Measurement of Atmospheric Aerosols using the Wide-Range Particle Spectrometer (WPSTM) Keung S. Woo, Francisco J. Romay, William D. Dick, and Benjamin Y. H. Liu MSP Corporation, Shoreview, MN 55126 Abstract Increases in ambient aerosol concentrations are associated with adverse health effects such as respiratory irritation and changes in pulmonary function. To test hypotheses regarding effects of atmospheric aerosols, it is essential to measure the atmospheric aerosol properties accurately and efficiently. Since atmospheric aerosols are contributed by multiple sources of aerosol with a wide range of particle diameters, it is important to measure atmospheric aerosols from a few nanometers to several microns in particle diameter. The Wide-Range Particle Spectrometer (WPS™) is a recently introduced commercial aerosol instrument with the unique capability to measure size distributions of aerosols over a diameter range of 0.01 to 10 µm (Liu et al. 2005). A Scanning Mobility Spectrometer (SMS) comprised of a Differential Mobility Analyzer (DMA) and a Condensation Particle Counter (CPC) is used to measure particles from 0.01 to 0.5 µm and a Laser Particle Spectrometer (LPS) is used to measure particles from 0.35 to 10 µm. These components are small enough to fit within a single portable cabinet (~25 kg) with all accompanying control hardware and electronics. No external pumps are required and power consumption is only 150 W. The WPS was tested in the field for semi-continuous atmospheric sampling during extended periods of time as a way to measure the time-evolution of the atmospheric aerosol size distribution in an urban setting. Results from these measurements show that the WPS works well in various ambient conditions with temperature ranging from 11 oC to 35 oC and relative humidity ranging from 10% to 70%. Detailed diurnal profiles and nucleation events of atmospheric aerosol size distributions were observed using the WPS in these measurements. With its relatively small size, light weight, low power consumption, and user friendliness, the WPS can perform unprecedented aerosol size distribution measurements in a wide particle diameter range (10 nm – 10,000 nm). The use of algorithms in the operation of the instrument (e.g., pressure and temperature compensation) and in the data analysis (e.g 1 particle refractive index correction), allows the user to easily interpret and display the experimental data, making the WPS a truly an invaluable tool for atmospheric aerosol studies. Introduction Size-resolved measurement of the atmospheric particulate matter (PM) is important as PM effects on health and human welfare can strongly depend on particle size. Particle size information can also provide clues on source attribution. Several studies have reported associations between particles and morbidity and mortality (Schwartz et al., 1996; Oberdörster et al., 1995; Seaton et al., 1995). Current air quality standards for atmospheric particulate matter in the United States are based on mass concentrations of particles with aerodynamic diameters of zero to 10 µm, or from zero to 2.5 µm (e.g. known as PM10 or PM2.5 mass fractions). The aerodynamic diameter is defined as the diameter of a spherical particle of unit density (i.e. 1 g/cm3 ) with the same settling velocity as the spherical particle in question. These particles are known to penetrate beyond the upper airways and are regulated. Particles smaller than 2.5 µm are referred to as “fine,” while particles larger than this are referred to as “coarse.” These two modes of particulate matter differ in sources, formation mechanisms, chemical composition and exposure relationships. Coarse mode particulate matter is formed by grinding, crushing and road traffic, as well as from natural sources such as dust, pollen and sea spray. Major components of coarse particles include crustal material such as iron, calcium and silicon; aluminosilicates in soil dust; sodium in sea spray and organic particulate matter from plant and animal materials. Coarse particles are removed from the atmosphere mainly by gravitational settling. Their lifetimes are short (in the range of minutes to hours) and they travel only short distances (< 10’s km). Thus, coarse particles tend to be unevenly distributed across urban areas and their effects are more localized. PM2.5 originates from both natural and anthropogenic sources. Important primary sources of PM2.5 include soot or semivolatile organic compounds emitted from combustion sources (e.g. fly ash, etc). Secondary PM2.5 is produced by atmospheric chemical transformations of precursor gases emitted by combustion or biogenic sources. Major components of fine particles include sulfates, nitrates, organic compounds and trace elements. Utrafine (< 50 nm) particles are removed by coagulation with accumulation mode (0.1 – 1.0 µm) 2 particles, which are removed from the atmosphere by forming cloud droplets or by dry deposition. Fine particles have long lifetimes (in the range of days to weeks) in the atmosphere and travel hundreds to thousands of kilometers. Thus, fine particles tend to be evenly distributed across urban areas. In addition, fine particles have a higher probability of being deposited in the alveoli and respiratory bronchioles of the lung, where removal mechanisms are inefficient. Several studies suggested that fine particles are more likely to cause adverse health effects (e.g. Oberdörster, 1996; Peters et al., 1997). Due to the differences in composition and respiratory deposition patterns of fine and coarse particles, it is necessary to measure these atmospheric particles in a wide size range. With the implementation of PM10 (1987 Particulate Matter NAAQS) and PM2.5 (1997 Particulate Matter NAAQS) particulate air quality standards based on particles smaller than 10 µm and smaller than 2.5 µm aerodynamic diameter respectively, interests in wide-range atmospheric aerosol measurement is now generally limited to particles of 10 µm and less in diameter for environmental research, health effects, air pollution, and air quality studies. Aerosol measurement in this more limited, though still wide size range, is still difficult for most experimental researchers due to the lack of a suitable instrument covering the entire size range of interest. Measurements of atmospheric aerosol size distribution have been conducted in many cities around the world. Li et al. (1993) measured particle size distributions in the size range of 17-866 nm in Taipei, China. Buzorius et al. (1999) measured particle size distributions in the size range of 10-500 nm in Helsinki, Finland. Hughes et al. (1998) measured particle size distributions in the size range of 17-3000 nm in Los Angeles, CA . Shi et al. (1999) studied the particle size distributions (10-400 nm) in Birmingham, UK. Also, particle size distributions in the size range of 65-900 nm in Santiago de Chile were measured by Trier (1997). All these researchers tried to cover the widest possible aerosol size range, so that they could have as much information as possible about the aerosols they were measuring. To cover the broad aerosol size range, these researchers had to combine several aerosol measurement instruments. This is because there was no instrument that could measure the aerosol size distribution in a wide size range. However, using several different instruments to measure atmospheric aerosol in a wide size range is difficult. The user has to: (1) design a flow system to bring the sample to each instrument, (2) design custom control software to communicate with each instrument, (3) use 3 several extra hardware components (e.g., pumps, valves and computers, etc), and (4) deal with specific instrument details from each instrument manufacturer, It is not uncommon that it takes researcher a year to incorporate several instruments as part of a single rack system before making aerosol size distribution measurements in a wide size range. This paper describes the use of a recently introduced commercial aerosol instrument with the unique capability to measure size distributions of aerosols over a diameter range of 0.01 to 10 µm (Liu et al. 2005) in atmospheric aerosol measurements. The overall design of the WPS™ as well as the experimentally obtained atmospheric size distribution measurements at several locations are described and presented in this paper. Wide-Range Particle Spectrometer (WPS™) The Wide-Range Particle Spectrometer (WPS™) is a new aerosol measuring instrument that combines several aerosol sizing and sensing techniques to measure aerosols over a wide size range. The individual sizing or sensing techniques used are well known and have been available for some years as stand-alone instruments. Taking advantage of recent advances in microelectronic and computer technology, and of the increased understanding of the underlying physical principles of aerosol measurement, MSP Corporation has succeeded in developing the WPS™ into a small compact device with advanced measurement capabilities. The techniques used in the WPS™ include laser light scattering, differential mobility analysis and condensation particle counting. Light scattering instruments are generally referred to as optical particle counters, laser particle counters, or as laser aerosol spectrometers (Liu et al., 1985; Barnard and Harrison, 1988). These instruments sense individual particles by light scattering and measure the scattered-light amplitude to determine particle size. Sizing and counting of particles by the differential mobility analyzer (DMA) and the condensation particle counter (CPC) involve classifying aerosol particles by electrical mobility with the DMA and then counting the classified particles by condensation and droplet growth with the CPC. The combination is usually referred to as a differential mobility particle sizer (known as DMPS) (Liu et al., 1978) or as a scanning mobility particle sizer (known as SMPS) (Wang and Flagan, 1990), depending on whether the voltage on the DMA electrode is changed in steps (DMPS) or varied continuously and scanned. In the latter case, the instrument is then referred to as a SMPS. 4 . The traditional stand-alone laser-light-scattering instrument typically has a lower size limit of 0.1 µm. The upper size limit is typically 10µm. For the DMPS or SMPS, the typical measurement range is 0.01 µm to 1.0 µm. In the WPS™, these measurement principles are combined to allow the measurement of particles from 10nm to 10,000nm (0.01µm to 10µm). In the WPS™, a Scanning Mobility Spectrometer (SMS) comprised of a Differential Mobility Analyzer (DMA) and a Condensation Particle Counter (CPC) is used to measure particles from 0.01 to 0.5 µm and a Laser Particle Spectrometer (LPS) is used to measure particles from 0.35 to 10 µm. These components are small enough to fit within a single portable cabinet (~55 lbs) with all accompanying control hardware and electronics. No external pumps are required and power consumption is only 150 W. WPS™ System Overview Figure 1 is a simplified system diagram for the WPS™. The sampled air flow enters the instrument through a common inlet at the sample flow rate of 1.0 liter per minute (L/min). Of this total flow, 0.70 L/min is sampled into the laser particle spectrometer (LPS) and the remaining 0.30 L/min is sampled for electrical mobility classification by the DMA followed by the CPC to count the single, individual particles. At the inlet of the DMA there is a single-stage impactor with a cutpoint of 0.5 µm and a radioactive Polonium 210 ionizer for aerosol charge neutralization (Liu et al., 1986). The impactor provides a known upper limit on the particle size that can be sampled by the DMA and the charge neutralizer imparts an equilibrium bipolar charge distribution on the aerosol particles sampled by the DMA. Flow Measurement and Control—An aerosol counting and sizing instrument such as the WPS™ requires accurate flow rate measurement and control because both the particle size and concentration are dependent on the flow rate. An important consideration in the design of the WPS™ was to insure that all flow rates were controlled and measured accurately by the system. A critical part of the flow control and management system is the use of separate miniature DC pumps to control the individual flow rates to the desired set-point values. Four miniature variable speed DC pumps are used: two for establishing the sample flow rates for the LPS, and for the DMA and CPC, and two for providing the clean, re-circulating sheath air flows needed by the LPS and by the DMA. Three flow loops use laminar-flow elements connected to a 5 differential pressure transducer for sensing the pressure drop and providing an electrical output for flow measurement and control. The pumps are protected at the inlet and outlet by high efficiency air filters to minimize particle contamination and to prolong pump life. The DMA and CPC flow loop uses an orifice as the pressure-drop element. In the case of the sample flow pumps, the high efficiency filters at the pump outlet serve to filter the exhaust gas before it is discharged to the ambient. For flow control, the pump speed is adjusted with pulse-width modulation, or PWM, so that just the right amount of electrical power is supplied to each pump to keep the DC pumps running at the required, but greatly reduced speed to provide the set-point flow and to overcome the flow resistance in each particular flow circuit. This reduces both the energy consumption and prolongs pump life. The nominal flow rates and control accuracies for the WPS™ are summarized below in Table 1. Table 1 WPS Sample and Sheath Flows Flow Loop Flow Rate (L/min) LPS sample flow 0.700 LPS sheath flow 3.00 DMA and CPC sample flow 0.300 DMA sheath flow 3.00 Accuracy (L/min) ± 0.02 ± 0.06 ± 0.006 ± 0.06 Differential Mobility Analyzer—The differential mobility analyzer in the WPS™ has a cylindrical geometry with an annular space for the laminar aerosol and sheath air flows (see Figure 2). The key dimensions of the DMA are given in Table 2. These critical dimensions were optimized to obtain size classification of particles between 5 and 500 nm with a maximum voltage smaller than 10,000 volts and a mobility-classifying length of only 12.25 cm. A precision absolute pressure transducer is connected to the sheath flow inlet on the DMA to measure the internal DMA pressure. A temperature transducer is mounted near the mobility tube to measure the DMA temperature. These transducers provide the necessary output to provide automatic temperature and pressure compensation needed for accurate DMA mobility and particle size analysis. The DMA has all the flow ports (with the exception of the aerosol inlet port) on the base. The high-voltage connector is also located on the DMA base. Without temperature and pressure compensation, the DMA measurement is subject to error that can vary 6 greatly due to elevation and temperature changes. For instance, if the flow rate of a DMA is measured with a thermal mass flow transducer and given in standard liters per minute, the difference in volumetric flow between sea level, and an elevation of 5,000 ft, for instance, can cause a volumetric flow difference and measurement error of as much as 20%. For high accuracy aerosol measurement, such errors are not acceptable even for routine measurements. Table 2 Critical Dimensions of Classifying Region in the DMA Description Dimension Dimension (inches) (cm) Outer diameter of annular classifying region 1.750 4.444 Inner diameter of annular classifying region 1.250 3.175 Longitudinal distance between aerosol inlet and outlet 4.820 12.24 The high voltage power supply generates the DMA operating voltage that can vary from 10 VDC to 10,000 VDC. This power supply has two output voltage monitors, one for the low voltage range (0 to 1000 VDC) and the other for the high voltage range (1,000 to 10,000 VDC) to achieve high accuracy over the wide voltage range needed for accurate aerosol size measurement by the DMA. The DMA has been evaluated experimentally using the TDMA (Tandem Differential Mobility Analysis) technique reported by Birmili et al. (1997) and more recently by Martinsson et al. (2001). TDMA measurements were made for 18 particle diameters from 15 nm to 450 nm while operating the DMAs at 3.10 L/min sheath air (i.e. Qs = 3.10 L/min) and 0.310 L/min aerosol flow (i.e. Qa = 0.310 L/min) (Liu et al. 2005). By incorporating the measured DMA transfer function as a function of particle diameter in the data inversion algorithm, it is possible for the WPS™ to obtain accurate aerosol size distribution measurements down to 10 nm without systematic errors in the measurement. Automatic data analysis is important in modern aerosol measurement. It allows for automatic correction of any known systemic error, (i.e. diffusion, time response, temperature, pressure, etc. ) provided the magnitude of the error can be calculated and/or measured for incorporation into the data analysis software. A NIST traceable calibration using the Standard Reference Materials (SRM) from the National Institute of Standards and Technology is performed for the DMA to ensure accurate size measurement. This involves generating a known sized PSL-sphere aerosol with the SRM and performing a DMA scan with a CPC. The results are then used to determine an empirical 7 factor (the ratio of scanned peak voltage to the calculated peak voltage based on the diameter of the SRM, the DMA flow rates and dimensions) that is subsequently included as a calibration factor to further eliminate any known or unknown systematic error in the DMA (Liu et al. 2005). Condensation Particle Counter— Figure 3 is a schematic diagram of the condensation particle counter (CPC). The CPC counts the individual particles and together with the measured flow, determines the number concentration of particles in the aerosol coming from the DMA. The CPC has a dual reservoir design, one for the working fluid (1-butanol referred from now on as butanol) and the other for the condensate. The CPC is of the thermal diffusion type, with a saturator maintained at 35 ºC and the condenser at 10 ºC. The aerosol is pre-heated by passing through a tubular passageway in the saturator heating block and thus at the saturator temperature upon entering the saturator tube. The saturator is made of porous metal. One end of the porous metal tube is dipped into the butyl alcohol reservoir. Through capillary surface tension, the interstitial pore space of the porous metal is filled with butanol. As aerosol flows along the inside surface of the porous metal tube, butanol is evaporated from the surface of the tube, causing the aerosol stream to become saturated with butanol vapor. In the condenser, the aerosol stream cools and becomes supersaturated. The vapor then condenses on the particles to form droplets. The droplets are detected by optical light scattering and counted individually as pulses by the pulse counting circuitry. Both the saturator and condenser temperatures are adjustable allowing different saturation to be achieved for vapor condensation on the particles. The saturator and condenser temperatures are adjusted to 35 and 10 ºC, respectively. The optics block is located at the condenser exit and is maintained at 40 ºC. The light scattering optics is of the conventional design. It includes a diode laser light source and a solid-state photo-detector. The detector circuit board is mounted on the optics block and provides pulses to be counted by the signal processing electronics and through the digital system board (DSB) communicate with the single board computer for data storage and analysis. A feedback flow control system maintains the CPC flow rate at a constant value of 0.300 L/min. The aerosol leaving the CPC is mixed with a dilution clean air flow of about 3.0 L/min before it exits the exhaust port on the back panel of the instrument. 8 One feature of the current CPC—a feature that is absent in other conventional CPC’s—is the use of a separate reservoir to collect condensate from the condenser. In ambient measurements, if the dew point is higher than the set-point temperature of the condenser, water vapor will also condense in the condenser tube. For instance, at a condenser temperature of 10ºC, moisture will condense if the ambient dew point is higher than 10 oC. For an ambient temperature of 30 ºC and a dew point of 10 oC, the relative humidity is 28%. Under such conditions, water will condense along with the butanol vapor in the condenser when the ambient relative humidity is higher than 28%, which is a common occurrence during the summer. In a conventional CPC, the condensed water as well as the condensed butanol is returned to the alcohol reservoir. Because of the high surface tension of water, water in the porous saturator will displace alcohol from the interstitial pore space, leading to incomplete saturation of the aerosol stream, causing incorrect data to be generated. The new CPC design makes sure that the condensate, including any water, if present, is returned to a separate reservoir to avoid possible inaccurate aerosol measurement due to water vapor condensation and the presence of water in the alcohol reservoir. A liquid level sensor in the condensate reservoir warns the user when the reservoir is full and needs to be drained. There is a small drainage hole at the bottom of the condenser tube to allow the condensate to drip into the reservoir below. This hole is normally filled with a few microliters of condensate with the weight of the condensate being supported by capillary surface tension, thus providing an effective seal between the condenser tube and the reservoir air space below. This liquid seal is effective in preventing air exchange between the aerosol stream above and the stagnant gas volume below. The condensation particle counter described above has been evaluated experimentally to determine its relative counting efficiency as a function of particle size (Liu et al. 2005). This was done by comparing the particle concentration of the CPC with the particle concentration measured with another CPC (TSI Model 3025 CPC). For these measurements monodisperse aerosols of silver were generated and classified by a DMA. The ultrafine silver particles were generated using the technique of Scheibel and Porstendorfer (1983). Figure 4 shows the relative counting efficiency of the WPS CPC from which we can determine that the 50% cutoff size is about 5 nm. 9 Laser Particle Spectrometer—The Laser Particle Spectrometer (LPS) is shown in Figure 5. The LPS includes a single-particle, wide-angle optical sensor used for sensing and detecting particle from 350nm to 10,000 nm in diameter. Particles are drawn into the aerosol inlet at a rate of 0.70 L/min and focused aerodynamically with a 3.0 L/min flow of sheath air to provide a narrow aerosol stream that passes through the center of a ribbon-shaped laser beam generated by a laser diode (785 nm, 5 mW). Light scattered by each particle is collected with a spherical mirror over a 20 to 100° range at a mean scattering (polar) angle 60o in the forward direction of the laser. The scattered light is projected onto a red-enhanced photomultiplier tube (PMT). In response, the PMT generates a pulse of electrical current which is then converted into voltage by the LPS Preamplifier board. The LPS Amplifier board conditions each lightscattering pulse from the Preamplifier board for input into the next stage of electronics in which pulses are sorted and binned according to their amplitude. Due to the wide-angle collection optics of the LPS, its response is monotonic with respect to particle size for a moderate range of particle refractive indexes. The response of the LPS described before has been modeled using Mie theory. Figure 6 shows the theoretical response versus particle diameter as a function of the particle refractive index. Due to the wide-angle collecting optics of the sensor, the LPS response in nearly independent of refractive index for particles from 1 to 10 microns. Between 0.3 and 1 microns there is some dependence on the particle refractive index but the response is monotonic with respect to particle diameter. For n = 1.7 there is flat are on the response curve between 0.8 and 1.1 microns. This characteristic allows particle diameter measurements made using the assumption of a particular refractive index value to be converted to a diameter scale based upon a different refractive index value, as may be done with the WPS Commander software. Conversely, in the embedded software, real refractive index n may be selected from a range of 1.30 to 1.60 in 0.05 intervals (including n=1.585 for PSL) for setting thresholds that are used in binning pulses with respect to diameter. The user may select the particle refractive index most appropriate for the aerosol sample to obtain more accurate measurement of the geometric diameters based on light scattering. Each LPS response curve has been subdivided in 24 particle diameter channels. The effect of particle refractive index is demonstrated in Figure 7. Figure 7 shows the average surface area distribution of 200 room air samples. The darker color shows the distribution data obtained by the LPS and the light color shows the distribution data obtained by the DMA and CPC. Note that the difference in the overlapping region (0.35 µm to 0.5 µm) is 10 quite large in Figure 7a. Since the LPS is typically calibrated by PSL, which has a particle refractive index of 1.585, the size response will be different when the LPS is used to measure particles (e.g., room air) other than PSL. If we use a more representative particle refractive index value (e.g. 1.45) for room air to analyze the LPS data, the difference in the overlapping region will be reduced. This is shown in Figure 7b. To confirm the theoretical response of the LPS experiments were performed with NIST traceable PSL spheres classified by a DMA. Figure 8 shows the measured voltage outputs as a function of particle diameter and also shows the scaled theoretical response for PSL spheres (i.e. n = 1.585). The agreement between the theoretical and experimental responses is very good, confirming the validity of the LPS predicted response. Signal Processing and Control Electronics—The WPS hardware system is controlled with MSP proprietary software executed by a low-power Pentium® single-board computer operating on a dedicated Windows® XP platform. The computer has an LCD display, a 60 GB hard drive, a serial communication card and a mouse. Two controller boards were used to communicate with and to control all system components in the WPS. One board, the Analog System Board (ASB) is used primarily for analog I/O. Two digital boards, or the Digital System Boards (DSB) are dedicated to digital I/O. The computer interfaces with the controller boards using one RS-485 serial communication port. Control functions include operating flow controllers, monitoring flow rates, pressures, temperatures, relative humidity, flow levels, setting and operating temperature controllers, setting and monitoring high voltage levels, operating solenoid valves, and receiving/processing particle counts from the CPC and the LPS. One low-voltage switching power supply and one high-voltage DC power supply provide all the power required by the instrument. The low-voltage power supply has several DC output voltages (i.e. +12 VDC, +5VDC, -12VDC, -5 VDC) that supply power to the different circuit boards of the system. Table 3 lists the main circuit boards along with a short description of the function that each performs on the WPS. Table 3 Circuit Boards on the WPS Circuit Board Function 11 CPC Detector CPC Counter CPC/LPS Master Laser Driver PSB Board AH&T Board LPS Slave LPS Detector and Preamp LPS HVPS LPS Amplifier CPC photo-detector board CPC counting board (slave) Communicates with CPC Counter and LPS Slave Provides controlled power to the CPC laser diode and LPS laser diode Board with absolute and differential pressure transducers Powers and signal conditions the ambient temperature and relative humidity sensor LPS counting board (slave) LPS PMT and pre-amplifier board High voltage power supply for LPS PMT LPS amplifier board Atmospheric Aerosol Size Distributions Measured with the WPS One of the most common applications for the WPS is the monitoring of atmospheric aerosol size distributions, both in urban polluted environments and in pristine environments (e.g. protected National Parks). For this application the WPS is particularly useful as it can be used with semi-continuous sampling schedules to measure the time-evolution of the aerosol size distribution. This type of monitoring allows us to detect the presence of nucleation events, whether they are of natural or anthropogenic origin (Woo et al., 2001). The WPS was used to measure ambient aerosol size distributions in different countries around the world, such as Hong Kong, Japan, and the United States. These measurements were performed under different ambient conditions. For instance, ambient temperature ranged from 11 oC to 35 oC, ambient relative humidity ranged from 10% to 70%. The WPS worked well in these environments. Also, the WPS was shown to be very easy to transport. In a recent sampling campaign, the WPS was used to measure atmospheric aerosol size distributions in Hong Kong and Japan. The WPS was shipped in a manufacturer provided hard traveling case as a regular airline-checked luggage. After the WPS arrived at the sampling site, the operator spent less than 15 minutes to unpack and setup the WPS. Setting up the WPS includes supplying the WPS with an AC power source (90 – 264 VAC, 50/60 Hz), and after the WPS is turned on, filling the CPC working fluid (the CPC working fluid must be completely drained before shipping). The WPS needed only ~ 15 minutes warm up time for the CPC temperatures to reach their set points. After this warm up time, the operator just needed to load a recipe (i.e., to tell the 12 WPS what measurement mode, sampling time, and measurement size range, etc…) and start sampling. In this particular sampling campaign, the WPS was set to measure continuously for sixty one-minute samples. Thus, the total sampling time was one hour. After the operator started the measurement, he just left the WPS unattended and came back an hour later when the measurement was completed. To pack the WPS for shipment to another sampling site was also very easy. It just included draining the CPC working fluid, installing a shipping plug inside the WPS (to isolate the CPC optics) and putting the WPS inside the hard traveling case. Typically, it takes less than 15 minutes to pack the WPS for shipment. Atmospheric Aerosol Size Distributions in Hong Kong The WPS was used to sample aerosol size distributions on a weekday (Monday) and on a weekend day (Saturday). For each day, the WPS was installed in two sampling locations in Hong Kong, one in Tsuen Wan and one in Yuen Long. Both sampling locations are the air sampling stations of the Hong Kong Environmental Protection Department. Tsuen Wan is a mixed residential/commercial/industrial city. We would expect that the aerosol size distribution will be different for weekdays and weekends. Figure 9 shows the average of 60 1-minute scan samples measured at the Tsuen Wan sampling station. Figure 9a shows the average aerosol size distribution obtained on a weekday and Figure 9b shows the average aerosol size distribution obtained on a weekend. Both average aerosol size distributions show a bi-modal distribution function. Note that the aerosol number size distribution on a weekday (Figure 9a) has a higher concentration than that of on a weekend (Figure 9b). The geometric mean diameter is larger on a weekday. This suggests that the smaller particles agglomerated to bigger particles, increasing the mean of the accumulation mode. Figure 10 shows the average of 60 1-minute scan samples measured at the Yuen Long sampling station. Figure 10a shows the average aerosol size distribution obtained in a weekday and Figure 10b shows the average aerosol size distribution obtained in a weekend. Yuen Long is mainly a residential city that is growing. As seen from the figures, the difference in concentration between the average distribution on a weekday (Figure 10a) and on a weekend (Figure 10b) is not significant. Note also that both average aerosol size distributions show a bimodal distribution function. However, in addition to these two modes, there is a number concentration peak at around 10 nm. It is believed that particles in the 10 nm peak are 13 contributed by nucleation events. One of such nucleation events was observed at the Yuen Long sampling station is shown in Figure 11. Figure 11 shows a sequence of four 1-minute long size distribution measurements taken at around noon time. Frame (a) shows the size distribution of the typical atmospheric aerosol during that part of the day. In frame (b) a nucleation event takes place giving a high aerosol concentration at 10 nm. In frame (c) the freshly nucleated aerosol agglomerates to larger sizes. In frame (d) the aerosol size distribution returns to the original one shown in frame (a). This demonstrates that the WPS is able to measure the time-evolution of atmospheric aerosol size distributions, as it can detect the presence of nucleation events, whether they are of natural or anthropogenic origin. Atmospheric Aerosol Size Distributions in Japan After the measurements in Hong Kong, the WPS was shipped to Japan. Similar to the Hong Kong measurement, the WPS was used to sample aerosol size distributions on a weekday and a weekend day. For each day, the WPS was installed in two sampling locations in Tokyo: Noge Park and Kawasaki. Both sampling locations are considered to be urban areas and are located at the intersections of major roadways. Kawasaki has a similar volume of passenger vehicular traffic on weekdays and weekends, but has several times more light to heavy duty vehicular traffic on a weekday than on a weekend. On the other hand, Noge Park has two times higher volume of passenger vehicular traffic on a weekend than on a weekday, but three times lower volume of light to heavy duty vehicular traffic on weekends. The average size distributions of 60 1-minute scan samples measured at Kawasaki on a weekday and on a weekend are shown in Figure 12. While the size distributions show similar peak diameters, the weekday distribution (Figure 12a) shows about 2.5 times higher number concentration than that of weekend (Figure 12b). This is consistent with the higher volume of light to heavy duty vehicular traffic on weekday. Figure 13 shows the average of 60 1-minute scan samples measured at Noge Park. Figure 13a shows the average aerosol size distribution obtained on a weekday and Figure 13b shows the average aerosol size distribution obtained on a weekend. Although Noge Park has higher volume of passenger vehicular traffic on a weekend, the expected increase in number concentration is offset by the lower volume of light to heavy duty vehicular traffic on a weekend. 14 Overall, the average size distribution obtained on a weekday still has a higher number concentration than on a weekend. Note that both sampling locations show similar peak diameters. This suggests that the sources of particles for these two locations are similar. It is believed that particulate matter from vehicle emissions is the major source for these two sampling locations. Conclusions Advances in aerosol instrumentation, together with advances in microelectronics and computer technology, have led to reduced size, weight and power requirements of aerosol measuring instruments. The WPS can perform unprecedented aerosol size distribution measurements in a wide particle diameter range (10 nm – 10,000 nm). The use of algorithms in the operation of the instrument (e.g., pressure and temperature compensation) and in the data analysis (e.g particle refractive index correction), allows the user to easily interpret and display the experimental data, making the WPS a truly an invaluable tool for atmospheric aerosol studies. The Wide-Range Particle Spectrometer (WPS) has been used to measure atmospheric aerosol size distributions in the range of 10 nm to 10,000 nm. Measurements were made in several urban locations in Hong Kong and Japan, with ambient temperature ranging from 11 oC to 35 oC and relative humidity ranging from 10% to 70%. The experimental results show that the WPS worked well in these environments. It has been shown that the WPS can measure detailed timeseries atmospheric aerosol size distributions, allowing the user to detect diurnal profiles of the aerosol size distribution and the detection of nucleation events. REFERENCES Barnard, J.C., and L.C. Harrison, “Monotonic Responses from Monochromatic Optical Particle Counters”, Appl. Opt. 27:584-592 (1988) Birmili, W. , F. Stratmann, A. Wiedensholer, D. Covert, L.M. Russell and O. Berg, “Determination of Differential Mobility Analyzer Transfer Functions Using Identical Instruments in Series,” Aerosol Sci. and Tech. 27:215-223 (1977). 15 Buzorius, G., K. Hämeri, J. Pekkanen and M. Kulmala, “Spatial variation of aerosol number concentration in Helsinki City.” Atmospheric Environment 33: 553-565 (1999). Hughes, L., G. R. Cass, J. Gone, M. Ames and I. Olmez, “Physical and chemical characterization of atmospheric ultrafine particles in the Los Angeles area.” Environmental Science and Technology 32: 1153-1161 (1998). Li, C. S., W. H. Lin and F. T. Jenq, “Characterization of outdoor submicrometer particles and selected combustion sources of indoor particles.” Atmospheric Environment 27B: 413-424 (1993). Liu, B.Y.H., D.Y.H Pui, K.T. Whitby, D.B. Kittelson, Y. Kousaka and R.L. McKenzie , "The Aerosol Mobility Chromatograph: A New Detector for Sulfuric Acid Aerosols," Atmos. Environ. 12:99-104 (1978). Liu, B.Y.H., D.Y.H. Pui and B.Y. Lin , "Aerosol Charge Neutralization by a Radioactive Alpha Source," Particle Characterization 3:111-116 (1986) Liu, B.Y.H., F.J. Romay, W. Dick and K. S. Woo, “A Wide-Range Particle Spectrometer for Aerosol Measurement from 0.10 µm to 10 µm”, Submitted to Aerosol Science and Technology (2005). Liu, B.Y.H., W.W. Szymanski, and K.H. Ahn, “On Aerosol Size Distribution Measurement by Laser and White Light Optical Particle Counters”, J. Environ. Sci. 28:29-24 (1985). Martinsson, B. M.N.A. Karlsson, and G. Frank, “Methodology to Estimate the Transfer Function of Individual Differential Mobility Analyzers,” Aerosol Sci. and Tech. 35:815-823 (2001). Oberdörster, G., R. Gelein, J. Ferin, and B. Weiss, “Association of particulate air pollution and acute mortality: Involvement of ultrafine particles?” Inhal. Toxicol. 7: 111-124 (1995). Oberdörster, G., “Significance of particle parameters in the evaluation of exposure-doseresponse relationships of inhaled particles.” Particulate Science and Technology 14: 135-151 (1996). Peters, A., E. Wichmann, T. Tuch, J. Heinrich and J. Heyder, “Respiratory effects are associated with the number of ultrafine particles.” Am. J. Respir. Crit. Care Med. 155: 13761383 (1997). 16 Scheibel, H.G., and J. Porstendorfer, “Generation of Monodisperse Ag- and NaClAerosols with Particle Diameters between 2 and 300 nm.” J. Aerosol Sci. 14:113-126 (1983). Schwartz, J., D. W. Dockery and L. M. Neas, “Is daily mortality associated specifically with fine particles?” Journal of Air & Waste Management Association 46: 927-939 (1996). Seaton, A., W. MacNee, K. Donaldson and D. Godden, “Particulate air pollution and acute health effects.” The Lancet 345: 176-178 (1995). Shi, J. P., A. A. Khan and R. M. Harrison, “Measurements of ultrafine particle concentration and size distribution in the urban atmosphere.” The Science of the Total Environment 235: 51-64 (1999). Trier, A., “Submicron particles in an urban atmosphere: A study of optical size distributions - I.” Atmospheric Environment 31: 909-914 (1997). Wang, S. C. and R. C. Flagan, “Scanning electrical mobility spectrometer.” Aerosol Science and Technology 13: 230-240 (1990). Woo, K.S., D.R. Chen, D.Y.H. Pui and P.H. McMurry, “Measurement of Atlanta Aerosol Size Distributions in the 3 nm to 2 µm Diameter Range,”, Aerosol Science and Technology 34:75-87 (2001) 17 Figure 1. Schematic Flow Diagram of the WPS™ Figure 2 Differential Mobility Analyzer 18 Figure 3 Condensation Particle Counter 1.20 Relative Counting Efficiency 1.00 0.80 0.60 0.40 0.20 Condenser Temperature: 10 °C, Saturator Temperature: 35 °C 0.00 0 5 10 15 20 Particle Diameter (nm) Figure 4 Relative Counting Efficiency of CPC 19 25 Figure 5 Laser Particle Spectrometer LPS Theoretical Threshold Values 1.E+03 n=1.30 n=1.40 1.E+02 n=1.50 2 1E+9*Csca(cm ) n=1.60 n=1.70 1.E+01 1.E+00 1.E-01 1.E-02 0.1 1.0 Sphere Diameter (µm) Figure 6 Theoretical and Experimental LPS Response 20 10.0 (a) Refractive Index = 1.585 (b) Refractive Index = 1.45 Figure 7 Average of 100 Room Air Surface Area Distributions Analyzed with Different Particle Refractive Indexes: (a) 1.585 (b) 1.45 21 LPS Amplifier Output (Volts) 1.E+01 Hi Gain Theory 1.E+00 Hi Gain Meas Lo Gain Theory 1.E-01 Lo Gain Meas 1.E-02 1.E-03 0.1 1.0 10.0 PSL Sphere Diameter (µm) Figure 8 Theoretical and Experimental LPS Response 22 (a) (b) Figure 9 Average of 60 1-minute Atmospheric Size Distributions in Tsuen Wan: (a) Weekday (b) Weekend 23 (a) (b) Figure 10 Average of 60 1-minute Atmospheric Size Distributions in Yuen Long: (a) Weekday (b) Weekend 24 (a) (b) (c) (d) Figure 11 Example of an atmospheric nucleation event observed at Yuen Long: (a) Before nucleation (b) During nucleation event (c) Agglomeration of freshly nucleated aerosol (d) Back to initial sizze distribution before nucleation took place 25 (a) (b) Figure 12 Average of 60 1-minute Atmospheric Size Distributions in Kawasaki: (a) Weekday (b) Weekend 26 (a) (b) Figure 13 Average of 60 1-minute Atmospheric Size Distributions in Noge Park: (a) Weekday (b) Weekend 27 28