Annex to IECEx PTB 04.0008X Issue No.: 1 Radar sensor type series VEGAPULS PS6*.CI***H*** or VEGAPULS PS61/62/63.CI***D*** The Radar sensor type series VEGAPULS PS6*.CI***H*** or VEGAPULS PS61/62/63.CI***D*** are used for level measurement in potentially explosive atmospheres. The radar sensor consist of an electronics housing with the corresponding analyzing electronic system, the process connectors and the sensor. In the Radar sensor type series VEGAPULS PS6*.CI***H*** the electronic assembly PS60HC or PS60HK and in the Radar sensor type series VEGAPULS PS61/62/63. CI***D*** the electronic assembly PS60HS is used. The enclosure may be optionally fitted with the control and display module “A/B module” or “PLICSCOM” for either parameterization or visualization. Type code VEGAPULS PS 61. Approval CI IECEx Ex ia IIC T6 Version/Material/Process temperature A with encapsulated horn antenna Process connection/Material Threads, Clamps, Bolting (only metal industry types, 316L) Electronics H 4...20mA HART® D 4...20mA HART® with sensitive electronics P Profibus PA L Profibus PA with sensitive electronics F Foundation Fieldbus K Foundation Fieldbus with sensitive electronics Housing/Protection K Plastic/IP66/IP67 A Aluminium/IP66/IP67 3 Aluminium/IP68 D Aluminium double chamber/IP66/IP67 4 Aluminium double chamber/IP68 V Stainless steel 316L/IP66/IP67 5 Stainless steel 316L/IP68 Cable entry/Plug connection M M20x1.5/without N ½NPT/without Indicating-adjustment module (Plicscom) X without A top mounted PS61. 1 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 VEGAPULS PS 62. Approval CI IECEx Ex ia IIC T6 Version/Material * horn antennas, standpipe, parabolic antenna (only metal industry types, 316L) Process connection/Material Threads, Flanges, swivelling (only metal industry types, 316L) Seal/Process temperature 2 Viton/-40...130°C 3 Kalrez 6375/-20...130ºC 4 Viton/-40...200°C with temperature adapter 5 Kalrez 6375/-20...200°C w. temperature adapter Electronics H 4...20mA HART® D 4...20mA HART® with sensitive electronics P Profibus PA L Profibus PA with sensitive electronics F Foundation Fieldbus K Foundation Fieldbus with sensitive electronics Housing/Protection K Plastic/IP66/IP67 A Aluminium/IP66/IP67 3 Aluminium/IP68 D Aluminium double chamber/IP66/IP67 4 Aluminium double chamber/IP68 V Stainless steel 316L/IP66/IP67 5 Stainless steel 316L/IP68 Cable entry/Plug connection M M20x1.5/without N ½NPT/without Indicating-adjustment module (Plicscom) X A PS62. 2 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 VEGAPULS PS 63. Approval CI IECEx Ex ia IIC T6 Version/Material/Process temperature * with encapsulated horn antennas Process connection/Material Tri-Clamps, Bolting, Flanges, Hyg. connections, Tuchenhagen, Varivent, Neumo biocontrol L (only metal industry types, 316L) Electronics H 4...20mA HART® D 4...20mA HART® with sensitive electronics P Profibus PA L Profibus PA with sensitive electronics F Foundation Fieldbus K Foundation Fieldbus with sensitive electronics Housing/Protection K Plastic/IP66/IP67 A Aluminium/IP66/IP67 3 Aluminium/IP68 D Aluminium double chamber/IP66/IP67 4 Aluminium double chamber/IP68 V Stainless steel 316L/IP66/IP67 5 Stainless steel 316L/IP68 Cable entry/Plug connection M M20x1.5/without N ½NPT/without Indicating-adjustment module (Plicscom) X A PS63. 3 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 VEGAPULS PS 65. Approval CI IECEx Ex ia IIC T6 Version/Material/Process temperature * Rod antennas, PTFE Process connection/Material Threads, Flange (only metal industry types, 316L) Electronics H 4...20mA HART® P Profibus PA F Foundation Fieldbus Housing/Protection K Plastic/IP66/IP67 A Aluminium/IP66/IP67 3 Aluminium/IP68 D Aluminium double chamber/IP66/IP67 4 Aluminium double chamber/IP68 V Stainless steel 316L/IP66/IP67 5 Stainless steel 316L/IP68 Cable entry/Plug connection M M20x1.5/without N ½NPT/without Indicating-adjustment module (Plicscom) X A PS65. 4 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 VEGAPULS PS 66. Approval CI IECEx Ex ia IIC T6 Version/Material * horn antennas, standpipe (only metal industry types, 316L) Process connection/Material Flanges (only metal industry types, 316L) Seal/Process temperature 2 Viton/-40...150°C 3 Kalrez 6375/-20...150°C G Graphite and ceramic -60...250°C H Graphite and ceramic -60...400°C Electronics H 4...20mA HART® P Profibus PA F Foundation Fieldbus Housing/Protection K Plastic/IP66/IP67 A Aluminium/IP66/IP67 3 Aluminium/IP68 D Aluminium double chamber/IP66/IP67 4 Aluminium double chamber/IP68 V Stainless steel 316L/IP66/IP67 5 Stainless steel 316L/IP68 Cable entry/Plug connection M M20x1.5/without N ½NPT/without Indicating-adjustment module (PLICSCOM) X A PS66. For the relationship between the temperature class and the maximum permissible temperature at the sensor and the maximum permissible ambient temperature for the electronic system, reference is made to the following table. Radar-sensors type series VEGAPULS PS6*.CI***H*** temperature class temperature at the sensor T6 -60 ... + 85 °C T5 -60 ... + 100 °C T4* -60 ... +135 °C T3* -60 ... +200 °C T2* -60 ... +300 °C T1* -60 ... +400 °C *from 130 °C with temperature distance piece ambient temperature for the electronic system -40 ... +55 °C -40 ... +70 °C -40 ... +85 °C -40 ... +85 °C -40 ... +85 °C -40 ... +85 °C When the sensors of the VEGAPULS PS6*.*** are operated with higher temperatures than 5 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 indicated in the table above, it shall be guaranteed by suitable measures that no ignition hazard is caused by such hot surfaces. In this case the temperature at the electronics housing shall not exceed the respective values of the table above. For the permissible operating temperatures and pressures, reference shall be made to the specifications provided by the manufacturer. Radar-sensors type series VEGAPULS PS61/62/63.CI***D*** temperature class temperature at the sensor T5 -60 ... +100 °C T4* -60 ... +135 °C T3, T2, T1* -60 ... +200 °C *from 130 °C with temperature distance piece ambient temperature for the electronic system -40 ... +45 °C -40 ... +80 °C -40 ... +85 °C When the sensors of the VEGAPULS PS6*.*** are operated with higher temperatures than indicated in the table above, it shall be guaranteed by suitable measures that no ignition hazard is caused by such hot surfaces. In this case the temperature at the electronics housing shall not exceed the respective values of the table above. For the permissible operating temperatures and pressures, reference shall be made to the specifications provided by the manufacturer. Electrical data Supply and signal circuit (terminals 1 [+], 2 [-] in the compartment, for the 2-cell enclosure version in the terminal compartment) Type of protection Intrinsic Safety Ex ia IIC For connection to a certified intrinsically safe circuit. Maximum values: Ui = 30 V Ii = 131 mA Pi = 983 mW Ci negligibly low or in the version VEGAPULS PS6*.CI***H/D3/4/5*** Ci’core/core = 58 pF/m, Ci’core/screen = 270 pF/m Li negligibly low or in the version VEGAPULS PS6*.CI***H/D3/4/5*** Li’ = 55 µH/m Control and display circuit (terminals Nos. 5,6,7,8 in the electronics compartment or plug connector for the 2-cell enclosure version) Type of protection Intrinsic Safety Ex ia IIC For connection to the intrinsically safe supply and signal circuit of the corresponding external VEGA display unit VEGADIS61 (IECEx PTB 06.0048 X). The rules for interconnection of intrinsically safe circuits between the radar sensors, type series VEGAPULS and the external VEGADIS61 display unit are complied with if the total inductance and capacitance of the connecting line between radar sensors, type series VEGAPULS and the external VEGADIS61 display unit (Lcable = 100 µH and Ccable = 2.8 µF) is not exceeded. 6 of 7 Annex to IECEx PTB 04.0008X Issue No.: 1 A control and display module (A/B module or PLICSCOM) installed in the radar sensors, type series VEGAPULS and a connected VEGACONNECT3 have been considered. Communication circuit (I²C-bus socket in the electronics compartment, additionally for the 2-cell-enclosure version in the terminal compartment) Type of protection Intrinsic Safety Ex ia IIC For connection to the intrinsically safe signal circuit of a VEGA interface converter VEGACONNECT3. Control and display module circuit (spring contacts in the electronics compartment, additionally for the 2-cell-enclosure version in the operating terminal compartment) Type of protection Intrinsic Safety Ex ia IIC For connection to the VEGA control and display module (A/B module or PLICSCOM) With the 2-cell-enclosure version the and display module may either be fitted in the electronics compartment or in the terminal compartment. The metal elements of the radar sensors are electrically connected to the earth terminals. The intrinsically safe circuits are safely electrically isolated from elements that may be earthed. 7 of 7