Isranalytica2006 Unexpected Peaks in Chromatograms - Are They Related Compounds, System Peaks or Contaminations? From the Diary of an HPLC Detective SHULAMIT LEVIN ©Dr. Shulamit Levin 2005 Isranalytica2006 HPLC in Pharmaceutics Σ ©Dr. Shulamit Levin 2005 Isranalytica2006 Stability Indicating Methods Extra Sensitive to Extraneous Peaks • Accurately quantifies the active pharmaceutical ingredient without interference from: – – – – – Impurities Degradation products Excipients Other potential impurities Other active ingredients FDA Guidelines: Where an analytical method reveals the presence of impurities in addition to the degradation products (e.g., impurities arising from the synthesis of the drug substance), the origin of these impurities should be discussed. ©Dr. Shulamit Levin 2005 Isranalytica2006 Accounting for Every Peak in the Related Compounds’ Profile - Even Down to 0.003 %Area Butamben Compound Retention Time (Minutes) Area % Signal-toNoise Impurity 1 2.174 0.0054 4.1 Impurity 2 2.231 0.0229 16.4 Butamben 2.263 99.954 64544 Impurity 3 2.504 0.0037 2.6 Impurity 4 2.549 0.0106 5.8 Impurity 5 2.714 0.0034 2.2 >LOQ ©Dr. Shulamit Levin 2005 Isranalytica2006 Sample vs. Blank (Diluent) Unexpected peaks can appear in blanks and in all following injections In this case they are excluded from the processing of the sample 50.00 40.00 Blank mAU 30.00 20.00 ? ? ? ? ? 10.00 0.00 50.00 40.00 Sample mAU 30.00 20.00 10.00 0.00 0.00 10.00 20.00 30.00 40.00 50.00 Minutes ©Dr. Shulamit Levin 2005 Isranalytica2006 INJECTION OF PURE SOLVENT: Why would there be peaks? “System Peak” ??? Injection Carry over ??? Contamination ??? ©Dr. Shulamit Levin 2005 Isranalytica2006 System Peaks “Legitmate” System Peaks Originate from the Mobile Phase Components Going Through Re-Equilibration ©Dr. Shulamit Levin 2005 Isranalytica2006 CONDITIONS FOR APPEARANCE OF REAL SYSTEM PEAKS • Mobile phase is multi-component (n=>2) • Mobile phase contains adsorbable components • Mobile phase’s components respond to the detector (high background) • Sample or sample-diluent is different from the mobile phase, enough to create equilibrium perturbation. ©Dr. Shulamit Levin 2005 Isranalytica2006 Mechanism Mechanismof ofSystem SystemPeaks PeaksFormation Formation Vacancy EXAMPLE: Two additives in the mobile phase k' = tR - t 0 t0 k’ =φ Cs Cm Example: k’(1) = 1 Example: k’(2) = 2 Step 1: Step 1: Equilibrium: Cs=1; Cm = 1 Equilibrium: Cs=2; Cm = 1 Step 2: Step 2: Injection of Vacancy: Cs=1; Cm=0 Injection of Vacancy: Cs=2; Cm=0 Step 3: Step 3: Re-equilibration: Cs=0.5; Cm=0.5 CHROMATOGRAM Re-equilibration: Cs=1.33; Cm=0.67 t0 k’=1 k’=2 ©Dr. Shulamit Levin 2005 Isranalytica2006 Injection of free amino acids into mobile phase containing: Ac buffer, CuAc, and Heptsulfonate. Diluent: Water Levin and Grushka © Dr. Shulamit Levin 2005, Medtechnica Isranalytica2006 Legitimate System Peaks Result from Mobile phase components due to their absence in the injected sample Levin and Abu-Lafi Injection of Blank = Water Conclusion: Use mobile phase composition as the diluent © Dr. Shulamit Levin 2005, Medtechnica Isranalytica2006 An Example for Real System Peaks in Gradients with Trifluoroacetic Acid (TFA) in the Mobile Phase Peptide Peak Sample=Peptide Dissolved in Blank = Mobile phase Water TFA Blank = Mobile phase Zoomed ACN Blank =Water System peaks appear because diluent does not contain TFA Isranalytica2006 Real System Peaks in Multi-component Mobile Phase H2O, MeCN, MeOH, THF Chromatogram and Peaks' UV-VIS Spectra 220.00 240.00 260.00 280.00 300.00 320.00 340.00 Cpd 2 - 2.364 0.00 AU -0.10 -0.20 -0.30 -0.40 -0.50 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.817 nm 2.981 nm 220.00 240.00 260.00 280.00 300.00 320.00 340.00 2.817 2.981 220.00 Cpd 2 - 2.364 nm 240.00 260.00 280.00 300.00 320.00 340.00 THF Wavelength: 215nm 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 Minutes Diluent does not contain all mobile phase components 5.80 6.00 Isranalytica2006 Contaminations Peaks - Not System Peaks! Phosphate Buffer and Acetonitril Do Not Produce Such System Peaks! 0.006 0.005 Mobile phase: ACN : Phosphate Buffer pH 9.5 1:1 Wavelength: 280 nm 3.404 0.004 AU 0.003 0.000 6.477 4.274 2.373 2.731 0.001 18.274 0.002 -0.001 Sample Name: Diluent -0.002 -0.003 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Minutes ©Dr. Shulamit Levin 2005 30.00 Isranalytica2006 Use of Diode-Array Detector for Troubleshooting of Contamination Peaks in Multicomponent Mobile Phase Chromatogram and Peaks' UV-VIS Spectra Cpd 2 - 2.188 250.00 nm 2.748 250.00 nm 300.00 2.867 250.00 nm 300.00 3.180 250.00 nm 300.00 4.226 250.00 nm 300.00 300.00 Aromatic (not in mobile phase) 0.000 4.226 0.005 3.180 AU 0.010 Wavelength: 254.0 nm 2.748 2.867 0.015 Cpd 2 - 2.188 Negative UV spectrum 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 3.20 3.40 3.60 3.80 4.00 4.20 4.40 4.60 4.80 5.00 5.20 5.40 5.60 5.80 6.00 Minutes ©Dr. Shulamit Levin 2005 Isranalytica2006 Carry Over: Chemical Residues in the system q Residues in the injector q Non-specific irreversible adsorption on column q Accumulation on system’s surfaces ©Dr. Shulamit Levin 2005 Isranalytica2006 Carry Over in the Injector Washed by Blank Injection A1100 VWD AU - A1100 AU 286 nm 0.0004 0.0003 Blank Injection 1 11.334 AU 0.0002 0.0001 0.0000 -0.0001 A1100 VWD AU - A1100 AU 286 nm 0.0004 0.0003 Blank Injection 2 Contamination is cleaned AU 0.0002 0.0001 0.0000 -0.0001 8.00 8.50 9.00 9.50 10.00 10.50 11.00 11.50 12.00 12.50 13.00 13.50 14.00 14.50 Minutes ©Dr. Shulamit Levin 2005 Isranalytica2006 Blank Injections Signals are reduced from Injection to Injection indicate carry over in the injector 100.00 Series of Blank Injections 80.00 mAU 60.00 40.00 20.00 0.00 0.00 10.00 20.00 30.00 40.00 Minutes ©Dr. Shulamit Levin 2005 50.00 Isranalytica2006 Carry-Over in the Injector LC-MS/MS 1. 2. Peak elutes at RT of main component MRM of the main component Blank Sample after wash Blank Sample before wash Injection-Clean-up with EDTA solved the problem Isranalytica2006 Carry Over: Residues in the system q Residues in the injector q Non-specific irreversible adsorption on column q Accumulation on system’s surfaces Isranalytica2006 Accumulation on Column AU MAIN1 0.0010 Sample 0.0005 0.0000 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 Minutes SampleName: SST; Vial: 1; Injection: 1; Name: MAIN1; Area: 833913 0.0005 Blank #1 MAIN1 AU 0.0010 0.0000 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 Minutes AU 0.0010 Not Cleaned Cleaned 0.0005 MAIN1 SampleName: DILUENT; Vial: 2; Injection: 1; Name: MAIN1; Area: 5063 Blank #2 0.0000 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 Minutes SampleName: DILUENT; Vial: 2; Injection: 2; Name: MAIN1; Area: 4784 14.00 15.00 Isranalytica2006 Contamination Peaks TFA Containing Mobile Phase: TFA, H2O, MeCN - Do Not Contain Aromatic Components! Chromatogram and UV-VIS Spectra of Contamination Peaks Aromatic group Aromatic group 0.000 -0.001 -0.002 41.142 41.142 AU 0.001 8.510 0.002 11.604 11.604 nm 8.510 340.00 320.00 300.00 280.00 260.00 240.00 220.00 200.00 Diluent at 280 nm 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 Minutes Isranalytica2006 Peaks of aromatic compounds in non aromatic mobile phase: Cannot be Legitimate System Peaks Chromatogram and UV -VIS Spectra of a Blank Run 1.066 1.553 2.362 3.695 6.237 6.418 24.320 26.940 27.026 nm 350.00 300.00 24.320 6.237 6.418 3.695 2.362 26.940 27.026 AU 0.20 1.066 1.553 250.00 0.00 -0.20 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 30.00 Minutes Sample Name: blank ; Wavelength : 214 Mobile phase: Gradient of 0.1% TFA in water with ACN Isranalytica2006 Contamination Peaks - Residues Contamination on-column and/or in the mobile phase Chromatogram and UV-VIS Spectra of Contaminations’ Peaks 1.274 1.514 1.731 8.500 11.633 34.345 36.595 37.612 41.018 42.128 340.00 Aromatic compounds – not legitimate components of the mobile phase 320.00 nm 300.00 280.00 260.00 240.00 220.00 0.025 0.005 8.500 0.010 1.514 1.7311.274 AU 0.015 11.633 Blank=Diluent at 220 nm 0.020 41.018 42.128 0.030 36.595 37.612 34.345 200.00 0.000 5.00 10.00 15.00 20.00 25.00 30.00 Minutes Gradient with TFA 35.00 40.00 45.00 50.00 55.00 Isranalytica2006 Carry Over: Residues in the system q Residues in the injector q Non-specific irreversible adsorption on column q Accumulation on system’s surfaces ©Dr. Shulamit Levin 2005 Isranalytica2006 Contaminations Origin – Outside the Sample q Non-pure solvents q Vials’ leachables q Reservoir’s leachables ©Dr. Shulamit Levin 2005 Isranalytica2006 Baseline at Gradient Change of Baseline with Time at Various Wavelengths: Indication for non pure solvent 100.00 0.030 220 nm 0.025 0.020 60.00 AU 0.015 Gradient’s Profile 0.010 0.005 40.00 20.00 230 nm 254 nm 0.000 0.00 280 nm 5.00 15.00 25.00 35.00 45.00 55.00 Minutes ©Dr. Shulamit Levin 2005 % Composition 80.00 Isranalytica2006 Non-Pure/Contaminated Solvents Used in Gradients 2.919 PG Contaminations can also come from the Parafilm used to seal reservoirs!!! 10.633 BHT - 6.896 BHA 4.525 - Sample TBHQ 0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Minutes 30.00 Blank 0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 Minutes ©Dr. Shulamit Levin 2005 28.00 30.00 Isranalytica2006 Sample vs Blank If Solvents are not entirely pure their impurities are adsorbed on the Stationary Phase at the beginning of the run and then elute from the column as the gradient develops. 50.00 40.00 mAU 30.00 This is the reason why there are “Gradient Grade” solvents Blank 20.00 10.00 0.00 50.00 40.00 Sample mAU 30.00 20.00 10.00 0.00 0.00 10.00 20.00 Minutes 30.00 40.00 50.00 ©Dr. Shulamit Levin 2005 Isranalytica2006 Peak’s Origin: Contamination accumulated on pump’s in-line filter Pressure Jump Bubble-Detection Unexpected Baseline Perturbation ©Dr. Shulamit Levin 2005 Isranalytica2006 Contaminations Origin – Outside the Sample q Non-pure solvents q Vials’ leachables q Reservoir’s leachables ©Dr. Shulamit Levin 2005 Isranalytica2006 Contamination from Vials’ Surfaces 0.0012 0.0010 Non-Certified 0.0008 AU 0.0006 0.0004 0.0002 0.0000 -0.0002 1.00 2.00 3.00 4.00 Minutes 5.00 6.00 7.00 8.00 0.0014 0.0012 0.0010 Certified AU 0.0008 Change of vial brand In the same experiment! 0.0006 0.0004 0.0002 0.0000 -0.0002 1.00 2.00 3.00 4.00 Minutes 5.00 6.00 7.00 8.00 Isranalytica2006 Contamination is developed during the experiment in complex Diluents (Sample’ (Sample’s Solvent) 0.020 Contamination AU 0.015 0.010 0.005 0.000 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 6.50 7.00 6.50 7.00 Minutes SampleName: Sample; Vial: 1:F,1; Injection: 1; Name: Contamination 0.020 Contamination AU 0.015 0.010 0.005 0.000 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00 5.50 6.00 Minutes SampleName: Diluent ; Vial: 1:E,2; Injection: 1; Name: Contamination Isranalytica2006 Contaminations Origin – Outside the Sample q Non-pure solvents q Vials’ leachables q Reservoir or Column’s leachables Isranalytica2006 Contamination Peaks in Size Exclusion Chromatography (SEC) Contamination is NOT a monomer! Chromatogram at 210 nm 0.030 Monomer 0.025 Sample 0.020 Contamination AU 0.015 0.010 0.005 0.000 Blank -0.005 -0.010 0.00 Higher Molecular Weight 5.00 10.00 15.00 20.00 25.00 Minutes 30.00 35.00 40.00 45.00 50.00 Isranalytica2006 Other Reasons for Extraneous Peaks Isranalytica2006 Column’s Packing Collapse All Peaks are split ©Dr. Shulamit Levin 2005 Isranalytica2006 Suggested Flow-Chart for Troubleshooting In a Regulated Environment (no change in method!) “System Peak” ??? Injection Carry over ??? Contamination ??? Yes? No Mobile phase: Multicomponent? Adsorbed components (ion pair)? Response in detector (background)? Yes? Legitimate System Peaks •Dissolve samples in mobile phase •Adjust diluent with mobile phase components ©Dr. Shulamit Levin 2005 Isranalytica2006 Suggested Flow-Chart for Troubleshooting Contd. Steps: by ease of use 1. Review and revise diluent’s composition 2. Replace vials’ batch and/or brand 3. Replace in-line filters 4. Replace solvents batch and/or brand 5. Try a new/fresh column 6. Wash system, especially injector Not System Peaks? Not washed by blanks? Make sure to try a fresh blank each test! Peaks still persist? Replace injector’s surfaces Peaks still persist? Try another instrument/Operator/Lab ©Dr. Shulamit Levin 2005 Isranalytica2006 Many times the extraneous peak disappears on its own and remains an unsolved mystery… ©Dr. Shulamit Levin 2005