451 Input Impedance of the Systemic Circulation in Man WILMER W. NICHOLS, C. RICHARD CONTI, WILLIAM E. WALKER, AND WILLIAM R. MILNOR SUMMARY To determine the systemic input impedance, pulsatile pressure and flow were measured in the ascending aorta in 16 human subjects who were undergoing diagnostic cardiac catheterization. Blood flow was measured with a catheter-tip electromagnetic velocity meter, and pressure with an external transducer connected with the fluid-filled lumen of the catheter. Five subjects were found to have no evidence of cardiovascular disease (group A, mean age 32 ± 2 years, mean aortic pressure 97 ± 4 mm Hg). Seven had clinical and angiographic signs of coronary arterial disease, and mean pressures less than 100 mm Hg (group B, mean age 48 ± 2 years). Four subjects had signs of coronary disease and mean pressures greater than 100 mm Hg (group C, mean age 48 ± 3 years). The frequency spectra of impedance were qualitatively similar in all three groups and resembled those previously observed in the canine aorta. Characteristic impedance was lower in the normal subjects (group A, average 53 dyn sec cm"5) than in the subjects with coronary artery disease (groups B and C, average 129 dyn sec cm 5 ) . Among the subjects with coronary disease, characteristic impedance was higher in the hypertensive subjects (group C, average 202 dyn sec cm"5) than in those with lower mean pressures (group B, average 95 dyn sec cm"5). External left ventricular work per unit time (hydraulic power) averaged 1715 milliwatts (niYV) in group A, 1120 mVV in group B, and 2372 mVV in group C. Cardiac outputs were within normal limits in all subjects, but tended to be lower in group B than in group C. These results suggest that the subjects of group C were better able to meet the increased energy demands imposed by an abnormally high aortic input impedance. Further investigation is needed to learn whether the high impedances in subjects with coronary disease represent an increase with age and transmural pressure alone, or whether some additional factor is involved. The data on relatively normal subjects permit a tentative definition of the normal limits for aortic input impedance in man: 26-80 dyn sec cm"5. CLINICAL investigations of cardiovascular function have until recently been based on what might be called "steady flow hemodynamics." in that mean blood flow (cardiac output) has been measured, rather than pulsatile flow. The time-varying contours of pressure pulsations have been studied intensively, but the lack of methods for measuring instantaneous blood flow in human subjects prevented similar studies of flow waves. The first direct measurements of pulsatile flow in the ascending aorta of man were obtained with a perivascular electromagnetic probe placed around the vessel at the time of surgery.U2 Advances in design have now made it possible to place the probe at the tip of a cardiac catheter that can be inserted into a peripheral vessel and then advanced to the site where the flow is to be measured.3"5 As a result, pulsatile pressure and flow can be measured simultaneously during clinical aortic catheterization. and from these data the input impedance of the arterial tree and the total hydraulic energy required to move blood into the system can be calculated. The input impedance of an arterial system, defined as the ratio of pressure harmonics to flow harmonics at the entrance to the system, depends on the dimensions and viscoelasticity of the artery involved, on the physical properties of the blood, and on waves reflected from more distal parts of the arterial tree. Measurements of aortic input impedance provide information about the physical state of the arteries, an assessment of the external "afterload" faced by the left ventricle.6"8 and complete data for calculating the external work of the ventricle.9"" The present investigation was undertaken to determine the aortic input impedance spectra, and the hydraulic power associated with aortic blood flow, in human subjects who had no cardiovascular disease, to the extent such individuals could be found among patients referred to a diagnostic catheterization laboratory, and in patients with coronary arterial disease. Studies of this kind have been reported previously on only a few subjects.1'2- 12> 13 Our results provide a tentative definition of the normal impedance spectrum, based on five relatively normal subjects, and show higher values of impedance moduli in 11 patients with coronary arterial disease. The linearity of pulsatile pressure-flow relationships was tested in some of these patients by measuring aortic impedance at different heart rates. From the Department of Medicine, University of Florida. College of Medicine, Gainesville, Florida, and the Departments of Medicine and Physiology, The Johns Hopkins University School of Medicine, Baltimore, Maryland. Supported in part by Grants 5 R01 HL 17717-02 and HL 12607 from the National Institutes of Health. Dr. Walker's present address is: Department of Surgery, Vanderbilt University School of Medicine, Nashville, Tennessee. Received July 14, 1976; accepted for publication October 23, 1976. Methods The observations reported here were made on adults who were undergoing diagnostic cardiac catheterization. Patients with evidence of congenital or valvular heart disease were excluded, but the subjects were otherwise unselected. The study was approved by the appropriate institutional committees for clinical investigation, and informed consent was obtained from each patient. Sodium pentobarbital (100 mg p.o.) was given prior to catheterization. Clinical and angiographic signs of coronary arterial disease were found in 11 of the 16 patients studied. The remaining five gave a history of nonspecific chest pain, but had normal coronary angiograms. normal hemodynamics. and no objective indications of cardiovascular disease. Velocity of blood flow was measured with electromagnetic catheter-tip velocity transducers. A Mills catheter-tip probe3'13"16, and flowmeter (model 275. S.E. Laborato- 452 CIRCULATION RESEARCH ries. Feltham. Middlesex. England) were used in nine cases. The frequency response of this system at the filter settings used in this study, measured by applying an appropriate electrical input signal to the system," was constant in amplitude (±5%) from 0 to 32 Hz. Phase shift was approximately linear and equivalent to a time delay of 17 msec. This response is adequate for accurate measurement of pulsatile velocities in the human aorta because more than 98% of the variance of the pulsations is included in the first seven harmonics (unpublished observations). In the seven other subjects, a Carolina probe flowmeter4' 17~19 was used (model 601D, Carolina Medical Electronics). The dynamic response of this system was constant (±5%) in amplitude from 0 to 14.5 Hz. with a time delay of 35 msec. The catheter containing the Carolina probe included a flexible radiopaque tail19 extending 5 cm beyond the velocity sensor. The tail was advanced through the aortic valve and into the left ventricle, so that the sensing electrodes were at approximately the upper border of the sinuses of Valsalva. This arrangement tends to stabilize the probe in the central axis of the ascending aorta, eliminates the spurious signal that appears if the probe comes to lie against the wall of the vessel, and minimizes artifacts in the recorded velocity waveform caused by motion of the catheter. Measurements of blood flow were calibrated by one of two different methods. In the first method, the velocity calibration of the probe was determined in vitro in a hydraulic model, and the internal cross-sectional area of the vessel was measured by radiography at the time of catheterization. Since the velocity profile of the ascending aorta is relatively flat."- 2°-23 the product of measured velocity and cross section is volume of blood flow per unit time. In the second method, the output signal of the velocity meter was calibrated in cm3/sec by reference to a simultaneous determination of cardiac output by the dyedilution method. A comparison of the two calibration techniques in eight subjects showed excellent correlation (r = 0.97. y = l.O&t - 0.33 liters/min. P < 0.001). The velocity signal averaged over the last third of diastole was taken to represent zero flow. Both kinds of velocity probe have been shown to be thermally and electrically safe for use in man.24 Under local anesthesia with 1 % lidocaine. a sterilized velocity catheter was inserted percutaneously into a femoral artery, or else through a brachial arteriotomy. With the aid of fluoroscopy. the velocity probe was then positioned in the ascending aorta. Aortic pressure was measured through the fluid-filled lumen of the velocity catheter with a Millar strain gauge transducer attached externally. The frequency response of this sytem was determined by the free vibration technique.6 and the damped natural frequency in these studies ranged from 27 to 36 Hz. with a damping ratio of 0.120.17. The pressure and velocity signals were recorded on analog magnetic tape (Hewlett-Packard, model 3960) and later digitized at a sampling interval of 10 msec by an analog-to-digital converter (Technical Instruments, model 400B). Data analysis was carried out on a programmable calculator (Hewlett-Packard, model 9820A) which converted pressure and velocity data to Fourier series, applied VOL. 40, No. 5, MAY 1977 corrections for the measured dynamic responses of the transducers, and computed aortic impedance and hydraulic power as functions of frequency. Input impedance modulus at each harmonic frequency was computed by dividing flow modulus into pressure modulus, and impedance phase by subtracting the phase angle of flow from that of pressure." The impedance of 0 Hz. or "input resistance," was calculated by dividing mean flow into mean pressure. Impedances were not calculated for harmonics in which the pressure modulus was less than 0.6 mm Hg. or flow modulus less than 1 cm3/sec; these are values that probably represent the noise levels of our measurement systems. In effect, this procedure eliminated all data above about 1 2 Hz. Characteristic impedance was estimated by averaging impedance moduli between 2 and 12 Hz." Theoretically, characteristic impedance depends on the physical properties of the vessel under study, while input impedances oscillate around the characteristic value because of waves reflected from more distal points."'7> 25'2(i The hydraulic energies associated with aortic blood flow were computed by methods previously reported.10 Results The data obtained from each subject are summarized in Tables 1-3. Clinical and angiographic signs of coronary arterial disease were found in 11 of the 16 subjects studied, including two who had previously been treated surgically for coronary disease. No cardiovascular abnormality was found in the other five individuals. For purposes of comparison, the subjects were divided into three groups: Group A ("normals"): Five subjects who had given a history of nonspecific chest pain, but who were found to have normal hemodynamics. normal coronary angiograms, and no objective indications of cardiovascular disease. Their ages ranged from 28 to 37 years and averaged 32 years. Mean aortic pressure averaged 97 ± 4 (SEM) mm Hg. Group B: Seven subjects with evidence of coronary disease and mean aortic pressures <100 mm Hg. Their average age was 48 years, and average aortic pressure 85 ± 1 (SEM) mm Hg. Group C: Four subjects with coronary disease, and mean aortic pressures >100 mm Hg. Average age was 48 years, and average pressure 120 ± 5 (SEM) mm Hg. Peripheral vascular resistance in these subjects was not significantly different from that in group B. but cardiac outputs were higher in group C than in group B. The waveforms of the measured flow curves were similar to those previously recorded by electromagnetic flowmeters in the ascending aorta of the dog8 and m a n i.2. i3. is. i6. is, 19 A n e x a m p l e from each group is shown in Figures 1-3. Peak flows were much lower in subjects with coronary artery disease (average in groups B and C = 433 ± 27 ml/sec) than in the normal subjects (average = 712 ± 53 ml/sec). The low peak flows in group C were associated with prolonged ejection times (0.367 ± 0.022 seconds as compared to 0.313 ± 0.019 seconds for the normals). The average aortic input impedance spectrum for each group is shown in Figures 4-6. Because of the large differ- AORTIC INPUT IMPEDANCE/Mdiofr et al. TABLE 453 1 Hemodynamic Data Subject 1 2 3 4 5 Mean ± SEM 6 7 8 9 10 11 12 Mean ± SEM 13 14 15 16 Mean ± SEM Age, sex Diagnosis Ejection fraction 37 M 32 M 28 M 35 M 30 M 32 ± 2 52 M 45 F 52 F 57 M 43 F 44 F 40 M 48 ± 2 48 F 42 M 57 F 46 M 48 ± 3 Normal Normal Normal Normal Normal 0.71 0.65 0.61 0.66 0.57 0.64 ± 0.02 0.60 0.74 0.53 0.65 0.30 0.23 0.34 0.48 ± 0.07 0.82 0.67 0.72 0.82 0.76 ± 0.04 CAD CAD CAD CAD* CAD CAD CAD CAD CAD CAD CAD* Cardiac index (liters/min/m!) Aortic radius (cm) 1.42 1.43 1.54 1.91 1.71 1.60 ± 0.10 1.42 1.45 2.6 5.3 2.5 3.0 2.9 3.3 ± 0.52 3.0 2.68 3.4 2.6 3.2 2.8 2.2 Heart rate (beats/min) 2.8 ± 0.20 4.3 3.7 3.5 4.0 3.9 ± 0.20 2.28 2.06 2.23 2.10 1.90 2.11 ± 0.07 64 64 90 83 77 76 ± 5 79 81 98 69 80 97 102 3.4 1.15 1.34 1.60 1.43 1.39 ± 0.05 1.52 1.34 1.42 1.33 1.40 ± 0.06 Surface area (m2) 1.7 1.39 1.94 2.19 1.03 1.97 1.96 1.74 ± 0.15 1.86 1.86 1.99 1.70 1.85 ± 0.06 87 ± 5 79 75 59 80 73 ± 5 CAD = coronary artery disease. ' Post-saphenous vein bypass. ences in heart rate, the aortic input impedance data were grouped for averaging by frequency rather than harmonic number. In all cases, the moduli of the input impedance fell steeply from a high value at zero frequency (the input resistance) to a minimum between 2.3 and 5.3 Hz. and then rose to a less well defined peak at 6-9 Hz. In the subjects of group A the minimum occurred at frequencies between 2.5 and 4.2 Hz (average. 3.1); in group B, between 2.3 and 5.3 Hz (average. 4.1); and in group C, between 3.7 and 5.0 Hz (average. 4.5). The impedance phase was negative (i.e., flow led pressure) for the first three harmonics in 14 of the 16 cases. The phase angle was positive for the higher harmonics, crossing zero at approximately the frequency of the modulus minimum. The average characteristic impedances of the three groups differed to a striking degree. The value in the normal group was 53 ± 4 dyn sec cm*5, while the average in group B was almost twice that value, and the average in group C was 4 times the normal level (Table 2). Input resistances were approximately the same in all groups. Total hydraulic power associated with aortic blood flow (including kinetic energy) averaged 1715 ± 240 milliwatts (mW) in group A (Table 3). Average total power was significantly lower in group B than in the normals (group TABLE 2 Hemodynamic Data — continued Aortic pressure (mm Hg) Aortic flow Mean (ml/ sec) Mean Systolic Diastolic Pulse SV (ml) Peak (ml/ sec) 86 92 97 102 106 103 111 116 114 126 67 73 80 84 86 36 38 36 30 40 94 172 62 77 72 632 917 628 684 698 97 ± 4 114 ± 4 78 ± 4 36 ± 2 95 + 20 712 ± 53 6 7 8 9 10 11 12 81 88 83 84 88 83 89 101 119 111 109 117 105 101 67 65 72 65 67 65 74 34 54 39 44 50 40 27 66 47 67 83 41 58 42 434 274 560 410 442 420 295 87 62 110 95 55 92 72 Mean 85+1.3 109 ± 3 68 ± 1.4 41 ± 3 58 + 6 405 ± 36 82 + 7 15 16 131 115 109 125 187 150 161 147 90 84 74 95 97 66 87 52 101 93 117 85 533 439 437 524 133 115 116 113 Mean 120 ± 5 161 + 9 86 ± 5 76 ± 10 99 ± 7 483 ± 26 119 ± 5 Subject 1 2 3 4 5 Mean Ejection time (sec) R 0.267 0.380 0.320 0.297 0.300 114 ± 17 0.313 ± 0.019 1395 1298 1530 1218 ± 147 0.306 0.295 0.257 0.346 0.226 0.243 0.232 0.272 ± 0.017 1242 1860 1000 1180 2132 1204 1653 1390 ± 163 0.372 0.378 0.387 0.330 0.367 ± 0.022 133 115 96 182 93 105 92 1200 671 Zo 51 50 67 43 53 53 ± 4 ± SEM 115 118 60 124 81 95 85 95 + 12 ± SEM 13 14 1254 1484 1344 ± 49 273 160 214 95 202 ± 32 ± SEM R = input resistance in dyn sec cm SV = stroke volume. 5 (mean aortic pressure divided by mean flow); Z,, = characteristic input impedance modulus (see text) in dyn sec cm 5 ; CIRCULATION RESEARCH 454 TABLE 1 2 3 4 5 Mean 40, No. 5, MAY 1977 3 External Hydraulic Power of the Left Ventricle (Milliwatts) Kinetic power Potential power Subject VOL. Steady flow Pulsatile 1256 2233 1203 1436 1304 1486 ± 190 94 371 162 216 169 Combined 192 ± 37 Steady flow Pulsatile 1350 2550 1365 1652 1473 1678 ± 224 2.6 Total power Combined Steady flow Pulsatile power 2 37 39 1258 131 414 8 107 115 2241 176 15 1 14 1204 232 17 1 16 1437 11 179 1 10 1305 ± 1 . 4 37 ± 18 39 ± 19 1489 ± 192 226 ± 50 Pulsatile/ Combined total power total power 1389 2655 1380 1699 1484 1715 ± 240 10 16 13 14 12 (%) 13 ± 1 ± SEM 6 7 8 9 10 11 12 Mean ± SEM 13 14 15 16 Mean 947 762 1121 860 174 178 259 213 139 165 95 928 ± 75 175 ± 20 1220 1065 628 1021 940 1479 1278 767 1177 955 2330 474 381 1747 535 1697 283 1891 1916 ± 144 418 ± 55 1103 ± 90 1 13 1 5 2 36 3 31 1 14 1 7 1 5 1.4 ± 0 . 3 16 ± 5 2804 2128 2232 2174 2335 ± 158 4.0 3 21 4 28 6 53 3 33 ± 0.7 34 ± 7 14 6 38 34 15 8 6 34 ± 8 929 ± 75 190 ± 24 1120 ± 94 17 ± 1.3 2828 2160 2291 2210 2372 ± 154 18 19 26 14 1222 1068 629 1013 17 ± 5 24 32 59 36 1135 861 187 183 294 244 153 172 100 948 763 495 2333 409 1751 588 1703 316 1894 1920 ± 143 452 ± 58 946 1516 1312 782 1185 961 17 19 19 18 20 15 10 19 ± 2.5 ± SEM A), and there was little overlap in the individual values. Six of the seven subjects in group B had levels of hydraulic power lower than any observed in the normal subjects. Average total power in group C, however, was significantly higher than in group A. The group A average probably overestimates the total power to be expected in normal subjects, because of the relatively high cardiac output (11.0 liters/min) in one individual (subject 2). If this subject were excluded, the average would be 1,488 mW. Adopting this value for the normal average would not alter the significance of the differences discussed above. The "pulsatile component" of hydraulic power.10 which expresses the energy entailed in pulsations and depends predominantly on the physical properties of the aorta, constituted a larger part of total power in the subjects with coronary disease than in the normal subjects (Table 3). Kinetic energy accounted for approximately 2% of the total power in all three groups. LINEARITY The valid representation of a continuous function such as a pressure or flow wave by Fourier series requires that these waves be periodic, and that the system be in a steady state. These conditions are satisfied in the dog27 and there is no reason to believe that they are not satisfied in man. The use of Fourier analysis to define impedance also assumes linearity of the system in which the pressure and flow pulses are measured. Several techniques have been employed to test this assumption in the arterial system of the dog,28"32 and all of them have shown approximately linear relationships. To test the linearity of the pressureflow relationships in the human aorta, we measured impedance as the input waveforms were altered by electrically pacing the heart at different rates28- 31>32 in five subjects. Changes in the fundamental frequency, or heart rate, in the range 1.4-2.6 Hz produced no consistent changes in the impedance spectrum, demonstrating that the system behaves in an almost linear fashion in this FLOW MODULI ( Normal) FLOW MODULI (P<IOO) 0 (Normal) 2 4 6 8 Harmonics FIGURE 1 Pressure and flow curves obtained with a Mills velocity catheter in the ascending aorta of a young adult subject with normal cardiovascular hemodynamics (group A). The bar graph shows the harmonic content of the flow curve. The mean value of the curve is represented by the dark-shaded bar at the 0 harmonic. Heart rate was 1.07/sec and ejection time, 0.267 second. 2 4 Harmonics FIGURE 2 Pressure and flow curves obtained with a Carolina velocity catheter in the ascending aorta of a subject with coronary artery disease (CAD) and normal blood pressure (group B). Heart rate was 1.62/sec and ejection time, 0.243 second. AORTIC INPUT IMPEDANCE/Mc/io/5 et al. FLOW MODULI (P>IOO) 1400 455 CAO, P<IOO (N = 7) 1200 300" 2 4 200- Harmonics FIGURE 3 Pressure and flow curves obtained with a Carolina velocity catheter in the ascending aorta of a subject with coronary artery disease (CA D) and elevated mean aortic pressure (group C). Heart rate was 1.25/sec and ejection time, 0.378 second. range. Pacing was carried out in at least one subject from each group and an example from group A is shown in Figure 7. The relatively small shifting of points on the impedance spectrum with pacing appeared to be a random variation associated with errors of measurement rather than a systematic change related to the fundamental rates. Another conceivable source of error lies in the respiratory swings of intrathoracic pressure, which might alter aortic compliance and hence impedance. In all subjects studied there were small variations in the pressure and flow during the respiratory cycle. In four subjects pulses were analyzed throughout the respiratory cycle, including NORMAL (N=5) -1200- m E > \ 100- 4 6 Freq (Hz) - o -2 FIGURE 5 Average aortic input impedance of seven subjects with coronary artery disease (CA D) and mean aortic pressure less than 100 mm Hg (group B). The resistance was 1,390 ± 163 dyn sec cm~b, and the characteristic impedance, 95 ± 12 dyn sec cm'5. samples at maximum inspiration and expiration. The results of each test were similar to the example shown in Figure 8. demonstrating that respiration per se does not influence aortic input impedance. 300 \ 1400 » CAD, P>IOO 200- 4 6 Freq (Hz) FIGURE 4 Average aortic input impedance of five normal adult subjects (group A). The standard errors of the means of modulus, phase, and frequency are represented by vertical and horizontal bars. Z o is the estimated characteristic impedance (53 ± 4 dyn sec cm~s) obtained by averaging the moduli above 2 Hz. The input impedance modulus (lop panel) falls from a high value at zero frequency (mean aortic pressure divided by mean flow, 1,218 ± 147 dyn sec cm~h) to a minimum and then rises to a maximum. The impedance phase (lower panel), which is initially negative (flow leads pressure), crosses zero in the neighborhood of the first modulus minimum and becomes positive (pressure leads flow). -2J FIGURE 6 Average aortic input impedance of four subjects with coronary artery disease (CA D) and elevated mean aortic pressure (group C). The resistance was 1,344 ± 49 dyn sec cm~b and the characteristic impedance, 202 ± 32 dyn sec cm~b. CIRCULATION RESEARCH 456 I4OO PACING • E 1200- i 3003 200 100- 4 6 8 Freq (Hz) 2i -2 (Normal) FIGURE 7 Input impedance of the systemic circulation in a normal young adult at five different heart rates, showing no significant change of impedance pattern. Resting heart rate was 84 beats /min, and faster rates were produced by electrical pacing. The relative constancy of the impedance patterns with changes of heart rate is evidence that the arterial system is approximately linear. There is a well defined minimum in the modulus at 2.5 Hz. Discussion NORMAL IMPEDANCES The aortic input impedance spectra in our subjects closely resemble the examples published previously by other investigators.1-2- l2f 13 In spite of the variety of methods and subjects, certain distinctive features of the aortic impedance spectrum have been noted consistently. Impedance moduli are usually less than '/io the amplitude of the input resistance, and there is a steep decline in modulus at low frequencies to a minimum between 2 and 6 Hz. followed by a maximum at about twice the frequency of the minimum, and only small oscillations of amplitude at higher frequencies. The impedance phase angle is negative (denoting that flow leads pressure) at low frequencies, becomes zero at approximately the frequency of the minimum modulus, and is usually positive at higher frequencies. A similar impedance pattern has been found in the canine aorta. 6 ''• 9 although the impedance moduli are greater in magnitude in that species. The difference in magnitude is related to the general constancy of pulse pressures in mammals of different size, combined with the direct correlation between flow amplitudes and body size. The spectrum of impedance vs. frequency is thus qualitatively the same for the human as for the canine aorta, and conclusions that have been drawn from such spectra on the basis of experiments in the dog presumably apply also to man. Two examples may be cited. First, the fundamental harmonic for heart rates in the physiological range falls on the steep, low frequency portion of the impedance modulus curve. Consequently, at rates below about 120 beats/min. the slower the rate, the greater the external VOL. 40, No. 5, MAY 1977 cardiac work needed to eject a given pulsatile flow.10 Second, because the steepness of this curve depends on reflections from the peripheral vascular tree, the vasomotor state of the peripheral vessels can influence significantly the impedance in the ascending aorta.6- 9-25 The characteristic impedance in our normal subjects (53 ± 4 dyn sec cm"5) is lower than that in previously published human aortic impedance spectra.1- 2- 12> 13 but the spectra previously reported were from subjects with cardiovascular disease. Patel and his colleagues2 measured impedance in the ascending aorta in three subjects with atrial septal defects, and the average characteristic impedance was approximately 82 dyn sec cm"5 (our estimate from their figures). The ages of their subjects were 24-35 years; mean aortic blood pressures were 74-94 mm Hg. Their data were obtained with a rigid velocity probe around the aorta, and under conditions of open thoracotomy. Gabe and his associates12 catheterized the aorta with a double-lumen catheter in three subjects with rheumatic mitral valvular disease, measured flow by a pressure-gradient method, and computed impedance. Their figures indicate an average characteristic impedance of about 100 dyn sec cm"5. Age ranged from 37 to 39 years, and mean pressure from 83 to 98 mm Hg. Mills and his group.13 who used methods similar to those in the present investigation, presented impedance spectra for the ascending aorta in two subjects. 40 and 51 years of age. both diagnosed as cases of ischemic heart disease. (Data from 21 other subjects were reported in Mills' paper, but ascending aortic impedance measurements were presented for only two.) The characteristic impedance in these subjects was apparently much higher than in our normal subjects. Their impedances were expressed as pressure-velocity ratios. RESPIRATORY CYCLE 400 \ \ 1200 \ • Inspirotion \ o Expiration \ 300 200 \ O o °*\ \\ «• o • -""'^ \ 100 •bo~"~ o 4 6 Freq ( H z ) J—i (CAD) FIGURE 8 Input impedance of the systemic circulation in a subject with coronary artery disease (CA D). Pressure and flow for analysis were selected at intervals throughout one complete respiratory cycle. The impedance data points at each frequency did not change with respiration, even though there were changes in pressure and flow during the cycle. AORTIC INPUT IMPEDANCE/Mc/io/s et al. and we estimate the characteristic impedance at approximately 600 dyn sec cm"3. Converted to the same units, our group A average is 416 dyn sec cm"3. Blood pressures and cardiac outputs were normal in their subjects. We conclude from these reports and our own observations that many different pathological conditions can elevate the characteristic impedance in the ascending aorta. The number of normal subjects studied is as yet too small to permit firm conclusions about normal limits, but we suggest as a tentative guide that in the absence of cardiovascular disease or elevated blood pressure, under basal conditions, characteristic impedance in the ascending aorta in man is usually between 26 and 80 dyn sec cm"5 (group A mean ± 3 SD). 457 _ 300 0 70 80 20 IOO PA 0 (MM HG) 80 AGE (YEARS) sion. In any case, the response to vasoactive agents is not an appropriate test, because they alter not only the distending pressure but also the active tension of smooth muscle in the vessel being examined. The evidence is inconclusive, and it may be that transmural pressure plays only a small part in determining characteristic impedance in vivo. Conceivably, our results may depend less on transmural pressure than on a correlation between disease of the coronary arteries and some pathological process in the aortic wall, perhaps an exaggeration of the intramural effects of aging on the whole arterial tree. TOTAL EXTERNAL POWER t V r 00 oscillatory terms O milliwatt on OD Transmural pressure and age are known to have an important influence on the stiffness of the vascular wall.33-34 and for that reason it has been widely assumed that these variables affect the characteristic impedance of arteries. Consideration of the characteristic impedance (Zo) in relation to mean aortic blood pressure and age in our subjects suggests that these two factors may account for the relatively high characteristic impedances in the individuals with coronary disease (Fig. 9). The graph on the left side of Figure 9 shows that the difference in Z,, between groups A and C appears to be related to the higher pressures in the latter group. The right side of the figure indicates that the difference between groups A and B is associated with the older age of the subjects in the latter group. In addition, groups B and C have about the same age distribution, and the subjects with higher pressures tend to have the highest characteristic impedance. Whether age and distending pressure completely account for the observed differences, however, is a question that cannot be answered until data are available on older normal subjects, and on subjects with coronary disease who are under the age of 40. A direct relation between age and arterial impedance is to be expected. Arteries unquestionably become less compliant with age.6'33 and experimental stiffening of the aortic wall by external constraints has been shown to increase characteristic aortic impedance.9 The effect of increased transmural pressure on vascular impedance is far from certain, however. Arteries become stiffer as they are distended, presumably because more and more of the wall stress is borne by collagen as the diameter increases.34 Vascular impedance is directly proportional to the stiffness (elastic modulus) of the vascular wall, but it is also inversely proportional to the cross-sectional area of the vessel.s-25 The net effect of increased transmural pressure on impedance is therefore difficult to predict, and the literature on the subject gives no definitive answer. The elevation of arterial pressure that follows intravenous infusion of norepinephrine. for example, is not accompanied by an increase in characteristic aortic impedance in the dog.7-35 though the impedance minimum is shifted to higher frequencies. Nevertheless, in two of the three subjects studied by Gabe and his associates12 there was a slight increase in aortic impedance during norepinephrine infu- 60 FIGURE 9 Variations in characteristic impedance (Zo) with age and mean arterial pressure. Values for the subjects with normal cardiovascular hemodynamics (group A) are indicated by the open circles (O). Values for the subjects with coronary artery disease and mean arterial pressures less than 100 mm Hg (group B) are indicated by the closed circles (%). Values for the subjects with coronary artery disease and mean arterial pressures greater than 100 mm Hg (group C) are indicated by the crosses (x). See text for discussion. LIC POWER ( EFFECTS OF AGE AND PRESSURE 40 16 24 GROUP A PATIENTS 1-5 6-9 10-16 FIGURE 10 Total external hydraulic power of the left ventricle in the three groups of subjects. The upper portion of the figure indicates the amount of power required to move blood through the systemic circulation in a pulsatile manner (oscillatory terms) and the lower portion indicates the amount of power associated with mean blood flow (mean terms). CIRCULATION RESEARCH 458 HYDRAULIC POWER The hydraulic power, or work per unit time, associated with blood flow at the root cf the aorta depends not only on the ability of the left ventricle to do external work, but also on the properties of the arterial tree into which blood is ejected. Aortic input impedance is an expression of these properties. On the one hand, the stiffer the aorta, the higher the impedance moduli and the greater the amount of work required to produce a given blood flow. On the other hand, given a constant impedance spectrum, the smaller the pressure and flow generated by the ventricle, the lower the external work and power. Consequently, the performance of the heart and the state of the aorta must both be taken into account in interpreting measurements of hydraulic power. A useful "rule of thumb" for this purpose, derived from the equations for computing hydraulic power,10 states that power equals flow squared multiplied by resistance (or impedance). The "steady flow" component of power is calculated by using mean flow and input resistance as the elements in this expression. The pulsatile component for any one harmonic is calculated by inserting the appropriate flow modulus and real impedance. The relatively low total hydraulic power in the subjects of group B (average, 1,120 mW, as compared to 1,715 mW in group A) (Table 3 and Fig. 10) thus indicates that the increased characteristic impedance in Group B was outweighed by a relatively low blood flow in that group. The subjects in group C, in contrast, had mean flows that were not significantly different from those in the normal subjects, in spite of a high impedance. They maintained normal cardiac outputs in spite of an increased afterload. in other words, though at the cost of a marked increase in work per unit time (2,372 mW). This ability to generate more power in the face of a high impedance suggests that ventricular performance in group C was in some sense better than in group B. The possibility that this relationship might be used to evaluate ventricular behavior is worth exploring. The distinction between "oscillatory" and "mean" or "steady flow" components of hydraulic power10 is useful because the steady flow terms represent energy that is dissipated primarily in the arterioles and capillaries, while the pulsatile terms depend mainly on the elasticity of the aorta. One manifestation of diminished aortic compliance is an increase in pulsatile power, as is evident in group C. In the oldest subject of that group the pulsatile component was 588 mW, or 26% of total power. Whatever the cause of the elevated aortic impedance in these subjects, the physical state of the aorta has clearly increased the energy that must be supplied by the left ventricle to move blood into the systemic circulation. References 1. Patel DJ:, Austen WG, Greenfield JC Jr. Tindall GT: Impedance of certain large blood vessels in man. Ann NY Acad Sri 115: 1129-1139. 1964 2. Patel. DJ, Greenfield JC Jr. Austen WG, Morrow AG, Fry DL: Pressure-flow relationships in the ascending aorta and femoral artery of man. J Appl Physiol 20: 459-463, 1965 3. Mills CJ: A catheter-tip electromagnetic velocity probe. Phys Med Biol 11: 323-324. 1966 4. Bond RF, Barefoot CA: Evaluation of an electromagnetic catheter-tip VOL. 40, No. 5, MAY 1977 velocity sensitive blood flow probe. Appl Physiol 23: 403-409, 1967 5. Kolin A: A new principle for electromagnetic catheter flowmeters. Proc Natl Acad Sci USA 63: 357-363, 1969 6. McDonald DA: Blood Flow in Arteries, ed 2. Baltimore, Williams & Wilkins, 1974, pp 101-351 7. O'Rourke, MF, Taylor MG: Input impedance of the systemic circulation. Circ Res 20: 365-380, 1967 8. Milnor WR: Arterial impedance as ventricular afterload. Circ Res 36: 565-570, 1975 9. O'Rourke MF: Steady and pulsatile energy losses in the systemic circulation under normal conditions and in simulated arterial disease. Cardiovasc Res 1: 313-326, 1967 10. Milnor WR, Bergel DH, Bargainer JD: Hydraulic power associated with pulmonary blood flow and its relation to heart rate. Circ Res 19: 467-480, 1966 11. Milnor WR, Conti CR, Lewis KB, O'Rourke MF: Pulmonary arterial pulse wave velocity and impedance in man. Circ Res 25: 637-649, 1969 12. Gabe IT, Karnell J, Porje IG, Rudewald B: The measurement of input impedance and apparent phase velocity in the human aorta. Acta Physiol Scand 61: 73-84, 1964 13. Mills CJ, Gabe IT, Gault JH, Mason DT, Ross J Jr. Braunwald E, Shillingford JP: Pressure-flow relationships and vascular impedance in man. Cardiovasc Res 4: 405-417, 1970 14. Mills CJ, Shillingford JP: A catheter-tip electromagnetic velocity probe and its evaluation. Cardiovasc Res 1: 263-273, 1967 15. Gabe IT, Gault J, Ross J Jr, Mason DT, Mills CJ, Shillingford JP. Braunwald E: Measurement of instantaneous blood flow velocity and pressure in conscious man by a catheter-tip velocity probe. Circulation 40: 603-614, 1969 16. Mason DT, Gabe IT, Mills CJ, Gault JH, Ross J Jr, Braunwald E, Shillingford JP: Applications of the catheter-tip electromagnetic velocity probe in the study of the central circulation in man. Am J Med 49: 465-471,1970 17. Warbasse JR, Hellman BH, Gillian RE, Hawley RR, Babitt HI: Physiologic evaluation of a catheter-tip electromagnetic velocity probe; a new instrument. Am J Cardiol 23: 424-433, 1969 18. Jacobs RR, Williams BT, Anderson MN, Schenk WG: An accurate method of measuring instantaneous blood flow in patients. J Trauma 11: 178-186, 1971 19. Uther JB, Peterson KL, Shabetai R, Braunwald E: Measurement of ascending aortic flow patterns in man. J Appl Physiol 37: 513-518, 1973 20. Ling, SC, Atabek HB, Fry DL, Patel DJ, Janicki JS: Application of heated-film velocity and shear probes to hemodynamic studies. Circ Res 23: 789-801, 1968 21. Schultz DL, Thunstall-Pedoe DS. Lee GDeJ. Gunning AJ, Bellhouse BJ: Velocity distribution and transition in the arterial system. In Ciba Foundation Symposium on Circulatory and Respiratory Mass Transport. London, Churchill. 1969, pp 172-199 22. Seed WA, Wood NB: Velocity patterns in the aorta. Cardiovasc Res 5: 319-330, 1971 23. Schultz DL: Pressure and flow in large arteries. In Cardiovascular Fluid Dynamics, edited by DH Bergel. New York. Academic Press, 1972, pp 287-314 24. Buchanan JW Jr, Shabetai R: True power dissipation of catheter-tip velocity probes. Cardiovas Res 6: 211-213, 1972 25. McDonald DA, Taylor MG: Hydrodynamics of the arterial circulation. In Progress in Biophysics, vol 9, edited by JAV Butler, B Katz, Oxford, Pergamon, 1959, pp 105-173 26. O'Rourke MF, Taylor MG: Vascular impedance of the femoral bed. Circ Res 18: 126-139, 1966 27. Attinger EO, Anne A, McDonald DA: Use of Fourier series for the analysis of biological systems. Biophys J 6: 291-304, 1966 28. Bergel DH, Milnor WR: Pulmonary vascular impedance in the dog. Circ Res 16: 401-415, 1965 29. Taylor MG: Use of random excitation and spectral analysis in the study of frequency-dependent parameters of the cardiovascular system. Circ Res 18: 585-595. 1966 30. Dick DE, Kendrick JE, Matson GL. Rideout VC: Measurement of nonlinearity in the arterial system of the dog by a new method. Circ Res 22: 101-112, 1968 31. Noble MIM. Gabe IT. Trenchard D. Guz A: Blood pressure and flow in the ascending aorta of conscious dogs. Cardiovasc Res 1: 9-20, 1967 32. Nichols WW, McDonald DA: Wave velocity in the proximal aorta. Med Biol Eng 10: 327-335, 1972 33. Learoyd BM, Taylor MG: Alterations with age in the viscoelastic properties of human arterial walls. Circ Res 28: 278-292, 1965 34. Bergel DH: Dynamic elastic properties of the arterial wall. J Physiol (Lond) 156: 458-469, 1961 35. Cox RH, Pace JB: Pressure-flow relations in the vessels of the canine aorta arch. Am J Phvsiol 228: 1-10, 1975