RADFORD LABPACKS The Radford Labpack laboratory bench power pack appeared in 1959 and was one of the first portable low-voltage power supplies for school use. Many school physics labs were equipped with Labpacks which provided a wide range of ac and dc outputs particularly suitable for work with thermionic valves. During late 1995 and early 1996 a number of accidents were reported to CLEAPSS in which pupils and staff received serious electric shocks from Labpacks. CLEAPSS had already recommended that the 300V dc output be disabled or key locked, since it could deliver dangerously high currents. However, some of the shocks reported were received from the low-voltage outputs. This was due to deterioration or damage to the internal insulation caused by overloading which may not have been detected by the standard electrical portable-appliance test. Further investigation has led us to conclude that several aspects of the original design of Labpacks would not satisfy the safety requirements for the design of school laboratory power supplies today. 1. CLEAPSS recommendations 1.1 If at all possible, schools should take Radford Labpacks out of use now and replace them. 1.2 If a Labpack must continue in use, it needs a thorough internal inspection, a special electrical test and may need immediate modification to meet higher safety standards. 1.3 The regular portable appliance test must now include internal inspection and a more severe electrical test. 2. Replacement The Labpack has proved to be a very versatile power supply, different variants providing all or some of the following outputs. AC 6.3 V 3A (fixed) AC 12 V 8A (fixed) AC 0 to 22 V 8A (variable in 0.2 V steps) DC 0 to 22 V 8A (variable in 0.2 V steps) HT DC 300 V 150 mA (fixed) It is understandable that some science staff are reluctant to dispose of Labpacks. Modern power supplies are rarely so versatile. However, for most school laboratory activities, particularly day-to-day experiments, a simpler low-voltage power supply is adequate and preferable. Modern power packs have more straight-forward controls and weigh considerably less. Schools now intending to replace Labpacks will find the CLEAPSS guide Generalpurpose low-voltage power supplies, R59a, very useful. Low-voltage power supplies cost between £80 and £180 each, depending on facilities. Where an HT supply is still required, usually for fine beam tube or thermionic investigations in A-level Physics, most schools PS29 JRE 8/97 Page 1 of 4 ® © CLEAPSS , Brunel University, Uxbridge UB8 3PH Tel: 01895 251496; Fax: 01895 814372; E-mail: science@cleapss.org.uk; Web site: www.cleapss.org.uk would only need one unit. HT supplies should include a 6.3 V output for valve heater filaments. Prices from the major education suppliers range from £120 to £360, so it is worth shopping around. We believe that money must be spent on replacing Radford Labpacks if safety is not to be compromised. Power supplies are essential and often daily-used pieces of equipment in school laboratories. Our investigations have shown that a very large number of variants of the Radford Labpack left the factory; even those with identical model numbers had different features. Some Labpacks we have seen have been repaired and/or modified several times by science staff, technicians or repair companies. The quality of the repairs and modifications is variable. As a result, it is hard to find two completely identical Labpacks, even within the same school. Some are intrinsically safe, but others have fallen well short of basic standards. This ad hoc situation is unacceptable. 3. Continued use If, due to financial limitations, a school is unable to replace its Radford Labpacks immediately, a phased replacement is suggested with as short a time scale as possible. Those Labpacks which are to continue to be used must be inspected, tested and modified according to sections 4 and 5 below. The extensive nature of the modifications and the requirement for specialised components means that modifications should not be undertaken by school staff, but by an electrical repairer with proven experience of Radford Labpacks. This is likely to cost up to £50 per Labpack. At this time we are only able to recommend one repairer (see section 6). However, we hope to approve at least one other and would welcome information to enable us to locate more. 4. Inspection and test Where it is decided that a Radford Labpack will continue to be used, the following inspection and test procedures must be carried out. 4.1 The presence of the modifications listed in section 5 below must be confirmed. 4.2 A thorough, internal inspection must look for signs of overheating of components or wiring, particularly where wires are laced together. 4.3 A normal visual and mechanical portable-appliance test must be carried out. 4.4 A more stringent class 1 electrical test must be carried out by connecting all the outputs together to an earth point before using a standard earth-bond and insulation tester. The insulation resistance should be greater than 1 MΩ. Subsequently, the Labpack should be inspected and tested every term following the procedures in paragraphs 4.2, 4.3 and 4.4. PS29 JRE 8/97 Page 2 of 4 ® © CLEAPSS , Brunel University, Uxbridge UB8 3PH Tel: 01895 251496; Fax: 01895 814372; E-mail: science@cleapss.org.uk; Web site: www.cleapss.org.uk 5. Modifications Each Labpack which remains in use must be checked and then modified where necessary by a competent electrical repairer with experience of Radford Labpacks as follows. 5.1 Mains plug A 3 A fuse must be fitted in the mains plug. 5.2 Cable entry to Labpack Either (if fitted), remove miniature mains socket and hard wire a mains lead to the Labpack with a strain-relief bush, or if the mains lead enters via a grommet, replace the grommet with a strain-relief bush. 5.3 Internal wiring Inspect for signs of damage or charring. Replace any wires showing signs of damage. 5.4 Mains fuse holder, front panel If removable without tools, replace with flush holder having screwdriver slot removal only. 5.5 Mains indicator (neon) If connected between live and earth, reconnect between live and neutral. 5.6 Mains voltage selector Remove voltage selector and hard wire taps on transformer primary to give fixed input voltage. The nominal mains voltage in the UK is now 230 V. 5.7 Transformer Inspect visually for signs of burning or severe overheating. Very occasionally, the transformer needs to be replaced. 5.8 AC 6.3 V output Check wires back to transformer for charring or burning of nearby wires (especially those carrying mains power). Either disable by cutting off wire back to transformer or Replace wires with thicker gauge and install self-resetting thermal-overload circuit breaker. 5.9 AC 12 V output Check wires to transformer for charring or burning. Replace wires if necessary. Install selfresetting thermal-overload circuit breaker. 5.10 AC and DC 0 to 22 V outputs These outputs are both protected by a single circuit breaker. Check operation. Repair or replace as necessary. 5.11 HT DC 300 V output Either disable by cutting off wire to within a few cm of transformer and putting insulators on end of wire or add a key switch to prevent unauthorised use of this output. 5.12 300 V fuse holder, front panel If removable without tools, replace with flush holder having screwdriver slot removal only. 5.13 Blanking plugs Plastic blanking plugs must be fitted with Loctite or similar adhesive to all exterior holes created in the Labpack by any modifications. PS29 JRE 8/97 Page 3 of 4 ® © CLEAPSS , Brunel University, Uxbridge UB8 3PH Tel: 01895 251496; Fax: 01895 814372; E-mail: science@cleapss.org.uk; Web site: www.cleapss.org.uk 6. Recommended repair company Harding Scientific Instruments 163 West Malvern Road Malvern Worcs WR14 4AY Tel: Fax: 01684 575974 01684 893960 PS29 JRE 8/97 Page 4 of 4 ® © CLEAPSS , Brunel University, Uxbridge UB8 3PH Tel: 01895 251496; Fax: 01895 814372; E-mail: science@cleapss.org.uk; Web site: www.cleapss.org.uk