DENTAL LIGHT
IO 5000TA
TYPE AL-820S
INSTALLATION
AND
OPERATION
IMPORTANT
After installation is completed, check all
the bolts, screws and fasteners to confirm
that they are securely fastened.
INSTRUCTIONS
TABLE OF CONTENTS
Page
[1] MAJOR PARTS IDENTIFICATION---------------------------------------------- 1
[2] SPECIFICATIONS-------------------------------------------------------------------- 1
[3] CLASSIFICATION-------------------------------------------------------------------- 1
[4] WIRING DIAGRAM----------------------------------------------------------------- 2
[5] DIMENSIONS-------------------------------------------------------------------------- 2
[6] INSTALLATION
6-1. Attach the light assembly to the light pole------------------------------------- 3
6-2. Attch the light pole to the light post----------------------------------------------- 3
6-3. Attach the wire harness----------------------------------------------------------- 3
[7] OPERATION
7-1. Mode selection switch------------------------------------------------------------ 4
7-2. Touch-less switch ----------------------------------------------------------------- 4
7-3. Intensity switch ---------------------------------------------------------------------- 4
7-4. Switching to the composite mode ------------------------------------------------- 4
[8] ADJUSTMENT OF THE BALANCE ARM
8-1. Adjustment of the tension of the balance arm--------------------------------- 5
8-2. Adjustment of the angle of the light head-------------------------------------- 5
8-3. Adjusting the tension of the light head -------------------------------------------- 6
8-4. Balance arm adjustment (lateral movement)-------------------------------------- 6
8-5. Converting the balance arm for left handed -------------------------------------- 6
[9] REPLACEMENT OF THE BULB
9-1. Removal of the reflector cover -------------------------------------------------- 8
9-2. Replacement of the bulb---------------------------------------------------------- 8
9-3. Inserting a new bulb -------------------------------------------------------------- 9
9-4. Attaching the reflector cover----------------------------------------------------- 9
[10] CLEANING---------------------------------------------------------------------------10
[11] MAINTENANCE AND INSPECTION------------------------------------------11
[12] BEFORE ASKING FOR REPAIRS----------------------------------------------12
[13] ELECTROMAGNETIC COMPATIBILITY (EMC)-------------------------13
Intended Use of the Product
This product is intended for the exclusive use for diagnoses, treatments and relative procedures of dentistry, and must be operated
or handled by the qualified dentists or by dental staffs under the supervision of the dentist. Such dentists or dental staffs should
instruct and/or assist the patients to approach to and leave from the product. Patients should not be allowed to operate or handle
the product unless he/she is so instructed.
Environmental condition for Operation
Temperature : 5 ~ 40℃
Humidity : 10 ~ 80%
Pressure : 600 ~ 1060 hpa
Environmental condition for Storage
Temperature : -10 ~ 50℃
Humidity : 10 ~ 80%
Pressure : 600 ~ 1060 hpa
Environmental condition for Transportation
Temperature : -10 ~ 50℃
Humidity : 10 ~ 80%
Pressure : 600 ~ 1060 hpa
Important Notes
In case of the troubles, please contat Takara Belmont offices or your dealers.
Do not disassemble or attempt to repair.
Disassembly, repair or modifications shoud only be done by a qualified repair technician.
Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.
In case of disposal of equipment
When disposing the dental light, appropriately dispose complying with all current applicable regulations and local codes.
In EU area, EU directive 2002/96/EC on waste electrical and electronic equipment (WEEE) is applied on this product. In this
directive, environment conscious recycling/abandonment is obligated.
Symbols
Alternating-current
㻨㻦
㻵㻨㻳
Authorized
representative in
the European
community
Separate
collection for
electrical and
electronic
equipment
Power ON
Manufacturer
Refer to instruction
manual/booklet
Power OFF
Date of manufacture
Non-ionizing
radiation
Caution
It means “caution,
warnings, or
possibility to
danger”.
WARNING
Operation of a dental light under Electro Magnetic Interferences.
This equipment may malfunction by electromagnetic interferences.
Do not install dental light close to the equipment that generates electromagnetic interference
(i.e. communication equipment, elevator)
Do not use the device that creates interference (i.e. cellular phone) near this equipment.
Do not use electric surgical knife or laser knife with dental light.
Light may turns on/off by itself due to malfunction of a sensor caused by electromagnetic interferences.
WARNING
• To avoid the risk of electric shock, this equipment must only be connected to supply mains with protective earth.
• Lamp cover becomes very hot during use.To avoid burning fingers, do not a touch lamp cover.
• Always put a lamp cover on to avoid burning fingers.
• Position a light head by holding light handle(s).
• Installation and service work should be conducted by an authorized installation/service personnel only.
WARNING: The followings are prohibited.
• To modify this equipment.
• To use the equipment under any failure condition.
• To use the equipment without doing the daily and periodical check-up.
• To wipe the plastic covers with any disinfectant or detergent that contains organic solvent.
Precautions for Installation
• Keep the equipment away from water.
• Keep in circumstances safe from influence by temperature, humidity, wind, sun light, air containing salts
and minerals.
• Care about stability such as inclination, vibration and impact, including handling and transportation.
• Do not keep the equipment in a place where chemicals are or where gas is emitted.
• During lifting and unpacking of the light, make sure to hold only the designated parts.
• Do not drop or hit the light.
• Ground light properly prior to turning power on.
• When the installation process has been completed, verify that all the mechanical and electrical functions
are working properly.
• Thick gloves are highly recommended at unpacking.
• Do not modify this equipment.
CAUTION
Before use
• Check connection of switches and make sure that the device functions properly.
• Make sure that grounding wire is connected.
• Make sure that cables are properly and perfectly connected.
During use
• Do not use the light longer than required for examination or treatment.
• Always watch the patient and the equipment to make sure nothing is wrong.
• If anything wrong is observed with the equipment or the patient, take a proper action, such as stopping
the use of equipment as well as keeping the patient in safe.
• Keep an eye on the patient not to touch the equipment.
After use
• Turn off the light.
• Clean the equipment and get it ready for use.
CAUTION
Do not spray liquids directly onto light surfaces.
In order to prevent damage to electrical components and systems, do not apply excess cleaning solution onto
light surfaces.
NOTE
Warranty does not cover damage to equipment caused by disinfectant solutions
Replacement of parts (Except the Bulb and Lamp cover)
Replacement must be done by an professional technician(s) of our company or a company authorized by us.
[1] MAJOR PARTS IDENTIFICATION
1
2
3
4
5
6
8
7
1
Light Pole Joint
2
Balance Arm Joint
3
Arm Joint Cover
4
Balance Arm
5
Arm Joint Cover
6
Yoke Joint
7
First Yoke
8
LED
9
Second Yoke
10 Intensity Switch
9
10
11 Handle (Left)
12 Touchless Sensor Switch
13 Reflector Cover
16
14 Light Bulb Cover
15 Handle (Right)
16 Yoke Rear Cover
15
13
14
18
11
17 Bracket Frame
18 Reflector
12
17
[2] SPECIFICATIONS
1. Power Requirement..................................................... AC23.7V 50/60Hz
2. Bulb Type..................................................................... Tungsten Halogen Type (JA 24V 60W)
3. Focal Distance............................................................. 700mm
4. Color Temperature....................................................... 5000Kelvin at 28,000 Lux
5. Light Intensity.............................................................. H : 28,000 Lux
M : 22,000 Lux
L : 8,000 Lux
6. Light Pattern................................................................ 60mm x 170mm - 84mm x 217mm
7. Net Weight................................................................... 5.2kg
8. Service Life.................................................................. 10 years
[3] CLASSIFICATION
a. Protection against electric shock : Class I Equipment
b. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
-1-
[4] WIRING DIAGRAM
Remark
Connector Symbol
NMF・・・New Mini Fit Connector
VL・・・VL Connector
YL・・・YL Connector
SV・・・SV Connector
VH・・・VH Connector
PH・・・PH Connector
NH・・・NH Connector
Black dot
Black dot indicates
male pin side
Plug
Mode Selection Switch
OFF
Cap
Sensor
Manual
CN7
2
1
3
CN8
2
3
4
1
CN2
5V 1
VH4P
Red
1
2
3
Red
Black
Brown
1A0190**
Lead Wire(Light Output)
1
2
Black
Brown
1
2
Red
Yellow
Orange
Purple
Blue
Green
SV2P
1 2
CN3
1
2
3
Low Intensity Indicate 4
GND 5
Intensity Selection
(Alternate)
Intensity Selection
(Momentary)
Hi Intensity Indicate
1A018K**
Inside Arm Lead Wire
CAGP81**
Lead Wire for
Lamp
Lead Wire for
Sensor SW
Setting condition
Change Intensity
Off : Disable On : Enable
2 Duration of time switching to Composite Mode
Off : 1second On : 2seconds
Intensity setting (When turning light on from off condition by using a sensor)
Off : Switch to composite mode when a hand is held in front of a sensor for
3
more than 2 seconds
On : Stay at normal treatment mode though a hand is held in front of a sensor
for more than 2 seconds
Baud Rate
4
Off : 4800bps On : 9600bps
Default
setting
1
OFF
ON
SW2
ON
On : When a hand move out
2 Composite Mode
Off : Enable On : Disable
Intensity setting switch
3
Off : Alternate On : Momentary
4
(42)
232.8
160°
20°
45°
160°
°
67.5
340
657
340
40°
40°
φ38.5
10.3
160
125
13
0°
°
10
-2-
OFF
ON
46.5
530
Default
setting
OFF
OFF
45°
369
Sensitivity Setting
OFF
[5] DIMENSIONS
(73.5)
CN1
NH3P
(CGGN64**)
Setting condition
Sensor activation method by movement of a hand
Off : When a hand move in
1 5V
2
3 GND
Sensor P.C.B.
(CGCN37**)
1
Lamp
Round Connectors
White/Gray
White
White/Yellow
3
2
1
1
2
3
4
5
6
LED P.C.B.
Setting of PCB Dip Switch
50
Faston Slide Connectors
White
White
1
2
YL2P
Sensor Input 2
GND 3
CN1
1 Power Input
2
SW1
Black
Dental Light PCB
Orange
1A0191**
Light SW Section(DL-820)
Head
Yoke
Arm
272
[6] INSTALLATION
6-1. Attach the light assembly to the light pole (Fig.6-1)
Pass the light cable through the light pole and insert the light pole joint into the light pole.
Fasten with two button screws (M6x8 ) .
Light Pole Joint
Button Screw (M6x8)
Light Pole
Light Cable
Light Head
Fig. 6-1
6-2. Attch the light pole to the light post (Fig.6-2)
Insert the light pole to the light post. Fasten with screws.
Please refer to a installation instruction of the unit.
Light Pole
Light Post
Fig. 6-2
6-3. Connect the wire harness (Fig.6-3)
Attach the connectors of the light cable onto the dental light PCB and harness inside the cuspidor unit.
Wire colours and connectors for connection are shown in Fig.6-3.
See also the wiriring diagram on page2.
2P Female
2P Male
1. Black
2. Bown
3P Female
1. Red
2. Yellow
3. -
Connect directly to a dental light PCB
5P Female
1. 2. Orange
3. Purple
4. Blue
5. Green
Fig. 6-3
-3-
Light Cable
[7] OPERATION
7-1. Touchless Switch (Fig.7-1)
Approach a hand to the touchless switch within the reaction
zone to turn on/off for the light.
LED
Intensity Switch
Touchless Sensor
Reaction Zone : 65mm
Fig. 7-1
7-2. Intensity Switch (Fig.7-2)
Push intensity switch.
High
High beam : Green LED turns on.
Low
Low beam : Amber LED turns on.
Light Intensity
28,000Lux
22,000Lux
LED
Green
Amber
Fig. 7-2
7-3. Switching to the composite mode (Fig.7-3)
During normal treatment mode, while the light is on, place
and hold your hand within the reaction zone for 2 seconds.
The light intensity will be automatically reduced to the
composite mode.
To switch back to the normal treatment mode, place your
hand by the sensor for 2 seconds again.
Green or Amber LED blinks in the composite mode.
7-1. Mode Selection switch
Mode can be changed by dental light select switch on the unit.
Please refer to operating instruction of the unit.
-4-
Composite
Light Intensity
Fig. 7-3
8,000Lux
LED
Blink
[8] ADJUSTMENT
8-1. Adjustment of the tension of the balance arm
Position the angle of the balance arm so the tension adjustment nut becomes visible just under slot A.
Turn the nut and adjust the tension with the adjustment bar supplied with the light.
Adjustment Bar
Balance Arm Joint
Light Pole Joint
Hole B
Slot A
Balance Arm
Tension Adjustment Nut
Slot C
Slot A (for Balance)
Direction of rotation of the nut
When the light head tends to go up, turn the nut to + direction.
When the light head tends to go down, turn the nut to – direction.
8-2 Adjustment of the angle of the light head
(1) Use a 2mm hex key wrench to loosen socket screw in front of Hole B.
CAUTION
Adjusting the head without loosening the socket screw will damage the arm.
(2) Position the angle of the balance arm so the head angle adjustment nut becomes visible just under the slot C.
Turn the nut and adjust the tension with the adjustment bar supplied with the light.
When the light head tilts upward,
turn the nut to + direction.
When the light head tilts downward,
turn the nut to - direction.
Slot C Hole B
Slot C
Hole B
Adjustment Nut
Socket Screw
(3) After adjustment is done, re-tighten the socket.
CAUTION
Unless this socket screw is re-tightened, the angle of the head will be changed during use.
-5-
8-3. Adjusting the tension of the light head
1. With a No.2 Phillips head screwdriver, unscrew the M3x5 screws, and remove the left and right yoke caps.
2. Adjust the tension by equally tightening or loosening the two socket head screws using the supplied 1.5mm
Allen key.
3. Upon completion of the adjustment, reattach the left and right yoke caps that were removed in Step 1 with
a No.2 Phillips head screwdriver and the M3x5 screws.
Socket Screw
Yoke Cap
Screw (M3x5)
Yoke Cap
CAUTION
Ensure that adjustments are made equally to both left and right sides.
Failure to do so may result in excess slack or damage.
8-4. Balance arm adjustment (lateral movement)
If the movement of the balance arm is either too loose or too tight, follow the procedures below to adjust.
1. Loosen the four M5 socket head screws (E) with the supplied 2.5mm Allen key.
CAUTION
When adjusting, make sure that the socket head screws are loosened.
Failure to do so may cause damage to the threads of the M8x20 screws.
Light Pole Joint
(D) M8x20 Screw (Top)
Balance Arm Joint
Balance Arm
Light Pole
(E) M5 Socket Screw x 4pcs.
(D) M8x20 Screw (Bottom)
2. Adjust the lateral tension of the balance arm by tightening or loosening the top and bottom M8X20
screws of the balance arm joint (D). (Use a No.3 (big) Phillips head screwdriver.)
3. Upon completion of the adjustment, retighten the four M5 socket head screws (E) that were loosened in
Step 1 with the supplied 2.5mm Allen key.
CAUTION
Upon completion of the adjustment, make sure to retighten the socket head screws.
Failure to do so may render the balance arm immobile or may damage the M8x20 screws.
-6-
8-5. Converting the balance arm for left handed use
When converting the balance arm from right to left handed use, follow the procedures below.
1. Loosen the four M5 socket head screws (E) of the balance arm joint with the supplied 2.5mm Allen key.
2. Unscrew the M8x20 screws (D), both top and bottom. (Use a No.3 (big) Phillips head screwdriver.)
Light Pole Joint
(D) M8x20 Screw (Top)
Balance Arm Joint
Balance Arm
Light Pole
(E) M5 Socket Screw x 4pcs.
(D) M8x20 Screw (Bottom)
3. Remove both M6 bolts that hold the light pole and light pole joint together with the supplied 3mm Allen
key.
4. Detach the balance arm joint from the light pole joint.
5. Rotate the light pole joint 180 degrees, then screw both M6 bolts back in using the supplied 3mm Allen
key.
M6 Button Bolt
Light Pole Joint
Light Cable
Light Pole
M6 Button Bolt
Light Pole Joint
Balance Arm Joint
6. Place a metal washer on the end of the upper side of the light pole joint, and move the balance arm in place.
7. Loosely tighten both top and bottom M8x20 (D) screws.
8. Tighten both top and bottom M8x20 (D) screws, and adjust the lateral tension of the balance arm.
9. To ensure that the four M5 bolts (E) of the light pole joint do not inadvertently become loose, securely
tighten them using the supplied 2.5mm Allen key.
Metal Washer
(D) M8x20 Screw (Top)
Ligh Pole Joint
(E) Socket Screw x 4pcs.
(D) M8x20 Screw (Bottom)
CAUTION
After adjustment, fasten the socket head screws.
Failure to do so may render the balance arm immobile or may damage the M8x20 screws.
-7-
[9] REPLACEMENT OF THE BULB
CAUTION
Make sure the power supply is turned off.
Halogen bulb and surrounding parts be hot immediately after the lamp goes off. Wait until they get cool down.
Do not touch glass with bare hand. Halogen bulb surface must be clean. Oil or body moisture will affect
bulb span of life. If glass surface is touched, clean with alcohol.
9-1. Removal of the reflector cover
(1) Turn the light bulb cover fixing screw counterclockwise by 90 degrees with a coin. See Fig.9-1-1
(2) Remove the reflector cover by pulling it toward you. See Fig.9-1-2
Light bulb cover fixing screw
Coin
Light bulb cover
Cut out
Reflector cover
Fig.9-1-1
Fig.9-1-2
Reflector cover
fixing gap
9-2. Replacement of the bulb
(1) Disconnect the connectors. See Fig.9-2-1
(2) Push the plastic-coated parts of the bulb retaining spring, and remove it from the hook. See Fig.9-2-2
Pull off
Flat connector
Plastic coated
Fig.9-2-1
Hook
Fig.9-2-2
(3) Remove the bulb. See Fig.9-2-3
Bulb
Fig.9-2-3
-8-
Bulb retaining spring
9-3. Inserting a new bulb
IMPORTANT
Replace halogen lamp only with type JA-24V 60W cbtaainable through your local dealer.
(1) At inserting the bulb, make sure that the projection of the bulb holder and the cut-out of the bulb are
aligned. Misalignment will cause an improper light pattern.
Cut-out of bulb
Projection
of the bulb holder
Fig.9-3-1
(2) Close the bulb retaining spring. Make sure to place the springs behind the hook. See Fig.9-3-2
(3) Firmly attach wire connectors. See Fig.9-3-3
Connect firmly
Fig.9-3-3
Fig.9-3-2
9-4. Attaching the reflector cover
(1) Insert the cut-out of the reflector cover into the gap at the handle (right). See Fig.9-4-1
(2) Be careful not to pinch the wires when attaching cover. See Fig.9-4-2
(3) Hold the reflector cover and tighten the screw by turning it clockwise by 90 degrees.
Make sure to tighten it by pushing it forward until it clicks. See Fig.9-4-3
Light bulb cover fixing screw
Light bulb cover fixing screw
Be careful not to pinch the wires
Thread hole
Cut out
* Push and tighten
Reflector cover
fixing gap
Fig.9-4-1
until it clicks
Fig.9-4-2
-9-
Fig.9-4-3
[10] CLEANING
CAUTION
Allow light to cool prior to cleaning.
CAUTION
All surfaces can be cleaned with DURR FD333 cleaner.
Spray the cleaner (DURR FD333) on cloth and wipe the surfaces with the cloth.
Wipe all surfaces dry after cleaning.
REFLECTOR : Extreme care should be taken to prevent scratching reflector surfaces, as this will degrade
the performance of the light.
-10-
[11] MAINTENANCE AND INSPECTION
11-1.Guide for daily maintenance and inspection (Maintenance and inspection by user)
Management of maintenance and inspection of medical equipment should be implemented by the user (medical
inspection). In case the user does not implement such management, it is permitted that such management is
outsourced to a qualified entity such as a medical equipment repair company.
For safe use of this product, it is necessary that inspection should be conducted in the specified frequency on
the items described below.
No. Item
Frequency
Inspection Method and diagnosis
Influence if inspection is
not conducted
Maintenance required
in case of
nonconformity
1
Appearance of Lamp
Cover
Before start
Check deformation, scratches by
eyes
Deterioration of optical
performance
2
Lamp cover is firmly
attached or not
Before start
Pull the lamp cover. Make sure that
the lamp cover doesn't fall
Lamp cover may fall.
Reattach a lamp
cover
3
Light bulb
Before start
Check whether the light brinks or not
due to the poor connection.
Light bulb turns very hot.
This could shorten bulb
life.
Securely insert a light
bulb into the bulb
holder
4
Light switch
Before start
Set Sensor/Manual switch to Manual.
Confirms the light turns on.
Light doesn't function.
Replace a light bulb
5
Touchless sensor
Before start
Set Mode Selection switch to Sensor.
Confirms the light turns on.
Light doesn't function.
Clean the sensor
surface.
6
Intensity switch
Before start
Change intensity.
It may cause the difficulty
at dental practice.
Turn off the light.Contact
your dealer or our office
7
Functionality of
Balance Arm
Before start
Swing the balance arm up/down.
Swing the balance arm right/left.
Make sure the balance arm stops and
holds its position.
Light head doesn't stay at
the desired position.
8
Vertical rotation of the
Light head
Before start
Rotate the light head up/down. Make
sure the light head stops and holds its
position.
Light head doesn't stay at
the desired position.
9
Light head angle
Before start
Confirms the light head is vertically
aligned.
Light head doesn't stay at
the desired position.
Adjust the angle of the
light head.
Light head rotation
Before start
Rotate the light head right/left. Make
sure the light rotate 160° to each
direction.
Light head rotates freely.
This may cause the
snapping of a wire.
Contact your dealer or
our office
Discoloration and
deterioration to the
exterior, and corrosion and
rusting to metallic
components may arise.
Executes wiping in
accordance with
"Cleaning
Instructions"
Light may not function
right.
Turn off the light.Contact
your dealer or our office
10
11
Care exterior
Before start
Chemical and dirt on the product
exterior must be cleaned.
12
Other 1
Once every
week
Make sure that no abnormal noise
occurs when the product is operated.
13
Other 2
㻃As needed
If the light has not been used for a
long time, make sure the light
functions correctly and safely.
-11-
Change a lamp cover
Adjust the tension of
the balance arm.
Adjust the tension of
vertical rotation.
If the light
malfunctions, Contact
your dealer or our office
11-2. Guide for Periodical Check-up
Some parts and components of the products are degraded or deteriorated depending on the frequency of
use. Annual check-up and maintenance, as well as replacement of consumable parts, are required.
The required parts (including consumable parts) are listed below. It may be different from the following
list depending on the option of the light.
For check-up and repair, call a technician of our authorized dealer.
Parts and components that require periodical check-up
Standard
Lifetime
No.
Moving part
7 years
3
Switches
5 years
Electric wiring of moving part
5 years
4
Control PCBs.
5 years
No.
Parts Description
1
2
Parts Description
Standard
Lifetime
Consumable parts
No . Parts Description
1
Bulb
2
Lamp Cover
WARNING
Execute the maintenance in accordance with this instraction manual and operating manual attached
to each individual equipment ( Dental unit, Handpiece, etc..) .
Failure to maintain this product may lead to physical injury or property damage.
[12] BEFORE ASKING FOR REPAIRS
If any of phenomena described below has occurred, make the following checks before asking for repairs.
Phenomenon
The light does When an indicator
not light up.
is not displayed.
When the indicator
is on.
Check point and result
Action to be taken
Unit main switch is not on.
Turn on the unit main switch.
Dental light main switch is not on.
Turn on the dental light main switch to
“ I ” mark.
Light bulb had been burned out.
Replace with a new light bulb.
Refer to replacement of the bulb
section in this manual.
Wrong light bulb is used.
Use specifed light bulb described in
specification section in this manual.
Touchless sensor surface is dirty.
Clean the touchless sensor surface.
Confirm distance of a touchless
switch.
Hand must pass by within 65mm from
the touchless sensor.
If the dental light does not normally work even if actions were taken upon checkup stated above, then
stop using the unit, turn off the main switch and contact your dealer or our office.
-12-
[13] ELECTROMAGNETIC COMPATIBILITY(EMC)
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect medical electrical equipment.
The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
Guidance and manufacture’s declaration - electromagnetic emissions
The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the
IO5000TA should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
The IO5000TA uses RF energy only for its internal function.
CISPR 11
Group 1
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
The IO5000TA is suitable for use in all establishments,
Class B
CISPR 11
including domestic establishments and those directly
Harmonic emissions
connected to the public low-voltage power supply network
Class A
IEC 61000-3-2
that supplies buildings used for domestic purposes.
Voltage fluctuations/
Flicker emissions
Complies
IEC 61000-3-3
Guidance and manufacture’s declaration - electromagnetic immunity
The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the
IO5000TA should assure that it is used in such an environment.
IEC 60601
Electromagnetic environment Immunity test
Compliance level
test level
guidance
Electrostatic
Floors should be wood, concrete or
± 6 kV contact
± 6 kV contact
discharge (ESD)
ceramic file. If floors are covered
± 8 kV air
± 8 kV air
IEC 61000-4-2
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
Mains power quality should be that
± 2 kV for power
± 2 kV for power
transient/burst
supply lines
supply lines
of a typical commercial or hospital
IEC 61000-4-4
environment.
± 1 kV for input/output
± 1 kV for input/output
lines
lines
Surge
± 1 kV differential mode ± 1 kV differential mode Mains power quality should be that
IEC 61000-4-5
of a typical commercial or hospital
± 2 kV common mode
± 2 kV common mode
environment.
Voltage dips, short
<5% UT
<5% UT
Mains power quality should be
interruptions and
(>95% dip in UT)
(>95% dip in UT)
that of a typical commercial or
voltage variations
for 0.5 cycle
for 0.5 cycle
hospital environment. If the user of
on power supply
40% UT
40% UT
the IO5000TA requires continued
input lines
(60% dip in UT)
(60% dip in UT)
operation during power mains
IEC 61000-4-11
for 5 cycle
for 5 cycle
interruptions, it is recommended that
70% UT
70% UT
the IO5000TA be powered from an
(30% dip in UT)
(30% dip in UT)
uninterruptible power supply or a
for 25cycle
for 25cycle
battery.
<5% UT
<5% UT
(>95% dip in UT)
(>95% dip in UT)
for 5 s
for 5 s
Power frequency
3 A/m
3 A/m
Power frequency magnetic fields
(50/60 Hz)
should be at levels characteristic
magnetic field
of a typical location in a typical
IEC 61000-4-8
commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to applications of the test level.
-13-
Guidance and manufacture’s declaration – electromagnetic immunity
The IO5000TA is intended for use in the electromagnetic environment specified below. The customer or the user of the
IO5000TA should assure that it is used in such an environment.
Compliance
Immunity test
IEC 60601 test level
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the IO5000TA,
including cables, than the recommended separation
distance calculated from the equation applications to
the Frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bandsa
3 Vrms
d = 1.2 √ P
Radiated RF
IEC 61000-4-3
3V/m
80 MHz to 2.5 GHz
3 V/m
d = 1.2 √ P 80 MHz to 800 MHz
d = 2.3 √ P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a should
be less than the compliance level in each frequency
range.b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1 At 80 MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by adsorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the IO5000TA is used exceeds the applicable RF
compliance level above, the IO5000TA should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the IO5000TA.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Essential performance (purpose of IMMUNITY testing)
There is no essential performance.
-14-
Recommended separation distances between
Portable and mobile RF communications equipment and the IO5000TA
The IO5000TA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the IO5000TA can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the IO5000TA as recommended
below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output power
m
of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
W
d = 1.2 √ P
d = 1.2 √ P
d = 2.3 √ P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by adsorption and
reflection from structures, objects and people.
-15-
NOTE
Takara Belmont (U.K.) Ltd.
Belmont House
One St.Andrews Way,Bow,
London E3 3PA U.K.
Tel: (44)20 7515 0333
Fax:(44)20 7987 3596
TAKARA BELMONT CORPORATION
2-1-1, Higashishinsaibashi,Chuo-ku, Osaka, 542-0083, Japan
TEL : 81-6-6213-5945
FAX : 81-6-6212-3680
Book No. 1E014RC0
Printed in Japan 2012.04