DENTAL LIGHT IO 5000TA TYPE AL-820S INSTALLATION AND OPERATION IMPORTANT After installation is completed, check all the bolts, screws and fasteners to confirm that they are securely fastened. INSTRUCTIONS TABLE OF CONTENTS Page [1] MAJOR PARTS IDENTIFICATION---------------------------------------------- 1 [2] SPECIFICATIONS-------------------------------------------------------------------- 1 [3] CLASSIFICATION-------------------------------------------------------------------- 1 [4] WIRING DIAGRAM----------------------------------------------------------------- 2 [5] DIMENSIONS-------------------------------------------------------------------------- 2 [6] INSTALLATION 6-1. Attach the light assembly to the light pole------------------------------------- 3 6-2. Attch the light pole to the light post----------------------------------------------- 3 6-3. Attach the wire harness----------------------------------------------------------- 3 [7] OPERATION 7-1. Mode selection switch------------------------------------------------------------ 4 7-2. Touch-less switch ----------------------------------------------------------------- 4 7-3. Intensity switch ---------------------------------------------------------------------- 4 7-4. Switching to the composite mode ------------------------------------------------- 4 [8] ADJUSTMENT OF THE BALANCE ARM 8-1. Adjustment of the tension of the balance arm--------------------------------- 5 8-2. Adjustment of the angle of the light head-------------------------------------- 5 8-3. Adjusting the tension of the light head -------------------------------------------- 6 8-4. Balance arm adjustment (lateral movement)-------------------------------------- 6 8-5. Converting the balance arm for left handed -------------------------------------- 6 [9] REPLACEMENT OF THE BULB 9-1. Removal of the reflector cover -------------------------------------------------- 8 9-2. Replacement of the bulb---------------------------------------------------------- 8 9-3. Inserting a new bulb -------------------------------------------------------------- 9 9-4. Attaching the reflector cover----------------------------------------------------- 9 [10] CLEANING---------------------------------------------------------------------------10 [11] MAINTENANCE AND INSPECTION------------------------------------------11 [12] BEFORE ASKING FOR REPAIRS----------------------------------------------12 [13] ELECTROMAGNETIC COMPATIBILITY (EMC)-------------------------13 Intended Use of the Product This product is intended for the exclusive use for diagnoses, treatments and relative procedures of dentistry, and must be operated or handled by the qualified dentists or by dental staffs under the supervision of the dentist. Such dentists or dental staffs should instruct and/or assist the patients to approach to and leave from the product. Patients should not be allowed to operate or handle the product unless he/she is so instructed. Environmental condition for Operation Temperature : 5 ~ 40℃ Humidity : 10 ~ 80% Pressure : 600 ~ 1060 hpa Environmental condition for Storage Temperature : -10 ~ 50℃ Humidity : 10 ~ 80% Pressure : 600 ~ 1060 hpa Environmental condition for Transportation Temperature : -10 ~ 50℃ Humidity : 10 ~ 80% Pressure : 600 ~ 1060 hpa Important Notes In case of the troubles, please contat Takara Belmont offices or your dealers. Do not disassemble or attempt to repair. Disassembly, repair or modifications shoud only be done by a qualified repair technician. Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents. In case of disposal of equipment When disposing the dental light, appropriately dispose complying with all current applicable regulations and local codes. In EU area, EU directive 2002/96/EC on waste electrical and electronic equipment (WEEE) is applied on this product. In this directive, environment conscious recycling/abandonment is obligated. Symbols Alternating-current 㻨㻦 㻵㻨㻳 Authorized representative in the European community Separate collection for electrical and electronic equipment Power ON Manufacturer Refer to instruction manual/booklet Power OFF Date of manufacture Non-ionizing radiation Caution It means “caution, warnings, or possibility to danger”. WARNING Operation of a dental light under Electro Magnetic Interferences. This equipment may malfunction by electromagnetic interferences. Do not install dental light close to the equipment that generates electromagnetic interference (i.e. communication equipment, elevator) Do not use the device that creates interference (i.e. cellular phone) near this equipment. Do not use electric surgical knife or laser knife with dental light. Light may turns on/off by itself due to malfunction of a sensor caused by electromagnetic interferences. WARNING • To avoid the risk of electric shock, this equipment must only be connected to supply mains with protective earth. • Lamp cover becomes very hot during use.To avoid burning fingers, do not a touch lamp cover. • Always put a lamp cover on to avoid burning fingers. • Position a light head by holding light handle(s). • Installation and service work should be conducted by an authorized installation/service personnel only. WARNING: The followings are prohibited. • To modify this equipment. • To use the equipment under any failure condition. • To use the equipment without doing the daily and periodical check-up. • To wipe the plastic covers with any disinfectant or detergent that contains organic solvent. Precautions for Installation • Keep the equipment away from water. • Keep in circumstances safe from influence by temperature, humidity, wind, sun light, air containing salts and minerals. • Care about stability such as inclination, vibration and impact, including handling and transportation. • Do not keep the equipment in a place where chemicals are or where gas is emitted. • During lifting and unpacking of the light, make sure to hold only the designated parts. • Do not drop or hit the light. • Ground light properly prior to turning power on. • When the installation process has been completed, verify that all the mechanical and electrical functions are working properly. • Thick gloves are highly recommended at unpacking. • Do not modify this equipment. CAUTION Before use • Check connection of switches and make sure that the device functions properly. • Make sure that grounding wire is connected. • Make sure that cables are properly and perfectly connected. During use • Do not use the light longer than required for examination or treatment. • Always watch the patient and the equipment to make sure nothing is wrong. • If anything wrong is observed with the equipment or the patient, take a proper action, such as stopping the use of equipment as well as keeping the patient in safe. • Keep an eye on the patient not to touch the equipment. After use • Turn off the light. • Clean the equipment and get it ready for use. CAUTION Do not spray liquids directly onto light surfaces. In order to prevent damage to electrical components and systems, do not apply excess cleaning solution onto light surfaces. NOTE Warranty does not cover damage to equipment caused by disinfectant solutions Replacement of parts (Except the Bulb and Lamp cover) Replacement must be done by an professional technician(s) of our company or a company authorized by us. [1] MAJOR PARTS IDENTIFICATION 1 2 3 4 5 6 8 7 1 Light Pole Joint 2 Balance Arm Joint 3 Arm Joint Cover 4 Balance Arm 5 Arm Joint Cover 6 Yoke Joint 7 First Yoke 8 LED 9 Second Yoke 10 Intensity Switch 9 10 11 Handle (Left) 12 Touchless Sensor Switch 13 Reflector Cover 16 14 Light Bulb Cover 15 Handle (Right) 16 Yoke Rear Cover 15 13 14 18 11 17 Bracket Frame 18 Reflector 12 17 [2] SPECIFICATIONS 1. Power Requirement..................................................... AC23.7V 50/60Hz 2. Bulb Type..................................................................... Tungsten Halogen Type (JA 24V 60W) 3. Focal Distance............................................................. 700mm 4. Color Temperature....................................................... 5000Kelvin at 28,000 Lux 5. Light Intensity.............................................................. H : 28,000 Lux M : 22,000 Lux L : 8,000 Lux 6. Light Pattern................................................................ 60mm x 170mm - 84mm x 217mm 7. Net Weight................................................................... 5.2kg 8. Service Life.................................................................. 10 years [3] CLASSIFICATION a. Protection against electric shock : Class I Equipment b. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. -1- [4] WIRING DIAGRAM Remark Connector Symbol NMF・・・New Mini Fit Connector VL・・・VL Connector YL・・・YL Connector SV・・・SV Connector VH・・・VH Connector PH・・・PH Connector NH・・・NH Connector Black dot Black dot indicates male pin side Plug Mode Selection Switch OFF Cap Sensor Manual CN7 2 1 3 CN8 2 3 4 1 CN2 5V 1 VH4P Red 1 2 3 Red Black Brown 1A0190** Lead Wire(Light Output) 1 2 Black Brown 1 2 Red Yellow Orange Purple Blue Green SV2P 1 2 CN3 1 2 3 Low Intensity Indicate 4 GND 5 Intensity Selection (Alternate) Intensity Selection (Momentary) Hi Intensity Indicate 1A018K** Inside Arm Lead Wire CAGP81** Lead Wire for Lamp Lead Wire for Sensor SW Setting condition Change Intensity Off : Disable On : Enable 2 Duration of time switching to Composite Mode Off : 1second On : 2seconds Intensity setting (When turning light on from off condition by using a sensor) Off : Switch to composite mode when a hand is held in front of a sensor for 3 more than 2 seconds On : Stay at normal treatment mode though a hand is held in front of a sensor for more than 2 seconds Baud Rate 4 Off : 4800bps On : 9600bps Default setting 1 OFF ON SW2 ON On : When a hand move out 2 Composite Mode Off : Enable On : Disable Intensity setting switch 3 Off : Alternate On : Momentary 4 (42) 232.8 160° 20° 45° 160° ° 67.5 340 657 340 40° 40° φ38.5 10.3 160 125 13 0° ° 10 -2- OFF ON 46.5 530 Default setting OFF OFF 45° 369 Sensitivity Setting OFF [5] DIMENSIONS (73.5) CN1 NH3P (CGGN64**) Setting condition Sensor activation method by movement of a hand Off : When a hand move in 1 5V 2 3 GND Sensor P.C.B. (CGCN37**) 1 Lamp Round Connectors White/Gray White White/Yellow 3 2 1 1 2 3 4 5 6 LED P.C.B. Setting of PCB Dip Switch 50 Faston Slide Connectors White White 1 2 YL2P Sensor Input 2 GND 3 CN1 1 Power Input 2 SW1 Black Dental Light PCB Orange 1A0191** Light SW Section(DL-820) Head Yoke Arm 272 [6] INSTALLATION 6-1. Attach the light assembly to the light pole (Fig.6-1) Pass the light cable through the light pole and insert the light pole joint into the light pole. Fasten with two button screws (M6x8 ) . Light Pole Joint Button Screw (M6x8) Light Pole Light Cable Light Head Fig. 6-1 6-2. Attch the light pole to the light post (Fig.6-2) Insert the light pole to the light post. Fasten with screws. Please refer to a installation instruction of the unit. Light Pole Light Post Fig. 6-2 6-3. Connect the wire harness (Fig.6-3) Attach the connectors of the light cable onto the dental light PCB and harness inside the cuspidor unit. Wire colours and connectors for connection are shown in Fig.6-3. See also the wiriring diagram on page2. 2P Female 2P Male 1. Black 2. Bown 3P Female 1. Red 2. Yellow 3. - Connect directly to a dental light PCB 5P Female 1. 2. Orange 3. Purple 4. Blue 5. Green Fig. 6-3 -3- Light Cable [7] OPERATION 7-1. Touchless Switch (Fig.7-1) Approach a hand to the touchless switch within the reaction zone to turn on/off for the light. LED Intensity Switch Touchless Sensor Reaction Zone : 65mm Fig. 7-1 7-2. Intensity Switch (Fig.7-2) Push intensity switch. High High beam : Green LED turns on. Low Low beam : Amber LED turns on. Light Intensity 28,000Lux 22,000Lux LED Green Amber Fig. 7-2 7-3. Switching to the composite mode (Fig.7-3) During normal treatment mode, while the light is on, place and hold your hand within the reaction zone for 2 seconds. The light intensity will be automatically reduced to the composite mode. To switch back to the normal treatment mode, place your hand by the sensor for 2 seconds again. Green or Amber LED blinks in the composite mode. 7-1. Mode Selection switch Mode can be changed by dental light select switch on the unit. Please refer to operating instruction of the unit. -4- Composite Light Intensity Fig. 7-3 8,000Lux LED Blink [8] ADJUSTMENT 8-1. Adjustment of the tension of the balance arm Position the angle of the balance arm so the tension adjustment nut becomes visible just under slot A. Turn the nut and adjust the tension with the adjustment bar supplied with the light. Adjustment Bar Balance Arm Joint Light Pole Joint Hole B Slot A Balance Arm Tension Adjustment Nut Slot C Slot A (for Balance) Direction of rotation of the nut When the light head tends to go up, turn the nut to + direction. When the light head tends to go down, turn the nut to – direction. 8-2 Adjustment of the angle of the light head (1) Use a 2mm hex key wrench to loosen socket screw in front of Hole B. CAUTION Adjusting the head without loosening the socket screw will damage the arm. (2) Position the angle of the balance arm so the head angle adjustment nut becomes visible just under the slot C. Turn the nut and adjust the tension with the adjustment bar supplied with the light. When the light head tilts upward, turn the nut to + direction. When the light head tilts downward, turn the nut to - direction. Slot C Hole B Slot C Hole B Adjustment Nut Socket Screw (3) After adjustment is done, re-tighten the socket. CAUTION Unless this socket screw is re-tightened, the angle of the head will be changed during use. -5- 8-3. Adjusting the tension of the light head 1. With a No.2 Phillips head screwdriver, unscrew the M3x5 screws, and remove the left and right yoke caps. 2. Adjust the tension by equally tightening or loosening the two socket head screws using the supplied 1.5mm Allen key. 3. Upon completion of the adjustment, reattach the left and right yoke caps that were removed in Step 1 with a No.2 Phillips head screwdriver and the M3x5 screws. Socket Screw Yoke Cap Screw (M3x5) Yoke Cap CAUTION Ensure that adjustments are made equally to both left and right sides. Failure to do so may result in excess slack or damage. 8-4. Balance arm adjustment (lateral movement) If the movement of the balance arm is either too loose or too tight, follow the procedures below to adjust. 1. Loosen the four M5 socket head screws (E) with the supplied 2.5mm Allen key. CAUTION When adjusting, make sure that the socket head screws are loosened. Failure to do so may cause damage to the threads of the M8x20 screws. Light Pole Joint (D) M8x20 Screw (Top) Balance Arm Joint Balance Arm Light Pole (E) M5 Socket Screw x 4pcs. (D) M8x20 Screw (Bottom) 2. Adjust the lateral tension of the balance arm by tightening or loosening the top and bottom M8X20 screws of the balance arm joint (D). (Use a No.3 (big) Phillips head screwdriver.) 3. Upon completion of the adjustment, retighten the four M5 socket head screws (E) that were loosened in Step 1 with the supplied 2.5mm Allen key. CAUTION Upon completion of the adjustment, make sure to retighten the socket head screws. Failure to do so may render the balance arm immobile or may damage the M8x20 screws. -6- 8-5. Converting the balance arm for left handed use When converting the balance arm from right to left handed use, follow the procedures below. 1. Loosen the four M5 socket head screws (E) of the balance arm joint with the supplied 2.5mm Allen key. 2. Unscrew the M8x20 screws (D), both top and bottom. (Use a No.3 (big) Phillips head screwdriver.) Light Pole Joint (D) M8x20 Screw (Top) Balance Arm Joint Balance Arm Light Pole (E) M5 Socket Screw x 4pcs. (D) M8x20 Screw (Bottom) 3. Remove both M6 bolts that hold the light pole and light pole joint together with the supplied 3mm Allen key. 4. Detach the balance arm joint from the light pole joint. 5. Rotate the light pole joint 180 degrees, then screw both M6 bolts back in using the supplied 3mm Allen key. M6 Button Bolt Light Pole Joint Light Cable Light Pole M6 Button Bolt Light Pole Joint Balance Arm Joint 6. Place a metal washer on the end of the upper side of the light pole joint, and move the balance arm in place. 7. Loosely tighten both top and bottom M8x20 (D) screws. 8. Tighten both top and bottom M8x20 (D) screws, and adjust the lateral tension of the balance arm. 9. To ensure that the four M5 bolts (E) of the light pole joint do not inadvertently become loose, securely tighten them using the supplied 2.5mm Allen key. Metal Washer (D) M8x20 Screw (Top) Ligh Pole Joint (E) Socket Screw x 4pcs. (D) M8x20 Screw (Bottom) CAUTION After adjustment, fasten the socket head screws. Failure to do so may render the balance arm immobile or may damage the M8x20 screws. -7- [9] REPLACEMENT OF THE BULB CAUTION Make sure the power supply is turned off. Halogen bulb and surrounding parts be hot immediately after the lamp goes off. Wait until they get cool down. Do not touch glass with bare hand. Halogen bulb surface must be clean. Oil or body moisture will affect bulb span of life. If glass surface is touched, clean with alcohol. 9-1. Removal of the reflector cover (1) Turn the light bulb cover fixing screw counterclockwise by 90 degrees with a coin. See Fig.9-1-1 (2) Remove the reflector cover by pulling it toward you. See Fig.9-1-2 Light bulb cover fixing screw Coin Light bulb cover Cut out Reflector cover Fig.9-1-1 Fig.9-1-2 Reflector cover fixing gap 9-2. Replacement of the bulb (1) Disconnect the connectors. See Fig.9-2-1 (2) Push the plastic-coated parts of the bulb retaining spring, and remove it from the hook. See Fig.9-2-2 Pull off Flat connector Plastic coated Fig.9-2-1 Hook Fig.9-2-2 (3) Remove the bulb. See Fig.9-2-3 Bulb Fig.9-2-3 -8- Bulb retaining spring 9-3. Inserting a new bulb IMPORTANT Replace halogen lamp only with type JA-24V 60W cbtaainable through your local dealer. (1) At inserting the bulb, make sure that the projection of the bulb holder and the cut-out of the bulb are aligned. Misalignment will cause an improper light pattern. Cut-out of bulb Projection of the bulb holder Fig.9-3-1 (2) Close the bulb retaining spring. Make sure to place the springs behind the hook. See Fig.9-3-2 (3) Firmly attach wire connectors. See Fig.9-3-3 Connect firmly Fig.9-3-3 Fig.9-3-2 9-4. Attaching the reflector cover (1) Insert the cut-out of the reflector cover into the gap at the handle (right). See Fig.9-4-1 (2) Be careful not to pinch the wires when attaching cover. See Fig.9-4-2 (3) Hold the reflector cover and tighten the screw by turning it clockwise by 90 degrees. Make sure to tighten it by pushing it forward until it clicks. See Fig.9-4-3 Light bulb cover fixing screw Light bulb cover fixing screw Be careful not to pinch the wires Thread hole Cut out * Push and tighten Reflector cover fixing gap Fig.9-4-1 until it clicks Fig.9-4-2 -9- Fig.9-4-3 [10] CLEANING CAUTION Allow light to cool prior to cleaning. CAUTION All surfaces can be cleaned with DURR FD333 cleaner. Spray the cleaner (DURR FD333) on cloth and wipe the surfaces with the cloth. Wipe all surfaces dry after cleaning. REFLECTOR : Extreme care should be taken to prevent scratching reflector surfaces, as this will degrade the performance of the light. -10- [11] MAINTENANCE AND INSPECTION 11-1.Guide for daily maintenance and inspection (Maintenance and inspection by user) Management of maintenance and inspection of medical equipment should be implemented by the user (medical inspection). In case the user does not implement such management, it is permitted that such management is outsourced to a qualified entity such as a medical equipment repair company. For safe use of this product, it is necessary that inspection should be conducted in the specified frequency on the items described below. No. Item Frequency Inspection Method and diagnosis Influence if inspection is not conducted Maintenance required in case of nonconformity 1 Appearance of Lamp Cover Before start Check deformation, scratches by eyes Deterioration of optical performance 2 Lamp cover is firmly attached or not Before start Pull the lamp cover. Make sure that the lamp cover doesn't fall Lamp cover may fall. Reattach a lamp cover 3 Light bulb Before start Check whether the light brinks or not due to the poor connection. Light bulb turns very hot. This could shorten bulb life. Securely insert a light bulb into the bulb holder 4 Light switch Before start Set Sensor/Manual switch to Manual. Confirms the light turns on. Light doesn't function. Replace a light bulb 5 Touchless sensor Before start Set Mode Selection switch to Sensor. Confirms the light turns on. Light doesn't function. Clean the sensor surface. 6 Intensity switch Before start Change intensity. It may cause the difficulty at dental practice. Turn off the light.Contact your dealer or our office 7 Functionality of Balance Arm Before start Swing the balance arm up/down. Swing the balance arm right/left. Make sure the balance arm stops and holds its position. Light head doesn't stay at the desired position. 8 Vertical rotation of the Light head Before start Rotate the light head up/down. Make sure the light head stops and holds its position. Light head doesn't stay at the desired position. 9 Light head angle Before start Confirms the light head is vertically aligned. Light head doesn't stay at the desired position. Adjust the angle of the light head. Light head rotation Before start Rotate the light head right/left. Make sure the light rotate 160° to each direction. Light head rotates freely. This may cause the snapping of a wire. Contact your dealer or our office Discoloration and deterioration to the exterior, and corrosion and rusting to metallic components may arise. Executes wiping in accordance with "Cleaning Instructions" Light may not function right. Turn off the light.Contact your dealer or our office 10 11 Care exterior Before start Chemical and dirt on the product exterior must be cleaned. 12 Other 1 Once every week Make sure that no abnormal noise occurs when the product is operated. 13 Other 2 㻃As needed If the light has not been used for a long time, make sure the light functions correctly and safely. -11- Change a lamp cover Adjust the tension of the balance arm. Adjust the tension of vertical rotation. If the light malfunctions, Contact your dealer or our office 11-2. Guide for Periodical Check-up Some parts and components of the products are degraded or deteriorated depending on the frequency of use. Annual check-up and maintenance, as well as replacement of consumable parts, are required. The required parts (including consumable parts) are listed below. It may be different from the following list depending on the option of the light. For check-up and repair, call a technician of our authorized dealer. Parts and components that require periodical check-up Standard Lifetime No. Moving part 7 years 3 Switches 5 years Electric wiring of moving part 5 years 4 Control PCBs. 5 years No. Parts Description 1 2 Parts Description Standard Lifetime Consumable parts No . Parts Description 1 Bulb 2 Lamp Cover WARNING Execute the maintenance in accordance with this instraction manual and operating manual attached to each individual equipment ( Dental unit, Handpiece, etc..) . Failure to maintain this product may lead to physical injury or property damage. [12] BEFORE ASKING FOR REPAIRS If any of phenomena described below has occurred, make the following checks before asking for repairs. Phenomenon The light does When an indicator not light up. is not displayed. When the indicator is on. Check point and result Action to be taken Unit main switch is not on. Turn on the unit main switch. Dental light main switch is not on. Turn on the dental light main switch to “ I ” mark. Light bulb had been burned out. Replace with a new light bulb. Refer to replacement of the bulb section in this manual. Wrong light bulb is used. Use specifed light bulb described in specification section in this manual. Touchless sensor surface is dirty. Clean the touchless sensor surface. Confirm distance of a touchless switch. Hand must pass by within 65mm from the touchless sensor. If the dental light does not normally work even if actions were taken upon checkup stated above, then stop using the unit, turn off the main switch and contact your dealer or our office. -12- [13] ELECTROMAGNETIC COMPATIBILITY(EMC) Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual. Portable and mobile RF communications equipment can affect medical electrical equipment. The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the equipment or system should be observed to verify normal operation in the configuration in which it will be used. Guidance and manufacture’s declaration - electromagnetic emissions The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the IO5000TA should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance RF emissions The IO5000TA uses RF energy only for its internal function. CISPR 11 Group 1 Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions The IO5000TA is suitable for use in all establishments, Class B CISPR 11 including domestic establishments and those directly Harmonic emissions connected to the public low-voltage power supply network Class A IEC 61000-3-2 that supplies buildings used for domestic purposes. Voltage fluctuations/ Flicker emissions Complies IEC 61000-3-3 Guidance and manufacture’s declaration - electromagnetic immunity The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the IO5000TA should assure that it is used in such an environment. IEC 60601 Electromagnetic environment Immunity test Compliance level test level guidance Electrostatic Floors should be wood, concrete or ± 6 kV contact ± 6 kV contact discharge (ESD) ceramic file. If floors are covered ± 8 kV air ± 8 kV air IEC 61000-4-2 with synthetic material, the relative humidity should be at least 30%. Electrical fast Mains power quality should be that ± 2 kV for power ± 2 kV for power transient/burst supply lines supply lines of a typical commercial or hospital IEC 61000-4-4 environment. ± 1 kV for input/output ± 1 kV for input/output lines lines Surge ± 1 kV differential mode ± 1 kV differential mode Mains power quality should be that IEC 61000-4-5 of a typical commercial or hospital ± 2 kV common mode ± 2 kV common mode environment. Voltage dips, short <5% UT <5% UT Mains power quality should be interruptions and (>95% dip in UT) (>95% dip in UT) that of a typical commercial or voltage variations for 0.5 cycle for 0.5 cycle hospital environment. If the user of on power supply 40% UT 40% UT the IO5000TA requires continued input lines (60% dip in UT) (60% dip in UT) operation during power mains IEC 61000-4-11 for 5 cycle for 5 cycle interruptions, it is recommended that 70% UT 70% UT the IO5000TA be powered from an (30% dip in UT) (30% dip in UT) uninterruptible power supply or a for 25cycle for 25cycle battery. <5% UT <5% UT (>95% dip in UT) (>95% dip in UT) for 5 s for 5 s Power frequency 3 A/m 3 A/m Power frequency magnetic fields (50/60 Hz) should be at levels characteristic magnetic field of a typical location in a typical IEC 61000-4-8 commercial or hospital environment. NOTE UT is the a.c. mains voltage prior to applications of the test level. -13- Guidance and manufacture’s declaration – electromagnetic immunity The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the IO5000TA should assure that it is used in such an environment. Compliance Immunity test IEC 60601 test level Electromagnetic environment - guidance level Portable and mobile RF communications equipment should be used no closer to any part of the IO5000TA, including cables, than the recommended separation distance calculated from the equation applications to the Frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz outside ISM bandsa 3 Vrms d = 1.2 √ P Radiated RF IEC 61000-4-3 3V/m 80 MHz to 2.5 GHz 3 V/m d = 1.2 √ P 80 MHz to 800 MHz d = 2.3 √ P 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by adsorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the IO5000TA is used exceeds the applicable RF compliance level above, the IO5000TA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the IO5000TA. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m. Essential performance (purpose of IMMUNITY testing) There is no essential performance. -14- Recommended separation distances between Portable and mobile RF communications equipment and the IO5000TA The IO5000TA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the IO5000TA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the IO5000TA as recommended below, according to the maximum output power of the communications equipment. Separation distance according to frequency of transmitter Rated maximum output power m of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz W d = 1.2 √ P d = 1.2 √ P d = 2.3 √ P 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by adsorption and reflection from structures, objects and people. -15- NOTE Takara Belmont (U.K.) Ltd. Belmont House One St.Andrews Way,Bow, London E3 3PA U.K. Tel: (44)20 7515 0333 Fax:(44)20 7987 3596 TAKARA BELMONT CORPORATION 2-1-1, Higashishinsaibashi,Chuo-ku, Osaka, 542-0083, Japan TEL : 81-6-6213-5945 FAX : 81-6-6212-3680 Book No. 1E014RC0 Printed in Japan 2012.04