Antibody/Cell Binding and Magnetic Transport in a Microfluidic Device A Thesis Presented in Partial Fulfillment of the Requirements for the Degree Master of Science in the Graduate School of The Ohio State University By Shauna Adams, B.S. Graduate Program in Mechanical Engineering The Ohio State University 2013 Master’s Examination Committee: A.T. Conlisk, Advisor Derek Hansford Shaurya Prakash Abstract The advancement of micro-total analysis systems is increasing the ability to perform multiple functions in one microfluidic device. These systems have several advantages in biomedical applications, including lower equipment and personnel costs, reduced power requirements, faster separations, and smaller sample and reagent volume requirements. Because of this, there has been a growing interest in the study of particle motion in microand nanochannels. The Automated Cell to Biomolecule Analysis (ACBA) device labels, sorts, and analyzes cancer cells in an enriched blood sample. Understanding transport methods and the behaviour of particles in microfluidic lab-on-a-chip or micro-total analysis systems is important to the advancement of system applications. Stochastic equations are used to characterize magnetic microbead motion, and the effect of the Stokes drag force on magnetic microbeads is analyzed. Stochastic differential equations specifically the Langevin and Fokker-Planck equations are used to characterize the probability density distribution of a microbead in a microfluidic channel. The results show that due to the size of the microbead these equations the binding probability of a magnetic microbead to a circulating tumor cell. However the results do indicate that under different conditions the probability equation can be used if the Peclet number, P e = 10 − 100. The capture area of a magnetic microbead with and without a cell attached to it is analyzed to see if the effect of the change of Stokes drag force on the magnetic microbead effects the magnetic capture area. It is concluded that the added drag from the attached ii circulating tumor cell does in fact affect the capture area of the magnetic microbead in the presence of a magnetic with a constant magnetic field. iii Acknowledgments First and foremost I will like to thank God, my spiritual source of encouragement, strength, and perseverance. I will also like to offer my sincere thanks to my advisor A. T. Conlisk for his support and criticism. I thank him for inviting me to join his research team and providing me with the opportunity to work on a project that will effect the lives of many people to come. His patience, encouragement, and willingness to explore new concepts has made this experience one of absolute growth. I will also like thank my thesis committee members, Professor Hansford and Professor Prakash for their feedback, suggestions, and time throughout these two years. Thank you to my team members, Cong Zhang , Harvey Zambrano, and Zhizi(Jessica) Peng in the Computational Nano and Microfluidics Laboratory, especially Cong for all of your help, insight, and willingness to explain a topic over and over until I understood it. I want to thank my family and friends for everything they have done and said over the past two years. I would especially like to thank my mom, Delores, and my sister, Imani for pushing and motivating me along this journey. I would also like to acknowledge my dad, Melvin Sr. who was always one of my greatest cheerleaders. Lastly, thank you Sandia National Laboratories for providing me with this opportunity to obtain my master’s degree on your time and money. Thank you to the Nanoscale Science and Engineering Center for the research resources. Thank you to GEM. iv Vita June 2007 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Phoebus High School May 2011 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B.S Mechanical Engineering North Carolina Agricultural and Technical State University Greensboro, NC May 2011 to Present . . . . . . . . . . . . . . . . . . . . . . . . . Sandia National Laboratories Technical Engineer May 2011 to Present . . . . . . . . . . . . . . . . . . . . . . . . . Sandia National Laboratories Masters Fellowship Program and GEM Fellow September 2011 to Present . . . . . . . . . . . . . . . . . . . Graduate Research Fellow Computational Nano and Microfluidics Laboratory Columbus, OH Publications Adams, Shauna, Zhang, Cong, Zambrano, Harvey, Conlisk, A.T. January 2013. Antibodyantigen Binding in a Flow-through Microfluidic Device, talk presented at the 51st AIAA Aerospace Sciences Conference, Grapevine, TX, Chapter DOI: 10.2514/6.2013-1114. Fields of Study Major Field: Mechanical Engineering Studies in Microfluidics: A. T. Conlisk v Table of Contents Page Abstract . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii Acknowledgments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv Vita . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v List of Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ix List of Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x 1. 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 2. Microfluidics . . . . . . . . . . . . . . . . . . . . . Electrolyte Solutions and the Electric Double Layer Poiseuille Flow . . . . . . . . . . . . . . . . . . . . Electrokinetic Phenomena . . . . . . . . . . . . . . Automated Cell to Biomolecule Analysis System . . Magnetically Induced Flow . . . . . . . . . . . . . Biophysics of the Biological Cell . . . . . . . . . . Present Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 3 . 6 . 8 . 9 . 13 . 15 . 17 Electroosmosis and Electrophoresis . . . . . . . . . . . . . . . . . . . . . . . 20 2.1 2.2 2.3 2.4 Introduction . . Electroosmosis Electrophoresis Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 21 24 27 3. Stochastic Phenomena . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 3.1 3.2 3.3 4. 4.4 28 31 32 35 36 38 39 . 43 . 44 Introduction . . . . . . . . . . . . . Motion of a Circulating Tumor Cell Binding Probability . . . . . . . . . 4.3.1 Results . . . . . . . . . . . Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 46 49 50 53 Magnetic Phenomena . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 5.1 5.2 5.3 5.4 5.5 6. . . . . . . . Binding Probability of Magnetically Labeled Antibody Binding and Circulating Cancer Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 4.1 4.2 4.3 5. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1.1 Markov Process . . . . . . . . . . . . . . . . . . . . . . . . . 3.1.2 Langevin Equation . . . . . . . . . . . . . . . . . . . . . . . . 3.1.3 Fokker-Planck Equation . . . . . . . . . . . . . . . . . . . . . 3.1.4 Derivation of the Fokker-Planck Equation . . . . . . . . . . . . Fokker-Planck Equation Solved for Binding Probability . . . . . . . . 3.2.1 The Wiener Process . . . . . . . . . . . . . . . . . . . . . . . 3.2.2 Non-Dimensional Transition Probability Density and Probability Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Magnetostatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.2.1 Definition of the Magnetic Field . . . . . . . . . . . . . . . . . . 5.2.2 Maxwell’s Equations . . . . . . . . . . . . . . . . . . . . . . . . 5.2.3 Electric and Magnetic Dipoles . . . . . . . . . . . . . . . . . . . 5.2.4 The Force Induced by a Magnetic Material on a Magnetic Particle Magnetically Induced Cell Transport . . . . . . . . . . . . . . . . . . . Characterization of the Magnetic Field . . . . . . . . . . . . . . . . . . 5.4.1 Trapping Efficiency of a Magnetic Bead . . . . . . . . . . . . . 5.4.2 Trapping Efficiency of a Magnetic Bead Bound to a Cell . . . . . 5.4.3 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 54 55 57 58 61 62 65 65 67 68 71 Summary and Future Work . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 vii Appendices 76 A. MATLAB Calculation of Important Constants . . . . . . . . . . . . . . . . . . 76 B. MATLAB Code for Transition Probability Density and Probability Density . . 79 C. MATLAB Code for Magnetic Bead Capture at Different Alpha and y Positions 83 Bibliography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86 viii List of Tables Table Page 1.1 Summary of important parameters in the mixing and labeling stage of the ACBA System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 1.2 Summary of breast cells, cancerous and non-cancerous . . . . . . . . . . . 17 3.1 Parameters for calculating . . . . . . . . . . . . . . . . . . . . . . . . . . 34 4.1 Test parameters for binding probability for ACBA system. . . . . . . . . . 50 ix List of Figures Figure 1.1 Page This lab-on-a-chip device was created to detect HIV infected cells from a blood sample. Research being conducted at UC Davis by Prof. Alexander Revzin. Photo from http://www.bme.ucdavis.edu/articles/2010/08/12/alexrevzin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 The electric double layer (EDL) is made up of the diffuse layer and the Stern layer. The Stern layer is comprised of only counter-ions of the wall charge. This layer is immobile.The diffuse layer, though majority counterions also contains some co-ions. These co-ions have the same charge of the wall. It is these mobile ions in the diffuse layer that move in the presence of an electric field. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Schematic of 2-D Poiseuille flow in a rectangular channel using equation (1.3). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.4 Overall schematic of the entire ACBA device. . . . . . . . . . . . . . . . . 9 1.5 CAD representation of entire ACBA System. Stage 1 is where magnetic beads coated in antibodies bind with circulating tumor cells through antibody/antigen binding as they are mixed together using chaotic mixing. Stage 2 captures and transports the magnetically labelled cells and magnetic beads using magnet disk arrays controlled by external electromagnets. Stage 3 uses an aqueous solution and oil to encapsulate individual cells for mRNA analysis. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 1.6 A magnetic field is created when an unpaired electron is orbiting the nucleus of an atom. When an electron is paired no magnetic field is created. The direction of spinning of the elctron determines the direction if the magnetic field. Photo from http://www.ndt-ed.org/EducationResources /HighSchool/Magnetism/reviewatom.htm . . . . . . . . . . . . . . . . . . 13 1.2 1.3 x 1.7 Size comparison chart of circulating tumor cells to red and white blood cells. Though actual cells range in size, CTC are normally smaller than the actual tumor cells. Picture recreated from (Kim et al., 2012) . . . . . . . . 16 1.8 Stained MCF-10 breast cells. Photos from Sung et al. (2009) and http://www. mcb.arizona.edu/azcc/confocal/examples.htm . . . . . . . . . . . . . . . . 17 1.9 Stained MCF-7 breast cancer cells. Photos from (Ehrhart et al., 2008) and http://www.rndsystems.com . . . . . . . . . . . . . . . . . . . . . . . . . 18 1.10 MDA-MB-231 breast cancer cells. Photos from atcc.org . . . . . . . . . . 18 2.1 Velocity profile of electroosmotic flow. A curved velocity profile exist within the electric double layer to account for the no-slip boundary condition and the change in potential. In the bulk fluid the fluid is charge neutral so the potential is equal to 0. This creates a vertical velocity profile within the bulk fluid. The bulk fluid maintains the velocity of the flow at the EDL boundary to satisfy the no slip condition. . . . . . . . . . . . . . . 23 2.2 Electrophoresis takes place when an electrically charged particle is placed in an electric field. The particle moves in the field. This can take place in a stationary fluid or relative to the fluid bulk velocity. . . . . . . . . . . . . . 24 3.1 Brownian motion of the antibody that collides with the fluid molecules. . . 30 3.2 Representation of the antibody position from x0 to x from time t0 to t. . . . 39 3.3 The transition probability, P, shows the probability density of particle being at position x at time t in an unbounded region relative to its initial position x = x0 at t = t0 , D =1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40 3.4 The probability density of an particle being at position x at time t in an unbounded region, D = 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 3.5 The transition probability shows the probability density of a particle being at position x at time t in a region with no flux boundary on one end given its initial position x0 = 0 at t0 = 0, D= 0.01. . . . . . . . . . . . . . . . . . 42 3.6 The probability density of a particle being at position x at time t in a region with no flux boundary at one end for D=0.01. . . . . . . . . . . . . . . . . 43 xi 4.1 Due to the size of the circulating tumor cell (yc ∼ h), the velocity of the cell can be determined by using the average velocity equation for the Poiseuille flow of the bulk fluid. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 4.2 The limits a and b defined in the probability equation to find the antibody/bead to CTC/antigen binding probability when a circulating tumor cell is at position, x. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 4.3 The probability density curves from equation (3.39) at different time instances for a 161 node simulation with the conditions described in Table 4.1 including a Peclet number, P e = 3.36 × 106 . . . . . . . . . . . . . . . 50 4.4 These plots show the dimensionless probability density, W ∗ , versus the dimensionless position, x∗ , at different dimensionless time, t∗ , values for different Peclet numbers. a) Pe = 10−2 , b) Pe = 10−1 , c) Pe = 1, d) Pe = 10, e) Pe = 102 , f) Pe = 103 , g) Pe = 104 , h) Pe = 105 . . . . . . . . . . . . . . . 52 5.1 Dipole diagram of magnetic disk in microfluidic system. . . . . . . . . . . 55 5.2 Geometry for calculating the electric potential field due to a dipole at the point P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59 5.3 Geometry for calculating the magnetic field due to a magnetic dipole at the point P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60 5.4 Electromagnets and solenoid control magnetic fields on magnetic disk. Picture from (Henighan et al., 2010) from NSEC Faculty Dr. Sooryakumar’s group here at OSU, who investigates the use of “magnetic tweezers” in microfluidic systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 5.5 Particle moving along magnetic disk array in a process known as “magnetic tweezers”. Picture from (Yellen et al., 2007) at Duke University. . . . . . . 64 5.6 The x ranges for a particle’s motion at y = 1 and at y = 0 when y0 = −1, γ = 0.33, β = −1, and α = −0.5. . . . . . . . . . . . . . . . . . . . . . . 69 5.7 The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −0.1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 xii 5.8 The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −0.5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70 5.9 The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 xiii Chapter 1: Introduction 1.1 Microfluidics The desire to study and design small scale fluid systems has increased over the last twenty-five years. With the advancement of Micro-Electro Mechanical Systems (MEMS) taking off in the early to mid 1980’s, the questions started to emerge on the ability and value of shrinking fluid systems. The application of small scale micro and nano fluid systems is vastly diverse and can be used over a scope of fields. In the medical community lab-on-a-chip technology allows lab diagnostic processes that took days to process, several hundreds of thousands of dollars, and many pieces of equipment now can be performed in a system barely visible to the human eye in minutes or hours and at a cost more accessible to a larger number of people. The design of microvalves, micropumps, micromixers, and other fluid mechanical systems are allowing engineers and scientist to create systems very similar to their macro-scale counterparts. Due to the complexity of many micro and nano-fluidic systems and their uses, molecular biology, chemistry, materials, and engineering knowledge are all being combined. From water purification systems, to artificial organs, and bio-chemical detection systems the possibilities and capabilities of the uses and applications of micro and 1 Figure 1.1: This lab-on-a-chip device was created to detect HIV infected cells from a blood sample. Research being conducted at UC Davis by Prof. Alexander Revzin. Photo from http://www.bme.ucdavis.edu/articles/2010/08/12/alex-revzin nano-scale fluid systems is nearly limitless (Nguyen & Wereley, 2002). This thesis will describe and analyze different particle transport processes within a lab-on-a-chip microfluidic device that is used to detect cancer from a person’s blood sample. A microfluidic system is defined as any system where a critical dimension is either on the micro scale or between a range of 1-100 µm. Macro systems most commonly are pressure driven whereas in micro systems electric and magnetic transport is most commonly used. As shown in Figure 1.1 as the critical dimensions decrease on the micro and nano scale the pressure drop needed to sustain a reasonable flow rate drastically increases, whereas the voltage needed to sustain a reasonable flow rate slightly increases as the critical dimensions decrease (Conlisk, 2013). 2 It should be mentioned here that the equations that govern the behaviour of fluids are still applicable on the micro scale and down to about a minimum dimension of six nanometers on the nano scale (Zhu et al., 2005). This means that energy, momentum, and mass are still conserved and the continuity and Navier-Stokes equations can be used. 1.2 Electrolyte Solutions and the Electric Double Layer Aqueous electrolyte solutions are the primary solutions used in many microfluidic devices. These solutions have the ability to conduct electricity thus making them ideal for electrically induced flow explained later in the chapter. Some examples of commonly used electrolyte solutions are sodium chloride (NaCl) and water and potassium chloride (KCl) and water. Both solutions are considered strong electrolytes and good conductors of electricity (Wright, 2007). An electrolyte solution is created when ionic salt is added to a solvent, in many cases water. Due to thermodynamic interactions of the molecules the salt dissociates in the solvent. This process creates positive and negative ions in the solvent; in the case of sodium chloride the sodium and the chloride disassociates into monovalent ions N a+ and Cl− . The valence, z, of the ion indicates how many electrons an ion has gained or lost and is normally identified by the whole number in front of the + or - sign. The valence in N a+ is positive one and the valence in Cl− is negative one (Wright, 2007). For example NaCl completely dissociates in water according to the reaction N aCl → N a+ + Cl− (1.1) In this reaction equation the sodium, N a, molecule has given its one electron in its outer most shell to chloride, Cl, thus denoting the + sign and chloride received an electron in its 3 outter most shell thus denoting the − sign. The sodium ion is now positively charged and the chloride ion is now negatively charged thus making the creation of an electric double layer possible. The electric double layer (EDL) develops in a channel when a dissociated electrolyte solution makes contact with the charged walls of the channel. A thin layer of oppositely charged ions called counter-ions, from the electrolyte solution, will attract to the wall surface. This layer of ions is called the Stern layer and is assumed to be immobile at all times. The diffuse layer, also known as the Gouy-Chapman diffuse layer, is the layer right above the Stern layer (Devasenathipathy & Santiago, 2005). This layer has both co-ions, ions that have the same charge as the channel wall, and counter-ions. The counter-ions outnumber the co-ions in this layer making the overall layer the charge of the counter-ions. It is this layer that becomes mobile in the presence of an electric field (Devasenathipathy & Santiago, 2005). The Stern layer and the diffuse layer make up the EDL. The remainder of the ions are located in the bulk fluid. Figure 1.2 describes the different layers. The primary length scale of the electric double layer is calculate by, λ= √ e RT 1 FI 2 (1.2) and λ is commonly known as the Debye length (Conlisk, 2013). In this equation e is J the electrical permittivity of the medium, R is the universal gas constant (8.314 mol·K ), T C is the temperature, F is Faraday’s constant (96, 485 mol ), and I is the ionic strength. The P 2 ionic strength is defined by, I = zi ci where, zi is the valence of species i and, ci is the i concentration of the electrolyte species at a specific location. This equation is only true when the species are monovalent and the concentrations are equal (Conlisk, 2013). 4 Figure 1.2: The electric double layer (EDL) is made up of the diffuse layer and the Stern layer. The Stern layer is comprised of only counter-ions of the wall charge. This layer is immobile.The diffuse layer, though majority counter-ions also contains some co-ions. These co-ions have the same charge of the wall. It is these mobile ions in the diffuse layer that move in the presence of an electric field. 5 1.3 Poiseuille Flow A common way to transport fluid in a fluidic system is to use a pressure difference across the system. Fluid travels from highest pressure to lowest pressure which makes transport method ideal. The movement of fluid using a pressure gradient is known as pressure driven flow or Poiseuille flow. In a system where fluid is viscous and incompressible the Navier Stokes equation can be used to characterize the fluid flow. The Navier Stokes equations in rectangular coordinates for a channel where W >> h can be written as, 2 ∂ux ∂ux ∂p ∂ ux ∂ 2 ux ∂ux + ux + vy =− +µ + ρ ∂t ∂x ∂y ∂x ∂x2 ∂y 2 2 ∂uy ∂ uy ∂ 2 uy ∂uy ∂uy ∂p ρ + uy + uy +µ + =− ∂t ∂x ∂y ∂y ∂x2 ∂y 2 (1.3) (1.4) Under the following assumptions equations (1.3) and (1.4) can be simplified a great deal. It is assumed that the fluid is only moving in the x direction as a function of y, the fluid flow is fully developed, and the fluid is in a steady state. The coordinate reference can be viewed in Figure 1.3. The Navier-Stokes equations reduce to 2 ∂ ux ∂p +µ 0=− ∂x ∂y 2 ∂p 0=− ∂y (1.5) (1.6) ∂p Since − ∂y = 0 the pressure drop is only dependent on x. The velocity can be calculated using the following boundary conditions, the no-slip boundary condition at the walls and due to symmetry of the system dux dy = 0 at y=0 and dp dx = − ∆p . With these conditions in L place the velocity becomes, ux = 1 ∆p 2 y − h2 2µ L 6 (1.7) Figure 1.3: Schematic of 2-D Poiseuille flow in a rectangular channel using equation (1.3). here µ is the fluid viscosity, ∆p = p2 − p1 , L is the length L − L0 , and h is half the height of the channel. This parabolic flow is known as Poiseuille flow and can be seen in Figure 1.3. The pressure driven volumetric flow rate through a rectangular channel can be calculated by taking the integral of the velocity in terms of the length and width of the channel, Zh ZW Q= ux dzdy −h 0 Q=− 2W h3 ∆p 3µ L (1.8) In the case where L stays the same and h and W decrease it can be seen in equation (1.8) that in order to keep the same flow rate the pressure will have to significantly increase. It is because of this fact that alternative methods to induce flow are at times considered for microfluidic and nanofluidic systems. 7 1.4 Electrokinetic Phenomena As mentioned in the previous section there are alternate methods to transport fluids in microfluidic systems. One other method of transporting fluid is through electrokinetic transport. Electrokinetic transport occurs when an electric field causes or is created by the motion of a fluid or particle in a fluidic system. There are three main electrokinetic phenomena that occur in microfluidics, they are electroosmosis, electrophoresis, and streaming potential. Sedimentation potential is also an electrokinetic phenomena, but it will not be discussed in detail in this thesis (Conlisk, 2013; Devasenathipathy & Santiago, 2005). The three electrokinetic phenomena may be defined as follows. • Electroosmosis occurs as a result of an electric field introduction to a system that causes the counter-ions in the diffuse layer of a microfluidic system to flow in a certain direction causing bulk motion of the fluid in the channel. • Streaming potential is caused when a pressure gradient creates an electric field in electrolyte solutions causing a potential along the charged wall when current is not introduced. • Electrophoresis occurs because of the particles suspended in the fluid in the channel. If an electric field is introduced to charged particles within a solution the particles move in a certain direction causing fluid flow. This can occur in solutions that are not electrolyte solutions These electrokinetic phenomena indicate that in the presence of an electric field, motion of either a charged particle or the bulk fluid electrolyte can occur. The steaming potential however creates an electric potential along the surface of the channel wall where there was 8 Figure 1.4: Overall schematic of the entire ACBA device. not one due to motion of a fluid. Electroosmosis and electrophoresis will be discussed in the next chapter of the text. 1.5 Automated Cell to Biomolecule Analysis System The Automated Cell to Biomolecule Analysis (ACBA) device is being developed by the Nanoscale Science and Engineering Center (NSEC), Center for the Affordable Nanoengieering of Polymeric Biomedical Devices (CANPBD) and is a micro total analysis system (µ-TSA) system designed to detect and analyze circulating tumor cells (CTCs) in a blood sample. The ACBA device is made up of four stages: labeling the cancer cells with a magnetically tagged antibody; separation of these cells bound to the antibody; transport and analysis. These stages can be seen in Figure 1.4. The fluid flow in the first stage of the ACBA device is a pressure driven flow controlled by manual dispensing from a syringe. One focus of this thesis is on the binding of the magnetically labeled antibodies to a circulating tumour cell. Table 1.1 shows the properties of the functions that make up the first stage of the ACBA system. Circulating tumor cells are cancer cells found in the blood stream indicating the presence of a tumor in the body. These cells are very rare as their presence is one CTC cell for every 109 blood cells in a whole blood sample (Nagrath et al., 2007). The ACBA system 9 Figure 1.5: CAD representation of entire ACBA System. Stage 1 is where magnetic beads coated in antibodies bind with circulating tumor cells through antibody/antigen binding as they are mixed together using chaotic mixing. Stage 2 captures and transports the magnetically labelled cells and magnetic beads using magnet disk arrays controlled by external electromagnets. Stage 3 uses an aqueous solution and oil to encapsulate individual cells for mRNA analysis. 10 MCF-7 Cell Diameter, d, µm 25 m2 −14 Diffusion Coefficient, D, s 1.67 × 10 Concentration, CC , cells 10 mL Surface Potential, σ, mV −20.32 Mass, mC , pg 208 Antibody Coated Magnetic Bead Diameter, d, µm 2.8 m2 −13 Diffusion Coefficient, D, s 1.49 × 10 beads Concentration, CA , mL 6 − 7 × 108 Mass, mA , pg 15 Bulk Fluid (PBS) Properties µL Flow Rate, Q, min 0.1 Viscosity, µ, P a · s 10−3 System Inlet Demensions length × width × height, µm 1370 × 200 × 30 Mixing Channel Dimensions length × width × height, µm 2500 × 400 × 30 Table 1.1: Summary of important parameters in the mixing and labeling stage of the ACBA System. 11 is designed to identify these cells using magnetically labeled antibodies that bind to the cells in the mixing process. The blood samples used in the system are enriched so that the possible CTC to blood ratio becomes 1:10,000. Chaotic mixing is used in the first stage of the ACBA system on the two laminar flow streams. Fluid flow folding mixes the solution containing the magnetically labelled antibodies with the enriched blood sample possibly containing the CTCs. Once the antibodies bind with the CTC cell they are detected using magnetic disk located in the second stage of the system. These magnetic disks are controlled by external electromagnets. The control of these magnetic fields allow control of the magnetic particle containing the CTC in two main dimensions, left/right and up/down. When these magnets are turned on they capture the particle in the induced magnetic field pulling the particle from the natural flow of the system. This process separates the CTCs from the remaining cells in the system (Chen et al., 2013). After the cells are captured they are transported along the array of magnetic disks to an area where they can be transported individually for analysis, this encompasses Stage III of the ACBA system. Using two-phase flow the CTCs are encapsulated for mRNA analysis. Micro-RNA are messenger molecules that send genetic information between DNA and ribosomes in a cell The last stage of the ACBA system uses electrical systems and biosensing technology to analyse the messenger RNA, mRNA, of the cell to correctly identify the cell based on its biological fingerprint. An actual diagram of the ACBA system can be seen in Figure 1.5. The first two stages of this system will be the main focus of this thesis. The fluidics, physics, and applied mathematics determining the behavior of the fluid and the particles within the fluid will the analysed and/or modelled in great detail in the following chapters. 12 Figure 1.6: A magnetic field is created when an unpaired electron is orbiting the nucleus of an atom. When an electron is paired no magnetic field is created. The direction of spinning of the elctron determines the direction if the magnetic field. Photo from http://www.ndted.org/EducationResources /HighSchool/Magnetism/reviewatom.htm 1.6 Magnetically Induced Flow The use of magnets and magnetic particles are another way the control the behavior of the fluid or a particle in the fluid of a microfluidic device. The ACBA system, defined in Section 1.5, uses both magnetic particles and magnets to control the behavior of selective particles in the device. Magnets in microfluidic systems serve several different functions. They can be used to induce a bulk fluid flow, or to capture and transport particles and biological materials within a bulk fluid (Pamme, 2006). Both of these occur when a magnetic material are in the presence of a magnetic field. The magnetic field is the area around a magnet where a magnetic force is present. The magnetic field is created by the spinning of unpaired electrons rotating around the nucleus of the atom as seen in Figure 1.6. The magnetic field is indicated by magnetic field lines that run from the two poles of the magnet. This field can span the length of an entire system or be contained and controlled within a small area. 13 There are two types of magnets. Those who produce their own magnetic fields and those that produce magnetic fields as a result of an electric current. The latter is known most commonly as an electromagnet. The former can be either a permanent or temporary magnet. Both types of magnets are used in microfluidic systems for a variety of functions. Magnets used to control magnetic material or magnetized fluids are utilized as pumps, switches, valves, mixers, trappers, and sorters in microfluidic systems as detailed by Pamme (Pamme, 2006). Magnets can be very beneficial in systems specialized for isolation processes and systems designed to use fluids with different electrical properties. The electrical properties such as the conductivity of a fluid are not affected by the presence of an magnetic field. Magnet use in microfluidic systems can be either active or passive. An active magnetic system is attached directly to the microfluidic device and all the equipment needed to produce a magnetic field is attached to the system. This system either has permanent magnets or electromagnets microfabricated into the design of the system. A passive magnet requires the use of external magnets to control the magnetic particles in the fluid system. Similar to electrically induced flow, magnetically induced flow takes place with the introduction of a magnetic field to a bulk fluid concentrated with nanoparticles. The motion of the bulk fluid caused by the direction of the magnetic fields is described as Hartmann flow and will be discussed later in this thesis. Magnetic particles have been used to sort, transport, and identify objects in a microfluidic system (Yellen et al., 2005). This method’s advantage over its counterparts of electrically and optically induced flow is that it does not harm cells and other living objects. Other methods have the ability to heat up the system causing the living objects and cells to become damaged or even die (Yellen et al., 2005). 14 1.7 Biophysics of the Biological Cell The ACBA device is specifically designed to detect and analyze cancer cells found in a blood sample. Cancer cells like many other cells are very complex. Cancer is the title given to over a 100 diseases where mutations in a cell causes a cell to divert from its usual proliferation and survival cycle (Alberts et al., 1998; Peng, 2011). The mutation causes the cell to no longer follows the life cycle pattern it was designed to follow and instead the gene that prompts cell production in an existing cell can become too active producing cells in excess; this gene is called an oncogene (Alberts et al., 1998). The regeneration process naturally established to replenish cells as other cells die is the catalyst for cancer cell growth. Cancer cells produce other cancer cells during cell division. The mutation found in the original cell is often carried over to its progeny or child cells. A tumor forms when the mutated cells cluster together to form a mass. If this mass is found in the part of the tissue where it is suppose to be found it is benign or non cancerous to the living being. These cells can spread to other parts of the body using the circulatory and lymphatic systems as their carrier. It is at this point that the cancer cells become circulating tumour cells (CTCs) (Mavroudis, 2010; Alberts et al., 1998). CTCs are the most dangerous of the mutated cells since they have the ability to invade tissue where they are not suppose to be found and continue to regenerate and form tumors. These cells prevent healthy cells around them from producing identical healthy cells, and they destroy neighboring tissue. A size comparison of CTCs to blood cells is in Figure 1.7. CTCs are normally smaller than the actual tumor cells (Kim et al., 2012). All cells have individual markers called antigens. Proteins called antibodies are produced by B cells, a class of white blood cells, and are designed to bind to antigen to identify 15 Figure 1.7: Size comparison chart of circulating tumor cells to red and white blood cells. Though actual cells range in size, CTC are normally smaller than the actual tumor cells. Picture recreated from (Kim et al., 2012) a cell for destruction when the immune system considers it a foreign object (Alberts et al., 1998). Each antigen has its own distinctive antibody. This means that an antibody will only bind to an antigen it is designed to bind with. For this reason antibodies are widely used in many microfluidic tagging or labelling applications. The antibodies used for the ACBA microfluidic device are designed to detect three types of human mammary epithelial cells. Each are a different type of breast cell line that is either cancerous or non-cancerous. They include the MCF-10a, MCF-7, and the MDA-MB-231 breast cell line and breast cancer cell lines respectively (Peng, 2011). The MCF-10 cell line are normal non-cancerous breast epithelial cells. They average in size between 20 30µm in diameter and have a surface potential of around -31.16 mV (Peng, 2011; Zhang et al., 2009). One of the most tested and utilized breast cancer cell lines is the MCF-7 cancer cells. They are spherical in shape, range in size between 10 - 50 µm in diameter and have a surface charge potential around -20.32 mV (Peng, 2011; Zhang et al., 2009). These 16 Figure 1.8: Stained MCF-10 breast cells. Photos from Sung et al. (2009) and http://www. mcb.arizona.edu/azcc/confocal/examples.htm cell characteristics are used for analysis throughout this thesis. The MDA-MB-231 breast cancer line is a progressive strains of cancer. These cells are rod-like in shape and have a surface potential between -24 to -31 mV (Peng, 2011; Trickler et al., 2008). Figures 1.8 1.10 provide visual representations of these different cells. Table 1.2: Summary of breast cells, cancerous and non-cancerous Cell Size(µm) Surface Potential (mV ) MCF - 10 20 − 30 (diameter) −31.16 MCF - 7 10 − 15 (diameter) −20.32 MDA-MB-231 10 × 70 − 90 (width × length) −24 to −31 1.8 Present Work This thesis will describe the different fluid and particle transport modes in microfluidic devices including pressure driven flow, electroosmosis and electophoresis. The binding behavior of antibody coated magnetic beads and circulating tumor cells will then be analyzed. 17 Figure 1.9: Stained MCF-7 breast cancer cells. Photos from (Ehrhart et al., 2008) and http://www.rndsystems.com Figure 1.10: MDA-MB-231 breast cancer cells. Photos from atcc.org 18 Lastly, the trapping ability of a magnetic bead and a magnetic bead with cell attached will be studied in detail. In chapter 2 two types of electrokinetic forms of transport will be looked at in detail. The velocity equations for electrophoretic and electroosmotic flow will be dervived, and examples of how these two phenomena are used in microfluidic applications are then discussed. In chapter 3 stochastic motion is analyzed in great detail. More specifically the Langevin and Fokker-Plank stochastic equations are derived. The Langevin equation describes the random motion of the magnetic bead, and the Fokker-Planck equation describes the transition probability density distribution of the magnetic bead along the length of the channel. A specific form of the Fokker-Plank equation known as the Wiener Process is then applied as a simplified model to describe the probability density of a simplified 1-D problem undergoing Brownian motion. In chapter 4 the probability density equation derived in the previous chapter will be used along with the known motion of the larger circulating tumor cell to predict the probability of a bead and cell binding at different times in the system. In chapter 5 an analysis of magnetic fields will be conducted and the capture area of the magnetic bead with and without a cell attached by an array of magnets will be looked at. The concept of “magnetic tweezer” will also be discussed. This thesis focuses on the use of stochastic differential equations to predict particle binding in a microfluidic system as well as the study of the capture efficiency of a magnetic bead with and without attached cells are both added knowledge in the field of microfluidics. 19 Chapter 2: Electroosmosis and Electrophoresis 2.1 Introduction While the focus of this thesis is on the ACBA system which uses pressure driven flow, electrokinetic phenomena are important for similar applications. As discussed in the previous chapter electrokinetic techniques are used in a variety of microfluidic systems. They are used to pump and mix fluids as well as separate particles and microorganisms in electrolyte and neutral solutions (Haeberle & Zengerle, 2007; Nguyen & Wereley, 2002; Mitra & Chakraborty, 2012; Conlisk, 2013). Electrokinetic techniques are ideal to use in microfluidic lab on a chip applications because they scale down well so their results are not greatly affected as a system is miniaturized (Conlisk, 2013). They are also fairly easy in integrate into micron-sized systems (Nguyen & Wereley, 2002). As noted in the previous chapter there are three main types of electrokinetic phenomena found in microfluidic devices electroosmosis, electrophoresis, and streaming potential (Nguyen & Wereley, 2002). In this chapter eletroosmosis and electrophoresis will be discussed in greater detail. The disadvantage to the use of electrokinetic techniques is that the electric fields either used or generated in the processes can negatively affect biomaterials in the case of lab-on-a-chip applications, especially if heat is produced in excess (Yellen et al., 2005). 20 2.2 Electroosmosis Electroosmosis is used to pump electrically charged fluids in several lab-on-a-chip applications (Conlisk, 2013). The ability to move electrolyte solutions using an electric field is advantageous with the limited space afforded on all lab on a chip platforms. As mentioned above, the creation of the electric double layer (EDL) when electrolyte solutions and naturally charged surfaces like silca and glass combine make it possible for electroosmotic flow to occur (Conlisk, 2013; Nguyen & Wereley, 2002). Electroosmotic flow is caused by movement of ions in the diffuse layer of the EDL in the presence of an electric field. Due to the electroneutrality of the bulk fluid the velocity of the bulk fluid is determined by at the boundary between the EDL and the bulk fliud. This means that to fully understand the velocity profile of electroosmotic flow the velocity profile has to be broken down into two sections: the velocity profile within the EDL and velocity profile outside the EDL. To calculate the velocity of a fluid in electrosomotic flow within a rectangular cross area when the width of the channel is significantly larger than the height of the channel some assumptions are made. First, the electric double layer is thin compared to the height of the total channel. Secondly, the zeta potentials are equal at the walls. We also assume the fluid flow is steady-state and fully developed and the velocity the velocity profile is one-dimensional and only a function of y in the x-direction. The velocity can be calculated from the Navier Stokes equations and Poisson equation, assuming a Boltzmann distribution within the EDL (Conlisk, 2013; Nguyen & Wereley, 2002; Kirby, 2010). The Navier Stokes equations are ρ ∂u + u · ∇u ∂t = −∇P + µ∇2 u + ρe E 21 (2.1) and the Poisson equation is e ∇2 φ = −ρe (2.2) By inserting Equation (2.2) into Equation (2.1) and applying the assumptions mentioned above the N-S equation becomes Ee ∇2 φ = µ∇2 u (2.3) Equation (2.3) can be solved by integrating both sides and plugging in the boundary conditions within the EDL, φ = ζ at y = 0 and du dy = dφ dy = 0 at y = ∞, ζ is the potential at the shear plane between the Stern and diffuse layer. The velocity of the fluid due to the electric field E within the EDL,is (Nguyen & Wereley, 2002) ueof EDL = e E(φ − ζ) µ (2.4) In the bulk fluid φ = 0 due to its neutrality, so to satisfy the no slip condition boundary condition the constant velocity of the bulk fluid has to be the velocity of the fluid at the boundary. This means that the bulk fluid velocity outside the EDL is ueof = − e Eζ µ (2.5) In the above equations u is the electroosmotic velocity of the fluid within the EDL or in the bulk fluid, ρE is the electric potential, E is the electric field, µ is the fluid viscosity, and φ is the potential. The velocity profile in electroosmotic flow is “plug” like (Devasenathipathy & Santiago, 2005). This means that the majority of the profile for this flow is straight as seen in Figure 2.1. This is ideal for many visualization applications due to uniformity and consistent distribution of particles throughout the system that is not obtainable with the parabolic profile produced in pressure driven flow (Weigl et al., 2003). 22 Figure 2.1: Velocity profile of electroosmotic flow. A curved velocity profile exist within the electric double layer to account for the no-slip boundary condition and the change in potential. In the bulk fluid the fluid is charge neutral so the potential is equal to 0. This creates a vertical velocity profile within the bulk fluid. The bulk fluid maintains the velocity of the flow at the EDL boundary to satisfy the no slip condition. Electroosmosis has several applications to lab-on-a-chip applications. One of the applications of electroosmotic flow is electroosmotic pumping. Electroosmotic pumping is controlled by the switching on and off of an electric field. Takamura et al. (2003) performed work on controlling pumping using low voltage portable sources and multiple sources in one device (Takamura et al., 2003). Many devices require multiple voltage sources to control the flow of an electrolyte solution in different parts of a microfluidic device. Due to the plug-like profile of electroosmotic flow Takamura et al. (2003) were able to control to motion of fluids in a uniform manner. With the use of the low voltage sources embedded within the system they were able to increase their pumping pressure by decreasing their channel size. 23 Figure 2.2: Electrophoresis takes place when an electrically charged particle is placed in an electric field. The particle moves in the field. This can take place in a stationary fluid or relative to the fluid bulk velocity. 2.3 Electrophoresis Electrophoresis is used in many lab on a chip applications (Haeberle & Zengerle, 2007; Nguyen & Wereley, 2002). Since the velocity of a particle is determined by both its size and charge strength, electrophoresis is ideal for separating different types of particles. It is used to analyse DNA, tag biomolecules, and manipulate proteins. Electrophoresis describes the motion of an electrically charged particle in the presence of an electrical field relative to the motion of the bulk fluid. The bulk solution surrounding the particle can be either electrically neutral, thus eliminating the creation of an electric double layer at the walls of the channel, or an electrolyte solution. In the case where the charged particle is in an electrolyte solution the particle will move relative to the bulk fluid (Conlisk, 2013). An electric double layer will develop around the surface of the particle, and the EDL on the charged particle induces the velocity of the particle in a microsystem (Nguyen & Wereley, 2002; Conlisk, 2013). A representation of this flow can be seen in Figure 2.2 24 In many cases electrophoresis is used in the analysis phase of many lab-on-a-chip devices. Because electrophoresis is the movement of a charged object relative to its surrounding in the presence if an electric field, its has the ability to control the movement of an object based on its electrical properties and it is primarily used in separation of bio-material for chemical and biochemical analysis (Ugaz & Christensen, 2007). There are four different types of electrophoretic methods that can be used in microfluidic devices. They include free solution electrophoresis, gel electrophoresis, isoelectric focusing, and micellar electrokinetic chromatography (Ugaz & Christensen, 2007). Free solution electrophoresis uses the difference in charge strength to separate material in a microfluidic device. The stronger the charge of the material the faster it will move through the system, thus causing a gradient of material based on charge strength. In gel electrophoresis the charged material travel through a gel like medium. This medium allows the charged objects to not only be separated by charge but also by size. In this case if two or more objects have similar charges but differ in size, the smaller object will move faster through the system thus causing a size and charge gradient of particles in the system. Isoelectric focusing uses a pH gradient to neutralize charged objects as they travel through the system. Once charged material reach a particular pH the material becomes charge neutral thus hindering its mobility through the system. Assuming the Reynolds number is smaller than one and that the Debye length is much much larger than the particle radius (λ a) the electrical force is equivalent to the Stokes drag equation qE = 6πµueph a 25 (2.6) In equation (2.6), q is the total charge on the particle, E is the electric field, µ is the fluid viscosity, ueph is the velocity of the particle due to electrophoresis, and a is the radius of the particle. Solving for the electrophoretic velocity ueph = qE 6πµa (2.7) The total charge on the sphere is (Conlisk, 2013) q = 4πe a 2 Z∞ d dr 2 dφ r2 2 dr dr (2.8) a 2 = 4πe a ζ 1 1 + λ a (2.9) This equation is derived by EDL calculations on the charged sphere. Inserting equation (2.8) into equation (2.7) we find ueph = E (4πe aζ) 6πµa (2.10) 2Ee ζ 3µ (2.11) which simplifies to ueph = This form of the electrophoretic velocity equation is known as the Debye - Huckel equation it can be used when λ a (Conlisk, 2013). In the case where the radius of the particle is much larger than the Debye length (a λ) the electrophoretic velocity is ueph = Ee ζ µ (2.12) this equation is known as the Helmholtz-Smouluchowski equation and is derived using the same principles to derive the velocity from electroosmotic flow equation (Nguyen & Wereley, 2002). This equation can best be applied to particles larger than 100 nm in size (Nguyen & Wereley, 2002) 26 An application of electrophoresis in a system include DNA analysis. Liu & Guttman (2004) use electrophoresis to analyze DNA on a microchip (Liu & Guttman, 2004). Compared to the conventional method of analyzing DNA which uses different types of systems and over several different steps, the entire DNA analysis process can be performed on one microchip. As a result the reaction time between the reactants is faster, less material and chemicals are used thus reducing cost, and results are obtained quicker. Electrophoresis allows for different size DNA stands to to separated for analysis (Liu & Guttman, 2004). 2.4 Summary In this chapter the two main electrokinetic phenomena, electroosmosis and electrophoresis, have been discussed in great detail. The velocity behavior for both phenomena was derived. Electrophoresis involves the velocity of a charged particle in a microfluidic system in the presences of an electric field. Electroosmosis is the movement of bulk fluid in the presence of an electric field. The velocity profile for this phenomena produces a vertical velocity in bulk fluid making it ideal to use in studies involving visual detection. The practical applications of both of these phenomena in existing technologies was then described. It was noted that electrophoresis and electroosmosis can occur in the same system and also be combined with other processes. 27 Chapter 3: Stochastic Phenomena 3.1 Introduction The instantaneous mechanical state of a system consisting of many particles described by classical mechanics requires only the specification of a set of positions and momenta of the particles. If the particles in the system are heavy enough compared to surrounding molecules the classical mechanical approach will provide an accurate description of the physical state of a system. Then the usual approach to describing the time evolution of this mechanical state of such many-body system is by use of a coupled set of the Newton’s equations, which describes the individual dynamics of all the particles in the system. This approach has led to the development of the widely used numerical technique called molecular dynamics. Molecular dynamics simulations have provided numerous insights into the behavior of fluidic systems at the nanoscale (Zambrano et al., 2009, 2012). However, in microfluidic systems, the Brownian motion of particles with size of tens of microns, is outside the reach of the molecular dynamics technique due to the typical long time and large spatial scales inherent in microfluidics systems. Moreover, the deterministic problem for more than a few particles exhibits non-deterministic, chaotic behavior (Conlisk et al., 1989). There are alternative methods to describe dynamics of particle motion which is in contrast with the deterministic approach. These are called the stochastic methods, which were 28 pioneered by Einstein (1956) and Langevin (1908), among many others. In this approach a many body system is treated by using the equations of motion describing the dynamics of only some selected particles moving in the presence of the other particles in the system which are now regarded as a background or bath whose detailed dynamics is not explicitly treated. In stochastic methods, the dynamic variables of interest in a many body system, such as the position and velocity of a particle in a fluid, are discussed in great detail and some other aspects of the problem are treated by theory of random processes. Methodologies of this type have been used to study the properties of fluidic systems ranging from atomic and molecular liquids to colloidal suspensions and macroscale fluidic systems. Brownian motion is an example of a stochastic process (Snook, 2007). In 1827, it was first described in detail by Robert Brown, a botanist, while observing pollen under a microscope (Coffey et al., 2004). Eighty years later, Einstein combined the Maxwell-Boltzmann distribution and the idea of random walk, a stochastic principle, to mathematically explain Brownian motion as a diffusion equation (Coffey et al., 2004). Brown’s discoveries were the early stages for what is now the Fokker-Planck equation, a partial differential equation that combines Einstein’s diffusion equation and a deterministic term to obtain the probability density function of a stochastic process over time (Coffey et al., 2004; McKane, 2009). A few years later, Langevin was the first to derive a stochastic differential equation to describe Brownian motion (Coffey et al., 2004). Called the Langevin equation, he used Newton’s Law as the foundation of his reasoning (Coffey et al., 2004). It is the principle of these equations that are the foundation for determining the motion behavior of the antigen in the ACBA system. The motion of the antibodies within the ACBA device is assumed to be driven by a random force, which is the net effect of interaction forces between the antibody and the 29 surrounding fluid molecules. The time evolution of the motion is described probabilistically with a time-dependent random variable (Figure 3.1). The time revolution of the random variable is the spatial position in one dimension x of the anti-body at different time instances. Furthermore, a Markov process is assumed to determine the antibody locations, meaning the particle position x, at time t, is determined by its previous position at x0 , at its previous time t0 , and does not depend on the positions earlier than time t0 , as long as the time interval ∆t = t − t0 is much smaller than the characteristic time of the process but larger than the time interval between two successive collisions of the anti-body with the surrounding molecules. This is the definition of a Markov process which will discussed in the next section (McKane, 2009). Figure 3.1: Brownian motion of the antibody that collides with the fluid molecules. 30 3.1.1 Markov Process A Markov process is a process in which the present state of a system is determined only by the state in the immediate past. For a Markov process, the probability density function(pdf), W , depends only on the previous state For a general ith state then Z W (xi+1 , ti+1 ) = P (xi , ti )P (xi+1 , ti+1 )dxi (3.1) where P is the transition probability. The transition probability is the probability that the antibody will be at location xi+1 at time ti+1 given that it was at xi at a previous time ti . Now for a Markov chain (McKane, 2009), Z P (xi+2 , ti+2 , xi , ti ) = P (xi+2 , ti+2 |xi+1 , ti+1 ) × P (xi+1 , ti+1 |xi , ti )dxi+1 (3.2) The variables W and P are both non-negative and if they satisfy (3.1) and (3.2) they define a Markov process. In many cases, numerical solutions can be used to model physical processes, as that time and the dependent variable computed is a discrete set of numbers. A discrete Markov process is called a Markov chain. In this case the integrals in equations (3.1) and (3.2) become sums. A stationary process is one in which the trasition probability depends only on t − t0 . An example of a Markov chain is a random walk in one-dimension. The simplest random walk is such that the “walker” must move at every time step. With “n” denoting the step to emphasize the discrete nature of the Markov chain, the transition probability is given by Pnn0 P q = 0 if n = n0 + 1 if n = n0 − 1 otherwise 31 with P + q = 1. If there are boundaries, then boundary conditions are required. If Pn describes the transition probability at a boundary then Pn = 1 defines an absorbing boundary and Pn = 0 describes a reflecting boundary- meaning the particle can not stay at the boundary. In general a Markov chain is defined by the equation P (n, t + 1) = X Pnn0 P (n0 , t) (3.3) n0 with P being the transition probability density. At large times assuming that the system tends to a stationary or steady state P (n) = X Pnn0 P (n0 ) (3.4) n0 3.1.2 Langevin Equation In general, particle motion or in this case antibody bead motion is governed by Newton’s Law ma = F and this is given by dX(t) d2 X(t) = Fext (t) + β ū(X(t), t) − + R(t) m dt2 dt where X(t) is the location of the particle at time t, and dX(t) dt (3.5) is the particle velocity up . β = 6πµa is the Stokes drag coefficient, µ is the viscosity, a is the radius of the particle, dX(t) and m is the particle mass. Here Fext (t) is the external body forces, β ū(X(t), t) − dt is the Stokes Drag force and R(t) is a random force. Here ū(X(t), t) is the local average velocity of the fluid (Peng, 2011). 1 ū(X(t), t) = V Z u(x, t)δd (x − X(t))dV (3.6) This equation including the random force is termed a Langevin equation; if R(t) = 0 the motion of the particle is deterministic. The variable R(t) is independent of x and varies 32 rapidly compared to the variation of x(t) (Coffey et al., 2004). Statistically the average of all the random forces equal zero, hR(t)i = 0, where hi denote average. This can also be shown as 1 hR(t)i = t − t0 Zt R(t)dt = 0 (3.7) t0 After initial molecular collisions the R(t) value becomes independent of its previous value and no correlations exist over time between the values. Using the delta-function to characterize this behavior (Barrat & Hansen, 2003; McKane, 2009) hR(t)R(t0 )i = 2Dδ(t − t0 ). where D is the diffusion coefficient, (3.8) kT β and δ is the Dirac Delta Function in a 2D system. + ∞ if x2 + y 2 = 0 δ(x, y) = (3.9) 0 if x2 + y 2 6= 0 and Z∞ Z∞ δ(x, y)dxdy = 1 (3.10) −∞ −∞ The Langevin equation assumes the time scale of the collisions between the particle and surrounding molecules is much shorter than the time scale for the velocity of the particle (Barrat & Hansen, 2003). A deterministic analysis of particle motion is only valid if the mass of the particle is large so that any oscillations due to variations in temperature are negligible. The thermal velocity of the particle is given by r kT vth = m and thus if the particle is “small” the thermal velocity is large and the deterministic approach is not valid. Here k is the Boltzmann constant, T is the thermal temperature of the fluid, and m is the mass of the particles. To insure that the mean energy of the particle is correct (i.e. E = 21 kT ), a random force is required. 33 The system of particles in the present case is that consisting of the water molecules, the ions in the PBS solution, the antibodies and the biological cell. The sizes of the antibody and the cell are provided in Table 1.5. If we choose the velocity scale U0 as the average bulk flow velocity in a channel, then the dimensionless velocity is given by ū∗ = ū(X,t) . U0 The dimensionless particle displace- ment in each of the coordinate directions is defined as X∗ = X/L where L is the channel length. The dimensionless time is t ∗ = t/t0 with scale t0 = L/U0 . Table 3.1: Parameters for calculating . m 1.5 × 10−17 kg U0 2.1 × 10−4 m s β 2.6 × 10−8 Ns m L 2400 × 10−6 m 2.2 × 10−4 We may also define the force scale βU0 = 6πµaU0 . Therefore all the forces now can be represented by a dimensionless force ratio in the form of F∗ = F . βU0 Equation (3.5) can dX(t) X(t) = ū(X(t), t) − + Fext + R(t) dt2 dt (3.11) be rewritten in a dimensionless form (the * is dropped) as 2d where 2 = mU0 . βL 2 The values are in Table 3.1. Typically the coefficient ∼ 10−4 and so the acceleration term is usually negligible. This means that the particle paths, apart from a very short initial time period are governed by a first order differential equation dX(t) = ū(X(t), t) + Fext + R(t) dt (3.12) which is subject to initial conditions on each particle. If there are N particles, there are 3N spatial variables corresponding to the positions of each particle in three-dimensional space. 34 All of these variables are random variables because of the presence of the random force R(t). 3.1.3 Fokker-Planck Equation The Fokker-Planck equation has been applied to several biological systems in literature. Plant et al. (1993) uses the Fokker-Planck equation to describe the probability of a certain number of bonds at a particular time on a multivalent liposome during dissociation (Plant et al., 1993). The Fokker-Planck equation is used to model the time-dependent bacteria population size probability distribution for bacteria when fluctuated doses of immunoglobulin G (IgG) is administered in an effort to study the effect of bacterial infections under varying IgG subsitution therapies(Figge, 2009). The general Fokker-Planck equation is an equation of motion for a transition probability function, p that is a function of a collection of random variables. Consider the Fokker-Planck equation in an N particle system; in general form, corresponding to the N variables x1 , ...xN the Fokker-Planck equation is given by " # N N 2 X X ∂ ∂ ∂p = − Ai (x) + Bij (x) P ∂t ∂xi ∂xi ∂xj i=1 i,j=1 (3.13) where x = [x1 , ..., xN ], is the vector of random variables, in this case, the three-dimensional location of the N antibodies, Ai is called the drift vector, and Bij is termed the diffusion tensor,which is non-negative, definite and symmetric, and P is the transition probability of the system. It is thus seen that the Fokker-Planck equation represents an equation of motion for the transition probability of stochastically fluctuating quantities. The Fokker-Planck equation can be derived from the Langevin equation and further details are available in (Risken, 1984) and (Schuss, 1980). It should be pointed out that equation (3.13) is entirely 35 equivalent to the system of ordinary differential equations (Risken, 1984) dxi = Ai (x) f or i = 1, ...N dt (3.14) where N is the total number of antibodies. Comparing this equation with equation 3.12 it is seen that for a single particle in a single spatial dimension, A = ū + Fext + R where ū is the average fluid velocity in the immediate neighborhood of an antibody. Consider the form of the Fokker-Planck equation for a single random variable, the position in one dimension, of a single particle (i.e. antibody), x. Then we write the equation for the transition probability density, P (x, t|x0 , t0 ), that a particle that starts at the point x0 and ends at the point x at time t later as ∂P ∂ 1 ∂2 = − (A(x, t)P ) + [(B(x, t)P ) ∂t ∂x 2 ∂x2 (3.15) Here A and B are defined by 1 A(x, t) = lim ∆t→0 ∆t Z 1 ∆t→0 ∆t Z ∞ 0 (x − x)P (x|x0 , ∆t)dx0 −∞ ∞ 0 (x − x)2 P (x|x0 , ∆t)dx0 B(x, t) = lim −∞ where the scalars A and B have replaced the drift vector and the diffusion tensor respectively. Equation (3.15) needs to be solved subject to the initial condition P (x|x0 , 0) = δ(x − x0 ) and with appropriate boundary conditions. 3.1.4 Derivation of the Fokker-Planck Equation The Fokker-Planck equation can be derived from the Chapman-Kolmagorov equation under certain conditions. This derivation comes from (McKane, 2009). A jump moment 36 for a given system is defined as Z Mj (x, t, ∆t) = (ζ − x)j P (ζ, t + ∆t|x, t) (3.16) Consider the Chapman-Kolmagorov equation (3.1) in the form Z W = P (x, t + ∆t|x0 , t)P (x0 , t)dx0 (3.17) This is the continuous version of the equation P (n3 , t3 |n1 , t1 ) = X P (n3 , t3 |n2 , t2 )P (n2 , t2 |n1 , t1 ) (3.18) n2 Now, the integrand of (3.17), with x0 = x − ∆x can be written as P (x, t + ∆t|x0 , t)P (x0 , t) = P ([x − ∆x] + ∆x, t + ∆t|x − ∆x, t)P (x − ∆x, t) (3.19) Recall that a Taylor series in one dimension is defined by ∞ X f n (x)∆xn ∆x3 ∆x2 + f 000 (x) +··· = (3.20) f (x + ∆x) = f (x) + f (x)∆x + f (x) 2 6 n! n=0 0 00 Thus equation(3.19) is equivalent to the Taylor series P ([x − ∆x] + ∆x, t + ∆t)P (x − ∆x, t) = ∞ X (−1)j j=0 j! ∆xj jj (P (x + ∆x, t + ∆t)P (x, t)) δxj (3.21) Integrating equation(3.17) over x0 = x − ∆x, ∞ X (−1)j ∂ j [Mj (x, t, ∆t)P (x, t)] P (x, t + ∆t) = j j! ∂x j=0 (3.22) P (ξ, t|x, t) = δ(ξ − x) (3.23) Now, so that lim Mj (x, t, ∆t) = 0 ∆t→0 37 for j≥1 (3.24) so that Mj (x, t, ∆t) = Dj (x, t)∆t + O(∆t2 ) (3.25) Thus P (x, t + ∆t) = ∞ X (−1)j ∂ j [Dj (x, t)∆tP (x, t)] j j! ∂x j=0 (3.26) The j = 0 term is just P (x, t) since Mj (x, t, ∆t) = 1 so that, ∞ X (−1)j ∂ j ∂P = [Dj (x, t)P (x, t)] j ∂t j! ∂x j=1 (3.27) Equation(3.27) is the Kramers-Moyal expansion. In many cases the jump moment may be neglected for j > 2 and the result of this truncation is the Fokker-Planck equation ∂ 1 ∂2 ∂P = − (A(x, t)P (x, t)) + (B(x, t)P (x, t)) ∂t ∂x 2 ∂x2 (3.28) The jump moments can be calculated by the following procedure. The stochastic variable, x is the position of the antibody and its average is defined by Z hx(t)i = xP (x, t|x0 , t0 )dx = x0 (3.29) In particular for any function f Z hf (x(t))i = f (x)P (f (x), t|x0 , t0 )dx (3.30) So that Mj (x, t, ∆t) = h(x(t + ∆t) − x(t))j i. 3.2 Fokker-Planck Equation Solved for Binding Probability The Fokker-Planck equation is used to determine the transition probability of an antibody position at a given time, given its previous position at previous time instance. Assume the antibody is at position x0 at time t0 , then the transition probability of the antibody at position x after a time interval at time t is governed by the one-dimensional Fokker-Planck 38 Figure 3.2: Representation of the antibody position from x0 to x from time t0 to t. equation (Sinaiski & Zaichik, 2008). If the time interval t−t0 = ∆t 1, A and B become independent of t. ∂ 1 ∂2 ∂P (x, t | x0 , t0 ) = − (A(x)P (x, t | x0 , t0 )) + (B(x)P (x, t | x0 , t0 )) (3.31) ∂t ∂x 2 ∂x2 3.2.1 The Wiener Process As a first approximation and to gain a basic understanding of the transition probability distribution the classical diffusion equation will be solved. Using the initial condition mentioned in the previous section, the boundary condition P → 0 as x → ±∞, and the conditions A = 0 and B = 1, equation (3.31) can be rewritten as ∂P 1 ∂ 2P = ∂t 2 ∂x2 (3.32) The Fokker-Planck equation under these conditions is known as the Wiener Process (Schuss, 1980). The physical meaning of A and B are clarified by comparing the one-dimensional Fokker-Plank equation (where A and B become scalars) with the molecular diffusion equation. Thus B 2 corresponds to the diffusion coefficient, D, and the drift A corresponds to the average velocity of antibody displacement. The transition probability, P , and the probability density, W , of this form of the equation is (Risken, 1984) (x−x0 )2 1 − P (x, t | x0 , t0 ) = p e 4πD(t−t0 ) 2 πD(t − t0 ) 39 (3.33) Figure 3.3: The transition probability, P, shows the probability density of particle being at position x at time t in an unbounded region relative to its initial position x = x0 at t = t0 , D =1. and Z+∞ W (x, t) = P (x, t | x0 , t0 )W0 dx0 (3.34) −∞ where W0 is the initial distribution of the particles in the system. Equations (3.33) and (3.34) explain the most basic one dimensional probability density distribution of a particle experiencing Brownian motion. Figure 3.3 shows the plot of the transition probability of a particle in an unbounded region. In this case, the initial condition on the transition probability is W0 = 1 2 within the region [−1, 1], and zero elsewhere.Also, D = 1. From the plot we can see that the position of the particle is a maximum in a region about x = 0 and near t = 0. This can be understood because the initial position of the antibody is x0 = 0 at t0 = 0. The distribution extends to x = ∞ and x = −∞, 40 Figure 3.4: The probability density of an particle being at position x at time t in an unbounded region, D = 1. which corresponds to an unbounded region. Figure 3.4 shows the probability density of the antibody after integrating over the space of all initial positions x0 . In the microfluidic system of interest, the initial condition is similar to the case for the unbounded problem; however the boundary conditions change to P → 0 as x → ∞ and dP dx = 0 at x = 0. This means that the probability density flux vanishes for x = 0 to take into account that the antibodies cannot move backwards in the system because of the fluid flow. With the application of the new boundary conditions the transition probability becomes 1 P (x, t | x0 , t0 ) = p 2 πD(t − t0 ) 41 (x−x )2 0 − 4πD(t−t e 0) (x+x )2 +e 0 − 4πD(t−t 0) (3.35) Figure 3.5: The transition probability shows the probability density of a particle being at position x at time t in a region with no flux boundary on one end given its initial position x0 = 0 at t0 = 0, D= 0.01. The integral of equation (3.35) defines the probability distribution for the new boundary conditions and is given by Z∞ P (x, t | x0 , t0 )W0 dx0 W (x, t) = (3.36) 0 = W0 2 r t 1+x −1 + x √ √ − erf erf 2Dπ 2 πDt 2 πDt where the initial condition on the transition probability is W0 = 0.1 within the region [0, 1], and zero elsewhere. Figure 3.5 shows the plot of the transition probability of the antibody in a region with no flux boundary at x = 0 for D = 0.01. From the plot we can see that the position of the antibody again shows a maximum in a region about x = 0. Figure 3.6 shows the probability density of the antibody being at position x at time t. Still in this case, the initial condition is similar to the Wiener process above. The transition probability and the probability density both increase with the implementation of the new boundary conditions. This makes sense 42 Figure 3.6: The probability density of a particle being at position x at time t in a region with no flux boundary at one end for D=0.01. because the particles can only move in one direction thus minimizing the spread of the particles in the system and increasing the number of particles in the distribution. 3.2.2 Non-Dimensional Transition Probability Density and Probability Density To apply equations (3.35) and (3.34) to the ACBA system the equations need to be non-dimensionalized. The non-dimensional parameters are, x∗ = x L t∗ = U t L P∗ = PL and the scaling parameters are L and U . L is the hydraulic diameter which is two times the height of the binding channel and U is the velocity of the particle located at the center of the channel. 43 The non-dimensional transition probability density becomes 1 P (x∗ , t∗ | x∗0 , t∗0 ) = q 2 π UDL (t∗ − t∗0 ) to account for the Peclet number, P e = LU , D √ ∗ ∗ P (x , t | x∗0 , t∗0 )L Pe = p 2 π(t∗ − t∗0 ) e − 2 L2 (x∗ −x∗ 0) 4π D (t∗ −t∗ 0) UL − +e 2 L2 (x∗ +x∗ 0) 4π L (t∗ −t∗ 0) UL ! (3.37) equation (3.37) can be rewritten as ! ∗ ∗ 2 ∗ ∗ 2 − e P e(x −x0 ) 4π(t∗ −t∗ 0) − +e P e(x +x0 ) 4π(t∗ −t∗ 0) (3.38) with the use of the Peclet number velocity is introduced into the equations. This is beneficial because now the equations account for velocity of the particle in the system. The non-dimensional probability density is √ π W ∗ (x∗ , t∗ ) = W0 2 s erf (1 + x∗ ) Pe 4π(t∗ − t∗0 ) s ! − erf (x − 1) !! Pe 4π(t∗ − t∗0 ) (3.39) It is derived by taking the integral in terms of x0 of the product of P and the W0 from 0 to some large value, a0 . 3.3 Summary In this chapter we have studied stochastic ordinary and partial differential equations. Two types of stochastic equations studied in this chapter are the Langevin equation and the Fokker-Planck equation. The Langevin equation identifies the randomness in a system, and the Fokker-Plank equation provides a transition probability distribution of a random particle in a system. One form of the Fokker-Planck equation is known as the Wiener Process. The transition probability was calculated for a distribution of particles under two conditions: in an unbounded region and in a region with no flux boundary at one end. The no-flux model describes the conditions in the microfluidic system. The transition probability is the probability that the antibody will be at location x at time t given that it was at x0 at a 44 previous time t0 . The probability density is the integral of the transition probability over all possible starting positions x0 for a given antibody. 45 Chapter 4: Binding Probability of Magnetically Labeled Antibody Binding and Circulating Cancer Cells 4.1 Introduction In order to bind with the antigens on a cell, the antibodies must reside within a location very close to the antigen. The probability of this event happening at a certain location at a certain time is equivalent to the probability of the antibody showing up in any area within the bead radius distance from the surface of the cell as they both move through the system. Suppose the cell/antigen is stationary relative to the antibody, then the random motion of the antibody solely determines the probability of binding. Using the probability of an antibody being within a certain range of the channel at the same time the cell is also in the location will provide a rough estimate on the binding probability. The actual binding between the antibodies and antigen involve biomolecular interactions at specific binding sites on the circulating tumor cell. This analysis looks at the probability that the antibody is within the location of the circulating tumor cell so that binding can occur. 4.2 Motion of a Circulating Tumor Cell Before using the probability density function to obtain the probability that a particle will be within a certain area at a certain time the movement of the circulating tumor cell 46 first has to be understood. In this system the CTCs follow a deterministic trajectory. Due to the size of the cell and its ability to take up the entire height of the microchannel, the position and velocity of the cell can be evaluated as the average of the bulk fluid velocity. It has been shown in previous work that the hindered effect caused by the microchannels walls are negligible in this case (Peng et al., 2012). The velocity of the fluid is given by the Poiseuille flow as seen in Figure 4.1 and can be derived using the Navier Stokes equations as u(y) = 1 dp 2 (y − h2 ) 2µ dx (4.1) so the average velocity of the cell is uavg uavg = 1 = 2yc Zyc udy (4.2) −yc 1 dp 1 (−h2 + yc2 ) 2µ dx 3 (4.3) If yc ∼ h then uavg = − 1 dp 2 h 3µ dx (4.4) and the position of the cell is xc = xc (0) + uavg t In these equations µ is the viscosity of the fluid, dp dx (4.5) is the pressure gradient, y is the position in the microchannel, 2h is the total height of the channel, and ±yc are the upper and lower bounds of the cell position in the vertical direction. 47 Figure 4.1: Due to the size of the circulating tumor cell (yc ∼ h), the velocity of the cell can be determined by using the average velocity equation for the Poiseuille flow of the bulk fluid. Figure 4.2: The limits a and b defined in the probability equation to find the antibody/bead to CTC/antigen binding probability when a circulating tumor cell is at position, x. 48 4.3 Binding Probability Using the information from equation (4.6) and the probability density calculated in equation (3.39) the probability of binding at a certain time can be calculated. When the center of the cell is at position x at t = t the probability density at that same t = t can be integrated from x − R − r to x − R. This integral results in the probability that the antibody bead will be within that specific range behind the cell at that time. The R and r mentioned above account for the radius (µm) of the cell and the antibody bead respectively. The pictorial representation of this point can be viewed in Figure 4.2. Analytically the binding probability is x−R ZL Pbinding@t∗ =t∗ = W ∗ (x∗ , t∗ )dx∗ (4.6) x−R−r L In this case, x−R−r L and x−R L are the dimensionless limits and W ∗ (x∗ , t∗ ) is the dimension- less probability density. The binding probability calculated assumes the actual binding will occur when the cell and the antibody bead make contact. This is a simplified assumption seeing as though binding will occur only at specific binding sites on the cell that may or may not be at that location. To accurately predict the probability of binding in the system the problem has to be solved in both the dimensional and non-dimensional realm. The dimensional time, t, is needed to solve equation (4.5) for the physical position of the circulating tumor cell in the microfluidic system. The position, x, and time, t, are then non-dimensionalized to x∗ and t∗ using scaling factors L and U = umax , the maximum velocity of the Poiseuille flow. These non-dimensional values are then used in equation (3.39). This binding probability is obtained by integrating equation (3.39)over the nondimensional boundaries described above. 49 12 t*=0.0186 t*=0.559 t*=0.994 10 W* 8 6 4 2 0 0 0.2 0.4 0.6 0.8 1 x* Figure 4.3: The probability density curves from equation (3.39) at different time instances for a 161 node simulation with the conditions described in Table 4.1 including a Peclet number, P e = 3.36 × 106 . 4.3.1 Results Using Mathematica and MATLAB, two analytical and mathematical software packages, and the dimensionless probability density the binding probability is calculated for specific time and position values using equation (4.6). The scenario described in Table 4.1 accounts for the first 2400 µm of the ACBA channel. Figure 4.3 shows the probability density curves for different time instances in the large Peclet number limit. Table 4.1: Test parameters for binding probability for ACBA system. L = 2400µm U = 208.3 µm s ab = 1.4µm P e = 3.36 × 106 0 0 W (x∗ , t∗ ) = 0.1 In this example binding occurs almost instantaneously for the given initial distribution. Since the initial condition is W0 = 0.1 for 0 ≤ x∗ ≤ 0.1 and W0 = 0 elsewhere. If 50 the initial condition for the probability density distribution was to decrease the probability density values for the larger time values will decrease making the area underneath the probability density curve equal to one as it should be. Since the Peclet number is so large the initial condition persists over the time scale presented. It was observed that as parameters such as the velocity, bead size, and hydraulic diameter changed the probability density plots changed. The probability density distribution increases as the velocity decrease, the particle size decrease to nano-sized particles, or the hydraulic diameter is decreased. This observation implies that this method of calculating probability might be applicable on a different model where the parameters mentioned above are considered. To find out under which model this method for calculating probability of binding might be best utilized, a study of different Peclet numbers was conducted and the results are discussed below. In Figure 4.4 a-h the Peclet number was increased by a factor 10. This was done to find out within what Peclet number range will the binding equation mentioned above work best. From the The probability density distribution at different dimensionless times was calculated for Pe numbers ranging from 10−2 to 105 . This was conducted to see at which Pe numbers is the probability density the probability of binding equation suitable to use. The Peclet number range for this probability method to best be utilized is between 10 - 100. Between this range the model is not solely controlled by either diffusion or the initial velocity of the particle. When the Peclet number is small diffusion controls the model, and when the Peclet number is large convection controls the motion of a particle and the diffusion coefficient has little influence on the system as in our experimental case. With a Peclet number between 10 - 100 the particle positions are dependent on both the diffusion and velocity and the plots can be used to calculate the binding probability. 51 Figure 4.4: These plots show the dimensionless probability density, W ∗ , versus the dimensionless position, x∗ , at different dimensionless time, t∗ , values for different Peclet numbers. a) Pe = 10−2 , b) Pe = 10−1 , c) Pe = 1, d) Pe = 10, e) Pe = 102 , f) Pe = 103 , g) Pe = 104 , h) Pe = 105 52 4.4 Summary The probability of binding equation was introduced as the integral of the probability density. Values from the deterministic equation for cell motion were non-dimensionalized and used a parameters in this equation. The probability of binding is lower as the Peclet number is decreased; this means that the diffusion has a significant effect, or the velocity is significantly slowed. There might be some application under these conditions. 53 Chapter 5: Magnetic Phenomena 5.1 Introduction In the ACBA system magnets play an important part in the total functionality of the system. The labelling and trapping of the magnetic particles in the system allow for encapsulation and analysis in later stages of the device. But, in order for this to occur it is important to understand how and why the particles are being trapped within the system. Several papers explain the behavior of magnetic particles in a microfluidic system and the way they react in the presence of a magnetic field, but to my knowledge no one has looked at the behavior of the magnetic particle once it is attached to the cell in this case and how does this additional object effect the capture efficiency of the magnetic particle in the system. In this chapter the different roles magnetic particles play in microfluidic systems will be addressed. A study will also be conducted to analyze how the diameter of the cell in the drag calculations effect the capture efficiency of the magnetic particle in the system under different magnetic strengths and volumetric flow rates. 5.2 Magnetostatics In the ACBA, the magnetically labeled cell will interact with the externally imposed magnetic field due to the magnetized disk. The situation is depicted on Figure 5.1. In using 54 Figure 5.1: Dipole diagram of magnetic disk in microfluidic system. the magnetic field to move the magnetized cell, we seek to determine the cell trajectory and velocity. In order to determine the trajectory and hence the speed of the cell we must determine the force on the cell due to the magnetic field. We assume that the external magnetic field is governed by the laws of magnetostatics. In this section we discuss the basics of magnetostatics, derive the force exerted by a suparamagnet and then calculate the magnetic field. Magnetic fields are generally viewed as originating as a result of orbital rotation of electrons (Kirby, 2010). 5.2.1 Definition of the Magnetic Field Recall that the Electric Field at any point is defined as the force acting on a single charge at that point, or E= N F q = 0 2 q 4πe r C (5.1) where r is the distance from the point charge. This equation is known as Coulomb’s Law. A magnetic field arises as a result of charges moving in a domain. In analogy with the 55 electric field, the magnitude of the magnetic induction field is given by FM Vq B∝ (5.2) and B is called the magnetic induction. The direction of B is such that the force is given by Fm = q(V × B) and is called the Lorentz force. Note that a force is exerted on a given charge only if the magnetic induction is in a direction perpendicular to the direction of motion of the charge and that V • FM = 0 indicating that the magnetic force does no work. Thus the general formula for the magnitude of the magnetic induction field is B= The unit of the magnetic induction is Fm V qsinθ W eber m2 where W eber = sometimes called the magnetic flux density and the unit W eber m2 N msec . Coul The quantity B is = T esla. Two other vectors may be defined in terms of the magnetic induction, the magnetic field intensity H and the magnetization density M which indicates the extent to which the magnetic medium is polarized. These vectors are defined as B = µ0 (H + M) (5.3) where µ0 is the magnetic permeability of a vacuum, and has the magnitude µ = 4π × mkg 10−7 Coul 2 = H m where H stands for Henry. Analogous to the electrical permeability, we may define the relative permeability as µr = µm . µ0 The magnetic permeability is entirely analogous to the electrical permittivity and is a transport property. Most materials have a magnetic permeability close to that of a vacuum. Thus magnetostatics is characterized by 56 the the three vectors B, H, M. Equation (5.3) means that a permanent magnet of magnetization M in an applied magnetic field H produces the magnetic induction field B. Another property of a magnetic material that measures the extent with which a body can be magnetized is called the magnetic susceptibility and is defined by M = χH (5.4) and the susceptibility is usually small for non-magnetic materials; on the other hand for highly magnetized materials such as permalloy, the susceptibility is very large. This means that B = µ0 (1 + χ)H = µm H 5.2.2 (5.5) Maxwell’s Equations At steady state the magnetic fields and electric field are independent with the magnetic field being proportional to the electric current. However, when the magnetic field is time-varying, the electric and magnetic fields are coupled and the resulting equations are Maxwell’s equations. Maxwell’s equations, which describe the flow of charge, the electric and magnetic fields in moving media are ∇ · B = 0 N o magnetic sources ∂B F araday 0 s Law ∇×E=− ∂t ∂D Ampere0 s Law ∇ × B = µm J + ∂t ∇ • D = ρe Gauss0 s Law (5.6) (5.7) (5.8) (5.9) Here, H is the intensity of the magnetic field; E, the electric field; D, the displacement field; B, the magnetic induction field; J, the current density; and ρe , the free charge density. 57 If the material is an isotropic, permeable dielectric, then D = e E, B = µm H (5.10) ∂E ∇ × B = µm J + e ∂t (5.11) and Ampere’s Law becomes where the magnetic permeability µm is related to the electrical permittivity by c = (µm e )−1/2 and c is the speed of light. The magnitude of magnetic induction fields that are measured can be on the order of 1T esla. Even when the electric field is time varying, the magnetic field term in Faraday’s Law is negligible; for example; an electric field of 1V over a 10µm channel will require a very short time scale of about 10−10 seconds for a magnetic field of B = 1T to balance Faraday’s Law. Similarly, the left side of Ampere’s Law dominates unless the electric field is very large. Thus in most cases of practical interest, the electric and magnetic fields can be decoupled. Note that since ∇•B=0 (5.12) there is no such thing as a magnetic monopole; all magnetic materials are made up of dipoles, having north and south poles as depicted on Figure 5.3. The magnetic moment of a material is defined in the next section. 5.2.3 Electric and Magnetic Dipoles An electric dipole is set up as a combination of a positive and negative charge and the expression for the potential is a linear superposition of each according to q φ= 4πe 58 1 1 − r1 r2 (5.13) Figure 5.2: Geometry for calculating the electric potential field due to a dipole at the point P. From Figure 5.2 using the Law of Cosines, r1 = r − d2 cosθ and r2 = r + d2 cosθ, so that φ= dcosθ q 4πe r2 − d42 cos2 θ (5.14) qdcosθ 4πe r2 (5.15) Now for d << r φ= The quantity p = qd is called the electric dipole moment and the unit of dipole moment is 1Debye = 1D = 3.336 × 10−30 Cm. There is an entirely analogous magnetic potential for a magnetic dipole whose potential may be calculated as follows. It is well known that the magnitude of the magnetic force 0 between two point poles of strength p and p in a vacuum is given by 0 pp Fd = 4πµ0 r2 59 (5.16) Figure 5.3: Geometry for calculating the magnetic field due to a magnetic dipole at the point P . 0 and the force acts along the line in the plane of the dipole. The convention is that if p is a unit pole pointing north, then the magnetic field is defined by H= pr 4πmu0 r3 The magnetic induction field is then B = µ0 H = p . 4πr2 (5.17) If the dipole is located at a surface of area A, then the strength of the dipole is given by p = Aρm where ρm is the surface density of the magnetic poles. Now consider the situation depicted on Figure 5.3 similar to the situation depicted on Figure 5.2. The magnetic field is written as the linear sum of the two magnetic fields as r1 r2 ρm A (5.18) − 3+ 3 H= 4πµ0 r1 r2 Using the law of cosines as with the electrical dipole and the binomial theorem for d << r we obtain ρm A H= 4πµ0 1 3d 1 3d − d−r 1 − cosθ + − d + r 1 + cosθ 2 2r 2 2r 60 (5.19) so that H= ρm Ad ˆ − d + 3cosθr̂ 4πµ0 r3 (5.20) where dˆ and r̂ are unit vectors. Now cosθ = d̂ • r̂ and ρm = µ0 M so that with V = Ad we have H= MV ˆ ˆ • r̂r̂ − d + 3 d 4πr3 (5.21) Since the dipole moment m = ρm Ad we can also write the result in terms of the magnetic moment as H= 1 m m • rr − 3 +3 4πµ0 r r5 (5.22) It is this dipole distribution that creates a magnetic force that acts on a magnetic particle in the channel. 5.2.4 The Force Induced by a Magnetic Material on a Magnetic Particle The induced motion of a magnetized particle due to a magnetic force called the Kelvin force, is termed magnetophoresis (Kirby, 2010). Consider a dipole oriented as on Figure 5.1 (Rosensweig, 1985). The force experienced by this volume element is thus −Hρm A + (H + dH)ρm A = dHρm A (5.23) The dH on the right side is the directional derivative in the direction denoted by d dH = d • ∇H = d (M • ∇)H M (5.24) Since ρm = µ0 M and V = Ad the force per volume is given by Fd = µ0 (M • ∇H) 61 (5.25) To find the total force, substitute equation (5.4) into equation (5.25) and multiply by the volume Fd = µ0 V χ(H • ∇H) (5.26) There has been some controversy in the literature about the nature of the magnetic field in equation (5.26). Recall that the electrical body force is given by FE = ρe E (5.27) where ρe is the volume charge density and E is the electric field, both quantities taken in the fluid. On the other hand, Petit et al. (2011) and Bakuzis et al. (2005) argue that the H in equation (5.26) should be given by that force associated with the magnetic dipole given by equation (5.22), and this approach is used here. 5.3 Magnetically Induced Cell Transport The goal of the ACBA device is to manipulate an object within a fluid flow through the introduction of magnetic particles. Magnetic particles are normally superparamagnetic, lacking magnetic memory, and can be manipulated only in the presence of a magnetic field. Outside of that field they behave as normal particles in a fluid flow (Pamme, 2006). Most magnetic particles used in microfluidic systems are a few nm to several µm in size. This is advantageous because biomolecules such as antibodies, antigen, DNA, and mRNA can be attached to the particles to be used in different applications. In the ACBA system magnetic particles called Dynabeads are 2.8 µm in diameter and have antibodies attached to them (Henighan et al., 2010). These antibodies are what causes the magnetic particle to attach to the antigens on the identified circulating tumour cells. Magnetic particles are used first to label then trap, transport, or separate biological material (Pamme, 2006). Permanent magnets or electromagnets either microfabricated inside 62 of the micofluidic device or placed outside of the microfluidic device are used to control the movement of the magnetic particles in the system. Permanent magnets produce their own magnetic field and have the ability to attract a magnetic particle without the use of an electrical current. An electromagnet, on the other hand, uses electrical currents to produce a magnetic field thus attracting the magnetic particle. The ACBA system uses external electromagnets to control the magnetic fields of an array of permalloy disk 5 µm in diameter and 40 nm in depth located on the internal silicone surface of the microfluidic device (Henighan et al., 2010). The orientation of the electromagnets located on the outside of the ACBA system control the magnetic fields of the permalloy disk. Electromagnets control the magnetic fields in the x, y, and z directions, Hx , Hy , Hz , on the permalloy disk creating a three dimensional magnetic field as seen in Figure 5.4. By controlling the maxima locations of the magnetic fields on the disk the magnetic particle can be moved from disk to disk. This method of trapping and transporting magnetic particles is called “magnetic tweezers”. A detailed analysis of the magnitude of the forces acting on a particle in the ACBA system was conducted in Peng (2011). It was shown that when analyzing a particle in a micrfluidic system in the presence of body forces the primary forces acting on the particle are Stokes drag force, electrostatic force, and magnetic forces. The secondary forces of the random force, the EDL force, van der Waals force, and the force due to gravity were negligible. Without the presence of an electric field the electostatic force is not present in this system. The forces remaining are the magnetic force and the Stokes drag force. The Stokes drag force accounts for the force the fluid exerts on the particle as it moves through the fluid. This force for a spherical object is calculated by FS = 6πµap u 63 (5.28) Figure 5.4: Electromagnets and solenoid control magnetic fields on magnetic disk. Picture from (Henighan et al., 2010) from NSEC Faculty Dr. Sooryakumar’s group here at OSU, who investigates the use of “magnetic tweezers” in microfluidic systems. Figure 5.5: Particle moving along magnetic disk array in a process known as “magnetic tweezers”. Picture from (Yellen et al., 2007) at Duke University. 64 In the equation above µ is the viscosity of the fluid, u is the relative velocity of the cell, u = uf − uc where f and c stand for the fluid and cell respectively. 5.4 Characterization of the Magnetic Field The magnetic field in the microchannel produced by the permalloy disks are characterized directly using the Landau-Lifshitz-Gilbert equation in micromagnetic simulation software (Neudecker et al., 2006; Liu et al., 2007; Ha et al., 2003). Sooryakumar’s research group uses this equation incorporated in a micromagnetic software called OOMMF to characterize their magnetic fields for the magnetic tweezers (Henighan et al., 2010). The magnetic fields in Sooryakumar’s group papers (Henighan et al., 2010; Chen et al., 2013) have four major inputs that effect the magnetic field generated by the micromagnetic software they include the magnetic strength in all three directions and the rotating frequency of the oscillating field. Other inputs include the material and magnetic properties of the disk and the microbead. 5.4.1 Trapping Efficiency of a Magnetic Bead In this section the binding efficiency of a magnetic particle in a constant magnetic field in a rectangular channel undergoing Poiseuille flow will be explained. From the previous section the force balance for the system assuming the acceleration is negligible is FS = −FM where we assume one-dimensional fully developed flow; on the other hand, the magnetic force is two-dimensional with components in both the x− and y− directions. In the x−direction, 6πµap (uf − up ) = −FM x 65 Solving for the velocity of the bead, ub is given by, ub = uf + FM x 6πµab (5.29) Using a process similar to that found in (Haverkort et al., 2009) the fluid velocity for this system is characterized by a Poiseuille flow where the velocity is only the x direction, uf can be described as uf = umax y2 1− 2 h (5.30) In this equation umax = Q( 4W3 h ), where Q is the volumetric flow rate and W is the width of the channel. Substituting equation(5.30) into equation(5.29) results in ub = umax y2 1− 2 h + FM 6πµab (5.31) The velocity of the particle can be broken down into its different coordinates so that a relationship between the motion of the particle in the x and y directions y2 dx FM x = umax 1 − 2 + dt h 6πµab dy Fmy = dt 6πµab By dividing dx dt by dy dt (5.32) (5.33) an equation is obtained that describes the change in position of the particle in the x direction with that of the y direction. dx umax 6πµab = dy FM y y2 FM x 1− 2 + h FM y (5.34) Actually the above equation is already in dimensionless form with both x and y assumed to be scaled on the channel height. Thus we write dx 1 = 1 − y2 + β dy α 66 (5.35) with α= FM y FS β= FM x FM y Integrating between the limits x to x0 and y to y0 , the following equation is obtained to describe the particle motion y03 y 3 − + β(y − y0 ) α(x − x0 ) = y − y0 + 3 3 (5.36) Equation (5.36) has a four-parameter family of solutions depinding on the values of (x0 , y0 , α, beta). We consider the computational domain to be −l ≤ x ≤ l and −1 ≤ y ≤ 1, where l = l∗ . h For example we can start with l = 4. In this formulation we consider the magnetic disk to be placed at y = −1 and −γ ≤ x ≤ γ. We look for curves for which the trajectories of the particles end up at a location on the disk. Actually the minimum distance that any particle can attain is y = −1 + ap . h In the actual case we know that the magnetic forces will vary with (x, y); however, here we assume that both forces are constant so that α and β are constant. This same method described above to calculate the capture efficiency of a just a magnetic bead will be used in the next section. However, instead of using ab , the bead radius, for the radius in the Stokes drag equation ac will be used to account for the cell radius. The reason for this change will be explained in the next section. 5.4.2 Trapping Efficiency of a Magnetic Bead Bound to a Cell The capture efficiency of a magnetic bead in a microfluidic system using different fluid transport methods is documented well in literature The capture efficiency of a magnetic bead in a microfluidic system using different fluid transport methods is documented well in literature (Furlani et al., 2007; Sinha et al., 2007; Haverkort et al., 2009). These papers however, fail to assess the capture efficiency of a magnetic bead attached to a larger object 67 in a fluidic system. In the case of the ACBA system the magnetic bead is attached to the circulating tumor cell. In this section the capture efficiency of magnetic bead will be assessed with the magnetic properties of the bead and the drag properties of the circulating tumor cell. Previous authors have either calculated the magnetic and drag force on the magnetic bead or treated the cell as a magnetic particle, but they have not combined the two for a analysis of a magnetic bead attached to a larger particle. Here the characteristics from the magnetic particle and the circulating tumor cell will be used to predict magnetic particle behavior under the influence of a magnetic field. The analysis for the magnetic force on the object will come from the parameters of the magnetic bead while the force due to drag will be calculated using parameters of the cell. The equations and method used to calculate the efficiency equation in the previous section will be the same in this situation. The main difference between this case and the previous case is that the particle used in the drag equation is the cell instead of the particle. This means that the Stokes drag force in α and β now becomes 6umax πµac . The equations (5.36) stays the same. 5.4.3 Results As mentioned in the previous section a MATLAB code was created to calculate the capture area of the magnetic bead using equation (5.36). Any bead located between the minimum and maximum x positions for a given y value will be captured. The edges of the magnetic disk or in 1-D, magnetic strip were used to find the initial x position of a bead that will allow it to intersect the specific γ and y0 values representing the borders of the disk for different values of α. In the results below the β = −1, γ = 0.33 to represent a 10 µm disk with the x orgin at the center of the disk, and y0 = −1 to represent the bottom of 68 1 Particle position y vs x at α = −0.5 Max. Particle Trajectory @ y=1 Min Particle Trajectory @ y=1 Max. Particle Trajectory @ y=0 Min Particle Trajectory @ y=0 0.8 0.6 0.4 y 0.2 0 −0.2 −0.4 −0.6 −0.8 −1 −5 −4 −3 −2 −1 0 1 x Figure 5.6: The x ranges for a particle’s motion at y = 1 and at y = 0 when y0 = −1, γ = 0.33, β = −1, and α = −0.5. the channel. The α and β values are negative to account for the negative magnetic force on the particle. The results are in Figures 5.6 - 5.9. The trajectory of a particle is the same for different y values however the x position range varies depending on the y value. This can be seen in Figure 5.6. As the initial y position of the magnet bead decreases the x position needed for capture increases. At a starting position at y = 1 the range of capture for a magnetic bead is between −4.33 and −5 and for a particle whose initial y position is y = 0 the x range of capture is −2 to −2.67. The distance between the x values do not change due to the similarity of the curve. Figures 5.7 through 5.9 show the x positions for bead capture increase as the α value increases. This makes sense because as the α value increases the magnetic force on the particle decreases so the particle has to be closer to the magnetic strip to be captured. The smaller the α values magnetic force has a greater influence on the particle motion than the 69 1 Particle position y vs x at α = −0.1 Max Particle Trajectory Min Particle Trajectory 0.8 0.6 0.4 y 0.2 0 −0.2 −0.4 −0.6 −0.8 −1 −16 −14 −12 −10 −8 −6 −4 −2 0 2 x Figure 5.7: The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −0.1. 1 Particle position y vs x at α = −0.5 Max Particle Trajectory Min Particle Trajectory 0.8 0.6 0.4 y 0.2 0 −0.2 −0.4 −0.6 −0.8 −1 −5 −4 −3 −2 −1 0 1 x Figure 5.8: The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −0.5. 70 1 Particle position y vs x at α = −5 Max Particle Trajectory Min Particle Trajectory 0.8 0.6 0.4 y 0.2 0 −0.2 −0.4 −0.6 −0.8 −1 −3 −2.5 −2 −1.5 −1 −0.5 0 0.5 x Figure 5.9: The x range for a particle’s motion at y = 1, y0 = −1, γ = 0.33, β = −1, and α = −5. stokes drag force thus it has the ability to trap the magnetic bead earlier. A large value for α could also mean that the Stokes drag force is very small. This is taken into account when looking at the drag caused by just the magnetic particle versus the drag caused by the circulating tumor cell that is attached to the bead. 5.5 Summary In this chapter the equation for magnetic force has been derived using the characteristics of a dipole magnet. How magnets are used in the ACBA device was described in detail. Magnetic force along with the Stokes drag force was then used to describe the behavior of a particle in a microfluidic channel in the presence of a permalloy magnetic disk. This disk had a constant magnetic field. The results were as expected whereas the capture efficiency 71 of the bead with and without the CTC increased as the volumetric flow rate decreased and as the magnetic strength increased. 72 Chapter 6: Summary and Future Work In this thesis the different types of fluid and particle transport in microfluidic systems have been discussed. These transport processes include pressure driven flow, electrokinetic flow, and magnetically induced flow. Electrokinetic applications to lab-on-a-chip technology is described and the use of magnetic materials to induce biological cell transport in microfluidic systems is investigated. In this thesis, the Langevin and Fokker-Plank stochastic equations are derived. The Langevin equation describes the random motion of a magnetic bead that contains an antibody that is designed to bind to a cancer cell. The Fokker-Planck equation describes the transition probability density distribution of the magnetic bead along the length of the channel. A specific form of the Fokker-Plank equation known as a Wiener Process is then used as a simplified model to describe the probability density of a simplified 1-D problem undergoing Brownian motion. Though not used in the ACBA device, electroosmosis and electrophoresis have many applications in the success of a many lab-on-a-chip devices and in the future it is possible that these transport phenomena may be employed in the ACBA. In Chapter 4 the probability density equation derived in Chapter 3 is used along with the known motion of the larger circulating tumorcell to predict the probability of a bead and cell binding in the system. In Chapter 5 an analysis of magnetic fields is performed 73 and results for the capture area of the magnetic bead with and without a cell attached by an array of magnets are producecd. The conclusions of this thesis are as follows, 1. Stochastic differential equations, more specifically the Langevin equation and the Fokker-Planck equation are useful in describing the random behavior of small particle in microfluidic systems. The Wiener Process, a form of the Fokker Planck equation, is an simplified equation used to characterize the behavior of a particle undergoing Brownian motion. 2. The dimensional and nondimensional transition probability densities of the Wiener Process were used to describe an microfluidic system that was not bound on either side with limits ∞ to −∞ and a microfluidic system that had a no flux boundary condition. The no flux condition modelled a system where flow only occurred in one direction. 3. The probability of binding can be calculated by taking the integral of the probability density from two different position values at a certain time. 4. For the model described in this thesis the binding probability obtained did not characterize the binding behavior of the particles. The ACBA system particles are too large to observe the stochastic behavior (Pe = 3.6 × 106 ). However the principles of this model may be used to characterize other models with smaller in value parameters than the ones used in this model. 5. The ACBA system might be better characterized using a deterministic model to describe magnetic bead behaviour. 74 6. The magnetic strength of a magnet varies according to the type of magnet a system. The magnetic field of a permalloy is actually characterized using the LandauLifshitz-Gilbert equation. To simplify the model the magnetic field of a magnetic block is used instead of the disk. 7. Magnetic force and particle size can be used to find that theoretical capture area of a magnetic particles in a microfliuidic device. 8. The attachment of a circulating tumor cell to a magnetic bead affects the capture area of the bead in the system. Several areas mentioned in this thesis can be studied in greater detail. In the future, 1. The Ornstein-Uhlenbeck process, a more advance model to describe Brownian motion may be used to better characterize the Brownian motion in the system for smaller particles. 2. A more advanced model the describe the magnetic particles in the ACBA system needs to be developed to include their deterministic trajectories. 3. The theoretical capture area of a magnetic particle attached and not attached to a circulating tumor cell can be compared to experimental data and a efficiency calculation can be derived. 75 Appendix A: MATLAB Calculation of Important Constants 76 clc;clear all %Stokes Einstein Equation (Diffusion Coefficient) %D=(kb*T)/(6*pi*u*r) %Variables Name Unit %kb Boltzmann constant J/K %T Temperature K %u Viscosity Pa*s %r Particle Radius m %************************************************* % kb=1.3806488e-23; T=284.15; u=0.001; r1=12.5e-6; %Radius of Cell r2=1.4e-6; %Radius of Antibody %Calculate the Diffusion Coefficient D1=(kb*T)/(6*pi*u*r1) D2=(kb*T)/(6*pi*u*r2) %************************************************* %% %Peclet Number %Pe=LU/D %Variables Name Unit %L Characteristic Length m %U Velocity m/s %D Diffusion Coefficient mˆ2/s %************************************************* L=2e-6; %height of the channel U=2.0833e-4; %Calculate the Peclet Number Pe1=L*U/D1 Pe2=L*U/D2 %************************************************* %% %Diffusion Time Over a Certain Distance %t=xˆ2/(2D) %Variables Name Unit %x Distance of Diffusion m %************************************************* x=25e-6; %Diffuse half the width of the channel %Calculate the Diffusion Time t1=xˆ2/(D1*2) t2=xˆ2/(D2*2) %************************************************* %% %Concentration and Molar Concentration of Antibodies and Antigen %C=n/V %C_m=C/NA %Variables Name Unit %n #of objects in Solution %V Volume of Solution L %NA Avogadro’s number molˆ-1 %************************************************* n1=15; %Number of Cells in Solution n2=10000; %Number of Antibody in Solution V=7.5e-3; NA=6.0221415e23; %Calculate Concentrations C1=n1/V; C_m1=C1/NA 77 C2=n2/V C_m2=C2/NA %************************************************* %% %Thermal Velocity %vth=sqrt((kbT)/(m)) %Variables Name Unit %kb Boltzmann constant J/K %T Temperature K %m mass kg %************************************************* m1=2.08e-13; %Cell mass m2=1.5e-14; %Antibody mass %Calculate Thermal Velocity vth1=sqrt((kb*T)/m1) vth2=sqrt((kb*T)/m2) %************************************************* %% %Epsilon Squared %e_2=mU/BL %Variables Name Unit %kb Boltzmann constant J/K %T Temperature K %m mass kg %************************************************* B2=(6*pi*u*r2) LL=2400e-6 %Calculate Epsilon_2 e_22=sqrt((m2*U)/(B2*LL)) 78 Appendix B: MATLAB Code for Transition Probability Density and Probability Density 79 clear all; close all; clc %Stokes Einstein Equation (Diffusion Coefficient) %D=(kb*T)/(6*pi*u*r) %Variables Name Unit %kb Boltzmann constant J/K %T Temperature K %u Viscosity Pa*s %r Particle Radius m %************************************************************************** kb=1.3806488e-23; T=284.15; u=0.001; r1=12.5e-6; %Radius of Cell r2=1.4e-6; %Radius of Antibody %Calculate the Diffusion Coefficient D1=(kb*T)/(6*pi*u*r1) D2=(kb*T)/(6*pi*u*r2) %************************************************************************** L=2400e-6; %Diffusion length of the channel U=2.0833e-4; %Calculate the Peclet Number Pe1=L*U/D1 Pe2=L*U/D2 %Divide nodes into equal spacing % dx=lengthx/(n-1); lengthx=1; n=161; dx=lengthx/(n-1); x=0:dx:lengthx; m=n; lengtht=1; dt=lengtht/(m-1); for i=1:m t(i)=(i-1)*dt; tpr(i)=t(i); end for k=1:m for i=1:n %Dimensional transition probability density P(i,k)=1/(2*sqrt(pi*t(k)*D2))*exp(-x(i)ˆ2/(D2*4*t(k))); %Dimensionaless transition probability density P2(i,k)=sqrt(Pe2)*(1/(2*sqrt(pi*t(k))))*exp(Pe2*(-x(i)ˆ2/(4*t(k)))); %Call P or P2 depending on which form of the equation is used P2(i,k)=2*P(i,k); end end % Integrate to get probability for given initial probability % W(xpr,t)=.5 for abs xpr<1 0 otherwise lengthint=0.1*lengthx; nint=n; dxint=lengthint/(nint-1); xint=0:dxint:lengthint; % evaluate transition prob over integration domain for k=1:m 80 for i=1:n for j=1:n probint(j)=1/(2*sqrt(pi*t(k)*D2))*exp(-(x(i)-xint(j))ˆ2/(4*t(k)*D2)); probint2(j)=sqrt(Pe2)*(1/(2*sqrt(pi*t(k))))*(exp(Pe2*(-(x(i)-xint(j))ˆ2/ (4*t(k))))+exp(Pe2*(-(x(i)+xint(j))ˆ2/(4*t(k))))); end prob(i,k)=sum(probint)*dxint ... - 1/2*probint(1)*dxint-1/2*probint(n)*dxint; prob(i,k) = prob(i,k)/lengthint; % prob(i,k)=trapz(xint,probint2); end end % ppp = zeros(n,m); for k=2:m for i=2:n dummy(i,k)=trapz(x(1:i),prob(1:i,k)); end end dummy(end,1)=1; for i=2:n for k=2:m ppp(i,k)=trapz(t(1:k),dummy(i,1:k)); end end for k=1:m dummy_total(k)=trapz(x,ppp(:,k)); end max(max(ppp)) %Transition Probability Density Plot figure(1) set(gca,’FontSize’,15) surf(t,x,P2); % h = colorbar; % set(h, ’ylim’, [0 1.2]) % caxis([0 1.45]) colorbar; xlabel ’t’; ylabel ’x’; %title ’Transtition Probability Density’; %Probability Density Plot figure(2) set(gca,’FontSize’,15) surf(t,x,prob); % h = colorbar; % set(h, ’ylim’, [0 1.2]) % caxis([0 1.45]) colorbar; xlabel ’t’; ylabel ’x’; %title ’Probability Density’; %Probability Plot figure(3) set(gca,’FontSize’,15) surf(t,x,dummy); % h = colorbar; % set(h, ’ylim’, [0 1.2]) 81 % caxis([0 1]) colorbar; xlabel ’t’; ylabel ’x’; %title ’Probability’; axis equal %Probability desnity(PD) curve for specific time nodes figure(4) set(gca,’FontSize’,15) %PD Curve at node 3 plot(x,prob(:,3),’color’,[1 0 0], ’LineWidth’,2); hold on %PD Curve at node 90 plot(x,prob(:,90),’color’,[0 1 0], ’LineWidth’,2); %PD Curve at node 160 plot(x,prob(:,160),’color’,[0 0 1], ’LineWidth’,2); hold off %Legend displays time equivalent to node position legend(’t*=0.0186’,... ’t*=0.559’,’t*=0.994’) xlabel(’x*’) ylabel(’W*’) %title(’Pe=100000’) grid on xlim([0 1]) 82 Appendix C: MATLAB Code for Magnetic Bead Capture at Different Alpha and y Positions 83 clc;clear all dalph=0.1; %alpha=[0:dalph:2]; beta=-1; h=15e-6 %height of the channel from the centerline ab = 1.4e-6; y = (-h+ab)/h; ymax = (h-ab)/h; xo= 0; x = 5e-6/h; u = 0.001; y0=-1; x0=.33; al=2; N=21; dy=-al/(N-1); yy=[1:dy:-1]; x02=-x0 alpha=-5; for i=1:N x(i)=x0+(yy(i)-y0+y0ˆ3/3-yy(i)ˆ3/3)/alpha+beta*(yy(i)-y0); end for i=1:N x2(i)=x02+(yy(i)-y0+y0ˆ3/3-yy(i)ˆ3/3)/alpha+beta*(yy(i)-y0); end alpha2=-0.1; for i=1:N x3(i)=x0+(yy(i)-y0+y0ˆ3/3-yy(i)ˆ3/3)/alpha2+beta*(yy(i)-y0); end for i=1:N x4(i)=x02+(yy(i)-y0+y0ˆ3/3-yy(i)ˆ3/3)/alpha2+beta*(yy(i)-y0); end % Plot the path figure(1) plot(x,yy,x2,yy,’Linewidth’, 3) x = x(1); y = 1; line(’XData’, [x x x(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); x = x2(1); y = 1; line(’XData’, [x x x2(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); xlabel(’$x$’,’Interpreter’,’LaTex’,’FontSize’,18); ylabel(’$y$’,’Interpreter’,’LaTex’,’FontSize’,18); title(’Particle position $y$ vs $x$ at $\alpha = -5$ ’,’Interpreter’,’LaTex’,’FontSize’,18); legend(’Max Particle Trajectory’,’Min Particel Trajectory’); figure(2) plot(x3,yy,x4,yy,’LineWidth’,3) x = x3(1); y = 1; line(’XData’, [x x x3(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); x = x4(1); y = 1; line(’XData’, [x x x4(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); xlabel(’$x$’,’Interpreter’,’LaTex’,’FontSize’,18); ylabel(’$y$’,’Interpreter’,’LaTex’,’FontSize’,18); 84 title(’Particle position $y$ vs $x$ at $\alpha = -0.1$ ’,’Interpreter’,’LaTex’,’FontSize’,18); legend(’Max Particle Trajectory’,’Min Particle Trajectory’); al=1; N=21; dy=-al/(N-1); y3=[0:dy:-1]; for i=1:N x5(i)=x0+(y3(i)-y0+y0ˆ3/3-y3(i)ˆ3/3)/alpha2+beta*(y3(i)-y0); end for i=1:N x6(i)=x02+(y3(i)-y0+y0ˆ3/3-y3(i)ˆ3/3)/alpha2+beta*(y3(i)-y0); end figure(3) plot(x3,yy,’g’,x4,yy,’c’,x5,y3,’b:’,x6,y3,’r:’,’LineWidth’, 3) % x = x5(1); % y = 0; % line(’XData’, [x x x5(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... % ’LineStyle’, ’-.’, ’Color’, ’k’); % x = x6(1); % y = 0; % line(’XData’, [x x x6(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... % ’LineStyle’, ’-.’, ’Color’, ’k’); xlabel(’$x$’,’Interpreter’,’LaTex’,’FontSize’,18); ylabel(’$y$’,’Interpreter’,’LaTex’,’FontSize’,18); title(’Particle position $y$ vs $x$ at $\alpha = -0.5$ ’,’Interpreter’, ’LaTex’,’FontSize’,18); legend(’Max. Particle Trajectory @ y=1’,’Min Particle Trajectory @ y=1’, ’Max. Particle Trajectory @ y=0’,’Min Particle Trajectory @ y=0’); x = x3(1); y = 1; line(’XData’, [x x x3(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); x = x4(1); y = 1; line(’XData’, [x x x4(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); x = x5(1); y = 0; line(’XData’, [x x x5(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); x = x6(1); y = 0; line(’XData’, [x x x6(1)], ’YData’, [-1 y y], ’LineWidth’, 3, ... ’LineStyle’, ’-.’, ’Color’, ’k’); 85 Bibliography A LBERTS , B., B RAY, D., J OHNSON , A., L EWIS , J., R AFF , M., ROBERTS , K. & WAL TER , P. 1998 Essential Cell Biology: An Intorduction to the Molecular Biology of the Cell. New York: Garland Publishing, Inc. BAKUZIS , A. F., C HEN , K., L UO , W. & Z HUANG , H. 2005 MAGNETIC BODY FORCE. International Journal of Modern Physics B 19 (07n09), 1205–1208. BARRAT, J. & H ANSEN , J. 2003 Basic Concepts for Simple and Complex Liquids. Cambridge, United Kingdom: Cambridge University Press. C HEN , A., B YVANK , T., C HANG , W.-J., B HARDE , A., V IEIRA , G., M ILLER , B. L., C HALMERS , J. J., BASHIR , R. & S OORYAKUMAR , R. 2013 On-chip magnetic separation and encapsulation of cells in droplets. Lab on a Chip 13 (6), 1172. C OFFEY, W., K ALMYKOV, Y. P. & WALDRON , J. 2004 The Langevin Equation, 2nd edn. River Edge, New Jersey: World Scientific Publishing Co. C ONLISK , A. T. 2013 Essentials of Micro and Nanofluidics with Applications to the Biological and Chemical Sciences. Cambridge University Press. C ONLISK , A. T., G UEZENNEC , Y. G. & E LLIOTT, G. S. 1989 Chaotic motion of an array of vortices above a flat wall. Physics of Fluids A: Fluid Dynamics 1 (4), 704–717. 86 D EVASENATHIPATHY, S. & S ANTIAGO , J. 2005 Electrokinetic flow diagnostics. In Microscale Diagnostic Techniques (ed. K. Breuer), pp. 113–154. New York: Springer Verlag. E HRHART, J.-C., B ENNETAU , B., R ENAUD , L., M ADRANGE , J.-P., T HOMAS , L., M ORISOT, J., B ROSSEAU , A., A LLANO , S., TAUC , P. & T RAN , P.-L. 2008 A new immunosensor for breast cancer cell detection using antibody-coated long alkylsilane self-assembled monolayers in a parallel plate flow chamber. Biosensors and Bioelectronics 24 (3), 467–474. E INSTEIN , A. 1956 Investigations on the Theory of Brownian Movement. New York, NY: Dover Publications. F IGGE , M. T. 2009 Optimization of immunoglobulin substitution therapy by a stochastic immune response model. PloS one 4 (5), e5685. F URLANI , E. P., S AHOO , Y., N G , K. C., W ORTMAN , J. C. & M ONK , T. E. 2007 A model for predicting magnetic particle capture in a microfluidic bioseparator. Biomedical Microdevices 9 (4), 451–463, PMID: 17516176. H A , J. K., H ERTEL , R. & K IRSCHNER , J. 2003 Micromagnetic study of magnetic configurations in submicron permalloy disks. Physical Review B 67 (22), 224432. H AEBERLE , S. & Z ENGERLE , R. 2007 Microfluidic platforms for lab-on-a-chip applications. Lab Chip 7 (9), 10941110. H AVERKORT, J. W., K ENJERES , S. & K LEIJN , C. R. 2009 Magnetic particle motion in a poiseuille flow. Physical review. E, Statistical, nonlinear, and soft matter physics 80 (1 Pt 2), 016302. 87 H ENIGHAN , T., C HEN , A., V IEIRA , G., H AUSER , A., YANG , F., C HALMERS , J. & S OORYAKUMAR , R. 2010 Manipulation of magnetically labeled and unlabeled cells with mobile magnetic traps. Biophysical Journal 98 (3), 412–417. K IM , M. S., S IM , T. S., K IM , Y. J., K IM , S. S., J EONG , H., PARK , J.-M., M OON , H.-S., K IM , S. I., G UREL , O., L EE , S. S., L EE , J.-G. & PARK , J. C. 2012 SSAMOA: a novel CTC isolation platform using selective size amplification (SSA) and a multi-obstacle architecture (MOA) filter. Lab on a chip 12 (16), 2874–2880. K IRBY, B. J. 2010 Micro- and Nanoscale Fluid Mechanics Transport in Microfluidic Devices. New York: Cambridge University Press. L ANGEVIN , P. 1908 The theory of brownian movement. C. R. Acad. Sci. (Paris) 146, 530– 533. L IU , S. & G UTTMAN , A. 2004 Electrophoresis microchips for DNA analysis. TrAC Trends in Analytical Chemistry 23 (6), 422–431. L IU , Y., G LIGA , S., H ERTEL , R. & S CHNEIDER , C. M. 2007 Current-induced magnetic vortex core switching in a permalloy nanodisk. Applied Physics Letters 91 (11), 112501112501. M AVROUDIS , D. 2010 Circulating cancer cells. Annals of oncology: official journal of the European Society for Medical Oncology / ESMO 21 Suppl 7, vii95–100, PMID: 20943650. M C K ANE , A. J. 2009 Stochastic processes. Encyclopedia of Complexity and System Science pp. 8766–8783. 88 M ITRA , S. & C HAKRABORTY, S. 2012 Microfluidics and Nanofluidics Handbook Chemistry, Physics, and Life Science Principles. Boca Raton, FL: CRC Press Taylor & Francis Group. NAGRATH , S., S EQUIST, L. V., M AHESWARAN , S., B ELL , D. W., I RIMIA , D., U LKUS , L., S MITH , M. R., K WAK , E. L., D IGUMARTHY, S., M UZIKANSKY, A., RYAN , P., BALIS , U. J., T OMPKINS , R. G., H ABER , D. A. & T ONER , M. 2007 Isolation of rare circulating tumour cells in cancer patients by microchip technology. Nature 450 (7173), 1235–1239. N EUDECKER , I., P ERZLMAIER , K., H OFFMANN , F., W OLTERSDORF, G., B UESS , M., W EISS , D. & BACK , C. H. 2006 Modal spectrum of permalloy disks excited by in-plane magnetic fields. Physical Review B 73 (13), 134426. N GUYEN , N.-T. & W ERELEY, S. T. 2002 Fundamentals and Applications of Microfluidics. Norwood, MA: Artech House, Inc. PAMME , N. 2006 Magnetism and microfluidics. Lab on a Chip 6 (1), 24. P ENG , Z. 2011 Modeling of particle and biological cell transport in microchannels. PhD thesis, The Ohio State University. P ENG , Z., Z AMBRANO , H. & C ONLISK , A. T. 2012 A theoretical study of biological cell/colloidal particle transport in microchannels. In AIAA paper 2012-88-910. P ETIT, M., K EDOUS -L EBOUC , A., AVENAS , Y., TAWK , M. & A RTEGA , E. 2011 Calculation and analysis of local magnetic forces in ferrofluids. PRZEGL\kAD ELEKTROTECHNICZNY (Electrical Review) p. 115119. 89 P LANT, A. L., G RAY, M. & H UBBARD , J. B. 1993 A fokker-planck description of multivalent interactions. Biophysical chemistry 48 (1), 7589. R ISKEN , H. 1984 The Fokker-Planck Equation. Berlin, Germany: Springer Verlag. ROSENSWEIG , R. E. 1985 Ferrohydrodynamics. Cambridge, UK: Cambridge University Press. S CHUSS , Z. 1980 Theory and Applications of Stochastic Differential Equations. New York: John Wiley & Sons Inc. S INAISKI , E. & Z AICHIK , L. 2008 Statistical Microhydrodynamics. Germany: WileyVCH. S INHA , A., G ANGULY, R., D E , A. K. & P URI , I. K. 2007 Single magnetic particle dynamics in a microchannel. Physics of Fluids 19 (11), 117102–117102–5. S NOOK , I. 2007 The Langevin and Generalised Langevin Approach to the Dynamics of Atomic, Polymeric and Colloidal Systems. Oxford, United Kingdom: Elsevier. S UNG , Y. M., X U , X., S UN , J., M UELLER , D., S ENTISSI , K., J OHNSON , P., U RBACH , E., S EILLIER -M OISEIWITSCH , F., J OHNSON , M. D. & M UELLER , S. C. 2009 Tumor suppressor function of syk in human MCF10A in vitro and normal mouse mammary epithelium in vivo. PLoS ONE 4 (10), e7445. TAKAMURA , Y., O NODA , H., I NOKUCHI , H., A DACHI , S., O KI , A. & H ORIIKE , Y. 2003 Low-voltage electroosmosis pump for stand-alone microfluidics devices. Electrophoresis 24 (1-2), 185–192. 90 T RICKLER , W. J., NAGVEKAR , A. A. & DASH , A. K. 2008 A novel nanoparticle formulation for sustained paclitaxel delivery. AAPS PharmSciTech 9 (2), 486–493. U GAZ , V. M. & C HRISTENSEN , J. L. 2007 Electrophoresis in microfluidic systems. In Microfluidic Technologies for Miniaturized Analysis Systems (ed. S. Hardt & F. Schnfeld), pp. 393–438. Springer US. W EIGL , B. H., BARDELL , R. L. & C ABRERA , C. R. 2003 Lab-on-a-chip for drug development. Advanced Drug Delivery Reviews 55 (3), 349–377. W RIGHT, M. 2007 An Introduction to Aqueous Electrolyte Solutions. West Sussex, England: John Wiley & Sons Inc. Y ELLEN , B. B., E RB , R. M., S ON , H. S., H EWLIN , J R ., R., S HANG , H. & L EE , G. U. 2007 Traveling wave magnetophoresis for high resolution chip based separations. Lab on a Chip 7 (12), 1681. Y ELLEN , B. B., H OVORKA , O. & F RIEDMAN , G. 2005 Arranging matter by magnetic nanoparticle assemblers. Proceedings of the National Academy of Sciences of the United States of America 102 (25), 8860–8864. Z AMBRANO , H. A., P INTI , M., C ONLISK , A. T. & P RAKASH , S. 2012 Electrokinetic transport in a waterchloride nanofilm in contact with a silica surface with discontinuous charged patches. Microfluidics and Nanofluidics 13 (5), 735–747. Z AMBRANO , H. A., WALTHER , J. H., KOUMOUTSAKOS , P. & S BALZARINI , I. F. 2009 Thermophoretic motion of water nanodroplets confined inside carbon nanotubes. Nano Letters 9 (1), 66–71. 91 Z HANG , Y., YANG , M., PARK , J.-H., S INGELYN , J., M A , H., S AILOR , M. J., RU OSLAHTI , E., O ZKAN , M. & O ZKAN , C. 2009 A surface-charge study on cellular- uptake behavior of f3-peptide-conjugated iron oxide nanoparticles. Small (Weinheim an der Bergstrasse, Germany) 5 (17), 1990–1996. Z HU , W., S INGER , S. J., Z HENG , Z. & C ONLISK , A. T. 2005 Electro-osmotic flow of a model electrolyte. Phys. Rev. E 71, 041501. 92