Microbiology 345: Setting up a gel filtration column The column has been mounted on a ring stand and the outlet closed. 50mL Tris buffer is added to the column. The column is filled with a slurry of 50% sephadex G-75. The slurry must be stirred thoroughly before pouring. More sephadex is added frequently to pack the column. The final surface of the gel bed should be 1-2 cm below the top of the column (note that ours is lower so the top of the gel bed is visible). Notice a distinct difference between packed material and unpacked slurry. Video shots were taken at several points during the packing. The outlet is lowered throughout the packing procedure to slow the rate of dripping. The outflow is closed and the Mariotte flask disconnected. The cap is taken off and the buffer above the gel is removed with a pipette. Once the sephadex has settled, the outlet is taped just below the top of the column and opened to allow buffer to flow through the column. The top of the column is closed tightly. The Mariotte flask is attached through the tubing to the top of the column. The column is washed with 300 mL buffer. The sample to be run is gently loaded by using a pasteur pipette. The tip of the pipette is held against the wall of the column and moved in a circular motion so the sample slowly layers onto the gel bed. Do NOT break the pasteur pipette! The effluent is collected in a beaker. Once the wash is complete the column set-up is taken to the cold room and run overnight to equilibrate. Note: it is vital that that the column be set up properly so it does not dry out, especially if left unattended. The pressure head (space between the buffer and outlet) should never be more than 20 cm. The sample is drawn into the gel bed. Notice the height of the liquid decreases during this part. The column is washed 2 -3 times with 2 mL aliquots of buffer. Once washed, the outlet is closed and the column is filled with buffer. The Mariotte flask with 400-500 mL buffer is attached. A pressure head of 15 cm is used. The outlet is atteched to an automatic fraction collector. The column should be run slowly and should be set up so it does NOT dry out.