Handheld On Site Tester Operator Manual 1 All Content Copyright © WiMAC Ltd Project No : 76 Document Title : Handheld Testers Operating Instructions Document No : 0076 / 08 Document Revision :6 Author : J.Marples / R. Adelsberg Date : 15/12/2010 Modification Record Revision Rev 1 Rev 2 Rev 3 Rev 4 Rev 5 Rev 6 Author Date John Marples John Marples John Marples Richard Adelsberg John Marples Richard Adelsberg 14/08/2009 27/10/2009 05/05/2010 23/11/2010 02/12/2010 15/12/2010 Definitions Conduit Node : Otherwise known as a Branch Node or Collector 2 All Content Copyright © WiMAC Ltd Contents 1.0 Introduction. 2.0 Getting Started. 3.0 Operational Modes. 4.0 Operational Mode Descriptions. 5.0 Node Detection Mode. 6.0 Automatic Single Node Testing. 7.0 Automatic Nema Node Testing. 8.0 Manual Single Node Operation Configuration. 9.0 Manual Nema Node Testing. 10.0 Dual Automatic Operation Configuration. 11.0 Automatic Dual Node Testing 12.0 Conduit Node Search Option. 13.0 Termination of Tests in the Manual Mode 3 All Content Copyright © WiMAC Ltd Handheld Tester Operating Instructions 1.0 Introduction The Handheld Tester is a self contained battery operated instrument capable of retrieving data from and simple testing of WIMAC enabled ballasts. It is intended as an aid to installers and as a simple diagnostic tool. The following operating instructions detail how the Handheld Tester should be used to enable data retrieval and simple tests to be carried out on a WiMAC compatible Node system. Note Wherever possible pictures have been used to aid in the clarity of the information provided. On opening the box please ensure that you have all of the items which you ordered and none are damaged. Handheld Tester part number: WMSNIFF-868A Contents: 1 2 3 4 A Handheld Tester. A set of rechargeable batteries A battery charger. A set of operating instructions. If any of the items are missing or damaged then please contact either: LeafNut Customer Support Tel: +44 (0) 113 383 1000 then choose option 5. Email: support@harvardeng.com Vizion Customer Support Tel: +44 (0) 1422 317 337 Email: enquiries@zodionltd.eu.com Note Power should NOT be applied to the column under test until the Handheld Tester has been configured for the required mode of operation and is waiting for the Nodes test message. The test message from the Node is only transmitted for a limited time during power up. If the Tester is not ready then this signal could be missed. This could result in a false fail message being displayed, or the message being missed altogether. 4 All Content Copyright © WiMAC Ltd Note WiMAC Nodes with software version of less than 25, will only reply with their serial number, Node type and the software version. The Nodes with these earlier versions of software will terminate communications with the Handheld Tester after they have transmitted the above. This is normal and does not signify a problem with either the lamp, WiMAC Node or the Handheld Tester. In response to this the Handheld Tester may display a communication lost message (see Fig. below). 2.0 Getting Started Please remove the charger from its packaging and connect it to a suitable supply, insert the batteries into the charger ensuring they are the correct way round. Power on the charger and wait until the batteries are fully charged this is indicated by a green light on the charger. Now remove the Handheld Tester from its packaging and install 2 AA cells into the battery holder located in the compartment on the rear of the product. This can be done as follows:a) Remove the securing screw and save b) Remove the cover battery cover c) Carefully extract the battery holder from the unit d) Install two AA cells the correct way round into the battery holder e) Carefully refit the battery holder into the compartment 5 All Content Copyright © WiMAC Ltd f) Replace the securing screw g) The unit is ready for operation On power up the unit will display the following splash screen. Note In the bottom right hand corner of the WiMAC logo will appear the operating frequency of the handheld tester. This is dependant on the legal requirements of the country where the unit is to be operated. Please ensure that this matches your countries legal requirements. Example: 868 = 868 MHz for EU. 6 All Content Copyright © WiMAC Ltd 3.0 Operational Modes The unit is capable of operating in the following different modes 1 Node detection. 2 Automatic single Node test. 3 Manual single Node test. 4 Dual Node automatic test mode. 5 Conduit detection mode. Screen Display The operator MUST select which mode is required by using the (center) key to cycle between options and then using the (right) key to select the option. Once selected that mode of operation will be maintained until the instrument is power cycled, this enables consecutive columns to be tested without the need to constantly reconfigure the instrument. Note If no activity is detected by the Handheld Tester for 2 minutes then the unit will automatically power down to prolong battery life. Note Power should NOT be applied to the column under test until the Handheld Tester has been configured for the required mode of operation and is waiting for the Nodes test message. The test message from the Node is only transmitted for a limited time during power up. If the Tester is not ready then this signal could be missed. This could result in incorrect failure messages being displayed, or the test message being missed altogether. 7 All Content Copyright © WiMAC Ltd After Configuration of the instrument into the required mode of operation, the unit will wait to receive a valid test message from the WiMAC enabled Node. The actual tests are dependant upon the type of Node detected. Note If after 2 minutes the Handheld unit has not received a valid message then the unit will power down. If in automatic or manual single Node of operation the unit will undertake the following tests in the order shown. 4.0 Operational Mode descriptions 4.1 Node detection mode 4.1.1 This mode of operation is intended as a simple quick check of a power cycled WiMAC enabled node. WiMAC Nodes with software version of less than 25, will only reply with their serial number, Node type and the software version 4.2 Automatic single node Test Mode 4.2.1 Initialise the communications 4.2.2 Test ballast communications 4.2.3 Request ballast serial number and identification 4.2.4 Test turning the lamp OFF 4.2.5 Test turning the lamp ON 4.2.6 Request the lamp voltage 4.2.7 Test the signal strength WiMAC Nodes with software version of less than 25, will only reply with their serial number, Node type and the software version 4.3 Manual single node Test Mode 4.3.1 Initialise the communications 4.3.2 Test ballast communications 4.3.3 Request ballast serial number and identification 4.3.4 Test turning the lamp OFF 8 All Content Copyright © WiMAC Ltd 4.3.5 Test turning the lamp ON 4.3.6 Request the lamp voltage 4.3.7 Test the signal strength 5.0 Node Detection Mode 5.1 In this mode of operation the handheld unit will simply display any WiMAC enabled node data it receives. 5.2 It will not attempt any communications with the WiMAC enabled node. Screen Display 6.0 Automatic Single Node Test Mode In order to operate in this mode the following steps should be followed: 6.1 Press the power button on the front of the unit 6.2 The unit will beep and vibrate and the display will show the splash screen below. 9 All Content Copyright © WiMAC Ltd 6.3 After 5 seconds unit will display the test options menu 6.4 Using the keys detailed earlier select the Automatic menu option. 6.5 The unit will now enter automatic single node test mode of operation 6.6 The instrument will now display the original splash screen with the addition of a letter “A” in the top right hand corner and vertical incrementing bar graph in the bottom right hand corner showing that the unit is working correctly 6.7 The Instrument is now in automatic single Node Test mode. 6.8 The Tester is now waiting for a valid transmission from a WiMAC enabled Node. 6.9 Automatic WiMAC Node Testing 6.9.1 On reception of a valid message from a leaf Node the unit will again beep, vibrate and the back light will illuminate. 10 All Content Copyright © WiMAC Ltd 6.9.2 The display will show the following information 6.9.3 The top line is the Serial Number of the unit detected 6.9.4 The next line is Node type followed by the software revision number of the unit. 6.9.5 The unit will again beep, vibrate, the back light will stay illuminated and the display will show the additional message “Initialising” 6.9.6 During this phase the unit is initialising communications between itself and the detected Node. 6.9.7 Having initialised the communications the unit will again beep, vibrate and display the appropriate message dependent upon Node type detected. (See 9.0 for Nema Node). 11 All Content Copyright © WiMAC Ltd 6.9.8 If a WiMAC Node has been detected by the Handheld Tester then the ballast communications test will be initiated and the following screen will be displayed: 6.9.9 If communications with the ballast is established and all is ok then the Unit will display the “Ballast Comms OK” message. Note This test can take a while dependant upon the ballast type, so please be patient. 6.9.10 If the there is no response from the Node then two further attempts are made. 12 All Content Copyright © WiMAC Ltd 6.9.11 If there is still no response from the WiMAC Node then the Tester will display the following message screen. 6.9.12 This could signify a problem with the ballast or the wiring between the ballast and the WiMAC Node, so all further tests will be aborted. 6.9.13 If all is ok then the Handheld Tester will send a request for the ballasts serial number and type Identification. The following request message will be displayed. 6.9.14 On receiving a valid response the Handheld Tester will display the ballast serial number and Identification number as shown below. 13 All Content Copyright © WiMAC Ltd 6.9.15 If no response is received then a further 2 attempts are undertaken before the test is aborted. 6.9.16 On receiving a valid message the unit will automatically issue a “Lamp OFF Test” command to the Node, informing the operator as follows. Fig. 12 6.9.17 On receipt of a response the Handheld Tester will display the results of the test if ok. 14 All Content Copyright © WiMAC Ltd 6.9.18 If an invalid or No response is received then the unit will make 2 more attempts before reporting the failure 6.9.19 If all is ok then the unit will vibrate again and beep and progress to the next test which is turning on the lamp. 6.9.20 If the lamp turn on tests are OK then the unit will inform the user with the following message on the display. 6.9.21 If the lamp turn on tests, are NOT OK the unit will attempt up to three times to turn the lamp ON. If it is not possible then “Strike Failure” will be displayed on each attempt (Fig. 17). 15 All Content Copyright © WiMAC Ltd Fig. 17 Note: The above problem can be caused by lamps which have previously been running and are hot. This is a characteristic of the Lamp and is not a failure of the Node. 6.9.22 As before if an invalid response or no reply is received the unit will make two more attempts after which it will display the following: If you suspect this to be the case then allow the lamp to cool and repeat the test. 6.9.23 The unit will then exit the test procedure as there is a problem with the lamp. 6.9.24 If the above test was valid then the unit will proceed with requesting the current lamp voltage. This will be indicated on the display see below. 6.9.25 If a valid response is received then the current lamp voltage will be displayed. 16 All Content Copyright © WiMAC Ltd 6.9.26 If an invalid message or no response is received then the unit will make a further 2 attempts before displaying the following message. 6.9.27 If a valid response is received then the unit will proceed to the signal strength Test and will display the following message: 6.9.28 If the test returns a valid response then the unit will display the following: 17 All Content Copyright © WiMAC Ltd 6.9.29 If the unit returns an invalid response or none at all then the unit will indicate this as follows. 6.9.30 If the message is ok then this completes the automatic testing of the column. 7.0 Automatic Nema Node Testing. 18 All Content Copyright © WiMAC Ltd 7.1 Once detected the Tester will initialise communications between itself and the Nema Node which will be indicated on the display. 7.2 On successful initialisation the unit will initiate a Lamp OFF test. 7.3 On successful completion of the test the unit will display the following message: 7.4 The Tester will now initiate a Lamp on test as shown below 19 All Content Copyright © WiMAC Ltd 7.5 If the test completes successfully then the following message is displayed. 7.6 On completing the above successfully the Tester will request the power data from the Nema Node. 7.7 If the response is valid then the Unit will display the following additional data Supply Voltage : 296 (Fig. 32) Real Power : 263 Status bits : 0000000010000000 20 All Content Copyright © WiMAC Ltd Fig. 32 7.8 8.0 If all of the above tests have been completed successfully then the unit will display the following and return to waiting for the next Node to test. Manual Single Node Operation Configuration In order to operate in this mode the following steps should be followed: 8.1 Press the power button on the front of the unit. 8.2 The unit will beep and vibrate and the display will show the splash screen 8.3 After 5 seconds the unit will beep and vibrate. 21 All Content Copyright © WiMAC Ltd 8.4 The display will show the test mode selection screen 8.5 The operator will now need to select manual operation using the same key as before. 8.6 This is achieved by pressing the Manual (center) key, the cursor will scroll round. 8.7 Set the cursor to the Manual Selection and press the key the instrument will beep and vibrate again. 8.8 The instrument will then display the original splash screen as above with the addition of a letter “M” in the top right hand corner and a vertical incrementing bar graph in the bottom right hand corner showing that the unit is working correctly. 8.9 The Tester is now in Manual single Node test mode. 8.10 It is now waiting for a valid transmission from a WiMAC enabled Node. 22 All Content Copyright © WiMAC Ltd 8.11 Manual WiMAC Node Testing Note In Manual mode of operation, all of the above appropriate tests are undertaken. This mode only allows the operator the option of delaying the start of the next test, to allow time to record the retrieved data. It does not allow tests to be included or excluded. 8.11.1 On reception of a valid message from a WiMAC Node the unit will again beep, vibrate and the back light will illuminate. 8.11.2 The display will show the following information (Fig. 37) 8.11.3 The top line is the serial number of the unit detected. 8.11.4 The next line is mode type followed by the software revision number of the unit 8.11.5 The unit will beep again, vibrate, the back light will stay illuminated and the display will show the additional message “Initialising”, During this phase the unit is initialising communications between itself and the detected Node. 8.11.6 Having initialised the communications the unit will beep again, vibrate and display the appropriate message dependent upon Node type detected. ( see 9.0 for Nema Node ) 23 All Content Copyright © WiMAC Ltd 8.11.7 If a leaf Node has been detected by the Handheld Tester then the ballast communications test will be initiated and the following screen will be displayed. 8.12 If communication with the ballast is established and all is ok then the unit will display the “Ballast Comms OK” (Fig. 40) message. Note This test can take a while dependant upon the ballast type, so please be patient. 8.13 If there is no response from the Node then two further attempts are made. 8.14 If there is still no response from the WiMAC Node then the Tester will display the following message screen (Fig. 41): 24 All Content Copyright © WiMAC Ltd 8.15 This could signify a problem with the ballast or the wiring between the ballast and the Node, so all further tests will be aborted. 8.16 If the response is valid the Handheld Tester will now request operator input before moving on to the next test. 8.17 Once the manual key has been pressed the Handheld Tester will send a request for the ballast’s serial number and type identification. The following request message will be displayed. 8.18 On receiving a valid response the Handheld Tester will display the ballast serial number and identification number as shown below (Fig.44). 25 All Content Copyright © WiMAC Ltd 8.19 If no response is received then a further two attempts are undertaken before the test is aborted. 8.20 If the response is valid the Handheld Tester will now request operator input before moving on to the next test. 8.21 On receiving a valid key press the unit will issue a Lamp OFF test command to the Node, informing the operator as follows. 26 All Content Copyright © WiMAC Ltd 8.22 On receipt of a valid response the Handheld Tester will display the results of the test if ok. 8.23 If an invalid or no response is received then the unit will make another two attempts before reporting the failure and aborting further tests. 8.24 If all is ok then the unit will vibrate and beep again then request input from the operator before progressing onto the next test which is turning on the lamp. 27 All Content Copyright © WiMAC Ltd 8.25 On a valid key press the unit will initiate the lamp turn on test. 8.26 If the lamp turn on tests is OK then the unit will inform the user with the following message on the display: 8.27 The unit will attempt to turn the lamp on up to three times. If it is not possible then the following message will be displayed on each attempt. Strike Failure 28 All Content Copyright © WiMAC Ltd 8.28 As before, if an invalid response or no reply is received or the lamp fails to strike then after three attempts the unit will display the following. 8.29 If the above test was valid the unit will again wait for valid operator input before proceeding with requesting the current lamp voltage. This will be indicated on the display as seen below. 8.30 If a manual key press is detected then the unit will display the following: 29 All Content Copyright © WiMAC Ltd 8.31 If a valid response is received then the current lamp voltage will be displayed. 8.32 If an invalid message or no response is received then the unit will make a further two attempts before displaying the following message. 8.33 If a valid response is received then the unit will display the following message; awaiting user input before proceeding. 30 All Content Copyright © WiMAC Ltd 8.34 The unit will send the signal strength test request to the Node displaying the following message. 8.35 If the test returns a valid response then the unit will display the following. 8.36 If the unit returns an invalid response or none at all then the unit will indicate this as follows. 31 All Content Copyright © WiMAC Ltd 8.37 If the message is ok then this completes the automatic testing of the column. 32 All Content Copyright © WiMAC Ltd 9.0 Manual Nema Node Testing. 9.1 Once detected the Tester will initialise communications between itself and the Nema Node. 9.2 On successful initialisation the unit will initiate a Lamp OFF test. 9.3 On successful completion of the test the unit will display the following message. 33 All Content Copyright © WiMAC Ltd 9.4 The unit will signal that user input is required by displaying the following: 9.5 The Tester will now initiate a Lamp on test as shown below. 9.6 If the test completes successfully then the following message is displayed: 34 All Content Copyright © WiMAC Ltd 9.7 On correct completion of the above test the unit will request user input. 9.8 On correct user input the Tester will request the power data from the Nema Node. 9.9 If the response is valid then the unit will display the following additional data Supply Voltage Real Power Status bits : 323 : 314 : 0000000010000000 35 All Content Copyright © WiMAC Ltd 9.10 On correct completion of the above test the unit will request user input. 9.11 If all of the above tests have been completed successfully then the unit will display the following and return to waiting for the next Node to test. 36 All Content Copyright © WiMAC Ltd 10.0 Dual Automatic Operation Configuration Dual In DUAL mode of operation ONLY THE TOP TWO TESTS are undertaken. This mode will be expanded in future updates to the product 10.1 In order to operate in this mode the following steps should be followed. 10.2 Press the power button on the front of the unit. 10.3 The unit will beep and vibrate and the display will show the splash screen (Fig. 75) below. 10.4 After 5 seconds the unit will beep and vibrate. 10.5 The display will show the manual or dual mode selection screen (Fig. 76) 10.6 The operator will now need to select dual operation. 10.7 This is achieved by pressing and holding the dual key while the menu is being displayed. 10.8 After 5 seconds the instrument will beep and vibrate again. 37 All Content Copyright © WiMAC Ltd 10.9 The instrument will then display the original splash screen in Fig. 77 with the addition of a letter “D” in the top right hand corner and a vertical incrementing bar graph in the bottom right hand corner showing that the unit is working correctly. 10.10 The Tester is now in dual automatic operating mode and awaiting a valid transmission from WiMAC enabled Nodes. 11.0 Automatic Dual Node Testing 11.1 On receipt of a valid message the display will momentarily show the following. 38 All Content Copyright © WiMAC Ltd 11.2 If a second Node is detected then the display will again momentarily show the following. 11.3 If the unit detects a second Node then it will show the following 11.4 It will initialise communications between the unit and the Tester and the detected Nodes. 39 All Content Copyright © WiMAC Ltd 11.5 After initialising the communications the test ballast command will be issued to each Node in turn and the display will show each stage. 11.6 If the ballast tests on each Node completes ok then the following will be displayed. 40 All Content Copyright © WiMAC Ltd 12.0 11.7 If no problems are detected then the unit will turn off the back light and return to waiting for valid Node transmissions. 11.8 If problems were detected with the ballast tests then a failure message for the appropriate Node will be displayed. Conduit Node Search Option 12.1 This search option allows installers to determine which Conduit Nodes can be detected by the Node. 12.2 This display consists of up to fourteen Conduit serial Node numbers followed by a bar graph of their RSSI Level. 12.3 The bar graph has a peak reading at the bottom which is held for up to 30 seconds and above this is displayed the latest reading. 12.4 Both the peak and current reading will be refreshed every time the Handheld receives a new reading from that particular Conduit. 41 All Content Copyright © WiMAC Ltd 13.0 12.5 The current reading is decayed away automatically by the unit. If a new reading is not received within 30 seconds then the peak is also cleared. 12.6 The menu has the following options: 12.7 The current selection is indicated by a flashing Cursor on the left of the list, this can be cycled around by using the key marked with a Hand. 12.8 Having selected the required option the user should then press the key with a lamp icon and the unit will enter the required test mode. Termination of Tests in the Manual Mode 13.1 When operating the unit in manual mode it is possible to terminate the testing early. 13.2 This can be achieved by pressing the power Button when the unit displays the following screen: 42 All Content Copyright © WiMAC Ltd 13.3 This will cause the unit to display the message “TEST COMPLETE” and exit the tests. 13.4 If the power off button is pressed again then the unit will fully power off. 13.5 If the power off button is not pressed then unit is ready to test another node when the indicator in the bottom left hand of the screen is incrementing. This is to allow the operator chance to note the serial number being displayed. End. N.B. FCC warning statement: This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. 43 All Content Copyright © WiMAC Ltd LeafNut Customer Support Tel: +44 (0) 113 383 1000 Email: support@harvardtechnology.com Vizion Customer Support Tel: +44 (0) 1422 317 337 Email: enquiries@zondionltd.eu.com