04-XX-00-1 FT- i40 Ex2-22.fm

advertisement
i 40Ex 2-22 1
Options
Communication
Communication by field bus
Weight thresholds
• 4 digital outputs 30 V AC/DC - 100 mA galvanically isolated.
• Gross weight or net weight threshold definition.
• Output activated if the current weight is >= or <= threshold
value.
• Thresholds on a positive or negative weight value.
Remote control
•
•
•
•
•
PROFIBUS-DP*
DeviceNet*
Ethernet Modbus-TCP*
EtherNet/IP*
CANopen
The CANopen bus also allows connecting a d 20 remote display
and integrating an i 40 Ex 2-22 into an I 400 network.
Analog output*
•
•
•
•
4-20mA or 0-10V analog output
1500V RMS galvanic isolation
Reading of negative weight values in the 0-4 mA range
Instant gross or net weight image
2 galvanically isolated digital inputs:
• Voltage range...................................................................................... 8-30 V DC
• Max. current..........................................................................................
15 mA
* Exclusive options: Either one of field bus connections marked with an asterisk,
or an analog output.
Allow remote zero and tare setting operations.
Serial communication
Setting
The indicator's metrological and functional settings can be made:
• Via the indicator's front panel;
• Via PC and Internet browser.
Parameter backup/restore via optional μSD card or PC.
Preventive maintenance help
• Recorded weight overload counter.
• Alert display on 3 levels according to the number of weighings taken.
Weight indicator for process applications
WWW.PRECIAMOLEN.IN
WWW.PRECIAMOLEN.COM
• RS232/RS485 configurable port.
• Protocols: Slave RTU Modbus, PM A+ slave and PM A+
master repeater.
WWW.PRECIAMOLEN.IN
Features
• Peripheral type USB port
• RS 232 port
• Protocols: Slave RTU Modbus, PM A+ slave and PM A+
master (repeater).
TCP/IP communication
• 10/100 Mb RJ 45 Ethernet port
• PM A+ slave and PM A+ master (repeater) protocols.
μSD card
• Allows parameter backup / restore.
• Data Storage Device (DSD) with 300,000 records for trade
use.
Panel mounting kit
• For integrating an i 40 Ex 2-22 indicator into a cabinet or
console door..
i 40 Ex 2-22 weight indicator is especially suitable for process weighing and integrates easily
into any industrial process requiring weight measurement.
< The
< In
accordance with directive 94/9/CE, this indicator may be installed in a area that is dangerous
due to risks related to the presence of dust (area 22) and/or of gases (area 2).
< Various
connectivity options enable direct connection of the i 40 Ex 2-22 indicator to most PLCs
and industrial PCs.
Precia Molen India Ltd
(formerly Nova Weigh India Ltd)
L-13, Dr.V.S.I. Estate, Phase II, (Near S.R.P Tools) Thiruvanmiyur, Chennai-600 041. India
Tel : +91 44 2454 2009 / 10 / 11 Fax : +91 44 2454 4013
E-mail : info@preciamolen.in | www.preciamolen.in
• FSC (Field Sequential Colour) segments technology
• Up to 200,000 divisions for non-trade use.
• Configurable 7 colour backlighting
• Sensitivity ..............................................................................................
2 mV/V
• Weight display on 7 x 25 mm high characters
• Minimum voltage step per step........................................
0.5 ȝV
• 10 x 8 mm high character text field
• Load cell power ................................................................................
5V
• Weight unit: g, kg, t and lb.
• Minimum load cell impedance............................................
43 ȍ
• Field bus metrological and activity indicator
• Maximum load cell impedance ..........................................
1245 ȍ
• High viewing angle and contrast
• 4/6 wire connection
• French / English bilingual messages
• Up to 100 measurements / second
WWW.PRECIAMOLEN .IN
Metrological characteristics
WWW.PRECIAMOLEN .IN
Display
Physical characteristics
• Cascadable digital filtering on 2 configurable levels
Keyboard
Metrological certification
• Touch control keys.
• 3 metrological keys: Zero, tare and gross weight recall
Test certificate issued by a notified body in accordance with
Directive 2009/23/CE for Non-Automatic Weighing instruments.
• 3 menu access keys
• Class III IPFNA
• 4 key directional keypad
• Single range 1 x 3 000 or 1 x 6 000
• Dual range 2 x 3 000 or 2 x 6 000
Environmental characteristics
• Power supply.......................................................................
9 to 30 V DC*
ATEX Certification 1
Dimensions in mm
• Maximum consumption:
Typical ........................................................................................
1.4 W
With 4-20 mA / 0-10 V option ............................
1.8 W
With field bus option....................................................
2.7 W
• Comply with the requirements of ATEX directive 94/9/CE
related to protection instruments and systems designed
for use in explosive atmospheres.
• Marking according to EN 60079-0:2012
II3G Ex nA IIC T6 Gc
•
•
•
•
•
304L stainless steel housing
IP66 - IP69K protection rating
180 ° swivel bracket
Weight 1.5 kg
Cable gland ISO16 (for cable of 4,5-10mm)
• Connections on spring loaded terminals
• Operating temperature............................................. - 20°C / + 55°C**
(at 85 % max. relative humidity, non-condensing)
• Storage temperature .................................................. - 20 °C / + 65 °C
(at 90 % relative humidity, non-condensing)
* if option 4-20 mA : minimum 22 V.
** - 10 °C / + 40 °C in trade use.
II3D Ex tc IIIC T85°C Dc
Panel mounting kit (option)
Electrical certification
• Electromagnetic Compatibility complying with the
requirements of Directive 2004/108/CE.
Cut out template
Max thickness : 6 mm
Download