Int. J. Computational Biology and Drug Design, Vol. 3, No. 3, 2010 Fast algorithm in estimating high intensity focused ultrasound induced lesions Yufeng Zhou Division of Engineering Mechanics, School of Mechanical and Aerospace Engineering, Nanyang Technological University, 639798, Singapore and Laboratory of Modern Acoustics of Ministry of Education, Nanjing University, Nanjing 210093, China Fax: 65-6792-4062 E-mail: yfzhou@ntu.edu.sg Abstract: High Intensity Focused Ultrasound (HIFU) has been emerging as a new and effective modality in non-invasive thermal ablation for cancers and solid tumours. Although a theoretical model is available for calculating thermal field and the consequent lesion production, its computation is too time-consuming (~1 day) for HIFU treatment planning on the site. In this study, several approximations were made on the BioHeat equation to estimate HIFU-induced lesions. Experimental results, estimation and theoretical calculation were compared with each other in a good agreement. However, the computation time for 25 treatment spots was only 1 min. Altogether, the developed fast algorithm provides an accurate outcome in a timely fashion and can push the wide acceptance of HIFU technology in clinics. Keywords: HIFU; high intensity focused ultrasound; BioHeat equation; temperature elevation; approximation. Reference to this paper should be made as follows: Zhou, Y. (2010) ‘Fast algorithm in estimating high intensity focused ultrasound induced lesions’, Int. J. Computational Biology and Drug Design, Vol. 3, No. 3, pp.215–225. Biographical notes: Yufeng Zhou is an Assistant Professor at Nanyang Technological University in Singapore and an Adjunct Professor at the Laboratory of Modern Acoustics at Nanjing University in China. He received his PhD in Bioacoustics from Duke University (2003). His research interests include bubble dynamics, therapeutic ultrasound technology (shock wave lithotripsy for stone treatment, high-intensity focused ultrasound for solid tumour/cancer ablation, ultrasound-enhanced drug delivery, shock wave therapy), non-destructive evaluation and acoustic wave propagation. Copyright © 2010 Inderscience Enterprises Ltd. 215 216 1 Y. Zhou Introduction High Intensity Focused Ultrasound (HIFU) is emerging as a new modality for solid cancer/tumour treatment, such as prostate, kidney, liver, breast and pancreatic cancers (Kenney, 2005). In Asia and Europe, more than 100,000 cases have already been performed with great success. The principle of HIFU therapy is focusing a high-intensity ultrasound beam into a tumour within the body for ablation. The acoustic intensity at the focus is high enough to achieve temperatures reaching over 65°C within a few seconds to coagulate the tissue. In comparison with traditional cancer therapies, such as surgery, radiotherapy and chemotherapy, HIFU ablation has the advantage of being a completely non-invasive therapy without exposing the patient to ionising radiation, resulting in highly targeted therapy with fewer complications (Zhou, 2011). Despite promising results, HIFU remains in its infancy stage with several technical problems preventing it from becoming a widely adopted therapeutic procedure (Zhou, 2011). For example, current HIFU systems, both clinical and experimental models, have very limited capabilities of treatment planning. Under the guidance of either sonography or Magnetic Resonance Imaging (MRI), the geometrical information (i.e., contour and anatomical position) of to-be-treated cancers/solid tumours and vital tissues (i.e., vessels and nerves) possible in the treatment region can be figured out, which is critical in non-invasive therapy. Afterwards, most systems provide physician or operator the HIFU treatment information (i.e., locations of the treatment spots and the scanning pathway), but no information of the consequent ablation lesions is available. Therefore, it becomes the physician’s responsibility to input the treatment parameters (i.e., acoustic power, pulse parameters and interval spacing between the treatment spots), which determine the lesion shape and size. Unfortunately, most physicians do not have a sufficient background and understanding of physics underlying HIFU to set up those parameters rationally. To improve the safety and efficacy of HIFU therapy, a more effective method for treatment planning, which assists the user to determine the input parameters and estimate the treatment outcome prior to beginning the therapy, is essential and critical for the wide acceptance of HIFU technology. The ablation mechanism of HIFU is the absorption and transfer of acoustic energy into heat when an ultrasound wave propagates in tissue and the subsequent high-temperature rise as high as 65°C within a few seconds at the focus of the therapy transducer. At this temperature, the proteins are denatured and internal structure located in the focal area can be precisely destroyed without any incision or injury to the intervening tissue, as the suddenness and rapidity of this phenomenon prevent the dissemination of heat around the focal point. It is, therefore, desirable to have complete knowledge of the temperature distribution. In the fundamental studies of HIFU, theoretical model has already been established to simulate the acoustic fields in the focal region of any HIFU transducer. Briefly, the Khokhlov-Zabolotskaya-Kuznetsov (KZK) non-linear evolution equation is first used to model numerically high-intensity acoustic beams (Bakhvalov et al., 1987). Then, with knowledge of the acoustic fields, the Pennes BioHeat equation is used to predict the temperature response and Thermal Dose (TD) resulting from HIFU exposure in the tissue fairly accurately (Pennes, 1948), although some limitations are included. The BioHeat equation can be solved with initial and boundary conditions by numerical methods. These methods include the Finite Element Method (FEM) (Das et al., 1999), the Finite Difference Time Domain (FDTD) method (Curra et al., 2000) and solution by a semi-discrete scheme of Galerkian FEM Fast algorithm in estimating high intensity 217 (Sohrab et al., 2006). These simulated results were found to be in a good agreement with ex vivo and in vivo experimental observations (Chapelon et al., 1993). However, all of these theoretical models are computationally intensive (several CPU hours). Therefore, simulating the HIFU outcome just prior to the clinical treatment is not feasible using current theoretical computation approaches. In this study, the solution to the BioHeat equation was approximated to reduce the computation time. A total of 25 treatment spots with different scanning pathways (raster, spiral scanning from the outside to inside and spiral scanning from the inside to outside) were evaluated. Calculation time of the fast algorithm was only about 1 min. Approximation results were then compared with the produced lesion in the transparent polyacrylamide tissue phantom embedded with Bovine Serum Albumin (BSA). A quite good agreement was found between our estimation, theoretical calculation and experimental outcome. Altogether, our approximation algorithm provides a fast HIFU planning and evaluation option for physician and operators before clinical treatment. 2 Methods 2.1 BioHeat transfer equation The mathematical model for temperature elevation in the tissue is based on the BioHeat Transfer Equation (BHTE) (Pennes, 1948), G T − T0 Q(r ) F (t ) ∂T , = k ∆T − + (1) ∂t Cv τ where t is the time, τ is the perfusion time, T(r, t) is the tissue temperature, T0 is the equilibrium temperature, k = K/Cv is the local tissue temperature conductivity, K is the heat conductivity, Cv is the heat capacity of a unit volume, ¨ is the Laplacian operator, G F(t) is the temporal function indicating an HIFU pulse, and Q(r ) is the absorbed ultrasound energy G G Q(r ) = 2α I (r ), (2) G where α is the ultrasonic absorption coefficient of the soft tissue and I (r ) is the acoustic intensity. A solution to equation (1) for an infinite, homogeneous and isotropic medium can be found using transform techniques (Davies, 1978; Carnes et al., 1991) as 1 G T (r , t ) = Cv t ³³ 0 V G G G G Q(r ′) F (ξ )G (r − r ′, t − ξ )dr ′ d ξ , G G 2 ª r − r′ º exp « − t − ξ − » 4k (t − ξ ) ¼» G G ¬« τ , G (r − r ′, t − ξ ) = [4π k (t − ξ )]3/ 2 (3) (4) G where G(*) is the appropriate Green’s function (Stakgold, 1979), r ′ is the source variable coordinate and ξ is the temporal variable of the heat source. The corresponding TD to HIFU pulses was calculated using 218 Y. Zhou t TD 43°C (t ) = ³ R 43−T (t ') dt ′ 0 (5) with R = 0.25 if T(t) < 43°C and 0.5 otherwise (Sapareto and Dewey, 1984). The value required to create a thermally irreversible damage in tissues is equivalent to the TD of a 240-minute exposure at 43°C (Lele, 1967; Damaniou and Hynynen, 1994). 2.2 Approximation The lateral distributions of the acoustic intensity are assumed in Gaussian function − G I (r ) = I 0 e G G ( r − r0 ) 2 2σ 2 , (6) where I0 is the spatial-peak pulse-average acoustic intensity at the focus and calculated from the pressure waveforms measured using a Fibre Optic Probe Hydrophone (FOPH) (Zhou et al., 2006), r0 is the location of the treatment spot and σ is the variance and determined by σ= (HPBW/2) 2 ln 2 (7) where HPBW is 3 dB or half-power beam width. Although the pressure distribution could be described more closely using a transverse beam profile 2 J1 (ar )/ar , where J1(*) is the first-order Bessel function and a is the radius of HIFU transducer, the contribution of side-lobes in the thermal ablation is usually neglected and Gaussian function is much easier in calculation. Since the delivery of HIFU energy is in the format of pulse, G the equivalent absorbed ultrasound energy by the tissue, Qe (r ), can be approximated as G G Qe (r )U (t ) = Q(r ) ⋅ dc% ⋅ F (t ), (8) where dc is the duty cycle of HIFU pulses and U(t) is the unit step function. Afterwards, larger time step (0.5 s) can be used in the calculation of temperature field to save the computation time without a reduction in its accuracy. Two-dimensional thermal field was calculated first and then extended to 3D space under the assumption that the ratio of the length of the generated lesion to its diameter is same as the ratio of the measured −6 dB beam size in the axial direction to its lateral direction. 2.3 Experimental set-up The HIFU system used in this study (FEP-BY02, Yuande Bio-Engineering Ltd., Beijing, China) consists of 251 individual PZT elements, driven all in phase and arranged in a concave spherical holder. The HIFU transducer has a centre frequency of ~1 MHz, an outer diameter of 33.5 cm and an inner diameter of 12 cm with integrated ultrasound imaging probe (S3, Logiq 5, GE, Seongnam, Korea) mounted in the central hole co-axial to the HIFU beam. An optically transparent gel phantom (L × W × H = 5.5 × 5.5 × 5 cm3), composed of polyacrylamide hydrogel and BSA that becomes optically opaque when denatured by heat (Lafon et al., 2005), was surrounded by a tissue phantom that contains 6.5% Alginate (Jeltrate, Dentsply International, York, PA). The centre of the Fast algorithm in estimating high intensity 219 transparent gel phantom was aligned with the HIFU focus under the guidance of B-mode ultrasound imaging. A LabVIEW (National Instruments, Austin, TX) program was written and run on a PC to control the motion of the treatment table and delivery of HIFU pulses (see Figure 1). After the treatment, the HIFU phantom was taken out and the lesions were recorded photographically for comparison. Figure 1 Experiment set-up of generating thermal lesions inside tissue phantom using extracorporeal High-Intensity Focused Ultrasound system (see online version for colours) Three scanning pathways were used in this study: a conventional raster scan, spiral scanning from the centre to the outside and spiral scanning from the outside to the centre (see Figure 2) (Zhou et al., 2007). There were total 25 treatment spots arranged in the shape of a diamond with a grid size of 4 mm. The treatment parameters are same for each spot: the HIFU on time is 150 ms, the HIFU off time is 150 ms, there are 60 pulses per spot, interval time between the treatment spots is 6 s (including the mechanical movement time) and the absorbed acoustic energy at the target is 1000 J. The electrical output power of the HIFU transducer was calculated using our proposed method (Hwang et al., 2009). Figure 2 3 Motion pathway used in HIFU treatment in: (a) raster scanning; (b) spiral scanning from the centre to the outside and (c) spiral scanning from the outside to the centre for a total of 25 spots with 4 mm grid space (see online version for colours) Results Approximation of the effective heat source as in equation (8) was validated and shown in Figure 3. HIFU exposure consists of 60 pulses with pulse duration of 300 ms and duty 220 Y. Zhou cycle of 50%. In each pulse, the temperature will increase when HIFU energy is delivered and decrease at the HIFU-off stage, respectively, seeming as saw-teeth. Using the effective heat source approximation and much longer temporal step size (500 ms), the temperature profile fits well with the exact solution although without saw-teeth. Because of the low-temperature elevation induced by each HIFU pulse (low amplitude of each saw-tooth), its contribution to the overall TD can be ignored. Figure 3 Approximation of the effective heat source in the temperature elevation calculation. Blue line: HIFU pulses (150 ms on, 150 ms off, 60 pulses) with calculation time step of 30 ms; red line: continuous HIFU exposure with effective heat source approximation and calculation time step of 500 ms (see online version for colours) The lesions generated in the gel phantom were found to have similar pattern and characteristics as predicted by the theoretical simulations (see Figure 4) as well as using the fast algorithm proposed in this study. By using the raster scanning method, the first visual lesion was the third treatment spot and with the progress of the HIFU treatment the lesion size became larger because of the thermal diffusion from nearby spots. Merging of lesions occurred towards the end of the treatment. When using the spiral scanning outwards pathway, only the first lesion was not visible. However, the other lesions were smaller in comparison with those generated in the raster scanning pathway, and no merging of lesions was observed. The lesion pattern using the spiral scanning inwards had different characteristics with all treatment spots along the outside boundary (first 12 spots) not being visible. Since thermal energy is concentrated towards the centre of the treatment area, the last a few lesions merged to form a large lesion. The patterns of lesion production and their characteristics can also be assessed by observing the lesions from the lateral direction (see Figures 5 and 6). The great differences lie in the computation time. It took more than 24 h in computation using the accurate FTFD code running in a workstation (Hewlett-Packard, Palo Alto, CA, USA) (Zhou et al., 2007). However, the computation time reduced to about 1 min with the fast algorithm in a PC with 2.0 GHZ CPU and 1GB memory run in Matlab (Mathworks, Natick, MA, USA). The maximum lesion lengths in the gel phantom using these scanning pathways were 10.1, 6.9 and 12.2 mm, respectively. In comparison, the corresponding values using the fast algorithm were 10.1, 5.2 and 9.5 mm, respectively, whose errors are within an acceptable range. Fast algorithm in estimating high intensity 221 Figure 4 Thermal field calculation results using finite-time finite-difference method in (left) raster, (middle) spiral scanning outwards and (right) spiral scanning inwards Figure 5 Representative photos of lesions generated inside the gel phantom from: (a) top view, lateral view in (b) X direction and (c) Y direction using pathway in (left) raster, (middle) spiral scanning outwards and (right) spiral scanning inwards Furthermore, Matlab environment provides several useful features for end-users. For example, the predicted lesions can be viewed in 3D space from an arbitrary angle (see Figure 7(a)), which provides direct impression on the clinical HIFU outcome if 222 Y. Zhou combined with the reconstructed anatomies from sonographs and MRI sections. The temperature profile at any position is also available (see Figure 7(b)) to examine the potential for the occurrence of boiling phenomenon (T > 100°C). The boiling bubbles are of orders larger than those HIFU pulses-induced ones, blocking the propagation of the following HIFU pulses, and subsequently distorting the lesions from a symmetric cigar shape to an asymmetric tadpole shape (Watkin et al., 1996), which makes the HIFU-induced lesions not expectable as in the stage of treatment planning and estimated by theoretical model. Figure 6 Simulated lesions generated with the fast algorithm from: (a) top view, lateral view in (b) X direction and (c) Y direction using pathway in (left) raster, (middle) spiral scanning outwards and (right) spiral scanning inwards (see online version for colours) Figure 7 Features of simulation package in Matlab environment: (a) 3D view of generated lesions in a raster scanning and (b) the temperature profile of the 23rd spot in spiral scanning inwards way (see online version for colours) (a) (b) Fast algorithm in estimating high intensity 4 223 Discussion and conclusion HIFU is a new and effective modality in solid tumour/cancer treatment. Because of its inherent advantages of non-invasiveness and non-ionisation, its application is increasing rapidly. However, the high dependency of the clinical outcomes on input parameters, which is difficult for most physicians because of their absence in understanding of HIFU physics, limits the adoption. Although the theoretical model is available in calculating thermal field and subsequent lesion size, it takes too much time (in the order of a day) in the computation to be applied on the site. A fast algorithm was developed in this study based on BioHeat equation and its solution using Green’s function. Comparison between experimental, theoretical and estimated results was carried out. It is shown that the fast algorithm produces a rather satisfactory outcome, but in a much shorter time (in the order of a minute). Several approximations were made in calculating the solution of BioHeat equation. Despite that, the systematical error is within a satisfactory range because the purpose of this algorithm is to provide an estimation of the sizes and shapes of HIFU-induced lesions for physician/operator in a timely fashion. One of the limitations of this approach is the need of calculating heat source from measured pressure waveform. Numerical method is available to solve the KZK equation for finite-amplitude acoustic wave propagation through multiple tissue layers (i.e., the abdominal layer of humans) both economically and computationally (Soneson, 2008). Coupling this module with tissue information obtained from either ultrasound or MRI images for the guidance of HIFU therapy, in vivo acoustic pressure, distribution, and the consequent heat source could be available with higher accuracy than those derrated using free field measurement and attenuation coefficients of the tissue in the propagation path that was usually summarised from ex vivo experiments. Because the detectable tumour size (centimetres) is much larger than the HIFU focal size (millimetres), the HIFU focus needs to be scanned throughout the entire target volume. The common scanning pathway is a raster mode and HIFU parameters for each spot are kept the same. Both gel phantom and theoretical studies have illustrated that the energy delivered at the beginning of the HIFU ablation is insufficient to generate visual lesions; however, with the HIFU ablation progress the lesion size will increase due to the thermal diffusion effect and treatment spots later in the treatment may be over-treated (see Figures 4 and 5). From the viewpoint of the physician, there are three basic requirements for the ideal HIFU ablation: • ability to generate a predictable lesion for every treatment spot • all generated lesions need to be uniform • complete coverage of the entire treated volume. Therefore, the delivered energy should be adjusted dynamically with the use of different scanning pathways (Zhou et al., 2008). The fast algorithm proposed in this study can be developed into an HIFU treatment planning software that runs prior to the therapy when the patient lying at the table in ready status. HIFU planning software may include the functionalities of having a user-friendly interface and an intuitive operation that requires minimal training, assisting the HIFU operator to evaluate the treatment protocols, determining the HIFU 224 Y. Zhou parameters without influencing the other steps (tumour diagnosis, HIFU ablation and post-treatment observation) and reviewing the estimated lesions in 3D space. As a result, it will make HIFU operation more easy and reduce the outcome variation due to operator dependency. The most advantage of this fast algorithm is its universe application for all commercial or experimental HIFU systems (different centre frequency, transducer geometries and treatment parameters) and tissue targets (different tumour types and anatomic structures on the HIFU propagation pathway). Furthermore, other relevant information (i.e., instantaneous temperature, pressure and acoustic intensity distribution) will also be available in considering the safety of HIFU therapy. There are at least 11 major HIFU manufacturers in the market (5 in China, 2 in USA, 1 in Korea and 3 in Europe) and also several smaller ones worldwide. Since the HIFU technology is still in its infancy stage and a large amount of clinical trials have been performed with promising results, more manufacturers may penetrate this cutting-edge medical device market. Although the healthcare fee related with HIFU treatment is unknown, it is believed that with more acceptance of this novel technology this number will increase significantly in the near future. According to our knowledge, there is no capability of the HIFU treatment planning in currently available HIFU systems, neither clinical nor experimental type. Therefore, our proposed fast algorithm targets this niche market and may be able to push the wide acceptance of HIFU technology, to enhance its efficiency and safety, and to facilitate the FDA clearance of HIFU device for solid tumour treatment. There are other FDA-approved ablation devices that utilise energy sources of Radiofrequency (RF), laser and microwave to create thermal injury to tissue in conjunction with a percutaneous approach. While thermal ablative techniques are rapidly gaining acceptance in the treatment of inoperable tumours, incomplete treatments are found common because of the thermal diffusion effect as in HIFU (Goldberg et al., 1998). Although the energy sources are different, the principle of thermal lesion production is same as HIFU. Therefore, the fast algorithm proposed here can also be applied to the other thermal ablation methods. In summary, the proposed fast algorithm is able to produce satisfactory accurate results as the theoretical model but with reasonable computation time, which is feasible for onsite evaluation of the treatment protocols and allows the development of HIFU planning software for commercial system. References Bakhvalov, N.S., Zhileikin, Y.M. and Zabolotskaya, E.A. (1987) Nonlinear Theory of Sound Beams, New York, AIP. Carnes, K.I., Drewniak, J.L. and F.D. (1991) ‘In utero measurement of ultrasonically induced fetal mouse temperature increases’, Ultrasound in Medicine and Biology, Vol. 17, No. 4, pp.373–382. Chapelon, J.Y., Faure, P., Plantier, M., Cathignol, D., Souchon, R., Gorry, F. and Gelet, A. (1993) ‘The feasibility of tissue ablation using high intensity electronically focused ultrasound’, IEEE Ultrasonics Symposium, pp.1211–1214. Curra, F.P., Mourad, P.D., Khokhlova, V.A., Cleveland, R.O. and Crum, L.A. (2000) ‘Numerical simulations of heating patterns and tissue temperature response due to high-intensity focused ultrasound’, IEEE trans. Ultra. Ferro. Freq. Control, Vol. 47, No. 4, pp.1077–1089. Fast algorithm in estimating high intensity 225 Damaniou, C.A. and Hynynen, K. (1994) ‘The effect of various physical parameters on the size and shape of necrosed tissue volume during ultrasound surgery’, Journal of the Acoustical Society of America, Vol. 3, pp.1641–1649. Das, S.K., Clegg, S.T. and Samulski, T.V. (1999) ‘Computational techniques for fast hyperthermia temperature optimization’, Med. Phys., Vol. 26, No. 2, pp.319–328. Davies, B. (1978) Integral Transforms and Their Applications, Springer-Verlag, New York. Goldberg, S.N., Gazelle, G.S., Solbiati, L., Livraghi, T., Tanabe, K.K., Hahn, P.F. and Mueller, P.R. (1998) ‘Alabtion of liver tumors using percutaneous RF therapy’, American Journal of Roentgen, Vol. 170, pp.1023–1028. Hwang, J.H., Wang, Y.N., Warren, C., Upton, M.P., Starr, F., Zhou, Y.F. and Mitchell, S. (2009) ‘Pre-clinical in vivo evaluation of an extracorporeal HIFU device for the treatment of pancreatic cancer’, Ultrasound in Medicine and Biology, Vol. 35, No. 6, pp.967–975. Kenney, J.E. (2005) ‘High-intensity focused ultrasound in the treatment of solid tumors’, Nature Review – Cancer, Vol. 5, pp.321–327. Lafon, C., Zderic, V., Noble, M.L., Yuen, J.C., Kaczkowski, P.J., Sapozhnikov, O.A., Chavrier, F., Crum, L.A. and SVaezy, S. (2005) ‘Gel phantom for use in high-intensity focused ultrasound dosimetry’, Ultrasound Med. Biol., Vol. 31, No. 10, pp.1383–1389. Lele, P.P. (1967) ‘Production of deep focal lesions by focused ultrasound current status’, Ultrasonics, Vol. 5, pp.105–112. Pennes, H.H. (1948) ‘Analysis of tissue and arterial blood temperatures in the resting human forearm’, J. Appl. Physiol., Vol. 1, No. 2, pp.93–122. Sapareto, S. and Dewey, W. (1984) ‘Thermal dose determination in cancer therapy’, International Journal of Radiation Oncology, Biology Physics, Vol. 10, No. 6, pp.787–800. Sohrab, B., Farzan, G., Ashkan, B. and Amin, J. (2006) ‘Ultrasound thermotherapy of breast: theoretical design of transducer and numerical simulation of procedure’, Japanese Journal of Applied Physics, Vol. 45, No. 3A, pp.1856–1863. Soneson, J.E. (2008) ‘A user-friendly software package for HIFU simulation’, 8th International Symposium on Therapeutic Ultrasound, Minneapolis, MN. Stakgold, I. (1979) Green’s Functions and Boundary Value Problems, John Wiley and Sons, New York, pp.488–489. Watkin, N.A., ter Haar, G.R. and Rivens, I. (1996) ‘The intensity dependence of the site of maximal energy deposition in focused ultrasound surgery’, Ultrasound in Medicine and Biology, Vol. 22, No. 4, pp.483–491. Zhou, Y.F. (2011) ‘High intensity focused ultrasound in clinical tumor ablation’, World J. Clin. Oncol., Vol. 2, No. 1, pp.8–27. Zhou, Y.F., Kargl, S.G. and Hwang, J.H. (2008) ‘Producing uniform lesion pattern in HIFU ablation’, 8th International Symposium on Therapeutic Ultrasound, Minneapolis. Zhou, Y.F., Kargl, S.K., Kim, K. and Hwang, J.H. (2007) ‘Comparison of pathway in high intensity focused ultrasound (HIFU) lesion production’, 154th Meeting of Acoustical Soceity of America, New Orleans, USA. Zhou, Y.F., Zhai, L., Simmons, R. and Zhong, P. (2006) ‘Measurement of high intensity focused ultrasound field by a fiber optic probe hydrophone’, J. Acoust. Soc. Am., Vol. 120, No. 2, pp.676–685.