Proposed Scope of Accreditation

advertisement
Certificate # L1145-1
Scope of Accreditation
For
ICC Instrument Co. Inc
1483 E Warner Ave
Santa Ana, CA 92705
Erica Bender
714-540-4966
In recognition of a successful assessment to ISO/IEC 17025:2005 and ANSI Z540-1 to the following
Calibration and Measurement Capabilities, accreditation has been granted to ICC Instrument Co. Inc for
the following:
Accreditation granted through: April 24, 2018
Calibration
Amount of Substance – pH/Conductivity
Calibration
Range
Parameter/Equipment
Conductivity – Source
pH
Electrical – Capacitance 1
Calibration
Parameter/Equipment
Capacitance - Source
(0.05 to 1) kHz
(0.05 to 1) kHz
(0.05 to 1) kHz
(0.05 to 1) kHz
(50 to 400) Hz
(50 to 200) Hz
(50 to 100) Hz
Capacitance – Measure
(0.1 to 1) kHz
Form 28.8—Rev 11 – 3/1/14
1
Expanded Uncertainty of
Measurement (+/-)
Remarks
1 µS/cm
0.35 µS/cm
10 µS/cm
0.37 µS/cm
100 µS/cm
2.3 µS/cm
1000 µS/cm
6.4 µS/cm
10 mS/cm
0.06 mS/cm
(4 to 10) pH
0.02 pH
pH Solutions
Range
Expanded Uncertainty of
Measurement (+/-)
Remarks
(0.33 to 11) nF
(11 to 110) nF
(0.11 to 3.3) µF
(3.3 to 11 µF
(11 to 33) µF
(33 to 110) µF
(0.11 to 1.1) mF
15 mF/F + 0.02 nF
5 mF/F + 0.2 nF
8 mF/F + 0.3 nF
8 mF/F + 0.3 nF
6 mF/F + 16 nF
8 mF/F + 120 nF
14 mF/F + 270 nF
Multifunction
Calibrator and LCR
Meter
1 nF to 10 000 nF
0.2% of reading
Digital RLC Meter
Conductivity Solutions
Page 1 of 9
Certificate # L1145-1
Electrical – Current 1
Calibration
Parameter/Equipment
DC Current - Source
DC Current – Measure
AC Current - Source
Range
(22 to 220) µA
(0.22 to 2.2) mA
(2.2 to 22) mA
(22 to 220) mA
(0.22 to 2.2) A
(2.2 to 10) A
(10 to 100) µA
(0.1 to 1) mA
(1 to 10) mA
(10 to 100) mA
(0.1 to 1) A
(22 to 220) µA
(40 to 1 000) Hz
(1 to 5) kHz
(5 to 10) kHz
(0.22 to 2.2) mA
(40 to 1 000) Hz
(1 to 5) kHz
(5 to 10) kHz
(2.2 to 22) mA
(40 to 1 000) Hz
(1 to 5) kHz
(5 to 10) kHz
(22 to 220) mA
(40 to 1 000) Hz
(1 to 5) kHz
(5 to 10) kHz
AC Current – Measure
Form 28.8—Rev 11 – 3/1/14
(0.22 to 2.2) A
(40 to 1 000) Hz
(1 to 5) kHz
(5 to 10) kHz
(2.2 to 10) A
(40 to 1 000) Hz
(10 to 100) µA
(20 to 45) Hz
(45 to 1 000) kHz
(0.1 to 1) mA
(20 to 45) Hz
(45 to 1 000) Hz
(1 to 10) mA
(20 to 45) Hz
(45 to 1 000) Hz
(10 to 100) mA
(10 to 20) Hz
(45 to 1 000) Hz
(0.1 to 1) A
(20 to 45) Hz
(45 to 1 000) Hz
Expanded Uncertainty of
Measurement (+/-)
70 µA/A + 0.01 µA
69 µA/A + 0.01 µA
70 µA/A + 0.1 µA
81 µA/A + 1 µA
110 µA/A + 36µA
0.7mA/A + 0.4 mA
20 µA/A + 2 nA
21 µA/A + 11 nA
21 µA/A + 110 nA
38 µA/A + 1 µA
122 µA/A + 20 µA
Remarks
Multifunction
Calibrator and DMM
480 µA/A + 0.03 µA
800 µA/A + 0.06 µA
2 mA/A + 0.1 µA
220 µA/A + 0.04 µA
814 µA/A + 0.6 µA
2 mA/A + 1.2 µA
500 µA/A + 0.43 µA
815 µA/A + 6 µA
2 µA/A + 12 µA
Multifunction
Calibrator and DMM
240 µA/A + 4 µA
815 µA/A + 60 µA
2 mA/A + 120 µA
900 µA/A + 44 µA
1 mA/A + 110 µA
1.2 mA/A + 230 µA
2.2 mA/A
5 mA/A + 35 nA
715 µA/A + 35 nA
5 mA/A + 0.3 µA
0.4 mA/A + 0.2 µA
5 mA/A + 2 µA
0.4 mA/A + 2 µA
8 ½ Digit
Multimeter
0.41 mA/A + 22 µA
0.4 mA/A + 20 µA
4 mA/A + 0.2 mA
5 mA/A + 0.2 mA
Page 2 of 9
Certificate # L1145-1
1
Electrical – Inductance
Calibration
Parameter/Equipment
Range
Expanded Uncertainty of
Measurement (+/-)
Remarks
Inductance – Measure
1 mH to 1 H
300 µH/H
Digital RLC Meter
Inductance – Source
100 mH
0.12 mH
Digital RLC Meter and
Standard Inductor
Range
Expanded Uncertainty of
Measurement (+/-)
Electrical – Resistance 1
Calibration
Parameter/Equipment
Resistance - Source
1 Ω, 1.9 Ω
10 Ω
19 Ω
100 Ω
190 Ω
1 kΩ
1.9 kΩ
10 kΩ
19 kΩ
100 kΩ
190 kΩ
1 MΩ
1.9 MΩ
10 MΩ
19 MΩ
100 MΩ
Resistance –Measure
(0.1 to 10) Ω
(10 to 100) Ω
(0.1 to 1) kΩ
(1 to 10) kΩ
(10 to 100) kΩ
(0.1 to 1) MΩ
(1 to 10) MΩ
(10 to 100) MΩ
260 µΩ
410 µΩ
710 µΩ
2.4 mΩ
4.8 mΩ
19 mΩ
38 mΩ
180 mΩ
430 mΩ
1.8 Ω
4Ω
31 Ω
58 Ω
550 Ω
1.2 kΩ
15 kΩ
18 µΩ/Ω + 62 µΩ
16 µΩ/Ω + 570 µΩ
14 µΩ/Ω + 520 µΩ
13 µΩ/Ω + 540 µΩ
14 µΩ/Ω + 56 mΩ
20 µΩ/Ω + 2.2 Ω
61 µΩ/Ω + 110 Ω
600 µΩ/Ω + 2.1 kΩ
(-200 to 0) °C
(0 to 630) °C
(630 to 800) °C
0.08 °C
0.08 °C
0.27 °C
Range
Expanded Uncertainty of
Measurement (+/-)
(0 to 220) mV
(0.22 to 2.2) V
(2.2 to 22) V
(22 to 220) V
(220 to 1 100) V
10 µV/V + 1 µV
10 µV/V + 1.4 µV
10 µV/V + 9 µV
11 µV/V + 100 µV
13 µV/V + 700 µV
RTD –
Pt 385 Simulate
Electrical – Voltage 1
Calibration
Parameter/Equipment
DC Volts- Source
Form 28.8—Rev 11 – 3/1/14
Remarks
Multifunction
Calibrator and DMM
8 ½ Digit Multimeter
Multifunction
Calibrator and
Electronic
Thermometer
Remarks
Multifunction
Calibrator and DMM
Page 3 of 9
Calibration
Parameter/Equipment
DC Volts- Measure
AC Volts – Source
Form 28.8—Rev 11 – 3/1/14
Range
Expanded Uncertainty of
Measurement (+/-)
Certificate # L1145-1
Remarks
(0.1 to 100) mV
(0.1 to 1) V
(1 to 10) V
(10 to 100) V
(100 to 1 100) V
11 µV/V + 0.4 µV
13 µV/V + 0.5 µV
9 µV/V + 0.6 µV
13 µV/V + 55 µV
13 µV/V + 440 µV
8½ Digit Multimeter
(1.1 to 10) kV
0.33% of reading
High Voltage Probe
(0.1 to 2.2) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(2.2 to 22) mV
(10 to 20) Hz
(20 H to 40) Hz
40 Hz to 20 kHz
(20 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(22 to 220) mV
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(0.22 to 2.2) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(2.2 to 22) V
(10 to 20) Hz
(20 to 40) Hz
40 Hz to 20 kHz
(20 to 100) kHz
(100 to 300) kHz
(300 to 500) kHz
(0.5 to 1) MHz
(22 to 220) V
(10 to 40) Hz
40 Hz to 20 kHz
(220 to 1 100) V
(50 to 1 000) kHz
700 µV/V + 6 µV
260 µV/V + 6 µV
140 µV/V + 6 µV
420 µV/V + 6 µV
1.1 mV/V + 9 µV
1.5 mV/V + 18 µV
2.2 mV/V + 35 µV
700 µV/V + 7 µV
280 µV/V + 7 µV
150 µV/V + 7 µV
475 µV/V + 7 µV
1 mV/V + 9 µV
1.5 mV/V + 17 µV
2 mV/V + 35 µV
700 µV/V + 19 µV
290 µV/V + 11 µV
290 µV/V + 11 µV
150 µV/V + 11 µV
1 mV/V + 30 µV
2 mV/V + 46 µV
2 mV/V + 60 µV
Multifunction
Calibrator and DMM
700 µV/V + 120 µV
210 µV/V + 35 µV
110 µV/V + 6 µV
170 µV/V + 22 µV
325 µV/V + 100 µV
560 µV/V + 170 µV
3 mV/V + 1.1 mV
700 µV/V + 1 mV
210 µV/V + 340 µV
100 µV/V + 80 µV
160 µV/V + 230 µV
330 µV/V + 1 mV
700 µV/V + 2 mV
1.6 mV/V + 6 mV
700 µV/V + 12 mV
105 µV/V + 1.2 mV
105 µV/V + 5 mV
Page 4 of 9
Calibration
Parameter/Equipment
AC Volts – Measure
AC Volts – Measure
Thermocouple Simulation
Type J
Type K
Type R
Form 28.8—Rev 11 – 3/1/14
Range
(0.1 to 10) mV
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(10 to 100) mV
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 2) MHz
(0.1 to 1) V
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 2) MHz
(1 to 10) V
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(0.3 to 1) MHz
(1 to 2) MHz
(10 to 100) V
40 Hz to 1 kHz
(1 to 20) kHz
(20 to 50) kHz
(50 to 100) kHz
(100 to 300) kHz
(100 to 1 000) V
(40 to 1 000) Hz
(1 to 10) kV
(40 to 1 000) Hz
(-210 to –100) °C
(-100 to –30) °C
(-30 to 150) °C
(150 to 760) °C
(760 to 1 200) °C
(-200 to –100) °C
(-100 to 120) °C
(120 to 1 000) °C
(1 000 to 1 372) °C
(0 to 250) °C
(250 to 1 000) °C
(1 000 to 1 767) °C
Expanded Uncertainty of
Measurement (+/-)
335 µV/V + 6.6 µV
235 µV/V + 6 µV
360 µV/V + 6 µV
5.6 mV/V + 3.6 µV
46 mV/V + 3.9 µV
Certificate # L1145-1
Remarks
8½ Digit Multimeter
141 µV/V + 9 µV
140 µV/V + 8.5 µV
180 µV/V + 9.2 µV
350 µV/V + 7.5 µV
920 µV/V + 5.8 µV
3 mV/V + 15 µV
11 mV/V + 64 µV
100 µV/V + 48 µV
95 µV/V + 24 µV
170 µV/V + 24 µV
360 µV/V + 26 µV
930 µV/V + 25 µV
3 mV/V + 130 µV
12 mV/V + 180 µV
8½ Digit Multimeter
100 µV/V + 480 µV
95 µV/V + 240 µV
170 µV/V + 240 µV
360 µV/V + 260 µV
1 mV/V + 250 µV
3 mV/V + 1.3 mV
11 mV/V + 2.2 mV
0.25 mV/V + 5 mV
0.24 mV/V + 2.4 mV
0.24 mV/V + 2.5 mV
0.42 mV/V + 2.3 mV
1.4 mV/V + 2.3 mV
0.50 mV/V + 25 mV
0.6 % of reading
0.31 °C
0.19 °C
0.17 °C
0.2 °C
0.27 °C
0.38 °C
0.19 °C
0.3 °C
0.46 °C
0.66 °C
0.41 °C
0.47 °C
High Voltage Probe
Multifunction
Calibrator and
Electronic
Thermometer
Page 5 of 9
Calibration
Parameter/Equipment
Type T
Range
(-250 to –150) °C
(-150 to 0) °C
(0 to 400) °C
Length – Artifacts and Standards 1D
Calibration
Range
Parameter/Equipment
Expanded Uncertainty of
Measurement (+/-)
Certificate # L1145-1
Remarks
0.73 °C
0.28 °C
0.19 °C
Expanded Uncertainty of
Measurement (+/-) 2
Remarks
(35 + 2L) µin
Laser Micrometer
Expanded Uncertainty of
Measurement (+/-) 2
Remarks
(0 to 4) in
(56 + 7L) µin
Gage Blocks
(0 to 1) in
(52 + 2.3L) µin
Gage Blocks, Optical
Flats
(1 to 8) in
(46 + 11L) µin
Height Gages 1
(0 to 24) in
(1 000 + 5L) µin
Gage Blocks, End Rods
Calipers 1
(0 to 24) in
(510 + 9L) µin
Gage Blocks, End Rods
Steel Rules
(0 to 36) in
0.015 in
Certified Steel Rule
Range
Expanded Uncertainty of
Measurement (+/-)
Remarks
(0 to 110) scfh
3.4 scfh
Digital Flowmeter
Range
Expanded Uncertainty of
Measurement (+/-)
Remarks
(10 to 1 000) g1
(1 to 10) kg1
(0.5 to 100) lb1
(50 to 400) lb
0.09 g + 0.00007 g / g
1.2 g
0.013 lb + 0.00006 lb / lb
0.12 lb
Cylindrical Plugs
(0.01 to 1) in
Length – Hand Tools and Precision Gages 1D
Calibration
Range
Parameter/Equipment
Indicators 1
Micrometers – OD
1
Mass – Flow
Calibration
Parameter/Equipment
Variable Area Flowmeter
Mass – Force
Calibration
Parameter/Equipment
Force Gauges
(Tension Only)
Mass – Pressure/Low Vacuum 1
Calibration
Range
Parameter/Equipment
Pressure – Source
Form 28.8—Rev 11 – 3/1/14
(0.01 to 1) inH2O
(1 to 10) inH2O
(10 to 300) inH2O
(300 to 650) inH2O
(10 to 1 000) psi
Expanded Uncertainty of
Measurement (+/-)
0.002 inH2O
0.03 inH2O
0.09 inH2O
0.2 inH2O
0.027 psi + 0.00029 psi / psi
ASTM Class 1 & 2
weights, NIST Class S,
Q & F weights
Remarks
Pressure Module
Dead Weight Tester
Page 6 of 9
Calibration
Parameter/Equipment
Certificate # L1145-1
Remarks
Range
Expanded Uncertainty of
Measurement (+/-)
Pressure – Source
(1 000 to 2 000) psi
(2 000 to 5 000) psi
(5 000 to 10 000) psi
2.4 psi
4.3 psi
6.8 psi
Digital Pressure
Calibrator
Vacuum
(0 to 29.9) inHg
0.0 inHg
Digital Vacuum
Calibrator
Mass – Scale and Balances 1
Calibration
Parameter/Equipment
Range
Expanded Uncertainty of
Measurement (+/-)
(0.01 mg Resolution)
(0 to 100) g
0.3 mg
(0.01 mg Resolution)
(100 to 200) g
0.6 mg
(1 mg Resolution)
(200 to 2 000) g
6 mg
(1 mg Resolution)
(2 to 5) kg
14 mg
(0.001 lb Resolution)
(0 to 100) lb
0.001 lb/lb
(0.1 lb Resolution)
(0 to 820) lb
0.003 lb/lb
(1 lb Resolution)
(0 to 3 280) lb
2 lb
Range
Expanded Uncertainty of
Measurement (+/-)
(1 to 20) lbf·in
(20 to 500) lbf·in
(1 to 5) lbf·in
(5 to 3 000) lbf·in
0.005 lbf·in + 0.0005 lbf·in / lbf·in
0.0007 lbf·in + 0.0007 lbf·in / lbf·in
0.005 lbf·in + 0.005 lbf·in / lbf·in
0.0025 lbf·in + 0.006 lbf·in / lbf·in
Mass – Torque 1
Calibration
Parameter/Equipment
Torque Transducer
Torque Wrench
Thermodynamics – Infrared (IR) Temperature
Calibration
Range
Parameter/Equipment
IR Thermometer
Thermodynamics – Humidity
Calibration
Parameter/Equipment
Humidity
Measure 1
Source
Form 28.8—Rev 11 – 3/1/14
Remarks
ASTM Class1, Class 2
Weights
NIST Class F Weights
Remarks
Torque Arm & Weights
Digital Torque
Calibrator
Expanded Uncertainty of
Measurement (+/-)
Remarks
(20 to 50) °C
(50 to 200) °C
(200 to 350) °C
1.5 °C
2.2 °C
3 °C
Blackbody
Range
Expanded Uncertainty of
Measurement (+/-)
Remarks
(20 to 90) %RH
1.3 % RH
Vaisala HM70
33.1 %RH
75.5 %RH
1.3 % RH
Vaisala HM70 and
Humidity Salts
Page 7 of 9
Certificate # L1145-1
Thermodynamics – Thermometers and Probes
Calibration
Range
Parameter/Equipment
Thermodynamic
Measure/Source
Expanded Uncertainty of
Measurement (+/-)
(-35 to 400) °C
0.04 °C
(400 to 1 000) °C
3.6 °C
Time and Frequency – Frequency / Period
Calibration
Range
3
Parameter/Equipment
Remarks
7 ½ Digit Multimeter
with SPRT
Type R Thermocouple
with Digital
Thermometer
Expanded Uncertainty of
Measurement (+/-)
Remarks
9 x 10-7 ∙ f
Electronic Counter
1 mHz to 1.05 GHz
9 x 10-7 ∙ f
Frequencies generated
by Signal
Generator/Level
Generator monitored by
a counter
(300 to 1 800) RPM
0.11 RPM
(10 to 100 000) RPM
0.03 RPM
Stopwatch
Up to 24 hrs
0.05 sec/day
Electronic Counter
Process Timers 1
Up to 24 hrs
0.6 sec/day
Certified Stopwatch
Expanded Uncertainty of
Measurement (+/-)
Remarks
Frequency Measure
Frequency Source
Tachometers 1
Mechanical
Optical
1 mHz to 1.3 GHz
Tachometer Calibrator
Function Generator
Time and Frequency – Oscilloscopes
Calibration
Range
Parameter/Equipment
Oscilloscopes
Time Markers
DC Voltage Source
Oscilloscopes
into 50 ohms
DC Voltage Source
Oscilloscopes
Into 1M ohms
Square Wave, Source
Oscilloscopes
into 50 ohms
Square Wave, Source
Oscilloscopes
into 1M ohms
Leveled Sine Wave
Amplitude
50 kHz (reference)
Form 28.8—Rev 11 – 3/1/14
2 nS to 1 µS
2 µS to 5 S
12 mS/S
14 mS/S
(0 to 2.2) V
3 mV/V + 0.12 mV
(0 to 33) V
3 mV/V + 0.12 mV
Fluke 5500A/SC300
1.8 mVp-p to 2.2
Vp-p
4 mV/V + 0.17 mV
1.8 mVp-p to 105
Vp-p
4 mV/V + 0.17 mV
5 mV to 5. 5V
23 mV/V + 0.24 mV
Page 8 of 9
Calibration
Parameter/Equipment
Range
Expanded Uncertainty of
Measurement (+/-)
5 mV to 5.5 V
41 mV/V + 0.35 mV
5 mV to 3 V
50 mV/V + 0.35 mV
5mV to 5.5V
18 mV/V + 0.13 mV
5 mV to 3 V
27 mV/V + 0.16 mV
Level Sine Wave
Frequency
50 kHz to 300 MHz
6 kHz/ MHz
(0.3 to 1) GHz
24 kHz/ MHz
Leading Edge Risetime
1 kHz to 1 MHz
1.2 ns
Leveled Sine Wave
Amplitude
50 kHz to 100 MHz
Leveled Sine Wave
Amplitude
100 MHz to 300 MHz
Leveled Sine Wave
Flatness
50 kHz to 100 MHz
Leveled Sine Wave
Flatness
100 MHz to 300 MHz
Certificate # L1145-1
Remarks
Fluke 5500A/SC300
Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range,
expanded uncertainty of measurement and remarks. The expanded uncertainty of measurement is expressed as the standard
uncertainty of the measurement multiplied by a coverage factor of 2 (k=2), corresponding to a confidence level of
approximately 95%.
Notes:
1) Laboratory offers calibration services at the laboratory's own facilities and at the client or other agreed upon facilities.
2) L = length in inches
3) f = frequency in Hz
Approved by:
Date: June 10, 2015
R. Douglas Leonard
Chief Technical Officer
Re-Issued: 6/10/15
Form 28.8—Rev 11 – 3/1/14
Page 9 of 9