CORTECS 2.7 µm Columns CONT ENT S I.INT RODUC T ION II. G E T T ING STA RT E D a. Column Connection b. Column Installation c. Minimizing Band Spread Volume d. Column Equilibration e. eCord™ Installation f. Functional Tests for Benchmarking a New Column g. VanGuard™ Cartridge Columns III. COLUMN US E a. Sample Preparation b. pH Range c. Solvents d. Pressure e. Temperature IV. COLUMN C L EANING, R EG EN E RAT ION, AND STO RAG E a. Cleaning and Regeneration b. Storage for Reversed-Phase and HILIC Columns V. eCO RD INT E L LIG ENT C HI P T EC HNOLOGY a. Introduction b. Installation c. Column Use Information V I. A DDIT IONA L INFO RMAT ION a. Tips for Maximizing Column Lifetime b. Getting Started with CORTECS® HILIC Columns c. Troubleshooting Questions V II. C AUT IONA RY NOT E I. INT RODUC T ION Thank you for choosing a CORTECS Column. CORTECS 2.7 µm Columns are based on the same solid-core particle technology as CORTECS UPLC® 1.6 µm Columns, enabling seamless method transfer between UPLC, UHPLC, and HPLC platforms. All CORTECS Columns are manufactured in cGMP, ISO 9001:2000 certified facilities. Strict quality procedures for the manufacture of the solid-core particle, the bonded phase synthesis, and the column packing result in a high quality and reproducible final product. Every column is individually tested and a Performance Chromatogram and Certificate of Batch Analysis are provided on the eCord Intelligent Chip. The eCord is available on all 2.1 mm and 3.0 mm column dimensions. CORTECS 2.7 µm Columns will exhibit maximum chromatographic performance when used on a member of the ACQUITY UPLC® System family since they were created and designed to operate with it. II. G E T T ING STA RT E D c. Minimizing Band Spread Volume Each CORTECS 2.7 µm Column comes with a Certificate of Analysis, which includes the bonded phase batch number and the analytical test results for the unbonded and bonded particle. T he included Performance Test Chromatogram summarizes the performance of each individual column and provides batch number, column serial number, USP plate count, USP tailing factor, retention factor, and chromatographic test conditions. These data should be recorded and stored for future reference. When available, the information can be accessed via the ACQUITY UPLC Console using the attached eCord. Band spreading is the measurement of the system dispersion, which impacts the chromatographic performance. Internal tubing diameter and fluidic connections can significantly impact system band spreading and chromatographic performance. Larger tubing diameters cause excessive peak broadening and reduced sensitivity (Figure 1). 0.005 inches 0.020 inches 0.040 inches a. Column Connection CORTECS 2.7 µm Columns are designed to operate on any HPLC, UHPLC, or UPLC system. Due to the absence of an industry standard, be aware that the type of fittings and connection on each system will vary by manufacturer. All tubing connections must be reseated to match the new column when it is installed. If the style of the column end fitting does not properly match the system connections, the chromatographic performance of your new column will be negatively impacted, or leaking can occur. b. Column Installation Note. The flow rates given in the procedure below are described for a 2.1 mm I.D. column. Adjust the flow rate up or down based on the column pressure and system limits. 1. Purge the pumping system of any buffer-containing mobile phases using 100% HPLC-grade water. 2. Purge the pumping system to 100% organic mobile phase (methanol or acetonitrile). 3. Connect the inlet of the column to the chromatographic system. 4. Flush the column with 100% organic mobile phase (methanol or acetonitrile) by setting the pump flow rate to 0.1 mL/min. Increase the pump flow rate to 0.5 mL/min over 5 minutes. 5. Once the mobile phase is flowing from the column outlet, stop the flow. 6. Attach the column outlet to the detector. This prevents air from entering the detector flow cell. 7. Increase the flow rate as described in step 4. Diluted/distorted sample band Figure 1. Impact of tubing diameter on band spread. d. Column Equilibration CORTECS 2.7 µm Columns are shipped in 100% acetonitrile. It is important to ensure mobile phase compatibility before changing to a different mobile phase system. To avoid precipitating mobile phase buffers within the column or system, flush the column with five column volumes of a water/organic solvent mixture, using the same, or lower, solvent content as in the desired buffered mobile phase (for example, flush the column and system with 60% methanol in water prior to introducing 60% methanol/40% buffer mobile phase). Note. If mobile phase additives (i.e., ion-pairing reagents) are present in low concentrations (<0.2% v/v), 100 to 200 column volumes may be required for complete equilibration. In addition, mobile phases that contain formate (i.e., ammonium formate, formic acid) may require extended equilibration times. For reversed-phase separations, equilibrate the column with a minimum of 10 column volumes of the mobile phase to be used (refer to Table 1 for a list of column volumes). The column may be considered fully equilibrated once a constant backpressure and a stable detector baseline is achieved. 8. Monitor until a steady backpressure and baseline have been achieved. CORT ECS 2.7 µm Columns 2 Table 1. Empty Column Volumes (mL) I.D. Column length (mm) 1.0 mm 2.1 mm 30 0.024 0.10 50 0.039 75 0.059 100 0.079 150 0.12 3.0 mm 4.6 mm 0.21 0.50 0.17 0.35 0.83 0.26 0.53 1.25 0.35 0.71 1.66 0.53 1.07 2.49 For CORTECS HILIC Columns, flush with 50 column volumes of 50:50 acetonitrile/water with 10 mM final buffer concentration. Prior to the first injection, equilibrate with 20 column volumes of initial mobile phase conditions. If gradient conditions are used, equilibrate with 8–10 column volumes between injections. Failure to appropriately equilibrate the column could results in drifting retention times. See “Getting Started with CORTECS HILIC Columns” under “Additional Information”. e. eCord Installation eCord technology represents a significant advancement in column usage tracking management when connected to an ACQUITY UPLC System. The eCord Intelligent Chip can be read by connecting the yellow fob to the reader/writer located on the right hand side of the ACQUITY UPLC column heater module. Embedded information such as the column manufacturing QC data and Certificates of Analysis may be accessed via the ACQUITY UPLC console. The eCord is available on all 2.1 mm and 3.0 mm column dimensions. f. Functional Tests for Benchmarking a New Column Waters recommends performing a benchmarking test upon receipt of your column and throughout lifetime usage. By using a standard mix, such as the appropriate QC Reference Material, you can: ■■ Verify the performance of your column upon receipt. ■■ Monitor the health of your column and system over time. ■■ Troubleshoot separation difficulties that arise. 1. Run an initial benchmark test to monitor key performance criteria such as retention time, peak area, peak tailing, resolution, response, system pressure, etc. 2. Repeat test daily to track column and system performance over time. For more information on benchmarking performance, visit www.waters.com/QCRM Note. If an analyte mixture found in the Performance Test Chromatogram is used to benchmark, the isocratic efficiencies measured in your laboratory may be less than those given on the Waters Performance Test Chromatogram. This is normal and expected. The Waters isocratic column testing systems have been modified in order to achieve extremely low system dispersion. This presents a more challenging test of how well the column was packed. This also guarantees the highest quality packed column. These special testing systems have been modified to such an extent that they are not commercially viable and have limited method flexibility other than isocratic column testing. 3. Determine the number of theoretical plates (N) and use this value for periodic comparisons. 4. Repeat the test periodically to track column performance over time. Slight variations may be obtained on different LC systems due to the quality of the connections, operating environment, system electronics, reagent quality, column condition, and operator technique. g. VanGuard Cartridge Columns VanGuard Cartridge Columns are a universal solution to extend all HPLC, UPLC and UHPLC CORTECS column lifetimes. The chemistries offered for VanGuard Cartridge Columns are identical to the available CORTECS chemistries. The reusable holder is designed with an adjustable ferrule so that it can be attached to any column inlet. VanGuard Cartridge Columns come in two sizes, 2.1 mm I.D. for columns I.D.’s 3.0. mm or less, and 3.9 mm I.D. for column I.D.’s 3.9 mm or greater. The Neutrals QC Reference Material (p/n 186006360) or HILIC QC Reference Material (p/n 186007226) are appropriate mixtures with which to benchmark your CORTECS Column and system. Other QC reference materials are available depending on the application and detection mode. CORT ECS 2.7 µm Columns 3 VanGuard Cartridge Column Installation Instructions VanGuard Cartridge Columns are recommended to provided superior column protection for all HPLC and UHPLC columns up to 4.6 mm inner diameter. VanGuard Cartridge Columns require the VanGuard Cartridge Holder (p/n 186007949) for proper assembly. Cartridge Holder Parts list Two Piece Ferrule Holder Fitting Protective Plug Tools required / " wrench / " wrench 3 8 / " wrench 5 16 7 16 O-Ring Compression Screw ATTENTION: READ AND UNDERSTAND ALL ASSEMBLY INSTRUCTIONS PRIOR TO USE. 1. Remove protective o-ring. Keep assembly upright to keep ferrule from falling off. DO NOT remove the protective plug from the holder assembly until Step 4. T he plug is used to seat the tubing securely prior to final assembly. In order to ensure void-free and leak-free connections, the VanGuard Cartridge Holder is shipped with the ferrule NOT permanently attached. Care must be taken when assembling the holder to the column. Transition Tube Assembly Two Piece Ferrule 2. T hread the holder assembly finger tight into the column inlet. 3. With holder on top, tighten holder to the column inlet using wrenches (1/4 turn after finger tight). Hold assembly securely to prevent the holder from slipping from the column inlet. Ensure that the holder is firmly seated into the column inlet before setting the ferrule. HPLC column ⁄ " wrench 5 16 4. Remove plastic plug. 5. Insert guard cartridge into holder. Holder housing should rotate freely until cartridge is tightened. 6. Tighten guard cartridge. ⁄ " wrench 7 16 ⁄ " wrench 38 Guard cartridge Guard Cartridge Replacement Instructions 1. Loosen guard cartridge and remove from holder. 2. Insert new guard cartridge into holder. The guard cartridge can be replaced without removing the holder from the column inlet. 3. Tighten guard cartridge. ⁄ " wrench 7 16 ⁄ " wrench 7 16 ⁄ " wrench 38 ⁄ " wrench 38 CORT ECS 2.7 µm Columns 4 III. COLUMN US E 3. If the sample is not prepared in the mobile phase, ensure that the sample, solvent, and mobile phases are miscible in order to avoid sample precipitation or buffer precipitation. To ensure the continued high performance of CORTECS 2.7 µm Columns, follow these guidelines: a. Sample Preparation 1. Sample impurities and/or particulates often contribute to column contamination. One option to avoid column contamination is to use Oasis ® or Sep-Pak ® solid-phase extraction (SPE) devices. To select the appropriate sorbent for a specific sample type, visit www.waters.com/sampleprep 2. It is preferable to prepare the sample in the initial mobile phase conditions or a weaker solvent for the best peak shape and sensitivity. 4. Filter sample with a 0.2 µm membrane to remove particulates. If the sample is dissolved in a solvent that contains an organic modifier (e.g., acetonitrile, methanol, etc.) ensure that the membrane material is compatible with the solvents in use. Alternatively, centrifuge the sample for 20 minutes at 8,000 RPM, followed by the transfer of the supernatant to an appropriate vial. 5. For hydrophilic-interaction chromatography (HILIC) separations, the samples must be prepared in a high percentage of organic solvent (e.g., 95% acetonitrile). See “Getting Started with CORTECS HILIC Columns” under “Additional Information”. Table 2: CORTECS Chemistry Characteristics Name of column Ligand type Surface charge modification End cap style Carbon load Ligand density pH limits Temp. limits C18 Trifunctional C18 None Proprietary 6.6% 2.7 µmol/m2 2–8 45 °C 5.7% 2.4 µmol/m 2 C18 + Trifunctional C18 + Proprietary 2–8 45 °C Proprietary 4.5% 3.4 µmol/m 2 C8 Trifunctional C 8 None 2–8 45 °C None Proprietary 4.7% 1.6 µmol/m 2 T3 Trifunctional C18 2–8 45 °C Shield RP18 Monofunctional Embedded Polar C18 None Proprietary 6.4% 3.2 µmol/m2 2–8 45 °C Phenyl Trifunctional phenyl None Proprietary 5.9% 3.2 µmol/m2 2–8 45 °C None None None Unbonded n/a 1–5 45 °C HILIC Table 3. Mobile Phase Buffer Recommendations for CORTECS 2.7 µm Columns Volatility (±1 pH unit) Used for mass spec 0.30 Volatile Yes Ion pair additive, can suppress MS signal, used in the 0.02–0.10% range. Acetic Acid 4.76 Volatile Yes Maximum buffering obtained when used with ammonium acetate salt. Used in 0.1–1.0% range. Formic Acid 3.75 Volatile Yes Maximum buffering obtained when used with ammonium formate salt. Used in 0.1–1.0% range. Acetate (NH4 CH2COOH) 4.76 3.76–5.76 Volatile Yes Used in the 1–10 mM range. Note that sodium or potassium salts are not volatile. Formate (NH4 COOH) 3.75 2.75–4.75 Volatile Yes Used in the 1–10 mM range. Note that sodium or potassium salts are not volatile. Phosphate 1 2.15 1.15–3.15 Non-volatile No Traditional low pH buffer, good UV transparency. Phosphate 2 7.20 6.20–8.20 Non-volatile No Above pH 7, reduce temperature/concentration and use a guard column to maximize lifetime. Additive/Buffer pK a TFA Buffer range Comments CORT ECS 2.7 µm Columns 5 b. pH Range Column lifetime will vary depending on the combination of temperature, mobile phase pH, and type of buffer or additive used. Table 3 lists the recommended buffers and additives for CORTECS Columns. Note. Working in combinations of extreme pH, temperature, and pressure may result in reduced column lifetime. solvent to remove the non-polar contaminant(s), taking care not to precipitate any buffered mobile phase components. If this flushing procedure does not solve the problem, purge the column with the following cleaning and regeneration procedures. 1. Use a cleaning routine that matches the properties of the samples and stationary phase type (reversed-phase, normalphase, or HILIC) to help solubilize the suspected contaminate. 2. Flush with 20 column volumes of solvent at 45 °C. c. Solvents To maintain maximum column performance, use high quality HPLC or MS-grade solvents. Filter all aqueous buffers prior to use through a 0.2 µm filter. Solvents containing suspended particulate materials will generally clog the outside surface of the inlet of the column. This may result in higher backpressure or distorted peak shape. d. Pressure CORTECS 2.7 µm Columns are compatible with HPLC, UHPLC, and UPLC pressures. Table 4 provides the maximum operation pressure for each column dimension. Table 4. Maximum Operation Pressure Column I.D. Maximum operating pressure 2.1 mm 15,000 PSI [1034 bar] 3.0 mm 15,000 PSI [1034 bar] 4.6 mm 9,000 PSI [620 bar] e. Temperature CORTECS 2.7 µm Columns can be used at temperatures up to 45 °C to enhance selectivity, reduce solvent viscosity, and increase mass transfer rates. IV. COLUMN C L EANING, R EG EN E RAT ION, AND STO RAG E a. Cleaning and Regeneration Changes in peak shape, peak splitting, shouldering peaks, shifts in retention, change in resolution, or increasing backpressure may indicate contamination of the column. Flush with a neat organic 3. Return to the initial mobile phase conditions by reversing the sequence. 4. If using a reversed-phase column, purge the column with a sequence of progressively more non-polar solvents (i.e., waterto-methanol-to-tetrahydrofuran-to-methylene chloride). 5. If using a HILIC column, purge the column with a sequence of progressively more polar-organic solvents (i.e., acetonitrileto-acetonitrile/methanol-to-acetonitrile/water-to-water). 6. If column performance has not improved after regeneration/ cleaning procedures, contact your local Waters representative for additional support. b. Storage after Reversed-Phase and HILIC Use For periods longer than four days, store the column in 100% acetonitrile. For separations utilizing elevated temperature, store immediately after use in 100% acetonitrile. Do not store columns in buffered eluents. If the mobile phase contained a buffer salt, flush the column with 10 column volumes of HPLC-grade water (see Table 1 for column volume information) followed by 10 column volumes of acetonitrile. Failure to perform this intermediate step could result in precipitation of the buffer salt in the column when 100% acetonitrile is introduced. Completely seal the column to avoid solvent evaporation and drying out of the chromatographic bed. Note. If a column has been run with a formate-containing mobile phase (e.g., ammonium formate, formic acid, etc.) and is purged with 100% acetonitrile, slightly longer equilibration times may be necessary when the column is re-installed and re-wetted with that same formate-containing mobile phase. CORT ECS 2.7 µm Columns 6 V. eCO RD INT E L LIG ENT C HI P T EC HNOLOGY c. Column Use Information a. Introduction The eCord intelligent chip provides the user with specific column information as well as column use data including; chemistry type, column dimension, serial number, and part number. The overall column use information includes; total number of samples injected, total number of injections as well as the maximum pressure, and temperature that the column has been exposed to. Additionally, detailed column history includes the sample set start date, user name, and system name. The eCord intelligent chip provides a paperless tracking history of the column’s performance and usage throughout its lifetime. The eCord is permanently attached to the column body via a tether that cannot be removed. This ensures that the history of the column is always accessible to the user of that column. The eCord intelligent chip is available on CORTECS Columns with 2.1 mm and 3.0 mm inner dimensions. eCord– intelligent chip V I. A DDIT IONA L INFO RMAT ION a. Tips for Maximizing Column Lifetime Figure 5. eCord intelligent chip identification. At the time of manufacture, the Performance Test Chromatogram, analytical data for the particles, and Certificate of Batch Analysis is downloaded onto the eCord. This information may then be accessed via the ACQUITY UPLC console once the column is installed. b. Installation of the eCord The eCord device can be read by connecting the yellow fob to the heater compartment receptacle located on the right hand side of the ACQUITY UPLC Column heater module. Once the eCord is connected, column identification, and overall column usage information can be accessed. eCord fob Figure 6. Installing the eCord intelligent chip. 1. To maximize column lifetime, pay close attention to: ■■ Water quality (including water purification systems) ■■ Solvent quality ■■ Mobile phase preparation, storage, and age ■■ Sample, buffer, and mobile phase solubility ■■ Sample quality and preparation 2. When problems arise, systematically troubleshoot potential cause one variable at a time in a systematic fashion. 3. Always remember to: ■■ Use an ACQUITY UPLC Column In-Line Filter Kit (p/n 205000343) or a suitable Vanguard Cartridge Column. ■■ Discourage bacterial growth by minimizing the use of 100% aqueous mobile phases where possible. ■■ Reduce the chances of mobile phase contamination or degradation, prepare enough mobile phase to last for 3–4 days. Alternatively, store excess bulk quantities in a refrigerated environment. ■■ Discard and re-prepare aqueous mobile phase every 24–48 hours (if 100% aqueous mobile phase is required). ■■ Add 5–10% organic modifier to aqueous buffer to minimize bacterial growth (adjust gradient profile as necessary). ■■ Filter aqueous portions of mobile phase through a 0.2 µm filter. ■■ Routinely maintain your water purification system to ensure it is functioning properly. ■■ Only use ultra-pure water (18 Ω-cm) and highest quality solvent possible. ■■ Consider sample preparation (e.g., SPE, filtration, centrifugation, etc.) when possible. CORT ECS 2.7 µm Columns 7 4. Avoid when possible: ■■ 100% aqueous mobile phases. ■■ HPLC-grade bottled water. ■■ “Topping off” your mobile phases. ■■ Using phosphate salt buffer in combination with high acetonitrile concentrations (e.g., >70%) due to precipitation. b. Getting Started with CORTECS HILIC Columns Note. CORTECS HILIC Columns are designed to retain very polar bases. Acidic, neutral, and non-polar compounds will have limited retention. Mobile Phase Considerations: 1. Always maintain at least 3% polar solvent, such as water, in the mobile phase, or gradient. T his ensures that the CORTECS HILIC particle is always hydrated. 2. Maintain at least 40% organic solvent (e.g., acetonitrile) in your mobile phase or gradient. 3. Avoid phosphate salt buffers to avoid precipitation in high organic content mobile phases used for HILIC separations. Phosphoric acid is acceptable. 4. Buffered mobile phases such as ammonium formate or ammonium acetate will produce more reproducible results compared to unbuffered additives such as formic acid or acetic acid. For best peak shape, maintain a buffer concentration of 10 mM. 5. If using an ACQUITY UPLC System, the weak needle wash solvent should closely match the percent organic solvent present in the initial mobile phase conditions, otherwise, analyte peak shape distortion can occur. Injection Solvent Considerations: 1. Whenever possible, injection solvents should contain 95% acetonitrile with the polar solvent (i.e., water, methanol, isopropanol) component to be no more than 25% of the total volume. A generic injection solvent is 75:25 acetonitrile/ methanol. T his is a good compromise between analyte solubility and peak shape. 2. Avoid water and dimethyl sulfoxide (DMSO) in the injection solvents. These solvents will produce very poor peak shapes. In HILIC, it is important to remember that water is the strongest eluting solvent. Therefore, it must be eliminated or minimized in the injection solvent. Exchange water or DMSO with acetonitrile by using reversed-phase SPE. If this is not possible, dilute the water or DMSO with organic solvent. c. Troubleshooting Questions 1. Are you using 100% aqueous mobile phases? 2. W hat is the age of the mobile phase? 3. Is the mobile phase filtered through a 0.2 µm membrane? 4. Was the mobile phase prepared fresh or topped off? 5. Is the water source of adequate quality? 6. W hen was the last time the water system was serviced or was the bottle of water unopened? 7. Is bacterial growth a possibility (pH 7 phosphate buffer is susceptible to bacterial growth within 24 hours)? 8. If a neat standard is prepared in the initial mobile phase conditions and injected, are the problems still observed? 9. If the sample is filtered/purified (i.e., SPE, filtration… etc.) is the problem still observed? 10. Has the quality of the samples changed over time? V II. C AUT IONA RY NOT E Some products may be hazardous during and after use and are to be used by professional laboratory personnel trained in the competent handling of such materials. The responsibility for the safe use of products rests entirely with the purchaser and user. The safety data sheets (SDS) for these products are available at www.waters.com /SDS. Waters, T he Science of W hat’s Possible, CORTECS, UPLC, ACQUITY UPLC, Alliance, Oasis, and Sep-Pak are registered trademarks of Waters Corporation. eCord and VanGuard are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. ©2016 Waters Corporation. Produced in the U.S.A. September 2016 720005024EN IH-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com