DIL 402 PC Operating Instructions Operating Instructions DIL 402 PC Contents 02/05 Chapter I General Information Chapter II Installation Chapter III System Components Chapter IV Operation Chapter V Appendix J:\Dilatometer\DIL402PC\neu\content_e.doc DIL 402 PC Chapter I General Information 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc DIL 402 PC General Information Information In the design of your instrument, we endeavor to take individual solutions into account and to include these in the documentation. However, in order to keep the scope of the technical documentation at a reasonable level, we must limit the description to a standard model. We ask for your understanding, if additional information particular to your instrument is not included within the scope of the standard instructions. This additional information can always be found on the corresponding information sheets. Prior written permission from NETZSCH-Gerätebau GmbH is required for electronic or mechanical duplication and distribution of these instructions. All technical data, instrument features and other information described in these operating instructions are presented to the best of our knowledge and in accordance with the technical standards of the instrument at the time of printing. We welcome any comments, suggestions or new ideas concerning the instrument and these operating instructions. Please address them to: NETZSCH-Gerätebau GmbH Wittelsbacherstraße 42 D - 95100 SELB Telephon: 09287/881- 0 Telefax: 09287/881- 144 E-Mail: at@ngb.netzsch.com Internet: http://www.netzsch-thermal-analysis.com Maintenance and service must be carried out by NETZSCH Customer Service personnel. A service contract is available for our customers. This instruction manual is provided to give the customer information on proper operation of the instrument. NETZSCH-Gerätebau GmbH will accept no liability for damage resulting from improper use. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc 1 DIL 402 PC General Information Notes on the Use of this Manual In this manual, the symbols described below are used to simplify orientation. NOTE This sets particularly important information apart from the rest of the text. ATTENTION! These instructions must be followed exactly to avoid injury to the user and damage to the instrument. This symbol refers to more detailed information which can be found elsewhere, e.g. in the Software Manual. The tools listed after this symbol are required for the installation or modification of your instrument. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc 2 DIL 402 PC General Information Basic Safety Information State-of-the-Art • Your instrument has been produced with state-of-the-art technology and is safe to operate. Authorized Operation • Any operation of the instrument other than as authorized requires consultation with NETZSCH. Any use exceeding the (expanded) authorized operation is considered unauthorized. The manufacturer will not be liable for any damage resulting from such use. Manufacturer's Requirements • Authorized operation of the instrument includes compliance with manufacturer's requirements regarding installation, commissioning, operation and maintenance. Training of Personnel • Your instrument may only be operated and maintained by authorized, trained and individually instructed personnel who have been informed of possible sources of danger. Responsibility • Responsibility for commissioning, operation and maintenance must be clearly defined and compliance must be ensured. The resulting responsibilities can be clearly resolved under the aspect of safety. Unauthorized Access • The operator must make sure that only trained personnel work on the instrument. Improper Operation • Any type of operation which reduces the safety of the user and the operability of the instrument should be avoided. Unauthorized Changes to the System • Unauthorized modifications and changes which effect the safety of the instrument are not permitted. Obligation to Report Changes to the System • The user is obligated to immediately report to the operator in charge any changes to the instrument which effect the safety of the system. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc 3 DIL 402 PC General Information Maintenance Obligation • The operator of the system must ensure that the instrument is operated in perfect condition at all times. Proper Set-up of the Work Stations • The operator must guarantee the openness and cleanliness of the work stations at the instrument through appropriate instruction and inspections. Shut-down • For all maintenance work, the instrument must be switched off and unplugged. Removal of Protective Devices • Protective devices may only be removed when the instrument is switched off and unplugged. It is imperative that the protective devices be replaced before starting the instrument. Checks following Maintenance or Repair Work • After maintenance or repair work, a check should be made to ensure that all protective devices are in place and operate properly. Only then should the instrument be started. Industry-specific Accident Prevention Regulations • The operator must observe the relevant regulations and protective measures when handling the required gases. In addition, possible reactions with the materials used must be considered. At operating temperatures above 55°C, protective gloves must be worn. In any case, the industryspecific and local accident prevention regulations are also valid for the instrument. Disposal of Production Materials • Production materials are to be disposed of according to local regulations. Maintenance or Repair • Products sent in for maintenance or repair should, to the extent possible, be free of harmful substances (e.g. radioactive, toxic, caustic or microbiological materials). Otherwise, the type of contamination must be declared. Products not explicitly declared to be "free of harmful substances" will be decontaminated at the expense of the sender. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc 4 DIL 402 PC Operating Instructions General Information • Using the Operation Manual, the operator should prepare operating instructions which specify the actions and tasks required for safe operation. The operating instructions should be placed in a suitable location in the work place and complied with by the employees. Gases • Observe the relevant regulations and protective measures when handling the required gases. Consider possible reactions with the materials. Warning signs should be hung in the appropriate places in accordance with national and regional regulations. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 1.doc 5 DIL 402 PC Chapter II Installation 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc DIL 402 PC Installation CONTENTS INSTALLATION ......................................................................................................1 PACKING AND DELIVERY ............................................................................................... 1 LOCATION REQUIREMENTS ............................................................................................ 2 INSTALLATION SCHEMATIC............................................................................................. 3 ELECTRICAL CONNECTIONS ............................................................................................ 4 IMPORTANT NOTES FOR SOFTWARE INSTALLATION ......................................................... 5 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc DIL 402 PC Installation Installation In most cases, your DIL 402 PC is set up and commissioned by one of our customer service engineers. If you would like to set up your instrument yourself, please read the following sections. Packing and Delivery − We deliver all components in separate cartons. The measuring unit, electronics, computer system etc. are packed in formmolded foam to protect against damage during transport. − We recommend keeping the original cartons. − Should repairs or an extension of the system become necessary, the cartons can be reused, thus ensuring a relatively safe return to the manufacturer. − Prior to shipping, we carefully test all components of the system for mechanical and electrical operability. − After unpacking, please check all delivered components for possible transport damage and compare the individual items against the delivery note supplied. Should an item be missing, please contact us immediately. Telephone: Telefax: E-Mail: 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc 0049 9287/881- 155 (Service Hotline) 0049 9287/881- 144 service@ngb.netzsch.com 1 DIL 402 PC Installation Location requirements Select the optimal component arrangement for the space you have available. The location of your instrument must meet the following requirements: constant temperature conditions (room temperature) to the extent possible no direct sunlight on the instruments dust-free environment to the extent possible stable tables measuring system A stable, shock-absorbent table with a working surface of approx. 2.5 m x 1 m, is required for the measuring device (measuring unit, computer, printer). An installation schematic of the measuring system follows on the next page. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc 2 DIL 402 PC Installation Installation schematic 980 depth 450 470 500 depth 500 computer weight 10.5 Kg printer weight: 8 Kg 270 500 810 depth 260 measuring part 402 PC wide range power pack 100-240 V, 6 A weight: 20 Kg table monitor weight: 3.5 kg wide range power pack 100-240 V, 6 A 365 depth 500 200 option or provided by the customer power unit 230 V, 16 A weight: 36 Kg floor Figure 1: Optional arrangement – DIL 402 PC Figure 1 shows an arrangement of the instruments which has proven to be practical. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc 3 DIL 402 PC Installation Electrical connections • Connect the measuring unit, power unit and computer system to one another with the supplied cables according to Figure 2. Connect the power unit on a separate electrical phase. • Plug the measuring part, power unit and computer system into the power supply. L3 L2 L1 N PE fur nace c ontrol U SB pow e r s upply furnace control Figure 2: Cable diagram electrical connections 02/05 J:\Dilatometer\DIL402PC\neu\chapter 2.doc 4 DIL 402 PC Installation Important notes for Software installation Your DIL 402 PC instrument communicates with the PC using the USB interface of the PC. This requires the installation of suitable drivers delivered together with the Proteus Software. → Before initially connecting the DIL 402 PC the Proteus Software must be installed first. For details of installation, please read chapter 3 and chapter in the document “NETZSCH Proteus Software”. Notes 02/05 Do not use USB hubs (connect your DIL 402 PC directly to the PC). Use the original USB cable delivered from NETZSCH. If you accidentally change the USB interface on the PC the Plug&Play procedure will start again to install the drivers for this interface. Please note that this procedure has to be done for any “not yet” connected USB interface. J:\Dilatometer\DIL402PC\neu\chapter 2.doc 5 DIL 402 PC Chapter III System Components 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc DIL 402 PC System Components CONTENTS SYSTEM COMPONENTS ..........................................................................................1 MEASURING UNIT ........................................................................................................ 2 PRINCIPLE OF OPERATION .............................................................................................. 3 FRONT PANEL .............................................................................................................. 4 SAMPLE HOLDER.......................................................................................................... 5 SAMPLE THERMOCOUPLE .............................................................................................. 7 FURNACE - MIDDLE RANGE (6.215.2-15) .................................................................. 8 FURNACE - HIGH RANGE (6.215.2-26) ....................................................................... 8 PURGE GAS CONNECTION .............................................................................................. 9 POWER UNIT ............................................................................................................. 10 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc DIL 402 PC System Components System Components The DIL 402 PC measurement device consists of the following components: • Dilatometer (measuring unit) • Power unit • Computer system, screen, printer 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 1 DIL 402 PC System Components Measuring unit tube adjusting screw slider knob furnace support base plate control panel Figure 1: Measuring unit (closed) sample holder Figure 2: Measuring unit (open) 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 2 DIL 402 PC System Components Principle of operation sample thermocouple displacement transducer (LVDT) sample carrier pushrod control thermocouple on sample furnace power position 0 DIL 402 PC base plate / control panel Figure 3: Schematic diagram of DIL 402 PC (measuring unit closed) The sample is placed in the homogenous temperature zone of the furnace. The furnace is heated according to a preselected temperature program. The furnace temperature is controlled by the control thermocouple. The sample is submitted to a temperature-dependent change in length. The sample temperature is measured by the sample thermocouple. Sample carrier and pushrod are also submitted to a change in length. Thus there is measured and registered the sum of change in length of the sample, the sample carrier and the pushrod. The pushrod transmits this change in length mechanically to the displacement transducer (LVDT) and causes a displacement of the core of the displacement transducer (LVDT). The resulting change in voltage is transformed by a carrier frequency measuring amplifier to a d.c.voltage which is proportional to the displacement. The signal is is registered by a computer. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 3 DIL 402 PC System Components Front panel adjusting screw control panel on power position 0 DIL 402 PC Figure 4: control panel Control panel Label Function on Control LED: power supply to measurement unit on/off power Control LED: furnace heating on/off position LED´s: pushrod position Adjusting screw: The position of the pushrod can be adjusted with the adjusting screw. The direction of the movement is indicated by the luminous band. For starting a measurement the pushrod should be positioned so that only the LED in the middle (0) is on (central position of the displacement transducer, measurement range ± 2.5 mm). 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 4 DIL 402 PC System Components Sample holder The dilatometer can be equipped with a tube sample holder (Figure 5). The sample holder is available in aluminum oxide or fused silica. Because of its low coefficient of expansion, fused silica is especially suitable for the determination of very small length changes. The rod sample holder stands out due to a very homogeneous temperature distribution around the sample. Tube-type sample holder sample push rod tube on power position 0 DIL 402 PC Figure 5: Tube sample holder 02/05 − At temperatures above 900°C, fused silica tends towards irreversible transformations. Do not use the fused silica sample holder for the absolute determination of the expansion coefficient for measurements above 900°C. − Analyses, which serve only to compare individual samples, can also be run with the fused silica sample holder to 1100°C. − If application areas overlap (absolute determination, comparative measurements), we recommend acquisition of separate sample holder systems. J:\Dilatometer\DIL402PC\neu\chapter 3.doc 5 DIL 402 PC A pushrod System Components spacers sample B pushrod sample supports Figure 6: (A) spacers prevent the sticking from the sample, B supports for positioning of the sample (for samples: Ø 4mm, 6 mm and 8 mm) Special sample holders are available for testing pasty and powdery samples. The sample is placed between two stamps in a cylindrical container. container sample rod Figure 7: Sample holder for powdery and pasty samples 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 6 DIL 402 PC System Components Sample thermocouple • located in the sample holder • Pt-Pt/Rh thermocouple (Type S) • flexible, can be adjusted to different sample positions sample thermocouple sample Figure 8: Location of the sample thermocouple 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 7 DIL 402 PC System Components Furnace - Middle Range (6.215.2-15) insulation heating element protective tube Figure 9: Schematic of the middle range furnace • temperature range 25 ... 1200°C • tube furnace with resistance coiling • control thermocouple type S • recommended heating rate for dilatometer measurement: < 5 K/min • electrical supply: max. 180 V Furnace - High Range (6.215.2-26) insulation heating element protective tube Figure 10: Schematic of the high range furnace • temperature range 25 ... 1600°C • tube furnace with resistance coiling • with integrated protective tube for gas flow • control thermocouple type S • recommended heating rate for dilatometer measurement: < 5 K/min • electrical supply: max. 125 V 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 8 DIL 402 PC System Components Purge gas connection Measuring system and sample chamber can be purged with gas. The purge gas inlet is placed on left cover plate of the measuring system. Using a purge gas (dry, inert) avoids the contamination of the measuring system by reaction products evolved from the sample. Recommended gas flow rate: 100 – 150 ml/min Maximum input pressure: 0.5 bar purge gas connection gas supply gas flow (6.217.1-40) Figure 11: connecting the gas flow 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 9 DIL 402 PC System Components Power unit Dimensions in “mm”: (365 x 200 x 500) 1 2 3 4 5 Figure 12: Power unit - front 6 7 8 9 Figure 13: Power unit - rear 02/05 J:\Dilatometer\DIL402PC\neu\chapter 3.doc 10 DIL 402 PC 02/05 System Components 1. LED – "on line" instrument switched on /off; LED on /off 2. LED – "power" – indicates furnace heating 3. LED – "fuse" – indicates a defective fuse in the power pack 4. LED – "limit" – indicates an exceeding of the preadjusted current limit value 5. main switch (power unit on/off) 6. fan (instrument cooling) 7. furnace operation (plug connection from measuring unit) 8. furnace control system (connection to measuring unit) 9. AC line /230V (connection to power supply) J:\Dilatometer\DIL402PC\neu\chapter 3.doc 11 DIL 402 PC Chapter IV Operation 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc DIL 402 PC Operation CONTENTS OPERATING THE INSTRUMENT .................................................................................1 MEASUREMENT PREPARATIONS ............................................................................1 SAMPLE PREPARATION .......................................................................................1 SOLID SAMPLES ........................................................................................................ 1 PASTY SAMPLES ....................................................................................................... 2 POWDERY SAMPLES .................................................................................................. 2 INSTRUMENT PRESETTINGS ..................................................................................3 INSERT SAMPLE ........................................................................................................ 3 POSITIONING THE THERMOCOUPLE .............................................................................. 4 CLOSE MEASURING UNIT ........................................................................................... 4 GENERATION OF THE REQUIRED SAMPLE ATMOSPHERE (SAFETY NOTES) ............................ 5 START A MEASUREMENT .....................................................................................8 TROUBLE SHOOTING LIST ......................................................................................14 DIAGNOSIS .....................................................................................................15 OTHER SCREEN MESSAGES ................................................................................16 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc DIL 402 PC Operation Operating the instrument Measurement preparations Carefully select the test conditions prior to the measurement according to the composition of the sample material and the expected effects. The following parameters should be given special attention during preparation for the measurement: • sample preparation • temperature program • sample chamber atmosphere Sample preparation Sample composition: solid, plastic, powdery Form of sample: predominantly cubic or cylindrical, triangular or n-angular also possible Sample dimensions: rod sample holder: max. ∅ 8 mm, length: 0 to 25 mm (25 to 50 mm optional) tube sample holder: max. ∅ 12 mm, length: 0 to 25 mm (25 to 50 mm optional) Solid samples • The sample is sawn or cut from a larger piece of material. It can also be pressed or cast. The preparation of the sample can affect the measurement results. Information on sample preparation specifications can be found in the existing test standards (e.g. DIN 51045, DIN 52328, DIN 53752). • Grind the end surfaces of the sample plane parallel and smooth. Depending on the degree of hardness, use either emery paper or a grinding machine. Make sure that the end surfaces are even and at right angles to the longitudinal axis of the sample and free of ridges and foreign particles. • For measurement of the sample length, use a slide caliper rule or micrometer screw. According to DIN 51045, the sample length must be determined with a precision of 0.002 L0 (L0 = initial length of the sample at 20°C). 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 1 DIL 402 PC Operation Pasty samples These samples must be molded and predried. Powdery samples These sample can be pretreated by: • pressing • binding with adhesives or plastic clays (followed by molding and predrying) A special sample holder for testing pasty and powdery samples is shown in Figure 1 . The sample is placed between two stamps in a cylindrical container. container sample stamps Figure 1: Sample holder for powdery and pasty samples The sample holder is available in graphite, alumina and fused silica. picture 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc material order-no. alumina 6.219.1-60.1 fused silica 6.219.1-60.2 sapphire 6.219.1-60.3 2 DIL 402 PC Operation Instrument presettings • Switch on computer, measuring unit and printer about one hour before starting the measurement. Insert sample ATTENTION! Before opening the dilatometer, always make sure that sample temperature is no higher than 100°C. • Slide the furnace to the right to the limit stop. • Slide the pushrod to the left. • If necessary, remove the sample from the previous measurement. • If required, place the support rings in the sample holder. • Use tweezers to insert the sample. NOTE The pushrod is a sensor for recording the length-change signal. The sample is not held in place by the pushrod. The sample must not tilt when in the measurement position. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 3 DIL 402 PC Operation Positioning the thermocouple • Carefully bend the sample thermocouple as close to the center of the sample as possible. ATTENTION! The thermocouple must not touch the sample - danger of alloy formation or chemical corrosion. The thermoelectric wires must not come in contact with one another (except at the bead hot-junction). Close measuring Unit • Slide the furnace carefully to the left to the limit stop. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 4 DIL 402 PC Operation Generation of the required sample atmosphere (Safety Notes) ATTENTION! Under no circumstances may explosive gas mixtures be used. Information concerning the possibility of using various gases can be found in the appropriate technical literature or can be requested from the manufacturer or retailer of gases. The following is a selection of purge gases including their limiting properties. The list makes no claim on being complete. Helium (He) Chemically inert (inert gas) No technical safety limitations Argon (Ar) Chemically inert (inert gas) Asphyxiating No technical safety limitations Argon should not be used for low temperature investigations Nitrogen N2 Largely inert Asphyxiating No technical safety limitations, but sample reactions are possible in the high temperature range. Air Oxidizing No technical safety limitations, but sample reactions are possible. Application is possible above room temperature (maximum application temperature is determined by furnace material, sample holder type, crucible material). Oxygen (O2) Increases flammability (must not come into contact with fats or oils) No technical safety limitations, but hefty reactions are possible with the sample. Application is possible above room temperature (maximum application temperature is determined by furnace material, sample holder type, crucible material). 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 5 DIL 402 PC Operation Carbon dioxide (CO2) Non-toxic (MAK value 5000 vpm), Non-flammable Application is possible above room temperature. Hydrogen (H2) The instrument is not destined and suitable for use under hydrogen atmosphere. It can happen that explosible mixtures are formed inside the system during the respective experiments. Ammonia (NH3) Toxic (MAK value 50 vpm), Flammable Danger of explosion when in contact with oxygen. For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Corrosion of seals is possible. Carbon monoxide (CO) Toxic (MAK value 10 vpm), Flammable Corrosive For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Hydrogen sulphide (H2S) Toxic (MAK value 10 vpm), Flammable, corrosive For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Other reducing gases or gas mixtures For safety reasons it is not possible to use these gases as contact with oxygen or air could result in an explosive compound. However, certain gas compounds with inert gases are possible where the mixture ratio excludes all technical safety doubts while complying with the required analytical conditions. The manufacturers of gases and gas mixtures can provide the required information. Chlorine (Cl2) Very toxic (MAK value 5 vpm), Non-flammable, corrosive, caustic For safety reasons it is forbidden to use an instrument that has been installed in the normal way. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 6 DIL 402 PC Operation Hydrogen chloride (HCl) Toxic (MAK value 5 vpm), Non-flammable, corrosive, caustic For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Sulphur dioxide (SO2) Toxic (MAK value 2 vpm), Non-flammable, corrosive For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Fluorine (F2) Very toxic (MAK value 0.1 vpm), Encourages burning, corrosive, caustic For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Hydrogen fluoride (HF) Very toxic (MAK value 3 vpm), Very caustic, corrosive For safety reasons it is forbidden to use an instrument that has been installed in the normal way. Gaseous hydrocarbon Can form explosive gas mixtures with air For safety reasons it is forbidden to use an instrument that has been installed in the normal way. MAK values 1998 We recommend that the expelled purge gas be passed into a suitable exhaust hood. The measuring part must be sufficiently purged with inert gas once the measurement is finished. • Set the required flow rate for the purge gas on the flow meter (optional). • Enter the measurement parameters (software). • Start the measurement program. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 7 DIL 402 PC Operation Start a measurement • set up a measurement, • create a correction curve, • and then to measure a sample. Switch on the computer system. Switch on the power unit. Prepare the sample. Determine the sample length at room temperature. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 8 DIL 402 PC Operation Slide the furnace to the right limit stop. If necessary, place sample supports in the sample holder. Place the standard sample (e.g. Al2O3 sample) on the sample supports. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 9 DIL 402 PC Operation Using the adjusting screw, slide the pushrod towards the sample. Position the thermocouple close to the sample. The thermocouple must not come into contact with the sample. Slide the furnace carefully to the left to the limit stop. The sample holder must not come into contact with the protective tube. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 10 DIL 402 PC Operation Optional: Open the purge gas supply. purge gas connection gas supply gas flow (6.217.1-40) Start the measurement software. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 11 DIL 402 PC Operation File Menu New Proceed as follows to create a correction curve with a standard (e.g. Al2O3). DIL 402 PC Measurement Header Measurement Type: Correction Open Temperature Recalibration Open Table for Sample Holder Material (e.g. Al2O3.scl) Open Table for Calibration Material (e.g. Al2O3.scl) Define Temperature Program Define File Name The program switches to the Adjustment Dialog Set the transducer position, measuring range and temperature threshold values in the Adjustment dialog box. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 12 DIL 402 PC Operation Start the measurement. File Menu Open Use File from Correction Measurement! Define Test Parameters: Measurement Type: Sample+Correction Run a sample measurement according to the following flow chart, including the correction measurement: Open Temperature Recalibration Accept Temperature Program Define File Name Continue with Adjustment Dialog After completion of the measurement, you can begin the analysis. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 13 DIL 402 PC Operation Trouble shooting list error sporadic or permanent jumps in the dL curve (especially in the cooling curve) results seem to be incorrect possible reasons end faces of the sample not plane grind sample plan parallel sample is tilted put in sample again sample holder is defect (cracks) change sample holder thermocouple touches the sample readjust thermocouple sample holder may be contaminated by reaction products evolved by the sample measure standard sample (sapphire, Al2O3) contamination of sample thermocouple check thermojunction (magnifying glass) expansion of the measuring system (correction curve) wasn’t considered make a new measurement using a correction curve heating rate > 10 K/min make a new measurement with heating rate 5 K/min furnace does not heat check power supply of the power unit program „dies“ or „hangs up“ 02/05 possible remedy change fuse if necessary LED fuse (power unit) lights change thyristor fuse LED limit (power unit) lights call NETZSCH-customer service computer, and power unit are connected to the same phase reconnect the instruments: computer, to one phase, power unit to another phase install line filter trace reasons for trouble in surroundings: refrigerators, defect neon lamps, radiators J:\Dilatometer\DIL402PC\neu\chapter 4.doc 14 DIL 402 PC Operation Diagnosis In the status line of the computer screen there will appear several messages during normal operation or failure conditions (the values in the middle and right columns are only examples): 1. OFF 2. INITIALIZATION 3. ON e.g. 10,5 µm 21.0°C ON OVF 21,3°C ON UNF 22,0°C Ext. Res. 10,5 µm 21,0°C Ext. Res. 10,5µm UNF 4. 5. 6. 7. Causes: 1. Normal display for some seconds after connecting the power supply to line. Power supply not on line or defective. See LED bar „on line“ on the front panel of the 402 PC USB cable not connected. 2. Normal display after some seconds. Computer is looking for communication with dilatometer electronics. 3. Communication successful. Micrometers are displayed (e.g. 10.5 µm). 4. Dilatometer in (+) overflow. Switch to 5000 µm range and turn micrometer-screw clockwise. Problem with carrier frequency amplifier signal path. 5. Dilatometer in (-) underflow. Switch to 5000 µm range and turn the micrometer-screw counterclock-wise. Problem with carrier frequency amplifier (see 4.) No sample. 6. External reset. Control thermocouple not connected or interrupted (21.0°C temperature sample thermocouple). Power unit switched off. Power unit disconnected from 37 polar sub-D-socket. Transformer in power unit overheated. 7. External reset. Sample thermocouple not connected or interrupted. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 15 DIL 402 PC Operation Other screen messages When starting the measurement, one of the following pictures may appear on the screen: Causes: - most probable furnace not closed - switch defective or wiring interrupted. Furnace 1 = 1200°C furnace (middle) Furnace 2 = 1600°C furnace (high) Causes: Setup mismatch. Furnace changed without correction of the instrument setup. Go into instrument setup and select the furnace you want to use. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 4.doc 16 DIL 402 PC Chapter V Appendix DIL 402 PC Appendix CONTENTS TECHNICAL DATA ..................................................................................................1 ASSEMBLY OF THE SAMPLE HOLDER SYSTEM AND THERMOCOUPLE.................................2 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc DIL 402 PC Appendix Technical data Dilatometer 402 PC Electronic Dilatometer with digital data acquisition and digital control Measurement path: ± 2.5 mm Resolution of data acquisition: 16 Bit Measuring ranges: 5000/500 µm Contact pressure: 15-45 cN (25 cN preadjusted) Interface to PC: USB Type of furnace control: digital controller Sample holder Type: tube sample holder / rod sample holder Material: fused silica (RT … 1100°C) Al2O3 (RT … 1600°C) Sample dimensions Length: 0-25 mm (25-50 mm possible) Diameter: 3-12 mm Atmosphere in the sample chamber: dynamic gas flow (inert gases only) Temperature range: 20 ... 1600°C (different furnaces) middle range RG: 20 ... 1200°C high range 1: 20 ... 1600°C Heating rates: recommended for dilatometer measurements < 5 K/min Dimensions: width: approx. 810 / 980 mm Measuring unit (closed) height: approx. 270 mm depth: approx. 260 mm Power supply: 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 100-240 V, 50/60 Hz 1 DIL 402 PC Appendix Assembly of the sample holder system and thermocouple The sample holder system consists of following components: sample thermocouple pushrod sample holder tube A sample thermocouple B pushrod C sample holder tube Check that all parts are provided. 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 2 DIL 402 PC Appendix Assembly sample holder tube 1. Push the sample thermocouple through metal bush the hole into the metal bush. ATTENTION! Turn the thermocouple so that when looking from above the wires are placed horizontally. top view sample thermocouple wires of the sample thermocouple correct sample sample holder tube top view wires of the sample thermocouple incorrect sample sample holder tube 2. Affix the sample thermocouple on the metal bush with the allen screw. allen screw 02/05 metal bush J:\Dilatometer\DIL402PC\neu\chapter 5.doc 3 DIL 402 PC Appendix sample thermocouple 3. Affix the sample thermocouple on the wire sample holder tube by means of a short Pt wire. sample holder tube adjusting screw 4. Pull out the knob of the adjusting screw. knob 5. Loosen the hexagon head screw by means of a socket wrench. socket wrench adjusting screw 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 6. Remove the adjusting screw. 4 DIL 402 PC Appendix tube 7. Loosen both allen screws on the tube. allen screws tube 8. Remove the tube (push it forwards). ATTENTION! While removing the tube hold it tight with both hands so that it cannot fall onto mechanical parts of the measuring system! furnace 9. Loosen the allen screw on each side of the furnace. allen screw allen screw 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 5 DIL 402 PC furnace sample holder tube Appendix 10. Remove the furnace. 11. Push the pushrod into the sample holder tube. pushrod sample holder tube 12. Push the sample holder tube with pushrod into the measuring system. ATTENTION! Take care that the open area of the sample holder tube is placed parallel to the base plate! 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 6 DIL 402 PC Appendix allen screw 13. Affix the sample holder tube with the allen screw. cable fixing 14. Pass the sample thermocouple connecting cable as shown in the figure and connect the plug. plug 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 7 DIL 402 PC Appendix pushrod 15. Affix the pushrod on the displacement device with the allen screw. allen screw Adjustment of the pushrod: Height: allen screws • Loosen the allen screws and you can move the pushrod upwards and downwards Lateral position: allen screws • Loosen the three allen screws and you can change the lateral position of the hinge and pushrod. allen screw hinge Ideal position of the pushrod: pushrod sample support sample 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 8 DIL 402 PC Appendix furnace 16. Place the furnace on the set pins and fix it with the allen screws. furnace allen screw allen screw 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 9 DIL 402 PC Appendix tube 17. Push the tube over the instrument. tube 18. Affix the tube with the allen screws. allen screws adjusting screw 19. Place the adjusting screw. 20. Affix the adjusting screw by means of a socket wrench. socket wrench 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 10 DIL 402 PC Appendix adjusting screw 21. Push in the knob. knob 02/05 J:\Dilatometer\DIL402PC\neu\chapter 5.doc 11