OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
OSC-16 for imc CRONOS-PL/-SL/compact
16-channel isolated differential amplifier
Datasheet version 3.3 23.10.2012
The OSC-16 is a scanner module with optical relays, equipped with 16 isolated and differential input
channels. It comes with enhanced isolation strength of up to 60 V, and its 16 channels allow measurement
of voltage, current, thermocouples and Pt100. The OSC-16 is available as a plug-in module for imc
CRONOS-PL/compact and as a configuration module for CRONOS-SL.
Highlights:
Ideal for isolated measurement of passive sensors
optimal suppression of 50 Hz interference
support of TEDS (imc Plug & Measure, Transducer Electronic Data Sheets (IEEE 1451.4)
Order code:
Article number Remarks
CRC/OSC-16
CRC/OSC-16-ET
CRC/OSC-16-T
1170054
1171029
1170103
CRC/OSC-16-T-ET
CRC/OSC-16-R
CRC/OSC-16-R-ET
CRC/OSC-16-R-T
11710xx
1170118
1171077
11700xx
CRC/OSC-16-R-T-ET
11710xx
CRSL/OSC-16-D
1180023
CRPL/OSC-16
CRPL/OSC-16-ET
CRPL/OSC-16-T
1080177
1081123
1080178
CRPL/OSC-16-T-ET
1081124
for installation in imc CRONOScompact
extended environmental temperature range
with thermocouple sockets
for installation in imc CRONOScompact
extended environmental temperature range
for installation in imc CRONOScompact RACK
extended environmental temperature range
with thermocouple sockets
for installation in imc CRONOScompact RACK
extended environmental temperature range
for installation in imc CRONOS-SL
with DSUB terminal connection
for installation in an imc CRONOS-PL housing
extended environmental temperature range
with thermocouple sockets
for installation in imc CRONOS-PL
extended environmental temperature range
Physical structure:
Plug-in module for imc CRONOS-PL/compact systems, occupying two slots
Terminal connection:
4x DSUB-15 terminals (4 channels per connector)
16x thermocouple sockets (-T)
16x 7-pin LEMO terminals with 1 channel per connector plug (only with CRSL/OSC-16-L)
For this module, the use of DSUB terminals is recommended, since LEMO terminals do not support
temperature measurement with thermocouples.
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
1
OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
Included accessories for imc CRONOS-PL/compact:
Terminal connection (DSUB):
4x ACC/DSUB-T4, 15-pin DSUB connection terminal for each 4-channel group. For measurement
of voltages as well as temperatures with Pt100 and thermocouples (with integrated cold-junction
compensation)
Included accessories for imc CRONOS-SL:
------
Power supply:
provided by imc CRONOS-PL/-SL/compact unit
Additional power consumption due to installed module: 1.68 W
Operating conditions:
The module varieties' respective operating conditions (with or without an extended temperature
range) depend on the corresponding housing type.
Installed software:
The module is fully supported by the imc CRONOS-PL/-SL/compact operating software. The entire
functionality, particularly the parameterization, storage and online computations is provided.
Data storage:
handled via imc CRONOS-PL/-SL/compact
Optional accessories:
Sensor supply:
Integrated sensor supply module (requires no extra slot) with adjustable supply voltages, output on
DSUB terminals.
Connection terminals:
ACC/DSUB-U4 4-channel connector for voltage measurement
ACC/DSUB-I4, 15-pin DSUB terminal connector for each 4-channel group (50
measurement of currents up to 50 mA (scaling factor: 0.02 A/V)
shunt). For
ACC/DSUB-TEDS-U4, 15-pin DSUB terminal connector for each 4-channel voltage measurement
group, according to IEEE 1451.4 for use with TEDS
ACC/DSUB-TEDS-I4, 15-pin DSUB terminal connector for each 4-channel group (50 shunt). For
measurement of currents up to 50 mA (scaling factor: 0.02 A/V), conformant to IEEE 1451.4 for
use with TEDS
ACC/DSUB-TEDS-T4, 15-pin DSUB terminal connector for each 4-channel group. For
measurement of voltages as well as temperatures with Pt100 and thermocouples (with integrated
cold-junction compensation), compliant with IEEE1451.4 for use with TEDS
ACC/DSUB-U4-IP65, terminal connector adapted to CRONOS-SL for voltage measurement on 4
channels
ACC/DSUB-T4-IP65, 15-pin terminal connector adapted to CRONOS-SL for 4-channel groups.
For measurement of voltages as well as temperatures with Pt100 and thermocouples (with
integrated cold-junction compensation)
ACC/DSUB-I4-IP65, 15-pin DSUB terminal connector adapted to CRONOS-SL for each 4-channel
group (50 shunt). For measurement of currents of up to 50 mA (scaling factor: 0.02 A/V)
ACC/DSUB-TEDS-U4-IP65, 15-pin DSUB terminal connector adapted to CRONOS-SL, for each
group of 4 voltage measurement channels, conformant to IEEE1451.4 for use with TEDS
ACC/DSUB-TEDS-I4-IP65, 15-pin DSUB terminal connector adapted to CRONOS-SL for each 4channel group (50 shunt). For measurements of currents of up to 50 mA (scaling factor: 0.02 A/
V), conformant to IEEE1451.4 for use with TEDS
ACC/DSUB-TEDS-T4-IP65, 15-pin DSUB terminal connector adapted to CRONOS-SL for each 4channel group. For measurement of voltages as well as temperatures with Pt100 and
thermocouples (with integrated cold-junction compensation), compliant with IEEE1451.4 for use
with TEDS
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
2
OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
OSC-16 voltage / current / temperature (isolated)
Technical Datasheet version 3.3
Parameter
Inputs
Measurement (DSUB)
Measurement mode OSC-16-T
Terminal connection
Standard
Value (typ. min. / max.)
16
voltage
current
thermocouples, RTD (Pt100)
thermocouples type-K
4x DSUB-15 / 4 channels
or
16x thermo-sockets
or
16x LEMO
OSC-16-T
OSC-16-L
General
Parameter
Bandwidth / noise suppression
max. signal frequency
(f_s)
(f_filt)
1 Hz
50 Hz
1 Hz
50 Hz
0.5 Hz
48.5 Hz
0.25 Hz
48.5 Hz
1 Hz
max. 5 Hz (200 ms) / channel
Synchronicity (at sampling rate)
Resolution
TEDS
Isolation
nominal
test voltage
Overvoltage protection
Input coupling
Input impedance
Remarks
aliasing only in the frequency range:
f_s ... f_filt
aliasing-free for frequencies above:
f_filt
(noise suppression 60 dB)
1.0 Hz
1.0 Hz
0.5 Hz
0.25 Hz
10 Hz (100 ms)
Suppression @ 50 Hz ( 2 %)
5 Hz to 2 Hz
(200 to 500 ms)
1 Hz (1 s)
Max. settling time
ACC/DSUB-U4
ACC/DSUB-I4
ACC/DSUB-T4
for thermocouples only
CRSL/OSC-16-L
Value (typ. min. / max.)
Bandwidth
max.
Noise supp.
-1 dB
signal
60 dB
(fg)
at sampling rate:
5 Hz (200 ms)
2 Hz (500 ms)
1 Hz (1 s)
0.5 Hz (2 s)
Bandwidth
Sampling rate
Remarks
differential, isolated
Standard connector (ACC/DSUB-U4)
Current connector (ACC/DSUB-I4)
Thermo-connector (ACC/DSUB-T4)
Two pin thermo-sockets
-0.01 dB
internal: 2 Hz (500 ms)
with additional Interpolation;
max. allowable input signal
frequency: 1Hz
filter / bandwidth like at 2 Hz / 5 Hz,
additional values are duplicated
34 dB @ 49 Hz / 51 Hz
68 dB @ 49 Hz / 51 Hz
max. 1 s
sampling rate 5 Hz (200 ms)
complete settling as a response to
input step
constant time offset between two equally
configured channels:
sampling rate 2 Hz (200 ms)
max. 500 ms
16 Bit
conform IEEE 1451.4
ACC/DSUB-TEDS-U4
Class II MMI
ACC/DSUB-TEDS-I4
ACC/DSUB-T4
60 V
300 V (10 sec.)
60 V
ESD 2 kV
transient protection:
automotive load dump
ISO 7636, Test pulse 6
DC, isolated (differential
10 M
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
channel to frame (CHASSIS)
and channel to channel
diff. input voltage, (long term)
human body model
Test pulse 6 with max. -250 V
Ri=30 , td=300 µs, tr<60 µs
electrical isolation to system-GND
(frame, CHASSIS)
voltage mode (range 2 V),
3
OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
General
Parameter
Value (typ. min. / max.)
1M
50
Static input current
Dynamic input current
1 nA (typ.)
0.1 mA (typ.)
10 nA (max.)
1.5 mA (max.)
30 nA (typ.)
600 nA (max.)
Input current below overvoltage
Sensor supply voltage
1.5 mA
+5 V, 250 mA / 4 channels
Voltage and current measurement
Parameter
typ.
min. / max.
Voltage input range
50 mV / 100 mV / 250 mV / 500 mV /
1 V / 2 V / 5 V / 10 V / 25 V / 50 V /
60 V
Current input range
1 mA / 2 mA / 5 mA
10 mA / 20 mA / 40 mA
Gain uncertainty
<0.025 %
<0.05 %
<0.07 %
<0.15 %
Offset
<0.05 %
<3 µV
Non linearity
< 30 ppm
Gain drift
6 ppm/K
36 ppm/K
Offset drift
Noise voltage (RTI)
CMRR/ IMR
(isolation mode rejection)
3 ppm/K
<0.5 µVrms
<3.0 µVpkk (<1LSB)
all sampling rates
> 110 dB (50 Hz)
> 95 dB (50 Hz)
> 65 dB (50 Hz)
Remarks
temperature mode
voltage mode (range
5 V)
Current mode (current connector)
settled, during sampling
peak dynamic input current value
(typ. @100 mV, max. @2 V)
mean dynamic input current value
(typ. @100 mV, max. @2 V)
|Vin| > 7 V in the range 2 V
or device deactivated
non-isolated, short-circuit proof
(equipped as standard); replaceable
by optional supply module with
selectable voltage ranges
(SUPPLY)
on terminals „ SUPPLY“ , NOT on
„+5 V / GND“, not compatible with
IEPE (ICP) extension plug
Remarks
with current connector (50 shunt)
(ACC/DSUB-I4)
voltage, 23 °C
current with current connector
of input range
range: 10 V
over entire temp.
ranges
2V
range
ranges
5V
over entire temperature range
sampling rate 5 Hz (200 ms)
range
range
range
2V
2V
5V
Rsource = 0
Rsource = 100
Rsource = 100
Channel isolation
< 50 pF, <100 nA
Channel cross-talk damping
all sampling rates
> 116 dB (50 Hz)
> 101 dB (50 Hz)
range
range
2V
2V
Rsource = 0
Rsource = 100
>123 dB @ sampling rate 200 ms
range
2V
Rsource = 100
Suppression of square edges on
neighboring channels
Max. source impedance
Channel to protection ground
(CHASSIS); Channel-to-channel
5k
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
4
OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
Temperature measurement - Thermocouples
Parameter
typ.
min. / max.
Input range
R, S, B, J, T, E, K, L, N
Resolution
0.063 K (1/16 K)
Uncertainty
< 0.5 K
0.05 %
Drift
0.02 K/K Ta
Uncertainty of cold junction
compensation
Drift of cold junction temp.
Sensor breakage recognition
< 0.15 K
0.001 K/K Tj
Display: “-2000 °C”
0.05 %
Reference current (Pt100)
type-K, range -150 °C to 1200 °C
plus value indicated
Ta=|Ta -25 °C|;
Ta: ambient temperature
ACC/DSUB-T4, OSC-16-T
Tj = |Tj -25 °C|
cold junction temperature Tj
Temperature measurement – Pt100 (RTD)
Parameter
typ
min. / max
Input range
-200 °C to 850 °C
-200 °C to +250 °C
Resolution
0.063 K (1/16 K)
Uncertainty
< 0.1 K
Drift
Remarks
compliant with IEC 584
0.01 K/K
Ta
250 µA
Errors and changes excepted
For more information, contact:
Berlin: +49 - 30 - 46 70 90 - 0
indicating unconnected input
Remarks
-200 °C to +850 °C,
4-wire configuration
plus value indicated
Ta=|Ta -25 °C|;
Ta: ambient temperature
non-isolated (CHASSIS is Ground)
5
OSC-16 for imc CRONOS-PL/-SL/compact
Technical datasheet
Technical Specs sensor SUPPLY module
Parameter
Value (typ./ max.)
Configuration options
5 adjustable ranges
Output voltage
Isolation
Standard:
option, upon request:
Short-circuit protection
Accuracy of output voltage
Efficiency
Max. capacitive load
Voltage
(+2.5 V)
+5.0 V
+10 V
+12 V
+15 V
+24 V
( 15 V)
Current
580 mA
580 mA
300 mA
250 mA
200 mA
120 mA
190 mA
Netpower
1.5 W
special order, +12 V can be replaced by +2,5 V
2.9 W
set globally for all channels of an amplifier.
3.0 W
3.0 W
3.0 W
2.9 W
3.0 W
special order, +15 V can be replaced by 15 V
non isolated
isolated
unlimited duration
<0.25% (typ.)
<0.5% (max.)
<0.9% (max ).
typ. 72%
typ. 66%
typ. 55%
typ. 50%
>4000 µF
>1000 µF
>300 µF
Remarks
The sensor supply module always got 5
selectable voltage ranges.
Default ranges: +5 V to +24 V
output to case (CHASSIS)
nominal rating: 50 V, Test voltage (10 sec.): 300
V, not available with option 15 V.
to output voltage reference ground
at terminals, no load
25°C
25°C
over entire temperature range
10 V to 24 V none isolated
5V
10 V to 24 V isolated
5V
2.5 V to 10 V
12 V, 15 V
24 V
Errors and changes excepted