SPE-DEX® 4790 Extraction System Specifications/Characteristics and Ordering Information Physical Properties Minimum bench space for each SPE-DEX 4790 Station: 6.5” W x 13.5” D x 19.75” H (16.5 x 34.3 x 50.2 cm) Minimum bench space for the Envision Controller: 4.3” W x 12.5” D x 9.25” H (10.9 x 31.8 x 23.5 cm) Each Controller can operate up to eight extractors. At least one inch between each extractor is recommended. Approximately 24 inches of bench space is needed for the solvent bottles and waste recovery bottles. A counter depth of 30 inches is preferable but at least 24 inches is required. Support The table on which the SPE-DEX 4790 System sits must accommodate the combined weight of the Extractor Stations and Envision® Controller: Extractor 22 lbs (9.8 kg) Controller: 18 lbs (8.2 kg) Gas A clean, dry regulated source of nitrogen is required. The source can be either a tank and regulator, or house nitrogen. The pressure must be a minimum of 60 PSI and a maximum of 80 PSI. ⅛-inch male NPT fittings are supplied, which will adapt to most nitrogen regulators. The connection also can be made using a customer-supplied ⅛-inch Swage fitting. House air is not recommended for use. The vacuum pump draws solvent vapors from the solvent recovery bottle, so the exhaust from the vacuum pump must be adequately vented. Running the exhaust line from the pump to a fume hood is sufficient. If desired, the extractor units, solvent bottles, and waste containers can be placed in a hood. A vent or exhaust hose (P/N 02-0882) is required if the system will be operated on a bench. It is not required if system will be placed in a fume hood. The SPE-DEX 4790 system requires a vacuum pump (P/N 50-2480-01), capable of pulling a minimum vacuum of 10 inches of Hg at 1 SCFM. The vacuum source must remain constant to ensure adequate performance. House vacuum is not recommended for use. Extractor Power 28 VAC, from the Envision Controller to the extractor Ventilation Vacuum Pump Electrical Requirements Envision Controller Input Voltage (to power supply) 115 VAC, 1.6 Amp 230 VAC 0.8 Amp (Manually Selectable) Line Frequency 50-60 Hz Maximum Power 180 Watts Uses an IEC type connector to accept various international power cords. SPE Disks Customer-provided SPE disks are required. Atlantic® and Pacific™ disks from Horizon Technology are strongly recommended. Solvent Capability Solvents are required to condition disks and elute the captured semi/nonvolatile analytes. Dichloromethane, hexane, acetone, methanol and ethyl acetate are some examples. Computer Requirements PC with Windows Operating System software; Windows XP, SP3 to Windows 8.1 Intel Pentium III 1.0 GHz minimum or better 512 MB Ram (minimum) 4 MB free hard disk space (minimum) CD ROM Drive Ethernet 10/100 Port (10BaseT) Monitor with 1024 x 768 resolution Internet Web Browser software Ordering Information: See your local Price List for pricing information Part Number 4790-CE Description One SPE-DEX 4790 Extractor Station, including one (1) 19/22 VOA Vial Adapter with a single 40-mL VOA Vial, one (1) Blue Retaining Clamp for 19/22 Tapers, one (1) 125-mL 19/22 Erlenmeyer Flask with Stopper, one (1) kit, Solvent-Waste and Water-Waste Lines and one (1) Certificate of Final Test Disk Holder. For every 4790-CE purchased, the user may select (1) at no extra charge: 50-0807-01 or 47 mm Holder. Includes metal collar, riser, cup and support screen 50-2767 or EZ Flow Holder Assembly, 100 mm wide, uses standard 47-mm disks, Includes metal collar, riser, cup, support screen, fine mesh screen, 8-second rinse magnet, 5 µm prefilter sample pack, 1 µm prefilter sample pack 50-0809-01 90 mm Holder. Includes metal collar, riser, cup and support screen (used for Oil & Grease analysis) 4790 Assembly Kits. Depending on the number of 4790’s purchased, a single of the following parts numbers will needed: 4790-12 or This allows the user to set up either 1 or 2 extractors 4790-34 or This is for either 3 or 4 extractors 4790-56 or This is for either 5 or 6 extractors 4790-78 This is for 7 or 8 extractors 50-2145-01 Envison Controller (one required for up to 8 stations) 120V version, choose the power cord for the country 50-2145-05 Envison Controller (one required for up to 8 stations) 220V, choose the power cord for the country 50-2480-01 Vacuum Pump, This is the 120 V system 50-2480-02 Vacuum Pump, This is the 220 V system, choose the appropriate power cord Vent Hose: if the extractor(s) is to be placed outside the hood, choose a vent hose to purchase based on the number of stations: One (1) 02-0882-01, or Two (2) 02-0882-02, or three (3) 02-0882-03, or four (4) 02-0882-04, or five (5) 02-0882-05, or six (6) 02-0882-06, or seven (7) 02-0882-07, or eight (8) 02-0882-08 Useful Consumables – for Atlantic® SPE Disks, and Pacific Disks, please see the website (www.horizontechinc.com) under products for the latest disk availability 31-0454 Dirty Sample Fine Mesh Screen - Fits 47-mm Disk Holder Assembly 31-0457 Dirty Sample Fine Mesh Screen - Fits 90-mm Disk Holder Assembly 49-0665 Dirty Sample Fine Mesh Screen - 100 mm for use with 50-2767 disk holder FFAP-47-HT Atlantic Fast Flow Prefilter, 47 mm, coarse 5.0 μm (Pkg of 50) FFP-47-HT Pacific Oil & Grease Fast Flow Pre-filter 47 mm (Pkg of 50) FFP-90-HT Pacific Oil & Grease Fast Flow Pre-filter 100 mm (Pkg of 50) FFAP-100-HS1 Atlantic Fast Flow Prefilter 100 mm Fine 1.0 μm (Pkg of 50) For use with 50-2767 FFAP-100-HS5 Atlantic Fast Flow Prefilter 100 mm Coarse 5.0 μm (Pkg of 50) For use with 50-2767 49-2620-01 8270 Carbon Cartridge Max Detect 20CC, Kit 16 per pack 49-2620 Carbon Cartridge 20 CC, Kit 16 per pack 49-2793 Atlantic One Pass Hardware Kit (for use with 47-2346-11 SPE Disks) Spec0051508_01