Optional 2.5KV x100 Probe

advertisement
Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com
Optional 2.5KV x100 Probe
Type
Attenuation
Loading Input Bandwidth
R(MΩ) C(pF)
TT-HV 250
x100
100
4
Rise
Time
Cable Compensation
Length
Range
Max. Input
Voltage
(MHz)
(ns)
(m)
(pF)
(V)
300
(-3dB)
1.2
1.2
10 to 50
2500 pk*
*Max. Input Voltage, DC incl. Peak AC (derate with rising frequency)
Page 1 of 1
Newtons4th Ltd
30 Loughborough Rd
Mountsorrel
Loughborough
LE12 7AT UK Tel: +44 (0)116 2301066
Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com
Optional 15KV x1000 Probe
Type
Attenuation
Loading Input Bandwidth
R(MΩ) C(pF)
TT-HVP
15HF
x1000
(+/- 2% @
1MΩ)
100
1
Rise
Time
Cable Compensation
Length
Range
Max. Input
Voltage
(MHz)
(ns)
(m)
(pF)
(V)
50
(-3dB)
7
2,0
15 to 30
10kVrms*
(15kV DC
30kVpk)
*Max. Input Voltage derates with rising frequency
Dimensions (L x W x H): 350mm x 75mm x 75mm
IEC1010: CAT I Pollution Degree 2
Page 1 of 1
Newtons4th Ltd
30 Loughborough Rd
Mountsorrel
Loughborough
LE12 7AT UK Tel: +44 (0)116 2301066
Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com
Ultra Low Capacitance Probe
ULCP
Accessory for use
with the PPA range
power analyzer in
applications where
capacitance
to
ground must be
minimized.
APPLICATION
In order to minimize high frequency noise components that may be present in a power measurement
application, most power analysis instruments incorporate high voltage decoupling capacitors at the
input stage of the voltage and current measurement channels. The capacitance to ground of a power
analyzer input channel is therefore a combination of parasitic and decoupling capacitance and this is
typically in the order of 50pF.
While this capacitance provides useful high frequency noise attenuation and is not detrimental to most
power measurement applications, it can in certain cases be problematic. Examples of this include
applications with high frequency high voltage waveforms such as high switching speed inverter drives
or where capacitive loading must be minimized such as the start up of low power ballasts.
SOLUTION
The Ultra Low Capacitance Probe uses a differential input arrangement that is optimized for minimum
capacitance to achieve an input capacitance to ground less than 1pF while maintaining an input
voltage range up to ±3kVpk.
OPERATION
The ULCP inputs are connected directly to the voltage connection points of a DUT and a touchproof
BNC cable connects the output of the ULCP to the external voltage input of a PPA series power
analyzer.
SPECIFICATION
Capacitance to ground:
Voltage rating:
Voltage connections:
Dimensions:
Input power:
≤ 1pF
Bandwidth:
1kVrms, ± 3kVpk
Nominal attenuation:
4mm safety sockets
PPA Interface:
100 x 80 x 50 mm
Weight:
9V ac adaptor supplied standard or ± 12V dc
Newtons4th Ltd 30 Loughborough Rd Mountsorrel
dc to 2MHz ±1dB
1000:1
BNC connection
400 g
Loughborough LE12 7AT UK Tel +44 (0)116 2301066
Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com
PPA Range Accessories
ATT10 Wideband
Attenuator
The ATT10 Wideband attenuator provides a very flat frequency response over a
wide frequency range. Giving 10:1 attenuation allowing a 30V signal to be
attenuated to the input rating of the BNC external voltage input.
Voltage input:
BNC
Voltage rating: 30V pk(ATT10)
Connections to PPA: BNC
<0.5dB@25MHz <3dB@30MHz
Voltage output:
BNC
Frequency range DC-30MHz
Voltage scale factor: 10–1(ATT10)
Nominal accuracy : 0.1%
Note: Specification covers previously labelled DC-2MHz ATT20
Newtons4th Ltd
30 Loughborough Rd
Mountsorrel
Loughborough
LE12 7AT UK
Tel: +44 (0)116 2301066
Newtons4th Ltd
office@newtons4th.com
www.newtons4th.com
PPA Range Accessories
ATT20 Wideband
Attenuator
The ATT20 Wideband attenuator provides a very flat frequency response over a
wide frequency range. Giving 20:1 attenuation allowing a 60V signal to be
attenuated to the input rating of the BNC external voltage input.
Voltage input:
BNC
Voltage rating: 60V pk(ATT20)
Connections to PPA: BNC
<0.5dB@25MHz <3dB@30MHz
Voltage output:
BNC
Frequency range DC-30MHz
Voltage scale factor: 20–1(ATT20)
Nominal accuracy : 0.1%
Note: Specification covers previously labelled DC-2MHz ATT20
Newtons4th Ltd
30 Loughborough Rd
Mountsorrel
Loughborough
LE12 7AT UK
Tel: +44 (0)116 2301066