Product Specification Digital Alarm System Mixed Function Alarms Elements • • • • • Monitors medical gas and vacuum service and can include eight-signal modules for external switch monitoring Individual LCD display indicates pressure/ vacuum at all times (transducers sold separately) Local or remote pressure sensing Alarm module calculates high and low alarm settings automatically Primary power design accepts 115 or 230 VAC (50/60 Hz) Description Chemetron® mixed alarm panels are designed to meet the requirements of NFPA and CSA standards. Alarms shall be UL and CSA listed as an assembly and shall include all necessary gauges, factory wiring, transformers and circuitry requiring only 115 volt AC or 230 volt AC primary power. Internal voltage shall be stepped down to 5 volt DC and 20 volt DC control circuit power. Wiring to external switches shall also be at 5 volts. Voltage to external pressure or vacuum transducers (sold separately) will be 20 volt DC. Master alarm panels shall be modular in design and have one LED (Light Emitting Diode) to indicate each normalabnormal condition. Each gas monitored shall have an LCD (Liquid Crystal Display) to continuously indicate actual pressure. The master alarms shall include one or more 8-signal annunciator modules for wiring to remote switches. External switches may be either normally open (NO) or normally closed (NC) type, or, if more than one 8-signal annunciator module is used, both types can be used. The selection of NO or NC is selected at set-up and is set per module. Factory setup shall be for normally closed circuitry wiring, per NFPA 99. Each signal shall be labeled for its function using selfadhesive labels provided with the unit. Adjacent to each signal label will be a red LED indicator light to signify condition of the external switch. Activation of any switch will illuminate its LED and actuate the audio alarm. The 8-signal annunciator module shall contain an independent normal/abnormal LED (green for normal and red for abnormal). The Control Module will have a test switch and an alarm silence switch. The test switch shall test all modules one at a time. There will also be an LED on the Control Module to indicate status of the micro-controller (green is normal, red is for a malfunction). The back box shall contain factory installed copper tube extensions, 6" long with a 3/8" OD to accept installer furnished lines from the medical gas system. Each inlet tube shall have a 1/8" NPT connection to accept pressure/ vacuum transducers (sold separately). The signal distribution and power supply board shall have dual transformers and a selector switch to field select either 115 VAC or 230 VAC primary service. The alarm audio tone shall be continuous. Its sound intensity shall be 90 dBa at 2 meters. The audio signal shall be canceled only by the alarm silence button or fault correction. For faults identified by the Line Pressure Modules, the audio alarm condition shall reactivate every 30 minutes until the fault has been corrected. For faults identified by the Annunciator Module, the audio alarm shall reactivate every eight hours until the fault has been corrected. Regardless of the audio alarm condition, the LED shall indicate abnormal as long as a fault condition exists. The alarm shall automatically reset with the correction of the fault condition. The panel will be equipped with dry contacts for each remote signal. This will enable the alarm to interface with another alarm or central computer system. Digital Alarm System – Mixed Function Alarms Dimensional Data - Inches (cm) Face Plate Dimensions 3 Gas 14" (35.6) x 8-7/8" (22.6) 6 Gas 23" (58.4) x 8-7/8" (22.6) Product Specification Digital Alarm System – Mixed Function Alarms Ordering Information Alarm Panels with Dry Contacts for Remote Connection Catalog Number Quantity Description 74-14-1015 2 Modules, 8 Remote Signals 74-14-1020 1 Module, 8 Remote Signals, 1 Future 74-14-1025 3 Modules, 8 Remote Signals, 2 Future 74-14-1030 4 Modules, 8 Remote Signals, 1 Future 74-14-1040 5 Modules, 8 Remote Signals Transducers with Gas-Specific Fittings (sold separately) Catalog Number Quantity Description 74-14-9051 Oxygen 74-14-9052 Vacuum 74-14-9053 Nitrous Oxide 74-14-9054 Air 74-14-9055 Carbon Dioxide 74-14-9056 Nitrogen 74-14-9057 WAGD Product Specification Digital Alarm System – Mixed Function Alarms All specifications are nominal and subject to change without notice. Warranty: See Allied Statement of Warranties for details. Customer Service: USA: (800) 444-3940 FAX: (800) 477-7701 Canada/Mexico: (800) 446-0552 FAX: (800) 246-6201 International: (314) 268-1683 1720 Sublette Avenue St. Louis, MO 63110 USA (314) 771-2400 www.alliedhpi.com © 2003 Allied Healthcare Products, Inc. Printed in Product Specification Form 74-00-3006 June