Medical Engineering & Physics 26 (2004) 569–579 www.elsevier.com/locate/medengphy The effect of duty cycle and frequency on muscle torque production using kilohertz frequency range alternating current Alex R. Ward a,, Valma J. Robertson b, Harry Ioannou a a Department of Human Physiology and Anatomy, School of Human Biosciences, La Trobe University, Bundoora, Melbourne, Vic. 3086, Australia b School of Physiotherapy, La Trobe University, Vic. 3086, Australia Received 24 February 2003; received in revised form 11 March 2004; accepted 20 April 2004 Abstract We investigated the frequency and duty cycle dependence of maximal electrically induced torque (MEIT) of the wrist extensors. Fifty hertz burst modulated sinusoidal alternating current (AC) in the frequency range 0.5–20 kHz was used, with duty cycles ranging from a minimum (one cycle) to maximum (continuous AC). MEITs were similar at low frequencies but decreased markedly above 2.5 kHz. MEITs also decreased markedly above a 20% duty cycle. Subjective reports of discomfort were fewest at 4 kHz and at duty cycles in the range 20–25%. Our conclusion is that for maximum torque production, a frequency of 1 kHz and a duty cycle of 20% are indicated. When comfort is a major consideration, a frequency of 2.5 kHz provides an acceptable trade-off between MEIT and comfort. The findings also suggest that low duty cycle, burst modulated AC stimulation may be more effective than stimulation using conventional low-frequency pulsed current. # 2004 IPEM. Published by Elsevier Ltd. All rights reserved. Keywords: Electrical stimulation; Alternating current; Frequency; Duty cycle; Torque 1. Introduction Electrical stimulation is widely used in rehabilitation for pain control, muscle re-education, prevention of atrophy, and restoration of function (functional electrical stimulation or FES) [1,2]. Its potential application in athletic performance through muscle strengthening has also been explored using healthy subjects and athletes [3,4]. Both low-frequency rectangular pulsed current and sinusoidal alternating current (AC) in the low kilohertz frequency range are used. When kilohertz frequency AC is used, it is modulated at a low frequency [5], typically 50 Hz. For muscle strengthening, burst modulated kilohertz frequency AC has been claimed to be more effective than low-frequency rectangular pulsed current [1–4], however, there have been insufAbbreviations: AC, alternating current; MEIT, maximal electrically induced torque Corresponding author. Tel.: +61-3-9479-5787; fax: +61-3-94795784. E-mail address: a.ward@latrobe.edu.au (A.R. Ward). ficient comparative studies and the findings are either inconclusive or contradictory [5,6]. The most commonly used sinusoidal AC frequencies are 2.5 and 4 kHz, though there is only limited empirical evidence to justify the choice [1–3]. A frequency of 2.5 kHz is advocated for the production of maximal electrically induced torque (MEIT) [5,6]. A frequency of 4 kHz is claimed to be more comfortable and therefore a more appropriate choice if maximal torque is not the objective [1–3,7,8]. These claims are consistent with early qualitative observations of the effects of kilohertz frequency AC [9]. d’Arsonval [10] reported that when a constant voltage AC stimulus is applied transcutaneously, neuromuscular excitation becomes stronger up to 1.25–1.5 kHz, is constant between 1.5 and 2.5 kHz and decreases to 5 kHz (the highest frequency his apparatus could produce). He also noted that current with a frequency of 1.5 kHz is more painful than 5 kHz but much more comfortable than low-frequency AC stimulation (75 and 20 Hz were the comparisons). 1350-4533/$ - see front matter # 2004 IPEM. Published by Elsevier Ltd. All rights reserved. doi:10.1016/j.medengphy.2004.04.007 570 A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 Stimulation using 2.5 kHz AC gained popularity in the late 1970s, largely as a result of the activities of a Russian scientist, Kots, who claimed force gains of up to 40% in elite athletes as a result of what was then regarded as a new form of electrical stimulation [6,11,12]. He did not provide details of his published work, nor did he give references and what became known as ‘Russian current’ found popularity despite the lack of research in the English-language literature [5,6]. The AC frequency advocated by Kots was either 1 or 2.5 kHz, depending on whether stimulation was via the nerve trunk (1 kHz) or over the muscle (2.5 kHz) [6,12]. The recommended burst frequency was 50 Hz and the duty cycle (burst ‘on’ to ‘on þ off’ time), 50%. The features of the stimulus waveform are illustrated in Fig. 1(a). A frequency of 2.5 kHz is usual for commercially produced Russian current stimulators, and 1 kHz appears to have been overlooked. A recent study [13] which measured MEIT of the wrist extensors used AC frequencies in the range 1–15 kHz and found that greatest MEIT was elicited at 1 kHz. This study also used a 50 Hz burst frequency, a duty cycle of 50% and an electrode placement typical of those used clinically. Whether frequencies lower than 1 kHz might elicit greater MEIT was not examined, but Kots and co-workers [6,12] reported that the MEIT measured at 0.5 kHz was appreciably less than at 1 or 2.5 kHz. This suggests that 1 kHz is optimal for torque production but further experimental evidence is required. Few studies have examined the effect of duty cycle on torque production. Kots [6,12] stated that he adopted a 50% duty cycle as it produced a similar MEIT to a continuous sinewave stimulus of the same peak intensity, but resulted in half the average current, so the risk of tissue damage was reduced. This rationale appears valid: AC with a 50% duty cycle delivers one-quarter the power of a continuous stimulus of the same peak intensity. Kots apparently did not consider whether lower duty cycles might have a greater advantage. One study which did examine the effect of duty cycle was reported by Bankov [14]. Bankov used 5 kHz sinusoidal AC, modulated at 60 Hz with duty cycles of 6%, 12% and 30%. The modulation envelope was sinusoidal in shape. Biceps brachii were stimulated at an intensity which produced sufficient force to keep the v elbow joint at 90 flexion with the upper arm vertical. Subjects were asked to comment on the relative discomfort of the stimulus and consistently reported an increase in discomfort with increasing duty cycle. Another study which included duty cycle as a variable was that of Moreno-Aranda and Seireg [15], they investigated the effects of kilohertz frequency, burst frequency and duty cycle on finger flexor torque production and discomfort. Stimulus voltages were kept constant and the AC frequency, initially at 500 Hz, was increased in steps to 10 kHz. The duty cycle was varied in the range 10–100% and burst frequency in the range 25–400 Hz at each AC frequency. The authors found that the optimal parameters for torque production were an AC frequency of 4 kHz, a burst frequency of 100 Hz, and a duty cycle of 20%. The authors also found that discomfort decreased with increasing frequency over the range examined and concluded that for the best compromise between torque production Fig. 1. Burst modulated sinusoidal current with a burst frequency of 50 Hz and (a) a 50% duty cycle and (b) a 10% duty cycle. The period is 20 ms, so the burst frequency in each case is 50 Hz. A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 and comfort, the optimum frequency was 10 kHz with other parameters remaining the same. The studies by Bankov [14] and Moreno-Aranda and Seireg [15] each had important limitations. Bankov [14] chose a fixed, low level of torque production as the parameter to be kept constant. The choice has obvious relevance for clinical situations where muscle contractions are required, but maximum torque production is not a central objective. Moreno-Aranda and Seireg [15] chose a constant stimulus voltage across conditions and examined how both torque production and discomfort varied with duty cycle and frequency. Their focus was on efficiency, i.e. how to best evoke a reasonable amount of torque with minimal discomfort at a particular stimulation intensity. The existing literature indicates that no optimal duty cycle for medium frequency AC stimulation has been established. ‘Optimal’ may well depend on the chosen experimental constraints. Bankov [14] kept muscle torque at a constant level. Moreno-Aranda and Seireg [15] kept stimulus intensity constant. Both authors concluded that an optimal duty cycle was less than the 50% commonly used and was probably 20% or less. An important question is whether the same duty cycle range is optimal in terms of maximum torque production. The aim of the present study was to examine the effect of duty cycle on maximum electrically induced torque (MEIT) in normal, healthy subjects. A secondary aim was to examine the effect of AC frequency and whether the variation in MEIT with duty cycle was frequency dependent. Discomfort was qualitatively assessed to establish whether MEIT is limited by the discomfort associated with electrical stimulation. 2. Method Sixteen subjects, eight males and eight females, participated in the study. Subjects were recruited from staff and students of the School of Physiotherapy, La Trobe University. Subjects’ ages ranged from 19 to 52 years with a mean of 26.0 years (standard deviation 10.4 years). Prior to data collection, ethics approval was 571 obtained from the Human Research Ethics Committee of the Faculty of Health Sciences, La Trobe University. Inclusion criteria required that all subjects had previously used and experienced electrical stimulation. This meant that subjects would be familiar with electrical stimulation and more consistent when manually adjusting the stimulus intensity. Other inclusion criteria required that the subject had no pathology affecting the left forearm, no pacemaker and no damage to the skin overlying the wrist extensor muscles. Subjects were required for two sessions of electrical stimulation and were asked to experience all 41 conditions (combinations of six frequencies and eight duty cycles, see Table 1), applied in a different, randomised order at each session. Randomisation was important as an order effect (principally due to fatigue) was expected. Subjects were asked to inform the experimenter if any test conditions felt noticeably more uncomfortable than others. After the procedure was explained to each subject, and informed consent obtained, the skin of the posterior surface of the left forearm was cleaned using an alcohol swab and conductive rubber electrodes, measuring 44 40 mm, were attached using conductive, adhesive skin mounts (American Imex type 00200). The electrodes were positioned so as to efficiently stimulate the wrist extensors: on a line from the head of the radius to the distal radioulnar joint with the proximal electrode 1 cm distal to the head of the radius and the distal electrode 5 cm distal to the proximal electrode along this line. The electrode leads were attached, ensuring that the cathode was the distal electrode, i.e. the distal electrode was the negative terminal for the initial half-cycle of the sinewave burst. The stimulator was a purpose-built device designed to produce a burst of sinewaves with user selection of the burst frequency and number of sinewaves per burst. A zero-crossing detector was used to ensure that only complete sinewaves were gated. The burst frequency was set at 50 Hz. The number of sinewaves per burst was varied from a single sinewave (minimum duty cycle) to continuous output (100% duty cycle) in discrete steps. Test frequencies of 0.5, 1, 2.5, 4, 10 and 20 Table 1 The combinations of frequency and duty cycle used in this study. Whole numbers of sinewave cycles were used (numbers shown in parentheses). Duty cycles above 50% were not used at 10 and 20 kHz because of the risk of burns and blistering Frequency (kHz) 0.5 1 2.5 4 10 20 Duty cycle (%) 0.25 (1) 0.5 (1) 0.5 (2) 1.25 (1) 1 (2) 1.25 (5) 2 (1) 2.5 (2) 2.5 (5) 2.5 (10) 5 (1) 4 (2) 6.25 (5) 5 (10) 5 (20) 10 (1) 10 (2) 10 (5) 12.5 (10) 10 (20) 12.5 (50) 20 (2) 25 (5) 20 (10) 25 (20) 25 (50) 25 (100) 50 (5) 50 (10) 40 (20) 50 (40) 50 (100) 50 (200) 70 75 60 75 (7) (15) (30) (60) 100 (cont) 100 (cont) 100 (cont) 100 (cont) 572 A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 kHz were used. At 10 and 20 kHz, duty cycles greater than 50% were not tested as preliminary experimentation, using one of the investigators as a subject, indicated a risk of burns and blistering due to the high average currents involved. Table 1 lists the test combinations of frequency and duty cycle. The subject’s forearm was secured in a device built expressly for measuring wrist extensor torque, described previously [13]. The wrist axis was aligned with the rotational axis of the apparatus and the forearm and hand were secured with velcro straps. A force gauge (Mecmesin Inc. model MFG500: range 0–500 N, with a resolution of 0.2 N) connected to the pivoting arm of the apparatus measured wrist extensor torque, calculated as force distance from the axis of rotation (53 mm). Since the extremely high stiffness of the force gauge prevented rotation about the axis these measurements were essentially of isometric torque. Prior to data collection, each subject was asked to experiment with the stimulus intensity for familiarisation and to increase the intensity to a level which they regarded as the maximum tolerable. After two repeats, data collection began, following a randomised order of conditions produced for the particular subject. For each condition, the subject was asked to adjust the stimulus intensity to the ‘maximum tolerable’ level, indicating to the investigator when they had done so, so that 3-s measurements of torque, stimulus voltage, and current could be made. This was followed by a 3-s rest interval. The measurements were repeated twice so that three measurements were obtained for each test condition. 3. Results 3.1. The effect of frequency on MEIT To examine the effect of frequency on MEIT, the data were averaged across subjects at each particular frequency and duty cycle. Plots of average MEIT versus frequency at different duty cycles are shown in Fig. 2. Exact matching of duty cycles across frequencies was not possible because of the requirement that whole numbers of sinewaves be used. Errors, expressed as standard deviations, are not shown in Figs. 2, 3 and 6, because if ‘error’ is defined as the standard deviation for measurements across all subjects, it would be very large (approaching 100%). The true error is much smaller than a standard deviation would indicate. This is because the intra-subject variance in MEIT is large. The effect of intra-subject variation as a potentially confounding variable was eliminated in our study by the experimental design: measurements were made at all frequencies and duty cycles with each subject. Fatigue also adds to the vari- ation in measured MEIT values: torque measurements were lower if they occur later in a measurement session. The effect of fatigue as a potentially confounding variable was minimized by randomising the order of presentation of each frequency/duty cycle combination across subjects. The dashed line in Fig. 2(a) indicates that at 0.5 kHz, the lowest duty cycle possible (one sinewave) is 10%. Dashed lines in graphs (e) and (f) indicate that at duty cycles of 60/70/75% and 100%, the average current output, as indicated previously, was possibly too high to be safe and was therefore not tested. Fig. 2 shows that for all duty cycle ranges, maximum torque was produced at low frequencies (0.5, 1 and 2.5 kHz) and for all but one of the six frequencies, maximum torque was elicited at 1 kHz. Between 2.5 and 20 kHz, there was a rapid decrease in torque with increasing frequency. Establishing 1 kHz as the optimum frequency required a statistical comparison of torques produced at 0.5, 1 and 2.5 kHz. We compared the data using a non-parametric, binomial sign test (which is robust and makes no assumptions about sample size or normality of distributions). Comparison of data obtained at frequencies of 1 and 0.5 kHz yielded a p-value of 0.03. The results thus indicated that although the torques differed by only about 3%, which is small in clinical terms, the difference was statistically significant. The difference between the means at 1 and 2.5 kHz was almost 7% but did not reach statistical significance. A p-value of 0.06 was obtained, suggesting a problem of statistical power rather than there being no real difference. The evidence thus indicates that 1–2.5 kHz is an optimum frequency range for electrically induced torque production. A frequency of 0.5 kHz elicits significantly less torque than 1 kHz. Above 2.5 kHz, a rapid decrease in MEIT occurs with increasing frequency. 3.2. The effect of duty cycle on MEIT To investigate the effect of duty cycle on MEIT, graphs of torque vs. duty cycle were plotted at each frequency (Fig. 3). Fig. 3 shows that, at least in the range 20–100% duty cycle, there is a clear downward trend in MEIT with increasing duty cycle for all frequencies. Fig. 3(c), a frequency of 2.5 kHz, seems to show that maximum torque production occurs at 10% duty cycle. However, this is not reflected in the duty cycle vs. torque graphs for 0.5 kHz (Fig. 3(a)) and 1 kHz (Fig. 3(b)). Fig. 3(d) shows that at 4 kHz, maximum torque production appears to occur at the 12.5% duty cycle. Results for frequencies of 10 and 20 kHz (Fig. 3(e) and (f), respectively) at low duty cycles are equally inconclusive. A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 573 Fig. 2. Plots of average torque vs. frequency for all duty cycles. The dashed line in (a) indicates that at 0.5 kHz, testing was not possible as one sinewave at 0.5 kHz is a 10% duty cycle. Dashed lines in graphs (e) and (f) indicate that at the higher duty cycles, the average current output was possibly too high to be safe. To further examine the variation in MEIT at low duty cycles, lines of best fit were calculated at each frequency using the 20–100% data. These are the lines drawn in Fig. 3. Next, a predicted MEIT was calculated using the line of best fit for each measured duty cycle less than 20%. Calculated values in the range less than 20% were compared with measured values using a non-parametric, binomial sign test. The difference was found to be significant (p ¼ 0:008), indicating that the graphs depart from linearity at low duty cycles. This means that the graphs either curve, plateau or decrease below 20% duty cycle. To establish which of these applies, a linear regression was performed using data for all frequencies and duty cycles of 10% or less, stan- dardising to compensate for absolute changes with frequency. The gradient obtained (m ¼ 0:43 0:38, r ¼ 0:28) demonstrates a weak, barely significant downward trend in MEIT between 10% and 0% duty cycle. This indicates either a plateau in MEIT or a slight downward trend over the lowest duty cycles. 3.3. Discomfort Subjects were asked to comment if any conditions felt more uncomfortable than others. The reason for asking this question was a previous empirical observation we have made when using burst modulated kilohertz frequency AC stimulation. MEIT does not seem to be determined solely by pain evoked directly 574 A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 Fig. 3. Plots of average torque vs. duty cycle at all frequencies. Lines of best fit for data in the duty cycle range 20–100% are shown. r2 values greater than 0.96 were obtained at all duty cycles. Dashed lines in graphs (e) and (f) indicate that at the higher duty cycles, the average current output was possibly too high to be safe. by the stimulation, but partly by the perception of the forcefulness of the resulting contraction. That is, the limit is often set by a subject’s feeling that if the contraction were any stronger, muscle or tendon damage might occur. Hence, some variation in perceived discomfort was anticipated, even though the stimulation was to the ‘maximum tolerable’ level. The frequency of reports of discomfort is shown in Table 2, together with the number of different conditions within each duty cycle range (values in parentheses). In order to compensate for particular frequencies having a greater number of conditions than others, data were standardised by dividing by the number of conditions that were used within a particular duty cycle range. Standardised numbers were then summed across duty cycles to obtain an index of discomfort for each tested frequency. The 0.5 kHz total was multiplied by 6/5 to compensate for there being no conditions in the ‘less than 10%’ category. Similarly, 10 and 20 kHz totals were multiplied by 6/4 as they were only represented in four of the six duty cycle groups. A graph of these standardised totals vs. frequency is shown in Fig. 4. Fig. 4 indicates that 0.5 kHz is the most uncomfortable frequency, accounting for 35% of all comments. There is a gradual decrease with increasing frequency to 4 kHz. From 4 to 20 kHz, there is an increase in negative comments. A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 575 Table 2 Number of reports of discomfort categorized by frequency and duty cycle. Values in parentheses are the number of different conditions within the duty cycle range Frequency (kHz) 0.5 1 2.5 4 10 20 Duty cycle <10% 10/12.5% 20/25% 40/50% 60/70/75% 100% (0) 2 (1) 5 (2) 5 (3) 18 (4) 27 (5) 9 (1) 1 (1) 2 (1) 1 (1) 5 (1) 3 (1) 4 (1) 2 (1) 1 (1) 0 (1) 1 (1) 4 (1) 5 (1) 2 (1) 2 (1) 2 (1) 1 (1) 3 (1) 9 (1) 5 (1) 1 (1) 3 (1) (0) (0) 11 (1) 7 (1) 7 (1) 5 (1) (0) (0) To examine the variation in discomfort with duty cycle, standardised data were summed across frequencies. The <10% total was multiplied by 6/5 to compensate for there being no conditions in the 0.5 kHz category. Similarly, 60/70/75% and 100% totals were multiplied by 6/4 as they were only represented in four of the six frequencies. A graph of standardised number of comments vs. duty cycle was plotted (Fig. 5, lightly shaded columns). Also shown are totals of only those frequencies which had duty cycles in each category (1, 2.5 and 4 kHz, dark shading). The totals across all frequencies have fewer reports of discomfort in the <10% than in the 10/12.5% category. This is evidently because there is no contribution to the <10% category at 0.5 kHz and this is the frequency associated with a disproportionally large number of reports of discomfort (Fig. 4). When only those frequencies which have representation in all duty cycle categories are used (dark shaded columns in Fig. 5), the variation is more systematic. Whether the totals for all frequencies are used or the totals for the more restricted range, a minimum in reports of discomfort occurs in the 20/25% duty cycle category. Greatest discomfort is associated with duty cycles of more than 50%. Fig. 4. Number of reports of discomfort at each frequency. Values are standardised to take account of varying numbers of conditions tested at each frequency (see text). Fig. 5. Standardised number of reports of discomfort vs. duty cycle summed over all frequencies (lightly shaded columns) and only summed over frequencies of 1, 2.5 and 4 kHz (dark columns). 3.4. Stimulation efficiency Stimulation efficiency is reflected by the stimulus voltage needed to generate a given amount of torque. In order to investigate stimulation efficiency, mean stimulus intensity values were divided by their corresponding MEIT values to give a voltage/torque ratio (V/N m). The results for each test frequency are shown in Fig. 6. Stimulus efficiency is greatest when the stimulus voltage used is lowest for a given torque value. This is found in Fig. 6 at the minimum of the V/N m curves. Fig. 6(a) shows that at 0.5 kHz, there is little variation in the range 10–75% duty cycle. Only the 100% figure is appreciably higher. A minimum possibly exists at 50%. At 1 kHz (Fig. 6(b)), some variation can be seen, with a minimum around a 30–40% duty cycle. The graphs 576 A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 Fig. 6. Voltage/torque ratio vs. duty cycle at each frequency. Dashed lines in graphs (e) and (f) indicate that at the higher duty cycles, the average current output was possibly too high to be safe. for 2.5, 4 and 10 kHz appear to have minima at progressively lower duty cycles. Another feature, most apparent at 20 kHz, is the steepness of the initial drop in voltage/torque ratio. At 20 kHz, the voltage/torque ratio decreases to almost half as the duty cycle increases from 0.25% to 2.5%. This effect decreases with decreasing frequency, partly because one sinewave at a lower frequency has a higher duty cycle than a higher frequency sinewave so small duty cycles do not exist at lower frequencies. Nonetheless, when common duty cycles are compared, the rate of decrease measured at 20 kHz is higher than that observed at lower frequencies. The 10 kHz rate is lower, but still higher than that observed at 4 kHz. 4. Discussion 4.1. Frequency effects The present study found that the greatest torque was produced at a frequency of 1 kHz (Fig. 3). A previous study by Ward and Robertson [13] compared frequencies in the range 1–10 kHz and also found that the greatest torque was elicited at 1 kHz. Lower frequencies were not tested in the earlier study so it was not established whether 1 kHz or some lower frequency was optimal. The present study, which found significantly higher MEIT at 1 kHz than 0.5 kHz, confirms and extends the earlier findings. A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 The present study also found that the difference in MEIT between 1 and 2.5 kHz did not achieve statistical significance, but the calculated p-value of 0.06 suggests, as noted previously, a problem of statistical power rather than that there was no real difference. A reanalysis of the data reported by Ward and Robertson [13] shows that significantly greater MEIT was induced at 1 kHz than 2 kHz (binomial sign test, p ¼ 0:006), thus supporting the notion that a frequency of 1 kHz is optimal. Whatever the optimum frequency, the lower frequencies (0.5, 1, and 2.5 kHz) produced considerably more torque than the higher frequencies (4, 10, and 20 kHz) and there was a general downward trend of torque produced with increasing frequency from 2.5 kHz. The variation in MEIT between 0.5 and 1 kHz averaged only 3% and that between 2.5 and 1 kHz, 7%. This suggests that despite the statistical significance of the differences, the clinical significance is more questionable. The decrease in discomfort between 0.5 and 1 kHz is continued to 4 kHz (Fig. 4). This suggests that if one had to choose between 0.5, 1, and 2.5, 0.5 kHz would be rejected because of greater discomfort and lesser MEIT. A frequency of 1 kHz would be preferred if MEIT was the sole criterion. With relative comfort as an additional criterion, 2.5 kHz would be the preferred option. The findings suggest that the traditional Russian current frequency of 2.5 kHz is not optimal from the point of view of MEIT, but is close to optimal, and more comfortable than 1 kHz. Thus, 2.5 kHz might be chosen as a trade-off between MEIT and discomfort. Torque production is not much less than at 1 kHz and comfort is greater. Minimum discomfort occurs with a frequency of 4 kHz. This supports the notion that so-called ‘interferential currents’, which are commonly used clinically at this frequency [1,2], can provide a very comfortable form of electrical stimulation. Comfort evidently has the penalty of lesser muscle force production. Whether greater comfort translates into clinical effectiveness remains open to question. The variation in MEIT with frequency can be attributed, at least in part, to the different strength–duration behaviour of nerve fibres with different diameters. As Li and Bak [16] observed, the strength–duration graph for smaller diameter fibres is shifted to the right (to longer pulse durations). This is illustrated qualitatively in Fig. 7. An AC frequency of 1 kHz, for example, has sinewaves with a positive phase which is one-half of a millisecond duration (500 ls). An AC frequency of 10 kHz has sinewaves with a positive phase which is one-tenth of the 1 kHz value; that is, a pulse width of 50 ls. Fig. 7 indicates that with long duration pulses, thresholds for large and small diameter fibres are closer 577 Fig. 7. Strength–duration curves for different nerve fibre types. A-b, A-beta or class II fibres, A-d, A-delta or class III fibres, C, C or class IV fibres. Adapted from Li and Bak [16]. together. This means that with supra-threshold stimulation, fibres with a wider range of diameters will be stimulated. With short duration pulses, there would be less stimulation of smaller diameter fibres. The motoneurons innervating skeletal muscle have a range of fibre diameters, so it would be expected that a larger population of muscle fibres would be recruited with long duration pulses. This, then, would explain the increase in MEIT with decreasing frequency. The penalty of stimulating fibres with a wider range of diameters is a predicted excitation of more A-d (pain signalling) fibres. This would account for the progressive increase in discomfort below 4 kHz. 4.2. Duty cycle effects With increasing duty cycle, torque decreased (Fig. 2). This was clearly evident at all frequencies for duty cycles of 20% and above. Below 20% duty cycle, torque appeared to reach a maximum somewhere between 10% and 20%. An optimum within this range was not able to be determined due to the amount of scatter in the results. Analysis of the data showed a positive trend (increasing MEIT with increasing duty cycle) up to 10%. Moreno-Aranda and Seireg [15] suggest that a 20% duty cycle is optimal for maximum torque production. Bankov [14] suggests that 12% is optimal. The present findings are in agreement with those earlier studies, despite the different methods used. Although no optimal value has been determined, the results indicate that it is somewhere in the range 10–20% and that the difference over this range is not large. Importantly, the 50% duty cycle adopted by Kots [6,12] is not appropriate if maximal torque production is the objective. A 578 A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 duty cycle in the range 10–20% would be more effective. This study also found that duty cycle affected perceived discomfort. The most frequently reported as uncomfortable was the 100% duty cycle (Fig. 5). A 50% duty cycle was more comfortable but least discomfort was reported with a 20–25% duty cycle. The minimum in the discomfort graph at 20–25% duty cycle, together with the finding of greatest MEIT using a 10–20% duty cycle, suggests that a 20% duty cycle is the optimum for eliciting MEIT with minimum discomfort. A possible explanation for the findings is that with long duration bursts of AC, nerve fibre firing rates will be multiples of the 50 Hz burst frequency [17,18]. Maximum firing rates approaching 1 kHz have been observed in isolated a-motoneurons stimulated suprathreshold using continuous (unmodulated) stimuli with frequencies in the range 1–10 kHz [19]. It follows that with bursts of current using a 50% duty cycle, the maximum firing rate would be less than 500 Hz and at a 20% duty cycle, less than 200 Hz. While a nerve fibre firing rate of somewhat less than 200 Hz might elicit greater MEIT than 50 Hz, firing rates approaching 500 or 1 kHz would result in rapid fibre dropout due to neurotransmitter depletion, propagation failure [20,21] and/or nerve block [17–19]. Similarly, fibre firing rates could provide an explanation for minimum discomfort with a 20% duty cycle. Gating of pain fibre activity is believed to be optimal when sensory fibres are stimulated at rates of 100 Hz or so [1–3,22]. Firing rates in excess of 200 Hz would thus be less efficient than lower firing rates for the modulation of pain sensation. AC depended strongly on the burst duration, i.e. the number of cycles in the burst, and also on the AC frequency. The AC frequency is important because the nerve fibre membrane acts as a rectifier [25] so an AC stimulus is able to push the nerve fibre closer to threshold with each successive pulse in a burst. Membrane threshold is reached when successive pulses result in sufficient depolarization to produce an action potential. The Gildemeister effect, as it became known, was demonstrated quantitatively by Schwarz and Volkmer [26] who measured the change in membrane potential of isolated nerve fibres as a result of successive pulses in a burst of kHz frequency AC. The Gildemeister effect requires that the pulses occur sufficiently rapidly that the membrane does not have time to recover between them. Depolarization due to summation is thus expected to occur more readily at higher kilohertz frequencies. The large drop in voltage/torque ratio at 20 kHz which occurs over low duty cycles (<10% in Fig. 6(f)) is consistent with the notion of summation of successive pulses. The lesser drop in voltage/torque ratios at lower frequencies is explicable in terms of a greater time between AC pulses, allowing the membrane to recover between them. The progressive rise in the voltage/torque ratios with higher duty cycles (above 20%) in Fig. 6 is explicable by nerve fibre firing rates which are several multiples of the burst frequency. As noted previously, this would result in fibre dropout due to neurotransmitter depletion, propagation failure [20,21] and/or nerve block [18,19]. It would have the consequence of reduced MEIT for a given stimulus intensity and hence a higher voltage/torque ratio. 4.3. Stimulation efficiency Stimulation efficiency, assessed in terms of the voltage/torque ratio, varied markedly with duty cycle. Fig. 6 shows that at all but the lowest of the tested frequencies (0.5 kHz), a duty cycle of approximately 20% produced the best stimulation efficiency (lowest voltage/torque ratio). Results at the extremes of the frequency range were equivocal, but in the range 1–10 kHz, a 20% duty cycle was optimal from the perspective of maximum torque production with minimum stimulus intensity. An interesting observation is the drop in voltage/torque ratio in the range 0–20%. The ratio is small at the lowest frequencies (0.5 and 1 kHz) but becomes more marked at higher frequencies (Fig. 6). The increasing rate and amount of drop at higher frequencies are likely due to summation of subthreshold depolarizations. Summation was first proposed by Gildemeister [23,24] as an explanation for his observations that the subjective sensation, and whether there was any sensation, associated with bursts of kilohertz frequency 5. Conclusion The 50% duty cycle which is commonly used clinically is less than optimal from the point of view of maximum muscle force production. The greatest MEIT is elicited, not at 50%, but with a duty cycle of 20% or less. A possible explanation is that at higher duty cycles, nerve fibre firing rates are too high to maximize and sustain an electrically induced contraction. A duty cycle of 20% appears to be optimal when MEIT, comfort and stimulation efficiency are taken into account. For maximum torque production, a frequency of 1 kHz is indicated. When comfort is also a consideration, a frequency of 2.5 kHz provides a trade-off between MEIT and comfort and this is the frequency traditionally used for Russian current stimulation. Least discomfort occurs at 4 kHz, the frequency traditionally used for interferential current stimulation, but less MEIT is produced. Whether a 1 or 2.5 kHz stimulus with a duty cycle of 10–20% is more efficient than a low-frequency pulsed A.R. Ward et al. / Medical Engineering & Physics 26 (2004) 569–579 current stimulus remains open to question. The present findings strongly suggest that low duty cycle, burst modulated, AC stimulation produces at least the same amount of torque as low-frequency pulsed current but with less discomfort. References [1] Low J, Reed A. Electrotherapy explained, 3rd ed. Oxford: Butterworth-Heinemann; 2000. [2] Nelson RM, Hayes KW, Currier DP. Clinical electrotherapy, 3rd ed. Stamford, CT: Appleton & Lange; 1999. [3] Selkowitz DM. Electrical currents. In: Cameron MH, editor. Physical agents in rehabilitation from research to practice. Philadelphia: WB Saunders; 1999, p. 345–427. [4] Lloyd T, De Domeneco G, Strauss GR. A review of the use of electro-motor stimulation in human muscle. Aust J Physiother 1986;32:18–30. [5] Selkowitz DM. High frequency electrical stimulation in muscle strengthening. Am J Sports Med 1989;17:103–11. [6] Ward AR, Shkuratova N. Russian currents: the early experimental evidence. Phys Ther 2002;82:1019–30. [7] De Domenico G. Basic guidelines for interferential therapy. Ryde: Theramed Books; 1981. [8] Savage B. Interferential therapy. London: Faber & Faber; 1984. [9] d’Arsonval A. Action de l’électricité sur les êtres vivants. Exposé des titres et traveaux scientifiques du Dr A. d’Arsonval. Paris: Imprimerie de la cour d’appel; 1894, p. 37–77. [10] d’Arsonval A. Action physiologique des courants alternatifs. Comptes Rendus Société de Biologie, May 2, 1891:283–7. [11] Babkin D, Timtsenko N, (Trans.). Electrostimulation: notes from Dr Y.M. Kots’ (USSR) lectures and laboratory periods presented at the Canadian–Soviet Exchange Symposium on Electrostimulation of Skeletal Muscles, Concordia University, Montreal, Quebec, Canada, December 6–15, 1977 [Available from Dr Ward]. [12] Andrianova GG, Kots YM, Marmyanov VA, Xvilon VA. Primenenie elektrostimuliatsii dlia trenirovki mishechnoj sili. Novosti Meditsinskogo Priborostroeniia 1971;3:40–7. 579 [13] Ward AR, Robertson VJ. The variation in torque production with frequency using medium frequency alternating current. Arch Phys Med Rehabil 1998;79:1399–404. [14] Bankov S. Medium frequency modulated impulse current for electric stimulation of non-denervated muscles. Acta Medica Bulgarica 1980;7:12–7. [15] Moreno-Aranda J, Seireg A. Investigation of over-the-skin electrical stimulation parameters for different normal muscles and subjects. J Biomech 1981;14:587–93. [16] Li CL, Bak A. Excitability characteristics of the A- and C-fibres in a peripheral nerve. Exp Neurol 1976;50:67–79. [17] Ward AR, Robertson VJ. The variation in fatigue rate with frequency using medium frequency alternating current. Med Eng Phys 2001;22:637–46. [18] Ward AR, Robertson VJ. Variation in motor threshold with frequency using kHz frequency alternating current. Musc Nerve 2001;24:1303–11. [19] Bowman BR, McNeal DR. Response of single alpha motoneurons to high frequency pulse trains. Appl Neurophysiol 1986;49:121–38. [20] Jones DA. High- and low-frequency fatigue revisited. Acta Physiol Scand 1996;156:265–70. [21] Jones DA. Muscle fatigue due to changes beyond the neuromuscular junction. In: Porter R, Whelan J, editors. Human muscle fatigue: physiological mechanisms. London: Pitman Medical; 1981, p. 178–96. [22] Johnson M, Ashton CH, Bousfield DR, Thompson JW. Analgesic effects of different frequencies of transcutaneous electrical nerve stimulation on cold induced pain in normal subjects. Pain 1989;39:231–6. [23] Gildemeister M. Untersuchungen über die Wirkung der Mittelfrequenzströme auf den Menschen. Pflügers Arch 1944;247: 366–404. [24] Gildemeister M. Zur theorie des elektrischen Reizes. V. Polarisation durch Wechselströme. Berichte uber die Verhandlungen der Sachsischen Akademie der Wissenschaften zu Leipzig. Mathematisch–Physische Klasse 1930;81:303–13. [25] Sabah NH. Rectification in biological membranes. IEEE Eng Med Biol 2000;1:106–13. [26] Schwarz F, Volkmer H. Über die Summation lokaler Potentiale bei Reizung motorischer Nervenfasern mit elektrischen Wechselimpulsen. Acta Biol Med German 1965;15:283–301.