Thomas Bergauer HEPHY Vienna CMS Week, 25.9.2001, CERN ¡ Setup l l l ¡ Results l l ¡ Schematic Devices Picture Temp. & Humidity current Conclusion Thomas Bergauer, HEPHY Vienna 2 -400V R V SMU BIAS Backplane (1 times) GND (10 times) Thomas Bergauer, HEPHY Vienna 3 ¡ ¡ SMU Keithley 2410 Keithley 2700 Multimeter l ¡ ¡ ¡ with K7702 40-channel multiplexer card TRHX Temperature and Humidity Readout Box with 10 resistors Sensor Box (Strasbourg) l Apple Mac Serial TRHX Sensor 1 Sensor 2 GPIB K2410 HV K2700 10 R Box 10 Sensor Box Connection to 10 sensors with wirebonding Thomas Bergauer, HEPHY Vienna 4 Sensor Box (filled with Nitrogen) Instruments PowerMac G3 Resistor Box Thomas Bergauer, HEPHY Vienna 5 ¡ ¡ ¡ ¡ Software: Labview 2 sensors f. temp. & humid. Alarm if humidity exceeds limit (30%) Current compliance l l ¡ Hardware: 40 uA (all sensors) Software: 20 uA (each sensor) Single output file for each sensor Thomas Bergauer, HEPHY Vienna 6 35 30 25 20 humidity [%] temperature [C] 15 10 5 0 0 20 40 60 80 100 120 time [h] 4,5 days Thomas Bergauer, HEPHY Vienna 7 0,35 0,3 current [uA] 0,25 W6A-60532-30 W6A-60532-36 0,2 W6A-60532-37 W6B-60533-18 W6B-60533-25 0,15 W6B-60533-26 W6B-60533-27 W6B-60533-28 0,1 0,05 0 0 20 40 60 time [h] Thomas Bergauer, HEPHY Vienna 80 100 @400V 4 days 8 ¡ ¡ Measured 75 HPK sensors @ 400V bias Every sensor had very constant current for at least 96 hours l ¡ ¡ (one sensor has to be verified) Our setup works very well HPK sensors have no problem with longtime behavior!! Thomas Bergauer, HEPHY Vienna 9