology
 Temperature thermostat regulator.
 Temperature indicator lamp.
 Audible and visual alarms for high/low temperature, power failure, high/low
voltage, overheated
 condenser with push button alarm reset/stop and adjustable alarm set points
 Push button with quick freeze indicator.
 Acclimatized hermetically sealed cfc free compressor Cascade refrigeration
Technology.
 Automatic defrost with electrical water evaporation system.
 Adjustable shelves.
 Door lock
 Internal Voltage stabilizer.
 Laminar Cabinets Inner Chamber & Outer Chamber made of stainless
steel high efficiency particular air filters, to achieve the air purification
upto 0.3 Microns in Working area.
 Working area 4 ft x 2 ft x 2 ft
 Blower fitted with ¼ HP Motor, with RPM 1200 to 1400.
 Pre-filters made of high grade nylon Net fixed in S.S. frame for first
Stage air purification, through blower system.
 Closed Inner Chamber fitted with HEPA having very accurate
performance rate of air filtration, rated 99.99%, resulting in ceasing all
air bore molecule of particle upto 0.3 micron in working Area of
Laminar Bench.
7 
30
Ligation bath
31
Semi-automatic rotary
microtome
32
Animal Mortuary
chamber
33
Deep freezer Horizontal
34
Triple distillation
apparatus
35
Leica trinocular research
microscope along with
camera and computer set
Working area of Laminar Airflow Cabinets illuminated by fluorescent
light ;cabinets operated at 230V. Single Phase 50Hz. AC Supply.
 Fitted with UV Germicidal lamp for sterilization.
 Fitted with Acrylic Front Door sliding type
 Fitted with Manometer for Measurment of HEPA Filters Choking
system.
 Fitted with Cock for Gas Connection.
 automatic heating or cooling mode selection
 thermoelectric cell
 Temp. Range - + 5 ºC to 60 ºC
 LED display
 electronic solid temperature controller
 on site calibration facility
 block holds 4ml, 1.5 ml
 non corrosive body
 Leica RM2245
 Section thickness range-0.5 to 100µm
 Trimming section range-1 to 100 µm
 Section mode-2 manual modes (conventional full handwheel rotation
and rocking mode)
 -Specimen retraction 5 – 100 µm in 5 µm increments,can be turned off
 -Vertical specimen stroke -70 mm
 -Electric coarse feed 300 µm/s and 900 µm/s
 -Maximum specimen size (L x H x W) 50 x 60 x 40 mm
 -Specimen orientation horizontal: 8°, vertical: 8°
 Nominal supply voltages 100/120/230/240 V AC ± 10%
 -Nominal frequency 50/60 hz
 -Power draw 70 VA
Temperature- up to -20°C
-Exterior surface in chemically treated, antirust coated and duly finished with
powder coated paint
-Temperature controller is mounted on front panel, provided on front top, along
with pilot lamp, main switch etc. Supplied complete with trays, cord and plug.
-Power supply 220 V, single phase, 50 Hz, Ac supply. Supplied with suitable
Voltage Stabilizer.
Size-6X6 ft.
 Temperature Range- -20°C
 Design-Horizontal
 Control Panel- Touch Key Pad w/ LED Display
 Temperature Regulation- Microprocessor Controlled
 Alarm Type- Audible & Visible
 Refrigerant- CFC free refrigerant
 Construction- Double Walled w/ PUF insulation
 Door- Solid Door
 Inner Doors- Single and Double Doors w/ Tempered Glass
 Cabinet Housing- MS w/ Powder Coating (SS optional)
 Internal Chamber- Stainless Steel (316 / 304)
 Power Supply-220 - 240 Volts, 50 Hz. Single Phase
 Output Capacity (Approx.)-2 Litre/hr.
 -Electrical Power Reqd.- 220/240 V, Single Phase, 3.5 KW Heater
 -Cooling Water Consumption-65 lit/hr
 -Quartz distiller, Demountable boiler.
Compact design, Dual safety cutout(optional).
High purity, low conductivity, Pyrogen free distillate.
 4XCE CCD Metallographic Trinocular Microscope with Camera
-Three eyepiece tube- Tilting angle: 30°
 Adjustable range: 55-75mm
 -Eyepiece- Wide view eyepiece 10X (16mm), 12.5X, 16X, 20X, 10X
flat fielding graduation eyepiece (grid value 0.1mm)
8 
36
Micro Slide cabinet(5000
slide
37
Semi-automatic rotary
microtome
38
Cry- microtome
39
Hemo Cue WBC Diff.
System
Split Air Condition for
tissue preparation room
40
41
Leica trinocular research
microscope along with
camera and computer set
-Objective lens- Flat fielding achromatic objective lens10X, 25X, 40X,
100X (oil).
 -Photographic device -1. Image adapting lens, high resolution color
CCD camera static and dynamic capturing card, image analysis software,
EPSON color jet printer
 2. Digital camera, specific adapting lens
 Horizontal Manner Microslide Cabinet - For handling of 75 x 25 mm
slides. Made of mild steel sheet. Well designed for its appearance,
performance and durability. Powder coated paint outer finish outer finish
with smoothly working doors fitted with handle, lock and key.
 Leica RM2245
 -Section thickness range-0.5 to 100µm
 -Trimming section range-1 to 100 µm
 -Section mode- 2 manual modes (conventional full hand wheel rotation
and rocking mode)
 -Specimen retraction 5 – 100 µm in 5 µm increments, can be turned off
 -Vertical specimen stroke -70 mm
 -Electric coarse feed 300 µm/s and 900 µm/s
 -Maximum specimen size (L x H x W) 50 x 60 x 40 mm
 -Specimen orientation horizontal: 8°, vertical: 8°
 Nominal supply voltages 100/120/230/240 V AC ± 10%
 -Nominal frequency 50/60 hz
 -Power draw 70 VA
 Freestanding cryostat with encapsulated, splash-proof microtome.
- Stable, self-contained cryocabinet on casters.
- Cryochamber temperature selection from 0 °C to -30 °C,
adjustable in 1K increments at ambient temperature of 20 °C.
- Cryochamber defrosting- manually or automatically.
- Section thickness selection from outside the cryochamber.
 -Control panel with membrane-protected buttons
- Manufactured in compliance with c-CSA-US and CE standards.
CFC-free.
 -Accessories required to be quoted along with the instruments – 4
Specimen discs- 25 mm, 4 Specimen discs- 30 mm, 1 Section waste tray,
2 Storage shelves.
Analyzer 4-50 c (39-122 F .Micro cuvettes 15-35C (59-95)Single Pack micro
cuvettes must be used within 10 minutes of opening of individual pack.
 Split air conditioner ,1.5 ton
-Auto defrosting
-Energy Saving, Quiet operation
-Sleeping Model, Remote Control
-Recruit sole agent
 -AC Stabilizer
 4XCE CCD Metallographic Trinocular Microscope with Camera
-Three eyepiece tube- Tilting angle: 30°
 Adjustable range: 55-75mm
 -Eyepiece- Wide view eyepiece 10X (16mm), 12.5X, 16X, 20X, 10X
flat fielding graduation eyepiece (grid value 0.1mm)


42
43
-Objective lens- Flat fielding achromatic objective lens10X, 25X, 40X,
100X (oil).
-Photographic device –1. Image adapting lens, high resolution colour
CCD camera static and dynamic capturing card, image analysis software,
EPSON colour jet printer
 2. Digital camera, specific adapting lens
Material: PP ABS .Packaging & Delivery
Packaging Details: 120 pcs/carton
 Horizontal Manner Micro slide Cabinet –For handling of 75 x 25 mm
9 44
Slide box(100 slide)





45
Micro Slide cabinet(5000
slide)





46
Binocular microscope
47
MABW200 veterinary
urine analyzer
48
Automatic tissue
processor
slides. Made of mild steel sheet. Well designed for its appearance,
performance and durability. Powder coated paint outer finish outer finish
with smoothly working doors fitted with handle, lock and key.
Optical System- UIS2 (Universal Infinity-corrected) optical system
Illumination- Built-in transmitted Koehler illuminator 6V30W halogen
bulb 100-120V/220-240Vg 0.85/0.45A 50/60Hz
Revolving nosepiece- Quadruple positions fixed (Inward tilted)
Stage- 188 (W) x 134 (D) mm
Condenser- Abbe condenser, with built-in daylight filter,1.25 (when
immersed with oil)
Aperture iris diaphragm built in.
Principle of determination: by ultra-high degree of cold light reflection.
Scope of application : 8 item/10 item/ 11 item/ 12 item/14 item3.Product
determination
GLU/BIL/KET/SG/PH/BLD/PRO/URO/NIT/LEU/VC/CRE/CAL/MAL
Test speed: 60 copies of test/hour continuous test of 120/h
Programmable 12 stage timing sequence, having 24 hours calibrated disc, Delay action of 24 hours
MODEL MLXi PLUS complete with built-in 6V20W halogen light illuminator,
Quadruple ball bearing inward nosepiece, focusing by co-axial coarse and fine
focusing controls,
High resolution Infinity Corrected Plan Achromatic Objectives 4x, 10x, 40x
(spring), 100x oil (spring) with an inclined Seidentopf Binocular observation tube
rotatable through 360degree fitted with fungus resistant, anti-reflection coated
prism, compensating wide field
paired eyepiece WF10x (F.N.20) and with right hand co-axial low drive
mechanical stage,
sub stage Abbe condenser 1.25 N.A with iris diaphragm focusable with rack &
pinion and with blue Filter.
Plan Infinity Optics, Excellent Field Flatness, F.N.20, Seidentopf Head, hard
coated stage with double slide holder, Easily replaceable lamp, Inward facing
nosepiece,
upper limit stopper for prevent the specimen in high magnification observation,
UniversaPower supply ensures constant voltage & International Build Quality.
LCD high power laboratory biological microscope offers high quality optics at an
affordable price point. The included 9.7" tablet camera offers a touch screen for
capturing images and includes advanced software for making measurements and
adding annotations as well as capturing images.





Objective Lenses: Infinity Corrected DIN Plan Achromat 4x, 10x,
40xR, 100xR (oil immersion) objectives. All microscope objectives are
parfocalled, parcentered and color-coded. Quad objective turret is ballbearing mounted for smooth, precise positioning of objectives, features
textured rubber knurling for easy grip. Mounted in reverse position to
facilitate ease in changing slides.
Eyepieces: Widefield High Eyepoint Focusing 10x eyepieces / FN
20mm. Trinocular Siedentopf viewing head with 30° inclined binocular
eyepiece with interpupillary adjustment 50mm to 75mm. Vertical
trinocular tube for camera integration.
Total Magnification: 40x, 100x, 400x, 1000x.
Specimen Stage: Large 140mm x 135mm stage (5.5" x 5.25"), with
professional quality, built-in, low profile, ball-bearing mounted
mechanical specimen holder with low-position coaxial controls, 78mm
(x) by 50mm (y) movement with graduations of 0.1mm. 1.25 N.A. Abbe
condenser has rack & pinion focusing and iris diaphragm.
Focusing: Coaxial coarse and fine focusing controls, coarse has tension
10 adjustment, fine is graduated. Locking position for stage.
Illumination: 3.3v, 3W LED with variable rheostat control and on/off
switch.110~240V power supply.
 Tablet Camera: Includes c-mount adapter and 9.7" Android tablet with
built-in 5 mega pixel camera.
 Size & Weight: "7.25" W x 14.5" D x 16" H, 16 lbs.
 Warranty: 5 Year warranty, excluding bulb, cord and fuse. Tablet
camera has 1 year warranty.
Tablet Specifications
 9.7" tablet for display and capture.
 5 mega pixel camera, image capture resolution 2592 x 1944.
 2GB internal flash storage.
 SD/SDHC Card Slot
 HDMI Out, USB 2.0, USB Mini B, WiFi, Bluetooth enabled
 Android version 4.2.2 tablet with touch screen.
 Color camera with 1/2.5" Aptina MT9P001 Sensor.
 Capture images (.jpg) and video (.3gp)
 Video frame rate 15fps @ 1280 x 720, 30fps @ 640 x 480
 Live measurement
 Auto exposure/white balance
 Tablet has 1-year warranty

49
MAGNUS BINOCULAR
RESEARCH
MICROSCOPE
(Olympus)



Dimensions: 18 3/4”W x 15 3/4”D x 19”H
Weight: 43 pounds
DR1417-6G DRvwt Imaging System with 300,000 shutter exposure life
expectancy
 Imaging Media: DOUBLE SPEED Rare Earth green Phosphor
Scintillate Plate
 14X17 Image size, 12 mega pixel, 65,536 shades of gray, Full Frame
Transfer 16 Bit
 Includes: Power Cord, Computer Imaging System Interconnection
Connector Cables
X-Ray Generator
 kW Rating........................................40kW
 kVp Range........................................40 to 150kVp, 1kV step
 Time Range......................................0.001 to 10 sec.
 mA Range........................................10 to 900mA
 HV Cable..........................................8 meters
 kVp Accuracy.....................................<(1% + 1kV)
 mAs Accuracy....................................<(3% + 0.01 mAs)
 Generator Mounter............................Provided
 Line Normal Phase..................... 240V AC, 1, 60Hz @ 70 Amps
X-Ray Tube Assembly
 Model / Maker..................................Toshiba
 Target Angle....................................16 Degrees
 Focal Spot Size.................................6 / 1.2 mm
 Max. Tube Voltage............................125kV
 Max. Anode HU.................................140 kHU (100kJ)
X-Ray Collimator
 Control............................................Manual, 30 sec. timer
 Field Shape......................................Rectangular
 Max. Field Size.................................>43 x 43 cm / 100 cm SID
 Max. kV Shield.................................150 kV
 Lamp..............................................Quartz Halogen 160 Lux
Digital Image Receptor
11 


Active Matrix Size..............................14” x 17” (35 x 43 cm)
Image Matrix....................................12 Megapixels
Image Resolution..............................Exceeds FDA’s Human Radiology
Resolution Standard
 Acquisition Time................................10 - 20 Seconds
 Dynamic Range.................................16-bit Grayscale, 65,000 Shades of
Gray
 File Format........................................DICOM
Tube Stand & Float Top Table
 SID .................................................Variable 12” - 40”
 Grid .................................................200 lpi (ratio 10:1)
 Table Top Size...................................30” x 64”
 Tabletop Movement............................4 - way
 Tabletop Stroke (longitudinal)..............12” (5”)
 Tabletop Stroke (lateral)......................8” (4”)
 Dimension.........................................32”(H) x 42”(W) x 40”(D)
 Weight...............................................80kg (176 lb)
 software for canine, equine, feline, ovine, bovine etc applications.
Accessory Instruments-Lead Apron, Lead Divider, Gloves, Goggles, Film badges,
Cassettes-intensifying screen (Rare earth)
50
RICHTER OPTICA
U2LCD DIGITAL LCD
LAB MICROSCOPE
The unit should be for dedicated veterinary purposes for canine, feline, bovine
etc; compact easily transportable mobile C-ARM unit with 9” Image Intensifier
Unit with High Resolution Camera, Dual Flat Panel high resolution monitors of at
least 22” or more, image processing software and storage facility.separateTable
51
Digital X‐ray Machine
for large and small animal.
The equipment should be / have  Suitably designed for veterinary use (mainly for small / pet animals)
 Easy to clean, have good finishing and the material should be rust proof.
 Provision of mounting bracket for mounting two vaporizers.
 Supplied with brand new precision Isoflurane vaporizer of premium
quality which should be temperature, pressure and flow compensated.

Fitted with specific breathing circuits like non re-breathing system,
partial re-breathing system and total re-breathing system.

Circle system fitted with fresh gas inlet, corrugated breathing tubes,
reservoir bag and pressure gauge, unidirectional aspiratory and
expiratory valves; pop off valves as safety device and oxygen flush
system. Soda lime carbon dioxide gas absorber (capable of absorbing
CO2 expired from 2 kg to 100 kg body weight).
Scavenging system.
Fitted with unbreakable valve dome caps.



The CO2 absorber should be efficient in absorption, seal easily and be
leak free seal.

capable of working with vaporizer out of circle and vaporizer in the
circle.

Fresh gas outlet for easy use of a non re-breathing system.

Oxygen flow meters of measuring capacity up to 4 liter per minute or
more

Fitted with low oxygen audio alarm.

supplied along with Bain circuit.

supplied with different sized veterinary specific induction masks having
12 good fitting rubber diaphragms for variable sized small animals
52
C-ARM 
supplied with good quality cuffed silicon endotracheal tubes of sizes 214 FR along with adaptors of suitable sizes 10 each.

supplied with re-breathing bags of different sizes ranging from 0.5 . 1,
2, 3 and 5L 5 each.

compatible for working with IPPV.

The mobile trolley should have five legs for maximum stability (two
with brakes), stainless steel body, one drawer and platform build on the
top for mounting vital signs monitor.

The machine should have minimum 2 years warranty/ guarantee period
and subsequent AMC rates may be quoted.

Bain system and attachement tube 3 piece.
 One set laryngoscope suitable for small animal
The equipment should be / have  Suitably designed for veterinary use in large animal (mainly Bovine,
equine and small ruminants) with inbuilt ventilator
Ventilator should be/have -able to work on 110-220V-tidal volume
adjustable (with facility for momentary switch to flush up to 15ltrs)-electronically
controlled, time cycled, pneumatic powered.-adjustable frequency (minimum
range 0-90 BPM)-adjustable control of inspiratory flow-have varying Inspiration
/expiration ratio (range 1:1 to 1:4.5) with upgradation facility.


Provision of mounting bracket for mounting two vaporizers.
Supplied with brand new precision Isoflurane vaporizer (range 0%-5%) of
premium quality which should be temperature compensated.
Should be supplied with Foal Absorber, two 30 liter bag, one 15 ltr bag,
wye piece, and tubing required.
Absorber- circuit should have pressure gauze, APL valve (allowing to
make the circuit from completely closed to fully open), check valve for
insipratory and expiratory flow. The CO2 absorber should be efficient in
absorption.
Scavenging system.
Supplied with 4 set of cuffed/ without cuff endotracheal tubes of 16,18 and
20 size.
One set laryngoscope suitable for large animal.





53
Small Animal
Anaesthesia Machine







Unit must have medical gradenot less than 15” high resolution TFT LCD
monitor.
Unit should have Standard Parameters in Modular type: ECG, SpO2,
NIBP, Resp, 1 Temp
Quoted unit must have Parameters in Modular type :
12CH EKG, 4Temp, 3IBP, ICO, BIS
Phaseln’s Multi-gas Mainstream (IRMA AX+), Sidestream(ISA AX+,
OR+)
Respironics”s EtCO2 Mainstream (C5), Siderstream (LiFIo)
Unit should have Virtual Screen.
Unit should be fully equipped with software for canine, equine, feline,
ovine, bovine etc applications.
Unit should have 3CH ECG : Full 7 ECG & ST Segment
analysis&12CH ECG : Full 12 ECG & ST Segment analysis.
Unit should have more than 13 Arrhythmia analysis & Pacemaker
detection.
Unit should have more than 160 hoursof tabular and graphic trend data
save facility.
13 
54
Large animal
Anaesthesia Machine
with ventilator
Unit should have facility for Drug Dose Calculation, ECG Recall, Mini
Trend, ICO, Color change,NIBP STAT, NIBP VENOUS STAT, OXYCRG, Patient information Entry.
 Software upgradation must be on-site through SD Card of required
memory.
 Unit must have durable and light Long Li-ion battery operation with
more than; Std-1Pac: 2 / Opt – 2 Pack : 4 hours.
 Unit should have LAN and Wireless LAN Connection.
 Unit should have HRV, APG Detection, Touch Screen in option.
 It must have Built in Printer in option.
 Unit should be provided with the following accessories:
 5 lead ECG cable: 1No.
 ECG electrodes: 50no.
 NIBP CUFF (As require in veterinary application of different
species):2no.
 Temp. Probe: Rectal & skin.
 SPO2 Probe: 01no. (The SpO2 probe must be firm & should have tight
grip for best result in veterinary.)
Ultrasound phacoemulsification
Frequency: 28 kHz, piezo driver
Control: auto-tuning (50 times/s), linear, pulsed, burst, cool micropulse phacoemulsif
Power: 1-100% in steps of 1%, max. deflection 120 IJm
Pulse: pulse frequency 0.5-40 Hz, pulse duration 0.25-1800 ms
Counter: effective time in 0.1 s (relevant to energy delivered)
Memory: 9 surgeons with 3 memory settings each
I/A System
Continuous irrigation function
Pump system: peristaltic, linear control, proportional override
Reflux: pump reverse (150 mmHg)
Tubing System: single use or autoclavable 250 ml waste-container
Memory: 9 surgeons with 2 memory settings each
Performance: 0-50 ml/min, vacuum 0-600 mmHg
Diathermy
Frequency: 500 kHz, pulsed
Power: max. 8 Watt at 500 hm
HF Capsulotomy
Frequency: 500 kHz, pulsed
Power: max. 6.2 Watt at 500 hm, 2 ranges, automatic regulation
Vitrectomy
Speed: 30-1200 cuts/min
Drive: electric, push/pull
14 Power requirements
Electric: 100-240 VAC, 50-60 Hz, power consumption 200 VA
Dimensions & Weight
Dimensions: 220 x 241 (plus 50 mm for handle) x 154 mm (IN pole clamp adds 80 m
Weight: 5.6 kg
Linear pedal: 180 x 250 mm base plate 2.75 kg
Standards
IEC601-1, EN60601-1, UL 544, CE-Mark (CE-class 2B, protection class I)
EMC/RFI: EN 6100-4-1, EN 50011/ EN 55081-1 EN 50082















6-Step magnification system
LED lightsource Lux, life, Kelvin
20W LED coaxial illuminator
10x Wide Field Eyepiece (12.5x available)
250mm Objective Lens (200-400mm also available)
Apochromatic Objective Lens
Integrated Video Camera
Adjustable Eyecups
Rotatable Maneuvering Handles
Counter-balanced pantographic arms design and tension control
adjustable
Mounting Options: Fixed Floor, Wall, Ceiling, Table, Mobile Floor
Integrated Orange and Green Filter
Available foot-long extension arm
Upgradable Assistant Scope






Eirger 45° inclined or 0-180° inclinable binocular head for Dental
surgery
12.5x/18mm High Eye-point Wide-Field eyepieces
Range of Pupillary distance: 50mm to 70mm
Range of the Diopter of Eyepiece: +/- 7D
250mm APO-chromatic objective lens with fine focusing adjustment
knob
Magnification: 6-step, Total magnification indicator 0.33=2.8x,
0.5=4.2x, 0.8=6.9x, 1.22=10.40x, 2=17X, 3=25.6x
Diameter of object field: 8.7mm to 78.3mm
Arms: First big arm, 500mm length can be rotated 360°. Second small
arm 600mm in length with movement +/-300mm can be rotated +/-150°.
Third angled arm can be 3-D movement
Height: 1750mm
Base size: 610mm X 600mm
One 20W LED Illuminator
Maximum Object Illumination: 50,000 Lux Min.
Built-in Orange filter and Green filter changed by a knob.
Power supply: 220V/50Hz or 110V/60Hz


Pupillary distance adjustable: 50mm to 70mm
The device activates with the push of a button and results are displayed







15 clearly on the easy-to-read LCD screen.
micro strain gage technology and a 1.0 mm transducer tip.
microprocessor technology produces objective IOP measurements that
are less influenced by operator bias than Goldmann tonometry
 Wireless binocular indirect with all accessories
P D Range: 48 - 76mm
 High Speed Veterinary Dental Air Unit. This all-in-one veterinarian
dental cart, performs all necessary procedures from cutting-polishingscaling.
Features:
 400,000 RPM high speed drill
 20,000 RPM low speed handpiece
 Built in 25K ultrasonic scaler
 Auto-tune circuitry
 Air water syringe
 Includes oil-less compressor
 H-Frame mobil stand
 Independent water controls
 Built in water reservoir
 Switching power supply
 Automatic handpiece activators

Fiber Optics System with high speed handpiece
 Table and all accessory instumentscompatible with machine 5 set.
 Shortwave cum pulse diathermy unit-1000w trolleymodel  Major tense (pain reliever)- 4 channel digital  Mini tense (pain reliever)- double channel  Interferntial therapy unit with stimulator(lcd  Ultrasonic stimulator/therapy- micro controlled based (lcd)  Infra red combined with uvr with timer Dorsal/Lateral Scissor Base Equine Table
Load Rating-2,500 lbs. (1,134 kg)
lateral surface area:
46” wide x 84” long (117cm x 213cm)


55
Veterinary Dental Air
Unit
56
Physiotherapy
instruments
57
Equine surgery Table
dorsal surface area:
center bed - 18” wide x 84” long
(46cm x 213cm)
58
Laparoscope for small
and large animals
Lower Height -15-1/2” (39cm)
Raised Height- 48” (122 cm)
2-way tilt (in either direction) by retraction of one pair of casters.
Endovision camera – 3 CCD video camera with PAL/ NTSC color system
consisting of
o Camera controller
o 3 CCD camera head with C mount adapter
o BNC video Cable 3 meters
o Shielded S- VHS cable – 2 meters
o Power cable
o Camera coupler
 CO2 insufflator capable of insufflation upto 20 liters per minute with
following accessories
 Insufflation tube2.5 meters long – 2
 Hygiene filter pack
 Power cable
 Halogen light source – 250 watt twin lamp with color correction and heat
dissipation.
16 






59
Milking machine for cow
and buffaloes
60
Chaff Cutter HC-801(Him
co)
61
Fodder
harvester(Harvesting of
fodder crops)
62
Trolley ( Double wheel
barrow)
Telescope – 10 mm, distortion free, universal ocular 0 degrees
Telescope – 10 mm, distortion free, universal ocular 30 degrees
Telescope – 5 mm, distortion free, universal ocular 0 degrees
Telescope – 5, distortion free, universal ocular 30 degrees
Fibreoptic light cable – 2.5 meters, fiber bundle diameter >5 mm. - 2
Veress Needle 20mm working length.
Trocar sleeve with magnetic ball valve, metal sleeve, straight distal
tip.10 mm capacity, 100 mm working ength.
 Trocar sleeve with magnetic ball valve, metal sleeve, straight distal tip.5
mm capacity, 100 mm working length.
 Trocar with pyramidal tip, capacity 5 mm,, working length 100 mm
 Trocar with pyramidal tip, capacity 10 mm,, working length 100 mm
 Hand Instruments – all swiveling type, doble action jaws, tungsten
carbide jaw inserts in scissors, forceps and needle holders, with
monopolar connector pin, and luer lock for cleaning.
 5 mm curved grasping and dissecting forceps
 Maryland dissector 5 mm
 Atraumatic grasping forceps 5mm both jaws opening
 Modular scissors 5mm, curved both blades opening,
Metzenbaum
 Universal grasping forceps 5mm both jaws opening
 Hook electrode diameter 5 mm, length >300mm
 Needle holder with tungsten carbide jaw inserts, 5 mm
 Micro hook scissors – 3 mm 20 cm rotating with
monopolar pin and luer lock.
 ALAIN-GROUSSEAU Pylorotome, distendable, size
3 mm, length 20 cm
 Palpation probe with cm markings 2mm, length 20 cm.
 Monopolar connecting cable . 2 meters long
 Suction and irrigation tube with stop valves
 Clip applicator 10 mm length > 300 mm
 Reducer adapter for 10 mm to 5 mm and 3 mm, 12.5 mm to 5 and 3 mm
and 7 mm to 5 mm and 3 mm.
 21” high definition flat screen monitor, aspect ratio 16:10, color system
PAL/NTSC, with power cable, mains cord, capable of being mounted on
stand, OT light arm or wall.
 Cabling for Video signals from the CCD endovision camera to be
transmitted to recording computer and DVD recorder.
 Separate flexible endoscope for equine, bovine,dogcystoscope,
rectoscope, thoracoscope compatible with machine
Specification:-for milking of two animals
At a time ,trolley milking system with two SS bucket with recirculation oil
lubricant vacuum pump 220 L/minute
Suitable for 20 to 50 cows
Specification:Cut all types grasses & straw
*cuts 6to7 tons of green fodder/hr. in 25 minutes
*cuts 1 to 1.5 tons of dry grass & hay/hr
Specification:Harvesting & windrowing of fodder crops
*field capacity-1 acre/hr
*engine:- 10.2 hp diesel air reverse
*gear:-4 forward& 1 reverse
*width &cut:- 4 feet
Bearinf-4 nos.
*Tyre:-Olympic rubber tyre
*Rod and Ring
17 63
64
Power tiller
Deep Freezer
Horizontal (-70°C)
with 1KVA voltage
stabilizer
Capacity : 383 ltrs
65
Automatic pipette
Washer
66
Immuno-electrophoresis
apparatus
67
Phase contrast
microscope
68
McIntosh Jar
69
Water bath with shaker
70 71 72 73 74 75 Chick Feeder Chick drinker -2 litter
capacity Jumbo chick drinker -4
litter capacity (plastic
made) Grower drinker 7 litter
capacity (plastic made) Jumbo automatic bell
drinker Chick & grower feeder – 2,
8, 10 kg Dimensions (w x d x h) : 28 x 61 x 35 inches
Environmentally friendly nature R refrigerant
Digital controller
Built-in internal and external alarm
i.
ii.
iii.
iv.
v.
vi.




Shall be made up of PP or HDPE material.
Works on siphoning principal. Proper inlet and out let.
Shall accommodate 1 ml, 2 ml, 5 ml and 10 ml pipettes.
Shall accommodate pipettes of length of more than 35 cm.
Shall be supplied with inlet and outlet tubing.
Shall consists of pipette jar, basket, washer-rinser
Made up of Acrylic material
Electrode holder of Acrylic plastic
Both electrode buffer chamber of approx 500 ml
Power Supply
Inverted Microscope with transmitted light LED illumination.
The Microscopes should include
- Fixed stage plate with left or right hand drive carrying standard inserts
for slide or various culture plates, flasks and Petri-dishes etc.
- A 4-fold revolving nosepiece with infinity optics, tube lens,
phototubes-condenser holder including collector, aperture diaphragm
and filter holder.
- Trinocular tube with 40-45° viewing angle
Slider for Phase Contrast with 4 positions, 1x Bright field and 3x light
rings for condenser.
 Objectives Plan Achromatic of 4x/5x, 10x, 20x and 40x with long
working distance for Bright field and Phase contrast applications.
For anaerobic culture








Digital
Temperature range (25 – 150 0 C)
Accuracy : ±0.1°C
Uniformity : ±0.1°C
Shaking Mode – Orbital-Reciprocating
Speed (slightly lower at 50 Hz) 20 to 400 rpm, 10 to 200 oscillations/min
Display : Digital tachometer
Timer: 99 hours and 59 minutes in 1 minute increments, or for continuous
operation
 Motor : Maintenance-free, brushless
 Heater: 1500 watts
 Overall, W x D x H: (31.5 x 16.0 x 15.7) in/ (80 x 41 x 40) cm
 auto cut thermostat and voltage stabilizer
2 kg, 3 kg Capacity
18 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 Layer feeder/drinker Brooder with chick guard Automatic vaccinator Debeaker Wing band/ Leg band Electric Incinerator Chick sexing machine Automatic pan feeding
system Nipple drinking system Special Breeder Nipple Vertical Manure removal
system Hanging Feeder
Laying/nest Fogger Automatic drinker Egg fumigation cabinet Egg carrying trolleys Other poultry equipments Pellet mills Dies & Rolls Horizontal /Vertical feed
mixer Feed grinder 5 kg, 7 kg, 8 kg
G.I. make
Different capacity
Good quality
Alluminium made
Automatic poultry feed
plant Condenser for Motor
40/80 Coupling Rubber
Clip Heater U type Contact Thermometer
Germany Dry & Wet Thermometer Fan blade for setter Fan blade for Hatcher Fan belt A-75 or A-89 Complete Gear Box Gear Wheel box
Gunny Ring with pan Gunny Guard (Thal) Hatcher Tray (10 egg
capacity) Hatcher Tray (202 egg
capacity) Heater U type (Stainless
steel) Humidity drop valve Humidity Transparent pipe
(Tubka) Indicator bulbs (6 pc set) Different capacity
Good quality
Two tier good quality
Different capacity
Different sizes 25000capacity
Different capacity
Different capacity
19 117 118 119 120 121 122 123 124 125 126 127 128 Kit Kat Fuse 15 Amp. Micro switch for door
contact Micro switch for Turning Micro switch for Heater Micro switch for Humidity Solenoid valve Solenoid valve for water &
Air cooling Solenoid valve for
Humidity Turning time clock
complete set Turning time clock motor Complete Thermostat
without capsule
In-vitro Gas Production
Apparatus
129 130 131 Hotplate Magnetic Stirrer
Co2 Cylinder
Air conditioner (Split) 132 Hp Computer (Desktop)
with all-in-one
(Printer+Copier+Scanner)
i.
HFT-glass Syringes (60 piece), grind with specific capillary attachmentwide execution
ii.
HFT-clips for silicon tube (500 piece)
iii.
HFT-silicon tube for HFT-glass syringes (25 metre)
With adaptor and accessories
With regulator, Medium size Capacity





TR (Ton Range): 1.5 T ,Star Rating: 3-5, Cooling (kW): 5-5.5
Power Supply (Volt/Phase): 230/Single ,Power Input (Watts): 1500-2000
Air Direction control: Auto, Dehumidification
HD Filter: 60, Anti corrosion finish
Display: LED/digital, Remote control
 Operating system: Windows‐7, Processor: 2.8 GHz, 8 MB cache Memory:  Memory: 4 GB DDR3 RAM, Memory slots: 2 or more Storage:  Internal drive: 4 GB or more Display and graphics:  Display: 27” widescreen Full HD LED, Graphics: HD Expansion: Ports: 2 USB 3.0 (front) , 4 USB 2.0 (back),1 headphone,1 microphone  Memory card device: 6‐in‐1 memory card reader Input device:  Key board and mouse: Wireless keyboard and wireless optical mouse Communications:  Network interface: Ethernet LAN  Wireless: Wireless LAN Specification of ALL‐IN‐ONE  Functions: Print, copy, scan
 Display: LCD/LED
Printer
 Method: Monochrome laser
 Print speed black: Normal: 18 ppm or better
 Print resolution: 600x600 dpi
 Monthly duty cycle: 8000 pages or better
 First printout time: 8.5 sec or better Copier:  Speed: 18 cpm or better 20 133 Bomb Colorimeter
134 Canon Photocopier
Machine
 Resolution: 600x400 or better  Max no. of copies: 90 or better Scanner:  Type: flatbed  Scan resolution: 600x600 dpi  Scan format: pdf, jpeg, tif, gif, BMP, word  Scan size: 216 x 297 mm Working Mode Isoperibol, Dynamic, Manual & Time-Controlled
External Power Supply:
Voltage Rating 100 -240 V AC
Input Power Max 120W
Measurement frequency 60 / 50 Hz
Calorimeter: Rated Voltage 24V DC 5A
Input Power Max 100W
Fuse 1 x 2.5 AT
On-Time Continuous Operation
Protection Class III
Protection Class acc. To DIN EN
60529 IP 21
Permissible ambient temperature 20 ºC - 25ºC (constant)
Permissible humidity 80%
System structure Manual
Energy Measurement range Max. 40.000 Joule
Reproducibility based on analysis of Isoperibol 0.05% RSD
1g benzoic acid NBS 39i Dynamic 0.1% RSD
Fundamental standards DIN 519000, ISO 1928, ASTM 240 D
Temperature Measuring Resolution 1 / 10.000K
Analysis Time : Isoperibol up to 17 mins
Dynamic up to 8 mins
Manual(isoperibol) up to 17mins
Time-Controlled 14 mins
Measuring accuracy ±0.1%
Corrections Halogen, sulphur, nitrogen, moisture and ash;Possible
only with Calwin software
Type of Bomb Removable
Standard Bomb C 5010
Halogen Resistant Bomb C 5012
O2 filling of decomposition vessel Manual using Oxygen filling station
Venting of decomposition vessel Manual
Water handling Manual
Interfaces 1 x serial (RS232)
1 x parallel (Centronics)
System expansion Calwin software
Network connection
Auxiliary energy Oxygen (Purity 99.95%)
Operating Oxygen Pressure 30 bar
Cooling medium Chilled water (17ºC and 25ºC)
Weight 21 kg
Different capacity
21 
Download