TSU12-20 - Optronis

advertisement
TSU12-20
Triggered Sweep Unit
Sweep unit to operate the OPTOSCOPE SC-20
streak camera.
> Temporal resolution
< 600 ps
> Speed 10 ns/mm .. 5 µs/mm
> Slow sweep option /S1
> Fast sweep option /F2
> Single-shot and repetitive mode
> Automatic Configuration
TSU12-20
Triggered Sweep Unit
Description
t
t
TSU12-20
t
t
The TSU12-20 sweep unit drives the deflection plates of the Optoscope SC-20 streak camera. After installation the sweep unit is automatically
identified and configured. Different operating modes (single-shot, repetitive, focus) mode can be selected. In single-shot mode the readout
camera is triggered and only one sweep per image is enabled. Combined with jitter correction techniques, signals with large jitter can be
measured without sacrifying temporal resolution.
trigger delay
t
TSU12-20/S1 (slow sweep option)
Sweep speeds
100 ns/mm
250 ns/mm
500 ns/mm
1 µs/mm
2.5 µs/mm
5 µs/mm
10 µs/mm
25 µs/mm
Time base *)
3 µs
7.5 µs
15 µs
30 µs
75 µs
150 µs
300 µs
750 µs
Trigger delay (typ.)
5 µs
10 µs
25 µs
50 µs
100 µs
200 µs
400 µs
700 µs
Specifications
Temporal system resolution
Sweep speed non-linearity
Sweep sweep accuracy
Trigger jitter
Trigger Freqeuncy
Trigger input signal
Trigger output signal
Dimensions
Environment
TSU12-20/F2 (fast sweep option)
Sweep speeds
<4 ns/mm
5 ns/mm
10 ns/mm
25 ns/mm
50 ns/mm
100 ns/mm
Time base *)
<120 ns
150 ns
300 ns
750 ns
1.5 µs
3 µs
Trigger delay (typ.)
190 ns
220 ns
350 ns
700 ns
1.4 µs
2.6 µs
< 600 ps system performance @ <4ns/mm
± 2.5 % (typ.) / ± 5 % (max.) over 30 mm on phosphor screen
±5%
corresponding to < 1 % of time base
single-shot and repetitive mode up to 10 Hz
>+3V, ± 7V max., < 10 ns rise time, > 50 ns duration, 50 Ohm, BNC connector
+ 2 V, 1.5 µs duration (typ.), 50 Ohm, BNC connector
300 (W) x 240 (H) x 60 (D outside main unit) mm / Weight 3.3 kg (typ.)
5 .. 35 °C / 20..80% rel. humidity, non-condensing
300
240
60
trigger output
measured
profile
trigger input
readout image
optical signal
*) at 30 mm readout
linear range
region of interest
Trigger delay (typ.)
800 ns
1.2 µs
2.5 µs
6.5 µs
13 µs
25 µs
65 µs
125 µs
250 µs
Time base *)
300 ns
750 ns
1.5 µs
3 µs
7,5 µs
15 µs
30 µs
75 µs
150 µs
sweep signal
Sweep speeds
10 ns/mm
25 ns/mm
50 ns/mm
100 ns/mm
250 ns/mm
500 ns/mm
1 µs/mm
2.5 µs/mm
5 µs/mm
OPTOSCOPE
All Dimensions in mm
The information given herein is believed to be reliable, however Optronis makes no warranties as to its accuracy or completeness. Data sheet is subject to modifications at any time. 05/2014
Optronis GmbH
Ludwigstr. 2
77694 Kehl, Germany
Telefon
Telefax
+49 (0) 78 51 / 91 26 -0
+49 (0) 78 51 / 91 26 -10
E-Mail: info@optronis.com
www.optronis.com
Download