Aspire HD Quality Control Program Manual Quality Control 1 Overview 2 Installation of FDR Mammography QC Program 3 Weekly Test 4 Quarterly Test 5 Semi-annual Test 6 Annual Test 7 Assistance Material 8 Technical Information 9 2nd Edition - October 2012 This Manual provides detailed information about how to use the FDR Mammography QC Program as well as important points to note when using it. Before using this product, be sure to read this Manual thoroughly. After reading this Manual, store it nearby so that you can refer to it whenever necessary. Please also read “FDR-1000DRSZ Operation Manual”, “FDR1000AWS Operation Manual”, “FDR Mammography QC Software Operation Manual” and “FCR 1Shot Phantom M Plus Operation Manual”. 897N101461A FDR Mammography QC Manual - 897N101461A Image Processing Parameters (for Mammography QC) ii FDR Mammography QC Manual - 897N101461A Introduction The FDR Mammography QC Program Manual (the “Manual” hereafter) provides the procedures for quality control and constancy test, technical explanation and other information necessary for managing the quality of the FDR digital mammography system. This quality control program primarily uses quantitative measurements and provides the ability to see gradual changes in X-ray equipment performances, as these changes are difficult to notice by visual checks alone. Exclusive Clauses 1. No part or all of this Manual (except Chapter 8) may be reproduced in any form without prior permission. 2. The information contained in this Manual may be subject to change without prior notice. 3. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from installation, relocation, remodeling, maintenance, and repair performed by other than dealers specified by FUJIFILM Corporation. 4. FUJIFILM Corporation shall not be liable for malfunctions and damages of FUJIFILM Corporation products due to products of other manufacturers not supplied by FUJIFILM Corporation. 5. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from remodeling, maintenance, and repair using repair parts other than those specified by FUJIFILM Corporation. 6. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from negligence of precautions and operating methods contained in this Manual. 7. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from use under environment conditions outside the range of using conditions for this product such as power supply, installation environment, etc. contained in this Guidebook. 8. FUJIFILM Corporation shall not be liable for malfunctions and damages resulting from natural disasters such as fires, earthquakes, floods, lightning, etc. Trademark FCR and FDR are trademarks or registered trademarks of FUJIFILM Corporation. Copyright © 2008 FUJIFILM Corporation. All rights reserved. FDR Mammography QC Manual - 897N101461A iii iv FDR Mammography QC Manual - 897N101461A Contents at a Glance Chapter 1 Quality Control 1 This chapter describes the relevant quality control standards. Chapter 2 Overview 2 This chapter describes the features of FDR Mammography QC Program and the purpose of each quality control test. Chapter 3 Installation of FDR Mammography QC Program 3 This chapter describes the operational procedure for installing the FDR Mammography QC Program. Chapter 4 Weekly Test 4 This chapter describes the Weekly Test procedure performed by the Technologist (and during MEE and Annual testing by the Medical Physicist). Chapter 5 Quarterly Test 5 This chapter describes the Quarterly Test procedure performed by the Technologist. Chapter 6 Semi-annual Test 6 This chapter describes the Semi-annual Test procedure performed by the Technologist. Chapter 7 Annual Test 7 The test procedures performed by the Medical Physicist during the Medical Equipment Evaluation (MEE) and Annual testing. Chapter 8 Assistance Material 8 This chapter consists of material that helps with quality control tests and includes explanations of the calculation functions, software operation quick guides and the worksheets. Chapter 9 Technical Information 9 This chapter consists of a brief specification of 1Shot Phantom M Plus and a glossary. FDR Mammography QC Manual - 897N101461A v Indications The following indications are used in the descriptions to show the supplementary information and what must be observed while using the software. NOTE Indicates operational information that should be noted. TIP Indicates operational information that may be helpful. Indicates an item that provides details of the procedure or related information. Indicates a procedure that is dependent on the situation. Conventions The names of the buttons in the window are shown as described below. (Example) vi [Start study.] The button name is shown in brackets ([ ]) regardless of the feature. FDR Mammography QC Manual - - 897N101461A Contents Chapter Chapter 1 Quality Control 1.1 Quality Control ................................................................................................................ 1.1.1 Acceptance Test ............................................................................................ 1.1.2 Constancy Test ............................................................................................... 1.1.3 Status Test .......................................................................................................... 1-2 1-2 1-2 1-2 1.2 Quality Control of the Mammography System .................................................... 1.2.1 Quality Control Tests and Frequency .......................................... 1.2.2 Types of Quality Control Tests .......................................................... 1-3 1-3 1-4 2 Overview 2.1 Product Outline .............................................................................................................. 2-2 2.2 QC Test Items.................................................................................................................... 2-5 2.2.1 Weekly Test........................................................................................................ 2-7 2.2.2 Quarterly Test .................................................................................................. 2-11 2.2.3 Semi-annual Test .......................................................................................... 2-11 2.2.4 Annual Test ........................................................................................................ 2-12 2.3 Tools ......................................................................................................................................... 2-13 2.4 QC Software Outline................................................................................................... 2-17 2.5 Notes on Conducting the Program .................................................................... 2-20 Chapter 3 Installation of FDR Mammography QC Program 3.1 Installation Procedure ............................................................................................. 3-2 3.2 Initial Settings .................................................................................................................. 3-3 3.3 Criteria Confirmation and Determination................................................................................. 3-5 3.4 Baseline Value Settings ............................................................................................ 3.4.1 Baseline Values for Annual Test ....................................................... 3.4.2 Baseline Values for Semi-annual Test.......................................... 3.4.3 Baseline Values for Weekly Test ....................................................... 3-12 3-13 3-17 3-21 3.5 Checking Equipment Conditions at the Time of Program Installation .......................................................................................................................... 3-29 FDR Mammography QC Manual - 897N101461A vii Chapter 4 Weekly Test 4.1 Test Flow .............................................................................................................................. 4-2 4.2 Test Items............................................................................................................................. 4-3 4.3 Tools ......................................................................................................................................... 4-4 4.4 Setting Confirmation ................................................................................................. 4-5 4.5 1 Shot Phantom Test 4-6 ................................................................................................ 4.6 ACR Phantom Test ....................................................................................................... 4-12 4.7 Printer Quality Control ............................................................................................. 4-18 4.8 Monitor Quality Control .......................................................................................... 4-20 Chapter Chapter Chapter viii 5 Quarterly Test 5.1 Test Flow .............................................................................................................................. 5-2 5.2 Test Items............................................................................................................................. 5-3 5.3 Repeat Analysis ............................................................................................................... 5-4 6 Semi-annual Test 6.1 Test Flow .............................................................................................................................. 6-2 6.2 Test Items............................................................................................................................. 6-2 6.3 Tools ......................................................................................................................................... 6-2 6.4 Compression Device Confirmation ............................................................... 6-2 7 Annual Test 7.1 Test Flow .............................................................................................................................. 7-2 7.2 Test Items............................................................................................................................. 7-3 7.3 Tools ......................................................................................................................................... 7-4 7.4 Setting Confirmation ................................................................................................. 7-5 7.5 Conducting Annual Test ........................................................................................ 7.5.1 Initial Performance Test .......................................................................... 7.5.1 Image Performance Test ........................................................................ 7.5.2 Spatial Resolution Test ............................................................................ 7.5.3 X-ray Equipment Performance Test ............................................. 7.5.4 AEC System Performance Test .......................................................... 7-7 7-7 7-19 7-27 7-31 7-45 FDR Mammography QC Manual -897N101461A Chapter 8 Assistance Material 8.1 Calculation Functions................................................................................................ 8-2 8.2 Quick Guide for Software Operation .......................................................................... 8-6 8.3 Worksheets ......................................................................................................................... 8-8 8.4 Report Forms .................................................................................................................... 8-20 Chapter 9 Technical Information 9.1 Specification Outline of 1Shot Phantom M Plus ................................ 9-2 9.2 Glossary ................................................................................................................................. 9-3 FDR Mammography QC Manual - 897N101461A ix Revision History Revision x Issue Date Reason First Edition 897N101461 11.2011 New Release 2nd Edition 897N101461A 10.2012 1) Revised Linearity/Beam Quality constancy test criteria for QL gap (4 step-5 step). 2) Revised corrective action limits for 1 Shot Phantom m Plus tests. 3) Revised compression plate heights for AGD Mode 1 testing. FDR Mammography QC Manual - 897N101461A Chapter 1 1 Quality Control Page 1.1 Quality Control ________________________________________ 1.1.1 1.1.2 1.1.3 1.2 Quality Control of the Mammography System_____________ 1.2.1 1.2.2 FDR Mammography QC Manual - 897N101461A Acceptance Test __________________________________ Constancy Test ___________________________________ Status Test _______________________________________ Quality Control Tests and Frequency__________________ Types of Quality Control Tests _______________________ 1-2 1-2 1-2 1-2 1-3 1-3 1-4 1-1 1.1 Quality Control This document provides information necessary for a facility using the FDR Mammography system to maintain an effective QA & QC program and meet the requirements of the MQSA regulations and Fujifilm Corp. 1 Quality Control Detailed instructions for carrying out quality control (e.g. when and who carries out quality control) are established as a quality assurance program. In addition to quality control techniques, training for providing adequate information on quality control is included so that any quality assurance program may be effectively implemented. International standards and guidelines regarding the quality control of the mammography system are as follows. • MQSA (Mammography Quality Standards Act) 21CFR Part 900 (hereafter MQSA). • ACR (American College of Radiology) Quality Control Manual • European guidelines for quality assurance in breast cancer screening and diagnosis; Fourth Edition (European Commission) (Hereafter abbreviated as EUREF Ed.4) • IEC 61223-3-2 Ed.2 (International Electrotechnical Commission) Always follow applicable laws and regulations for your jurisdiction. If anything in this manual is in conflict with applicable laws or regulations, the applicable law or regulation shall take precedence Tests for quality control are called performance tests. There are three types of performance tests, acceptance test, constancy test and status test, depending on their purpose or implementation frequency. 1.1.1 Acceptance Test The purpose of the acceptance test is to check the compliance of the equipment with specifications. An acceptance test is normally carried out when new equipment has been installed, when existing equipment has been remodeled, or when required by local juisdictions. 1.1.2 Constancy Test The constancy test is intended to monitor the constancy of the functional performance of the equipment by means of a test method that is simple, quick and easy to carry out, usually involving measurements of relative values. In FDR Mammography QC Program (Program hereafter), all tests are regarded as constancy tests. For some particular tests, the upper or lower limit is established by a law or guideline. 1.1.3 Status Test A status test is executed when components or sub-assemblies have been added, remodeled, replaced, or removed, or when the results of a constancy test indicate a substantial change in equipment performance. 1-2 FDR Mammography QC Manual - 897N101461 1.2 Quality Control of the Mammography System In this document, test items and their frequencies for the FDR Mammography system are deteremined so that facilities may be in compliance with the requirements of MQSA. Test items consist of quantitive tests and visual inspections, and the FDR 1Shot Phantom M Plus for quantitive measurement and QC Software are used for evaluation. Quality Control 1.2.1 Quality Control Tests and Frequency QC Software Installation Baseline Value Measurement Equipment Condition Check QC Software Operation 3 months 6 months 9 months Mammography E q u i p m e n t Evaluation (MEE) 1 year 15 months 18 months 21 months At time of FDR installation / major component replacement 1 2 years Chapter 7 Chapter 4 Every Week Weekly Test Chapter 4 Quarterly Test Chapter 5 Semi-annual Test Chapter 6 Chapter 7 Annual Test Chapter 4 Chapter 3 [1] Mammography Equipment Evaluation (MEE) - Medical Physicist The MEE tests should be performed when a new FDR Mammography system has been installed, and whenever changes that might affect performance (disassembly, major component repair, etc.) have been made to an existing FDR Mammography system. [2] Weekly Test - Technologist Weekly Tests using the FCR 1Shot Phantom M Plus and ACR Phantom are executed for detecting changes in image quality or subtle changes of the mammography system that may not reflected in actual images. These tests are executed as constancy tests for monitoring weekly changes in the mammography system. Quantitive tests can be executed by using the FCR 1Shot Phantom M Plus. However, a control limit needs to be determined for each quantitive test based on a series of measurements executed when installing this Program. For better performance, it is important that these control limits (criteria) are reviewed consistently taking into account other factors such as the scores of the tests using the ACR Phantom. [3] Quarterly Test - Technologist The Quarterly Test is designed to determine the number and cause of repeated radiographs. Analysis of this data will help identify ways to avoid repeat exposures and reduce costs, as well as reduce patient exposure. 4] Semi-annual Test - Technologist During the Semi-annual Test, performance verification tests of the compression system are executed.. [5] Annual Test- Medical Physicist During Annual Tests, performance verification tests of the components of the mammography system are executed using measuring instruments and materials such as a dosimeter and PMMA phantom. FDR Mammography QC Manual - 897N101461A 1-3 1.2.2 Types of Quality Control Tests All test items for quality control of the mammography system focus on change of measurement values. Accordingly, all test items are classified as constancy tests. However, there are two types of tests depending on evaluation methods. 1 [1] Constancy Test &AIL %STABLISHED #RITERIA 4EST2ESULTS 0ASS "ASELINE6ALUE &AIL )NSTALLATION -ONTHS 9EAR [2] Performance Verification Test In performance verification tests, the upper or lower limit is specified by a law or guideline. These tests need to be passed irrespective of the type of X-ray equipment. Items of the performance verification test include scores of the tests using the ACR Phantom and measured values of Missed tissue on chest wall edge). 7HENTHEUPPERLIMITISSPECIFIED &AIL 5PPER,IMIT 4EST2ESULTS )NSTALLATION 0ASS -ONTHS 9EAR 7HENTHELOWERLIMITISSPECIFIED 0ASS 4EST2ESULTS ,OWER,IMIT &AIL )NSTALLATION 1-4 -ONTHS 9EAR FDR Mammography QC Manual - 897N101461A 1 Quality Control Quality Control Constancy tests are intended to monitor the constancy of the functional performance of the equipment. Since the performance varies depending on the types of the X-ray equipment and measurement environment, the same test is executed multiple times and its average value is defined as a baseline value, and the variation of measured values is defined as an established criteria. When the measured value is within the acceptable range (baseline value plus or minus criteria), the test result is regarded as “Pass”. If it is out of the acceptable range, the test result is regarded as “Fail”. Chapter 2 2 Overview Page 2.1 2.2 Product Outline ________________________________________ 2-2 QC Test Items __________________________________________ 2-5 2.2.1 2.2.2 2.2.3 2.2.4 FDR Mammography QC Manual - 897N101461A Weekly Test ______________________________________ 2-7 Quarterly Test ____________________________________ 2-11 Semi-annual Test _________________________________ 2-11 Annual Test ______________________________________ 2-12 2.3 Tools _________________________________________________ 2-13 2.4 QC Software Outline____________________________________ 2-17 2.5 Notes on Conducting the Program _______________________ 2-20 2-1 2.1 Product Outline This Program is a dedicated program developed for implementing quality control of the FDR mammography system. Conducting this Program enables appropriate quality control on a FDR system in medical institutions to be carried out through the Weekly, Quarterly, Semi-annual and Annual Tests. • The Program consists of the following three components. • 2 Manual (FDR Mammography QC Program) It (this document) provides instructions for implementing quality control of the FDR mammography system. • Overview Exclusive Phantom (1Shot Phantom M Plus) 1Shot Phantom M Plus (1Shot Phantom hereafter) was developed for quality control of the FDR mammography system and can produce the image quality test results on 10 items with a single exposure. The 1Shot Phantom enables wide-ranging analysis of a system in a relatively short time with high reproducibility. • Exclusive Quality Control Software (FDR Mammography QC Software) FDR Mammography QC Software (hereafter QC Software) is designed for performing periodic quality control, data analysis, and QC test log management. [QC test categories] • • • • Weekly Test ........................................ Quantitative/Visual inspection with 1Shot Phantom/ ACR Phantom Quarterly Test ....................................Repeat analysis Semi-annual Test .............................Quantitative/Visual inspection Annual Test ........................................Quantitative/Visual inspection Comprehensive quality control on the FDR mammography system can be ensured by conducting these periodical tests and validating the results. The QC Software allows the auto-calculation of the CNR and AGD in addition to that of the items in each test. [Calculation functions] • • 2-2 CNR..........Used for calculating the values in AEC (Automatic Exposure Control) function adjustment or when determining the CNR baseline values. AGD .........Used for calculating the values in AEC. FDR Mammography QC Manual - 897N101461A [Equipment to be evaluated] 2 Overview [Test categories and their implementation frequencies] The QC test items are categorized by the required implementation frequency. The test categories, their implementation frequency and personnel responsibilities are as shown below. QC Software Installation QC Software Operation Equipment 3 Baseline Value Condition Measurement months Check Mammography Equipment Evaluation (MEE) 6 9 1 15 18 21 2 months months year months months months years At time of FDR mammography installation / major component replacement Chapter 7 Chapter 4 (Medical Physicist) Weekly Test Every Week (Technologist) Chapter 4 Quarterly Test Chapter 5 (Technologist) Semi-annual Test Chapter 6 (Technologist) Annual Test Chapter 7 (Medical Physicist) Chapter 4 Chapter 3 See Chapters 3 to 7 as necessary for details on each test procedure. FDR Mammography QC Manual -897N101461A 2-3 [The Program implementation procedure] 2 Overview 2-4 FDR Mammography QC Manual - 897N101461A 2.2 QC Test Items Test components in each test category are shown below. “Exposure Menus” are the menu items to be selected on a FUJIFILM workstation to start the corresponding tests. [Weekly Test] Test Items Exposure Menus Responsibility Test with 1Shot Phantom 1Shot PhantomM Good practice Test with ACR Phantom Technologist 2 ACR Phantom Conduct the Test with 1Shot Phantom and Good practice in the Daily Test, and the Test with ACR Phantom in the Weekly Test. Overview NOTE [Quarterly Test] Test Items Repeat analysis Exposure Menus Responsibility –– *1 Technologist *1 No Exposure Menu for the Repeat analysis in Quarterly Test [Semi-annual Test] Test Items Exposure Menus Responsibility Compression device confirmation –– *2 Technologist *2 No Exposure Menu for the Compression device confirmation in the Semi-annual Test FDR Mammography QC Manual - 897N101461A 2-5 [Annual Test] Test Items Image Basic Test Exposure Menus Responsibility Test Types Annual A Additive lag effects (Lag) Image performance Multiplicative lag effects (Ghost) Annual 1/6 Visual and Functional test Spatial Resolution (Magnification) 2 Annual 2/6 Spatial Resolution kVp accuracy and reproducibility Half Value Layer (HVL) Overview Annual 3/6 Medical Physicist Collimation assessment X-ray equipment performance Radiation output AEC reproducibility CNR Mode 1 Annual 4/6 AEC system performance 1 Shot Phantom M ACR Phantom Image Performance AGD Mode 1 Test with 1 Shot Phantom ACR Phantom Evaluation [Calculation functions] Test Items Exposure Menus CNR Calculation 1/2 AGD Calculation 2/2 When an exposure menu includes multiple test items/contents, perform all tests. (If any of the test items/contents remain untested, the QC Software will not calculate the results.) An exposure menu can be selected on the following window of a FUJIFILM workstation. When an exposure menu is selected, the relevant exposure submenu(s) is displayed (Example: 1Shot PhantomM). See the Operation Manual for FUJIFILM workstation for detailed FUJIFILM workstation operations. Explanations of test items/contents are shown on the following pages. 2-6 FDR Mammography QC Manual - 897N101461A 2.2.1 Weekly Test The Weekly Test is conducted to evaluate image quality by making an exposure using a 1Shot Phantom and the ACR Phantom, and to check that the X-ray equipment normally used in clinical practice is kept clean. The 1Shot Phantom is designed to identify subtle changes in FDR Mammography system performance. With a single exposure using this Phantom, the QC Software auto-calculates the necessary image quality parameters, enabling image quality variations to be monitored (some test items require visual inspection). As required by MQSA regulations, image evaluation using the ACR Phantom is to be conducted at least once a week., and, when used in combination with the QC Software, the results can be managed efficiently. 2 Overview * NOTE: The ACR phantom density measurement is required to be performed weekly if the facility routinely prints hardcopy images for primary interpretation. If images are only occasionally printed for final interpretation, such as upon request of a patient for images for interpretation elsewhere, then the test must be performed in advance of, but not more than one week before printing the films. FDR Mammography QC Manual -897N101461A 2-7 Phantoms Used for Exposure [1] Test with 1Shot Phantom (1) Missed tissue on chest wall edge 1Shot Phantom 2 ACR Phantom/Acrylic disk Overview NOTE The ACR phantom density measurement (using the acrylic disk) is required to be performed weekly if the facility routinely prints hardcopy images for primary interpretation. If images are only occasionally printed for final interpretation, such as upon request of a patient for images for interpretation elsewhere, then the test must be performed in advance of, but not more than one week before printing the films. Check that the degree of missed tissue on chest wall edge of the exposure table is within the criteria. (2) CNR Check that the CNR variation is within the criteria. 2-8 FDR Mammography QC Manual - 897N101461 (3) 1Shot Phantom sensitivity constancy (6) Uniformity 2 Check that the uniformity variation is within the criteria. (4) Geometric distortion (7) Dynamic range Overview Check that the 1Shot Phantom sensitivity variation is within the criteria. (ORIZONTAL DIRECTION 6ERTICAL DIRECTION Check that the image dimension variation (geometric distortion) in horizontal/vertical direction is within the criteria. Visually check that there is no distortion or jitter in the image. Check that the dynamic range variation is within the criteria. (8) Spatial Resolution (SR) (5) System artifact evaluation Check that the spatial resolution variation is within the criteria. Visually check that there are no artifacts affecting diagnosis in the image. FDR Mammography QC Manual - 897N101461 2-9 (9) Low Contrast Detectability (LCD) [2] Good practice Check that the following equipment is kept clean. • X-ray equipment • Softcopy output equipment • Viewbox [3] Test with ACR Phantom 2 Overview Check that the low contrast detectability meets or exceeds the criteria. (10) Linearity/Beam quality constancy Make exposures using the ACR Phantom and then output the images to check that the densities of the specified areas in the images are within the criteria*. In addition, visually check and evaluate the images. • Density at center of Phantom image* • Density difference (disk/outside)* • Fibers • Specks • Masses * NOTE Check that the X-ray linearity and radiation quality variations are within the criteria. 2-10 ACR phantom density measurements are required to be performed weekly if the facility routinely prints hardcopy images for primary interpretation. If images are only occasionally printed for final interpretation, such as upon request of a patient for images for interpretation elsewhere, then the test must be performed in advance of, but not more than one week before printing the films. FDR Mammography QC Manual - 897N101461A 2.2.2 Quarterly Test During the Quarterly Test, Repeat analysis is performed. [1] Repeat analysis 2.2.3 Semi-annual Test During the Semi-annual Test, Compression device confirmation is performed.. [1] Compression device confirmation Calculate data for analyzing the rejected images. 2 Overview Check that minimum and maximum compression forces can be applied. FDR Mammography QC Manual - 897N101461A 2-11 2.2.4 Annual Test The Annual Test contains test items [1] to [12] for X-ray equipment performance and test item [13] for display device performance. 2 Overview 2-12 FDR Mammography QC Manual - 897N101461A 2.3 Tools Tools to be used for QC test are shown in the table below. ...Required Tools Weekly Test Semi-annual Test Annual Test ...Optional Calculation Functions [ A ] QC Software [ B ] Worksheet (See Chapter 8) 2 [ C ] 1Shot Phantom [ D ] Viewbox (if film is used) [F] Overview [ E ] Timer or watch/clock ACR Phantom (including an acrylic disk) *1 [ G ] Dosimeter [ H ] Luminance meter [ I ] kVp meter [ J ] Force scale / Bathroom scale [ K ] Towels *2 *2 [ L ] Illuminance meter [M] Densitometer PMMA phantoms (available, in [ N ] combination, for 20, 40, 60 and 70 mm in thickness) [O] Aluminum plate (0.2 mm) for CNR measurement Aluminum plates for half value layer [ P ] measurement (0.3 and 0.5 mm) [ Q ] Scale (ruler) [ R ] Coins [ S ] Lead sheet *3 *3 *1 The ACR phantom density measurement is required to be performed weekly if the facility routinely prints hardcopy images for primary interpretation. If images are only occasionally printed for final interpretation, such as upon request of a patient for images for interpretation elsewhere, then the test must be performed in advance of, but not more than one week before printing the films. *2 Used to protect the exposure table and compression plate during Compression device confirmation *3 Placed on the exposure table to protect the image receptor when measuring the air kerma. The following equipment is used for the QC tests: A. FDA-cleared Hardcopy/Softcopy output equipment (laser imager/diagnostic monitor) B. X-ray equipment C. FUJIFILM workstation FDR Mammography QC Manual - 897N101461A 2-13 Tool Details [A] QC Software The Software is designed for performing periodic quality control, data analysis, and QC test log management. This is to be installed and used on a FUJIFILM workstation. [B] Worksheet Worksheets for recording or documenting QC test measurement and judgment results are contained in “8.3 Worksheets”. Make a copy of those worksheets as necessary. [C] 1Shot Phantom 2 The 1Shot Phantom enables an image quality test on 10 items by a single exposure, allowing analysis of the wide-ranging system quality in a short time with high reproducibility. NOTE Overview Scratches, deformation, or dirt on the Phantom not only affects the image evaluation or measured values but may damage the compression plate of the X-ray equipment. Make sure that there is no scratch, deformation, or dirt on the Phantom before use. [D] Viewbox A viewbox is used for visually checking an image output on film. Maintenance including surface cleaning or lamp replacement should be performed by following the manufacturer’s instructions. If using a viewbox having a luminance adjustment function, make sure to keep the luminance constant for every test. [E] Timer or watch/clock A timer or watch/clock is used for measuring the duration of the compression plate. [F] ACR Phantom The Phantom is used for image evaluation. The model 156 manufactured by RMI (Radiation Measurements Inc.), 18-220 by NA (Nuclear Associates), 15 by CIRS (Computerized Imaging Reference System, Inc.) or equivalent is necessary. The above mentioned Phantoms are qualified by ACR (American College of Radiology). NOTE Scratches, deformation, or dirt on the ACR Phantom not only affects the image evaluation or measured values but may damage the compression plate of the X-ray equipment. Make sure that there are no scratches, deformation, or dirt on the Phantom before use. [G] Dosimeter A dosimeter is used for measuring air kerma (exposure). Some dosimeters may require correction of the measured value according to the beam quality. See the operation manual for the dosimeter for details. The dosimeter should be calibrated using low energy (at least 10 keV to 40 keV), allowing air kerma rate or exposure time measurement, in addition to air kerma (exposure) measurement. [H] Luminance meter A luminance meter is used for measuring the luminance of a viewing box. The luminance meter allowing correct measurement of luminance around 3500 cd/m2 is required. [I] kVp meter A kVp meter is used for measuring X-ray tube voltage of the X-ray equipment. The required kVp meter is the one of non-contact type and having measurement accuracy within ± 1.5 kV and reproducibility within ± 0.5 kVp under the tube voltage range used for mammography. [J] Force scale A force scale is used for measuring the force of the compression plate. If the compression force cannot be applied properly on the measurement portion of the force scale, use soft rubber or towels so as to not damage the X-ray equipment and allow proper measurement. [K] Towels Towels should be used to protect the exposure table and compression plate during the Compression device confirmation. [L] Illuminance meter An illuminance meter is used for measuring illuminance under viewing conditions. 2-14 FDR Mammography QC Manual - 897N101461A [M] Densitometer A densitometer is used for measuring density of an image output on film. The densitometer should have a measurement range of 0 to 3.5 (0 to 4.0 is preferable) and measurement accuracy within ± 0.02. [N] PMMA phantom Acrylic plates large enough for covering the exposure table and 20, 40, 60 and 70 mm in thickness are required. These plates can be used in combination. The breast thickness where the average glandular dose becomes equal to the dose on each PMMA phantom is as shown in the table below. PMMA thickness (mm) Equivalent breast thickness (mm) 20 21 40 45 60 75 70 90 (Reference: EUREF Ed.4) NOTE Scratches, deformation, or dirt on the phantom not only affect the image evaluation or measured values but may damage the compression plate of the X-ray equipment. Make sure that there are no scratches, deformation, or dirt on the phantom before use. A 99.9% or higher purity aluminum plate of sufficient size for covering ROI is required. The plate can be also used as [P] Aluminum plates for half value layer measurement. Multiple plates of 0.2 mm or less in thickness can be used in combination. [P] Aluminum plates (0.3 and 0.5 mm) for half value layer measurement Overview [O] Aluminum plate (0.2 mm) for CNR measurement 2 A 99.9% or higher purity aluminum plate of sufficient size for covering the detection surface of the dosimeter is required. The plate can also be used as [O] Aluminum plate for CNR measurement. Multiple plates of 0.3 mm or less in thickness can be used in combination. [Q] Scale (ruler) A millimeter scale is required. [R] Coin Some familiar-sized coins are to be used. [S] Lead sheet A lead sheet is used to fully cover the receptor to completely block X-Rays when conducting the test on additive lag effects, and is also used as necessary to protect the image receptor when measuring the air kerma. FDR Mammography QC Manual - D 897N101461A 2-15 Equipment details A Hardcopy/Softcopy output equipment FDA-cleared hardcopy/softcopy output devices (laser imager/diagnostic monitor). B X-ray equipment The Fujifilm FFDM X-Ray exposure unit is to be used. NOTE In this Program, methods of exposure condition setting of an X-ray equipment are defined as follows: • “Manu (manual)” mode The mode allows manual setting of the kVp, mAs and target/filter of the X-ray equipment. 2 Overview • “Semi (semi auto)” mode The mode automatically specifies the mAs but allows manual setting of the kVp and target/filter of the X-ray equipment. • “Auto (full auto)” mode The mode automatically specifies kVp, mAs and target/filter of the X-ray equipment depending on the subject. C FUJIFILM workstation The workstation to be used is the Fujifim Acquisition Workstation (AWS). 2-16 FDR Mammography QC Manual - 897N101461A 2.4 QC Software Outline This section describes the operational procedure for the QC Software and the relevant window (“QC window” hereafter). &5*)&),-WORKSTATIONSTARTUP %XPOSURE 1#3OFTWARESTARTUP /PERATIONSONA&5*)&),-WORKSTATION 5SERINPUT ,OGDISPLAY Overview &ILMOUTPUTETC h1#WINDOWv 2 !UTOCALCULATION *UDGMENT /PERATIONSONTHE1#3OFTWARE The “QC window” composition is described below. %XPOSUREREGIONLIST %XPOSURE SUBMENULIST %XPOSUREMENULIST 0ATIENTINFORMATION DISPLAYFIELD FDR Mammography QC Manual - 897N101461A 2-17 2 Overview 2-18 FDR Mammography QC Manual - 897N101461A &5*)&),-WORKSTATIONSTARTUP %XPOSURE 1#3OFTWARESTARTUP 2 2 Overview Overview h1#WINDOWv 4HEFOLLOWINGFUNCTIONSCANBEEXECUTEDONTHEh1#WINDOWv4HERELEVANTBUTTONFUNCTIONSAREDESCRIBEDBELOW %XPOSURECONDITIONDISPLAYFIELD $ISPLAYSTHETESTLOG #HECKEXPOSURE CONDITIONS 3ETSTHEINITIALVALUES 1UITSTHE1#3OFTWARE 3AVESTHESTUDYRESULTS #HANGESTHECRITERIA /UTPUTSANIMAGE ONFILM %XECUTESMEASUREMENTBASEDONENTERED IMAGEANDVALUESANDDISPLAYS;0ASS&AIL= JUDGMENT /UTPUTSANIMAGE TOTHENETWORK /PENSTHEUSERINPUTSCREENFOR NECESSARYINFORMATION See the “FDR Mammography QC Software Operation Manual” and/or operation manual for a FUJIFILM workstation for detailed QC software operation. FDR Mammography QC Manual -897N101461A 2-19 5SERINPUT 4HERESULTSOFMEASUREMENTANDVISUALINSPECTIONCANBEENTEREDINTHEWINDOWBELOW%XAMPLE$AILY7EEKLY 4EST 2 2 Overview Overview !UTOCALCULATION %ACHTESTRESULTISAUTOCALCULATEDBASEDONTHEENTEREDIMAGEANDVALUESANDTHEJUDGMENT0ASS&AILIS DISPLAYEDINTHEWINDOWASSHOWNBELOW *UDGMENTRESULTS 0ASS&AIL #ALCULATEDRESULTS 2-20 FDR Mammography QC Manual - 897N101461A 2.5 Notes on Conducting the Program Operational notes regarding the Program are described below. Notes on Performing QC Tests (1) 1Shot/ACR Phantom position The positions of the Phantoms vary depending on the test item/content. See the test procedure for each item. 2 (3) Compression plate Some test items/contents require the compression plate to be installed, while some do not. See the test procedure for each item. Use a normal compression plate for mammography unless otherwise specified. 2 Overview Overview (2) Dosimeter Position the X-ray detector of the dosimeter 40 mm above the exposure table unless indicated otherwise. Make sure of the X-ray detection position of the dosimeter and adjust the height as specified by using an appropriate jig. In the X-ray reproducibility test, the dosimeter position differs since the air kerma is measured using AEC technique. (4) Solutions for failed test items Actions required when a test item judged as “Fail” are described for each test item. (5) Artifacts Artifacts may be seen in the Spatial Resolution chart on the 1Shot Phantom. This characteristic does not affect clinical images since there are no such special structures in a human body. Autocalculation is also not affected. (6) AGD AGD (Average glandular dose) values measured in this Program may be different from those in the test results displayed on a FUJIFILM workstation. In this Program, AGD values are measured based on the assumption that a uniform PMMA phantom is used and enable the user to evaluate the AEC (Automatic Exposure Control) performance of the X-ray equipment quantitatively. On the other hand, AGD values on a workstation are calculated for actual AGD values per mammogram. Due to this difference, the parameters used for calculations are different. Conduct tests by following the procedures in this Program. (7) Regular backup Regularly back up the results of the test conducted by using this Program. See “FDR-1000AWS Operation Manual” for details. (8) AEC exposure modes In this Program, only the (H-mode) exposure mode is used for clinical mammography (for test items that require AEC). Use this same exposure mode everytime you perform tests that require AEC. FDR Mammography QC Manual - 897N101461A 2-21 Chapter 3 Installation of FDR Mammography QC Program 3 Page FDR Mammography QC Manual - 897N101461A 3.1 Installation Procedure __________________________________ 3-2 3.2 Initial Settings _________________________________________ 3-3 3.3 Criteria Confirmation and Determination _________________ 3-5 3.4 Baseline Value Settings _________________________________ 3-12 3.4.1 Baseline Values for Annual Test ______________________ 3-13 3.4.2 Baseline Values for Annual A Test ____________________ 3-17 3.4.3 Baseline Values for Daily/Weekly Test _________________ 3-21 3.5 Checking Equipment Conditions at the Time of Program Installation ____________________________________________ 3-30 3-1 3.1 Installation Procedure The following steps shall be conducted by an authorized Fujifilm dealer prior to QC software use. These steps establish the criteria and baseline values that will be referenced for all QC tests supported by the QC software. Initial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 Specify the information for conducting the tests, such as dosimeter position. NOTE Make sure to perform these settings, otherwise the auto-calculation of the QC Software cannot run. Criteria Confirmation and Determination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 The reference criteria are preset. Check the criteria and change the values as necessary. 3 Installation of FDR Mammography QC Program Baseline Value Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.4 Measured values for some test items vary depending on the environment where the X-ray equipment is used and measurements are performed. For these items, conduct the tests several times when installing the Program and set the averages of the measured values as the baseline values. Checking Equipment Conditions at the Time of Program Installation . . . . . . . . . . . . . . . . . . . . . . 3.5 Conduct all of the test items provided in this Program except for Repeat Analysis to check the equipment conditions at the time of program installation. This will help correct a test item judged as [Fail] in a future QC test by providing the initial data for comparison. Conduct tests in the order of their potential influence on displayed images. Annual Test (Chapter 7) Semi-annual Test (Chapter 6) Weekly Test (Chapter 4) NOTE When new equipment or a system has been installed or existing equipment has been remodeled, perform the procedures described in “3.4 Baseline Value Settings” and later sections in Chapter 3. 3-2 FDR Mammography QC Manual - 897N101461A 3.2 Initial Settings The following items are required to be specified in the initial settings. • • • • Dosimeter - Exposure table distance Target / Filter and tube voltage (HVL) Air kerma unit Viewbox setting Setting Procedure 1 Enter patient information and press [Next]. 6 Press [Initial setting]. 7 Enter values for items requiring initial settings. For details on each item, see the following descriptions in (1) through (5). 3 (4) (5) (2) (3) 2 Select [QC/TEST] in the exposure region list and [1Shot PhantomM] in the exposure menu list, and then press [Start study.]. (1) Dosimeter - Exposure table distance • Measure the distance (mm) in the figure below and enter the value. The recommended value for h is 40 mm (This Manual assumes h is 40 mm in the following explanation). Set the distance l between the center of the dosimeter and chest wall-side edge of the exposure table as 60 mm. 2EFERENCEPOINT 3 The following exposure displayed. • 1Shot PhantomM submenu 4 Specify 23 kV, 2 mAs and Mo/Mo and make an exposure in “Manu” mode. is NOTE It may take some time until you can press [QC] after completing the exposure process. 5 Press [QC] to display the “QC window”. FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program (1) TIP Check the reference point of the dosimeter in its operation manual. (2) Magnification table - Exposure table’s top surface distance Enter 247 mm. 3-3 (3) Target / Filter and tube voltage (HVL) • Specify the kVp to be used for measuring half value layer in the Annual Test. The recommended kVp is 28 kV. (4) Air kerma unit • Select the measurement unit (mR or µGy) for the dosimeter. (5) Viewbox setting • Put a checkmark when using a viewbox. When the viewbox to be used has a luminance adjustment and/or masking function, also put a checkmark in the relevant box. 8 3 Press [OK]. Installation of FDR Mammography QC Program To continue to the criteria confirmation and determination, go to Step 6 in “3.3 Criteria Confirmation and Determination”. To complete the initial setting without setting the criteria, press [Close] to exit the window. 3-4 FDR Mammography QC Manual - 897N101461A 3.3 Criteria Confirmation and Determination The criteria should be specified on your own responsibility based on the measurement results obtained at the time of the Program installation. Enter the percentage or the absolute value of variations from the baseline value for each item. The reference criteria are shown on the “List of Reference Values for Criteria and Baseline Values”. The reference criteria are preset for each item. The criteria can be changed by following “Criteria Change Procedure” below. 3 The following displayed. • 1Shot PhantomM 4 Specify 23 kV, 2 mAs and Mo/Mo and make an exposure in “Manu” mode. Enter patient information and press [Next]. 2 Select [QC/TEST] in the exposure region list and [1Shot PhantomM] in the exposure menu list, and then press [Start study.]. FDR Mammography QC Manual - 897N101461A is 3 It may take some time until you can press [QC] after completing the exposure process. 5 Press [QC] to display the “QC window”. 6 Press [Criteria setting]. Installation of FDR Mammography QC Program 1 submenu NOTE Criteria Change Procedure When changing the criteria following “3.2 Initial Settings”, start from Step 6 . exposure 3-5 7 Select a tab and enter values for items to be changed. Daily/Weekly Tests 1/2 and 2/2 3 Installation of FDR Mammography QC Program [Items allowing criteria change] Test Item/Contents Judgment Items Criteria Missed tissue on chest wall edge (Right) [mm] 7 mm or less Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy Missed tissue on chest wall edge (Left) [mm] 7 mm or less CNR Baseline value ± 20 % System sensitivity Baseline value ± 35 % Dimension (Horizontal) [mm] Baseline value ± 2 % Dimension (Vertical) [mm] Baseline value ± 2 % Pixel Value (PV) ratio (Top-Right) [%] Baseline value ± 15 % Pixel Value (PV) ratio (Top-Left) [%] Baseline value ± 15 % Pixel Value (PV) ratio (Bottom-Right) [%] Baseline value ± 15 % Pixel Value (PV) ratio (Bottom-Left) [%] Baseline value ± 15 % SNR ratio (Top-Right) [%] Baseline value ± 15 % SNR ratio (Top-Left) [%] Baseline value ± 15 % SNR ratio (Bottom-Right) [%] Baseline value ± 15 % SNR ratio (Bottom-Left) [%] Baseline value ± 15 % Average QL at thinnest step wedge [QL] Baseline value ± 400 [QL] 2lp/mm [%] Baseline value ± 6 % Geometric distortion Uniformity Dynamic range Spatial Resolution (SR) 4lp/mm [%] Baseline value ± 15 % Low Contrast Detectability (Light) [%] > Baseline value - 14 % Low Contrast Detectability (LCD) Low Contrast Detectability (Dark) [%] > Baseline value - 14 % QL gap (1 step-2 step) [QL] Baseline value ± 50 [QL] QL gap (2 step-3 step) [QL] Baseline value ± 50 [QL] Linearity/Beam quality constancy QL gap (3 step-4 step) [QL] Baseline value ± 50 [QL] QL gap (4 step-5 step) [QL] Baseline value ± Baseline value [QL] Density at center of Phantom image 1.40 ± 0.2 Density difference (disk/outside) 0.4 or more and baseline value ± 0.05 Fibers (ACR Phantom) Specks (ACR Phantom) Test with ACR Phantom Masses (ACR Phantom) 3-6 4 or more and no more than 0.5 lower than baseline 3 or more and no more than 0.5 lower than baseline 3 or more and no more than 0.5 lower than baseline Visible step wedge (Step Phantom) *1 10 steps Specks (Step Phantom) *1 4 steps or more Masses (Step Phantom) *1 5 steps or more FDR Mammography QC Manual - 897N101461A Quarterly Test There is no item related to the criteria. Semi-annual Test There is no item related to the criteria. Annual A Test 3 Test Items Judgment Items Criteria Image Basic Test Relative sensitivity (S value) Baseline value ± 40 % Compressed breast thickness accuracy [mm] ± 5 mm Compression force accuracy [N] ± 20 N Compression device confirmation Maximum compression force [N] 111 to 200 N Luminance [cd/m2] 3500 cd/m2 or more Interpretation room illuminance (Ambient light) [lx] 50 lx or less Viewbox maintenance FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program [Items allowing criteria change] 3-7 Annual Test 1/3 (Annual Test consists of tabs 1/3 to 3/3) 3 [Items allowing criteria change] Installation of FDR Mammography QC Program Test Items Judgment Items Criteria Additive lag effects (Lag) Lag factor [QL] 75 QL or less Multiplicative lag effects (Ghost) Ghost factor Below 0.045 MTF (4lp/mm)[%] Baseline value ± 12 % Spatial Resolution (Magnification) kVp accuracy and reproducibility 3-8 MTF (8lp/mm)[%] Baseline value ± 15 % Low volt. accuracy (LFS) [kV] Specified kVp ± 1 kV Mid. volt. accuracy (LFS) [kV] Specified kVp ± 1 kV High volt. accuracy (LFS) [kV] Specified kVp ± 1 kV Low volt. accuracy (SFS) [kV] Specified kVp ± 1 kV Mid. volt. accuracy (SFS) [kV] Specified kVp ± 1 kV High volt. accuracy (SFS) [kV] Specified kVp ± 1 kV Mid. volt. reproducibility (LFS) [kV] Reproducibility ± 0.5 kV or less FDR Mammography QC Manual - 897N101461A Annual Test 2/3 3 [Items allowing criteria change] Test Items Criteria Total gap criteria = SID X 0.02 or less X-ray / Light field gap (Chest/Nipple) [mm] Total gap criteria = SID X 0.02 or less X-ray / Image receptor field gap (Left) [mm] SID X 0.02 or less X-ray / Image receptor field gap (Right) [mm] SID X 0.02 or less X-ray / Image receptor field gap (Nipple) [mm] SID X 0.02` or less X-ray field / Exposure table gap [mm] 5 mm or less Reproducibility Variation coefficient 0.05 or less Air kerma rate [mGy/s] 7 mGy/s or more Specific radiation output [μGy/mAs] 30 μGy/mAs or more Radiation output FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program Collimation assessment Judgment Items X-ray / Light field gap (Right/Left) [mm] 3-9 Annual Test 3/3 3 [Items allowing criteria change] Test Items Installation of FDR Mammography QC Program 3-10 Judgment Items Criteria Entrance air kerma accuracy [%] Average ± 15 % or less Entrance air kerma reproducibility Variation coefficient 0.05 or less CNR relative value 20mm [%] Baseline value 105 % or more CNR relative value 40mm [%] Baseline value 95 % or more CNR relative value 60mm [%] Baseline value 63 % or more CNR relative value 70mm [%] Baseline value 60 % or more AGD 20 mm [mGy] 1 mGy or less AGD 40 mm [mGy] 2 mGy or less AGD 60 mm [mGy] 4.5 mGy or less AGD 70 mm [mGy] 6.5 mGy or less AGD-ACR Phantom 3 mGy or less AEC reproducibility CNR mode 1 AGD mode 1 AGD-ACR Phantom TIP • [CNR_H-mode] in the window corresponds to CNR mode 1. • As with CNR mode, AGD mode 1 corresponds to H-mode. FDR Mammography QC Manual - 897N101461A 8 Complete checking and changing all the necessary items, then press [OK]. 3 Press [Close] to exit the “QC window”. NOTE If “The test result is not saved.” is displayed, press [OK]. 10 Select [Study completed] to finish the test. FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program 9 3-11 3.4 Baseline Value Settings This section describes how to determine and specify the baseline values of each test item as the basis of judgment. Make sure to perform these settings, otherwise judgment cannot be made correctly. In this Program, the judgment is made based on how much the measured values deviates from the baseline values. Since the measured values for some test items depend on the X-ray equipment and/or exposure conditions, it is necessary to specify the baseline values for the test items. [Test items requiring baseline value setting] The items in the shadowed cells of the below table require the baseline settings. Test Categories Exposure Menus Test Items/Contents 3 Installation of FDR Mammography QC Program 3-12 Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy Geometric distortion System artifact evaluation 1Shot PhantomM Test with 1Shot Phantom Weekly Test Uniformity Dynamic range Spatial Resolution (SR) Low Contrast Detectability (LCD) Linearity/Beam quality constancy ACR Phantom Semi-annual Test Test with ACR Phantom Compression device confirmation Annual A Image Basic Test Annual Test Annual 2/6 Spatial Resolution Spatial Resolution (Magnification) [Points to be noted] • Baseline values vary depending on the exposure environment. • Measurements must be conducted several times under uniform conditions to specify the baseline values. However, the operational procedure for the QC Software is different in the last measurement. • Test items are categorized into an exposure menu as shown in the above table. • Enter baseline values for all of the test items in an exposure menu, including the items not requiring the baseline setting, to proceed with the operation in the QC software. It is recommended to enter “0” in order not to mistake the entered value from the measured value when checking the setting later. • It is recommended to specify the baseline values in the order of Annual Test → Semi-annual Test → Weekly Test, according to their potential influence on images. • It is recommended to save image data based on which baseline values are determined to later confirm that weekly, semi-annual and annual tests have been conducted properly. FDR Mammography QC Manual - 897N101461A 3.4.1 Baseline Values for Annual Test Among the Annual Test items, the Spatial Resolution (Magnification) requires the baseline setting. Test Category Exposure Menu Annual Test Annual 2/6 [1] Test Item Spatial Resolution Spatial Resolution (Magnification) Spatial Resolution (Magnification) [Test flow] 0ATIENTINFORMATIONENTRYANDEXPOSURE MENUSELECTION 1#3OFTWARESTARTUP %XPOSURECONDITIONCONFIRMATION !UTOCALCULATION "ASELINEVALUESETTING 1#3OFTWARESHUTDOWN [Tools] 1Shot Phantom FDR Mammography QC Manual - 897N101461A 3 2EPEATSEVERALTIMES 2ECOMMENDEDNUMBER OFTIMES Installation of FDR Mammography QC Program 3PATIAL2ESOLUTION-AGNIFICATION EXPOSURE 3-13 [Baseline value setting procedure] 5 1. Patient information entry and exposure menu selection 1 Enter patient information and press [Next]. 2 Select [QC/TEST] in the exposure region list and [Annual 2/6] in the exposure menu list, and then press [Start study.]. Position the 1Shot Phantom being shifted to the right-hand side so that the area indicated by the broken line in the below figure is at the lateral center of the X-ray field. 3 Installation of FDR Mammography QC Program 3 The following exposure displayed. • 1ShotM(Magnification) submenu NOTE If the area is out of the X-ray field while the Phantom is positioned by pressing its corners against the chest wall-side edge of the magnification table, shift the Phantom to the front. Make sure that the shifted Phantom is parallel to the chest wall-side edge. is NOTE 2. Spatial Resolution (Magnification) (exposure) 4 3-14 The Phantom position must be the same for every exposure (the allowable positional error is within ± 5 mm). Record how much the Phantom is shifted to the right-hand side and front side. • Fill in the following items in the worksheet. 1 Shot Phantom position Lateral shift distance: _____mm Mount the magnification table of the Xray equipment and the compression plate normally used in magnification exposure in clinical practice. FDR Mammography QC Manual - 897N101461A 6 Specify “Manu” mode for the X-ray equipment and make a magnified exposure under the exposure conditions most often used in clinical practice (about 28 kV, 80 to 100 mAs and Mo/Mo is recommended). • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 5. Auto-calculation 9 Press [Measurement execution]. The results are displayed at the bottom of the window. TIP If “No baseline value is input.” appears, press [OK]. NOTE In the last time measurement, skip Steps 10 through 12 and go to Step 13 . When the results are displayed, press [Save] and then [Close] to exit the QC Software. 11 Select [Study completed] to finish the test. 12 Repeat Steps 1 through 11 several times (5 is recommended) to specify the averages of the multiple measured values as the baseline values. NOTE It may take some time until you can press [QC] after completing the exposure process. 3. QC Software startup 7 Press [QC] to display the “QC window”. 4. Exposure condition confirmation 8 Check the exposure conditions. NOTE Make sure to repeat the same procedure until the test is finished. Do not add an exposure submenu when repeating. 13 FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program 10 3 When the result is displayed, press [Save] and then [Test data log]. 3-15 6. Baseline value setting 14 Select the results to be used for calculating the baseline values among the measured results, and then press [Baseline value creation]. This completes the baseline value setting. Press [Close]. 3 Installation of FDR Mammography QC Program 3-16 7. QC Software shutdown 15 Press [Save] and then [Close] to exit the QC Software. 16 Select [Study completed] to finish the test. FDR Mammography QC Manual - 897N101461A 3.4.2 Baseline Values for Annual A Test Among the Annual A Test items, the Image basic test requires the baseline setting. Test Category Exposure Menu Annual A Test Annual A 1/1 Test Item Image Basic Test Compression device confirmation Viewing box maintenance [1] Image basic test [Test flow] 0ATIENTINFORMATIONENTRYANDEXPOSURE MENUSELECTION 3 )MAGEBASICTESTEXPOSURE %XPOSURECONDITIONCONFIRMATION 2EPEATSEVERALTIMES 2ECOMMENDEDNUMBER OFTIMES )MAGEBASICTESTUSERINPUT !UTOCALCULATION "ASELINEVALUESETTING 1#3OFTWARESHUTDOWN [Tools] A dosimeter, scale, and lead sheet [Baseline value setting procedure] 1 Determine the exposure conditions used in the test. Remove the compression plate, place a lead sheet on the exposure table, and then position the dosimeter at the lateral center of the exposure table, 60 mm away from the chest wallside edge, and also 40 mm above the exposure table. Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. Specify “Manu” mode for the X-ray equipment and make an exposure with 25 kV and Mo/Mo. Then record the mAs when the air kerma becomes approximate to 20 mR in the worksheet. Installation of FDR Mammography QC Program 1#3OFTWARESTARTUP • Fill in the following item in the worksheet. mAs: _____mAs 2 Specify “Manu” mode for the X-ray equipment and measure the air kerma 3 times with the mAs derived from the exposure with 25 kV and Mo/Mo in Step 1 . • Fill in the following item in the worksheet. mR mR Entrance air kerma: _____mR µGy , _____µGy , _____µGy 3 Remove the dosimeter. Select [Study completed] to finish the test. FDR Mammography QC Manual - 897N101461A 3-17 1. Patient information entry and exposure menu selection 2. Image basic test (exposure) 7 4 Enter patient information and press [Next]. 5 Select [QC/TEST] in the exposure region list and [Annual A] in the exposure menu list, and then press [Start study.]. Set [Grid] to [OUT], and then make an exposure with the same conditions as those used in Step 2 . • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ 3 Installation of FDR Mammography QC Program 3-18 NOTE It may take some time until you can press [QC] after completing the exposure process. 3. QC Software startup 8 Press [QC] to display the “QC window”. 4. Exposure condition confirmation 9 6 The following exposure displayed. • Uniform Expo(Image Basic) submenu Check the exposure conditions. is FDR Mammography QC Manual - 897N101461A 6. Auto-calculation 5. Image basic test (user input) 10 Press [User input], enter the measured entrance air kerma and visual inspection results (Pass/Fail), and then press [OK]. 11 Press [Measurement execution]. The results are displayed at the bottom of the window. TIP (4) If “No baseline value is input.” appears, press [OK]. (1) (2) NOTE In the last measurement, skip Steps 12 through 14 and go to Step 15 . (3) When the result is displayed, press [Save] and then [Close] to exit the QC Software. 13 Select [Study completed] to finish the test. 14 Some dosimeters may require correction of the measured value according to the beam quality. Repeat Steps 1 through 13 several times (5 is recommended) to specify the averages of the multiple measured values as the baseline values. NOTE NOTE Enter values or select [Pass/Fail] for all of the items. The Pass/Fail selection does not affect the baseline value determination. Make sure to repeat the same procedure until the test is finished. Do not add an exposure submenu when repeating. (1) Enter the measured entrance air kerma recorded in Step 2 . (2) Select either of [Pass/Fail] for [Artifact] and [White out]. (3)(4) Enter any values for these items. NOTE NOTE Enter any value (e.g. “0” for all items) except for negative or significantly large one. Otherwise, an error occurs. FDR Mammography QC Manual - 897N101461A 15 Installation of FDR Mammography QC Program 12 3 When the result is displayed, press [Save] and then [Test data log]. 3-19 7. Baseline value setting 16 Select the results to be used for calculating the baseline values among the measured results, and then press [Baseline value creation]. This completes the baseline value setting. Press [Close]. 3 Installation of FDR Mammography QC Program 3-20 8. QC Software shutdown 17 Press [Save] and then [Close] to exit the QC Software. 18 Select [Study completed] to finish the test. FDR Mammography QC Manual - 897N101461A 3.4.3 Baseline Values for Weekly Test The following Weekly Test items/contents require the baseline setting: CNR, 1Shot Phantom sensitivity constancy, Geometric distortion, Uniformity, Dynamic range, Spatial Resolution (SR) and Linearity/Beam quality constancy in Test with 1Shot Phantom, and Density difference (disk/outside) in Test with ACR Phantom*. For each item, make exposures several times using the 1Shot Phantom, or the ACR Phantom, Step Phantom and acrylic disk, and then derive the average from the calculated results to determine it as the baseline value. *NOTE The baseline value setting for the test with ACR Phantom/Acrylic Disk is only required if the facility uses hardcopy for primary interpretation Test Category Exposure Menu Test Items/Contents Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy System artifact evaluation 1Shot PhantomM Test with 1Shot Phantom Uniformity Daily/Weekly Test Dynamic range Spatial Resolution (SR) Low Contrast Detectability (LCD) Linearity/Beam quality constancy ACR Phantom Test with ACR Phantom [1]Test with 1Shot Phantom [Test flow] 0ATIENTINFORMATIONENTRYANDEXPOSURE MENUSELECTION 4ESTWITH3HOT0HANTOM EXPOSURE 1#3OFTWARESTARTUP Installation of FDR Mammography QC Program Geometric distortion 3 2EPEATSEVERALTIMES 2ECOMMENDEDNUMBER OFTIMES %XPOSURECONDITIONCONFIRMATION 4ESTWITH3HOT0HANTOMUSERINPUT !UTOCALCULATION "ASELINEVALUESETTING 1#3OFTWARESHUTDOWN [Tools] 1Shot Phantom FDR Mammography QC Manual - 897N101461A 3-21 [Baseline value setting procedure] 2. Test with 1Shot Phantom (exposure) 1. Patient information entry and exposure menu selection 1 4 Enter patient information and press [Next]. Position the 1Shot exposure table. Phantom on the NOTE Position the Phantom at the lateral center of the exposure table by pressing the corners against the chest wall-side edge of the exposure table. If there are obstacles at the time of positioning, the test may not be conducted accurately. 3HOT0HANTOM 3 Installation of FDR Mammography QC Program 3-22 2 Select [QC/TEST] in the exposure region list and [1Shot PhantomM] in the exposure menu list, and then press [Start study.]. %XPOSURETABLE #ORNEROFTHE 3HOT0HANTOM 5 3 The following exposure displayed. • 1Shot PhantomM submenu Position the compression plate 40 mm above the exposure table. is FDR Mammography QC Manual - 897N101461A 6 Specify 28 kV and Mo/Mo and make an exposure in “Semi” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 10 Select either of [Pass/Fail] for [1Shot Phantom image confirmation] and [Good practice] items, and then press [OK]. NOTE Enter values or select [Pass/Fail] for all of the items. The Pass/Fail selection does not affect the baseline value determination. NOTE It may take some time until you can press [QC] after completing the exposure process. 3. QC Software startup 7 3 Press [QC] to display the “QC window”. 8 Check the exposure conditions. 6. Auto-calculation 11 Press [Measurement execution]. The results are displayed at the bottom of the window. TIP If “No baseline value is input.” appears, press [OK]. 5. Test with 1Shot Phantom (user input) 9 Press [User input]. NOTE In the last time measurement, skip Steps 12 through 14 and go to Step 15 . FDR Mammography QC Manual - 897N101461A 12 When the result is displayed, press [Save] and then [Close] to exit the QC Software. 13 Select [Study completed] to finish the test. Installation of FDR Mammography QC Program 4. Exposure condition confirmation 3-23 14 Repeat Steps 2 through 13 several times (5 is recommended) to specify the averages of the multiple measured values as the baseline values. 17 Create baseline values for “Low Contrast Detectability (Light)” and “Low Contrast Detectability (Dark)” in “LCD”. NOTE Make sure to repeat the same procedure until the test is finished. Do not add an exposure submenu when repeating. 15 When the result is displayed, press [Save] and then [Test data log]. 3 Installation of FDR Mammography QC Program 7. Baseline value setting 16 3-24 Select the results to be used for calculating the baseline values among the measured results, and then press [Baseline value creation]. 1. Average measured results for “Low Contrast Detectability (Light)”. This is the baseline value for “Low Contrast Detectability (Light)”. If measured results for “Low Contrast Detectability (Light)” are respectively w1, w2, w3, w4 and w5, the average can be calculated by the following formula: (w1+w2+w3+w4+w5)/5. Round the calculated value to the whole number. • Fill in the following item in the worksheet. Baseline value for “Low Contrast Detectability (Light)”: _____% E.g.: Assuming that measured results for “Low Contrast Detectability (Light)” are respectively 59.66, 60.33, 55.25, 62.27 and 65.19. (59.66+60.33+55.25+62.27+65.19)/5 = 60.54 Round to the whole number. = 61 2. Similarly, average measured results for “Low Contrast Detectability (Dark)”. This is the baseline value for “Low Contrast Detectability (Dark)”. If measurement results for “Low Contrast Detectability (Dark)” are respectively b1, b2, b3, b4 and b5, the average can be calculated by the following formula: (b1+b2+b3+b4+b5)/5. Round the calculated value to the whole number. FDR Mammography QC Manual - 897N101461A • Fill in the following item in the worksheet. Baseline value for “Low Contrast Detectability (Dark)”: _____% E.g.: Assuming that measured results for “Low Contrast Detectability (Dark)” are respectively 67.03, 68.65, 63.03, 67.65 and 68.70. (67.03+68.65+63.03+67.65+68.70)/5 = 67.012 Round to the whole number. = 67 18 Press [Criteria setting]. FDR Mammography QC Manual - 897N101461A Select the [Daily/Week/y test (1/2)] tab. Enter the criteria for “Light” and “Dark” in “LCD” and press [OK]. 3 8. QC Software shutdown 21 Press [Save] and then [Close] to exit the QC Software. 22 Select [Study completed] to finish the test. Installation of FDR Mammography QC Program 19 Create criteria to be entered in “Low Contrast Detectability (Light)” and “Low Contrast Detectability (Dark)” in “LCD”. 1. Subtract 14 from the baseline value for “Low Contrast Detectability (Light)” in consideration of variations in the measurement. This is the criteria for “Low Contrast Detectability (Light)”. • Fill in the following item in the worksheet. Criteria for “Low Contrast Detectability (Light)”: _____% E.g.: Assuming that baseline value for “Low Contrast Detectability (Light)” is 61. 61-14 = 47 2. Similarly, subtract 14 from the baseline value for “Low Contrast Detectability (Dark)”. This is the criteria for “Low Contrast Detectability (Dark)”. • Fill in the following item in the worksheet. Criteria for “Low Contrast Detectability (Dark)”: _____% E.g.: Assuming that baseline value of “Low Contrast Detectability (Dark)” is 67. 67-14 = 53 20 3-25 [2]Test with ACR Phantom The baseline value setting for the test with ACR Phantom/Acrylic Disk is only required if the facility uses hardcopy for primary interpretation [Test flow] 0ATIENTINFORMATIONENTRYANDEXPOSURE MENUSELECTION !#20HANTOMEXPOSURE 1#3OFTWARESTARTUP 2EPEATSEVERALTIMES 2ECOMMENDEDNUMBER OFTIMES %XPOSURECONDITIONCONFIRMATION 3 )MAGEOUTPUT Installation of FDR Mammography QC Program !#20HANTOMUSERINPUT !UTOCALCULATION "ASELINEVALUESETTING 1#3OFTWARESHUTDOWN [Tools] ACR Phantom, an acrylic disk, and a densitometer 3-26 FDR Mammography QC Manual - 897N101461A [Baseline value setting procedure] 2. ACR Phantom (exposure) 1. Patient information entry and exposure menu selection 1 4 Position the ACR Phantom at the lateral center of the exposure table while aligning its edge with the chest wall-side edge of the exposure table. 5 Position the acrylic disk wherever it does not interfere with the visual inspection of the ACR Phantom (as shown in the figure below). Enter patient information and press [Next]. Nipple s ide 3 AC R P hantom Select [QC/TEST] in the exposure region list and [ACR Phantom] in the exposure menu list, and then press [Start study.]. C hes t wall s ide Acrylic dis k 6 Position the compression plate so that it comes in contact with the ACR Phantom. NOTE Make sure that excessive pressure is not applied to the Phantom. 3 The following displayed. • ACR Phantom exposure submenu is 7 Installation of FDR Mammography QC Program 2 Make an exposure in “Auto” or “Semi” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N NOTE It may take some time until you can press [QC] after completing the exposure process. 3. QC Software startup FDR Mammography QC Manual - 897N101461A 3-27 8 Press [QC] to display the “QC window”. 13 Enter the densities measured in Step 11 . NOTE 4. Exposure condition confirmation 9 Check the exposure conditions. Make sure to enter values for [Density inside the disk/Density outside the disk] in the correct position. Enter any values for [Density at center of Phantom image] and the image evaluation, then press [OK]. NOTE Enter values for all of the items. The values except those for [Density inside the disk/Density outside the disk] do not affect the baseline value determination. 3 Installation of FDR Mammography QC Program 5. Image output 10 Select an image exposed using Exposure Submenu [ACR Phantom] from the exposure condition display field and press [Film output]. 11 Measure the following items using the image output on film. • Fill in the following items in the worksheet. Density inside the disk: _____ Density outside the disk: _____ 6. ACR Phantom (user input) 12 3-28 Press [User input]. FDR Mammography QC Manual - 897N101461A 7. Auto-calculation 14 Press [Measurement execution]. The results are displayed at the bottom of the window. 8. Baseline value setting 19 Select the results to be used for calculating the baseline values among the measured results, and then press [Baseline value creation]. This completes the baseline value setting. Press [Close]. NOTE Make sure that the results for the Test with ACR Phantom are selected. They are displayed below those for the Test with 1Shot Phantom. TIP If “No baseline value is input.” appears, press [OK]. NOTE 3 In the last measurement, skip Steps 15 through 17 and go to Step 19 . When the result is displayed, press [Save] and then [Close] to exit the QC Software. 16 Select [Study completed] to finish the test. 9. QC Software shutdown 17 Repeat Steps 1 through 16 several times (5 is recommended) to specify the averages of the multiple measured values as the baseline values. 20 Press [Save] and then [Close] to exit the QC Software. 21 Select [Study completed] to finish the test. NOTE Make sure to repeat the same procedure until the test is finished. Do not add an exposure submenu when repeating. 18 When the result is displayed, press [Save] and then [Test data log]. FDR Mammography QC Manual - 897N101461A Installation of FDR Mammography QC Program 15 3-29 3.5 Checking Equipment Conditions the Time of Program Installation at Conduct all of the test items provided in this Program except for Repeat analysis to confirm that the all judgment items result in [Pass]. 3 This enables to find defects in the X-ray equipment or hardcopy/softcopy output equipment when installed. In addition, this will help correct a test item judged as “Fail” in a future QC test by providing the initial data for comparison. For this reason, it is recommended to save the image data when equipment conditions are checked at the time of the Program installation. Installation of FDR Mammography QC Program The specified baseline values or reference criteria are shown on the “List of Reference Values for Criteria and Baseline Values”. For visual inspections, it is recommended to clearly relate an output image and the judgment result in the condition check performed when the Program has been installed. Furthermore, it is advisable to observe an output image in the condition check at the time of the Program installation with some people (conductors of the QC tests in the Program are recommended) to share the points for evaluation in visual inspections. See the relevant Chapters for the test procedures. TIP Conduct the tests in the order of their potential influence on displayed images. Annual Test (Chapter 7) Semi-annual Test (Chapter 6) Daily/Weekly Test (Chapter 4) Conducting these Tests complete the condition check at the time of the Program installation. 3-30 FDR Mammography QC Manual - 897N101461A 3 Installation of FDR Mammography QC Program FDR Mammography QC Manual - 897N101461A 3-31 Chapter 4 Weekly Test Page 4.1 Test Flow ______________________________________________ 4-2 4.2 Test Items _____________________________________________ 4-3 4.3 Tools _________________________________________________ 4-4 4.4 Setting Confirmation ___________________________________ 4-5 4.5 Conducting Weekly 1 Shot Phantom Test ________________ 4-6 4.5.1 4.5.2 4.5.3 4.6 4 Procedure _______________________________________ 4-6 Test Result Evaluation and Judgment _________________ 4-11 Performance Criteria and Corrective Action ____________ 4-11 Conducting Weekly ACR Phantom Test ___________________ 4-12 Procedure _______________________________________ 4-12 Test Result Evaluation and Judgment _________________ 4-16 Performance Criteria and Corrective Action ____________ 4-17 4.6.1 4.6.2 4.6.3 FDR Mammography QC Manual - 897N101461A 4.7 Conducting Weekly Printer QC __________________________ 4-12 4.8 Conducting Weekly Monitor QC _________________________ 4-12 4-1 4.1 Test Flow Based on quantitative measurements that enable the identification of gradual changes in image quality over time, the Weekly test is performed : 1) Weekly by the Technologist 2) Annually by the Medical Physicist during the Medical Equipment Evaluation (MEE) and subsequent Annual testing. The Weekly Test is comprised of constancy tests and performance verification tests of the system. The constancy tests are designed for determining if variations of regularly-measured system performance values are within the allowable range (criteria) based on baseline values that were established at the time of QC program / system installation. The performance test is intended to check that system performance values are within the upper or lower limits specified by the baselines. 4 It is necessary to determine the criteria before conducting the Weekly Test. Weekly Test NOTE The reference criteria are shown on the “The List of Reference Vales for Criteria and Baseline Values”. See “3.3 Criteria Confirmation and Determination” for details. When conducting the Weekly Test for the first time after setting the criteria, specify the baseline values to be used in the future Weekly Tests. In the second or later Weekly Test, check that the variation from the specified baseline values is within the criteria. See “3.4 Baseline Value Settings” for details. 4-2 FDR Mammography QC Manual - 897N101461A 4.2 Test Items The Weekly Test consists of the following test items. Test Items/Contents Exposure Menus Exposure Submenus Items Autocalculated and Judged Based on Input Data 1. Exposure 2. Input Values Items Judged [Pass/Fail] Based on Visual Inspection by a User Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy ✓ Geometric distortion Test with 1Shot Phantom System artifact evaluation Uniformity 1Shot PhantomM 1Shot PhantomM ✓ ✓ Dynamic range Spatial Resolution (SR) Low Contrast Detectability (LCD) Linearity/Beam quality constancy Good practice ACR Phantom* ACR Phantom* ✓ ✓ Three judgment methods: 1. Auto-calculation by the QC Software based on the image data acquired by making an exposure. 2. Auto-calculation by the QC Software based on the input measurement data. 3. User selection of [Pass/Fail] based on visual inspection of the image. Weekly Test Test with ACR Phantom 4 ✓ - Each Weekly Test item is tested by normally using an output image. Most of the test results can be auto-calculated from the image data by the QC Software, though some items require visual inspection. *NOTE The Test with ACR Phantom is required to be conducted at a Weekly frequency. On the day that you would conduct both the Test with 1Shot Phantom and the Test with ACR Phantom, select Exposure Menu [1Shot/ACR]. In other cases, do not use the exposure menu. Otherwise, an incorrect test data log will be created. FDR Mammography QC Manual - 897N101461A 4-3 4.3 Tools Tools to be used for the Weekly Test based on this Guidebook are shown below. See “2.3 Tools” for details. Tools [ A ] QC Software [ B ] Worksheet (See Chapter 8) [ C ] 1Shot Phantom [ D ] Viewbox*1 [ F ] ACR Phantom (including an acrylic disk*1) [ M ] Densitometer (if film is used) *1 NOTE The ACR phantom density measurement (using the acrylic disk) is required to be performed weekly if the facility routinely prints hardcopy images for primary interpretation. If images are only occasionally printed for final interpretation, such as upon request of a patient for images for interpretation elsewhere, then the test must be performed in advance of, but not more than one week before printing the films. 4 Weekly Test 4-4 FDR Mammography QC Manual - 897N101461A 4.4 Setting Confirmation The following items described in Chapter 3 are required to be specified before conducting the Weekly Test. • • • Initial settings ...................................3.2 Initial Settings Criteria..................................................3.3 Criteria Confirmation and Determination Baseline values..................................3.4 Baseline Value Settings [Items allowing criteria change] Test Items/Contents Judgment Items Criteria Missed tissue on chest wall edge Missed tissue on chest wall edge (Right) [mm] Missed tissue on chest wall edge (Left) [mm] CNR CNR Baseline value ± 20 % 1Shot Phantom sensitivity constancy System sensitivity Baseline value ± 35 % Dimension (Horizontal) [mm] Baseline value ± 2 % 7 mm or less 7 mm or less Geometric distortion Dimension (Vertical) [mm] Baseline value ± 2 % Pixel Value (PV) ratio (Top-Right) [%] Baseline value ± 15 % Baseline value ± 15 % Baseline value ± 15 % Pixel Value (PV) ratio (Bottom-Left) [%] Baseline value ± 15 % SNR ratio (Top-Right) [%] Baseline value ± 15 % SNR ratio (Top-Left) [%] Baseline value ± 15 % SNR ratio (Bottom-Right) [%] Baseline value ± 15 % SNR ratio (Bottom-Left) [%] Baseline value ± 15 % Average QL at thinnest step wedge [QL] Baseline value ± 400 [QL] 2lp/mm [%] Baseline value ± 6 [%] Uniformity 1Shot PhantomM Dynamic range 4 Weekly Test Pixel Value (PV) ratio (Top-Left) [%] Pixel Value (PV) ratio (Bottom-Right) [%] Spatial Resolution (SR) Low Contrast Detectability (LCD) 4lp/mm [%] Baseline value ± 15 [%] Low Contrast Detectability (Light) [%] > Baseline value - 14 % Low Contrast Detectability (Dark) [%] > Baseline value - 14 % QL gap (1 step-2 step) [QL] Baseline value ± 50 [QL] QL gap (2 step-3 step) [QL] Baseline value ± 50 [QL] Linearity/Beam quality constancy ACR Phantom QL gap (3 step-4 step) [QL] Baseline value ± 50 [QL] QL gap (4 step-5 step) [QL] Baseline value ± Baseline value [QL] Density at center of Phantom image 1.4 ± 0.2 Density difference (disk/outside) 0.4 or more and baseline value ± 0.5 Fibers (ACR Phantom) 4 points or more Specks (ACR Phantom) 3 points or more Masses (ACR Phantom) 3 points or more Visible step wedge (Step Phantom) *1 10 steps Specks (Step Phantom) *1 4 steps or more Masses (Step Phantom) *1 5 steps or more TIP The criteria setting is not required for visual inspection. *1 When not checking these items, leave the preset values as they are. Do not delete the preset values. FDR Mammography QC Manual - 897N101461A 4-5 4.5 4.5.1 Conducting Weekly 1 Shot Phantom Test Procedure [Test flow] 0ATIENTINFORMATIONENTRYANDEXPOSUREMENUSELECTION 4ESTWITH3HOT0HANTOMEXPOSURE 'OODPRACTICECLEANING 1#3OFTWARESTARTUP %XPOSURECONDITIONCONFIRMATION 4 )MAGEOUTPUT Weekly Test 4ESTWITH3HOT0HANTOMUSERINPUT 'OODPRACTICEUSERINPUT !UTOCALCULATION 4ESTRESULTCONFIRMATION 1#3OFTWARESHUTDOWN 4-6 FDR Mammography QC Manual - 897N101461A 1. Patient information entry and exposure menu selection 1 Enter patient information and press [Next]. 2. Test with 1Shot Phantom (exposure) 4 Follow the instructions below to conduct the test item. [Tools] 1Shot Phantom 1. Position the 1Shot Phantom on the exposure table. NOTE Position the Phantom at the lateral center of the exposure table by pressing the corners against the chest wall-side edge of the exposure table. If there are obstacles at the time of positioning, the test may not be conducted accurately. 2 Select [QC/TEST] in the exposure region list and [1Shot PhantomM] in the exposure menu list, and then press [Start study.]. 3HOT0HANTOM 4 Weekly Test %XPOSURETABLE 3 The following exposure submenu is displayed. • 1Shot PhantomM #ORNEROFTHE 3HOT0HANTOM 2. Position the compression plate 40 mm above the exposure table. FDR Mammography QC Manual - 897N101461A 4-7 3. Viewbox cleaning Wipe clean the surface of the viewbox and check that the luminance is uniform. Follow the instructions given by the manufacturer if any. • Fill in the following items in the worksheet. X-ray equipment cleanliness: Pass/Fail Screen cleanliness: Pass/Fail Viewbox cleanliness: Pass/Fail 3. Specify the same kVp and target/filter as those determined in the “3.4.3 Baseline Values for Daily/Weekly Test”. Then make an exposure in “Semi” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N NOTE It may take some time until you can press [QC] after completing the exposure process. NOTE Make sure to specify the same exposure conditions as those used in the baseline value setting. If they are not certain, newly determine the exposure conditions for the 1Shot Phantom by following “3.4.3 Baseline Values for Daily/Weekly Test”. 4. QC Software startup 6 4 Press [QC] to display the “QC window”. 5. Exposure condition confirmation Weekly Test 3. Good practice (cleaning) 5 7 Check the exposure conditions. Follow the instructions below to conduct the test item. [Tools] See the respective operation manuals for the FDR-1000DRSZ, FDR-1000AWS, and viewbox for the necessary cleaning tools. 1. X-ray equipment cleaning Clean the following parts of the X-ray equipment. • Compression surface of the breast compression plate • Patient support and chest wall-side surface of the exposure table • Surfaces of X-ray tube assembly (incorporating X-ray tube and collimator) including the mounted and removable accessories • Armrests 2. Softcopy output equipment cleaning Wipe off any dust, finger marks and other dirt on all of the softcopy output equipment by using a lint-free soft cloth. Follow the instructions provided by the manufacturer. 4-8 FDR Mammography QC Manual - 897N101461A 6. Image output 8. Good practice (user input) Select Exposure Submenu [1Shot PhantomM] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. 8 11 Enter the results of Good practice, and then press [OK]. NOTE Make sure that the result is selected for all items. If there is an item left unentered, an error occurs. TIP For performing image diagnosis on a softcopy display, check the items on the display. 9. Auto-calculation 12 Press [Measurement execution]. The results are displayed at the bottom of the window. 4 Weekly Test Check or measure the following items using the image output on film. 9 • Fill in the following items in the worksheet. 1Shot Phantom Geometric distortion: Pass/Fail Artifact: Pass/Fail 7. Test with 1Shot Phantom (user input) 10 Press [User input]. inspection results. Enter FDR Mammography QC Manual - 897N101461A the visual 4-9 10. Test result confirmation 13 Check the displayed judgment results. If there is a [Fail], see “4.5.3 Performance Criteria and Corrective Action”. 4 Weekly Test 4-10 11. QC Software shutdown 14 Press [Save] and then [Close] to exit the QC Software. 15 Select [Study completed] to finish the test. FDR Mammography QC Manual - 897N101461A 4.5.2 Test Result Evaluation and Judgment Evaluate and judge the Weekly Test results. If all items are judged as [Pass], the Weekly Test is finished. If there is an item judged as [Fail], take corrective actions by following “4.5.3. Performance Criteria and Corrective Action”. 4.5.3 [1] Performance Criteria and Corrective Action Test with 1Shot Phantom [Any failed test item} 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUWASSELECTED s4HESAME8RAYEQUIPMENTSETTINGSTARGETFILTERK6PAND!%#SETTINGSHAVEBEEN USEDDURINGTHETEST s4HE3HOT0HANTOMWASCORRECTLYPOSITIONED s4HESAMECOMPRESSIONPLATEHASBEENUSEDDURINGTHETEST s4HECOMPRESSIONPLATEHASBEENINTHESAMEPOSITIONDURINGTHETEST )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM During the MEE / Annual survey performed by the Medical Phyiscist, If any of the items still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system. CNR. Performance Criteria: Baseline value ± 20 % Spatial Resolution. 2lp/mm: Baseline value ± 6 %, 4lp/mm: Baseline value ± 15 % Missed Tissue at Chest Wall. Less than 7mm. Uniformity. Performance Criteria: Baseline value ± 15 % Linearity/Beam Quality. Performance Criteria: Baseline value ± x [QL] Weekly Test During the Weekly test performed by the QC technologist, if any of the following items still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system: 4 During the Weekly test performed by the QC technologist, if an item other than those listed above still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. [2] Good practice Clean the equipment. FDR Mammography QC Manual - 897N101461A 4-11 4.6 4.6.1 Conducting Weekly ACR Phantom Test Procedure [Test flow] 4 Weekly Test 4-12 FDR Mammography QC Manual - 897N101461A 1. Patient information entry and exposure menu selection 1 Enter patient information and press [Next]. 2. Test with ACR Phantom (exposure) 4 Follow the instructions below to conduct the test item. [Tools] ACR Phantom, an acrylic disk*, and a densitometer*. 1. Position the ACR Phantom at the lateral center of the exposure table while aligning its edge with the chest wall-side edge of the exposure table. 2. Position the acrylic disk* wherever it does not interfere with the visual inspection of the ACR Phantom (as shown in the figure below). 2 Select [QC/TEST] in the exposure region list and [ACR Phantom] in the exposure menu list, and then press [Start study.]. *NOTE The acrylic sisk and densitometer are required only if the facility uses hardcopy (film) for primary interpretation. Nipple s ide Weekly Test AC R P hantom 4 C hes t wall s ide 3 The following exposure submenu is displayed. • ACR Phantom Acrylic dis k 3. Position the compression plate so that it comes in contact with the ACR Phantom. NOTE Make sure that excessive pressure is not applied to the Phantom. FDR Mammography QC Manual - 897N101461A 4-13 4. Make an exposure in “Auto” or “Semi” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 5. Image output 7 Select Exposure Submenu [ACR Phantom] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. 8 Check or measure the following items using the image output on film. Follow the recommended image evaluation method for the Phantom to evaluate the image. NOTE It may take some time until you can press [QC] after completing the exposure process. 3. QC Software startup 5 Press [QC] to display the “QC window”. 4. Exposure condition confirmation 4 6 Check the exposure conditions. Weekly Test TIP For performing image diagnosis on a softcopy display, check the items on the display. • 4-14 Fill in the following items in the worksheet. ACR Phantom Density at center of Phantom image: _____ Density inside the disk: _____ Density outside the disk: _____ Fibers (ACR Phantom): _____ points Specks (ACR Phantom) : _____ points Masses (ACR Phantom): _____ points FDR Mammography QC Manual - 897N101461A 6. Test with ACR Phantom (user input) Press [User input]. Enter the measured density and visual inspection results, and then press [OK]. 9 9. QC Software shutdown 12 Press [Save] and then [Close] to exit the QC Software. 13 Select [Study completed] to finish the test. NOTE Though the Step Phantom is not part of this QC program, values for it must be entered. Enter any integer from 0 to 10. If an item is left unentered, or a negative value or value over 11 is entered, an error occurs. 4 Weekly Test 7. Auto-calculation 10 Press [Measurement execution]. The results are displayed at the bottom of the window. 8. Test result confirmation 11 Check the displayed judgment results. If there is a [Fail], see “4.6.3 Performance Criteria and Corrective Action”. FDR Mammography QC Manual - 897N101461A 4-15 4.6.2 Test Result Evaluation and Judgment Evaluate and judge the Weekly Test results. If all items are judged as [Pass], the Weekly ACR Phantom test is finished. If there is an item judged as [Fail], take corrective actions by following “4.6.3. Performance Criteria and Corrective Action”. See “8.2 Quick Guide for Software Operation” for test log display procedure. 4 Weekly Test 4-16 FDR Mammography QC Manual - 897N101461A 4.6.3 [3] Performance Criteria and Corrective Action Test with ACR Phantom Performance Criteria: Fibers = 4 or more Specks = 3 or more Masses = 3 or more Density at center = 1.40 ± 0.2 Density difference disk / outside = 0.4 or more and baseline value ± 0.05 Make s ure that the AC R P hantom is c orr ec tly pos itioned and then edo r the tes t. P as s The tes t is finis hed. 4 Weekly Test P oints to be c hec ked • The c orr ec t expos ure s ubmenu was s elec ted. • The s ame X-ray equipment s ettings (target/filter , kVp and AE C s ettings ) have been us ed. • The AC R P hantom, and ac rylic dis k (if us ed) were c orrec tly pos itioned. • The s ame c ompr es s ion plate has been us ed during the tes t. • The c ompr es s ion plate has been in the s ame pos ition during the tes t. • The dens itometer was c orrec tly us ed. • The dens ity was meased ur on the c orr ec t pos ition. • The quality c ontr ol of the viewing c ondition is c onduc ted. If any of the above is notecc t/appropriate, orr c orre c t the pr oblem. F ail C ontac t a F UJ IF IL M dealer . If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system. FDR Mammography QC Manual - 897N101461A 4-17 4.7 4.7.1 Conducting Printer Quality Control Procedure [Test flow] 1. TP rinter Quality C ontrol 4 Weekly Test 4-18 FDR Mammography QC Manual - 897N101461A 3. Printer Quality Control (measurement) 1 Follow the instructions below to conduct the Printer Quality Control test. NOTE The printer used to produce films for final interpretation of mammography images must be cleared by FDA for that purpose. The objective of this test is to assure that the printer used for final interpretation is performing according to the manufacturer’s specifications. NOTE Follow the printer manufacturer’s QC program for test frequency, procedure, performance and corrective action. If the manufacturer does not provide a mammography printer QC program, use the following: [Tools] 1. Print the QC Pattern. 2. Measure the image to identify the step which has an optical density closest to, but not less than 1.20. Designate this step as the Mid-Density (MD) step. 3. Measure the image to identify the step with an optical density closest to, but not less than 2.20. Designate this step as the High-Density (HD) step. 4. Measure the image to identify the step with an optical density closest to, but not less than 0.45. Designate this step as the Low-Density (LD) step. 5. The difference between the HD step and the LD step is designated as the Density Difference (DD). 6. Measure the lightest (Base+Fog or unexposed) portion of the image. This is designated as B+F. 7. Plot the Mid-Density (MD), the Density Difference (DD) and the Base+Fog (B+F) on the control chart . FDR Mammography QC Manual - 897N101461A 4 Weekly Test QC pattern with a fixed digital value grayscale (e.g. TG18-QC or TG18-PQC (contact your printer manufacturer for details), calibrated densitometer, Report Form (available from printer manufacturer). 7. Determine if any of the data points exceed the control limits. If not, proceed to step 9. If so, print a second QC pattern, double checking that the correct procedure is followed. If the same results are obtained, proceed to step 8. 8. Circle any out-of-control data points and repeat the test. If any data point is still out of control, correct the cause of the problem and repeat the test to confirm that the problem has been corrected. Note the cause of the problem and the corrective action and plot the in-control point. 9. Determine if there are any trends, i.e., three or more data points moving in one direction (either upwards or downwards) in MD, DD or B+F. If trends are present but the data points have not, as yet, exceeded the control limits, clinical mammogram images can be printed. It is necessary to determine the cause of the trend and to monitor the processor closely to assure that control limits are not exceeded. 7. Write the data and time on the film. label the film with the printer ID if there is more than one printer. Retain the film for the MQSA prescibed period, e.g. the last full month for daily QC and the last 12 weeks for weekly QC. [PERFORMANCE AND CORRECTIVE ACTION] If the MD and DD are within ± 0.15 of their respective operating levels, and the B+F is within ± 0.03 of its operating level, the printer is in control, and no further action is required. If the MD and DD fall outside the above control limits, corrective action must be performed and then confirmed by performing the QC test again. If this test produces results that fall outside the action limits as specified above or by the manufacturer (whichever is applicable), the source of the problem shall be identified and corrected before the device can be used for mammographic image interpretation. Clinical image acquisition may be continued. If available, an alternative FDA-approved for mammography display device may be used for mammographic image interpretation, provided it has passed the applicable QC tests. 4-19 4.8 4.8.1 Conducting Monitor Quality Control Procedure [Test flow] 1. Monitor Quality C ontrol 4 Weekly Test 4-20 FDR Mammography QC Manual - 897N101461A 4. Monitor Quality Control (measurement) 1 Follow the instructions below to conduct the test item. NOTE The objective of this test is to assure that monitors are performing according to the monitor manufacturer’s specifications and that they are calibrated (or configured) to perform the DICOM Grayscale Standard Display Function (GSDF). Both the monitors used fro final interpretation of mammography images (primary monitors), and monitors used for acquisition QC (secondary monitors) must be tested. An additional requirement is that monitors used for final interpretation must have been cleared by the FDA for that purpose. Follow the instructions below to conduct the Monitor Image Quality Test. 1. For both primary and secondary monitors: display the QC pattern. 2. Record the maximum and minimum luminance levels as directed by your monitor or QC software manufacturer. The maximum luminance (Lmax) is typically specified by the monitor manufacturer as the highest value that can be used without compromising other performance characteristics, such as lifetime or resolution. The minimum brightness (Lmin) is influenced by Lmax. 3. Examine the image carefully to determine the visibility of the following features and record the results on the report form: • Verify that the 5% signal level inset is visible in the larger 0% signal field (a). Pass/Fail • Verify that the 95% signal level inset is visible in the larger 100% signal field (b). Pass/Fail NOTE Follow the monitor manuafacturer’s QC program for test frequency, procedure, performance and corrective action. If the manufacturer does not provide a mammography monitor QC program, use the following: [Tools] • • • • Monitor image quality pattern:. AAPM TG18-QC, SMPTE RP133 or other patterns appropriate for the monitor to tested (contact your monitor or QC software manufacturer for details) For monitor luminance response: gray scale images (supplied by the monitor or QC software manufacturer or AAPM TG18-LN). Calibrated photometer Monitor QC (Annual) report form, your own, or the QC software manufacturer-provided report form. FDR Mammography QC Manual - 897N101461A • Verify that each gray level step from 0% to 100% can be distinguished individually (c). Pass/Fail 4 Weekly Test NOTE Commercial calibration software programs supporting the American Association of Physicists in Medicine (AAPM) Assessment of Display Performance for Medical Imaging Systems (AAPM On-Line Report No. 03, www. aapm.org) are available for use with this test. 2 • Verify that the alphanumeric characters in the image appear sharp and in focus (d). Pass/Fail • Verify that the high-contrast Line-pair patterns in the center and at the corners of the QC pattern are distinguishable (e). The low-contrast patterns are not evaluated in this test. Pass/Fail 4-21 [PRECAUTIONS AND CAVEATS] Perform this test using only the test patterns described above. Some test patterns are specific to the image matrix size and bit depth. Using other patterns could yield erroneous results. [PERFORMANCE] If the following conditions are met, the Monitor QC (Annual) test is acceptable: 4 Weekly Test 4-22 • The maximum luminance value (Lmax) is within the reange specified by the manufacturer. For reference, the AAPM report recommends that once set, the target level (Lmax) is maintained within ±10% of the desired value .. • The 5% inset in the 0% field, and the 95% inset in the 100% field shall be visible, and the 0% to 100% gray level steps shall be distinguishable from the adjacent steps. • The alphanumeric characters appear sharp and focused. • The high contrast line-pair patterns are distinguishable at the center and at the corners of the display in both the horizontal and vertical orientations. NOTE] For monitors featuring DICOM GSDF display functionality, confirm that the monitors are set for DICOM DISPLAY (GSDF) and that the monitors successfully meet the criterai of Monitor Image Quality Test Procedure step. FDR Mammography QC Manual - 897N101461A Chapter 5 Quarterly Test Page 5.1 Test Flow ______________________________________________ 5-2 5.2 Test Items _____________________________________________ 5-3 5.3 Conducting Quarterly Test ______________________________ 5-4 5-4 5-6 5-6 5.3.1 5.3.2 5.3.3 Procedure _______________________________________ Test Result Evaluation and Judgment _________________ Performance Criteria and Corrective Action ____________ 5 FDR Mammography QC Manual - 897N101461A 5-1 5.1 Test Flow Performed by the technologist, the Quarterly Test is designed to determine the number and cause of repeated radiographs. Analysis of this data will help identify ways to avoid multiple X-raying and reduce costs, as well as reduce patient exposure. )NSTALLATION MONTHS MONTHS MONTHS YEAR MONTHS 1UARTERLY4EST Repeated images shall be evaluated quarterly. In order for the repeat rates to be meaningful, a patient volume of at least 250 patients or 1,000 exposures is needed. As described above, the Quarterly Test is neither a constancy test nor a performance test of the system. Specify the criteria when conducting the test, not in advance. The Retake Analysis software, a software module for the FDR-1000AWS, is convenient for organizing and managing the repeat analysis data. Detailed operation of the Repeat Analysis software module can be found in Aspire HD (Amulet) Options Operation Manual. %QUIPMENTUSE 2EJECTEDIMAGE ANALYSIS 5 1UARTERLY4EST Quarterly Test YES 2EPEATANALYSIS 4HERESULTISWITHIN THECRITERIA NO 4AKINGCORRECTIVEACTIONS #HECKIFTHECRITERIAARESATISFIED )F NOT SATISFIED TAKE CORRECTIVE ACTIONS BY FOLLOWING 3OLUTIONS FOR &AILED4EST)TEMS 5-2 FDR Mammography QC Manual - 897N101461A 5.2 Test Items The Quarterly Test has the following test item. Test Items Repeat analysis Exposure Menus --- Exposure Submenus --- Items Auto-calculated and Judged Based on Input Data 1. Exposure 2. Input Values --- --- Items Judged [Pass/Fail] Based on Visual Inspection by a User --- The Quarterly Test requires no settings since the QC Software is not used in the Test. The Test also requires no tool. Simply record the result of the Repeat analysis in the report form. 5 Quarterly Test FDR Mammography QC Manual - 897N101461A 5-3 5.3 5.3.1 Conducting Quarterly Test Procedure [Test flow] 2EPEATANALYSISCOLLECTINGREJECTEDIMAGES 2EPEATANALYSISCALCULATINGREPEATRATES 4ESTRESULTCONFIRMATION TIP 1. Repeat analysis (collecting rejected images) 5 Quarterly Test 1 Start by removing all existing rejected images in the department taken prior to the start of the analysis. 2 Take inventory of the image supply as a starting point to determine the total number of images consumed during this test. 3 Start collecting all rejected images. Continue to collect for the length of time needed to radiograph at least 250 consecutive patients. 4 Sort the rejected images into categories such as poor positioning, motion, compression, under exposure, (these might be due to exposure or processing), artifacts (streaks, spots, etc.). Good images (they appear to be acceptable mammograms when retrospectively evaluated during the Repeat analysis) may have also been repeated. Some images may not have resulted in an additional exposure of the patient but may have also been rejected. These include clear and QC images. Although it is appropriate to include wire localization images as part of the reject analysis, they should not be included in the repeat analysis because they are taken as part of the wire localization process. 2. Repeat analysis (calculating repeat rates) 5 Some facilities placing all images (repeated and good images) in the patient’s film jacket have no repeated images in the department. In this case, the reject/repeat analysis chart is completed as patient examinations are carried out. 6 Tabulate the counts from Steps 4 and 5 , determining the total number of repeated images, rejected images, and the total number of images exposed during the analysis period. NOTE Rejected images are all images that are in the reject bin, including repeated images. Repeated images are images that are retaken because of inadequate quality. The reject bin does not include additional views required to image selected tissue seen on the first image. It also does not include images taken for the purposes of including tissue that could not be positioned on the image receptor due to the size of the breast. For facilities using softcopy for final interpretation maintain a list of repeated images using the “REPEAT RATE ANALYSIS“ in “8.3 Worksheets”. 5-4 FDR Mammography QC Manual - 897N101461A 7 Determine the overall percentage of repeated images by dividing the total number of repeated images by the total number of images exposed during the analysis period, then multiply by 100. Next, determine the overall percentage of rejected images by dividing the total number rejected images by the total number of images exposed during the analysis period, and multiply by 100. 8 Determine the percentage of repeats in each “reason for repeat” category by dividing the repeats in the category by the total number of repeated images and multiply by 100. 3. Test result confirmation If the total repeat or reject rate changes from the previously determined rate by more than 2.0 percent of the total films included in the analysis, the reason(s) for the change shall be determined. Any corrective actions shall be recorded and the results of these corrective actions shall be assessed.. 5 Quarterly Test FDR Mammography QC Manual - 897N101461A 5-5 5.3.2 Test Result Evaluation and Judgment Evaluate and judge the Quarterly Test results. If the criteria are satisfied, the test is completed. If not satisfied, take corrective actions by following “5.3.3 Performance Criteria and Corrective Action”. 5.3.3 Performance Criteria and Corrective Action If the total repeat or reject rate changes from the previously determined rate by more than 2.0 percent of the total films included in the analysis, the reason(s) for the change shall be determined. Any corrective actions shall be recorded and the results of these corrective actions shall be assessed. Any corrective action should be recorded on the bottom of the “REPEAT RATE ANALYSIS” in “8.4 Report Forms”. The effectiveness of the corrective actions must be assessed by performing another repeat analysis after the corrective actions have been implemented. It is important to study films that are too dark or too light to determine if the underlying cause is the exposure equipment, image printer, patient positioning, technique or sub-optimal setting of digital image processing. If this test produces results that fall outside the action limits as specified, the source of the problem shall be identified and corrective action shall be taken within thirty days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. 5 Quarterly Test 5-6 FDR Mammography QC Manual - 897N101461A Chapter 6 Semi-annual Test Page 6.1 Test Flow ______________________________________________ 6-2 6.2 Test Items _____________________________________________ 6-2 6.3 Tools _________________________________________________ 6-2 6.5 Conducting Semi-annual Test ___________________________ 6.5.1 Procedure _______________________________________ 6.5.2 Test Result Evaluation and Judgment _________________ 6.5.3 Solutions for Failed Test Items _______________________ 6-2 6-2 6-4 6-4 6 FDR Mammography QC Manual - 897N101461A 6-1 6.1 Test Flow Performed by the technologist, the Semi-annual Test provides the basic test items for checking that performance of the X-ray equipment is maintained. Detailed tests for this purpose are conducted in the Annual Test. )NSTALLATION MONTHS YEAR YEARS YEARS YEARS 3EMIANNUAL4EST 6.2 Test Items The Semi-annual Test consists of the following test items. Items Judged Based on Input Data Test Items Exposure Menu Exposure Submenus Compression device confirmation --- --- 1. Exposure 2. Input Values Items Judged [Pass/ Fail] Based on Visual Inspection by a User --- ✓ Judgment methods: 1. User selection of [Pass/Fail] based on force / bathrrom scale measurements. 6 Semi-annual Test 6.3 Tools Tools to be used for the Semi-annual Test based on this Manual are shown below. See “2.3 Tools” for details. Tools [ B ] Worksheet (See Chapter 8) [ J ] Analog bathroom scale or Force scale [ K ] Towels 6.5 6.5.1 Conducting Semi-annual Test Procedure [Test flow] 1. C ompres s ion devic e c onfirmation (meas urment) 6-2 FDR Mammography QC Manual - 897N101461A 1. Compression device confirmation (measurement) 1 Follow the instructions below to conduct the test item. [Tools] A analog bathroom (or force) scale and several towels. 1. Place a towel on the cassette holder (to protect the cassette holder), then place the analog bathroom scale on the towel, with the dial or read-out positioned for easy reading. Locate the center of the scale directly under the compression device. . 2. Place several towels on top of the scale to prevent damage to the compression device.. Towels Bathroom Scale 4. Read the bathroom scale measurement and record it on the worksheet. • Automatic compression force measured value: _____lbs. 5. Release the compression device 6 Semi-annual Test 3. Using the power drive, activate the compression device and allow operation until it stops automatically. 6. Using the manual drive, move the compression device downward until it stops. 7. Read the bathroom scale measurement and record it on the worksheet. • Manual compression force measured value: _____lbs.l 8. Release the compression device. FDR Mammography QC Manual - 897N101461A 6-3 6.5.2 Test Result Evaluation and Judgment A compression force of at least 111 newtons (25pounds) shall be provided. The maximum compression force for the initial power drive mode must be between 111 newtons (25 pounds) and 200 newtons (45 pounds). If these requirements are not met, a qualified service engineer must make the appropriate internal adjustments and the system retested and pass the test before further exams are performed. . 6.5.3 [1] Performance Criteria and Corrective Action Compression device confirmation Performance Criteria: Compression force: At least 111 newtons (25 pounds) Maximum compression force: between111 newtons (25 pounds) and 200 newtons (45 pounds) 6 Semi-annual Test If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system 6-4 FDR Mammography QC Manual - 897N101461A Chapter 7 Annual Test Page FDR Mammography QC Manual - 897N101461A 7.1 Test Flow ______________________________________________ 7-2 7.2 Test Items _____________________________________________ 7-3 7.3 Tools _________________________________________________ 7-4 7.4 Setting Confirmation ___________________________________ 7-5 7.5 Conducting Annual Test ________________________________ 7.5.1 Initial Performance Test ____________________________ 1. Procedure ____________________________________ 2. Test Result Evaluation and Judgment ____________ 3. Performance Criteria and Corrective Action ______ 7.5.2 Image Performance Test____________________________ 1. Procedure ____________________________________ 2. Test Result Evaluation and Judgment ____________ 3. Performance Criteria and Corrective Action ______ 7.5.3 Spatial Resolution Test _____________________________ 1. Procedure ____________________________________ 2. Test Result Evaluation and Judgment ____________ 3. Performance Criteria and Corrective Action ______ 7.5.4 X-ray Equipment Performance Test ___________________ 1. Procedure ____________________________________ 2. Test Result Evaluation and Judgment ____________ 3. Performance Criteria and Corrective Action ______ 7.5.5 AEC System Performance Test _______________________ 1. Procedure ____________________________________ 2. Test Result Evaluation and Judgment ____________ 3. Performance Criteria and Corrective Action ______ 7-7 7-7 7-7 7-16 7-16 7-19 7-19 7-25 7-25 7-27 7-27 7-30 7-30 7-31 7-31 7-43 7-43 7-45 7-45 7-56 7-56 7 7-1 7.1 Test Flow Performed by the Medical Physicist during the Medical Equipment Evaluation (MEE) and subsequent Annual testing, these tests are designed for checking the overall performance of the FDR Mammography system. The Annual Test is comprised of constancy tests and performance verification tests of the system. The constancy tests are designed for determining if variations of regularly-measured system performance values are within the allowable range (criteria) based on baseline values that were established at the time of QC program / system installation. The performance test is intended to check that system performance values are within the upper or lower limits specified by the baselines. It is necessary to determine the criteria before conducting the Annual Test. NOTE The reference criteria are shown on the “List of Reference Values for Criteria and Baseline Values”. See “3.3 Criteria Confirmation and Determination” for details. When conducting the Annual Test the first time after setting the criteria, specify the baseline values to be used in the future Annual Tests. In the second or later Annual Test, check that the variation from the specified baseline values is within the criteria. See “3.4 Baseline Value Settings” for details. Initial s etting, criteria and bas eline value s ettings Meas urement and performance check Viewing conditions 7 Firs t Annual Tes t Meas urement Annual Test E quipment us e Annual Tes t YES The res ult is within the criteria NO C ompres s ion device confirmation E xpos ure and meas urement (1) Image performance (2) S patial R es olution (3) X-ray equipment performance (4) AE C s ys tem performance (5) 1 S hot P hantom tes ts (6) AC R P hantom tes t QC S oftware Us er input Auto-calculation J udgment Taking corrective actions C heck that the variation from the s pecified bas eline values is within the criteria, or that the values indicating s ys tem performance s atis fy the criteria. If the criteria are s atis fied, equipment can be us ed as is until the next Annual Tes t day. If not s atis fied, take corrective actions by following "P erformance C riteria and C orrective Action" in the relevant s ubs ection. 7-2 FDR Mammography QC Manual - 897N101461A 7.2 Test Items The Annual Test consists of the following test items. Test Types/Items Exposure Menus Exposure Submenus Items Autocalculated and Judged Based on Input Data 1. Exposure ✓ Viewing environment - - Viewbox maintenance Annual A - Printer Quality Control - - ✓ Monitor Quality Control - - ✓ ✓ Viewing Conditions Compression device confirmation Force, display accuracy and sustainability Image basic test Annual A Annual A Uniform Expo (Image Basic) ✓ ACR Phantom1(Lag) ✓ Additive lag effects (Lag) Image performance Multiplicative lag effects (Ghost) Spatial Resolution (Magnification) Half Value Layer (HVL) ✓ ✓ ACR Phantom2(Ghost) ✓ Uniform Expo.2(Ghost) ✓ Annual 2/6 Annual 3/6 Collimation assessment 1ShotM(Magnification) ✓ ✓ ✓ - ✓ - AEC reproducibility - AGD mode 1 ✓ ✓ ✓ ✓ Mode1 CNR (PMMA20mm) ✓ Mode1 CNR (PMMA40mm) ✓ Mode1 CNR (PMMA60mm) ✓ Mode1 CNR (PMMA70mm) ✓ Annual 4/6 AEC system performance ✓ ✓ Radiation output CNR mode 1 ✓ - Uniform Expo.3(Collimation) 7 ✓ - ✓ ✓ Constancy Performance 1 Shot Phantom tests 1Shot PhantomM - ✓ ✓ Test with ACR Phantom ACR phantom scoring ACR Phantom ✓ ✓ ✓ Three judgment methods: 1. Auto-calculation by the QC Software based on the image data acquired by making an exposure. 2. Auto-calculation by the QC Software based on the input measurement data. 3. User selection of [Pass/Fail] based on visual inspection of the image. Annual Test Calcualtion 2/2 AGD-ACR Phantom ✓ ✓ - kVp accuracy and reproducibility X-ray equipment performance Non Exposure(Lag) Uniform Expo.1(Ghost) Annual 1/6 Visual and Functional test Spatial Resolution 2. Input Values Items Judged [Pass/Fail] Based on Visual Inspection by a User . FDR Mammography QC Manual - 897N101461A 7-3 7.3 Tools Tools to be used for the Annual Test based on this Guidebook are shown below. See “2.3 Tools” for details. Tools [ A ] QC Software [ B ] Worksheet (See Chapter 8) [ C ] 1Shot Phantom [ D ] Viewing box (if film is used) [ E ] Timer or watch/clock [ F ] ACR Phantom [ G ] Dosimeter [ H ] Luminance meter [ I ] kVp meter [ J ] Force scale [ N ] PMMA phantoms (available, in combination, for 20, 40, 60 and 70 mm in thickness) [ O ] Aluminum plate (0.2 mm) for CNR measurement [ P ] Aluminum plates for half value layer measurement (0.3 and 0.5 mm) [ Q ] Scale [ R ] Coins [ S ] Lead sheet 7 Annual Test 7-4 FDR Mammography QC Manual - 897N101461A 7.4 Setting Confirmation The following items described in Chapter 3 are required to be specified before conducting the Annual Test. • • • Initial settings ....................................3.2 Initial Settings Criteria..................................................3.3 Criteria Confirmation and Determination Baseline values..................................3.4 Baseline Value Settings [Items allowing criteria change] Test Items Judgment Items Criteria Luminance (cd/m2) 3500 cd/m2 or more Interpretation room illuminance / ambient light (lx) 50 lx or less Compressed breast thickness accuracy (mm) ±5 mm or less Compression force accuracy (N) ± 20 N or less Maximum compression force (N) 111 to 200 N Image basic test Relative sensitivity (S value) Baseline value ± 40 % Additive lag effects (Lag) Lag factor [QL] 75 QL or less Multiplicative lag effects (Ghost) Ghost factor Below 0.045 MTF (4lp/mm) Baseline value ± 12 % MTF (8lp/mm) Baseline value ± 15 % Low volt. accuracy (LFS) [kV] Specified kVp ± 1 kV Mid. volt. accuracy (LFS) [kV] Specified kVp ± 1 kV High volt. accuracy (LFS) [kV] Specified kVp ± 1 kV Viewbox mainitenance Compression device confirmation Spatial Resolution (Magnification) kVp accuracy and reproducibility Half Value Layer (HVL) Specified kVp ± 1 kV Specified kVp ± 1 kV High volt. accuracy (SFS) [kV] Specified kVp ± 1 kV Mid. volt. reproducibility (LFS) [kV] Reproducibility ± 0.5 kV or less HVL (Mo/Mo) kVp / 100 HVL (Mo/Rh) kVp / 100 HVL (W/Rh) kVp / 100 X-ray / Light field gap (Right/Left) [mm] Total gap criteria = SID X 0.02 or less X-ray / Light field gap (Chest/Nipple) [mm] Total gap criteria = SID X 0.02 or less X-ray / Image receptor field gap (Left) [mm] SID X 0.02 or less X-ray / Image receptor field gap (Right) [mm] SID X 0.02 or less X-ray / Image receptor field gap (Nipple) [mm] SID X 0.02 or less X-ray field / Exposure table gap [mm] 5 mm or less Reproducibility Variation coefficient 0.05 or less Collimation assessment Radiation output FDR Mammography QC Manual - 897N101461A Air kerma rate [mGy/s] 7 mGy/s or more Specific radiation output [μGy/mAs] 30 μGy/mAs or more 7 Annual Test Low volt. accuracy (SFS) [kV] Mid. volt. accuracy (SFS) [kV] 7-5 [Items allowing criteria change] Test Items Judgment Items Criteria Entrance air kerma accuracy [%] Average ± 15 % or less Entrance air kerma reproducibility Variation coefficient 0.05 or less CNR relative value 20mm [%] Baseline value 105 % or more CNR relative value 40mm [%] Baseline value 95 % or more CNR relative value 60mm [%] Baseline value 63 % or more CNR relative value 70mm [%] Baseline value 60 % or more AGD 20 mm [mGy] 1 mGy or less AEC reproducibility CNR mode 1 AGD 40 mm [mGy] 2 mGy or less AGD 60 mm [mGy] 4.5 mGy or less AGD 70 mm [mGy] 6.5 mGy or less AGD-ACR Phantom 3 mGy or less AGD mode 1 AGD-ACR Phantom TIP • The criteria setting is not required for visual inspection. • CNR mode 1 corresponds to the CNR_H-mode • AGD mode 1 corresponds to the AGD_H-mode 7 Annual Test 7-6 FDR Mammography QC Manual - 897N101461A 7.5 Conducting Annual Test The Annual Test items are classified into “7.5.1 Initial Performance Test”, “7.5.2 Image Performance Test”, “7.5.3 Spatial Resolution Test”, “7.5.4 X-ray Equipment Performance Test”, “7.5.5 AEC System Performance Test” and “7.5.6 ACR Phantom Evaluation”; these five tests can be conducted separately. 7.5.1 Initial Performance Test 1. Procedure [Test flow] The Initial performance test consists of the Viewing environment evaluation, Viewbox maintenance, Printer Quality Control, Monitor Quality Control, Image basic test, and Compression device confirmation. 1. Viewing environment (meas urement) 2. Viewbox maintenanc e (meas urement) 3. P rinter Quality C ontrol (meas urement) 4. Monitor Quality C ontrol (meas urement) 5. Image bas ic tes t (meas urement) 6. P atient information entry and expos ure menu s elec tion 7. Image bas ic tes t (expos ure) 8. C ompres s ion devic e c onfirmation (meas urement) 9. QC S oftware s tartup 7 10. E xpos ure c ondition c onfirmation 12. Image bas ic tes t (us er input) Annual Test 11. Image output 13. C ompres s ion devic e c onfirmation (us er input) 14. Viewbox maintenanc e (us er input) 15. Auto-c alc ulation 16. Tes t res ult c onfirmation 17. QC S oftware s hutdown FDR Mammography QC Manual - 897N101461A 7-7 1. Viewing Environment (measurement) 1 Follow the instructions below to conduct the Viewing Environment evaluation. NOTE This evaluation is specific to the viewbox and reading room conditions, for images for final interpretation. NOTE [HARDCOPY] When interpreting hardcopy (film) images, it is suggested that you view the printed Aspire HD images under the same conditions as you would screen-film mammography images. Film masking devices and “hot” lights must be available for use by the interpreting physician. The 1999 ACR mammography Quality Control Manual includes a section on viewbox luminance, room illuminance, and msking, which may be used as a reference. NOTE [SOFTCOPY] When interpreting softcopy images, follow the monitor manufacturer’s quality control procedures. For monitors without a manufacturer-provided QC program, follow the monitor QC procedures provided in this manual. Set the final interpretation illumination to the lower of: environment • that recommended in the monitot manufacturer’s quality control program, if applicable, or • 20lux or lower. 7 Annual Test If the illumination is greater than the value set as described above, the source of the problem must be identified and corrective action taken before interpreting mammograms under these conditions. 2. Viewbox maintenance (measurement) 2 Follow the instructions below to conduct the test item. [Tools] A viewbox, luminance meter, and illuminance meter NOTE Check the viewbox performance under the normal use condition. 1. Turn on the viewbox and wait until the luminance becomes stable. 2. Measure the luminance at the center of the viewbox by using a luminance meter. • Fill in the following item in the worksheet. Luminance: _____cd/m2 3. Visually check if the viewbox luminance is uniform. • Fill in the following item in the worksheet. Viewbox uniformity: Pass/Fail 4. When the viewbox has a luminance adjustment function, check if the function works correctly. • Fill in the following item in the worksheet. Luminance adjustment function: Pass/Fail 5. When the viewbox has a masking function, check if the function works correctly. • Fill in the following item in the worksheet. Masking: Pass/Fail 6. Measure the illuminance near the viewbox, as a sample illuminance in the viewing condition, by using an illuminance meter. • Fill in the following item in the worksheet. Interpretation room illuminance (Ambient light): _____lx NOTE See the operation manuals for the viewbox, luminance meter or illuminance meter for how to measure the luminance or illuminance. NOTE It may take some time until you can press [QC] after completing the exposure process. 7-8 FDR Mammography QC Manual - 897N101461A 3. Printer Quality Control (measurement) 3 Follow the instructions below to conduct the Printer Quality Control test. NOTE The printer used to produce films for final interpretation of mammography images must be cleared by FDA for that purpose. The objective of this test is to assure that the printer used for final interpretation is performing according to the manufacturer’s specifications. NOTE Follow the printer manufacturer’s QC program for test frequency, procedure, performance and corrective action. If the manufacturer does not provide a mammography printer QC program, use the following: [Tools] QC pattern with a fixed geometric pattern (e.g. TG18-QC or SMPTE RP133 (contact your printer manufacturer for details)), Metric Ruler, Report Form (available from your printer manufacturer). 1. Print the QC Pattern. 2. Identify lines in each dimension that have a specified relationship (e.g. fixed length or distance between lines). Use these same lines for subsequent testing. 4. Using the ruler, confirm that the lines are straight and undistorted. 5. Inspect the image for artifacts. 6. Inspect the image for banding and jagged edges of the straight lines. [PERFORMANCE AND CORRECTIVE ACTION] The tolerance for film printer artifacts should be similar to the tolerance in screen-film mammography. According to the 1999 American College of Radiology Mammography Quality Control Manual, not all artifacts can be totally eliminated. It may be useful to use the concept of ALARA (as low as reasonably achievable) when assessing artifacts. If they can be easliy eliminated, they should. If the artifact is difficult or expensive to eliminate and is subtle (not mimicking or obscuring clinical information), it may be tolerable. The medical physicist should consult with the interpreting physician as to whether the artifact is tolerable. Tolerances for artifacts should be lower with new imaging equipment. If any of the following conditions exist: • The lines are bowed, bent or distorted • The lines exhibit jagged borders • The printer exhibits print or processing artifact that is judged intolerable • Measured distances vary by more than 1% from the tolerance standards • Artifacts are objectionable and can mimic or obscure clinical information The source of the problem shall be identified and corrective action shall be taken before the device can be used for mammographic image interpretation. Clinical image acquisition may be continued. If available, an alternative approved diagnostic display device may be used for mammographic image interpretation: 7 Annual Test 3. Measure the length of lines in the step above to establish a baseline. If the printer manufacturer provides a test image with specified distances, compare your measurement to the provided specification. Unless otherwise specified by the printer manufacturer, the visual observations should be made without magnification. 7. Note any unusual mechanical noises or other observations. [PRECAUTIONS AND CAVEATS] Perform this test using the QC pattern supplied (or recommended) by the manufacturer for that printer model. If none is available, use a QC test pattern recommended by the AAPM or by one of the approved MQSA accreditation bodies for this purpose. Using other patterns could yield erroneous results. FDR Mammography QC Manual - 897N101461A 7-9 4. Monitor Quality Control (measurement) 4 Follow the instructions below to conduct the test item. NOTE The objective of this test is to assure that monitors are performing according to the monitor manufacturer’s specifications and that they are calibrated (or configured) to perform the DICOM Grayscale Standard Display Function (GSDF). Both the monitors used fro final interpretation of mammography images (primary monitors), and monitors used for acquisition QC (secondary monitors) must be tested. An additional requirement is that monitors used for final interpretation must have been cleared by the FDA for that purpose. NOTE Commercial calibration software programs supporting the American Association of Physicists in Medicine (AAPM) Assessment of Display Performance for Medical Imaging Systems (AAPM On-Line Report No. 03, www. aapm.org) are available for use with this test. 5 Follow the instructions below to conduct the Monitor Image Quality Test. 1. For both primary and secondary monitors: display the QC pattern. 2. Record the maximum and minimum luminance levels as directed by your monitor or QC software manufacturer. The maximum luminance (Lmax) is typically specified by the monitor manufacturer as the highest value that can be used without compromising other performance characteristics, such as lifetime or resolution. The minimum brightness (Lmin) is influenced by Lmax. 3. Examine the image carefully to determine the visibility of the following features and record the results on the report form: • Verify that the 5% signal level inset is visible in the larger 0% signal field (a). Pass/Fail • Verify that the 95% signal level inset is visible in the larger 100% signal field (b). Pass/Fail NOTE Follow the monitor manuafacturer’s QC program for test frequency, procedure, performance and corrective action. If the manufacturer does not provide a mammography monitor QC program, use the following: 7 [Tools] • Annual Test • • • 7-10 Monitor image quality pattern:. AAPM TG18-QC, SMPTE RP133 or other patterns appropriate for the monitor to tested (contact your monitor or QC software manufacturer for details) For monitor luminance response: gray scale images (supplied by the monitor or QC software manufacturer or AAPM TG18-LN). Calibrated photometer Monitor QC (Annual) report form, your own, or the QC software manufacturer-provided report form. • Verify that each gray level step from 0% to 100% can be distinguished individually (c). Pass/Fail • Verify that the alphanumeric characters in the image appear sharp and in focus (d). Pass/Fail • Verify that the high-contrast Line-pair patterns in the center and at the corners of the QC pattern are distinguishable (e). The low-contrast patterns are not evaluated in this test. Pass/Fail FDR Mammography QC Manual - 897N101461A [PRECAUTIONS AND CAVEATS] Perform this test using only the test patterns described above. Some test patterns are specific to the image matrix size and bit depth. Using other patterns could yield erroneous results. [PERFORMANCE] If the following conditions are met, the Monitor QC (Annual) test is acceptable: • The maximum luminance value (Lmax) is within the reange specified by the manufacturer. For reference, the AAPM report recommends that once set, the target level (Lmax) is maintained within ±10% of the desired value .. • The 5% inset in the 0% field, and the 95% inset in the 100% field shall be visible, and the 0% to 100% gray level steps shall be distinguishable from the adjacent steps. • The alphanumeric characters appear sharp and focused. • The high contrast line-pair patterns are distinguishable at the center and at the corners of the display in both the horizontal and vertical orientations. NOTE] For monitors featuring DICOM GSDF display functionality, confirm that the monitors are set for DICOM DISPLAY (GSDF) and that the monitors successfully meet the criterai of Monitor Image Quality Test Procedure step 5. If so, you have successfully completed the Monitor QC (Annual) test. Proceed to step 7 FDR Mammography QC Manual - 897N101461A Follow the instructions below to conduct the Monitor Luminance Response Test. 1. Perform the luminance response test as described in the AAPM report. 2. Record the results of the evaluation on the report form. [PRECAUTIONS AND CAVEATS] Perform this test using only the test patterns described above. Some test patterns are specific to the image matrix size and bit depth. Using other patterns could yield erroneous results. [PERFORMANCE] Both primary and secondary monitors must be calibrated to DICOM Grayscale Standard Display Function (GSDF). The luminance differences between each measured value must agree with the expected difference associated with the GSDF and the measured contrast response at any given point must not exceed 10% of the standard for primary monitors and must not exceed 20% for the secondary monitors. [CORRECTIVE ACTION] For primary monitors (used for final interpretation): If the test produces results that fall outside the action limits as specified by the manufacturer, the source of the problem shall be identified and corrective action shall be taken before that device can be used for mammographic image interpretation. Clinical image acquisition may be continued. If available, an alternative approved diagnostic device may be used for mammographic image interpretation, provided it has passed the applicable QC tests. For secondary monitors (used for image acquisition QC): If this test produces results that fall outside the action limits as specified by the manufacturer, the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed using that monitor. If the monitor that failed is replaced by an alternative monitor that passes all tests, image acquisition may resume using the alternative monitor. 7 Annual Test For monitors that do not provide DICOM DISPLAY (GSDF), or fail the Monitor Image Quality Test Procedure step 5, perform the GSDF calibration and proceed to step 6 6 7-11 5. Image basic test (measurement) 7 Follow the instructions below to conduct the test item. 6. Patient information entry and exposure menu selection 8 Enter patient information and press [Next]. 9 Select [QC/TEST] in the exposure region list and [Annual A 1/1] in the exposure menu list, and then press [Start study.]. 10 The following exposure submenu is displayed. • Uniform Expo(Image Basic) [Tools] A dosimeter, scale, and lead sheet 1. Remove the compression plate. 2. Place a lead sheet on the exposure table, and then position the dosimeter at the lateral center of the table, 60 mm away from the chest wall-side edge, and also 40 mm above the table. .IPPLESIDE %XPOSURETABLE ,ATERALCENTER MM $OSIMETERPOSITION #HESTWALLSIDE %XPOSURETABLE MM #HEST WALL SIDE 7 MM .IPPLE SIDE Annual Test 3. Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. Specify the kVp, mAs and target/filter determined in “3.4.2 Baseline Values for Annual A Test”, and measure the air kerma 3 times in “Manu” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ mR mR µGy , _____µGy , _____µGy Air kerma: _____mR NOTE Make sure to specify the same exposure conditions as those used in the baseline value setting. If they are not certain, newly determine the exposure conditions for the test by following “3.4.2 Baseline Values for Annual A Test”. 4. Remove the dosimeter & lead sheet. Select [Study completed] to finish the test. 7-12 FDR Mammography QC Manual - 897N101461A 7. Image basic test (exposure) 11 Follow the instructions below to conduct the test item. 4. Remove the PMMA phantom, then position the force scale at the lateral center of the exposure table, about 60 mm away from the chest wall-side edge. Set [Grid] to [OUT], and then make an exposure with the same conditions as those used in Substep 3 in Step 7 . • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ 8. Compression device confirmation (measurement) 12 Follow the instructions below to conduct the test item. [Tools] A force scale, timer or watch/clock, PMMA phantom (40 mm), and scale 1. Position the PMMA phantom (40 mm) on the exposure table. 2. Apply compression force most frequently used in clinical practice and record the display value for compressed breast thickness on the X-ray equipment. • Fill in the following item in the worksheet. Compressed breast thickness displayed value: _____mm 60mm NOTE Ensure that the compression force is evenly applied to the measurement part of the force scale. If compression is not even, the measurement may result in an incorrect value. 5. Apply the same compression force as that used in Substep 2 and record the display value for the compression force on the X-ray equipment and the measured value. • Fill in the following items in the worksheet. Compression force display value: _____N Compression force measured value: _____N 6. Check that the compression applied in Substep 5 lasts 1 minute or more. • Fill in the following item in the worksheet. Compression force sustainability: Pass/Fail 3. Measure thickness of the compressed PMMA phantom (40 mm) by using a scale. • Fill in the following item in the worksheet. Compressed breast thickness measured value: _____mm FDR Mammography QC Manual - 897N101461A 7 Annual Test 7. Increment the compression force from that applied in Substep 6 and record the maximum compression force measured on the force scale. • Fill in the following item in the worksheet. Maximum compression force: _____N 7-13 9. QC Software startup 13 16 Press [QC] to display the “QC window”. Check for artifacts and white-out in the printed image. In image diagnosis on a softcopy display, check these items on the display. • Fill in the following items in the worksheet. Artifact: Pass/Fail White out: Pass/Fail 10. Exposure condition confirmation 14 )MAGEFRAME Check the exposure conditions. 7HITEOUT 12. Image basic test (user input) 17 Press [User input]. 18 Enter the measured air kerma (from Substep 3 in Step 7 .) and visual inspection results ([Pass/Fail]). 11. Image output 15 Select Exposure Submenu [Uniform Expo(Image Basic)] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. NOTE Some dosimeters may require correction of the measured value according to the beam quality. 7 Annual Test 7-14 FDR Mammography QC Manual - 897N101461A 13. Compression device confirmation (user input) 19 Enter the measured values, display values and visual inspection result (Pass/Fail). 16. Test result confirmation 22 Check the displayed judgment results. If there is a [Fail], see “7.5.3 Performance Criteria and Corrective Action”. 17. QC Software shutdown 14. Viewing box maintenance (user input) 20 23 Press [Save] and then [Close] to exit the QC Software. 24 Select [Study completed] to finish the test. Enter the measured value and visual inspection results, and then press [OK]. 7 Make sure that the values are entered for all items. If there is an item left unentered, an error occurs. 15. Auto-calculation 21 Annual Test NOTE Press [Measurement execution]. The results are displayed at the bottom of the window. FDR Mammography QC Manual - 897N101461A 7-15 2. Test Result Evaluation and Judgement Evaluate and judge the Annual A Test results. If all items are judged as [Pass], the Annual A Test is finished. If there is an item judged as [Fail], take corrective actions by following “6.5.3 Performance Criteria and Corrective Action”. 3. Performance Criteria and Corrective Action [1] Image basic test [Artifact] resulted in [Fail] Performance Criteria: No clinically relevant artifacts shall be visible. 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUWASSELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEAIRKERMAUSEDFORTHEEXPOSURESWASAPPROPRIATE s4HEREARENOSCRATCHESORDIRTONTHEEXPOSURETABLEORCOMPRESSIONPLATE s4HEQUALITYCONTROLOFTHESOFTCOPYHARDCOPYOUTPUTEQUIPMENTISCONDUCTED s4HEVIEWINGCONDITIONISAPPROPRIATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. 7 [Relative sensitivity] resulted in [Fail] Performance Criteria: Baseline value ± 40% Annual Test 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUWASSELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEAIRKERMAUSEDFORTHEEXPOSURESWASAPPROPRIATE s4HEDOSIMETERWASCORRECTLYPOSITIONED s4HEDOSIMETERWASCORRECTLYUSED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTOR DOSIMETERMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. 7-16 FDR Mammography QC Manual - 897N101461A [White out] resulted in [Fail] Performance Criteria: No white out shall be visible. 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUWASSELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEREARENOSCRATCHESORDIRTONTHEEXPOSURETABLEORCOMPRESSIONPLATE s4HEQUALITYCONTROLOFTHESOFTCOPYHARDCOPYOUTPUTEQUIPMENTISCONDUCTED s4HEVIEWINGCONDITIONISAPPROPRIATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. [2] Compression device confirmation Performance Criteria: Compressed breast thickness accuracy: ± 5mm or less Compression force accuracy (N): ±20N or less Maximum compression force (N): 111N to 200N 7 Annual Test If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system. FDR Mammography QC Manual - 897N101461A 7-17 [3] Viewing box maintenance [Viewing box luminance] resulted in [Fail] Performance Criteria: 3500 cd/m2 or more 0OINTSTOBECHECKED s4HELUMINANCEMETERWASCORRECTLYUSED s;CDM=ISSPECIFIEDASTHEUNITOFLUMINANCE s4HELUMINANCEWASMEASUREDATTHECORRECTPOSITION )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL #ONTACTADEALEROFTHEVIEWINGBOXMANUFACTURER [Interpretation room illuminance (Ambient light)] resulted in [Fail] Performance Criteria: 50 lx or less 0OINTSTOBECHECKED s4HEILLUMINANCEMETERWASCORRECTLYUSED s;LX=ISSPECIFIEDASTHEUNITOFILLUMINANCE s4HEILLUMINANCEWASMEASUREDATTHECORRECTPOSITION )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL )MPROVETHEVIEWINGCONDITION 7 Annual Test 7-18 FDR Mammography QC Manual - 897N101461A 7.5.2 Image Performance Test 1. Procedure [Test flow] The image performance test consists of the Additive lag effects (Lag), Multiplicative lag effects (Ghost), Missed tissue on chest wall edge (40 mm height) and Visual and Functional test. 0ATIENTINFORMATIONENTRYANDEXPOSUREMENUSELECTION !DDITIVELAGEFFECTS,AGEXPOSURE -ULTIPLICATIVELAGEFFECTS'HOSTEXPOSURE -ISSEDTISSUEONCHESTWALLEDGEMMHEIGHTEXPOSURE 6ISUALAND&UNCTIONALTESTCHECKING 1#3OFTWARESTARTUP %XPOSURECONDITIONCONFIRMATION -ISSEDTISSUEONCHESTWALLEDGEMMHEIGHTMEASUREMENT -ISSEDTISSUEONCHESTWALLEDGEMMHEIGHTUSERINPUT 6ISUALAND&UNCTIONALTESTUSERINPUT !UTOCALCULATION 4ESTRESULTCONFIRMATION 1#3OFTWARESHUTDOWN As Missed tissue on chest wall edge (40 mm height) is not required in this QC program, the above Step “8. Missed tissue on chest wall edge (40 mm height) (measurement)” can be skipped, but to satisfy the QC software input, Steps “4. Missed tissue on chest wall edge (40 mm height) (exposure)” and “9. Missed tissue on chest wall edge (40 mm height) (user input)” must be performed. FDR Mammography QC Manual - 897N101461A Annual Test NOTE 7 7-19 2. Additive lag effects (Lag) (exposure) 1. Patient information entry and exposure menu selection 1 4 Enter patient information and press [Next]. Follow the instructions below to conduct the test item. Make 2 exposures using the following exposure submenus: [ACR Phantom1(Lag)] [Non Exposure(Lag)] [Tools] ACR Phantom 1. Position the ACR Phantom at the lateral center of the exposure table while aligning its edge with the chest wall-side edge of the exposure table. 2. Remove the compression plate. .IPPLESIDE 2 %XPOSURETABLE Select [QC/TEST] in the exposure region list and [Annual 1/6] in the exposure menu list, and then press [Start study.]. !#20HANTOM #HESTWALLSIDE 3 7 Annual Test The following exposure submenus displayed. • ACR1: ACR Phantom1(Lag) • NON: Non Exposure(Lag) • UNI 1: Uniform Expo.1(Ghost) • ACR 2: ACR Phantom2(Ghost) • UNI 2: Uniform Expo.2(Ghost) • 1ST40: 1Shot PhantomM(40mm) are 3. Specify 28 kV, 56 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ 4. Place a lead sheet on the exposure table, specify 23 kV, 2 mAs and Mo/Mo, and then make an exposure in “Manu” mode. NOTE Make exposures in Substeps 3 and 4 consecutively without much interval between them to obtain high-quality results. 7-20 FDR Mammography QC Manual - 897N101461A 5. Remove the ACR Phantom. 3. Multiplicative lag effects (Ghost) (exposure) 5 6. Specify 28 kV, 6.3 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ Follow the instructions below to conduct the test item. Make 3 exposures using the following exposure submenus: [Uniform Expo.1(Ghost)] [ACR Phantom2(Ghost)] [Uniform Expo.2(Ghost)] [Tools] ACR Phantom 1. Remove the compression plate. 2. Specify 28 kV, 6.3 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ 3. Position the ACR Phantom on the righthand half of the exposure table while aligning the edges with the lateral center line and the chest wall-side edge of the exposure table as shown in the figure below. NOTE The ACR Phantom position differs from the normal position used for the exposure. Make sure that the phantom covers the Calculation Area 2 without protruding into the left-hand half of the exposure table. .IPPLESIDE %XPOSURETABLE !#20HANTOM MM MM 6 Follow the instructions below to conduct the test item. Make an exposure using the following exposure submenu: [1Shot PhantomM(40mm)] NOTE This test is not required as part of this QC Program, but you must make an exposure to satisfy the QC software. Make an exposure using Exposure Submenu [1Shot PhantomM(40mm above)] and go to Step 7 . [Tools] 1Shot Phantom and a jig for 1Shot Phantom NOTE The 1Shot Phantom must be positioned 40 mm above the exposure table in this test. Use the jig to position the Phantom at the specified height. TIP Use the same exposure conditions as those used for the Phantom exposure in the Daily/Weekly Test. See “3.4.3 Baseline Values for Daily/Weekly Test” for details. 7 Annual Test #ALCULATION!REAS MMXMM 4. Missed tissue on chest wall edge (40 mm height) (exposure) #HESTWALLSIDE 4. Specify 28 kV, 56 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ FDR Mammography QC Manual - 897N101461A 7-21 1. Position the 1Shot Phantom using the jig above the exposure table. NOTE 5. Visual and Functional test (Checking) 7 Position the Phantom by pressing the corners against the chest wall-side edge of the jig. Follow the instructions below to conduct the test item. Check the following points and enter the judgment results (Pass/Fail) in the worksheet. 2. Move the compression plate as close to the Phantom as possible without making contact. ■ Items under [X-ray equipment test 2 - Visual and Functional test] on the [Annual1-1] tab • • • • TIP Reinstall the compression plate removed for the Multiplicative lag effects (Ghost) test. • • • • • Verify mechanical stability Verify that all moving parts move smoothly All locks and detents work properly No sharp, rough edges or other hazards including electrical hazards Operator technique control charts are posted Operator protected during exposure by adequate radiation shielding All indicator lights working properly Auto decompression can be overridden and status displayed Compression can be manually released / in any failure NOTE Beware that the jig does not overlap with the measurement points of the Missed tissue on chest wall edge. 7 Annual Test 3. Specify the exposure conditions closest to those used for the 1Shot Phantom exposure in the Daily/Weekly Test in “Manu” mode and make an exposure. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ Compressed breast thickness: _____mm Compression force: _____N 6. QC Software startup 8 Press [QC] to display the “QC window”. NOTE It may take some time until you can press [QC] after completing the exposure process. 7-22 FDR Mammography QC Manual - 897N101461A 7. Exposure condition confirmation 9 Check the exposure conditions. 11 Measure the missed tissue on the right and left sides of the chest wall edge in the output image. The circles in Figure A indicate the measurement points of missed tissue on chest wall edge. Figure B shows the enlarged view of the measurement points. Measure how much of the points is reflected in the exposure image. Use the same Pass/Fail judgment criteria as those used for the Missed tissue on chest wall edge measurement in the Daily/Weekly Test. • Fill in the following items in the worksheet. Missed tissue on chest wall edge (Right) [mm]: Pass/Fail Missed tissue on chest wall edge (Left) [mm]: Pass/Fail 8. Missed tissue on chest wall edge (40 mm height) (measurement) The Missed tissue on chest wall edge (40 mm height) is not required as part of this QC program, go to Step 12 . 10 Select the Exposure Submenu [1Shot PhantomM(40mm)] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. A B FDR Mammography QC Manual - 897N101461A #HESTWALLSIDE 7 Annual Test MM MM MM MM MM MM MM MM MM MM 7-23 9. Missed tissue on chest wall edge (40 mm height) (user input) 12 Press [User input]. 13 Select the [Annual1-1] tab and enter the results (Pass/Fail). If not measuring the Missed tissue on chest wall edge (40 mm height), select either of [Pass/Fail] and go to Step 14 . 11. Auto-calculation 15 Press [Measurement execution]. The results are displayed at the bottom of the window. 12. Test result confirmation 16 Check the displayed judgment results. If there is a [Fail], see “7.5.1 3. Performance Criteria and Corrective Action”. 13. QC Software shutdown 10. Visual and Functional test (user input) 14 17 Press [Save] and then [Close] to exit the QC Software. 18 Select [Study completed] to finish the test. Select the results ([Pass/Fail]) and press [OK]. 7 Annual Test NOTE Make sure that the result is selected for all items. If there is an item left unentered, an error occurs. 7-24 FDR Mammography QC Manual - 897N101461A 2. Test Result Evaluation and Judgment Evaluate and judge the results of the image performance test in the Annual Test. If all items are judged as [Pass], the image performance test is finished. If there is an item judged as [Fail], take corrective actions by following “7.5.2 3. Performance Criteria and Corrective Action”. See “8.2 Quick Guide for Software Operation” for test log display procedure. 3. Performance Criteria and Corrective Action [1] Additive lag effects (Lag) Performance Criteria: 75 QL or less 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUSWERESELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEAIRKERMAUSEDFORTHEEXPOSURESWASAPPROPRIATE s4HETESTWASDONEFOLLOWINGCORRECTPROCEDURE s4HE!#20HANTOMWASCORRECTLYPOSITIONED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. [2] Multiplicative lag effects (Ghost) 7 Performance Criteria: Below 0.045 2EDOTHETEST 0ASS Annual Test 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUSWERESELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEAIRKERMAUSEDFORTHEEXPOSURESWASAPPROPRIATE s4HETESTWASDONEFOLLOWINGCORRECTPROCEDURE s4HE!#20HANTOMWASCORRECTLYPOSITIONED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 4HETESTISFINISHED &AIL #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. FDR Mammography QC Manual - 897N101461A 7-25 [3] Missed tissue on chest wall edge (40 mm height) This test is not required in this QC program. [4] Visual and Functional test #ONTACTA&5*)&),-DEALER 7 Annual Test 7-26 FDR Mammography QC Manual - 897N101461A 7.5.3 Spatial Resolution Test This test measures spatial resolution in magnification mode. For contact mode, spatial resolution is measured and performed as a constancy test on a weekly basis by the QC technologist. See section 4.5 of this QC program for details. 1. Procedure [Test flow] The spatial resolution test has only a single test item. 0ATIENTINFORMATIONENTRYANDEXPOSUREMENUSELECTION 3PATIAL2ESOLUTION-AGNIFICATIONEXPOSURE 1#3OFTWARESTARTUP %XPOSURECONDITIONCONFIRMATION !UTOCALCULATION 4ESTRESULTCONFIRMATION 1#3OFTWARESHUTDOWN 7 Annual Test FDR Mammography QC Manual - 897N101461A 7-27 1. Patient information entry and exposure menu selection 1 2. Spatial Resolution (Magnification) (exposure) Enter patient information and press [Next]. 4 Follow the instructions below to conduct the test item. Make an exposure using the following exposure submenu: [1ShotM(Magnification)] [Tools] 1Shot Phantom 1. Install the magnification table of the X-ray equipment and the compression plate normally used in magnified exposure in clinical practice. 2 Select [QC/TEST] in the exposure region list and [Annual 2/6] in the exposure menu list, and then press [Start study.]. 2. Position the 1Shot Phantom being shifted to the right-hand side so that the area indicated by the broken line in the below figure is at the lateral center of the X-ray field. NOTE Position the Phantom for exposure, the same as in “3.4.1 Baseline Values for Annual Test”. 3 7 Annual Test 7-28 The following exposure displayed. • 1ShotM(Magnification) submenu is NOTE If the area in the broken line is out of the Xray field when the Phantom is positioned by pressing its corners against the chest wallside edge of the magnification table, shift the Phantom to the front. Make sure that the shifted Phantom is parallel to the chest wall-side edge. FDR Mammography QC Manual - 897N101461A 4. Exposure condition confirmation 6 NOTE The Phantom position must be the same for every exposure (the allowable positional error is within ± 5 mm). Check the exposure conditions. 5. Auto-calculation 7 • Fill in the following items in the worksheet. Press [Measurement execution]. The results are displayed at the bottom of the window. 1 Shot Phantom position Lateral shift distance: _____mm 3. Specify “Manu” mode for the X-ray equipment and make an exposure under the same exposure conditions as those used in “3.4.1 Baseline Values for Annual Test”. • Fill in the following items in the worksheet. kVp: ______kV mAs: ______mAs Target/Filter: ______/______ Compressed breast thickness: _____mm Compression force: _____N 6. Test result confirmation 8 Check the displayed judgment results. If there is a [Fail], see “7.5.2 3. Performance Criteria and Corrective Action”. 7 It may take some time until you can press [QC] after completing the exposure process. 9 Press [Save] and then [Close] to exit the QC Software. 10 Select [Study completed] to finish the test. Annual Test 7. QC Software shutdown NOTE 3. QC Software startup 5 Press [QC] to display the “QC window”. FDR Mammography QC Manual - 897N101461A 7-29 2. Test Result Evaluation and Judgment Evaluate and judge the result of the spatial resolution test in the Annual Test. If all items are judged as [Pass], the spatial resolution test is finished. If there is an item judged as [Fail], take corrective actions by following “7.5.3 3. Performance Criteria and Corrective Action”. See “8.2 Quick Guide for Software Operation” for test log display procedure. 3. Performance Criteria and Corrective Action [5] Spatial Resolution (Magnification) Performance Criteria: 4lp/mm: Baseline value ± 12% 8lp/mm: Baseline value ± 15% 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUWASSELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HE3HOT0HANTOMWASCORRECTLYPOSITIONED 4HECALCULATIONAREAOFSPATIALRESOLUTIONISATTHECENTEROFTHE8RAYFIELD 4HE3HOT0HANTOMISPOSITIONEDINPARALLELWITHTHECHESTWALLSIDEOFTHE MAGNIFICATIONTABLE s4HECOMPRESSIONPLATEWASCORRECTLYPOSITIONED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 4HE3PATIAL2ESOLUTION32INTHE $AILY7EEKLY4ESTHASBEEN OCCASIONALLYJUDGEDASh&AILv #HANGETHE3HOT0HANTOM POSITIONANDREDOTHETEST 4HE3PATIAL2ESOLUTION32INTHE $AILY7EEKLY4ESTHASBEEN JUDGEDASh0ASSvINEVERYTEST 0ASS 4HETESTISFINISHED &AIL 7 Annual Test 7-30 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. Clinical imaging and mammographic image interpretation may be continued during this period. FDR Mammography QC Manual - 897N101461A 7.5.4 X-ray Equipment Performance Test 1. Procedure [Test flow] The X-ray equipment performance test consists of the kVp accuracy and reproducibility, Half Value Layer (HVL), Collimation assessment and Radiation output. TIP In this test, conduct the three test items not requiring an X-ray image first and then enter their measured values all at once to save time and simplify QC Software operation. K6PACCURACYANDREPRODUCIBILITY (ALF6ALUE,AYER(6, 2ADIATIONOUTPUT 0ATIENTINFORMATIONENTRYANDEXPOSUREMENUSELECTION #OLLIMATIONASSESSMENTEXPOSURE 1#3OFTWARESTARTUP %XPOSURECONDITIONCONFIRMATION #OLLIMATIONASSESSMENTMEASUREMENT K6PACCURACYANDREPRODUCIBILITYUSERINPUT (ALF6ALUE,AYER(6,USERINPUT #OLLIMATIONASSESSMENTUSERINPUT !UTOCALCULATION Annual Test 2ADIATIONOUTPUTUSERINPUT 7 4ESTRESULTCONFIRMATION 1#3OFTWARESHUTDOWN FDR Mammography QC Manual - 897N101461A 7-31 6. Make exposures with the selected kVps and measure the values by using a kVp meter. In this Substep, make 5 exposures with the middle kVp. • Fill in the following items in the worksheet. LFS - volt. Low: _____kV LFS - volt. Mid.-1: _____kV 1. kVp accuracy and reproducibility 1 Follow the instructions below to conduct the test item. [Tools] A kVp meter and lead sheet 1. Remove the compression plate. LFS - volt. Mid.-2: _____kV LFS - volt. Mid.-3: _____kV 2. Place a lead sheet on the exposure table, and then position the kVp meter on the sheet. LFS - volt. Mid.-4: _____kV LFS - volt. Mid.-5: _____kV LFS - volt. High: _____kV TIP Position the X-ray detector of the kVp meter at the lateral center of the exposure table and near the chest wall-side edge. The position must be the same for every measurement for reproducibility. 7. Mount the magnification table of the X-ray equipment and specify the Small Focal Spot (SFS) for it. NOTE Measurement with the Small Focal Spot (SFS) cannot be performed unless the magnification table is used. 3. Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. 8. Make exposures with the selected kVps and measure the values by using a kVp meter. In measurement with the Small Focal Spot (SFS), make one exposure with each kVp. • Fill in the following items in the worksheet. SFS - volt. Low: _____kV SFS - volt. Mid.: _____kV SFS - volt. High: _____kV 4. Specify the Large Focal Spot (LFS) and “Manu” mode for the X-ray equipment and select Mo/Mo for the target/filter. NOTE The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 7 Annual Test 5. Select 3 kVp values used for the measurement. • Fill in the following items in the worksheet. kVp used for the tests Low: _____kV Mid.: _____kV High: _____kV TIP Select the kVp most frequently used in clinical practice, and the adjacent upper and lower values (e.g. 25 kV, 28 kV, 31 kV). 2. Half Value Layer (HVL) 2 Following the instructions conduct the test item. below to [Tools] A dosimeter, aluminum plates (0.3 and 0.5 mm), and a lead sheet 7-32 FDR Mammography QC Manual - 897N101461A NOTE Make the size of the X-ray field approximately equal to that of the detector of dosimeter by using a diaphragm (e.g. lead mask). If the size cannot be reduced to that extent, make the Xray field as small as possible. TIP Position the aluminum plates within 300 mm away from the focal spot without using the compression plate as shown below. This may require careful effort. 3. Specify the kVp values for all the target/ filter combinations in “Manu” mode, and calculate the mAs producing the air kerma of 500 to 850 mR (≈ 4365 to 7420 µGy) without using an aluminum plate. Then measure the air kerma for each setting. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Entrance air kerma - w/o Al: _____mR µGy NOTE The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. &OCALSPOT MM !DDITIONALFILTERS !LUMINUMPLATESFORHALF VALUELAYERMEASUREMENT 4. Position the 0.3 mm aluminum plate and make an exposure under the condition calculated in Substep 3 using each of the kVp and target/filter to measure the air kerma after the X-ray transmitted through the plate. • Fill in the following item in the worksheet. Entrance air kerma – Al 0.3 mm: ____mR µGy $IAPHRAGM,EADMASK $ETECTOROFTHEDOSIMETER MM %XPOSURETABLE MM 1. Remove the compression plate. 2. Place a lead sheet on the exposure table, and position the dosimeter at the lateral center of the table, 60 mm away from the chest wall-side edge, and also 40 mm above the table. %XPOSURETABLE ,ATERALCENTER Wear gloves when using the aluminum plate to keep it clean. 5. Position the 0.5 mm aluminum plate and make an exposure under the condition calculated in Substep 3 using each of the kVp and target/filter to measure the air kerma after the X-ray transmitted through the plate. • Fill in the following item in the worksheet. Entrance air kerma – Al 0.5 mm: ____ 7 Annual Test .IPPLESIDE NOTE MM $OSIMETERPOSITION #HESTWALLSIDE %XPOSURETABLE MM #HEST WALL SIDE MM .IPPLE SIDE FDR Mammography QC Manual - 897N101461A 7-33 NOTE 3. Radiation output 3 A dosimeter allowing air kerma rate or exposure time measurement is necessary. Air kerma rate [mGy/s] can be calculated from exposure time [s] as follows: Air kerma rate [mGy/s] = air kerma [mGy] / exposure time [s] (1 mR = 8.73 × 10-3 mGy) Follow the instructions below to conduct the test item. [Tools] A dosimeter and lead sheet 1. Remove the compression plate. 2. Place a lead sheet on the exposure table, and position the dosimeter at the lateral center of the table, 60 mm away from the chest wall-side edge, and also 40 mm above the table. .IPPLESIDE 5. Select [Study completed] to finish the test. 4. Patient information entry and exposure menu selection 4 Enter patient information and press [Next]. 5 Select [QC/TEST] in the exposure region list and [Annual 3/6] in the exposure menu list, and then press [Start study.]. 6 The following displayed. %XPOSURETABLE ,ATERALCENTER MM $OSIMETERPOSITION #HESTWALLSIDE %XPOSURETABLE MM #HEST WALL SIDE MM .IPPLE SIDE 3. Specify 28 kV, 56 mAs and Mo/Mo for the X-ray equipment in “Manu” mode. NOTE 7 Annual Test 7-34 The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 4. Make an exposure and measure the air kerma, mAs and air kerma rate. In this Substep, measure air kerma 5 times under the same exposure conditions. If the air kerma rate cannot be measured, measure the exposure time for calculating the air kerma rate. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Entrance air kerma (1): _____mR µGy Entrance air kerma (2): _____mR µGy Entrance air kerma (3): _____mR µGy Entrance air kerma (4): _____mR µGy Entrance air kerma (5): _____mR µGy Air kerma rate: _____mGy/s (Exposure time: _____s) exposure submenu • Uniform Expo.3(Collimation) FDR Mammography QC Manual - 897N101461A is 3. Specify 23 kV, 2 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 5. Collimation assessment (exposure) 7 Follow the instructions below to conduct the test item. Make an exposure using the following exposure submenus: [Uniform Expo.3(Collimation)] NOTE Go to Step 12 when using the Image Reader, which uses a cassette exclusively for mammography, as well as the X-ray equipment. A more detailed test can be conducted. NOTE [Tools] NOTE A coin It may take some time until you can press [QC] after completing the exposure process. 1. Attach a coin to the bottom surface of the compression plate while aligning the edge with the chest wall-side edge of the plate. The “Semi” and “Auto” modes may not work correctly for this test. Make sure to use “Manu” mode for exposure. 6. QC Software startup 8 Press [QC] to display the “QC window”. 7. Exposure condition confirmation 9 Check the exposure conditions. 2. Move the compression plate down onto the exposure table. 7 NOTE Take care that the compression plate is not scratched by the coin. FDR Mammography QC Manual - 897N101461A 10 Annual Test 8. Collimation assessment (measurement) Select Exposure Submenus [Uniform Expo.3(Collimation)] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. 7-35 11 Visually check the gap between the image and compression plate edge by using the chest wall side as reference. Observe the position of the Coin in the output image and record the result. Judgment A part of Coin image is missing → Pass (no gap) Coin is completely imaged → Fail (with gap) A &AIL 0ASS NOTE #OIN #OIN • Fill in the following item in the worksheet. Compression plate edge (Chest wall side as reference): Pass/Fail Visual inspection result: Pass (no gap)/ Fail (with gap) NOTE Be careful not to position the Coin (a) where it will be overlapped with the coins to be positioned in Substeps 3 and 4 or the boundary of the cassettes placed in Substep 2. 2. Position 2 cassettes (Cassettes B1 and B2, hereafter) over the exposure table by aligning their chest wall-side edges. #ASSETTE" #ASSETTE" The Medical Physicist may perform collimation assessment using the methods that he/she typically uses, then, go to Step 17 . Steps 12 to 16 are the detailed test procedure for the X-ray equipment and the FCR Image Reader, which uses a cassette exclusively for mammography. 7 5. Collimation assessment (exposure) <details> 12 Annual Test Follow the instructions below to conduct the test item. Make an exposure using the following exposure submenus: [Uniform Expo.3(Collimation)] 3. Turn on the light field lamp of the X-ray equipment and position 4 coins (Coins (b) to (e), hereafter) respectively on the 4 sides of the light field on the Cassettes B1 and B2. [Tools] The same size 2 cassettes (no need to be QC exclusive), a scale and 6 coins (familiar sized) NOTE A cassette for general exposure cannot be used. 1. Position a coin (Coin (a), hereafter) on the exposure table while aligning the edge with the chest wall-side edge of the exposure table. 7-36 FDR Mammography QC Manual - 897N101461A 6. Specify 28 kV, 56 mAs and Mo/Mo and make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N ,IGHTFIELD C D E B NOTE The “Semi” and “Auto” modes may not work correctly for this test. Make sure to use “Manu” mode for exposure. NOTE Be careful not to position the Coins (b) to (e) where they overlap with the Coin (a) or on the boundary of the cassettes. 4. Attach a coin (Coin (f), hereafter) to the bottom surface of the compression plate while aligning the edge with the chest wall-side edge of the plate. NOTE It may take some time until you can press [QC] after completing the exposure process. F 6. QC Software startup 13 Press [QC] to display the “QC window”. 7. Exposure condition confirmation NOTE Be careful not to position the Coin (f) where it overlaps with the Coins (a) to (e) or on the boundary of the cassettes. 5. Move the compression plate down onto the Cassettes B1 and B2. 14 Check the exposure conditions. 7 Annual Test NOTE Take care that the compression plate is not scratched by the coins. FDR Mammography QC Manual - 897N101461A 7-37 8. Collimation assessment (measurement) <details> 15 16 Select Exposure Submenus [Uniform Expo.3(Collimation)] from the exposure condition display field and press [Film output]. For performing image diagnosis on a softcopy display, press [Network output] to output the image to the network. Output the images read from the 2 cassettes positioned on the exposure table on film or on a softcopy display in the same way. Measure the values required for judging the items below. .IPPLESIDE !D !B C D B ,EFT E F !F 2IGHT A #HESTWALLSIDE !E !A )MAGEREADFROM THE&$2MAMMOGRAPHY SYSTEM 2. Measure and record the distances between the coins and the adjacent edges of the X-ray field on the output images read from Cassettes B1 and B2. If a part of coin image is missing, measure the length of the missing part. 8RAYFIELD ,IGHTFIELD • X-ray/Light field gap (Right/Left, Chest/ Nipple) [mm] • X-ray/Image receptor field gap (Right/Left/ Nipple) [mm] !C .IPPLESIDE "C C D B ,EFT 2IGHT • Compression plate edge (Chest wall side) 1. Measure and record the distances between the coins and the adjacent edges of the output image. If a part of coin image is missing, measure the length of the missing part. TIP 7 Annual Test Coin (a) → Positioned on the chest wall-side edge of the exposure table. Coins (b), (c), (d), (e) → Positioned on the 4 sides of the light field. Coin (f) → Positioned on the bottom chest wall-side edge of the compression plate. • Fill in the following items in the worksheet. Coin (a) A-a: _____mm Coin (b) A-b: _____mm Coin (c) A-c: _____mm Coin (d) A-d: _____mm Coin (e) A-e: _____mm Coin (f) A-f: _____mm 7-38 E "E F "D #HESTWALLSIDE )MAGEREADFROM #ASSETTE" "B "F )MAGEREADFROM #ASSETTE" • Fill in the following items in the worksheet. Coin (b) B-b: _____mm Coin (c) B-c: _____mm Coin (d) B-d: _____mm Coin (e) B-e: _____mm Coin (f) B-f: _____mm 3. Make a record of the [X-ray / Light field gap (Right/Left, Chest/Nipple)]. TIP The values measured for Coins (b) to (e) in Substep 2 indicate the gap amount. Record the values measured in Substep 2. • Fill in the following items in the worksheet. Left: B-d: _____mm Right: B-b: _____mm Chest wall: B-e: _____mm Nipple side: B-c: _____mm FDR Mammography QC Manual - 897N101461A 4. Calculate the X-ray/Image receptor field gap (Right/Left/Nipple). Observe how the coins are reflected in the images read from the FDR mammography system and Cassette B1/B2 and judge which of the 4 examples the reflected images belong to. Then calculate the gap by using the corresponding formula. TIP When the X-ray field is inside of the image receptor edge in the image read from the FDR mammography system, as shown in Eg: 1, the size of the gap can be determined by measuring the distance (“x” in the figure) from the image receptor edge to the X-ray field. TIP TIP Check the following coin images. Left: Coin (d) Right: Coin (b) Nipple side: Coin (c) Assign the value recorded in Substeps 1 and 2 for each coin to A and B in the corresponding formula. • Fill in the following items in the worksheet. Left: _____mm Right: _____mm Nipple side: _____mm FDR mammography system Cassette B1/B2 X-ray/Image receptor field gap X calculation formula Not needed. The size of the gap can be calculated only from the image read from the FDR mammography system. X=x (Measure the distance x of the area not exposed to X-rays in the image read from the FDR mammography system.) )0 Eg: 1 ! X 8RAYFIELD )0 X=A+B Eg: 2 ! When the X-ray field is outside of the image receptor edge in the image read from the FDR mammography system as shown in Egs: 2, 3 and 4, the size of the gap can be calculated. Measure the distance “B” between the image receptor edge and that of Coin (on the chest wall-side edge) and the distance “A” between the image receptor edge and the edge of Coin (on the light field) in the image read from Cassette B, and the distance “X” between the edge of Coin and X-ray field edge, and then assign the measured values to the formula. 5. Calculate the X-ray field/exposure table gap. Observe how the coins are reflected in the images read from the FDR mammography system and Cassette B1/ B2 and judge which of the 3 examples the reflected images belongs to. Then calculate the size of the gap by using the corresponding formula (If the calculated value is a negative, derive the absolute value). " TIP Eg: 3 X=A-B " ! 7 Annual Test )0 Check the following coin images. Chest wall: Coins (a) and (e) Assign the value recorded in Substeps 1 and 2 for each coin to A and B in the corresponding formula. • Fill in the following item in the worksheet. Chest: _____mm )0 Eg: 4 X=B-A ! " FDR Mammography QC Manual - 897N101461A 7-39 FDR mammography system )0 8RAYFIELD Eg: 1 Y !A #OINA Cassette B1/B2 Not needed. The size of the gap can be calculated only from the image read from the FDR mammography system. 6. Visually check the gap between the image and compression plate edge by using the chest wall side as reference. Observe the position of the Coin (f) in the image read from the FDR mammography system and record the result. X-ray field/ exposure table gap Y calculation formula Y=y (Measure the distance y between the edges of Coin (a) and X-ray field in the image read from the FDR mammography system.) Judgment A part of Coin (f) image is missing → Pass (no gap) Coin (f) is completely reflected → Fail (with gap) Y=(A-e)-{(Aa)+(B-e)} Eg: 2 !E #OINE !A #OINA Eg: 3 #OINE "E !E #OINE Y=(A-e)+{(B-e)(A-a)} #OINF #OINF !A #OINA #OINE • Fill in the following item in the worksheet. Compression plate edge (Chest wall side as reference): Pass/Fail Visual inspection result: Pass (no gap)/ Fail (with gap) TIP When the X-ray field is inside of the image receptor edge in the image read from the FDR mammography system, as shown in Eg: 1, the size of the gap can be determined by measuring the distance (“y” in the figure) from the image receptor edge to the X-ray field. TIP 7 Annual Test 7-40 &AIL 0ASS "E When the X-ray field is outside of the image receptor edge in the image read from the FDR mammography system as shown in Egs: 2 and 3, the size of the gap can be calculated. Measure the distance ”A-a” between the image receptor edge and that of Coin (a) (on the chest wall-side edge) and the distance “A-e” between the image receptor edge and the edge of Coin (e) (on the light field) in the image read from Cassette B, and the distance “B-e” between the edge of Coin (e) and X-ray field edge, and then assign the measured values to the formula. 9. kVp accuracy and reproducibility (user input) 17 Press [User input]. 18 Select [Annual test 3-1]. Enter the display values and measured values. FDR Mammography QC Manual - 897N101461A 10. Half Value Layer (HVL) (user input) 19 Select [Annual Test 3-2]. Enter the measured entrance air kerma for each target/filter. 12. Radiation output (user input) 21 NOTE NOTE Some dosimeters may require correction of the measured value according to the beam quality. Some dosimeters may require correction of the measured value according to the beam quality. NOTE 11. Collimation assessment (user input) 20 Enter the measured entrance air kerma, mAs and air kerma rate, and then press [OK]. Make sure that the values are entered for all items. If there is an item left unentered, an error occurs. Select [Annual test 3-3]. Enter the measured values and visual inspection result (Pass/Fail). 13. Auto-calculation NOTE Enter all items even if the test is conducted without using cassettes. Although any value can be entered (e.g. “0” for all items), a negative value or significantly large one causes an error to occur. 22 Press [Measurement execution]. The results are displayed at the bottom of the window. 7 23 FDR Mammography QC Manual - 897N101461A Check the displayed judgment results. If there is a [Fail], see “7.5.4 3. Performance Criteria and Corrective Action”. Annual Test 14. Test result confirmation 7-41 15. QC Software shutdown 24 Press [Save] and then [Close] to exit the QC Software. 25 Select [Study completed] to finish the test. 7 Annual Test 7-42 FDR Mammography QC Manual - 897N101461A 2. Test Result Evaluation and Judgement Evaluate and judge the results of the X-ray equipment performance test in the Annual Test. If all items are judged as [Pass], the test is finished. If there is an item judged as [Fail], take corrective actions by following “7.5.4 3. Performance Criteria and Corrective Action”. 3. Performance Criteria and Corrective Action [6] kVp accuracy and reproducibility Performance Criteria: Accuracy ± 1kVp, Reproducibility ± 0.5kVp 0OINTSTOBECHECKED s4HEK6PMETERWASCORRECTLYPOSITIONED s4HEK6PMETERWASCORRECTLYUSED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PANDM!SWERECORRECT s4HEEXPOSUREWASAPPROPRIATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. [7] Half Value Layer (HVL) Performance Criteria: HVL > kVp/100 7 Annual Test 0OINTSTOBECHECKED s4HEDOSIMETERWASCORRECTLYPOSITIONED s4HEDOSIMETERWASCORRECTLYUSED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PANDM!SWERECORRECT s4HEEXPOSUREWASAPPROPRIATE s4HEUNITOFAIRKERMAM2OR«'YWASCORRECTLYUSED s4HEALUMINUMPLATEUSEDHASTHESPECIFIEDPURITY s4HEALUMINUMPLATETHICKNESSANDORPOSITIONWASCORRECTLYARRANGED s4HEREARENOSCRATCHESORDIRTONTHEALUMINUMPLATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. FDR Mammography QC Manual - 897N101461A 7-43 [8] Collimation assessment Performance Criteria: • Anterior edge of compression plate shall not appear on image, and shall not extend beyond chest wall edge of the image receptor by more than 1% of SID. • Collimation Total / Individual Gap Criteria = SID X 0.02 or less Additonal performance criteria can be found on page 7-5. 0OINTSTOBECHECKED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HECOINSWERECORRECTLYPOSITIONED s4HEDISTANCESWEREMEASUREDATTHECORRECTPOSITION s4HEUNITOFDISTANCEMMWASCORRECTLYUSED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. [9] Radiation output Performance Criteria: Variation coefficient = 0.05 or less, Air kerma rate = 7mGy/sec or more 7 Annual Test 0OINTSTOBECHECKED s4HEDOSIMETERWASCORRECTLYPOSITIONED s4HEDOSIMETERWASCORRECTLYUSED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PANDM!SWERECORRECT s4HEEXPOSUREWASAPPROPRIATE s4HEUNITOFAIRKERMAM2OR«'YWASCORRECTLYUSED s4HEUNITOFAIRKERMARATEM'YSWASCORRECTLYUSED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. 7-44 FDR Mammography QC Manual - 897N101461A 7.5.5 AEC System Performance Test 1. Procedure [Test flow] The AEC system performance test consists of AEC reproducibility, CNR mode 1, AGD mode 1, and AGD-ACR Phantom. The test items are intended to conduct constancy tests on the following items for checking the AEC system performance. Reproducibility of air kerma .................................... AEC reproducibility Change in CNR to target thickness ....................... CNR mode 1 Change in AGD to target thickness ....................... AGD mode 1 AGD (standard breast phantom) ........................ AGD-ACR Phantom TIP In this test, specify the baseline values for CNR mode 1, conduct the two test items not requiring an X-ray image, and then enter their measured values all at once to save time and simplify QC Software operation. 7 Annual Test FDR Mammography QC Manual - 897N101461A 7-45 [Tools] PMMA phantom (40 mm) 4 Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. 5 The following displayed. • CNR [Exposure condition setting procedure] 1 Determine the exposure conditions used in the test. Position the PMMA phantom (40mm) on the exposure table and the compression plate 45mm above the patient support. 2 Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. Select H-mode under “Auto” mode for the X-ray equipment and make an exposure. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N exposure submenu is 2. CNR mode 1 (exposure) [Tools] PMMA phantom (40 mm) and aluminum plate (0.2 mm) for CNR measurement 6 7 Annual Test 7-46 Position the PMMA phantom (40mm) on the exposure table. 1. Patient information entry and exposure menu selection 3 Enter patient information and press [Next]. FDR Mammography QC Manual - 897N101461A 7 Position the 0.2 mm aluminum plate for CNR measurement as shown in the figure below. 3. QC Software startup 10 Press [QC] to display the “QC window”. NOTE Make sure that the aluminum plate covers the Calculation Area 2 without protruding into the left-hand half of the phantom. .IPPLESIDE 0--! #ALCULATION!REAS MMXMM MM 4. Exposure condition confirmation 11 Check the exposure conditions. !LUMINUMPLATE MM #HESTWALLSIDE NOTE Wear gloves when using the aluminum plate to keep it clean. 8 Move the compression plate as close to the PMMA phantom as possible without making contact. !LUMINUMPLATE Specify the exposure conditions closest to those recorded in Step 2 in “Manu” mode. Then make an exposure. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 12 Press [Measurement execution]. The results are displayed at the bottom of the window. 13 When the result is displayed, press [Save] and then [Test data log]. 7 Annual Test 9 5. Auto-calculation NOTE It may take some time until you can press [QC] after completing the exposure process. FDR Mammography QC Manual - 897N101461A 7-47 6. Baseline value setting 14 Select the result calculated in Step 12 , and then press [Baseline value creation]. This completes the baseline value setting. Press [Close]. 2. Position the dosimeter aside the PMMA phantom (40 mm) as shown in the figure below. NOTE NOTE The “baseline values” above means the reference CNR for recognizing the change in image quality depending on the target thickness (target thickness dependence). 7. QC Software shutdown 15 7 16 Press [Save] and then [Close] to exit the QC Software. Select [Study completed] to finish the test. Annual Test 8. AEC reproducibility 17 Follow the instructions below to conduct the test item. [Tools] A dosimeter and PMMA phantom (40 mm) 1. Position the PMMA phantom (40 mm) on the exposure table. NOTE Position the PMMA phantom (40 mm) on the exposure table leaving a space on a side. 7-48 In the AEC reproducibility test, the dosimeter position differs since the air kerma is measured under the “Semi” mode. Position the dosimeter where it does not affect the AEC function but receives the X-ray exposure. See “FDR MS-1000 Operation Manual” for details on the AEC areas. 3. Move the compression plate as close to the PMMA phantom as possible without making contact. 4. Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. 5. Specify the kVp and target/filter to be used (28 kV and Mo/Mo is recommended) for the exposure in “Semi” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 6. Make 5 exposures and measure the air kerma. • Fill in the following items in the worksheet. Entrance air kerma (1): _____mR µGy Entrance air kerma (2): _____mR µGy Entrance air kerma (3): _____mR µGy Entrance air kerma (4): _____mR µGy Entrance air kerma (5): _____mR µGy FDR Mammography QC Manual - 897N101461A 9. AGD mode 1 18 Follow the instructions below to conduct the test item. [Tools] PMMA phantoms (20, 40, 60 and 70 mm), aluminum plates (0.3 and 0.5 mm) for half value layer measurement, a dosimeter, and a lead sheet 1. Position the PMMA phantom (20 mm) on the exposure table. 4. Using PMMA phantoms 60 mm and 70 mm, repeat the operation described in Substeps 1 to 3. For these thicknesses, the compression plate should be positioned at 75mm and 90mm, respectively, above the exposure table. TIP The exposure conditions for the PMMA phantom 40 mm has been determined in Step 2 . Enter the conditions in the worksheet. • Fill in the following items in the worksheet. PMMA phantom 40 mm exposure conditions kVp: _____kV kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 2. Position the compression plate 21mm above the exposure table. 3. Make an exposure in H-mode under “Auto” mode. • Fill in the following items in the worksheet. PMMA phantom 20 mm exposure conditions kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N PMMA phantom 60 mm exposure conditions kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N PMMA phantom 70 mm exposure conditions kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N .IPPLESIDE Annual Test 5. Remove the PMMA phantom, place a lead sheet on the exposure table, and then position the dosimeter as shown in the figure below. 7 %XPOSURETABLE ,ATERALCENTER MM $OSIMETERPOSITION #HESTWALLSIDE %XPOSURETABLE MM #HESTWALLSIDE FDR Mammography QC Manual - 897N101461A MM .IPPLESIDE 7-49 6. Move the compression plate as close to the dosimeter as possible without making contact. NOTE Wear gloves when using the aluminum plate to keep it clean. NOTE NOTE Do not reposition the compression plate during measurement. Make sure that the aluminum plate covers the detector of the dosimeter from the above. 10. Make an exposure with the exposure conditions used in Substep 7 and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma – Al 0.3 mm: _____mR µGy 11. Position the 0.5 mm aluminum plate for half value layer measurement in the same way. 7. Specify “Manu” mode for the X-ray equipment and make an exposure under exposure conditions for the PMMA phantom 20 mm recorded in Substep 3. If the mAs value determined in Substep 3 cannot be applied, specify the closest applicable value. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 12. Make an exposure with the exposure conditions used in Substep 7 and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma – Al 0.5 mm: _____mR µGy 13. Repeat the operation described in Substeps 7 to 12 using the exposure conditions determined in Substep 4 for PMMA phantoms 40 mm, 60 mm and 70 mm to measure and record the air kerma. 14. Select [Study completed] to finish the test. NOTE 7 Annual Test 7-50 The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 8. Make an exposure and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma - w/o Al: _____mR µGy 9. Position the 0.3 mm aluminum plate for half value layer measurement on the compression plate. FDR Mammography QC Manual - 897N101461A 22 10. Patient information entry and exposure menu selection 19 The procedure for testing CNR mode 1 is described below. Make 4 exposures using the following exposure submenus: [H-mode CNR(PMMA20mm)] [H-mode CNR(PMMA40mm)] [H-mode CNR(PMMA60mm)] [H-mode CNR(PMMA70mm)] Enter patient information and press [Next]. [Tools] PMMA phantoms (20, 40, 60, 70 mm) and aluminum plate (0.2 mm) for CNR measurement TIP In CNR mode 1, use the same H-mode of the AEC as that used in AGD mode 1. 11. CNR mode 1 (exposure) 20 Select [QC/TEST] in the exposure region list and [Annual 4/6] in the exposure menu list, and then press [Start study.]. NOTE Wear gloves when using the aluminum plate to keep it clean. 1. Position the PMMA phantom (20 mm) on the exposure table. 2. Position the 0.2 mm aluminum plate for CNR measurement as shown in the figure below. .IPPLESIDE 0--! #ALCULATION!REAS MMXMM 21 The following displayed. submenus HCN20: H-mode CNR(PMMA20mm) HCN40: H-mode CNR(PMMA40mm) HCN60: H-mode CNR(PMMA60mm) HCN70: H-mode CNR(PMMA70mm) FDR Mammography QC Manual - 897N101461A are MM MM 7 #HESTWALLSIDE NOTE Make sure that the aluminum plate covers the Calculation Area 2 without protruding into the left-hand half of the phantom. Annual Test • • • • exposure !LUMINUMPLATE 7-51 3. Move the compression plate as close to the PMMA phantom as possible without making contact. 12. QC Software startup 23 Press [QC] to display the “QC window”. 13. Exposure condition confirmation 24 Check the exposure conditions. 25 Press [User input]. !LUMINUMPLATE 4. Specify the exposure conditions for PMMA phantom 20 mm determined in Substep 3 of the AGD mode 1 test procedure. Then make an exposure in “Manu” mode. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N NOTE The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 7 Annual Test 5. Using PMMA phantoms 40 mm, 60 mm and 70 mm, repeat the operation described in Substeps 1 to 4. Use the exposure conditions for the PMMA phantom of each thickness determined in Substep 4 of the AGD mode 1 test procedure. 14. AEC reproducibility (user input) 26 Select [Annual test 4-1]. measured entrance air kerma. Enter the NOTE It may take some time until you can press [QC] after completing the exposure process. 7-52 FDR Mammography QC Manual - 897N101461A 15. AGD mode 1 (user input) 27 Select [Annual Test 4-2]. Enter the measured entrance air kerma and target/filter for each, and then press [OK]. 17. Test result confirmation 29 Check the displayed judgment results. If there is a [Fail], see “7.5.4 3. Performance Criteria and Corrective Action”. NOTE Some dosimeters may require correction of the measured value according to the beam quality. 18. QC Software shutdown 30 Press [Save] and then [Close] to exit the QC Software. 31 Select [Study completed] to finish the test. NOTE Make sure that the values are entered for all items. If there is an item left unentered, an error occurs. 16. Auto-calculation 28 Press [Measurement execution]. The results are displayed at the bottom of the window. 7 Annual Test FDR Mammography QC Manual - 897N101461A 7-53 ACR phantom, aluminum plates (0.3 and 0.5 mm), a dosimeter and a lead sheet. 6. Make an exposure with the exposure conditions used in Step 29 and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma - w/o Al: _____ 1. Select [QC/TEST] in the exposure region list and [Calculation 2/2] in the exposure menu list, and then press [Start study]. 7. Position the 0.3 mm aluminum plate on the compression plate so that it covers the detector of the dosimeter.. 19. AGD - ACR Phantom [Tools] 32 8. Make an exposure with the exposure conditions used in Step 29 and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma - Al 0.3mm: _____ 2. Select H-mode under “Auto” mode for the X-ray equipment and make an exposure of the ACR phantom. • Fill in the following items in the worksheet. kVp: _____kV mAs: _____mAs Target/Filter: _____/_____ Compressed breast thickness: _____mm Compression force: _____N 9. Position the 0.5 mm aluminum plate on the compression plate so that it covers the detector of the dosimeter. 10. Make an exposure with the exposure conditions used in Step 29 and measure the air kerma. • Fill in the following item in the worksheet. Entrance air kerma - Al 0.5mm: _____ 3. Remove the ACR phantom and cover the exposure table with the lead sheet. 4. Position the dosimeter at the lateral center of the exposure table, 60 mm away from the chest wall-side edge, and also 40 mm above the table. 33 Enter patient information and press [Next]. 34 Select [QC/TEST] in the exposure region list and [Calculation 2/2] in the exposure menu list, and then press [Start study.]. .IPPLESIDE %XPOSURETABLE ,ATERALCENTER MM $OSIMETERPOSITION 7 #HESTWALLSIDE %XPOSURETABLE MM Annual Test #HEST WALL SIDE MM .IPPLE SIDE 5. Move the compression plate as close to the dosimeter as possible without making contact. NOTE: Do not reposition the compression plate during measurement. 7-54 FDR Mammography QC Manual - 897N101461A 35 Make an exposure under any conditions (e.g. 23 kV, 2 mAs and Mo/Mo) to start up the QC Software. 39 Press [Measurement execution]. The results are displayed at the bottom of the window. 40 Confirm that the calculation result does not exceed 3mGy. If it does, see “7.5.4.3. Performance Criteria and Corrective Action”. 41 Press [Save] and then [Close] to exit the QC Software. 42 Select [Study completed] to finish the test. 43 To complete MEE / Annual tesing, please perform Sections 4-5 and 4-6 in Chapter 4 of this QC Manual. NOTE It may take some time until you can press [QC] after completing the exposure process. 36 Press [QC] to display the “QC window”. 37 Press [User input]. 38 Enter the target/filter used and measured entrance air kerma, and then press [OK]. NOTE Some dosimeters may require correction of the measured value according to the beam quality. 7 Annual Test FDR Mammography QC Manual - 897N101461A 7-55 2. Test Result Evaluation and Judgment Evaluate and judge the results of the AEC system performance test in the Annual Test. If all items are judged as [Pass], the AEC system performance test is finished. If there is an item judged as [Fail], take corrective actions by following “7.5.5 3. Performance Criteria and Corrective Action”. See “8.2 Quick Guide for Software Operation” for test log display procedure. 3. Performance Criteria and Corrective Action [10] AEC reproducibility Performance Criteria: Variation coefficient = 0.05 or less 0OINTSTOBECHECKED s4HEDOSIMETERWASCORRECTLYPOSITIONED s4HEDOSIMETERWASCORRECTLYUSED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEUNITOFAIRKERMAM2OR«'YWASCORRECTLYUSED )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 2EDOTHETEST 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. 7 Annual Test 7-56 FDR Mammography QC Manual - 897N101461A [11] CNR mode 1 Performance Criteria: CNR Relative Value 20mm [%] = 105% of baseline value or more CNR Relative Value 40mm [%] = 95% of baseline value or more CNR Relative Value 60mm [%] = 63% of baseline value or more CNR Relative Value 70mm [%] = 60% of baseline value or more 0OINTSTOBECHECKED s4HECORRECTEXPOSURESUBMENUSWERESELECTED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEEXPOSURESWEREMADEINTHECORRECTORDER s4HEALUMINUMPLATEUSEDHASTHESPECIFIEDPURITY s4HEALUMINUMPLATETHICKNESSANDORPOSITIONWASCORRECTLYARRANGED s4HEREARENOSCRATCHESORDIRTONTHEALUMINUMPLATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM 4HE#.2INTHE$AILY7EEKLY 4ESTHASBEENJUDGEDAS;0ASS=INEVERYTEST 2EDOTHETEST 2UN;#ALCULATION=ANDTHENCALCULATETHERATIOOF THE#.2OFEACH0--!PHANTOMTHICKNESSTOTHE BASELINEVALUEBYUSINGACALCULATORANDTHENJUDGE THERESULT 0ASS 4HETESTISFINISHED &AIL 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken within 30 days of the test date. 7 Annual Test FDR Mammography QC Manual - 897N101461A 7-57 [12] AGD mode 1 Performance Criteria: AGD 20mm [mGy] = 1 mGy or less AGD 40mm [mGy] = 2 mGy or less AGD 60mm [mGy] = 4.5 mGy or less AGD 70mm [mGy] = 6.5 mGy or less [13] AGD - ACR Phantom Performance Criteria: AGD-ACR Phantom [mGy] = 3 mGy or less 0OINTSTOBECHECKED s4HEDOSIMETERWASCORRECTLYPOSITIONED s4HEDOSIMETERWASCORRECTLYUSED s4HE8RAYEQUIPMENTSETTINGSTARGETFILTERK6PWERECORRECT s4HEUNITOFAIRKERMAM2OR«'YWASCORRECTLYUSED s4HEALUMINUMPLATEUSEDHASTHESPECIFIEDPURITY s4HEALUMINUMPLATETHICKNESSANDORPOSITIONWASCORRECTLYARRANGED s4HEREARENOSCRATCHESORDIRTONTHEALUMINUMPLATE )FANYOFTHEABOVEISNOTCORRECTAPPROPRIATECORRECTTHEPROBLEM #HECKTHERESULTOF!%#REPRODUCIBILITYTEST 0ASS 2EDOTHETEST 4HE8RAYEQUIPMENTMAYBEDEFECTIVE 0ASS 4HETESTISFINISHED &AIL 7 Annual Test 7-58 &AIL #ONTACTA&5*)&),-DEALER 4HE8RAYEQUIPMENTMAYBEDEFECTIVE #ONTACTA&5*)&),-DEALER If the item still results in [Fail], the source of the problem shall be identified and corrective action shall be taken before any further examinations are performed with the system. FDR Mammography QC Manual - 897N101461A Chapter 8 Assistance Material Page 8.1 Calculation Functions __________________________________ 8-2 8.2 Quick Guide for Software Operation _____________________ 8-6 8.3 Worksheets____________________________________________ 8-8 8.4 Report Forms __________________________________________ 8-20 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-1 8.1 Calculation Functions The CNR and AGD of PMMA phantom at a given thickness can be auto-calculated by the QC Software. Items Auto-calculated and Judged Based on Input Data Test Items Exposure Menus Exposure Submenus 1. Exposure CNR Calculation 1/2 CNR AGD Calculation 2/2 AGD 2. Input Values ✓ ✓ The measurement procedure of each item is described below. [1] CNR 3 The following displayed. • CNR 4 Position the PMMA phantom of the desired thickness to be measured and 0.2 mm aluminum plate as shown in the figure below. [Tools] exposure submenu is PMMA phantom (of a given thickness) and aluminum plate (0.2 mm) 1 Enter patient information and press [Next]. NOTE Make sure that the aluminum plate covers the Calculation Area 2 without protruding into the left-hand half of the phantom. 2 Select [QC/TEST] in the exposure region list and [Calculation 1/2] in the exposure menu list, and then press [Start study.]. .IPPLESIDE 0--! #ALCULATION!REAS MMXMM MM 8 !LUMINUMPLATE MM #HEST WALL SIDE NOTE Assistance Material Wear gloves when using the aluminum plate to keep the plate clean. 8-2 FDR Mammography QC Manual - 897N101461A 5 Move the compression plate as close to the PMMA phantom as possible without making contact. 10 Check the calculation result. 11 Press [Save] and then [Close] to exit the QC Software. 12 Select [Study completed] to finish the test. !LUMINUMPLATE 6 Specify the exposure conditions for the X-ray equipment and make an exposure in “Manu” mode. NOTE The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 7 Repeat Steps 4 to 6 as many times as necessary to capture the desired number of images to be calculated (max. 32 images). NOTE It may take some time until you can press [QC] after completing the exposure process. 8 Press [QC] to display the “QC window”. 9 Press [Measurement execution]. The results are displayed at the bottom of the window. 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-3 [2] AGD 5 Position the 0.3 mm aluminum plate on the compression plate. [Tools] NOTE A dosimeter and aluminum plates (0.3 and 0.5 mm) Wear gloves when using the aluminum plate to keep the plate clean. 1 Position the dosimeter at the lateral center of the exposure table, 60 mm away from the chest wall-side edge, and also 40 mm above the table. .IPPLESIDE NOTE Make sure that the aluminum plate covers the detector of the dosimeter from the above. 6 Make an exposure with the exposure conditions used in Step 3 and measure the air kerma. 7 Position the 0.5 mm aluminum plate in the same way. 8 Make an exposure with the exposure conditions used in Step 3 and measure the air kerma. 9 Enter patient information and press [Next]. 10 Select [QC/TEST] in the exposure region list and [Calculation 2/2] in the exposure menu list, and then press [Start study.]. %XPOSURETABLE ,ATERALCENTER MM $OSIMETERPOSITION #HESTWALLSIDE %XPOSURETABLE MM #HEST WALL SIDE 2 MM .IPPLE SIDE Move the compression plate as close to the dosimeter as possible without making contact. NOTE Do not reposition the compression plate during measurement. 8 3 Assistance Material Specify the desired exposure conditions for the X-ray equipment and make an exposure in “Manu” mode. NOTE The “Semi” and “Auto” modes may not work correctly for this measurement. Make sure to use “Manu” mode for exposure. 4 8-4 Record the air kerma. FDR Mammography QC Manual - 897N101461A 11 12 The following displayed. • AGD exposure submenu is 16 Press [Measurement execution]. The results are displayed at the bottom of the window. 17 Check the calculation result. 18 Press [Save] and then [Close] to exit the QC Software. 19 Select [Study completed] to finish the test. Make an exposure under any conditions (e.g. 23 kV, 2 mAs and Mo/Mo) to start up the QC Software. NOTE It may take some time until you can press [QC] after completing the exposure process. 13 Press [QC] to display the “QC window”. 14 Press [User input]. 15 Enter the target/filter used and measured entrance air kerma, and then press [OK]. NOTE Some dosimeters may require correction of the measured value according to the beam quality. 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-5 8.2 Quick Guide for Software Operation This section provides concise references for the QC Software main operations and the QC test log confirmation procedure. Make a copy of this section as necessary. Basic Procedure 1 Starting FUJIFILM workstation Start a FUJIFILM workstation and enter patient information, and then press [Next]. !UTOCALCULATESTHERESULTBASEDON ENTEREDIMAGEANDVALUESAND DISPLAYS;0ASS&AIL=JUDGMENT 2 3 Exposure Make an exposure using all displayed exposure submenus. 4 Starting QC Software Press [QC]. 5 Main button operations in “QC window” 8 Assistance Material 8-6 Starting QC test Select [QC/TEST] in the exposure region list and an exposure submenu displayed in the exposure menu list, and then press [Start study.]. of 3AVESTHESTUDY RESULTS 1UITSTHE 1#3OFTWARE /UTPUTSANIMAGE FORVISUAL INSPECTION 6 $ISPLAYSTHEUSERINPUT SCREENFORTHEMEASUREMENT ANDVISUALINSPECTIONRESULTS Finishing the test Press [Study completed] to finish the test. When finishing the test, the window in Step 1 is redisplayed. To conduct another test, repeat the procedure. the FDR Mammography QC Manual - 897N101461A Log Confirmation Procedure 1 Starting FUJIFILM workstation Start a FUJIFILM workstation and enter patient information, and then press [Next]. 4 Main button operations in “Test Data Log window” 3WITCHESTHEGRAPHDISPLAYSCALE 3WITCHESTHENUMBER OFLOGSDISPLAYEDIN THEGRAPH 3WITCHESINFORMATION DISPLAYEDINTHETEST ITEMFIELDBETWEEN MEASUREDVALUESAND JUDGMENTRESULTS $ISPLAYSEXPOSURE CONDITIONS 2 #REATESBASELINE VALUES Starting QC test Select [QC/TEST] in the exposure region list and press [Start study.] without selecting an exposure submenu. /UTPUTSTHETESTRESULTINTOA#36FILE 5 3 Finishing the test Press [Close] and then [Study completed] to finish the test. Starting QC Software Press [QC]. 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-7 8.3 Worksheets The forms on the following pages are prepared for recording the exposure conditions used in the tests and measured values. It is recommended to keep the record of each test conditions for reproducibility. Make a copy of these worksheets as necessary. 8 Assistance Material 8-8 FDR Mammography QC Manual - 897N101461A Baseline Value Settings <Baseline Values for Annual Test> Test Items Steps 5 [5] Spatial Resolution (Magnification) 6 Exposure Conditions/Measurement and Judgment Results 1Shot Phantom position Exposure conditions for Exposure Submenu [1ShotM(Magnification)] Purposes Lateral shift distance: mm kVp: kV mAs: mAs Target/Filter: / Compressed breast thickness: mm Compression force: Phantom positioning in Annual Test Exposure condition confirmation ( 8 ) N <Baseline Values for Semi-annual Test> Test Items [1] Image Basic Test Steps Exposure Conditions/Measurement and Judgment Results 1 mAs 2 Air kerma 7 Exposure conditions for Exposure Submenu [Uniform Expo(Image Basic)] Input Steps mAs mR µGy mR µGy mR µGy kVp: kV mAs: mAs Target/Filter: / Exposure condition determination ( 7 ) Result input ( 10 ) Exposure condition confirmation ( 9 ) White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [1] 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-9 <Baseline Values for Weekly Test> Test Items Steps Exposure Conditions/Measurement and Judgment Results Purposes kVp: kV mAs: [1] Test with 1Shot Phantom 6 1Shot Phantom exposure conditions Target/Filter: mAs / Compressed breast thickness: mm Compression force: N kVp: 8 [2] Test with ACR Phantom Exposure conditions for Exposure Submenu [ACR Phantom] kV mAs: Target/Filter: Compressed breast thickness: Compression force: Exposure condition confirmation ( 8 ) mAs / mm Exposure condition confirmation ( 10 ) N Density inside the disk 12 Result input ( Density outside the disk 14 ) White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [2] [1] 8 Assistance Material 8-10 FDR Mammography QC Manual - 897N101461A <Weekly Test> Test Items Steps Exposure Conditions/Measurement and Judgment Results Purposes kVp: [1] Test with 1Shot Phantom 4 -3 Exposure conditions for Exposure Submenu [1Shot PhantomM] kV mAs: mAs Target/Filter: / Compressed breast thickness: mm Compression force: [2] Good practice [1] Test with 1Shot Phantom N 5 -1 X-ray equipment cleanliness Pass Fail 5 -2 Screen cleanliness Pass Fail 5 -3 Viewing box cleanliness Pass Fail Distortion Pass Fail Artifact Pass Fail 9 4 -5 Exposure conditions for Exposure Submenu [ACR Phantom] kVp: kV mAs: mAs Target/Filter: Compressed breast thickness: Compression force: Exposure condition confirmation ( 7 ) Result input ( 11 ) Result input ( 10 ) Exposure condition confirmation ( 7 ) / mm N Density at center of Phantom image Density inside the disk [3] Test with ACR Phantom Density outside the disk 8 Fibers (ACR Phantom) points Specks (ACR Phantom) points Masses (ACR Phantom) points Visible step wedge (Step Phantom) Result input ( 10 ) steps Specks (Step Phantom) steps Masses (Step Phantom) steps White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [3] [1] [2] 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-11 Semi-annual Test Test Items [1] Compression device confirmation Steps 1 -4 Exposure Conditions/Measurement and Judgment Results Automatic Compression Date: Measured value: lbs. Acceptance Limits Automatic (25 45 lbs) PASS FAIL Corrective Action needed ? YES NO YES NO PASS FAIL Corrective Action description Corrective Action taken? Corrective Date: Results of new test 1 -7 Manual Compression Date: Measured lbs. value: Comments _______________________________________________________________________________________________________________ _________________________________________________________________________________________________________________________ _________________________________________________________________________________________________________________________ White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. 8 Assistance Material 8-12 FDR Mammography QC Manual - 897N101461A Annual Test (Initial Performance Test) Test Items Steps 1 -3 Exposure Conditions/Measurement and Judgment Results Exposure conditions for Exposure Submenu [Uniform Expo(Image Basic)] kV mAs: mAs Target/Filter: 5 Exposure conditions for Exposure Submenu [Uniform Expo(Image Basic)] / mR µGy Air kerma [1] Image Basic Test Purposes kVp: mR µGy mR µGy kVp: kV mAs: mAs Target/Filter: / Exposure condition determination ( 9 ) Result input ( 13 ) Exposure condition confirmation ( 9 ) Artifact Pass Fail Result input ( 13 ) White out Pass Fail Result input ( 13 ) Result input ( 14 ) Result input ( 15 ) 11 6 -2, 3 [2] Compression device confirmation [3] Viewibox maintenance Compressed breast thickness Display value: mm Measured value: mm Display value: mm 6 -5 Compression force 6 -6 Compression force sustainability 6 -7 Maximum compression force 7 -2 Luminance 7 -3 Viewing box uniformity Pass Fail 7 -4 Luminance adjustment function Pass Fail 7 -5 Masking Pass Fail 7 -6 Interpretation room illuminance (Ambient light) Measured value: Pass mm Fail N cd/m2 lx White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [1] [3] [2] 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-13 Annual Test (Image Performance Test) Test Items Steps Exposure Conditions/Measurement and Judgment Results Purposes kVp: [1] Additive lag effects (Lag) 4 -3 Exposure conditions for Exposure Submenu [ACR Phantom1(Lag)] kV mAs: mAs Target/Filter: 5 -2 Exposure conditions for Exposure Submenu [Uniform Expo.1(Ghost)] / kVp: kV mAs: mAs Target/Filter: / kVp: [2] Multiplicative lag effects (Ghost) 5 -4 Exposure conditions for Exposure Submenu [ACR Phantom2(Ghost)] kV mAs: mAs Target/Filter: 5 -6 Exposure conditions for Exposure Submenu [Uniform Expo.2(Ghost)] / kVp: kV mAs: mAs Target/Filter: [3] Missed tissue on chest wall edge (40 mm height) 6 -3 / kVp: kV mAs: mAs Target/Filter: Compressed breast thickness: 7 mm N Missed tissue on chest wall edge (Right) Pass Fail Missed tissue on chest wall edge (Left) Pass Fail Verify mechanical stability Pass Fail Verify that all moving parts move smoothly Pass Fail All locks and detents work properly Pass Fail No sharp, rough edges or other hazards including electrical hazards Pass Fail Operator technique control charts are posted Pass Fail Pass Fail All indicator lights working properly Pass Fail Auto decompression can be overridden and status displayed Pass Fail Compression can be manually released / in any failure Pass Fail Operator protected during adequate radiation shielding exposure by Exposure condition confirmation ( 9 ) / Compression force: 11 [4] Visual and Functional test Exposure conditions for Exposure Submenu [1Shot PhantomM(40mm)] Exposure condition confirmation ( 9 ) Result input ( 13 ) Result input ( 14 ) White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [3] [4] 8 Assistance Material 8-14 FDR Mammography QC Manual - 897N101461A Annual Test (Spatial Resolution Test) Test Items Steps 4 -2 [5] Spatial Resolution (Magnification) 4 -3 Exposure Conditions/Measurement and Judgment Results 1Shot Phantom position Exposure conditions for Exposure Submenu [1ShotM(Magnification)] Purposes Lateral shift distance: mm kVp: kV mAs: mAs Target/Filter: Exposure condition confirmation ( 6 ) / Compressed breast thickness: mm Compression force: N White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-15 Annual Test (X-ray Equipment Performance Test) Test Items Steps Exposure Conditions/Measurement and Judgment Results kVp used for the tests [6] kVp accuracy and reproducibility 1 -5, 6 1 -8 LFS - volt. SFS - volt. Low kV Mid. kV High kV Low kV Mid.-1 kV Mid.-2 kV Mid.-3 kV Mid.-4 kV Mid.-5 kV High kV Low kV Mid. kV High kV kVp [7] Half Value Layer (HVL) 2 -3, 4, 5 mAs Al 0.3 mm ) kV mAs mR µGy mR µGy mR µGy Target/Filter ( / ) kV mAs mR µGy mR µGy mR µGy Target/Filter ( / ) kV mAs mR µGy mR µGy mR µGy kVp: kV mAs: mAs Target/Filter: 3 -4 Entrance air kerma mR µGy (4) mR µGy mR µGy (5) mR µGy mR µGy (3) Exposure conditions mAs Target/Filter: / Compressed breast thickness: mm Compression force: 16 -3 8 Assistance Material 8-16 Distance between image edges X-ray / Light field gap X-ray / Light field gap mm A-b mm A-c mm A-d mm A-e mm A-f mm B-b mm B-c mm B-d mm B-e mm B-f mm Left: mm Right: mm Chest wall side: mm mm Left: mm Right: mm Nipple side: mm 16 -5 X-ray field / Exposure table gap Chest wall side: mm 16 -6 Compression paddle edge (Chest wall side) Pass X-ray / Image receptor field gap Exposure condition confirmation ( 9 ) N A-a Nipple side: 16 -4 21 ) kV mAs: 16 -2 Result input ( s kVp: [8] Collimation assessment 19 ) mGy/s (Exposure time) 16 -1 Result input ( / (2) Air kerma rate 7 -3 18 ) Al 0.5 mm / (1) Result input ( Air kerma w/o Al Target/Filter ( Exposure conditions [9] Radiation output Purposes Result input ( 20 ) Fail White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [6] [7] [8] [9] FDR Mammography QC Manual - 897N101461A Annual Test (AEC System Performance Test) Test Items Steps Exposure Conditions/Measurement and Judgment Results Purposes kVp: kV mAs: 2 Exposure conditions mAs Target/Filter: Compressed breast thickness: mm Compression force: [11] CNR mode 1 N kVp: 9 Exposure conditions kV mAs: mAs Target/Filter: / Compressed breast thickness: mm Compression force: kV mAs: Exposure conditions mAs Target/Filter: / Compressed breast thickness: [10] AEC reproducibility mm Compression force: 17 -6 N (1) mR µGy (2) mR µGy (4) mR µGy (5) mR µGy Entrance air kerma Exposure condition confirmation ( 11 ) N kVp: 17 -5 Exposure condition determination ( 9 ) / mR µGy (3) Result input ( 26 ) White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [10] Test Items Steps Exposure Conditions/Measurement and Judgment Results PMMA thickness [11] CNR mode 1 22 -4 Exposure Mode 1 kVp mAs Target/ Filter Purposes Compressed breast thickness Compression force 20 mm kV mAs / mm N 40 mm kV mAs / mm N 60 mm kV mAs / mm N mAs / mm N 70 mm kV Exposure condition confirmation ( 29 ) FDR Mammography QC Manual - 897N101461A 8 Assistance Material White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) 8-17 Test Items Steps Exposure Conditions/Measurement and Judgment Results PMMA thickness 18 -3, 4 Exposure 20 mm Mode 1: Exposure 40 mm condition determination 60 mm 70 mm [12] AGD Mode 1 PMMA thickness 18 -7, 8, 10, 12 Exposure Mode 1 20 mm 40 mm kVp mAs Target/ Filter Compressed breast thickness Compression force kV mAs / mm N kV mAs / mm N kV mAs / mm N kV mAs / mm N Compressed breast thickness Compression force mm N kVp kV kV mAs mAs mAs Target/ Filter / / mm Purposes Exposure condition ( for 18 -7, 13) determination Air kerma w/o Al Al 0.3 mm Al 0.5 mm mR µGy mR µGy mR µGy N mR µGy mR µGy mR µGy mR µGy mR µGy mR µGy mR µGy 60 mm kV mAs / mm N mR µGy 70 mm kV mAs / mm N mR µGy Result input ( 27 ) White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) [12] [12] 8 Assistance Material 8-18 FDR Mammography QC Manual - 897N101461A Test Items Steps 32 Exposure Conditions/Measurement and Judgment Results ACR Phantom: Exposure condition determination kV kVp 32 -6, 8, 10, kVp mAs mAs mAs Exposure ACR Phantom Target/ Filter / Target/ Filter Compressed breast thickness mAs Exposure condition ( for mm N Compressed breast thickness Compression force mm 32 -6, 8, 10) determination Air kerma w/o Al mR µGy / kV Purposes Compression force Al 0.3 mm Al 0.5 mm mR µGy mR µGy Result input ( 38 ) N White numbers on black squares indicate the relevant Steps in the test procedures described in Chapters 3 to 8. Dark shaded columns indicate that the items in the columns require user input in the QC Software. (Exposure conditions are acquired automatically.) 8 Assistance Material FDR Mammography QC Manual - 897N101461A 8-19 8.4 Report Forms The forms on the following pages are provided for recording the test results. It is important to record test results to follow the changes in performance of the X-ray equipment and/or other equipment. Make a copy of these report forms as necessary. The following report forms are provided on the following pages. • Daily/Weekly Test Report • Quarterly Test Report • Semi-annual Test Report • Annual Test Report 8 Assistance Material 8-20 FDR Mammography QC Manual - 897N101461A 1/1/ Facility Information Client Date Exposure Room Operator Time Department Manufacturer Name Model S/N Installation Date Manufacturer Model S/N Installation Date Manufacturer Model S/N Calibration Expiration Date X-ray Equipment FUJIFILM Workstation FUJIFILM Measurement Equipment/Tool Information Laser Imager Diagnostic monitor Viewing box 1Shot Phantom M FUJIFILM ACR Phantom Step Phantom Densitometer Density control strip Dosimeter Force scale Luminance meter Illuminance meter kVp meter Signature 1/3 Weekly Test Report Test with 1Shot Phantom Exposure Conditions kVp 1Shot PhantomM mAs Target/Filter Compressed Breast Thickness kV Compression Force mm S Value N Test Result Criteria Judgment Item Judgment Result Lower Limit Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy Geometric distortion System artifact evaluation Upper Limit Missed tissue on chest wall edge (Right) mm - PASS FAIL Missed tissue on chest wall edge (Left) mm - PASS FAIL CNR PASS FAIL System sensitivity PASS FAIL Horizontal direction mm PASS FAIL Vertical direction mm PASS FAIL Geometric distortion - - PASS FAIL Artifacts - - PASS FAIL Pixel Value (PV) ratio (Top-Right) % PASS FAIL Pixel Value (PV) ratio (Top-Left) % PASS FAIL Pixel Value (PV) ratio (Bottom-Right) % PASS FAIL Pixel Value (PV) ratio (Bottom-Left) % PASS FAIL SNR ratio (Top-Right) % PASS FAIL SNR ratio (Top-Left) % PASS FAIL SNR ratio (Bottom-Right) % PASS FAIL SNR ratio (Bottom-Left) % PASS FAIL Average QL at thinnest step wedge QL PASS FAIL 2 lp/mm % PASS FAIL 4 lp/mm % PASS FAIL Low Contrast Detectability (Light) % PASS FAIL Low Contrast Detectability (Dark) % PASS FAIL Uniformity Dynamic range Spatial Resolution (SR) Low Contrast Detectability (LCD) 2/3 Criteria Judgment Item Judgment Result Lower Limit Linearity/Beam quality constancy Upper Limit QL gap (1 step-2 step) QL PASS FAIL QL gap (2 step-3 step) QL PASS FAIL QL gap (3 step-4 step) QL PASS FAIL QL gap (4 step-5 step) QL PASS FAIL Remarks Signature 3/3/ Good practice Criteria Judgment Item Judgment Result Lower Limit Upper Limit X-ray equipment cleanliness - - Screen cleanliness - - Viewing box cleanliness - - PASS Remarks Signature FAIL 1/3/ Weekly Test Report Test with ACR Phantom Exposure Conditions kVp ACR Phantom mAs Target/Filter Compressed Breast Thickness kV Compression Force mm S Value N Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit Density at center of Phantom image PASS FAIL Density difference (disk/outside) PASS FAIL Fibers (ACR Phantom) - PASS FAIL Specks (ACR Phantom) - PASS FAIL Masses (ACR Phantom) - PASS FAIL Visible step wedge (Step Phantom) - PASS FAIL Specks (Step Phantom) - PASS FAIL Masses (Step Phantom) - PASS FAIL 2/3/ HARD COPY PHANTOM CONTROL/CNR CHART Room: Year: Month: Date: Initials: AEC SETTING: CNR Value: CNR OK: Density Difference +0.05 +0.05 Baseline -0.05 -0.05 Background Density +0.20 +0.20 +0.15 +0.15 Baseline -0.15 -0.15 -0.20 -0.20 No.Visible Fibers +0.5 Specks Masses +0.5 - 0.5 - 0.5 +0.5 +0.5 - 0.5 - 0.5 +0.5 +0.5 - 0.5 - 0.5 mAs (optional) +15% +15% Baseline -15% -15% 3/3/ SOFT COPY PHANTOM CONTROL/CNR CHART Room: Year: Month: Date: Initials: AEC SETTING: CNR Value: CNR OK: No.Visible Fibers +0.5 Specks Masses +0.5 - 0.5 - 0.5 +0.5 +0.5 - 0.5 - 0.5 +0.5 +0.5 - 0.5 - 0.5 S Value +20% +20% Baseline -20% -20% mAs (optional) +15% +15% Baseline -15% -15% Remarks Signature 1/1/ Quarterly Test Report Repeat analysis 4/4!,.5-"%2/&%8!-3 4/4!,.5-"%2/&&),-3 /4(%2 4/4!, 02/*%#4)/. 2%!3/.&/22%*%#4 ## -,/ -,OR,- !8),,!29 /4(%2 4/4!, 2%0%!43 0/3)4)/.).' 0!4)%.4-/4)/. #/-02%33)/. !24)&!#43 82!9%15)0 -!,&5.#4)/. 3/&47!2% -!,&5.#4)/. !%# -)30,!#%-%.4 5.$%2 %80/352% /6%2 %80/352% ).#/22%#4 0!4)%.4)$ 7!34% 35"4/4!, '2!.$4/4!, 2%0%!43 2%0%!42!4% 4/4!,&),-3).#,5$).'2%0%!43 2%0%!42!4% !,,2%*%#4&),-3 2%*%#42!4% 4/4!,&),-3).#,5$).'2%0%!43 2%*%#42!4% #/--%.43&/2#/22%#4)6%!#4)/.!.$'/!,3 3IGNATURE OF 2%0%!43 1/1/ Semi-annual Test Report Compression device confirmation Test Items [1] Compression device confirmation Exposure Conditions/Measurement and Judgment Results Automatic Compression Date: Measured value: lbs. Acceptance Limits Automatic (25 45 lbs) PASS FAIL Corrective Action needed ? YES NO YES NO PASS FAIL Corrective Action description Corrective Action taken? Corrective Date: Results of new test Manual Compression Date: Measured lbs. value: Remarks Signature 1/15 Annual Test Report Image basic test Exposure Conditions kVp mAs Compressed Breast Thickness Target/Filter Compression Force S Value S Value kV mm N Visual Inspection kV mm N Measured Values Entrance air kerma 1 mR (μC/kg) Entrance air kerma 2 mR (μC/kg) Entrance air kerma 3 mR (μC/kg) Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit Relative sensitivity (S value) PASS FAIL Criteria Judgment Item Judgment Result Lower Limit Upper Limit Artifact - - PASS FAIL White out - - PASS FAIL Remarks Signature 2/15 Compression device confirmation Measured Values Compressed breast thickness - Display value mm Compressed breast thickness - Measured value mm Compression force - Display value N Compression force Measured value N Maximum compression force N Test Result Criteria Judgment Item Judgment Result Lower Limit Compressed breast thickness accuracy Upper Limit mm PASS FAIL Compression force accuracy N PASS FAIL Maximum compression force N PASS FAIL PASS FAIL Compression force sustainability - - Remarks Signature 3/15 Viewing box maintenance Test Result Criteria Judgment Item Judgment Result Lower Limit Luminance cd/m2 Upper Limit - PASS FAIL Viewing box uniformity - - PASS FAIL Luminance adjustment function - - PASS FAIL Masking - - PASS FAIL PASS FAIL Interpretation room illuminance (Ambient light) lx - Remarks Signature 2/15 Annual Test Report MONITOR QUALITY CONTROL (ANNUAL) REPORT FORM Room: Workstation: Left Monitor Serial Number: Right Monitor Serial Number: Final interpretation? YES NO Cleared by FDA? YES NO N/A MONITOR IMAGE QUALITY REPORT Monitor has DICOM Display Functionality feature applied Left or Single Monitor Right Monitor YES / NO YES / NO % % Lmax target: Lmax measured: (Lmax measured - Lmax target)/Lmax target x 100% Lmax Within 10% ? YES / PASS NO / FAIL YES / PASS NO / FAIL 5% patch is visible in 0% patch? YES / PASS NO / FAIL YES / PASS NO / FAIL 95% patch is visible in 100% patch? YES / PASS NO / FAIL YES / PASS NO / FAIL Gray steps distinguishable? YES / PASS NO / FAIL YES / PASS NO / FAIL Alphanumeric characters sharp? YES / PASS NO / FAIL YES / PASS NO / FAIL Line-pair pattern (center) distinguishable? YES / PASS NO / FAIL YES / PASS NO / FAIL Line-pair patterns (corner) distinguishable? YES / PASS NO / FAIL YES / PASS NO / FAIL Overall, all tests pass ? YES / PASS NO / FAIL YES / PASS NO / FAIL MONITOR RESPONSE REPORT Room: Left or Single Monitor Right Monitor Measured contrast response at any given point does not deviate more than a) 10% of the standard for final interpretation or, b) 20% of the standard for other use. YES / PASS NO / FAIL Signature 4/15 Annual Test Report Additive lag effects (Lag) Exposure Conditions kVp mAs Target/Filter Compressed Breast Thickness Compression Force S Value ACR Phantom1(Lag) kV mm N Non Exposure(Lag) kV mm N Test Result Criteria Judgment Item Judgment Result Lower Limit Lag Factor QL Upper Limit - PASS Remarks Signature FAIL 5/15 Multiplicative lag effects (Ghost) Exposure Conditions kVp mAs Target/Filter Compressed Breast Thickness Compression Force S Value Uniform Expo.1(Ghost) kV mm N ACR Phantom2(Ghost) kV mm N Uniform Expo.2(Ghost) kV mm N Test Result Criteria Judgment Item Judgment Result Lower Limit Ghost Factor Upper Limit - PASS Remarks Signature FAIL 6/15 Missed tissue on chest wall edge (40 mm height) Exposure Conditions kVp 1Shot Phantom M(40 mm) mAs Compressed Breast Thickness Target/Filter kV Compression Force mm S Value N Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit Missed tissue on chest wall edge (Right) - - PASS FAIL Missed tissue on chest wall edge (Left) - - PASS FAIL Remarks Signature 7/15 Visual and Functional test Test Result Judgment Item Judgment Result Verify mechanical stability PASS FAIL Verify that all moving parts move smoothly PASS FAIL All locks and detents work properly PASS FAIL No sharp, rough edges or other hazards including electrical hazards PASS FAIL Operator technique control charts are posted PASS FAIL Operator protected during exposure by adequate radiation shielding PASS FAIL All indicator lights working properly PASS FAIL Auto decompression can be overridden and status displayed PASS FAIL Compression can be manually released / in any failure PASS FAIL Remarks Signature 8/15 Spatial Resolution (Magnification) Exposure Conditions kVp 1ShotM (Magnification) mAs Target/Filter Compressed Breast Thickness kV Compression Force mm S Value N Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit MTF (4lp/mm) % PASS FAIL MTF (8lp/mm) % PASS FAIL Remarks Signature 9/15 kVp accuracy and reproducibility Measurement Conditions kVp Display Values Low kV Mid. kV High kV Low kV Mid.-1 kV Mid.-2 kV Mid.-3 kV Mid.-4 kV Mid.-5 kV High kV Low kV Mid. kV High kV Measured Values Large Focal Spot (LFS) Small Focal Spot (SFS) Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit Low volt. accuracy (LFS) kV PASS FAIL Mid. volt. accuracy (LFS) kV PASS FAIL High volt. accuracy (LFS) kV PASS FAIL Low volt. accuracy (SFS) kV PASS FAIL Mid. volt. accuracy (SFS) kV PASS FAIL High volt. accuracy (SFS) kV PASS FAIL Mid. volt. reproducibility (LFS) kV PASS FAIL Remarks Signature 10/15 Half Value Layer (HVL) Measured Values w/o Al Al 0.3 mm Al 0.5 mm Measured HVL (Mo/Mo) mR(μC/kg) mR(μC/kg) mR(μC/kg) Measured HVL (Mo/Rh) mR(μC/kg) mR(μC/kg) mR(μC/kg) Measured HVL (W/Rh) mR(μC/kg) mR(μC/kg) mR(μC/kg) Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit HVL (Mo/Mo) mmAl - PASS FAIL HVL (Mo/Rh) mmAl - PASS FAIL HVL (W/Rh) mmAl - PASS FAIL Remarks Signature 11/15 Collimation assessment Exposure Conditions kVp Uniform Expo.3 (Collimation) mAs Target/Filter Compressed Breast Thickness kV Compression Force mm S Value N Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit X-ray / Light field gap (Right/Left) mm - PASS FAIL X-ray / Light field gap (Chest/Nipple) mm - PASS FAIL X-ray / Image receptor field gap (Left) mm - PASS FAIL X-ray / Image receptor field gap (Right) mm - PASS FAIL X-ray / Image receptor field gap (Nipple) mm - PASS FAIL X-ray field / Exposure table gap mm - PASS FAIL PASS FAIL Compression paddle edge (Chest wall) - - Remarks Signature 12/15 Radiation output Measured Values mAs Entrance air kerma 1 mR(μC/kg) Entrance air kerma 2 mR(μC/kg) Entrance air kerma 3 mR(μC/kg) Entrance air kerma 4 mR(μC/kg) Entrance air kerma 5 mR(μC/kg) Test Result Criteria Judgment Item Judgment Result Lower Limit Reproducibility Air kerma rate Specific radiation output Upper Limit - PASS FAIL mGy/s - PASS FAIL μGy/mAs - PASS FAIL Remarks Signature 13/15 AEC reproducibility Measured Values mAs Entrance air kerma 1 mR(μC/kg) Entrance air kerma 2 mR(μC/kg) Entrance air kerma 3 mR(μC/kg) Entrance air kerma 4 mR(μC/kg) Entrance air kerma 5 mR(μC/kg) Test Result Criteria Judgment Item Judgment Result Lower Limit Entrance air kerma accuracy % Entrance air kerma reproducibility Remarks Signature Upper Limit PASS FAIL PASS FAIL 14/15 CNR mode 1 Exposure Conditions Exposure Mode kVp mAs Target/Filter Compressed Breast Thickness Compression Force S Value Mode n CNR (PMMA20mm) kV mm N Mode n CNR (PMMA40mm) kV mm N Mode n CNR (PMMA60mm) kV mm N Mode n CNR (PMMA70mm) kV mm N Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit CNR relative value 20 mm % - PASS FAIL CNR relative value 40 mm % - PASS FAIL CNR relative value 60 mm % - PASS FAIL CNR relative value 70 mm % - PASS FAIL Remarks Signature 15/15 AGD mode 1 Measured Values Exposure Mode Mode 1 w/o Al Al 0.3 mm Al 0.5 mm Entrance air kerma (20 mm) mR(μC/kg) mR(μC/kg) mR(μC/kg) Entrance air kerma (40mm) mR(μC/kg) mR(μC/kg) mR(μC/kg) Entrance air kerma (60mm) mR(μC/kg) mR(μC/kg) mR(μC/kg) Entrance air kerma (70mm) mR(μC/kg) mR(μC/kg) mR(μC/kg) Test Result Criteria Judgment Item Judgment Result Lower Limit Upper Limit AGD 20mm mGy - 1 mGy PASS FAIL AGD 40mm mGy - 2 mGy PASS FAIL AGD 60mm mGy - 4.5 mGy PASS FAIL AGD 80mm mGy - 6.5 mGy PASS FAIL AGD - ACR Phantom Measured Values Exposure Mode Mode 1 w/o Al Entrance air kerma ( Al 0.3 mm mR(μC/kg) Al 0.5 mm mR(μC/kg) mR(μC/kg) Test Result Criteria Judgment Item AGD-ACR Phantom Judgment Result mGy Lower Limit Upper Limit - 3 mGy Remarks Signature PASS FAIL Chapter 9 Technical Information Page 9.1 Specification Outline of 1Shot Phantom M Plus ___________ 9-2 9.2 Glossary_______________________________________________ 9-3 9 FDR Mammography QC Manual - 897N101461A 9-1 9.1 Specification Outline of 1Shot Phantom M Plus Test Contents Missed tissue on chest wall edge CNR 1Shot Phantom sensitivity constancy Geometric distortion System artifact evaluation Uniformity (Uniform Exposure Region) Dynamic range Spatial Resolution (SR) Low Contrast Detectability (LCD) Linearity/Beam quality constancy Applicable Phantom Outline of Calculation Method Bar patterns (1 mm spacing) The amount of missed tissue on the chest wall-side edge is calculated by using coordinate information of the slit. Al (0.2 mm) CNR is calculated according to the definitional equation described in IEC 61223-3-2. A calculation area (20 x 20 mm) is set at the lateral center and also at 60 mm from the chest wall-side edge. Uniform exposure region at 60 mm from the chest wall- Based on the pixel value in this calculation area, the value side edge inversely proportional to the air kerma is calculated. Note: Although this value is calculated in the same way as S value, this value and S value need to be discriminated. Lines indicated with triangle markers (100 mm spacing) The distance between the two lines is calculated. Uniform exposure region made of metal and plastic (The transmission factor becomes equivalent to 40-mm PMMA N/A phantom at 28kVp, Mo/Mo.) A calculation area (10 x 10 mm) is set at the lateral center Uniform exposure region at 60 mm from the chest wall- and also at 60 mm from the chest wall. Based on this side edge and uniform exposure regions at four corners calculation region, the pixel value and relative SNR value of the phantom are calculated in the calculation regions (10 x 10 mm each) at four corners of the phantom. The lightest region of the step wedge The pixel value in the highest density region of the step wedge is calculated. According to IEC 61223-3-2, calculation is executed only in the highest density region. Bar patterns slanted at 45 degrees (2, 4 and 8 cycles/ SCTF is calculated according to the definitional equation mm) described in IEC 61223-3-2. PMMA phantom ( 2 mm, contrast 1.4%) A cross-correlation factor is calculated based on the reference image (ideal image). The pixel value is calculated in each step of the step Five step-wedge patterns forming a density region in the wedge and then the difference between the adjacent range of the center density plus or minus about one digit steps is calculated. 9 Technical Information 9-2 FDR Mammography QC Manual - 897N101461A 9.2 Glossary This section explains the terms used in this Guidebook. 1 Additive Lag Effect Additive lag effect is a phenomenon in which a faint image that should be present under normal conditions can be seen due to insufficient erasing performance of the X-ray equipment. 2 AGD: Average Glandular Dose In this Program, calculation formulas*1 adopted in IEC 61223-3-2 Ed.2 and EUREF Ed.4 are used. AGD is an index for adjusting or evaluating the performance of AEC. *1: DANCE, DR et al. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys. Med. Biol., 2000, Vol. 45, 3225-3240 3 Air Kerma (Kerma: kinetic energy released in materials) The air kerma is expressed as the sum of initial kinetic energy of all charged particles released in air by uncharged radiation divided by the air per unit mass. Since the air kerma (exposure) is obtained by measuring secondary electrons released as a result of interaction between radiation and air, the air kerma and secondary electrons are equivalent. The unit of the air kerma is Gy. 4 Air Kerma Rate In this Program, the air kerma rate means the exposure rate (exposure per unit time). 5 CNR: Contrast to Noise Ratio CNR is an index for adjusting or evaluating the performance of AEC. The higher the contrast or the lower the noise level, the greater the value of CNR. In this Program, measurement and calculation methods adopted in IEC 61223-3-2 Ed.2 are used. 6 Dynamic range In IEC 61223-3-2 Ed.2*2, visual evaluation on how many thin steps can be seen at the high dose side is executed using a stepped PMMA phantom. In this Program, based on the principle of IEC, a region for direct line has been reserved in the phantom and dynamic range is quantified by calculating the average pixel value in this region. If this value is the maximum value (in the case of 14 bits: 16384), there is a possibility of image saturation. *2: IEC 61223-3-2 Ed.2 Committee draft for vote 7 HVL: Half Value Layer HVL is defined as the thickness of the aluminum plate that reduces the air kerma rate to half. This is an index for representing the radiation quality. The harder the radiation quality, the greater the value of HVL. FDR Mammography QC Manual - 897N101461A Technical Information The ghost factor is an index for representing the degree of change in sensitivity of the flat panel detector (FPD) caused by the last X-ray exposure. The greater the degree of change in sensitivity, the greater the ghost factor. 8 9 Ghost Factor 9-3 9 Lag Factor The lag factor is an index for evaluating the erasing performance of the X-ray equipment. A calculated median value of the histogram of the remaining image that was not erased is displayed. In this Program, measurement and calculation methods adopted in AAPM Task Group 10 Report*3 are used. *3: Ehsan Samei et al. Performance evaluation of computed radiography systems. Medical Physics March 2001 Volume 28, Issue 3, pp. 361-371 Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems. Report of AAPM Task Group 10 October 2006 10 Low Contrast Detectability This is an index for representing the ability to display a low-contrast sample such as a Burger phantom. The cross-correlation factor between a column-shaped object (concave or convex) and ideal image data is shown quantitatively. 11 Multiplicative Lag Effect Multiplicative lag effect means change of sensitivity of the flat panel detector (FPD) caused by the last X-ray exposure. 12 Pixel Value Ratio The pixel value ratio is an index for evaluating image uniformity. (Average Pixel Value in the Intended Calculation Region) / (Average Pixel Value in the Base Calculation Region) Each pixel value (QL value) is converted to an antilog before calculating the average pixel value. 13 PMMA phantom: PolyMethylMethAcrylate phantom In general, the material of the PMMA phantom is called acrylic or polymethylmethacrylate. The PMMA phantom is used as an object for simulating breasts in standards and guidelines such as IEC 61223-3-2 Ed.2 and EUREF. The relationship between the thickness of the PMMA phantom and that of the compressed breast (mammary gland: 50/fat: 50) is shown in the table below. For example, AGD of 20-mm-thick PMMA and that of 21-mm-thick compressed breast is the same. PMMA thickness (mm) Equivalent breast thickness (mm) 20 21 40 45 60 75 70 90 (Source: EUREF Ed.4) 9 Technical Information 9-4 FDR Mammography QC Manual - 897N101461A 14 QL Value This Program executes various calculations regarding image data before EDR normalization. With the FCR 1Shot Phantom M Plus, log data equivalent to 4 digits/12 bits is output. 1 QL is equivalent to 0.90%. 12 bits/4 digits 1 QL = 4 (digits)/212 = 9.77 x 10-4 (digits) -4 When converted to the antilog: 10(9.77 x 10 ) = 1.00225 0.225% However, dynamic range calculation is performed at 14 bits. 1,6ALUE BIT M2 M2 M2 M2 2 $IGITS $OSE $IGIT Figure 1 Relationship between QL Value and Digit (before EDR Normalization) 15 Relative Sensitivity The relative sensitivity is a relative value calculated by using the average pixel value in the calculation region exposed under a certain condition, based on the medial value calculated by using the average pixel value in the specified calculation region of the flat panel detector (FPD) to which a standard amount of dose is exposed. 16 SID: Source to Image Distance SID is the distance between the X-ray focal spot and image receptor. Each X-ray equipment has a specific SID value. The SID is 65 cm for the FDR1000DRSZ. 17 SNR: Signal to Noise Ratio SNR represents the ratio between the signal (intended image) and noise. This is an index for representing image quality. The greater the SNR value, the less the intended image is buried in noise. SNR = (Average Pixel Value in the Calculation Region) / (Standard Deviation of Pixel Value in the Calculation Region) 18 Spatial Resolution Spatial resolution is an index for representing the ability of the imaging system to clearly define the shape of objects within an image. 9 19 System Sensitivity FDR Mammography QC Manual - 897N101461A Technical Information The system sensitivity is equivalent to the median value of the histogram of the digital image at the time of converting the dose on the flat panel detector (FPD) to a digital value. The system sensitivity is defined depending on the type of FPD. In this Guidebook, the high-resolution FPD is used. Accordingly, the system sensitivity is defined as follows. When uniform exposure is executed under the condition of 25 kVp and 0.03 mm Mo using X-ray equipment with a molybdenum target, if the dose on FPD is specified as y (mR) (1 (mR) ≈ 8.7 (µGy)): S Value = 2400/y 9-5 9 Technical Information 9-6 FDR Mammography QC Manual - 897N101461A Image Processing Parameters (for Mammography QC) This chapter describes the image processing parameters for study menus and exposure menus used in this Program. Page FDR Mammography QC Manual - 897N101461A How to Read This Chapter ______________________________ QCParam-2 EDR Mode Table ____________________________________ • Image Processing Parameters for Mammography QC _____ QCParam-3 QCParam-6 QCParam-1 How to Read This Chapter How to read descriptions included in this chapter is described below. 3 2 1 Exposure Menu MPM Code Study Menu 4 5 AP/PA 1 Menu Description Study Menu (Exposure Menu) The Study Menu is an aggregate of menus to be used for exposure of a series of study. The Exposure Menu included in a Study Menu, as well as order of performing studies is determined by the default, which, however, can be changed in the User Utility. (For details, see the descriptions related to the User Utility in the Operation Manual for the system.) * Exposure menu in other chapters is defined as study menu in this chapter. * The Study menu field is grayed out for the parameters for high-density film. 2 Exposure Menu (Exposure Submenu) The Exposure Menu is the name of a single study such as “L MAMMOGRAPHY, CC”. * Exposure submenu in other chapters is defined as exposure menu in this chapter. * Default settings for each exposure menu can be changed using the User Utility. 3 MPM Code The MPM Code is a four-digit code number assigned for the purpose of management of exposure menus. The MPM Code determines EDR (a function that corrects image density and contrast automatically) and image processing conditions to be applied. If the assigned MPM Code is the same, images will be output according to the same conditions even though the Exposure Menu used is not the same. Note that when an exposure menu (for example, L MAMMOGRAPHY, CC: MPM Code 0329) was subject to change of image processing conditions, the change will affect all the menus concerned if there is an exposure menu of the same MPM code (such as R MAMMOGRAPHY, CC). Should you wish to change the image processing condition only for the specified menu, change the 3rd-digit figure to make a different MPM code so that specific conditions are set up appropriately. 1st digit : Represents an exposure technique to be used. (ex. 0: General exposure, 1: Contrast exposure, etc.) 2nd digit : Represents an anatomical part to be exposed. (ex. 3: Breast) 3rd digit : Any alphanumeric selected from 0 to 9 and A to F. Even if a figure in this digit is changed, EDR will not be affected. 4th digit : Any alphanumeric that determines EDR. 4 AP/PA - Flipping setting for images to be output AP : Outputs an image as is without processing it. PA : Outputs an image flipped horizontally. : Outputs an image rotated by 180 degrees. : Outputs an image flipped vertically. * Default settings can be changed using the User Utility. It is also possible to change default settings by the currently used mode. 5 Menu Description The Menu Description describes anatomical parts and exposure techniques suited to a specific exposure menu. Brief precautions to be observed when performing exposures are also included. For details on those precautions to be observed, see the Operation Manual for the system. • When creating a new menu A new menu can be created using the User Utility. (For how to create a new menu, see the descriptions related to the User Utility in the Operation Manual for the system.) For a special exposure that cannot be handled by default exposure menus, the menu can be added as necessary. When doing so, select an exposure menu that involves similar images. See the descriptions related to the User Utility in the Operation Manual for the system to make sure that the MPM code to be used for the new menu is not used, and then determine proper image processing conditions. Also confirm that AP or PA is selected correctly. QCParam-2 FDR Mammography QC Manual - 897N101461A EDR Mode Table EDR mode (auto sensitivity adjustment system) applied to exposure menus pre-registered in the system are described below. S value : When recording a digital image, the X-ray dose that reaches the exposure unit is converted to a digital value. S value is the center X-ray dose of the histogram of the digital image. L value : A logarithmic value showing the range of X-ray dose when making an exposure. Semi Fix S Value Type L Value Type 1 PRIEF EDR mode Auto EDR Mode EDR mode consists of AUTO mode, SEMI AUTO mode, and FIX mode. A (AUTO mode) S (SEMI AUTO mode) F (FIX mode) : A mode which adjusts density and contrast automatically. (S and L values are dependent on this mode.) : A mode where the dynamic range (L value) of X-ray dose to be recorded as an image has been determined, and the center point (S value) used for the purpose of image recording is decided based on the average X-ray dose that enters the preset photometric area so that the density is adjusted automatically. : A mode where the range of X-ray dose to be recorded as an image has been determined. * Default settings for this mode can be changed using the User Utility. It is also possible to change the mode type currently being used. 2 Auto - Parameters used in AUTO mode PRIEF (Pattern Recognizer for Iris of Exposure Field) ..... This is a generic denomination of processing that recognizes split exposures and irradiated field automatically. PRIEF includes the following technique types: 1 1S 2 4 4S : : : : : : Does not recognize split exposures and irradiated field. (SEMI AUTO mode and type IV described below.) Does not recognize split exposures and judges a rectangular area as an irradiated field. Recognizes split exposures and judges individually recognized areas as rectangular irradiated fields. Recognizes irradiated field of a breast. Does not recognize split exposures and judges a protrusive area as an irradiated field. Recognizes split exposures and judges individually recognized areas as protrusive irradiated field. (normal mode) 4* : Judges a protrusive area as an irradiated field by split areas determined. AN : Auto neck algorithm SP : Activates AUTO mode based on a specially determined area, irrespective of the image size specified by DR equipment. TYPE ..... A type of technique for histogram or neuro analysis subjected after PRIEF processing. I II III IV V VI VII 3 : A mode that captures regions covering from the skin to the bone in an image (Note that direct X-rays are needed to activate this mode.) : A mode that is activated in a stable manner even if there are no direct X-rays. : A mode applied to contrast exposure. : A mode that attaches importance to improved representation of soft tissue. : A mode that attaches importance to improved representation of areas where X-rays are difficult to be penetrated. : Neuro analysis mode applied when variations become large on the shape of a histogram. : Neuro analysis mode applied when the position of a region of interest changes on a histogram. Semi Fix - Parameters used in SEMI AUTO mode and FIX mode TYPE ..... Determines layout and size of a photometric area preset in SEMI AUTO mode. I : A 10cm square located at the center of an image exposed by DR equipment. II : A 7cm square located at the center of an image exposed by DR equipment. III : A 5cm square located at the center of an image exposed by DR equipment III' : A 5cm square located at a position other than the center of an image when it is divided into nine portions up-and-down and right-and-left. IV : A special area determined for the chest. L value ........... A logarithmic value (L value) representing the width of an X-ray dose to be recorded as an image in SEMI AUTO and FIX modes. S value ........... A center point (S value) pre-determined in FIX mode for recording as an Image. * Default settings for this mode can be changed using the User Utility. It is also possible to change default settings by the currently used mode. FDR Mammography QC Manual - 897N101461A QCParam-3 1 2 Image Processing Parameters 1 GA GT GC GS MRB MRT MRE MDB MDT MDE PRN PTE PTC PRE 1.3 T 1.40 -0.04 G R 0.5 E F 0.3 B R O 1.0 Image Processing Parameters - MFP (Multi-Objective Frequency Processing) parameters Gradation processing ..... Processing that controls image gradation. GA : Adjusts contrast appropriately. As the numeric value increases, the contrast becomes enhanced. GT : A non-linear gradation curve. GC : Center of a density when the GA value is changed. GS : Adjusts density appropriately. As the numeric value increases, the density appears enhanced. Frequency processing ..... Processing that controls the image sharpness. MRB : A factor that determines the range of frequency bands when applying image enhancement. As the numeric value decreases, the range of frequency bands is widened toward lower-frequency bands. (See the figure below.) MRT : A non-linear curve that changes the degree of enhancement according to the image density. This parameter enhances specific density areas. Example : F : Applies enhancement uniformly in all density areas. R : Applies stronger enhancement as density rises. MRE : This factor adjusts the degree of enhancement. DR compression processing .... Processing to make image density areas that appear white or blackened easily visible, without affecting the density in the region of interest on an image. MDB : A factor that determines the smoothing mask for DR compression processing. MDT : A factor that determines density areas where DR compression processing is to be applied. MDE : A factor that determines the degree of DR compression processing to be applied. 2ESPONSE ! " # $ % & ' &REQUENCYCYMM Figure : MRB types and enhancement curves A stronger enhancement is applied in high-frequency bands as going from A to F. 2 Image Processing Parameters - PEM (Pattern Enhancement Processing for Mammography) parameters There are the following four PEM processing parameters : PRN, PTE, PTC, and PRE. PRN : This is a parameter to determine the frequency bands where enhancement is to be applied. A stronger enhancement is applied in low frequency bands as going from A to E. A : 8.0 cy/mm B : 5.66 cy/mm C : 4.0 cy/mm D : 2.83 cy/mm E : 2.0 cy/mm QCParam-4 FDR Mammography QC Manual - 897N101461A PTE : This is a parameter to determine edge detection sensitivity. The edge detection sensitivity decreases as going from A to T, making it difficult to detect portions that look like calcifications. PTC : This is a parameter to determine detection sensitivity of an isolated point. The isolated point detection sensitivity decreases as going from A to T, making it difficult to detect portions that seem to be calcifications. PRE : This is a parameter to determine the degree of enhancement. The larger the value is, the larger the degree of enhancement is. If this parameter is set to 0.0, no PEM processing will be applied. * Depending on the exposure conditions applied, radiographic object, and equipment being used, the PEM processing may be applied excessively, resulting in affecting reading of calcifications, etc. Prior to actually applying PEM processing, make sure that it does not affect image reading adversely. * If the PEM processing is judged to be affecting image reading, make sure that the applied parameters are appropriate by, for example, setting the PRE value smaller to lower the degree of enhancement. * PEM processing refers to the S value applied to the image and adjusts the degree of image detection, requiring you to check the value periodically. S value can be controlled effectively by using the QC Tool (optional software). If no QC Tool is avaialble, you can determine rough behaviors of the S value by using an appropriately controlled X-ray generator and checking the resulted S value of the image data acquired from an exposure made under certain conditions. Controllable range of the S value is about twice the resulted value. FDR Mammography QC Manual - 897N101461A QCParam-5 QCParam-6 : : : : Rotation Amount Gradation Type Rotation Center Gradation Shifting Amount 7375 7313 7313 7313 7313 7313 7302 7325 7375 7325 [Daily/Weekly] For checkup of mammography QC ACR Phantom AP AP AP AP AP AP [Annual] For checkup of mammography QC Lag1 [Annual] For checkup of mammography QC Lag2 [Annual] For checkup of mammography QC Ghost1 [Annual] For checkup of mammography QC Ghost2 [Annual] For checkup of mammography QC Ghost3 [Annual] For checkup of mammography QC Missed tissue on chest wall edge (40 mm height) [Semi-annual] For checkup of AP mammography QC Uniform Exposure(Image Basic) [Daily/Weekly] For checkup of mammography QC 1Shot Phantom M [Daily/Weekly] For checkup of AP mammography QC ACR Phantom AP AP [Daily/Weekly] For checkup of mammography QC 1Shot Phantom M S F F F F F S S S S S - - - - - - - - - - - V V V V V V V V V V V V III' I I I I I I III' III' III' III' I 2 2 2 2 2 2 1 2 2 2 2 1 200 200 200 200 200 200 200 200 200 200 200 200 1.2 1.3 1.3 1.3 1.3 1.3 1.0 1.2 1.2 1.2 1.2 1.0 T C C C C C A T T T T A 1.40 1.20 1.20 1.20 1.20 1.20 1.20 1.40 1.40 1.40 1.40 1.20 1.20 -0.05 0.00 0.00 0.00 0.00 0.00 0.00 -0.04 -0.05 -0.04 -0.05 0.00 0.00 G B B B B B B G G G G B B R R R R R R F R R R R F F 1.5 0.0 0.0 0.0 0.0 0.0 0.0 1.5 1.5 1.5 1.5 0.0 0.0 * High luminance is recommended in the environment where high-density film is used and normal luminance in the environment where a monitor, which does not support high-density film, is used. * It is not recommended that you use the PEM processing parameters for the tests other than the Test with ACR Phantom. 1Shot PhantomM (40mm) Uniform Expo.2 (Ghost) ACR Phantom2 (Ghost) Uniform Expo.1 (Ghost) Non Exposure (Lag) Annual 1/6 ACR Phantom1 (Lag) Semi-annual 1/1 Uniform Expo(Image Basic) ACR Phantom 1Shot/ACR 1Shot PhantomM MPM Code ACR Phantom ACR Phantom AP/PA AP EDR mode 7375 PRIEF 1Shot PhantomM 1Shot PhantomM Type - Type S L Value [Calculation] For checkup of mammography QC AGD S Value AP GA 7302 GT A GC 1.0 GS 200 MRB 1 MRT I MRE V E A A A A A A E E E E A A MDB - F A A A A A A F F F F A A MDT S 0.4 0.0 0.0 0.0 0.0 0.0 0.0 0.4 0.4 0.4 0.4 0.0 0.0 Image Processing Parameters (PEM) PEM Enhancement Frequency Ranking PEM Edge Information Detection Level PEM Calcification Image Information Detection Level PEM Enhancement Coefficient MDE Calculation 2/2 AGD [Calculation] For checkup of mammography QC CNR : : : : Image Processing Parameters (MFP) PRN PTE PTC PRE B A A A A A A B B B B A A PRN AP Semi Fix DRN : Dynamic Range Control Rank DRT : Dynamic Range Control Type DRE : Dynamic Range Control Enhancement Level R A A A A A A R R R R A A PTE 7302 Menu Description Auto RN : Frequency Rank RT : Enhancing Type RE : Frequency Enhancement O A A A A A A O O O O A A PTC Calculation 1/2 CNR Exposure Menu Study Menu • Image Processing Parameters for Mammography QC GA GT GC GS 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.7 0.0 0.7 0.0 0.0 0.0 PRE FDR Mammography QC Manual - 897N101461A FDR Mammography QC Manual - 897N101461A 7302 7302 7302 7302 7302 7302 7302 7302 7302 7302 7302 [Annual] For checkup of mammography QC W - mode CNR (20 mm height) [Annual] For checkup of mammography QC AP W - mode CNR (40 mm height) [Annual] For checkup of mammography QC AP W - mode CNR (60 mm height) [Annual] For checkup of mammography QC AP W - mode CNR (70 mm height) AP [Annual] For checkup of mammography QC AP L - mode CNR (40 mm height) [Annual] For checkup of mammography QC AP L - mode CNR (60 mm height) [Annual] For checkup of mammography QC AP L - mode CNR (70 mm height) [Annual] For checkup of mammography QC AP L - mode CNR (20 mm height) [Annual] For checkup of mammography QC H - mode CNR (20 mm height) [Annual] For checkup of mammography QC AP H - mode CNR (40 mm height) [Annual] For checkup of mammography QC AP H - mode CNR (60 mm height) [Annual] For checkup of mammography QC AP H - mode CNR (70 mm height) S S S S S S S S S S S S - - - - - - - - - - - - V V V V V V V V V V V V V I I I I I I I I I I I I I 1 1 1 1 1 1 1 1 1 1 1 1 2 200 200 200 200 200 200 200 200 200 200 200 200 200 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.0 1.3 A A A A A A A A A A A A C 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.20 1.40 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 -0.05 B B B B B B B B B B B B B G F F F F F F F F F F F F R R 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 1.5 * High luminance is recommended in the environment where high-density film is used and normal luminance in the environment where a monitor, which does not support high-density film, is used. * It is not recommended that you use the PEM processing parameters for the tests other than the Test with ACR Phantom. W-mode CNR (PMMA70mm) W-mode CNR (PMMA60mm) W-mode CNR (PMMA40mm) Annual 6/6 W-mode CNR (PMMA20mm) L-mode CNR (PMMA70mm) L-mode CNR (PMMA60mm) L-mode CNR (PMMA40mm) Annual 5/6 L-mode CNR (PMMA20mm) H-mode CNR (PMMA70mm) H-mode CNR (PMMA60mm) MPM Code H-mode CNR (PMMA40mm) AP/PA AP EDR mode 7302 PRIEF - Type S Type Annual 4/6 H-mode CNR (PMMA20mm) L Value [Annual] For checkup of mammography QC Collimation S Value AP GA 7303 GT T GC 1.2 GS 200 MRB 2 MRT III' MRE V A A A A A A A A A A A A A E MDB - A A A A A A A A A A A A A F MDT S 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.4 MDE Annual 3/6 Uniform Expo.3 (Collimation) [Annual] For checkup of mammography QC Spatial Resolution (Magnification) Image Processing Parameters (MFP) Image Processing Parameters (PEM) A A A A A A A A A A A A A B PRN AP Semi Fix A A A A A A A A A A A A A R PTE 7375 Menu Description Auto A A A A A A A A A A A A A O PTC Annual 2/6 1ShotM (Magnification) Exposure Menu Study Menu 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 PRE QCParam-7 United States Authorized Representative: FUJIFILM MEDICAL SYSTEMS U.S.A., INC. 419 WEST AVENUE, STAMFORD, CT 06902, U.S.A.