OSC-16 for imc CRONOS-PL/-SL/compact

advertisement
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
04.10.2013
OSC-16 for imc CRONOS-PL/-SL/compact
16-channel isolated differential amplifier
Data Sheet version 3.4
The OSC-16 is a scanner module with optical relays, equipped with 16 isolated and differential input
channels. It comes with enhanced isolation strength of up to 60 V, and its 16 channels allow
measurement of voltage, current, thermocouples and Pt100. The OSC-16 is available as a plug-in module
for imc CRONOS-PL/compact and as a configuration module for CRONOS-SL.
Highlights
Ideal for isolated measurement of passive sensors
optimal suppression of 50 Hz interference
support of TEDS (imc Plug & Measure, Transducer Electronic Data Sheets (IEEE 1451.4)
imc CRONOScompact - modular measurement system
imc CRONOScompact is a modular and reconfigurable hardware a
"rack"-based series of devices available in different housing sizes
and device frames.
imc CRONOScompact (CRC) plug-in-modules can be plugged in
imc CRONOScompact System (CRC-400 / CRC-2000G). This
assembly of such a system can be adapted fast and easily.
imc CRONOScompact plug-in-modules
Once the modules are built-in a portable or rack-based housing,
the modules will electrically connected to the imc
CRONOScompact (CRC) system and they will also be supplied by
the system. The data storage will also be managed by CRC.
The front panel of the rack-based modules ("-R") is different to the
front panel of the portable housing. The properties of both types is
the same.
imc CRONOScompact portable housing
Overview of available variants
Order code
Article no.
CRC/OSC-16
CRC/OSC-16-ET
CRC/OSC-16-T
1170054
1171029
1170103
CRC/OSC-16-T-ET
CRC/OSC-16-R
CRC/OSC-16-R-ET
CRC/OSC-16-T-R
11710xx
1170118
1171077
1170184
CRC/OSC-16-T-R-ET
CRSL/OSC-16-D
11710xx
1180023
CRPL/OSC-16
CRPL/OSC-16-ET
CRPL/OSC-16-T
1080177
1081123
1080178
Remarks
for installation in imc CRONOScompact, occupying 2 slots
extended environmental temperature range
with thermocouple sockets, for installation in
imc CRONOScompact, occupying 3 slots (12 TE)
extended environmental temperature range
for installation imc CRONOScompact RACK, occupying 2 slots
extended environmental temperature range
with thermocouple sockets, for installation in
imc CRONOScompact RACK, occupying 3 slots (12 TE)
extended environmental temperature range
for installation in imc CRONOS-SL
with DSUB terminal connection, occupying 2 slots
for installation in an imc CRONOS-PL, occupying 2 slots
extended environmental temperature range
with thermocouple sockets for installation in
imc CRONOS-PL, occupying 2 slots (8TE)
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
1
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
CRPL/OSC-16-T-ET
04.10.2013
1081124
extended environmental temperature range
Integrated sensor supply
Version with an integrated sensor supply, requires no extra module expansion. With adjustable
supply voltages (globally selectable for 8 channels), output on reserved pins of DSUB terminal.
Required software version
The module is fully supported by the imc operating software.
Included accessories
DSUB-15 connector
ACC/DSUB(M)-T4:
DSUB-15 plug with screw terminals for 4-channel measurement
of voltages as well as temperatures with PT100 and
thermocouples with integrated cold junction compensation (CJC).
Included accessories for imc CRONOS-SL:
------
Optional accessories
DSUB-15 connectors
ACC/DSUB(M)-TEDS-T4
ACC/DSUB(M)-U4
ACC/DSUB(M)-TEDS-U4
ACC/DSUB(M)-I4
ACC/DSUB(M)-TEDS-I4
ACC/DSUB-ICP4
version with TEDS support, according to IEEE 1451.4 for
use with imc Plug & Measure
DSUB-15 plug with screw terminals for 4-channel voltage
measurement
DSUB-15 plug with screw terminals for 4-channel voltage
measurement
DSUB-15 plug with screw terminals for 4-channel current
measurement of up to 50 mA (50 shunt, scaling factor: 0.02 A/V)
version with TEDS support, according to IEEE 1451.4 for
use with imc Plug & Measure
DSUB-15 plug with screw terminals for conditioning of
4 IEPE/ICP inputs
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
2
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
04.10.2013
Technical details - OSC-16
Data Sheet version 3.4
Parameter
Value
Inputs
Remarks
16
Measurement mode DSUB
voltage
Standard plug (ACC/DSUB(M)-U4)
current
Current plug (ACC/DSUB(M)-I4)
thermocouples, RTD (Pt100)
Thermo-plug (ACC/DSUB(M)-T4)
Measurement mode OSC-16-T
thermocouples type-K
Terminal connection
4x DSUB-15
or
16x thermo-sockets
OSC-16-T
Two pin thermo-sockets
4 channels per plug
for thermocouples only
Sampling rate, Bandwidth, Filter, TEDS
Parameter
Sampling rate
Value
Remarks
max. 5 Hz (200 ms) / channel
filter / bandwidth like at 2 Hz / 5 Hz,
additional values are duplicated
10 Hz (100 ms)
Bandwidth
1 Hz
Resolution
16 Bit
internal: 2 Hz (500 ms)
with additional Interpolation;
max. allowable input signal
frequency: 1 Hz
-0.01 dB
Suppression @ 50 Hz (±2%)
5 Hz to 2 Hz
(200 to 500 ms)
34 dB @ 49 Hz / 51 Hz
1 Hz (1 s)
68 dB @ 49 Hz / 51 Hz
Bandwidth / noise suppression
max. signal frequency
Bandwidth
max.
Noise supp.
60 dB
-1 dB
signal
(fg)
(f_s)
(f_filt)
1 Hz
50 Hz
at sampling rate:
5 Hz (200 ms)
1.0 Hz
2 Hz (500 ms)
1.0 Hz
1 Hz
1 Hz (1 s)
0.5 Hz
0.5 Hz
48.5 Hz
0.5 Hz (2 s)
0.25 Hz
0.25 Hz
48.5 Hz
Synchronicity (at sampling rate)
f_s ... f_filt
aliasing-free for frequencies above:
f_filt
50 Hz
Max. settling time
aliasing only in the frequency range:
max. 1 s
constant time offset between two
equally configured channels:
(noise suppression 60 dB)
sampling rate 5 Hz (200 ms)
complete settling as a response to
input step
sampling rate 2 Hz (200 ms)
max. 500 ms
TEDS
conform IEEE 1451.4
Class II MMI
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
ACC/DSUB-TEDS(M)-xx
3
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
04.10.2013
General
Parameter
Value
Remarks
±60 V
channel to frame (housing, CHASSIS,
protection ground) and channel to
channel
Isolation
nominal
test voltage
300 V (10 sec.)
Overvoltage protection
±60 V
diff. input voltage, (long term)
ESD 2 kV
human body model
transient protection:
automotive load dump
ISO 7636, Test pulse 6
Input coupling
Ri=30 , td=300 µs, tr<60 µs
DC, isolated (differential)
Input impedance
Static input current
Dynamic input current
Test pulse 6 with max. -250 V
electrical isolation to system-GND
(housing, CHASSIS, protection ground)
10 M
voltage mode (range ±2 V),
temperature mode
1M
voltage mode (range ±5 V)
50
Current mode (current connector)
1 nA (typ.)
10 nA (max.)
settled, during sampling
0.1 mA (typ.)
1.5 mA (max.)
peak dynamic input current value
(typ. @100 mV, max. @2 V)
30 nA (typ.)
600 nA (max.)
mean dynamic input current value
(typ. @100 mV, max. @2 V)
Input current below overvoltage
1.5 mA
|Vin| > 7 V in the range ±2 V
or device deactivated
Sensor supply 5 V (DSUB-15)
Parameter
Value typ.
min. / max.
Auxiliary supply
voltage
available current
internal resistance
Remarks
for IEPE (ICP)-extension plug
+5 V
>0.26 A
1.0
5%
>0.2 A
<1.2
independent of integrated
sensor supply, short circuit proof
power per DSUB-plug
Voltage and current measurement
Parameter
Value typ.
min. / max.
Voltage input range
50 mV / 100 mV / 250 mV / 500 mV /
1 V / 2 V / 5 V / 10 V / 25 V / 50 V /
60 V
Current input range
1 mA / 2 mA / 5 mA
10 mA / 20 mA / 40 mA
Gain uncertainty
Remarks
with current connector (50
(ACC/DSUB-I4)
shunt)
<0.025 %
<0.05 %
<0.07 %
<0.15 %
voltage, 23 °C
current with current connector
<0.05 %
of input range
Offset
<3 µV
Non linearity
< 30 ppm
Gain drift
Offset drift
Noise voltage (RTI)
range: 10 V
6 ppm/K
ranges
2V
36 ppm/K
ranges
5V
3 ppm/K
<0.5 µVrms
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
over entire temp.
range
over entire temperature range
sampling rate 5 Hz (200 ms)
4
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
04.10.2013
Voltage and current measurement
Parameter
CMRR/ IMR
(isolation mode rejection)
Value typ.
min. / max.
<3.0 µVpkk (<1LSB)
all sampling rates
> 110 dB (50 Hz)
range
2V
> 95 dB (50 Hz)
range
2V
> 65 dB (50 Hz)
range
5V
Channel isolation
< 50 pF, <100 nA
Channel cross-talk damping
all sampling rates
Suppression of square edges on
neighboring channels
Remarks
Rsource = 0
Rsource = 100
Rsource = 100
Channel to protection ground
(CHASSIS); Channel-to-channel
> 116 dB (50 Hz)
range
2V
> 101 dB (50 Hz)
range
2V
Rsource = 0
Rsource = 100
>123 dB @ sampling rate 200 ms
range
2V
Rsource = 100
Max. source impedance
5k
Temperature measurement - Thermocouples
Parameter
typ.
min. / max.
Input range
R, S, B, J, T, E, K, L, N
Resolution
0.063 K (1/16 K)
Uncertainty
Drift
< 0.5 K
0.05%
0.02 K/K Ta
Remarks
compliant with IEC 584
type-K, range -150°C to 1200°C
plus value indicated
Ta=|Ta -25°C|; Ta: ambient temp.
Uncertainty of cold junction
compensation
Drift of cold junction temp.
< 0.15 K
DSUB (ACC/DSUB(M)-T4)
< 0.5 K
thermo plug (green) Typ K
< 0.7 K
thermo plug (white) with type K
< 1K
thermo plug (white) other types
0.001 K/K Tj
Sensor breakage recognition
Tj = |Tj -25°C| cold junction temp. Tj
Display: “-2000°C”
indicating unconnected input
Temperature measurement – Pt100 (RTD)
Parameter
Input range
typ.
min. / max.
Remarks
-200°C to 850°C
-200°C to +250°C
Resolution
0.063 K (1/16 K)
Uncertainty
< 0.1 K
0.05%
Drift
Reference current (Pt100)
0.01 K/K Ta
250 µA
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
-200°C to +850°C,
4-wire configuration
plus indicated value
Ta=|Ta -25°C|;
Ta: ambient temperature
non-isolated (CHASSIS is Ground)
5
OSC-16 for imc CRONOS-PL/-SL/compact
Technical Data Sheet
04.10.2013
Technical Specs sensor SUPPLY module
Parameter
Configuration options
Value (typ./ max.)
5 adjustable ranges
Remarks
The sensor supply module always got 5
selectable voltage ranges.
Default ranges: +5 V to +24 V
Output voltage
Voltage
Current
Netpower
(+2.5 V)
+5.0 V
+10 V
+12 V
+15 V
+24 V
( 15 V)
580 mA
580 mA
300 mA
250 mA
200 mA
120 mA
190 mA
1.5 W
2.9 W
3.0 W
3.0 W
3.0 W
2.9 W
3.0 W
Isolation
Standard:
option, upon request:
non isolated
isolated
Short-circuit protection
unlimited duration
Accuracy of output voltage
Efficiency
Max. capacitive load
special order, +12 V can be replaced by +2,5 V
set globally for all channels of an amplifier.
special order, +15 V can be replaced by 15 V
output to case (CHASSIS)
nominal rating: 50 V, Test voltage (10 sec.): 300
V, not available with option 15 V.
to output voltage reference ground
at terminals, no load
<0.25% (typ.)
<0.5% (max.)
25°C
25°C
<0.9% (max ).
over entire temperature range
typ. 72%
typ. 66%
typ. 55%
typ. 50%
10 V to 24 V none isolated
5V
>4000 µF
>1000 µF
>300 µF
2.5 V to 10 V
12 V, 15 V
24 V
10 V to 24 V isolated
5V
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
6
Download