Power Supply Trouble Shooting The unit is powered by an internal rechargeable battery which gives up to 70 hours use between recharges. The battery has an expected service life of 200 charge cycles. Electronic faults with the Apalert are extremely rare. Problems are more likely to be due to one of the following. A recharge adaptor is supplied and the unit may be used concurrently with recharging. Sensor Failure L.E.D. on front panel. Indicated by total inability to respond to breath. Intermittent failure will cause random signals not correlated to breaths. These symptoms may be also caused by contamination of plug connecting sensor lead to machine. Consider cleaning plug with alcohol and cloth. Having a spare sensor on hand may avoid inconvenience. • 80% discharge indicated by digital display shutdown. Recharge Adaptor Failure • 100% discharge indicated by audible alarm. Low Battery Warnings • 60% discharge indicated by flashing yellow Recharging The recommended recharging power source is a 9 volt transformer producing 50 to 300 mA. Connect the recharge source to the power socket on the rear panel. The recharge indicator L.E.D. should flash to confirm correct operation. If the unit is being used concurrently the recharge indicator will not flash but the digital display changes mode from intermittant to constant illumination. With 50mA source recharging will take 14 hours. This reduces to 10 hours with higher powered sources. Over-charging cannot cause damage. Indicated by absence of red L.E.D. flashing when attempting to recharge battery. If transformer supplied has adjustable voltage ensure it is set at 9 volts. It may also be possible that polarity on the Recharge Transformer can be reversed so check this also. Veterinary Respiration Monitor Model RM5 Battery Worn Out Indicated by unit not running at all or only for short time after full recharge. Note : Display non-functioning and alarm continuous are symptoms of discharged battery but may be mistaken for a fault. Dealer The battery will self-discharge in 4 to 6 months. If it is allowed to remain completely discharged for any length of time it may become unserviceable. Sensor Sterilisation The sensor element is resistant to damage from most chemical sterilisation agents. The recommended treatment is Chlorhexidine in normal dilutions with or without alcohol. Options Zero Dead Space Sensor Miniature sensor fits directly into special ET tube adaptors or Y-piece (all supplied). Avoids adding to patient dead space. Allows monitoring of patients down to 250 gm bodyweight. Operating Instructions www.apalert.net Congratulations!! The purchase of your Apalert Respiration Monitor will change forever your approach to anaesthesia. The Apalert is the most advanced and comprehensive respiration monitor ever designed for veterinary patients. Almost all veterinary anaesthetic drugs depress respiration in a dose related manner. This makes respiration monitoring an ideal tool to assess anaesthetic depth. Most anaesthetic emergencies can be attributed to human error, simply a lack of vigilance. The Apalert is always vigilant, not only preventing deaths but allowing better management of every anaesthetic. As well as apnoea warning, the Apalert can signal the occurrence of slow or shallow breathing giving early notice of impending problems. It can also draw the anaesthetists attention to other problems such as hypothermia, exhausted gas supply and blockages or leakages in the gas anaesthesia equipment. Manual Alarm Reset Low Battery Indicator Respiration Indicator Green LED flashes with each breath. Yellow LED flashing indicates battery is 60% discharged. Digital Display Indicates time elapsing between breaths. Display is turned off when battery is 80% discharged though machine continues to operate as normal. Control Switch Activates Apalert and selects length of apnoea to trigger alarm. Recharge Indicator Sensitivity Control Volume Control Adjusts level of sound from audible signals. Sound level increases automatically if alarm is not reset. Adjusts sensitivity of Apalert to suit size of animal. Refer to sensitivity table to select correct adjustment. Features • • • Robust and reliable sensor, unaffected by moisture. Sensitivity adjusts to all patients down to 1 kg. Optional sensor available for patients as small as 250 gm. Audible apnoea alarm. • • Alarm Tones 1) Connect sensor to patient airway using adaptor provided. The audible apnoea alarm has 6 short pulses and a pause each second. 2) Activate Apalert and select time required with Control Switch. The display will begin counting seconds. The audible low battery alarm has 3 long pulses and a pause each second. For correct operation there must not be any leakage in the patients airway connections, particularly around the ET tube cuff. 3) Select and adjust sensitivity as detailed in Sensitivity Table below. 4) Adjust volume if desired. Controls & Indicators Small window on top surface contains optical switch. Reset apnoea alarm by passing hand or elbow above this window. Operation Audible and visual breath indicators. Apnoea Alarm adjustable from 10 to 60 seconds. • Digital display indicates time between breaths. • Hands free alarm reset. • • • Intensity of audible signals is adjustable. Alarms automatically increase in volume if ignored. Rechargeable battery. Indicators warn of 60%, 80% and 100% battery discharge. 5) Each breath should produce a short audible beep and a flash of the green L.E.D. after which the digital time display will reset to zero. Manual Alarm Reset Activated by a shadow falling on the optical switch. Move hand or elbow several centimetres above the machine to achieve this. The movement should be brisk - the switch is sensitive to rate of change in light intensity not the amount of change. The manual reset only operates when the alarm is active. It does not prevent Apalert use in low light. Alarm Procedure If there is no breath within the preset time, the apnoea alarm commences. A breath will reset the alarm or it may be reset manually. Alarm indicates breath has not been detected within preset period and the status of the patient should be evaluated. Check the patients vital signs. Consider why the alarm may have activated. Sensitivity Setting i) Respiratory depression from anaesthesia overdose. Experience will be the guide to setting the sensitivity control. To begin, select from the following table. SENSITIVITY TABLE The sensor is a thermal Patient Sensitivity transducer triggered by the temperature change 1 Kg 10 between inspired and expired gas. Greater 5 Kg 8 sensitivity is required at 10 Kg 6 higher ambient 20 Kg 4 temperature or when the 50 Kg 2 patient is hypothermic. 100 Kg 1 If the machine does not respond to breaths at the recommended setting, increase sensitivity one unit at a time until breaths are registered. If breaths are registered at the suggested setting, reduce setting 1 unit at a time until the machine no longer responds. Then increase sensitivity 1 unit. ii) Physiological apnoea, for example following hyperventilation. iii)Fault with anaesthesia equipment preventing patient breathing. iv)Sensor disconnected from patient or machine. v) Leakage from around cuff of E.T. tube. vi) Patient hypothermia. Decide if some action is needed. Possibly the machine needs re-adjusting to a longer time setting or higher sensitivity. During longer procedures an increase in sensitivity will be needed. Reasons include patient temperature falling, gas temperature in circuit increasing and respiratory pattern changing to deeper breaths less often. Installation IMPORTANT This procedure only takes a short time at the beginning of each anaesthetic and ensures the Apalert is adjusted to the patient. If the patients respiratory tidal volume drops significantly during the anaesthetic the machine can detect this and sound the alarm. A mounting bracket is supplied. It is suggested the bracket be used to attach Apalert onto anaesthesia machine, then being safe from accident damage and yet immediately ready to be used.