- Labo America, Inc.

advertisement
eVO 500/500D
Digital Slit Lamp
Contents
1. Introduction
6-7
2. Safety Instruction
8
3. Installation
9-11
4. Operating Procedure
12-13
5. Lamp and Fuse Replacement
14-15
6. Care & maintenance
16-18
7. Technical Specification
19
1 eVO500D
2 Parts Description
1. On/Off Button.
2. Slide Plate
3. Joystick
4. Light Intensity Regulator
5. Guide Rail Cover
6. Instrument Base Locking Knob
7. Microscope arm Locking Knob
8. Illumination arm locking arm
9. Lamp access door
10. Filter Dial
11. Slit Length Dial
12. Slit width rotation knob.
13. Yellow Filter
14. Magnification knob
15. Focusing rod
16. Back Light
17. Target Light
18. Binocular Head Locking Knob
19. Eyepiece
20. SD card slot
21. Monitor
22. Inlet for Lamp Supply
23. AC inlet
24. Inlet for Back light
25. Fuse Holder
3 eVO500
4 Parts Description
1. On/Off Button.
2. Slide Plate
3. Joystick
4. Light Intensity Regulator
5. Guide Rail Cover
6. Instrument Base Locking Knob
7. Microscope arm Locking Knob
8. Illumination arm locking arm
9. Lamp access door
10. Filter Dial
11. Slit Length Dial
12. Slit width rotation knob.
13. Yellow Filter
14. Magnification knob
15. Focusing rod
16. Back Light
17. Target Light
18. Binocular Head Locking Knob
19. Eyepiece
5 1 INTRODUCTION
Congratulations on the purchase of your new eVO500/ eVO500D!
This instruction manual is designed as a training and reference manual for the operation and
maintenance of your instrument. We recommend that you read it carefully prior to use and
follow the instructions to ensure optimum performance of your new instrument.
Please retain this manual for future reference and to share with other users. Additional copies
can be obtained from your authorized LABOMED dealer or from the LABOMED service
department. Contact information is provided at the end of this manual.
LABOMED eVO500/ eVO500D Slit Lamp is an all-in-one digital ophthalmic instrument for
observation, measurement and diagnosis of Contact lens fitting, Anterior eye segment,
Cataract, Glaucoma, Retina etc.
Its optimally structured design also allows the eVO500/ eVO500D to be used in laser therapy;
this has been achieved by the possibility of operating the Slit projector in the central position
either with the left hand or right hand. The excellent transmission of the optical system
minimizes light loss during therapy. Together with the excellent image quality, this provides
ideal conditions for crisp, crystal clear images of even the finest structure.
Considering standard work procedures, the design of this system have been modified to carry
an in-builtcamera. The integrated module is optically corrected unit, to deliver the best results
in its class. An eVO500D also has a micro SD slot and an HD touch screen panel, which provides
you with an optimum, ergonomically corrected configurations and is low on fatigue while in use.
The perfection of the observation system along with an outstanding operating convenience
ideally complements each other. The outstanding benefits of this design include an effortless
operation, short working distance, comfortable arm and body posture, easy grip controls for the
sensitive setting of the Slit length, width and rotation. Using the joystick, you can control the
eVO500/ eVO500D quickly and precisely in all three coordinates and set it exactly in any
position with a quick action locks. The joystick also has a push button at the top which is very
easy to use to click photos and videos in eVO500D.
6 Its slit angle of incidence makes it easier to view the inner side by rotating the illumination arm
(0° - 90°), giving it a 3 dimensional rigid view of the eye interrelated with slit housing. It makes
and rotate the slit at its 0° - 180° position with fine click stop smooth movement option for the
corneal view, vertically/horizontally reducing or increasing the size of slit from 0-14mm.With the
continuous slit measurement of 1.5 to 11mm and variable fixed measurement at 0.5, 5, 8,
12mm it has enhanced red free, cobalt blue and heat absorbing filters.
The user expects the slit lamp microscope to provide optimum stereoscopic observation with
selectable magnification. The size of the field of view and the depth of field are expected to be
as large as possible, and there should be enough space in front of the microscope for
manipulation on the eye. eVO500/ eVO500D is a powerful instrument with 5 step magnification
(6.5x, 10x, 16x, 25x& 40x) in combination with a binocular straight head, 12.5x/16mm wide
field eyepieces (Diopter adjustment +/-6 ) and a smooth adjustable IPD 48.5-80mm.
An eVO500D slit lamp is designed for all digital imaging requirements. Its fatigueless
examination with integrated 5.0 mega pixels (1080p) digital module supporting, image
capturing and option of recording live on the screen panel in HD quality makes the job even
easier while performing examination on the spot. Coupled with an integrated capture button to
capture images and make videos which can often be useful to explain to the patient, findings or
the condition and fit of a contact lens during an examination. This saves time-consuming
theoretical explanations later. This modern technique is well suited for documentation,
information and educational purposes and due to its low weight. It does not impair the mobility
and ease of handling of the slit lamp.
NTRODUCTION
7 2 Safety Instructions
DUCTION
•
This instrument is manufactured according to the safety norms as per CE regulations.
•
This instrument is intended for use only as prescribed in this manual.
•
Servicing and repair are allowed through authorized persons only.
•
Replace burnt out fuses by fuses of the same type only.
•
Do not keep SD cards in the slots for longer periods.
•
Do not touch or clean the panel with wet or rough cloth.
•
Use main plug and main socket with protective earth conductors only.
•
Do not use force while fixing connectors. If the male and female plugs do not readily
connect, make sure that they are appropriate for one another. If any of the connectors
are damaged, please contact LABOMED representative.
•
This instrument is meant to be used in dry rooms only. Care must be taken to ensure
that no product containing liquid be kept on the instrument at any time to avoid spillage
and seepage of liquids inside the instrument.
•
The manufacturer will not accept any liability for damage caused by unauthorized
persons tampering with the instrument; this will also forfeit any rights to claim warranty.
•
It is recommended to use the instrument only with the accessories supplied. In case you
wish to use other accessory make sure that LABOMED has certified that its use will not
harm the safety of the instrument.
NOTE : Its requested to retain in the Serial Number of the instrument for identification
by the service personnel.
8 3 Installation
eVO500/eVO500D Slit Lamp is shipped in a multi-layer foam container and it is divided
into major components: Microscope body, Binocular straight head, Base Platform, HD
Panel and Chin Rest Assembly, Illumination Column.
1. Take out the various parts after removing
the wrapping and save the packing.
Fig. 1
2. Mount the chin rest at the base platform
with the socket-head screws, hexagonal
wrench (provided) as shown in the Fig 2.
3. Plug the wire from the chin rest assembly
into the 5 pin connector on the back of the
power supply unit.
Fig. 2
4. Mount the base platform on appropriate
instrument stand arm or table base.
5. Position the Slit Lamp body gently on the
base platform so that the geared rollers
match with the teeth on the guide rails.
When the instrument is pushed forward or
pulled backwards, the wheels should roll
freely and evenly over the guide-rails
(Refer Figure 3).
Fig. 3
9 6. Tighten
the
microscope
base
locking
screw. Place the guide-rail covers over the
guide-rails (Refer Fig. 4).
Base Locking Knob
Rail Cover
Fig. 4
7. Mount
the
microscope
head
on
the
microscope carrying arm by sliding it into
position, making sure it is up against the
stop. Then, tighten lock-knob located on
the top of the microscope (Refer Fig.5).
Fig. 5
8. Plug the microscope illumination cable with
4 pin adaptor into the connector on the
back of the power supply as well as on the
back side of the microscope base (Refer
Fig. 6).
9. Connect the 3 pin inlet power cord to
power supply and plug into a wall outlet or
Fig.6
the motorized table stand arm.
10 10. Mount the touch screen panel on the
microscope arm along with the knob in the
mount. Screw the mount to the knob to
hold it tight incase of eVO500D(See Fig 7).
11. Connect the USB output at the back of the
panel and the power cord incase of
eVO500D
Fig. 7
11 4 Operating Procedure
1.
Turn the power using the ON/OFF switch located on the front of the power supply.
Brightness can be adjusted by rotating the illumination level knob situated on the
microscope base.
NOTE: The maximum position is for intermittent use only.
Continuous use will
shorten lamp life.
2.
Insert the projection rod in the pivot post of the instrument body to make rough PD
and focus adjustments. Position the light onto the flat surface of the projection rod
and adjust the pupillary distance and focus of the eyepieces till a comfortable
viewing is obtained. Remove the projection rod.
3.
To position a patient, adjust the chinrest height by turning the control knob on the
post of the chin rest assembly until the patient’s canthus is in line with the canthus
mark on the chin rest post.
4.
Microscope elevation is adjusted by rotating the joystick and observing the Slit
image through the microscope until the slit is centered on the patient’s cornea.
5.
Move the Slit lamp with the joystick held firmly and slightly angled towards the
patient, until the slit appears sharply on the cornea. The accuracy of this rough
adjustment should be checked by the naked eye. The fine adjustment is performed
while observing the slit through the microscope.
6.
Tilt the joystick, which is now held lightly at its upper end, until the slit appears
sharply at the depth of the eye which is to be observed. The horizontal motion of
the base can be locked by tightening the base locking screw. Lock the base
whenever the slit lamp is not in use.
7.
The slit width can be adjusted by rotating the slit width control knob on either sides
of the instrument.
8.
The angle between the illumination system and the microscope can be varied
between 0° to 90° to either the left or to the right. The angle is indicated on the
scale of the slit lamp arm.
12 9.
Magnification is altered by rotating the dial on the microscope head. The
magnification of each click stop position is engraved on the dial.
10. Three filters are provided for cool and comfortable view. This can be achieved by
rotating the filter dial.
An eVO500 is a basic and eVO500D is the high end model. So, technically what
separates evo500D from evo500 is the broad spectrum of optional accessories that
offer solutions best suited for the comfort level of both-the patient and the user of
the slit lamp. Following are the key features worth mentioning which makes an
eVO500D superior over its basic version eVO500.
1.
Mount the touch screen panel that allows the user to view live images. Use click
button provided on the joystick to capture images and record videos.
2.
Select the options on the touch screen panel as per your requirements. These
options are “Capture”, “Zoom In/Out” & “Exit”. A Capture option once selected
will show “Capture image” and “Capture Video”, this works only if the SD card is
inserted in the memory card slot. Exit selection will make all the options disappear
on the panel.
3.
Insert an SD card to capture live images and recording videos which can be used
later on to retrieve images and videos obtained for the patient with the slit lamp.
4.
A PixelPro* control software supports all the above features meant for eVO500D slit
lamp when attached with USB port.
13 5 Lamp & Fuse Replacement
1. Open the lamp house by loosen the
screw (1). See figure 8.
1
Fig. 8
2. Take out the bulb holder (1) by pulling
the metal pins on the holder(see fig 9)
1
Fig. 9
3. Loosen the screw and pull the bulb out
of the house assembly.(Refer Fig. 10)
1
Fig. 10
14 4. Remove the bulb from the housing.
Replace the blown lamp with new one.
Insert the bulb in the housing and
secure the same in the housing.
5. Insert the holder in the bulb slowly and
tighten the lamp cover after putting it
in its place.
Fig. 11
Replacement of Fuse
6. Open the fuse holder provided on the
back side of the power supply. See fig.
12
Fig. 12
7. One spare fuse is being provided in
the box. Replace the dead fuse with
the live one. Insert the fuse holder
back to it’s position.
Fig. 13
15 6 Care & Maintenance
•
Maintenance
1. Headrest/Chinrest Cleaning
For hygienic reasons, wipe the headrest and chin rest with an alcohol, wipe after every patient.
2. Changing Illumination Lamp
Ensure the Slit Lamp is switched off and remember that some parts may be hot. Do not
touch the glass of the new bulb.
Open the bulb access door. Swing the retaining spring towards the arm pivot and pull
the socket and bulb from the housing. Note the position of the notch in the metal collar
of the bulb. Replace the bulb with a new one and make sure the notch is in the same
position. Insert the bulb and the socket into the lamp housing and move the retaining
spring back into position. Close the bulb access door.
3. Projection Pros, Cleaning
Grasp the narrow shank of the projection prism and pull upwards. When cleaning the
prism, blow with clean dry air, then, gently wipe with a soft linen cloth.
4. Cleaning the Slide Plate
If the plate is dirty, it may cause a rough feeling when maneuvering the base of the
Slit Lamp. Clean the slide plate with soft cloth lightly dampened with a mild soap and
water solution.
CONTINUOUS POWER SUPPLY
The LABOMED eVO500/eVO500D Slit Lamp can be operated from 100 to 240V AC. This
is achieved by universal power supply. So, there is no need of adjusting the settings
according to the supply line voltage of the destination country.
16 NOTE: Two fuses are used in this container, one is used in the unit and the other one
is supplied as a spare part. It’s recommended to replace the blown out fuse with the
fuse of same rating and type i.e. T2 5L 250
More information on replacing the fuses can be found in the maintenance section of this
guide.
•
Care instructions
o
Keep accessories away from dust when not in use, e.g. protect them from dust
cover.
o
Remove dust with a pneumatic rubber bulb and soft brush.
o
Use special optics cleaning cloths and pure alcohol for cleaning lenses and
eyepieces.
o
Protect your slit lamp from moist, fumes, acids and cosmetic materials. Do not
store chemicals close to the instrument.
o
Protect it from improper handling. Never install other devices sockets or unscrew
optical system and mechanical parts unless explicitly instructed to do so in this
manual.
o
Protect the microscope from oil and grease. Never oil or grease the guide
surfaces or mechanical parts.
o
Remove coarse contamination using a damp disposable cloth.
o
Use disinfectants based on the following ingredients: aldehydes, alcohols,
quaternary ammonium compounds.
o
Camera: Keep optical components clean using a lint-free cloth. Soak the cloth
using a little methanol or glass cleaner. Do not use ethanol and spirit.
o
Do not clean products with optical components in a cleaning/disinfecting device
or ultrasonic bath.
o
LABOMED MaxLite coatings are very resistant. If you clean as described
above, the coatings will not be damaged.
17 •
Tropical environment/fungus:
LABOMED employs certain safety precaution in its manufacturing techniques and
materials. Other preventive measures include:
o
Keep optical part clean.
o
Use and store them in a clean environment only.
o
Store under UV light when not in use.
o
Use in continuously climate-controlled rooms only.
o
Keep moisture away and cover using a plastic cover filled with silica gel.
•
Occupational safety and health protection:
Observe work safety and health protection of persons responsible for processing
contaminated products.
Current regulations of hospital hygiene and prevention of infection must be
observed in the preparation, cleaning and disinfection of the products.
Instructions
•
Workplace:
Remove surface contamination with a paper towel.
•
Reprocessing:
Recommended: reprocess a product immediately after use.
•
Cleaning:
Needed: water, detergent, spirits, microfiber cloth
o
Flush the surface with running water (<40˚ C), using a little detergent if
necessary.
o
Also use spirits to clean optical components.
o
Dry optical components using a microfiber cloth, dry the rest of the product
using a paper towel.
•
Service:
Service as and when required should be informed to LABOMED after – sales
service department.
18 7 Technical Specification
Item
Specification
eVO 500
•
Microscope
Galilean
Galilean
•
Magnification
5 step rotation: 6.5x, 10x,
16x, 25x, 40x
5 step rotation: 6.5x, 10x, 16x,
25x, 40x
•
Eyepieces
•
Interpupillary
Distance
Wide field 12.5x, focusable,
anti fungus
48.5 – 80mm, Diopter
adjustment ± 6
Wide field 12.5x, focusable,
anti fungus
48.5
–
80mm,
Diopter
adjustment ± 6
•
•
•
Slit Length
Slit Width
Slit Apertures
•
•
Slit Rotation
Filters
•
Longitudinal
Movement
Lateral
Movement
Vertical
Movement
Chin Rest Range
Illumination
Digital Imaging
0 – 14mm
0 – 14mm
0.5, 5, 8, 12mm, 1.5-11mm
continuous,
slit
measurement
0 – 180°
Red-Free, Cobalt Blue, Heat
absorbing
90mm
0 – 14mm
0 – 14mm
0.5, 5, 8, 12mm, 1.511mm
continuous,
slit
measurement
0 – 180°
Red-Free, Cobalt Blue,
Heat absorbing
90mm
100mm
100mm
30mm
30mm
80mm
6V 20W Halogen
80mm
6V 20W Halogen
Integrated 5 MP Imaging
Touch screen panel for live
image display and capture
1080p HDMI output.
USB v2.0 output for PC
connectivity Integrated Capture
Button.
Online capturing on integrated
SD Card PixelPro* Camera
Control Software.
•
•
•
•
•
eVO 500D
19 20 