Technical Data Sheet Thermo Scientific Cryogenic Freezer Model ULT7150-9-M22 Thermo Fisher Scientific, Asheville, North Carolina Model Number ULT7150-9-M22 Application, Rating and Electrical Data Storage of General Laboratory Materials 6.8 Cubic Feet / 193 Liters -140 °C to -150 °C @ 15 ~ 25°C Ambient 380 VAC, 50 hz, 3 Phase 11/15 16A-6h/9 feet 63 @ 5ft in front, 5 ft high CE Indoor Use Only Non-Corrosive, Non-Flammable, Non-Explosive, Good Air Ventilation, 15° C - 25° C (59-77°F) Refrigeration Configuration CFC/HCFC Free Thermo Autocascade 3.5 hp Copeland* Scroll Compressor / 1 Enhanced Finned-Tube / Forced-Air Cooled / 1 Capillary Tube Cold Wall With Enhanced Heat Transfer Treatment Manual Defrost CFC/HCFC-Free Enviroementally Safe Refrigerant Mixture/ Non-Flammable Specifications Application Storage Volume Temperature Rating Electrical Power Rated Amps/Breaker Power Plug/Power Cord Length Sound Pressure Level (dBA) Agency Listings Indoor/Outdoor Usage Application Environment Refrigeration System Compressor / Number Condenser Type/Number Expansion Device Evaporator Type Defrost Method Refrigerant Charge/Flammability Controller/Electrical System Configuration and Features Controller Level Power Switch Controller Type Setpoint Security Compressor Safe Guard Control Sensor RS232/Remote Alarm Terminals Power Surge Protector Adjustable Warm/Cold Alarms Auto-Voltage Safeguard Interior Dimensions Exterior Dimensions Insulation Perimeter Heater Rack Capacity All-Direction Casters/Level Legs Ship Weight Other Options User Friendly Keyed On-Off Microprocessor-Based IntrLogic Control Yes Internal compressor pressure and temperature monitor. No audible or visible alert. Single RTD RS232/Remote Alarm Terminals N/A Standard Audible N/A Dimensions and Construction 26.5"H X 19" F-B" X 23.5" W(67 H cm X 48 F-B cm X 59 W cm) 42.75"H X 29" F-B X 58" W(109 H cm X 74 F-B cm X 148 W cm) 5 in. (12.7 cm) High Density HFC-blown Polyurethane Foam (R=42) Yes 12 X Catalog 5978 or 12 X Catalog 6096 Standard Approximately 825 lbs. / 374 Kg Electronic Recorder LN2 Back-Up System Typical Performance Characteristics in 25 ° C Ambient ULT7150-9-M22: -150C PD & Warm Up Performance temp C WU 40 temp C Avg-Ctr-TC ULT7150-9-M22: -150C Cycling Performance Max Min Avg TC31 -130 1 0 101 201 301 1 101 201 301 401 501 601 701 801 901 -40 -140 -80 -145 -120 -150 401 DeLog -135 Ambient: 25°C Load: Empty Sensor weighting: Empty Pull down (-50°C to -150°C): 210 min Avg. uniformity: 3.0 °C Avg. stability: 4.7 °C Peak variation range: 8.9 °C Cycle rate (on/off): 16 Minutes / 7 Minutes Avg. energy consumption: 72.6 kW-hr/day Avg. heat rejection: 10326 Btu/hr (3025 W) -155 -160 time (minutes) time (minutes) 1) Performance is nominal and individual units may vary. 2) Freezer performance will differ due to product amount, product size and operating conditions. 3) Continuous product enhancements may , without notice, result in amendments or ommisions to this specification. Thermo Fisher cannot accept responsibility for damage, injury, loss or expenses resulting from misapplication of the information herein. © 2011 Thermo Fisher Scientific Inc. Copeland is a trademark of Emerson Electric Co. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific ULT7150M(22)-cutsheet 10/5/2011