calibration jig for current probes

advertisement
PCJ 9201B
CALIBRATION JIG FOR CURRENT PROBES
The standards RTCA/DO-160 section 20 (Environmental Conditions and Test Procedures for Airborne
Equipment Section 20: Radio Frequency Susceptibility (Radiated and Conducted)), MIL-STD-461 (REQUIREMENTS FOR THE CONTROL OF ELECTROMAGNETIC INTERFERENCE CHARACTERISTICS OF SUBSYSTEMS
AND EQUIPMENT) and IEC/EN 61000-4-6 (Testing and measurement techniques –Immunity to conducted
disturbances, induced by radio-frequency fields) define a method for test level setting in a calibration
jig. The PCJ 9201B meets all the requirements and provides additionally a save way for measuring the
insertion loss of current sensing probes and monitoring probes. The enhanced VSWR offers measurements in range up to 400 MHz.
Technical specifications
PCJ 9201B calibration jig
Calibration jig for current probes
Meets ISO 11452-4, RTCA/DO-160
section 20, MIL-STD-461 and IEC/EN
61000-4-6
Ruggedly designed
Enhanced VSWR up to 400 MHz
Frequency range:
DC up to 400 MHz
Nominal impedance:
50 Ω
VSWR (without the current probe installed in the jig, typical values in the graph)
10 kHz
≤1.10 : 1
1 MHz
≤1.10 : 1
10 MHz
≤1.14 : 1
30 MHz
≤1.22 : 1
100 MHz
≤1.50 : 1
200 MHz
≤2.00 : 1
300 MHz
≤2.75 : 1
400 MHz
≤3.50 : 1
Dimension (LxWxH):
260 mm x 181 mm x 150 mm
Probe diameter max.:
115 mm
Probe window diameter min.:
16 mm
Probe wide max.:
67 mm
Connector:
N-type female
Weight:
approx. 4.5 kg
Operating temperature:
+5°C to +45°C
Detachable cover plate
52 mm
139 mm
N type
connector
67 mm
PCJ 9201B
CALIBRATION JIG FOR CURRENT PROBES
VSWR (without the current probe installed in the jig)
Maximum VSWR for RTCA/
typical value for PCJ 9201B
DO-160 section 20 and MIL-STD-461E,
4.0
3.5
3.0
2.5
VSWR (X:1)
2.0
1.5
0.09
PCJ 9201B calibration jig, side view
50
100
150
200
250
300
350
400
Frequency [MHz]
Test level setting in a 50 Ω system (e.g. BCI automotive, MIL-STD-461 and RTCA/DO-160
section 20)
Test generator e.g. Teseq NSG 4070
with built-in RF synthesizer and power meter
RF output
NSG 4070
RF out
Local
RF
7
8
9
MHz
dBµV
4
5
6
kHz
dBm
STO
RCL
1
2
3
Hz
V
StSize
Step
StSize
StSize
Step
.
Step
2
0
FRQ
LVL
MOD
ON/OFF
Help
Power meter
Amp in
ch.3 < +20 dBm
ch.2 < +20 dBm
Power meter with
input impedance 50 Ω
Power amplifier
input
1
2nd
3
Enter
Tuning
< +10 dBm
Hold
Amp out
ch.1 < +27 dBm
Back
Run
Stop
USB
Amp in
Power
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp out
Output
50 Ω Termination
DIRECTIONAL COUPLER
DCP 0100
0.01 ... 1000 MHz
Calibration jig PCJ 9201B
BCI probe
Ground plane
Directional coupler
PCJ 9201B
CALIBRATION JIG FOR CURRENT PROBES
Test level setting in a 150 Ω system (e.g. IEC/EN 61000-4-6 current clamp injection)
NSG 4070
Power amplifier
output
RF out
Local
RF
7
8
9
MHz
dBµV
4
5
6
kHz
dBm
STO
RCL
1
2
3
Hz
V
StSize
Step
StSize
StSize
Step
.
Step
2
0
FRQ
LVL
MOD
ON/OFF
Help
Power meter
Amp in
ch.3 < +20 dBm
1
2nd
3
Enter
Tuning
ch.2 < +20 dBm
< +10 dBm
Hold
Amp out
ch.1 < +27 dBm
150 Ω Termination
INA 721-150
Run
Stop
Back
USB
Power
Test generator e.g.
Teseq NSG 4070
with built-in
RF synthesizer,
power amplifier and
power meter
Power meter with 50 Ω input
INA 721-100, 100 Ω passageway
6 dB
Attenuator
150 Ω/ 50 Ω Adapter
INA 721-100
Calibration jig
PCJ 9201B
INA 721-150, 150 Ω termination
Current injection probe
CIP 9136
Ground plane
Insertion loss measurement
NSG 4070
RF out
7
8
9
MHz
dBµV
FRQ
LVL
4
5
6
kHz
dBm
STO
RCL
3
Hz
V
StSize
Step
Local
RF
MOD
ON/OFF
Help
Power meter
Amp in
ch.3 < +20 dBm
1
2
0
.
ch.2 < +20 dBm
StSize
StSize
Step
2
Step
1
2nd
3
Enter
Tuning
< +10 dBm
Hold
Amp out
ch.1 < +27 dBm
Back
Run
Stop
USB
Power
Power meter with 50 Ω input
Test generator e.g.
Teseq NSG 4070
with built-in
RF synthesizer,
power amplifier and
power meter
or
network analyzer
Termination
50 Ω
Calibration jig
PCJ 9201B
Monitoring
probe
Ground
plane
Delivery information
Part number
Description
252052
PCJ 9201B
Calibration jig for current probes, meets ISO 11452-4 ,RTCA/DO-160
section 20, MIL-STD-461 and IEC/EN 61000-4-6
INA 721-100
Adapter for IEC/EN 61000-4-6, 100 Ω passageway,
2x N connector
INA 721-150
150 Ω termination for IEC/EN 61000-4-6, N connector
403-403
403-404
Teseq GmbH
Landsberger Str. 255 · 12623 Berlin · Germany
T + 49 30 56 59 88 35 F + 49 30 56 59 88 34
desales@teseq.com www.teseq.com
82-252052 E01 Feb. 2008
Download