Band Gap of Germanium∗ Joel Ong† (Dated: April 4, 2014) I. INTRODUCTION Germanium is an intrinsic semiconductor. That is to say, its Fermi energy µ lies between its valence and conduction bands, and the band gap between its valence and conduction bands is small enough that conduction is possible at sufficiently high temperatures. In the absence of dopants and other impurities, its Fermi energy lies exactly in between the valence and conduction bands; consequently, the number density n of electrons and holes should be equal. Electrons and holes are described by Fermi-Dirac statistics. The law of mass action then demands that at sufficiently high temperatures, ] [ √ ∆E , (1) ne nh = nc ∝ exp − 2kT where nc is the total carrier concentration, k is Boltzmann’s constant and ∆E is the band gap between the conduction and valence bands. The conductivity of a material can be roughly described using the Drude model of electrical conductivity; in sum, the current density passing through a material is proportional to the carrier concentration and applied electric field. Therefore, we must have that the conductivity σ is proportional [ ∆E ] to the carrier concentration, and so σ ∝ exp − 2kT . A. Objectives In this experiment, we examine the temperature dependence of the conductivity of a germanium sample, and use it to estimate the band gap of germanium. II. EXPERIMENTAL PROCEDURE We were provided with a sample of germanium of dimensions 20 × 10 × 1 mm3 , which was attached to a heating unit (Figure 1). The circuit onto which these were mounted was capable of passing a constant current through the germanium sample while measuring both the temperature T and the potential difference V across the sample simultaneously. By maintaining a constant current I through the germanium sample, our measurements of the potential V across the sample should be inversely proportional to the conductivity σ. We recorded various values of V for constant I, at different values of T . This was achieved by letting the heating coil heat up the germanium sample while the computer was left running ∗ † Lab Summary for 8 Credit Hours Matric Number: A0098750U; Lab Partner: Kho Zhe Wei to collect data. We collected data in this manner over the range of temperatures 25◦ C to 131◦ C. FIG. 1: Experimental equipment: germanium sample, heating unit and sensors on a single chip III. RESULTS AND ANALYSIS We carried out the above procedure for I = 5.0(5) mA. Given the above relationship between the conductivity, band gap and temperature, we ] [ ∆Eenergy . Therefore, we should in then have V ∝ exp 2kT principle be able to measure the band gap by measuring V at different temperatures while keeping the current I constant, and fitting the resulting curve against the two-parameter model [ ] B V = A exp . (2) T Such a model should yield ∆E = 2kB. 4 V/V 3 2 1 T/K 320 340 360 380 400 420 FIG. 2: Fitted curve with experimental data for I = 5.0(5) mA (with error bars for instrument uncertainty) We performed this curve-fitting procedure on the data collected for this value of I (Figure 2 shows a plot of the fitted curve). The fitting procedure returned 2 parameter estimates of A = 4.5(9) × 10−6 V and B = 4120(6) K−1 to 95% confidence, with a correlation coefficient of r2 = 0.9992. This gave us an estimate of the band gap energy ∆E = 0.71(1) eV (to within a 95% confidence interval). However, we observe that the computer obtained samples of V against T at irregular intervals of temperature. Upon closer inspection, we found that values of V and T were recorded at regular intervals of time, at ∆t = 1 s. Therefore, we can improve the accuracy of our regression estimate by taking samples while the germanium sample was cooling, so as to obtain more data points at smaller temperature intervals. We recorded the same data with the germanium sample cooling down instead of heating up, and performed the same curve-fitting procedure. The curve fitting procedure returned parameter estimates of A = 2.25(4) × 10−6 V and B = 4321(6) K−1 to 95% confidence, with a correlation coefficient of r2 = 0.99997, which is much higher than that obtained from taking readings while heating the germanium sample. Using these parameter estimates we obtain an estimate for the band gap energy as ∆E = 0.745(1) eV, to within a 95% confidence interval. We see that the 95% confidence interval returned from performing the fitting procedure on the cooling curve is an order of magnitude smaller than that returned by the heating curve; moreover, we see that the plot of our experimental data is qualitatively smoother than the heating curve (Figure 3). V/V TABLE I: Parameter estimates with 95% confidence intervals for curves obtained from different values of I, using data collected while the sample was either being heated up or cooled down I/mA Heating 3.0(5) 5.0(5) 9.0(5) Cooling 3.0(5) 5.0(5) 9.0(5) B/K−1 ∆E/eV A/V 0.0000025(5) 4150(6) 0.72(1) 0.0000045(9) 4120(6) 0.71(1) 0.000008(1) 4150(6) 0.71(1) r2 0.999172 0.999199 0.999231 0.00000126(3) 4354(7) 0.750(1) 0.999966 0.00000225(4) 4321(6) 0.745(1) 0.999974 0.00000400(7) 4329(5) 0.7458(9) 0.999978 the sample was cooling (on average, 0.747(1) eV), as compared to those estimates calculated from data collected while the sample was being heated up (on average, 0.71(1) eV). IV. A. DISCUSSION Negative I In principle, we should have obtained the same results as above with the current flowing in the opposite direction (i.e. for negative I). However, when we performed the experiment in such a configuration, we observe a discontinuity at T = 310 K (Figure 4). V/V 4 0.0 - 0.5 3 - 1.0 2 - 1.5 - 2.0 1 T/K 320 340 360 380 400 FIG. 3: Fitted curve for cooling germanium sample with experimental data for I = 5.0(5) mA (with error bars for instrument uncertainty) However, we see that this parameter estimate differs significantly from the one we obtained when heating the germanium sample. To verify that this is a real observation and not a consequence of an error in our analysis, we decided to repeat the procedure outlined above for I = 3 mA and I = 9 mA; for each value of I we collected data both while the germanium sample was heated up, and while it was cooling down. We then performed the same fitting procedure as outlined above on all the data sets so obtained. We show our results below in Table I. We see that the estimates for the band gap energy ∆E are persistently higher for all data collected while - 2.5 T/K 320 340 360 380 400 420 440 FIG. 4: Data collected while the germanium sample was cooling, with fitted curve, for I = −5.0(5) mA (with error bars for instrument uncertainty) This discontinuity appears irrespective of whether the data was collected while the germanium sample was being heated up or cooled down. However, truncating data collected for temperatures below this discontinuity and then performing the same regression procedure as above returns the estimates for the band gap, for both the heating and cooling curves, that are consistent with the ones in Table I; this indicates that data collected at temperatures below this discontinuity are inaccurate when I is negative. This fact leads us to hypothesise that this observed discontinuity is an artifact of the instrumentation used to measure the voltage and current across the ger- 3 manium sample. The fact that the change is induced abruptly at a particular temperature, as well as the predictable behaviour of the discontinuity (nearly doubling the measured value of V ) indicates that a likely cause is that the combination of negative current and temperature change induces a change in the gain of an amplifier circuit used as part of the ammeter. However, given that the internal workings of the equipment provided was essentially a closed black box, there was little that we could do to verify this conjecture or investigate this phenomenon. B. Discrepancy with Literature Value All of our estimates differ from the literature value of 0.67 eV [1], with percentage discrepancies on the order of 6% for estimates obtained from the heating curves, and 12% for estimates collected from the cooling curves. However, we find also that the band gap energy varies with temperature, even in the literature, instead of being a constant as we have earlier supposed. In particular, at 0 K, the band gap of germanium is 0.74 eV, which is quite close to our measured values. Phenomenologically, this temperature dependence can be modelled as ∆E(T ) = ∆E(0) − αT 2 . T −β (3) Moreover, the literature values for these fitting parameters α and β are small enough that within the range of temperatures that we have examined, the variation of the band gap is approximately linear with temperature [2]. Consequently, the fitting procedure outlined above returns us not the actual band gap energy, but rather the band gap energy evaluated at 0[ K, since the additional linear term falls out as ] αT ≈ constant. Compared to that value, our exp 2kT parameter estimates exhibit far smaller discrepancies than when compared to the literature value for the room-temperature band gap. Aside from this, there must of course exist impurities within the germanium sample with which we were provided, whose electrons must also contribute to the conductance of the material. We therefore expect some degree of deviation from the literature value, alternative model notwithstanding. However, we were not able to accurately characterise the purity of the provided germanium sample; consequently, we were unable to draw any quantitative conclusions as to the [1] C. Kittel and P. McEuen, Introduction to solid state physics, Vol. 8 (Wiley New York, 1986). [2] P. Lautenschlager, P. Allen, and M. Cardona, Physical Review B 31, 2163 (1985). significance of any contribution from impurity carriers. C. Discrepancy between heating and cooling curves We see that we have larger estimates for the band gap energy when using data collected while the sample was being cooled down compared to data collected when it was being heated up. We can attribute this to the fact that all of the physical description outlined above (including Equation 1) applies only to systems in thermal equilibrium. Manifestly, when the germanium sample is being heated up rapidly by an external heating unit, it is not in thermal equilibrium with itself, since the side facing the heating unit is hotter than the side facing away (given that the heating unit was at a much higher temperature than the germanium sample, and we turned the heating unit off before equilibrium with it could be achieved in order to prevent heat damage to the sample). Conversely, for the data collected when the germanium sample was cooling down, the entire process of data collection took much, much longer, on the order of 12 minutes. During this time, the germanium sample was not in thermal equilibrium with the ambient air (as it was being cooled down). However, the relaxation time for the sample to come into equilibrium with itself is much smaller than the time taken for it to equilibrate to room temperature. Therefore, on sufficiently coarse time scales, we would expect that the cooling process is quasistatic, in that at each point in time the germanium sample is approximately in equilibrium with itself. Consequently, the analysis above should in principle hold more accurately for when the germanium sampled is being slowly cooled down that when it is being heated up; this should mostly account for the discrepancy in the results returned by the two different experimental methodologies. Therefore, we should privilege the values returned by cooling the germanium sample as being the more accurate of the two methodologies; this yields an average estimate for the band gap energy of 0.747(1) eV. V. CONCLUSION Using the temperature dependence of the conductance of a germanium sample with the current flowing through it held constant, we have determined that the band gap of germanium has a value of 0.747(1) eV.