Edwards Thermal Evaporator Standard Operating Procedure NOTE: These instructions are intended for reference only, and will not replace the thorough training required for proper system operation. Contact an experienced user or clean room staff member with questions or to report a system problem. 1. VERIFY SYSTEM STATUS You should find the system at high vacuum; the pressure reading on the LCD display should read “Fine Pumping.” 2. VENT To vent, first release the bell jar’s latching mechanism located on the hoist to the left of the bell jar. Press the “Vent” button on the system’s control panel to fill the chamber with nitrogen. When the system reaches atmospheric pressure, the seal at the bottom of the bell jar will break and you will hear the nitrogen gas leaking from the bell jar. Press the “Seal” button to stop the flow of nitrogen gas. 3. CLEAN SYSTEM Carefully raise the bell jar and rotate it away from the work area. Use a dedicated vacuum to clean up any debris that has been loosened on the very bottom of the chamber. Be careful to not touch any lose or small parts and the thickness monitor. 4. LOAD SAMPLE Mount your substrate on the adapter plate, as close to the center as possible. 5. LOAD SOURCE MATERIAL Place your source material in a turret slot. Rotate the dial on the front of the tool to move the deposition source into the proper evaporation position directly beneath the closed shutter. To avoid damaging the rotating turret mount, be careful to only turn the dial in the clockwise direction. 6. CLOSE CHAMBER Dampen a clean room cloth with IPA and wipe debris from the bell jar’s rubber seal and from the stainless steel base plate. Also, use this cloth to wipe the face of the tool’s ammeter to release any static charge build-up. Carefully rotate the bell jar back into position and lower it to the base plate, then re-clamp the latching mechanism. 7. EVACUATE CHAMBER Press the “Cycle” button on the control panel to begin the pumping process. To assure that the system is evacuating properly, remain with the tool at least until the control panel display changes from “Roughing” to “Fine Pumping”. As a courtesy to other users, set the attached “status” flag to indicate that the system is in use. After the system has reached fine pumping, you may fill the LN2 trap; it is best to wait till the system reaches a pressure below 9x10-6 Torr. 8. SET DEPOSITION PARAMETERS If you haven’t yet done so, you must now swipe into the card reader to activate the tool’s power supply. Assure that the correct material data (layer, density, and tooling factor) have been entered into the rate monitor control panel and that the source turret is correctly positioned. Use the “Data” button to scroll through menu items and the up and down arrows to adjust values. 9. TURN ON POWER Turn on the power supply by turning the voltage selector switch to “LT”; the green light will turn on. The “LT Selector” switch should not be touched and always remains at the “10V B” position. 10. DEPOSIT Slowly increase the current by rotating the power supply dial until the desired deposition rate is reached. Be sure to rotate the dial slowly; increase the power at a rate no faster than an increment of 0.5 every 30 seconds. When the desired deposition rate is reached, begin deposition. Simultaneously open the shutter and zero the rate monitor by pressing the “Run” button on the rate monitor control panel. Monitor the deposition rate and pressure during the deposition. Adjust the current as necessary to maintain the desired deposition rate and reduce pressure spikes. 11. POWER DOWN SYSTEM When the desired film thickness has been reached, close the shutter. Slowly reduce the source current to zero at a rate no faster than an increment of 0.5 every 30 seconds. Turn the voltage supply off by returning the voltage switch from the “LT” to “O” position. 12. RETRIEVE SAMPLE Allow the tool’s internal electrodes to cool for at least 5 minutes. Vent the system as before, by releasing the latching clamp and pressing the “Vent” button. Carefully raise the bell jar and rotate it away from the work area. Use caution – and perhaps a pair of pliers – when removing your source material as the electrodes may still be hot. Remove your coated sample and replace the sample mounting plate in the holder. 13. CLEAN AND PUMP DOWN SYSTEM LEAVE THE SYSTEM AS YOU FOUND IT. Clean the system of loose metal flakes with the vacuum cleaner as you did prior to deposition. Wipe the bell jar’s rubber seal with an IPA-dampened cloth. Carefully return the bell jar to its operational position, set the latching clamp, and evacuate the chamber by pressing the “Cycle” button on the control panel. Before leaving the tool, assure that it is in “Fine Pumping” mode, set the status flag to “Not in Use,” and punch out of the system’s card reader. Complete your entry in the tool’s log sheet.