Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis A51969AD January 2014 Beckman Coulter, Inc. 250 S. Kraemer Blvd. Brea, CA 92821 U.S.A. Application Guide PA 800 plus Pharmaceutical Analysis System Carbohydrate Labeling and Analysis A51969AD (January 2014) © 2009-2014 Beckman Coulter, Inc. All rights reserved. Made in U.S.A. Beckman Coulter, Inc. grants a limited non-exclusive license to the owner or operator of a PA 800 plus instrument to make a copy, solely for laboratory use, of any portion or all of the online help and electronic documents shipped with the PA 800 plus instrument. Beckman Coulter and the stylized logo are trademarks of Beckman Coulter, Inc. and are registered in the USPTO. No part of this document may be reproduced or transmitted in any form or by any means, electronic, mechanical, photocopying, recording, or otherwise, without prior written permission from Beckman Coulter, Inc. All other trademarks, service marks, products, or services are trademarks or registered trademarks of their respective holders. Find us on the World Wide Web at: www.beckmancoulter.com Printed in U.S.A. Revision History Initial Issue, A51969AA, April 2009 32 Karat Software version 9.1 PA 800 plus Software version 1.1 PA 800 plus Firmware version 9.0 First Revision, A51969AB, December 2009 Revised corporate address. Second Revision, A51969AC, February 2011 32 Karat Software version 9.1 patch PA 800 plus Software version 1.1 patch PA 800 plus Firmware version 9.2 Numerous syntax and grammatical edits Third Revision, A51969AD, January 2014 Formatting updates, Aminopyrene Trisulfonic Acid naming standardization. A51969AD iii Revision History iv A51969AD Safety Notices Symbols and Labels Introduction The following is a description of symbols and labels used on the Beckman Coulter PA 800 plus Pharmaceutical Analysis System or shown in this manual. WARNING If the equipment is used in a manner not specified by Beckman Coulter, Inc., the protection provided by the instrument may be impaired. General Biohazard Symbol This caution symbol indicates a possible biohazard risk from patient specimen contamination. Caution, Biohazard Label This caution symbol indicates a caution to operate only with all covers in position to decrease risk of personal injury or biohazard. A51969AD v Safety Notices Symbols and Labels Caution, Moving Parts Label This caution symbol warns the user of moving parts that can pinch or crush. CAUTION PARTS MOVE AUTOMATICALLY A015081L.EPS High Voltage Electric Shock Risk Symbol This symbol indicates that there is high voltage and there is a risk of electric shock when the user works in this area. 144557-AB A016352L.EPS Class 1 Laser Caution Label A label reading “Complies with 21 CFR 1040. 10 and 1040.11 except for deviations pursuant to Laser Notice No. 50, dated June 24, 2007” is found near the Name Rating tag. The laser light beam is not visible. vi A51969AD Safety Notices Symbols and Labels Sharp Object Label A label reading “CAUTION SHARP OBJECTS” is found on the PA 800 plus. CAUTION SHARP OBJECTS A16558-AA A016351L.EPS Recycling Label This symbol is required in accordance with the Waste Electrical and Electronic Equipment (WEEE) Directive of the European Union. The presence of this marking on the product indicates: 1. The device was put on the European Market after August 13, 2005. 2. The device is not to be disposed of via the municipal waste collection system of any member state of the European Union. It is very important that customers understand and follow all laws regarding the proper decontamination and safe disposal of electrical equipment. For Beckman Coulter products bearing this label, please contact your dealer or local Beckman Coulter office for details on the take back program that facilitates the proper collection, treatment, recovery, recycling, and safe disposal of this device. Disposal of Devices Containing Mercury Components This product contains a mercury-added part. Recycle or dispose of according to local, state, or federal laws. It is very important that you understand and comply with the safe and proper disposal of devices containing mercury components (switch, lamp, battery, relay, or electrode). The mercury component indicator label can vary depending on the type of device. A51969AD vii Safety Notices Symbols and Labels Restriction of Hazardous Substances (RoHS) Labels These labels and materials declaration table (the Table of Hazardous Substance’s Name and Concentration) are to meet People's Republic of China Electronic Industry Standard SJ/T11364-2006 “Marking for Control of Pollution Caused by Electronic Information Products” requirements. China RoHS Caution Label — This label indicates that the electronic information product contains certain toxic or hazardous substances. The center number is the Environmentally Friendly Use Period (EFUP) date, and indicates the number of calendar years the product can be in operation. Upon the expiration of the EFUP, the product must be immediately recycled. The circling arrows indicate the product is recyclable. The date code on the label or product indicates the date of manufacture. China RoHS Environmental Label — This label indicates that the electronic information product does not contain any toxic or hazardous substances. The center “e” indicates the product is environmentally safe and does not have an Environmentally Friendly Use Period (EFUP) date. Therefore, it can safely be used indefinitely. The circling arrows indicate the product is recyclable. The date code on the label or product indicates the date of manufacture. viii A51969AD Safety Notices Symbols and Labels Alerts for Warning, Caution, Important, and Note WARNING WARNING indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. The warning can be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products). CAUTION CAUTION indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. The caution can be used to indicate the possibility of erroneous data that could result in an incorrect diagnosis (does not apply to all products). IMPORTANT IMPORTANT is used for comments that add value to the step or procedure being performed. Following the advice in the IMPORTANT notice adds benefit to the performance of a piece of equipment or to a process. NOTE NOTE is used to call attention to notable information that should be followed during installation, use, or servicing of this equipment. A51969AD ix Safety Notices Symbols and Labels x A51969AD Contents Revision History, iii Safety Notices, v Symbols and Labels, v Introduction, v General Biohazard Symbol, v Caution, Biohazard Label, v Caution, Moving Parts Label, vi High Voltage Electric Shock Risk Symbol, vi Class 1 Laser Caution Label, vi Sharp Object Label, vii Recycling Label, vii Disposal of Devices Containing Mercury Components, vii Restriction of Hazardous Substances (RoHS) Labels, viii Alerts for Warning, Caution, Important, and Note, ix Carbohydrate Labeling and Analysis Kit, 1 Overview, 1 Safety Information, 1 General Precautions, 1 Intended Use, 2 Materials and Reagents, 2 Contents of this Kit, 2 Materials Needed but Not Provided in Kit, 2 Storage Conditions, 3 Overview of the Procedure, 3 Releasing Oligosaccharides From Glycoproteins, 4 Enzymatic Release of the N-Linked Oligosaccharides, 4 Working with the Supernatant, 4 Principle of the Labeling Method, 5 Preparing the Labeling Reagents, 5 Choosing the Proper Fluorophore, 5 Preparing the Reagents, 6 Preparing the Standards, 6 xi Contents Performing the Labeling Reaction, 7 Preparing the PA 800 plus Instrument, 9 Installing and Pre-conditioning the N-CHO Capillary, 10 Clean the Interface Block, 11 Insert the Cartridge, 11 Preparing the Buffer Trays, 11 Sample Vial Setup, 12 Running the Samples, 14 Initial Conditions, 14 LIF Detector Initial Conditions, 14 Time Program, 15 Evaluation of Results, 18 System Shutdown and Capillary Storage, 18 Short Term Storage (<24 hours) of Capillary, 18 Long Term Storage (>24 hours) of Capillary, 18 Troubleshooting Procedures, 19 Appendix, 20 LIF Detector Calibration Procedure, 20 Setting the Calibration Corrector Factors (CCFs), 20 Automatic Calibration, 22 CCF Troubleshooting Procedures, 25 CCF <1, 25 CCF >1, 25 No Step Change Detected, 25 xii Carbohydrate Labeling and Analysis Kit Overview Beckman Coulter’s Carbohydrate Labeling & Analysis Kit uses capillary electrophoresis to separate and quantify oligosaccharides released from glycoproteins. This kit has been developed to work with the PA 800 plus Pharmaceutical Analysis System, using the 488 nm laser and the 520 nm emission filter. This kit contains the reagents, buffer, and coated capillaries required to label, separate, and quantify oligosaccharides. This kit also provides a glucose size marker for relative size determination and a maltose standard for quantitation and mobility characterization of the released oligosaccharides. For further information, please refer to the following publication: Quantitative Analysis of Sugar Constituents of Glycoproteins by Capillary Electrophoresis, Fu Tai, A. Chen, Thomas S. Dobashi and Ramon A. Evangelista, “Glycobiology”, Vol. 8, No. 11, pp. 1045-1052, 1998. Safety Information NOTE This application guide has been validated for use in the PA 800 Enhanced Protein Characterization and the PA 800 plus Pharmaceutical Analysis Systems. IMPORTANT In this procedure 1 M sodium cyanoborohydride in THF is used, which is a very reactive compound. Please consult the MSDS before using or disposing of this reagent. CAUTION When installing or removing capillaries from the cartridge, use safety goggles. General Precautions • Prior to using the instrument, turn on the laser and let it stabilize for 15 minutes. Turn it off when not in use for longer periods of time. • Do not interchange the capillary (i.e. if a capillary is being used for carbohydrate analysis, do not use the same capillary in another application). A51969AD 1 Carbohydrate Labeling and Analysis Kit Materials and Reagents • When the capillary is not in use or after finishing with all experiments, it is recommended to run a shutdown method to rinse the capillary and leave the ends immersed in water to avoid capillary blockage and to keep the capillary from drying out. • Prior to using the capillary for the first time, it is recommended to rinse the capillary for 10 minutes at 30 psi pressure with DDI water and then rinse for 10 minutes at 30 psi with the Carbohydrate Separation Gel Buffer. Intended Use The Carbohydrate Labeling and Analysis Kit is for laboratory use only. CAUTION Refer to the Material Safety Data Sheets (MSDS) information, available at www.BeckmanCoulter.com, regarding the proper handling of materials and reagents. Always follow standard laboratory safety guidelines. Materials and Reagents Contents of this Kit Table 1 Kit Contents Reorder Number (477600) Component Quantity Reorder # 56 mL 477623 2 477601 Labeling Dye (APTS) 4x5 mg 501309 Labeling Dye Solvent 1 mL L3 50 mg G20 0.18 mg G22 Carbohydrate Separation Buffer N-CHO Coated Capillary Glucose Ladder Standard Quantitation/Mobility Marker (Maltose) APTS-M (monosaccharide grade) 20 mg Code L6 725898 Materials Needed but Not Provided in Kit Description Part Number LIF Performance Test Mix 726022 Universal Plastic Vials (pack 100) A62251 Universal Rubber Vial Caps - blue (pack 100) A62250 200 µL Micro Vials (pack 50) 144709 Assorted pipettors and pipette tips 2 A51969AD Carbohydrate Labeling and Analysis Kit Overview of the Procedure Description Part Number Spatula Analytical scale Vortexer Bench top micro-centrifuge Water bath Centrifugal vacuum evaporator Sodium Cyanoborohydride, 1 M/THF Sigma-Aldrich 296813 2-mercaptoethanol Sigma-Aldrich M7154 Double Deionized (DDI) water filtered through 0.2 µm filter N-glycosidase F (PNGase F) New England Biolabs P0704S Nonidet® NP-40 non ionic detergent Storage Conditions The kit components are stable for one year when stored under the following conditions: • Upon receipt, the kit should be stored at 2 to 8°C. • Reconstituted Quantitation Control (G22) and labeled glucose ladder standard (G20) should be stored at -35 to -15°C. • Reconstituted labeling reagent (APTS) shall be stored at -35 to -15°C. It is stable for 2 weeks after reconstitution. Overview of the Procedure The carbohydrate analysis involves the following three tasks: • Enzymatic release of the N-linked oligosaccharides from the glycoprotein. • Labeling of the mixture of released oligosaccharides. • Analysis of the flouorophore-labeled oligosaccharides by capillary electrophoresis. A51969AD 3 Carbohydrate Labeling and Analysis Kit Releasing Oligosaccharides From Glycoproteins Releasing Oligosaccharides From Glycoproteins IMPORTANT This labeling kit does not contain releasing enzymes. There are multiple enzymatic and chemical procedures for releasing oligosaccharides from proteins. In order to successfully label the released oligosaccharide, it is essential that the reducing termini of the oligosaccharide is not destroyed by the deglycosylation method. The following is a suggested protocol for N-deglycosylation, using N-Glycosidase F (PNGase F): Enzymatic Release of the N-Linked Oligosaccharides 1. Aliquot the sample volume in order to have between 25-300 µg of glycoprotein. 2. Dry the glycoprotein solution completely in a centrifugal vacuum evaporator. 3. Add 45 µL of 1X PBS buffer. 4. Add 1.0 µL of 5% SDS (or a volume that gives final concentration of 0.1% SDS). 5. Add 1.5 µL of 1:10 dilution of 2-mercaptoethanol in DDI water (or a volume that gives final concentration of 50 mM 2-mercaptoethanol). 6. Heat the sample for 5 minutes at 100°C. IMPORTANT If the denatured protein precipitates, discard the sample and restart this process, repeating steps 1 through 5. For step 6, denature the protein by incubating it at 37°C for 10 minutes. 7. Cool the sample to room temperature. 8. Add 5 µL of Nonidet® NP40 non-ionic detergent (or a sufficient volume that gives a final concentration of 0.75%). 9. Add PNGase F enzyme, according to its activity. 10. React for 15 hours at 37°C. 11. Add approximately 150 µL of cold ethanol (or 3 times the actual reaction mixture volume). 12. Vortex the mixture and place it on ice for complete protein precipitation. 13. Centrifuge the samples at 15,000 rpm for 5 minutes. 14. Withdraw and save the supernatant. IMPORTANT This solution contains the RELEASED N-LINKED OLIGOSACCHARIDES. 15. Discard the solid. This is the deglycosylated protein. Working with the Supernatant There are two options for working with the supernatant: For quantitative analysis, add an internal standard to the supernatant. See Preparing the Standards beginning with Step 1a. 4 A51969AD Carbohydrate Labeling and Analysis Kit Preparing the Labeling Reagents For qualitative analysis, bring the supernatant to dryness in the centrifugal vacuum evaporator. These oligosaccharides are ready to be labeled with APTS. See Performing the Labeling Reaction in this chapter. Principle of the Labeling Method After release (enzymatic or chemical), the oligosaccharides can be labeled with a fluorophore called 1-Aminopyrene-3,6,8 – Trisulfonic Acid (APTS). The stoichiometry of labeling reaction is one APTS molecule per molecule of oligosaccharide. Figure 1 illustrates the labeling reaction of an N-linked oligosaccharide with APTS. Figure 1 Labeling Reaction of an Oligosaccharide with APTS The efficiency of the labeling reaction is dependent on temperature and the amount of oligosaccharides. This protocol has been optimized for labeling 5 nmoles or less of total oligosaccharides. Samples with amounts higher than 5 nmoles may give a lower reaction yield. Use maltose (G22) as an internal labeling control and/or as an internal mobility marker. Preparing the Labeling Reagents Choosing the Proper Fluorophore APTS-M is a high-purity fluorophore intended for labeling monosaccharides. APTS-M contains citric acid as a catalyst. APTS (L6) can be used for oligosaccharides. A51969AD 5 Carbohydrate Labeling and Analysis Kit Preparing the Labeling Reagents Preparing the Reagents 1 Prepare the APTS labeling reagent. To a vial of APTS labeling dye (L6) a. Add 48 µL of Labeling dye solvent (15% acetic acid – L3). b. Vortex for 5 seconds until complete dissolution. c. Store at -35 to -15°C for up to 2 weeks when not in use. 2 Prepare the APTS-M (Monosaccharide grade) Labeling Dye a. Add 400 µL of DDI water to the APTS-M vial. b. Vortex the mixture for 5 seconds until all of the solid is dissolved. c. Store the reconstituted APTS-M at -35 to -15°C for up to two weeks. Preparing the Standards 1 Prepare the Maltose Quantitative Control/Mobility Marker (G22). a. Add 500 µL of DDI water to the Maltose quantitative control labeled G22. This makes a solution containing 500 nmoles of maltose, or concentration of 1 nmol/µL. b. For quantitation purposes, add 5 µL of G22 reconstituted to your released oligosaccharides. c. Dry down the sample in a centrifugal evaporator and proceed to labeling protocol, as described in the next section. d. Store at -35 to -15°C when not in use. NOTE When using the maltose solution for quantitation purposes, make sure that the solution is prepared fresh and used right away. Bacterial contamination can digest the sugars, leading to a miscalculation of your quantitation. 2 Prepare the Glucose Ladder Standard (G20). a. Weigh and dissolve 5 mg of Glucose Ladder Standard (G20) in 80 µL DDI water in 1.5 mL microcentrifuge tube. Sonicate if necessary. b. Aliquote at least ten 2 µL portions of the Glucose ladder standard solution to 0.5 mL microcentrifuge vials and dry them in a centrifugal vacuum evaporator. The dried glucose ladder can be stored at room temperature or used immediately. 6 A51969AD Carbohydrate Labeling and Analysis Kit Performing the Labeling Reaction Performing the Labeling Reaction 1 Label the Sample and/or Standard (G20). a. Add 2 µL of 1 M sodium cyanoborohydride/THF to the dried oligosaccharide sample. b. Add 2 µL of APTS Labeling Reagent to the sample. IMPORTANT Due to the reaction of sodium cyanoborohydride with water, avoid moisture and store this product under dry conditions. Use a dry needle to dispense this chemical. Pass dry argon into the chemical bottle through the septum while dispensing 2 µL from the bottle. c. Incubate the reaction mixture at 37°C for 4 hours, or at 60°C for 90 minutes, or at room temperature overnight (mildest reaction). NOTE A mild labeling condition is necessary to avoid losing the sialic acid residues from your oligosaccharide chain. You can leave the reaction mixture at room temperature overnight and in the dark to do this. For Glucose Ladder incubation, select any of the temperature/ time options. Figure 2 illustrates a summary of the overall labeling protocol. 2 Prepare the Sample and/or Standard for capillary electrophoresis. a. To the 4 µL reaction mixture prepared above add: • For Glucose Ladder Standard: add 96 µL of DDI water • For all other samples: add 46 µL DDI water This solution can be stored at 2 to 8°C. b. Take an aliquot of 5 µL of the solution above and add 195 µL of DDI water. c. Mix the contents well and transfer the solution into a micro vial for analysis in the PA 800 plus system see Sample Vial Setup. Figure 3 illustrates a summary of the sample preparation for injection. A51969AD 7 Carbohydrate Labeling and Analysis Kit Performing the Labeling Reaction Figure 2 Labeling Reaction Scheme 8 A51969AD Carbohydrate Labeling and Analysis Kit Preparing the PA 800 plus Instrument Figure 3 Sample Preparation for Injection Preparing the PA 800 plus Instrument Before proceeding, review the following procedures in the PA 800 plus System Maintenance User’s Guide (A51964AD). • Install the LIF Detector • Calibrate the LIF Detector • Capillary Cartridge Procedures • Cleaning the Interface Block and the Opening Levers A51969AD 9 Carbohydrate Labeling and Analysis Kit Installing and Pre-conditioning the N-CHO Capillary Installing and Pre-conditioning the N-CHO Capillary • Prior to using the PA 800 plus system for carbohydrate analysis, the N-CHO capillary must be installed in a capillary cartridge. To properly install the capillary, refer to the Capillary Cartridge sections in the PA 800 plus System Maintenance User’s Guide (A51964AD). IMPORTANT Be sure to pre-rinse a new capillary for 10 minutes at 30 psi pressure with DDI water and then rinse for 10 minutes at 30 psi with Carbohydrate Separation Buffer prior to the first run. For this analysis, you will be using a 50 µm I.D. N-CHO capillary. The length from injection inlet to detector should be 40 cm, with the total length being 50.2 cm. Be sure to measure the capillary dimensions accurately and record the dimensions in the capillary/performance tab of the Advanced Method Options. See Figure 4. This will be necessary for accurate mobility determinations. Figure 4 Capillary Performance Screen IMPORTANT To get good reproducibility from capillary to capillary and accurate mobility assignments, it is important to adhere to the capillary pre-measurement procedure. IMPORTANT The cut ends of capillaries should be inspected carefully under magnification. The cut must be clean (not jagged) and perpendicular to the capillary length (not angled). Poor cuts will result in poor resolution and poor sample loading. NOTE When not in use, submerge the capillary ends in DDI water to prevent the capillary from drying out. • Turn off the PA 800 plus system and install the LIF detection components. • Turn on the system and permit the laser to warm up for at least 30 minutes prior to calibration. 10 A51969AD Carbohydrate Labeling and Analysis Kit Preparing the Buffer Trays Clean the Interface Block Clean the electrodes, opening levers, capillary tips, and interface block carefully following the cleaning procedure as described in the PA 800 plus “System Maintenance” manual. Do this on a weekly basis or when changing chemistries on the PA 800 plus system. Accumulation on the interface block can lead to current leakage errors. Insert the Cartridge Insert the cartridge into the system. Close the cartridge and sample covers. Preparing the Buffer Trays 1 Perform the LIF Detector Calibration Procedure in the Appendix. 2 Fill the vials with the appropriate volumes of each reagent: • 1.5 mL of DDI water per H20 vial (4 vials) • 1.5 mL of Carbohydrate Separation Buffer per Gel-R vial (1 vial) • 1.3 mL of Carbohydrate Separation Buffer per Gel-S vial (2 vials) • 0.8 mL of DDI water per waste vial (1 vial) Figure 5 Universal Vials and Caps 1. Universal Vial Cap 2. Maximum Fill Level 3. Universal Vials A51969AD 11 Carbohydrate Labeling and Analysis Kit Sample Vial Setup 3 Place the reagent vials on the inlet and outlet buffer trays as indicated in Figure 6. NOTE During the separation, salts migrate from one vial of gel buffer to the other, changing the composition of the buffer. Therefore, it is recommended that the vials of buffer be replaced with fresh buffer after 20 runs. For unattended operation of more than 20 runs, simply program the increment function to increment the water and buffer vials after 20 runs. Vials will increment in a forward direction. Ensure that the tray is filled with the appropriate number of vials to manage the number of runs that are to be performed. NOTE The H20 vials are used to clean the capillary tips after the sample introduction step to prevent sample carryover. Replace the vials with fresh water if rinse water becomes contaminated. A schematic of the buffer tray setup is illustrated in Figure 6. Figure 6 Buffer Tray Configuration Sample Vial Setup Before placing the 200 microliter sample vials into the universal vials, ensure that no bubbles are at the bottom of the sample vials. If bubbles exist, centrifuge the sample vials for two minutes at 1,000 g and repeat if necessary. Place a cap on the universal vial and ensure a good seal. See Figure 7. Place the universal vials into the 48-position inlet sample tray positions A1 through C8. 12 A51969AD Carbohydrate Labeling and Analysis Kit Sample Vial Setup Figure 7 Micro Vial Inside a Universal Vial 1. Universal Cap 2. Micro Vial A51969AD 3. Universal Vial 4. Micro Vial inside Universal Vial 13 Carbohydrate Labeling and Analysis Kit Running the Samples Running the Samples Initial Conditions Table 2 Test Run: Initial Conditions Capillary Temperature: 20° Sample Storage Temperature: 10° Auxiliary Data Channel: Current Figure 8 Initial Condition Tab LIF Detector Initial Conditions Table 3 Test Run: LIF Detector Initial Conditions 14 Detection Laser Induced Fluorescence Wavelength Excitation - 488 nm, Emission - 520 nm Data Rate 4 Hz Dynamic Range 100 RFU (relative fluorescence units) Filter Setting Normal Peak Width 16 - 25 A51969AD Carbohydrate Labeling and Analysis Kit Running the Samples Figure 9 LIF Detector Initial Conditions Tab NOTE Prior to use, calibrate the LIF detector using the procedure under LIF Detector Calibration Procedure. For comparable results, the detection system should be calibrated at system start-up or whenever a capillary is replaced. Time Program 1 Rinse the capillary with buffer for three minutes at 30 psi from BI:B1 to waste at vial BO:B1. 2 Inject the sample at 0.5 psi for three seconds from sample vial to buffer vial BO:C1. NOTE Either pressure or electrokinetic sample introduction can be used for labeled carbohydrate samples. A 3 to 20 second 0.5 psi pressure injection is recommended. Longer injection times may cause peak shape and migration times to vary. For low concentration samples, electrokinetic injections may be the method of choice. The concentration and amount of sample will determine the conditions of injection - typically 0.1 to 1.0 Ws (for example,10 kV, 5 microamps, 2-20 seconds) should be used. NOTE Salts will interfere with electrokinetic injections. Samples should first be desalted before attempting this method. A51969AD 15 Carbohydrate Labeling and Analysis Kit Running the Samples 3 4 Wait 0.2 minutes with vials BI:A4 and BO:A4. See Figure 10. This step dips the capillary in water to protect against sample carry over. Change the rinse water vials if they are contaminated. Separate Step - 20 minutes from vial BI:C1 to vial BO:C1 (buffer). See Figure 10. The constant voltage should be at 30 kV, with reverse polarity and a 0.17 ramp time. 5 Autozero at 1.0 minute. 6 End at 20.0 minutes. Figure 10 Time Program Tab 16 A51969AD Carbohydrate Labeling and Analysis Kit Running the Samples Figure 11 Typical Current Profile Field Strength Generated 598 V/cm Typical Current. See Figure 11 under 20 µA NOTE Prior to use, calibrate the LIF detector using the procedure under LIF Detector Calibration Procedure. For comparable results, the detection system should be calibrated at system start-up or whenever a capillary is replaced. A51969AD 17 Carbohydrate Labeling and Analysis Kit Evaluation of Results Evaluation of Results The test mixture contains the APTS-labeled glucose oligomers consisting of at least 20 individual oligomers. An example of this test mix is highlighted in Figure 12. Figure 12 Electropherogram - Glucose Ladder System Shutdown and Capillary Storage Short Term Storage (<24 hours) of Capillary Perform a three minute, 30 psi rinse with water. The capillary may be stored on the instrument with the capillary ends immersed in water. Whenever the capillary has not been used for three hours or longer, rinse the capillary by performing a three minute, 30 psi rinse with water before performing a separation. Long Term Storage (>24 hours) of Capillary 1 18 Perform a 3 minute, 30 psi rinse with water. Repeat the rinse with Carbohydrate Separation Buffer for 3 minutes. A51969AD Carbohydrate Labeling and Analysis Kit Troubleshooting Procedures 2 3 Remove the capillary from the instrument and place in a cassette box with the capillary ends placed in vials of DDI water. Store the cartridge box at 2°C to 8°C in an upright position. Whenever the capillary has not been used for three hours or longer, rinse the capillary by performing a 3 minute, 30 psi rinse with water and then a 10 minute rinse with Carbohydrate Separation Buffer before performing a separation. Troubleshooting Procedures Symptom Corrective Action No peaks • Check if your LIF detector has been calibrated. If not, please refer to the LIF calibration section of this system manual to perform the LIF Calibration. • Make sure that your laser is ON. • Make sure the probe stabilizer is properly connected to the lock down bar (white bar that locks down your LIF cartridge). • Make sure that you set up voltage in the method properly. This separation should be run using the reverse polarity. Check if the proper polarity is being applied by confirming that the orange LED that indicates reverse polarity is on during the separation. • Make sure that there is no bubble trapped at the bottom of the sample vial preventing the sample from being injected. • Make sure the window of your capillary is intact. If the window is broken, change the capillary right away. Most importantly, clean the probe aligner with a cotton swab moistened in water. Low Intensity Peaks • Labeling reaction was not carried out properly. For a successful labeling reaction, make sure that: — the protein is in a non-tris buffer. Tris will compete with your carbohydrate for the reductive amination. — the carbohydrate pellets were fully dried. Any residue of water will interfere in the yield of your labeling reaction. • Enzyme activity may differ from lot to lot. You may need to adjust the suggested amount, if the reaction gives a yield lower than what you usually expect. A51969AD Low Current • Confirm N-CHO capillary with 50 cm length is in place. Confirm temperature. Replace reagent vials with new reagent vials. Shift in Migration time in between runs, within the same day • Migration time shifts, with the use of this application, are associated with the length of injection. Make sure that injections are between 3 - 20 s at 0.5 psi. The injection time depends on the concentration. 19 Carbohydrate Labeling and Analysis Kit Appendix Symptom Corrective Action Carryover • Change water dip vials or change buffer vials and increment these vials as necessary Spikes in Electropherogram • Check for air that is dissolved in the gel buffer. Sonicate the gel for 5 minutes. Extra Peaks • It is very important to use extra clean 0.5 mL micro-centrifuge vials, especially during the labeling step. Contaminants left over from the manufacturing process of these vials may react with APTS, giving extra peaks on separation. Appendix LIF Detector Calibration Procedure The PA 800 plus LIF detector calibration kit procedure is used to calibrate the detection system for carbohydrate analysis. This will ensure consistent results from day to day and from capillary to capillary, as detection is measured in relative fluorescence units (RFU). This calibration procedure allows you to normalize the signal generated from your analyte, relative to a known concentration of sodium fluorescein. Perform the calibration procedure when installing a new capillary or after detector and cartridge changes. IMPORTANT The LIF calibration procedure described in this application guide has changed from previous versions. The result is different RFU values than what were obtained from assays standardized on the previous LIF calibration procedure. The calibration correction factor (CCF) is a scalar multiple of all data points, so there is no impact on signal to noise performance. This change was made to reduce the risk of saturating the detector from excessively high fluorescence signals. Table 4 Reagents Needed to Perform the Procedure Materials LIF Performance Test Mix DDI water Part Number 726022 N/A Setting the Calibration Corrector Factors (CCFs) IMPORTANT The user must have instrument administration privileges. 20 1 Launch 32 Karat. 2 Go to Tools > Enterprise Login. 3 Enter the user name PA 800 and password plus and select Log in. A51969AD Carbohydrate Labeling and Analysis Kit Appendix 4 Select the CHO instrument icon and then Configure Instrument. 5 Select Configure on the instrument configuration dialog box. 6 Select the LIF Detector icon in the right pane of the PA 800 plus System Configuration dialog box. 7 The 32 Karat Software Configuration dialog box appears as shown in Figure 13. Figure 13 32 Karat Software Configuration Dialog Box 8 A51969AD Select LIF Calibration Wizard from the Instrument Configuration dialog box to display the Calibration Wizard. See Figure 14. 21 Carbohydrate Labeling and Analysis Kit Appendix Automatic Calibration 1 Choose Auto as the calibration option, as shown in Figure 14 and click Next. Figure 14 Calibration Wizard Screen - Auto Select 22 A51969AD Carbohydrate Labeling and Analysis Kit Appendix 2 Set up the target RFU value. The target value for N-CHO-coated capillary is 7, as shown in Figure 15, and click Next. Figure 15 Calibration Wizard Screen - Step 2 3 4 A51969AD Fill one buffer vial with 1.5 mL DDI water and another with 200 µL of diluted LIF Performance Test Mix (1:2 with DDI water) in a micro vial. Place them in the buffer vial positions as indicated in Figure 16. Place an empty vial in the waste position. 23 Carbohydrate Labeling and Analysis Kit Appendix IMPORTANT The system performance test kit (713360) contains instructions for performing detector noise tests on a bare-fused silica capillary, which uses borate buffer. Do not use the borate buffer for calibration of the N-CHO capillary. Use a 50% LIF Performance Test mix solution (100 µL of LIF Test Mix with 100 µL of DDI water) in a micro vial. 5 Click Next to perform the automatic calibration, as shown below. Figure 16 Calibration Wizard Screen - Step 3 6 After the calibration is completed, a dialog box displays as shown in Figure 17. A number will show in the Calibration Correction Factor field. If it is below 10, click Accept to complete the calibration procedure. If it is above 10. See CCF Troubleshooting Procedures. Figure 17 Calibration Wizard Screen - Step 4 24 A51969AD Carbohydrate Labeling and Analysis Kit Appendix CCF Troubleshooting Procedures CCF <1 Table 5 CCF <1 Symptom Corrective Action CCF is still below 1, but not below 0.1 • There is no problem with the system. Run a standard and verify acceptable performance of the system. System performance is not acceptable or the CCF is below 0.1 • Verify that the capillary for the carbohydrate analysis was used, and it is not broken. • Verify laser output for the laser in use on the PA 800 plus system. • Verify that the correct filters are installed in the LIF detector. • Contact a Beckman Coulter service representative. CCF >1 Table 6 CCF >1 Symptom Corrective Action CCF is more than 1.0 but less than 10 • Run a standard to determine if the system performance is still acceptable. CCF is more than 10 or if the system performance is not acceptable. • Verify laser output for the laser in use on the PA 800 plus system. • Verify that the correct filters are installed in the LIF detector. • Replace the chemistries and capillary and repeat the calibration. • Contact a Beckman Coulter service representative. No Step Change Detected The LIF Calibration method is basically a comparison of detector signals between a non-fluorescent solution to a known fluorescent solution. When rinsed with the non-fluorescent solution followed by a fluorescent solution rinse, you should see the first part of the detector signal near zero and the second part of the detector signal near the target fluorescent value. This detector output is step shaped in appearance. If no step change is observed, the appropriate solutions are not passing the detector or the detector is not able to detect them. 1 2 A51969AD Verify that the solution is rinsing through the capillary by performing a direct control 5 minute 20 psi pressure rinse of water, from buffer inlet A1 to an empty buffer vial in outlet B1. Once the rinse has begun, lift the sample cover. 25 Carbohydrate Labeling and Analysis Kit Appendix 3 Observe the outlet end of the capillary in B1. Droplets should be formed on the outlet end of the capillary. • If no droplets are formed, either the capillary is plugged or the instrument has a pressure failure. • Replace the capillary and repeat the rinse. If there is still no droplet observed, contact a Beckman Coulter service representative. Once flow through the capillary is verified, the detection system is the only other possible cause, when no step is detected. 4 Verify that the correct filters are installed in the LIF detector. 5 Verify that the laser provided with the PA 800 plus system is connected and on. 6 Run the calibration test with a new capillary and chemistries. If there is still no step change detected, contact a Beckman Coulter service representative. 26 A51969AD Carbohydrate Labeling and Analysis Kit Appendix A51969AD 27 www.beckmancoulter.com © 2014 Beckman Coulter, Inc. All Rights Reserved