L-9904-2390 Application note RLE plane mirror application note The RLE fibre optic laser encoder uses interferometry to provide high resolution, high linearity position feedback. Any RLE system comprises an RLU laser unit, RLD10 detector head and an additional target optic, either a plane mirror or a retroreflector, to complete the interferometer configuration. Mirror selection and installation should be carried out carefully to avoid degrading the metrological advantages of the interferometer system. In the case of an X-Y stage application mirrors can contribute to system errors via: • Surface non-uniformity (flatness), which can be minimised by using an optically flat target and approved mounting techniques. • Thermal expansion, which can be reduced by using the correct substrate and mounting materials. • Misalignment, which can be avoided through careful installation. This document discusses how these errors may be minimised through careful mirror specification, mounting and alignment. Mirror flatness In multi-axis applications, such as an X-Y stage, the interferometer’s laser beam will traverse the length of the mirror as the stage travels along the cross axis. An uneven surface results in a path length change between the target optic and the interferometer, which cannot be distinguished from actual motion. Global flatness Global flatness is macro distortion of a mirror. This figure is usually specified as a maximum peak to valley figure over a particular length. Global distortion can cause measurement errors and beam misalignment. Any distortion of mirror structure is very much dependant on the method of mounting used (mounting methods are discussed later in this document). The table below shows the measurement error possible with different flatness levels. Flatness* Maximum error (nm) λ/4 158 λ/8 79 λ/10 63 λ/12 53 λ/20 32 Mirror flatness, which generally describes undulating distortion, is specified as local or global (see Figure 1). Local flatness Global flatness Figure 1: Local and global flatness N.B. Diagram exaggerated for clarity Local flatness Local surface flatness distortions of a mirror can degrade the fringe contract in the interferometer and cause measurement errors. Local flatness is usually specified as a peak-to-valley figure, with the error quoted as a fraction of a wavelength. * λ = 633 nm Some motion controllers are capable of generating a flatness correction table, which means that it is possible to compensate for any residual distortion after mounting. (Error mapping is discussed laser in this document.) Cosmetic surface quality Local flatness can be reduced by purchasing mirrors with a very high flatness specification over the local area. This method is suitable when using small mirrors but notoriously expensive when longer mirrors are required. Surface flatness generally describes undulating distortions, whilst cosmetic surface quality refers to discontinuities (defects). Poor cosmetic surface quality can reduce signal intensity and possibly cause beam mis-alignment and measurement errors. A local flatness of λ/4 may reduce the fringe intensity to a level that renders interferometry impossible. Renishaw recommends using mirrors with a local flatness of not less than λ/10 over an area of 12 mm by 7 mm. Surface quality is described using the military scratch and dig specification MIL-0-13830A, e.g. a 60 - 40 specification allows scratches of up to 60 µm in width and digs of up to 0.4 mm in diameter. Renishaw recommends using a mirror with a scratch and dig specification of no more than 60 - 40. For a more detailed description of this specification, including maximum combined length figures, please contact a Renishaw representative or reference MIL-0-13830A. Thermal expansion error Mirror substrate material can contribute to measurement error through thermal expansion, although it will usually be negligible when compared to the expansion of the rest of the system. In order to minimise this error, a material with a low coefficient of thermal expansion (CTE) should be used. The table below shows the CTE (ppm/°C) for some common substrates and the maximum measurement error (µm) generated by a 25 mm thick substrate over a 1 °C change. CTE (ppm/°C) Maximum error (µm) Fused silica 0.5 0.013 Zerodur™ 0.1 0.003 8 0.2 Typical glass Angular deviation (arc sec) Measurement error (ppm) Measurement error* (ppm) 0.5 2 0.7 1 5 1.5 2 10 3 3 15 4 4 20 6 5 24 7 * for a 300 mm long mirror • Cosine error is caused when the mirror is not perpendicular to the axis of motion, or the laser beam is not parallel to the axis of motion (see Figure 4). Interferometer Stage travel Mirror Mirror misalignment In plane mirror interferometry, the position of the motion platform is obtained by direct reference to the mirrors, which eliminate the influence of any mechanical inaccuracies in the motion system slideways. If mirrors are properly aligned, linear errors, which would normally result from mechanical deviations in straightness, orthogonality, pitch and yaw, are removed (assuming no Abbé errors are present). Mirror misalignment can cause three types of error: parallel, orthogonal and cosine. • Parallel error is caused when the mirror is not aligned parallel to stage travel (see Figure 2). Stage travel Interferometer Figure 4: Cosine error A measurement error, x, is produced when a mirror is misaligned such that it has an angular deviation, ø, from the perpendicular motion axis of length L (x = L(1 - cosine ø)). The following table shows the measurement errors possible as a result of cosine misalignment on a 300 mm axis. Angular deviation (arc sec) Measurement error* (nm) 5 0.09 10 0.4 20 1.4 30 3.2 40 5.6 50 8.8 Mirror misalignment errors can be overcome in the following ways: Mirror Figure 2: Parallel error • Orthogonality error is caused when the mirror axes are not truly orthogonal to each other (see Figure 3). Stage travel Stage travel Interferometer Figure 3: Orthogonality error These misalignments mean that as an axis is traversed the path difference between the cross-axis interferometer and the mirror changes. This change is interpreted as stage movement, which results in a measurement error. A measurement error, x, is produced when a mirror of length L is misaligned such that it has an angular deviation, ø, from the parallel axis, or, ø±90, from the orthogonal mirror (x = L sine ø). The following table shows the measurement errors possible as a result of parallel or orthogonality misalignment. • Carefully following a thorough alignment procedure, such as the one detailed in this document. • Using a mono-lithic L-mirror with a high orthogonality specification, instead of two stick mirrors. N.B. L-mirrors are expensive and vendors may be unwilling to quote below 1 arcsec (1 arcsec can create a measurement error of 5 µm/m). • Approximately aligning the system and then compensating for any error through a controller or compensation unit. It is important to note that the accurate alignment of stick mirrors, or the use of L-mirrors, can produce a mirror system with a superior orthogonality to that of the stage axes. Mirror mounting Mounting is critical if mirror movement or distortion, which results in measurement errors, is to be avoided. Mirrors can be mounted directly to the stage, but it is more useful for them to be secured via a mount that can assist in alignment. Mounts are commercially available for stick mirrors, plate mirrors and L-mirrors. Renishaw stick mirror mounts Renishaw supply three point kinematic mounts for mirrors of up to 350 mm in length and 25 mm square cross section, which allow fine pitch and yaw adjustment. The mounts are constructed to minimise errors due to differential expansion and have reliable positional locks that provide clamping to overcome acceleration, deceleration and bump forces. Adjustment sensitivity Yaw sensitivity* Pitch sensitivity 0.5 arcsec / degree of rotation < 1 arcsec / degree of rotation • Finely adjust the pitch of the Y-mirror to allow the laser to reflect back into the interferometer output (set to a suitable resolution), move the X-axis along full travel and align the mirror yaw for minimum count difference between end limits. 4. Engage the locking mechanism on the Y-mirror. 5. Align the X-axis mirror and interferometer parallel to the Yaxis by repeating steps 1, 2, 3 and 4. 3-point levelling stand True square X-mirror * for a 350 mm long mirror Y-mirror For more information regarding Renishaw mirror mounts, please contact a Renishaw representative or refer to M-9904-2179. Y-travel Autocollimator X-travel Figure 6: Set-up terminology Figure 5: Renishaw stick mirror mount Typical alignment procedure The following instructions demonstrate one method of aligning two plane mirrors to an X-Y stage*. It is assumed that the person performing the procedure is familiar with the test equipment, all mechanical adjustments of the motion stages have been completed, the RLD10 detector heads are fully mounted and the axes are fully operational. The pitch of the mirrors is now aligned and they are parallel to the axes of travel. However, the mirrors are aligned to the mechanical orthogonality of the X and Y-axes, which may not provide you with the ultimate accuracy. The mirrors’ orthogonality can be aligned to the precision of a true square, which can provide accuracy to better than 0.5 arcsec. Orthogonality alignment Suitable equipment and recommended specification: • Optical autocollimator - should have an aperture > 75 mm and a resolution of at least 0.1 arcsec. Suitable equipment and recommended specification: • Master true square - should have an orthogonality of less than 0.5 arcsec and be traceable to a calibration standard, e.g. Starett-Weber Corporation produce a true square that is traceable to National Institute of Standard and Technology (NIST). • Electronic indicator with a resolution of at least 0.1 µm • Levelling stand. 1. Install the Y-mirror in its mount. 1. Align the true square onto the levelling plate in the centre of the X and Y mirrors. • Mount the autocollimator to the system base so that its field of view is over both the Y-mirror and true square front surface. It is important that the centre line field of view coincides with the mirror / true square boundary. Ensure the collimator is stable, free from vibration and that the equipment has had adequate time to thermally adjust to the measuring environment before performing any adjustments. * If a vertical axis of motion is located under the mirrors a different method of pitch alignment may be necessary. 2. Align the Y-interferometer parallel to Y-axis travel (remove Y-interferometer cosine error): • Move the Y-axis (with the X-axis movement locked) to achieve the shortest separation between the Y-axis interferometer and the mirror. Attach an alignment target on the mirror surface so that the laser beam is on the target. Move the axis to the further position from the RLD10 and adjust the pitch and yaw of the RLD10 so the laser spot does not exhibit any translation from the target. N.B. this can be an iterative process and may require multiple near and far field adjustment for the best alignment. 3. Align the Y-axis mirror parallel to the X-axis travel: • Remove the target and visually align the Y-mirror so it is parallel to the X-axis travel. (Alternatively lock the Y-axis and position an electronic indicator on the front surface of the mirror outside the optical aperture to prevent surface damage. Move the X-axis along full travel and align the mirror yaw for a zero reading at the end limits. • Visually align the Y-mirror pitch so the mirror is perpendicular to the X-Y plane. • Block the front surface of the square and align the collimator to the Y-mirror to obtain zero pitch and yaw measurements (0.0 arcsec). • Block the front surface of the mirror and view the pitch and yaw of the true square. Adjust the true square until a zero reading on the autocollimator is achieved. Take care not to move the true square again as it is now the reference for all remaining mirror adjustments. 2. Align the X-axis mirror parallel to the perpendicular surface of the true square. • Move and mount the autocollimator to the perpendicular side of the machine, allowing its field of view to fall over the X-axis mirror and true square. Ensure the centre line of the Renishaw plc Old Town, Wotton-under-Edge, Gloucestershire GL12 7DW United Kingdom T +44 (0) 1453 844302 F +44 (0) 1453 844236 E laser.products@renishaw.com www.renishaw.com field of view coincides with the mirror / true square boundary. • Cover the X-mirror and align the collimator to get a zero reading (0.0 arcsec) from the front surface of the true square. DO NOT move the true square. • Block the true square and, viewing the X-mirror through the collimator, adjust the yaw of the mirror for a zero reading. Identification and ordering Plane mirrors Renishaw supply plane stick mirrors up to 300 mm in length in multiples of 10, which meet the following specification (DS/9904/1492/01A): λ/10 over 100 mm < 0.5 µm over 500 mm Zerodur™ Hard oxide dielectric L* (+ ~ 20 mm) × 25 × 25 3. If a locking function is provided on the mirror mounts*, carefully tighten while reading the autocollimator to ensure the mirror has not moved. Local flatness Global flatness Substrate Mirror coating Dimensions * some locking mechanisms may have a slight effect on the mirror * L = optical aperture (working length) alignment and must be considered when performing adjustments. Upon completion of the above steps it is highly recommended to bring the collimator back to the Y-axis mirror set-up and perform the measurements again to verify nothing has changed to affect the orthogonality. Error mapping As mentioned earlier in this document, orthogonality errors and macro-mirror distortion can sometimes be compensated for using the motion controller. • Compensating for mirror distortion A typical method involves locking one axis in place and, using the stage interferometer and mirror, monitoring any cross axis displacement as the axis is traversed. At the same time, the stage motion is monitored independently of the mirror, for way straightness deviations. If these two readings are subtracted from each other any mechanical variations will be isolated from the mirror deformation, allowing the mirror’s distortion profile to be calculated. Way straightness can be monitored down to 0.1 µm using a calibration laser (or RLD10 with no internal optics (RLD10A3-XX)) and straightness optics (see Figure 7). If higher precision mapping is required, a master should be used. For a more detailed specification, please contact a Renishaw representative. Stick mirrors may be ordered directly from Renishaw using the part numbering protocol shown below. Y = modified for Renishaw mounts N = not modified for Renishaw mounts Part number RPM10 - Y X X X X If your application demands a higher specification, please contact a Renishaw representative. Mirror mounts Renishaw supply adjustable mirror mounts for stick mirrors of up to 350 mm total length and 25 mm square cross section. (For a more detailed specification, please contact a Renishaw representative.) Stick mirror mounts may be ordered directly from Renishaw using the part number below. S = stick mirror mount Part number Laser source Optical aperture (mm) (< 300 mm, multiples of 10 only) RAM10 - S X - 0 1 Mirrors used in the Renishaw mounts must have location features added to the base, so, if using mirrors not supplied by Renishaw, it is advisable to contact a Renishaw representative for further advice. Stage travel Interferometer Figure 7: Monitoring way straightness • Compensating for mirror distortion and orthogonality A typical method maps the system to a master on the stage, e.g. a four-point standard such as a wafer calibrated with fuducials. The master, in conjunction with a vision system, can be used to obtain the constants for a crossaxis algorithm that the motion controller can then employ to perform compensation. Renishaw reserves the right to change specifications without notice. © 2005 Renishaw plc. All rights reserved. For worldwide contact details, please visit our main website at www.renishaw.com/contact Printed in England 0805 Part No. L-9904-2390-01