CLINICAL RESEARCH Europace (2014) 16, 1787–1794 doi:10.1093/europace/euu079 Leads and lead extraction Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads Fatma Demirel, Ahmet Adiyaman, Peter Paul H.M. Delnoy, Jaap Jan J. Smit, Anand R. Ramdat Misier, and Arif Elvan* Department of Cardiology, Isala Klinieken, Dr. Van Heesweg 2, 8025 AB, Zwolle, The Netherlands Received 9 January 2014; accepted after revision 24 March 2014; online publish-ahead-of-print 19 May 2014 Aim ----------------------------------------------------------------------------------------------------------------------------------------------------------Keywords Intracardiac cardioverter-defibrillator † Lead failure † Externalization † Electrical lead failure Introduction In December 2011, the US food and drug administration classified a device advisory regarding the poor performance of the Riata intracardiac cardioverter-defibrillator (ICD) shock leads as a Class I recall.1 The manufacturer (St. Jude Medical) stopped the distribution of these defibrillation leads in 2010, of which the total number sold worldwide is estimated to be 227.000 over a period of 9 years. Insulation abrasion is the most common cause of ICD lead failure.2 Riata leads are prone to a unique form of failure mechanism with abrasions of the silicone insulation (inside-out abrasion). The high- and/or low-voltage conductors wear through the silicone insulation resulting in externalization and possible electrical dysfunction.3 The rates of conductor externalization have been noted to range between 11 and 22% with a higher prevalence seen among the 8 Fr. Riata leads.4 – 13 Moreover, in addition to structural failure, Riata leads have been shown to have higher rates of electrical failure when compared with standard diameter ICD leads.14,15 Nowadays, the clinical predictors and pattern in the time of the Riata lead for structural and electrical failure are not well known. The consequences of externalization without electrical abnormalities remain uncertain. In this study, we sought to: (i) identify the clinical predictors of Riata lead failure (electrical and/or structural). (ii) determine the association between structural and electrical failure of the Riata lead. (iii) analyse the rates of structural and electrical failure over time. * Corresponding author. Tel: +31 384242374; fax: +31 384243222. E-mail address: v.r.c.derks@isala.nl Published on behalf of the European Society of Cardiology. All rights reserved. & The Author 2014. For permissions please email: journals.permissions@oup.com. Downloaded from by guest on September 30, 2016 Riata implantable cardioverter-defibrillator leads are prone to failure by conductor externalization and/or electrical dysfunction. The objectives of this study were to determine the predictors of the Riata lead failure, to assess the association of conductor externalization and electrical lead failure, and to analyse the rates of lead failure over time. ..................................................................................................................................................................................... Methods Of 273 implanted Riata leads in our centre, 197 were investigated according to the Riata recall protocol, including electrical measurements by device interrogation and annually fluoroscopy. During a mean follow-up period of and results 5.6 + 1.4 years, Riata lead failure was 18.8% (37 of 197) for externalization and 17.3% (34 of 197) for electrical lead failure. Electrical lead failure was correlated with time after implant. Externalization and electrical dysfunction co-existed in only 6 of 197 (3%) patients and were not related (Phi’s coefficient 20.013, P ¼ 0.85). During the second annual screening, 145 (73.6%) patients underwent fluoroscopy and 9 patients had novel externalizations resulting in an incidence of 6.72%/patient/year which was higher than expected based on cross-sectional analysis. Besides, there was a significant increase in the extent of externalization (17.65 + 11.14 mm vs. 21.77 + 11.95 mm, P ¼ 0.001). In multivariate Cox regression analysis, non-ischaemic cardiomyopathy and impaired LVEF were independent predictors of externalization, and 7 Fr lead was a predictor of electrical lead failure. ..................................................................................................................................................................................... Conclusion Riata leads show progressive and high externalization rates without correlation between externalization and electrical lead failure. Non-ischaemic cardiomyopathy and impaired LVEF are independent predictors of structural lead failure in cross-sectional analysis, whereas 7 Fr lead is a predictor of electrical lead failure. 1788 What’s new? † We describe the results of longitudinal follow-up showing an increase of externalization rate and an increase in the extent of externalization over time and the lack of association between externalization and electrical dysfunction. † Identification of predictors for Riata lead failure, i.e. nonischaemic cardiomyopathy and impaired LVEF for structural lead failure and 7 Fr for electrical lead failure. Methods Chest radiography and Fluoroscopy Cine-fluoroscopy was performed in all patients using several projections including anterior– posterior, left-anterior-oblique 458, and rightanterior-oblique 458, and additional projections as needed for optimal identification of conductor externalization. The projections were examined at the time of image acquisition and reviewed by a team of cardiac electrophysiologists with extensive experience in device implantation and lead cine fluoroscopy. Intracardiac cardioverter-defibrillator interrogation In all patients, ICD interrogation was performed with measurement of low and high-voltage impedances, sensing and capture threshold values. Furthermore, all measurements were repeated during manipulation of the pocket/device and during isometric arm exercise. The device was additionally reviewed for inappropriate sensing events (noise and oversensing). During follow-up, several device interrogations (outpatient clinic visits with 6 months intervals or shorter when indicated, and remote-monitoring) took place. Definition of structural failure of the Riata lead Externalization was defined as conductor(s) visible outside the lead body on fluoroscopy in any of the projected views. The location of externalization was divided into four zones (Figure 1). The size of externalization was described in length by mm fluoroscopically calibrated for the thickness of the lead. The length of the externalization is measured on the projection with the largest visual length and controlled in the other available projections. Definition of electrical lead failure Electrical lead failure was considered if it met one of the following criteria: (i). presence of non-physiological signals on the intracardiac ventricular electrogram, (ii) pacing impedance outside the interval 200 – 2000 V or .100% increase or .50% decrease of the stable baseline impedance, (iii) change in high-voltage impedance to .200 or ,25 V, (iv) pacing threshold .5 V or .100% increase, (v) R-wave sensing ,3.0 mV or .50% reduction. Lead dislodgment, T-wave oversensing, physiological oversensing, and header problems were not considered as lead failures. Adverse events were defined as an inappropriate shock or failure of pacing or defibrillation, caused by lead dysfunction. For survival analysis, the date of lead failure was identified using device interrogation as the first occurrence of any lead abnormality fulfilling the definition of lead failure as stated above. Statistical analysis Continuous variables are expressed as mean + SD and significant differences were analysed by Student’s t-test or Mann– Whitney U test. Categorical data are expressed as number and percentages and compared with x2 test or Fischer’s exact test, when appropriate. We tested the variables for normally distribution by skewness and kurtosis and plotted with P– P and Q– Q plots. We used Spearman’s correlation for continuous variables, when not normally distributed and Phi’s coefficient for nominal binary variables. To identify the baseline characteristics associated with adverse outcome (electrical and/or structural lead failure), univariate and multivariate Cox proportional hazards regression analysis was performed. Before entering the covariates in the multivariate Cox regression model, we assessed co-linearity between covariates. We only entered covariates that were not significantly intercorrelated (R: 0.06 20.23 and VIF: 1.000– 1.089). Multivariate Cox regression analysis was performed with a forward conditional model of variables and removal of variables at a P-value of 0.15, with externalization or electrical dysfunction as dependent variables. Gender, lead size (7 or 8 French), type of cardiomyopathy (ischaemic vs. non-ischaemic) and left ventricular ejection fraction (LVEF) were used as covariates. Relative risk was expressed as hazard ratio (HR) with associated 95% confidence interval (CI). Statistical analysis was performed using SPSS version 20. A P-value of ,0.05 was considered statistically significant. Results Clinical characteristics At the time of recall we could analyse 197 patients, because 68 patients were deceased and 8 patients were lost to follow up (migration of the patient or inability of the patient to attend the check-up). In the deceased patients, there were no reports of lead externalization or lead dysfunction in the medical records. The study population consisted of 32 females (16.2%) and 165 males (83.8%), mean age was 70.4 + 10.1 years. The median follow-up was 5.8 years (IQR, 5.1 –6.4 years) and the mean follow-up was 5.6 + 1.4 years. The mean follow-up was significantly longer for the Riata 8 Fr than for the 7 Fr Riata ST (6.6 + 1.2 years vs. 5.2 + 1.3 years; P , 0.0001). The 7 Fr Riata ST lead was more prevalent (69%). The majority of the screened leads had a single coil design (59%). Twenty-seven of the 197 patients underwent lead revision prior to the formal Riata recall due to several reasons, i.e. 15 patients had electrical lead failure, 2 patients had lead dislocation, 2 patients had lead perforation, 7 patients had lead related infection, 1 patient had an isolation Downloaded from by guest on September 30, 2016 On 4 January 2012, a recommendation was given by the Device Advisory Committee of the Netherlands Heart Rhythm Association (NHRA) to identify all patients with an functional Riata or Riata ST lead and to perform fluoroscopic screening and electrical interrogation of these leads.16 According to the NHRA recommendation, a first screening with fluoroscopy took place in 2012 and a second annual screening in 2013. We investigated all 273 patients who underwent implantation of a Riata lead in our centre. This current study involves partly known data which is published by Theuns et al.10 This current study is an analysis of both retrospectively and prospectively collected data on all Riata (Model 1580, 1581, 1582) and Riata ST (Model 7000, 7001, 7002) leads implanted between February 2005 and February 2008 at the Isala Klinieken, Zwolle, the Netherlands. This prospective registry was approved by the Hospital board. The collected data include the lead model, date of implantation, date of screening, presence of externalized conductors, location of externalized conductors, extent of externalization, presence of electrical dysfunction, type of electrical dysfunction, and adverse events. F. Demirel et al. 1789 Mechanical and electrical dysfunction of Riata leads A C D B A3 A A2 A A1 B A N (%) N (%) of externalization 1(2.8) A1 N (%) A2 N (%) 16(44.4) 12(33.3) A3 N (%) B N (%) 5(13.9) 1(2.8) C N (%) 1(2.8) D N (%) 1(2.8) p-value <0.001 Figure 1 (A) Location of externalization divided into four zones; A: Distal to SVC-coil (dual coil), distal to superior vena cava (single coil), divided into A1: proximal RVC-coil; A2: tricuspid valve annulus; A3: distal SVC-coil to tricuspid valve annulus (dual coil), right atrium (single coil), B: distal to clavicle including SVC-coil (dual coil) superior vena cava (single coil), C: near clavicle, D: clavicle to CAN. (B) Table showing the rates of externalization per zone. defect that was observed during upgrading to a CRT system with placement of a LV-lead, and 1 patient underwent preventive lead replacement at the time of device replacement due to elective replacement interval. Riata leads were successfully extracted in 14 of 27 patients without adverse events. In the 13 patients without extraction, the Riata lead was capped in situ. At the time of revision or extraction fluoroscopy was available in all patients. First formal Riata recall screening During the formal Riata recall, 170 patients underwent fluoroscopic screening and ICD interrogation. The mean follow-up at the time of fluoroscopy was 5.0 + 0.7 years. In 26 patients Riata lead externalization was documented during fluoroscopic screening. Of these 26 patients, 25 patients underwent lead revision, i.e. a new shock lead was implanted and the Riata lead was capped. During follow-up, a total of 51 lead revisions were performed. Of these patients, 14 underwent preventive revision at ERI and in 2 patients Riata lead extraction was performed. Sixteen patients died during follow-up and 31 patients were lost to follow-up (migration, clinical condition prohibiting patients to adhere to the recall). Second formal Riata recall screening During the second annual screening, 1 year after the first screening, 134 patients underwent fluoroscopy of whom 90 patients had a functional Riata lead and 44 had a non-functional Riata lead. The mean follow-up at the time of fluoroscopy was 6.1 + 0.8 years. Additionally nine leads showed novel externalizations and five leads had developed electrical lead failure over a period of 1 year. Thus, a total of 37 structural lead failures and 34 electrical lead failures were found at the time of the last follow-up (Figure 2). Externalization Table 1 shows externalization rates and characteristics of the Riata/ Riata ST leads screened in 2012 and 2013. The prevalence of externalization was 14.2% (28 of 197) in 2012 and 18.8% (37 of 197) in 2013. Cross-sectional analysis showed a total externalization rate of 3.35%/patient/year. In the longitudinal follow-up, 6.72% (9 of 134) patients showed novel externalization, suggesting a higher externalization rate than in the cross-sectional analysis. Eighteen patients (64.3%) who had fluoroscopically documented externalization of the Riata leads in 2012, underwent a second fluoroscopic as part of the regular annual screening and showed an increase in the extent of externalization over a period of 1 year (17.65 + 11.14 mm vs. 21.77 + 11.95 mm, P-value ¼ 0.001). The four zones of externalization are shown in Figure 1. The most common place of externalization was in the A zone, proximal to the RVC coil in the right atrium, near the tricuspid valve (75.7%). Downloaded from by guest on September 30, 2016 Location of externalization 1790 F. Demirel et al. 273 riata leads implanted 68 pt deceased 8 pt lost to FUP 197 pt able to analyze subcutaneously implanted device (P ¼ 0.02) had significantly more externalization rates. In univariate and multivariate Cox regression analysis, nonischaemic cardiomyopathy (HR 6.33; 95% CI: 3.07 –13.03; P ≤ 0.001) and LVEF (HR 0.94; 95% CI: 0.91 –0.98; P ¼ 0.02) were identified as independent predictors of externalization of the Riata leads (Table 3). 27 pt revision before recall (14 lead extraction) –15 electrical lead failures ; 2 externalizations Electrical lead failure National screening of riata recall 2012 170ICD interrogations 173 fluoroscopic screening (3 screening while revision) 26 additional externalizations 14 additional electrical lead failures 51 lead revisions after recall (2 lead extraction) 16 pt deceased during 2012 31 pt lost to FUP for fluoroscopic screening National second screening of riata recall 2013 134 leads (90 active leads) (44 non-active leads) Total 37 externalizations 34 electrical lead failure Figure 2 Flowchart of patient follow-up. Table 2 shows that women had higher externalization rates than men (31.2 vs. 16.4%, P ¼ 0.05) and 8 Fr Riata leads had more externalization than 7 Fr Riata ST leads (34.4 vs. 11.8%; P , 0.001). Furthermore, patients with impaired LVEF (P ¼ 0.05), non-ischaemic cardiomyopathy (P , 0.001), biventricular ICD (P ¼ 0.003), and a Table 1 Rates of externalization in 2012 and 2013 Model Coils N (%) EX in 2012 (% within lead model) Dwell time 2012 (years + SD) EX in 2013 (% within lead model) Dwell time 2013 (years + SD) ............................................................................................................................................................................... 8 Fr Riata leads Riata 1580 Riata 1581 Dual Dual 28 (14.2) 12 (6.1) 5 (17.9) 5 (41.7) 6.1 + 0.3 5.5 + 0.9 7 (29.2) 5 (50.0) 7.2 + 0.4 6.4 + 1.2 Riata 1582 Single 21 (10.7) 6 (28.6) 5.3 + 1.3 9 (64.3) 6.0 + 1.6 7 Fr Riata ST leads Riata ST 7000 Dual 28 (14.2) 3 (10.7) 4.6 + 0.8 4 (21.1) 5.5 + 1.1 Riata ST 7001 Dual 12 (6.1) 2 (16.7) 3.7 + 1.7 2 (25.0) 4.7 + 2.0 Single 96 (48.7) 197 7 (7.3) 28 (14.2) 4.2 + 1.2 10 (14.3) 37 (18.8) 4.9 + 1.3 Riata ST 7002 Total Fr, French; EX, externalization. Downloaded from by guest on September 30, 2016 9 additional externalizations 5 additional electrical lead failures All functional Riata leads were monitored with device interrogation (clinical and remote monitoring) during a median follow-up of 5.8 years. In Table 4, the rates of electrical lead failure are analysed per lead type. In cross-sectional analysis, 34 (17.3%) of the leads showed signs of electrical lead failure which resulted in an incidence rate of 2.97%/patient/year and a significant correlation of dwell time (Spearman’s correlation 0.99, P , 0.001, Figure 3). None of the baseline characteristics were significantly different between patients with or without electrical lead failure. Of the 34 patients with electrical lead failure, 8 (23.5%) had detection of lead noise, 18 (52.9%) had abnormal pace/sense impedance changes, 8 (23.5%) abnormal shock impedance changes, 6 (17.6%) alteration in pacing threshold, and 1 (2.9%) patient had a drop in R-wave amplitude without dislocation (Figure 4). Eight patients had more than one feature of electrical lead failure. In the univariate and multivariate Cox regression analysis, only 7 Fr Riata lead (HR 3.89; 95% CI: 1.561 –9.708; P ¼ 0.004) was an independent predictor of electrical lead failure. Prospectively, between 2012 and 2013, 5 of the 107 functional Riata leads developed electrical lead failure which resulted in an incidence rate of 4.67%/year. Overall 4 of 197 (2.0%) patients had inappropriate shocks due to oversensing of non-physiological noise on the ventricular sensing channel. None of these patients showed signs of externalization and inappropriate shocks were not correlated to externalization (Phi’s coefficient 20.07, P ¼ 0.33). All of the patients with inappropriate shocks showed electrical lead failure, with a significant correlation between electrical dysfunction and inappropriate shocks (Phi’s coefficient 0.31, P , 0.001). 1791 Mechanical and electrical dysfunction of Riata leads Table 2 Clinical characteristics for externalization and electrical lead failure EX 2013 N 5 37 (18.8%) No-EX 2013 N 5 160 (81.2%) Male 27 (16.4) 138 (83.6) Female Mean age + SD (years) 10 (31.2) 71.4 + 8.4 24 (68.8) 70.6 + 10.1 21 (34.4) 16 (11.8) 40 (65.6) 120 (88.2) 32 (18.6) 5 (20.8) 3 (10.3) 11 (11.8) P-value ED 2013 N 5 34 (17.3%) No-ED 2013 N 5 161 (82.7%) P-value ............................................................................................................................................................................... Gender 0.05 29 (17.6) 136 (82.4) 0.79 0.62 5 (15.6) 70.7 + 10.3 27 (84.4) 70.8 + 9.7 0.98 ,0.001 10 (16.4) 24 (17.6) 51 (83.6) 112 (82.5) 0.83 140 (81.4) 19 (79.2) 0.79 30 (17.4) 4 (16.7) 142 (82.6) 20 (83.3) 0.93 26 (89.7) 82 (88.2) 0.003 6 (20.7) 15 (16.1) 23 (79.3) 78 (83.9) 0.85 Lead type 8 Fr 7 Fr ICD indication Primary Secondary Type ICD 1 Chamber 2 Chamber Biventricular LVEF + SD (%) Type CMP ICMP 23 (31.1) 51 (68.9) 13 (17.6) 61 (82.4) 25.7 + 8.5 28.9 + 9.0 0.05 28.2 + 6.2 28.4 + 9.4 0.91 ,0.001 22 (17.6) 103 (82.4) 0.90 12 (16.9) 59 (83.5) 114 (91.2) 26 (36.6) 45 (63.4) Subclavian 33 (18.6) 144 (81.4) Cephalic ICD pocket side 4 (26.7) 11 (73.3) Left Right Type of pocket 35 (19.3) 146 (80.7) 2 (15.4) 11 (84.6) Subcutaneous 33 (25.8) 95 (74.2) Subpectoral 3 (8.1) 34 (91.9) 0.45 0.73 0.02 33 (18.6) 144 (81.4) 1 (6.7) 14 (93.3) 31 (17.1) 150 (82.9) 3 (23.1) 10 (76.9) 23 (18.0) 105 (82.0) 9 (24.3) 28 (75.7) 0.24 0.59 0.39 EX, externalization; CMP, cardiomyopathy; ICMP, ischaemic cardiomyopathy; DCMP, dilated cardiomyopathy; HCMP, hypertrophic cardiomyopathy; VF, ventricular fibrillation; VT, ventricular tachycardia. Table 3 Univariate and multivariate Cox regression analysis for externalization Univariate HR 95% CI P-value Age (years) 0.99 0.97–1.04 0.962 Female gender 8 Fr 2.63 1.01 1.26–5.47 0.50–2.06 0.01 0.972 Primary indication 0.96 0.37–2.48 Non-ICMP 2/3-chamber device 5.35 1.39 2.63–10.87 0.43–4.54 Multivariate HR 95% CI P-value 2.00 0.95– 4.20 0.07 6.33 3.07– 13.03 ,0.001 0.94 0.91– 0.98 0.02 ............................................................................................................................................................................... 0.935 ,0.001 0.585 Left-sided pocket 1.08 0.26–4.51 0.916 Subcutaneous pocket Subcutaneous entrance 2.59 2.08 0.79–8.51 0.73–5.91 0.115 0.169 LVEF (%) 0.95 0.91–0.99 0.018 EX, externalization; CMP, cardiomyopathy; ICMP, ischaemic cardiomyopathy; DCMP, dilated cardiomyopathy; HCMP, hypertrophic cardiomyopathy; VF, ventricular fibrillation; VT, ventricular tachycardia. Downloaded from by guest on September 30, 2016 11 (8.8) Non-ICMP Venous entrance 1792 F. Demirel et al. Table 4 Rates of electrical lead failure and adverse events Model Coils N (%) ED (% within lead model) N=7 (24%) AD (% within lead model) N = 17 (59%) ................................................................................ N=2 (7%) 8 Fr Riata leads Riata 1580 Dual 28 (14.2) 5 (17.9) 1 (3.6) Riata 1581 Dual 12 (6.1) 0 (0.0) 0 (0.0) 21 (10.7) 5 (23.8) 1 (4.8) Riata 1582 Single 7 Fr Riata ST leads Riata ST 7000 Dual 28 (14.2) 8 (28.6) 1 (3.6) Riata ST 7001 Riata ST 7002 Dual Single 12 (6.1) 96 (48.7) 2 (16.7) 14 (14.6) 0 (0.0) 1 (1.0) 197 34 (17.3) 4 (2.0) Total impedance alteration impedance alteration & threshold alteration N=3 (10%) threshold alteration noise/oversensing Figure 4 Rates of different forms of electrical lead dysfunction. Ten patients had decreased impedance; nine patients had increased impedance, and one patient with varying impedance. Fr, French; ED, electrical dysfunction; AD, adverse events. Cumulative electrical lead failure (%) 20.00 Spearman’s rho 0.99, P<0.001 15.00 10.00 5.00 0 0 20 40 60 80 Time after implantation (months) B 100 Time 0 20 40 60 80 100 N remaining cases 147 145 143 125 27 0 N cumulative events 0 2 4 10 27 34 Figure 3 (A) Cumulative incidence of externalization which rises linearly between 47 and 80 months (10–90th percentile) of followup. (B) Cumulative incidence of electrical dysfunction which rises linearly between 47 and 90 months (10–90th percentile) of follow-up. Association of externalization and electrical lead failure Only 6 (16.2%) of the 37 externalized leads had electrical lead failure, whereas 28 (17.5%) of the 160 non-externalized leads had electrical lead failure (Fischer’s exact test P ¼ 1.00). Structural and electrical lead failure co-existed only in six patients (3.0% of the population), and did not correlate (Phi’s coefficient 20.013, P ¼ 0.85). Discussion In this study, we found that (i) there is a significant increase in electrical dysfunction and externalization rates, with prospective incidence Construction and mechanism of failure of the Riata lead Riata leads have a silicone insulation with a unique multilumen construction.17 The hollow multilumen design results in the fact that movement of internal conductors can occur, which increases the risk of externalization through the silicone insulation (e.g. ‘inside-out abrasion’) or conductor fracture. The inner pace-sense conductor is encased in polytetrafluoroethylene and the outer lumens contain conductors covered with ethylene tetrafluoroethylene (ETFE). The Riata lead contains, depending on the specific subtype, two to four conductors that operate as reinforcement for the lead and are electrically non-functional. Structural lead failure In our current report, a significant proportion of the Riata leads had signs of externalization, with a higher incidence rate in prospective longitudinal analysis (6.72%/year), compared with the suggested rate by cross-sectional analysis (3.35%/year). Current available data by the manufacturer (St. Jude Medical) and others reported comparable rates of externalization.1,10 – 13 However, the prevalence of externalization has been reported to range from 10 to 33% when systematic fluoroscopy was performed.7,9 Several reports showed that fluoroscopy should be considered as the golden standard for the diagnosis of externalization and both the NHRA and the FDA recommend systematic fluoroscopy.16,18,19 Downloaded from by guest on September 30, 2016 rates higher than suggested by cross-sectional analysis. Furthermore, the extent of externalization increased during follow-up. (ii) Based on cross-sectional analysis externalization and electrical lead failure were not related. Impaired LVEF and non-ischaemic cardiomyopathy were independent predictors of lead externalization and the 7 Fr lead was an independent predictor of electrical lead failure. These crosssectional data should, however, be interpreted with caution. For the best of our knowledge, this is the first study with both cross-sectional and longitudinal analysis of both externalization and electrical lead failure. A 1793 Mechanical and electrical dysfunction of Riata leads Electrical lead failure Riata leads showed high electrical failure rates, with prospective longitudinal incidence numbers higher (4.67%/year) than the linearly increasing rates suggested by cross-sectional analysis (2.97%/year). In previous studies, although the definition of failure has varied significantly, the rates of electrical lead failure ranged from ,1 to 57%. The most common form of electrical dysfunction was alteration in impedance, of which approximately half was abnormally low impedance, which can be caused by insulation abrasion. Other reports concluded that noise/oversensing was the main cause of electrical dysfunction.7,11,15 It should be stated that most of the groups investigated the association between electrical and structural lead failure, only few groups investigated the relationship with other factors.7,24,25 Cheungh et al.24 showed that female gender and younger age as independent predictors for electrical lead failure and showed a linear increase of electrical lead failure. Parkash et al. found a higher rate of electrical failure in 8 Fr Riata leads. In our report, 7 Fr lead was the only independent predictor for electrical lead failure. While controversy exists regarding the association between electrical and structural failures of the Riata lead, it is still unclear if externalization without electrical lead failure could be dangerous.8,9,11 Several reports have investigated the association of electrical lead failure with externalization in detail, which is the most important question in assessing the exact meaning and consequences of externalization.7 – 13,15,24,25 Longitudinal follow-up of patients with externalization without electrical dysfunction seems essential to answer this question. Liu et al.26 partially assessed this issue, by analysing electrical failure rates in the externalized leads, which was as high as 6.4% per year. Therefore, frequent electrical follow-up and where possible, home-monitoring, is mandatory in patients with externalization. We observed that only in a small proportion of patients with lead externalization (16.2%), electrical lead failure is present. Furthermore, we found no association of electrical lead failure with externalization. Our report is in line with previously published studies and the multicentre, prospective Riata Lead Evaluation study.4,8,9,13,23 In controversy, the national registry of the Netherlands Heart Rhythm Association found a significantly higher rate of electrical abnormalities in externalized leads than in non-externalized leads.10 Parvathaneni et al.7 did not see significant differences, but concluded that ‘serious’ electrical dysfunctions (noise, high threshold, lead fracture, high voltage short) were more likely to be present in leads with externalization. Conductor externalization without electrical dysfunction could be due to externalization of electrically non-functional conductors which are present in Riata leads or due to the intact ETFE coating of functional conductors which could have prohibited electrical abnormalities. It is not known whether the ETFE coating in externalized conductors can withstand high-voltage shocks, when low-voltage tests are normal.9 Adverse events due to Riata leads are rare, but we found a significant association with electrical lead failure and no association with externalization that is in line with previous studies.7,10,11 Previous studies found an accelerating or a linear increase of lead failure over time. In our analysis, we identified a higher incidence rate in the prospective analysis compared with the cross-sectional analysis, emphasizing the need for prospective data regarding this topic. This pattern of lead failure is also seen for Sprint Fidelis leads.27 It is likely that this pattern may be representative for most of the ICD leads and most leads require a minimum of dwell time before the failures occur. Limitations The partly cross-sectional design of the study could prohibit us to draw definite conclusions regarding the exact time of externalization and time dependency of lead failure, therefore the cross-sectional data should be interpreted cautiously. The longitudinal phases of the study provide additional insight on the time course of externalization and electrical failure. However, even in the longitudinal phase the exact time of externalization remains uncertain. Furthermore, like in most of the present reports, we did not have any information regarding deceased patients. Clinical implications This study demonstrates high rates of structural and electrical failure of the Riata lead. Despite the high rates, there was no relation between structural and electrical failure. Inconsistency exists about Downloaded from by guest on September 30, 2016 All studies are in line with the fact that externalization rates increase over time. In our follow-up, all functional and non-functional Riata leads were screened by fluoroscopy and seemed to show a significant correlation with dwell time. These results should be interpreted cautiously, since the exact time of externalization is unknown. However, this accounts for both prospective and crosssectional analysis, as also in the prospective study the exact time point of externalization remains unknown. We performed univariate and multivariate Cox regression analysis to determine predictors for externalization. Although we realize that this has limitations for retrospective data, of which partially the exact incidence dates are unclear, we believe that the analysis provides additive information and is hypothesis generating. These results remain to be confirmed in prospective analysis. Externalization is common; however, the amount of externalization seems to be very modest (virtually all ,2 cm).7 In our report we see small extents of externalization, except for one patient with extreme externalization without electrical dysfunction.20 Interestingly, a significant increase in the extent of externalization is seen during follow-up of both functional and non-functional Riata leads. This supports the current recommendation to perform serial followup with device interrogation and fluoroscopy in patients with a functional Riata lead and also in patients with non-functional Riata leads that were not extracted. Factors that have been associated with externalization in previous analyses are time since implantation, non-ischaemic cardiomyopathy, multiple leads, a higher percentage of RV pacing, lower LVEF, single coil leads, 8 Fr leads, lower age, apical lead position, and hypertrophic cardiomyopathy.7,9 – 11,21 – 23 In our present univariate and multivariate analysis, also non-ischaemic cardiomyopathy and impaired LVEF were associated with externalization. 8 Fr was not an independent predictor of externalization, while most of the large studies showed a significant association.10,11,22 This could be due to the larger amount (69%) of 7 Fr leads in our analysis compared with the large multicentre studies 53% (Theuns et al.), 33.4% (Hayes et al.), and 28.2% (Steinberg et al.), respectively.10,11,23 1794 the relationship between structural and electrical failure and routine fluoroscopic screening. How to handle leads with electrical dysfunction is clear. However, the key question is how to handle the leads that have externalized conductors without electrical dysfunction? Because of lack of correlation between structural and electrical failure, an appropriate individualized management should be taken for each patient by considering the risk of revision and or extraction vs. the risk of careful follow-up. Conclusion Riata leads show progressive and high rates of electrical and structural failure with progressive extent of externalization during follow-up. Based on the cross-sectional analysis, there was no association between externalization and electrical lead failure. Furthermore, we identified impaired LVEF and non-ischaemic cardiomyopathy as independent predictors for externalization and 7 Fr leads for electrical lead failure in the cross-sectional analysis. Acknowledgement The authors thank Vera Derks for her skillful secretarial assistance. Conflict of interest: none declared. 1. Carlson M, Tsung P. Medical device advisory. Sylmar, CA: St Jude Medical; 2011. 2. Kleemann T, Becker T, Doenges K, Vater M, Senges J, Schneider S et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of .10 years. Circulation 2007;115:2474 –80. 3. Hauser RG, McGriff D, Retel LK. Riata implantable cardioverter-defibrillator lead failure: analysis of explanted leads with a unique insulation defect. Heart Rhythm 2012;9:742 –9. 4. Erkapic D, Duray GZ, Bauernfeind T, De Rosa S, Hohnloser SH. Insulation defects of thin high-voltage ICD leads: an underestimated problem? J Cardiovasc Electrophysiol 2011;22:1018 – 22. 5. Kodoth VN, Hodkinson EC, Noad RL, Ashfield KP, Cromie NA, McEneaney DJ et al. Fluoroscopic and electrical assessment of a series of defibrillation leads: prevalence of externalized conductors. Pacing Clin Electrophysiol 2012;35: 1498 – 504. 6. Liu J, Rattan R, Adelstein E, Barrington W, Bazaz R, Brode S et al. Fluoroscopic screening of asymptomatic patients implanted with the recalled Riata lead family. Circ Arrhythm Electrophysiol 2012;5:809 –14. 7. Parvathaneni SV, Ellis CR, Rottman JN. High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: fluoroscopic grading scale and correlation to extracted leads. Heart Rhythm 2012; 9:1218 – 24. 8. Schmutz M, Delacretaz E, Schwick N, Roten L, Fuhrer J, Boesch C et al. Prevalence of asymptomatic and electrically undetectable intracardiac inside-out abrasion in silicon-coated Riata(R) and Riata(R) ST implantable cardioverter-defibrillator leads. Int J Cardiol 2013;167:264–257. 9. Shen S, Bhave P, Giedrimas E, Patel T, Arora R, Chicos AB et al. Prevalence and predictors of cable extrusion and loss of electrical integrity with the Riata defibrillator lead. J Cardiovasc Electrophysiol 2012;23:1207 –12. 10. Theuns DA, Elvan A, de Voogt W, de Cock CC, van Erven L, Meine M. Prevalence and presentation of externalized conductors and electrical abnormalities in Riata defibrillator leads after fluoroscopic screening: report from the Netherlands heart rhythm association device advisory committee. Circulation Arrhythm Electrophysiol 2012;5:1059 –63. 11. Steinberg C, Sarrazin JF, Philippon F, Bouchard MA, O’Hara G, Molin F et al. Detection of high incidence of Riata lead breaches by systematic postero-anterior and lateral chest X-ray in a large cohort. Europace 2013;15:402 –8. 12. Larsen JM, Riahi S, Nielsen JC, Videbaek R, Haarbo J, Due KM et al. Nationwide fluoroscopic screening of recalled Riata defibrillator leads in Denmark. Heart Rhythm 2013;10:821 –7. 13. Hauser RG, Abdelhadi RH, McGriff DM, Kallinen Retel L. Failure of a novel silicone-polyurethane copolymer (OptimTM ) to prevent implantable cardioverterdefibrillator lead insulation abrasions. Europace 2013;15:278 –83. 14. Sung RK, Massie BM, Varosy PD, Moore H, Rumsfeld J, Lee BK et al. Long-term electrical survival analysis of Riata and Riata ST silicone leads: National Veterans Affairs experience. Heart Rhythm 2012;9:1954 –61. 15. Abdelhadi RH, Saba SF, Ellis CR, Mason PK, Kramer DB, Friedman PA et al. Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm 2013;10:361 –5. 16. Inspectie voor de Gezondheidszorg - Patièˆnten met ICD-lead merk St. Jude Medical type Riata opgeroepen voor controle. http://www.igz.nl/actueel/nieuws/ patintenmeticdleadmerkstjudemedicaltyperiataopgeroepenvoorcontrole.aspx. 17. Epstein AE, Baker JH, Beau SL, Deering TF, Greenberg SM, Goldman DS. Performance of the St. Jude Medical Riata leads. Heart Rhythm 2009;6:204 – 9. 18. Safety Communications . FDA Safety Communication: Premature Insulation Failure in Recalled Riata Implantable Cardioverter Defibrillator (ICD) Leads Manufactured by St. Jude Medical, Inc. http://www.fda.gov/MedicalDevices/Safety/ AlertsandNotices/ucm314930.htm. 19. Lorvidhaya P, Mendoza I, Sehli S, Atalay MK, Kim MH. Prospective evaluation of cinefluoroscopy and chest radiography for Riata lead defects: implications for future lead screening. J Interv Card Electrophysiol 2013;38:131 – 5. 20. Demirel F, Adiyaman A, Delnoy PP, Elvan A. Extreme externalization of Riata intracardiac cardioverter defibrillator leads: a new peril of a troublesome lead. J Cardiovasc Electrophysiol 2014;25:216. 21. Moorman LP, Moorman JR, Dimarco JP, Malhotra R, Darby A, Bilchick K et al. Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads. J Interv Card Electrophysiol 2013;37:63–8. 22. Valk SD, Theuns DA, Jordaens L. Long-term performance of the St Jude Riata 1580 –1582 ICD lead family. Neth Heart J 2013;21:127 –34. 23. Hayes D, Freedman R, Curtis AB, Niebauer M, Neal Kay G, Dinerman J et al. Prevalence of externalized conductors in Riata and Riata ST silicone leads: results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm 2013;12:1778 –82. 24. Cheung JW, Al-Kazaz M, Thomas G, Liu CF, Ip JE, Bender SR et al. Mechanisms, predictors, and trends of electrical failure of Riata leads. Heart Rhythm 2013;10:1453–9. 25. Parkash R, Exner D, Champagne J, Mangat I, Thibault B, Healey JS et al. Failure rate of the Riata lead under advisory: a report from the CHRS Device Committee. Heart Rhythm 2013;10:692 –5. 26. Liu J, Qin D, Rattan R, Bazaz R, Adelstein E, Jain S et al. Longitudinal follow-up of externalized Riata leads. Am J Cardiol 2013;10:1616 –8. 27. Cheung JW, Tobin-Hess A, Patel A, Slotwiner DJ, Goldner BG. Trends in Fidelis lead survival: transition from an exponential to linear pattern of lead failure over time. Circ Arrhythm Electrophysiol 2012;5:906 –12. Downloaded from by guest on September 30, 2016 References F. Demirel et al.