electromagnetic damping

advertisement
ELECTROMAGNETIC DAMPING
© Institute of Lifelong Learning, University of Delhi
1
ELECTROMAGNETIC DAMPING
2
PHYSICS (LAB MANUAL)
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Introduction
In this unit, we are going to introduce you to an electrical instrument called
galvanometer. Galvanometers are instruments that are intended primarily to
indicate the existence of a current – steady or transient, in a circuit. These may
be capable of measuring the current, only under certain circumstances, when
calibrated.
The construction of galvanometers is based on the interaction between coils
carrying current and magnets. These may be, in general, divided into two types
as given below according to as the moving part is a magnet or a coil.
Moving magnet galvanometer: In this type of galvanometer, the current is
passed through a fixed coil, which produces a magnetic field under which a
magnet moves.
Moving coil galvanometer: In this galvanometer the current is passed through
a movable coil placed between the poles of a powerful magnet. The coil swings
under the action of a deflecting couple acting upon it. You will learn in detail
about such galvanometer and be able to explore many interesting features of its
construction and functioning.
Moving coil galvanometers can be further classified as:
Weston galvanometer or pivoted-coil galvanometer
In this galvanometer, the coil swings between two pivots and the deflection of the
coil is read by a pointer that moves on a circular scale. This type of galvanometer
is less sensitive but more convenient. These are extensively used in various
bridge-type circuits working with direct current, for obtaining null-point. You
might have seen or worked with such a galvanometer, in particular, while
performing an experiment with Carey Foster‟s Bridge.
D’ Arsonval galvanometer or suspended-coil galvanometer
This galvanometer consists of a coil which is suspended freely and the deflection
of the coil is read using a mirror attached to it. It shows very distinct and
interesting behaviour as a dead-beat galvanometer, a critically damped
galvanometer and a ballistic galvanometer, under different circuit conditions. The
behaviour of such galvanometers is elaborated in the “Theory” section.
Let us now look into the details of construction of such galvanometers. Figure 1
shows the sketch of D‟ Arsonval galvanometer or suspended-coil galvanometer.
It consists of a narrow rectangular coil C of many turns of fine insulated copper
wire suspended so as to move freely in a narrow annular space between the polepieces of a powerful horse-shoe magnet NS. The pole-pieces of the magnet are
hollowed out to make them cylindrically concave in shape. This makes the
magnetic field radial, i.e. the field is always parallel to the plane of the coil. A
cylinder of soft iron is fixed within the coil, which serves to concentrate the lines
of force in the gap, thus making the field in the gap strong and practically
uniform. The coil is wound on a light frame-work which is of metal (generally
aluminum) or of non-conducting material (bamboo, ivory or ebonite) according to
as a dead-beat or a ballistic action is required.
© Institute of Lifelong Learning, University of Delhi
3
ELECTROMAGNETIC DAMPING
PHYSICS (LAB MANUAL)
To binding
screw
To binding
screw
Figure 1: Construction of suspended-coil galvanometer
The suspension is a fine rectangular strip of phosphor-bronze, the upper end of
which is attached to a rigid support R (also known as movable torsion head). It
also serves as one of the current leads to the coil. The rectangular strip has very
small torsional rigidity and hence the twist for a given couple is much greater
and the stress on the material much less than with a wire of circular crosssection. It also offers a comparatively larger area for the dissipation of heat
produced by the current. Connection to the other end of the coil is made by
means of a very light metal spring K attached to the bottom of the coil. A small
concave mirror M is attached to the top of the coil. This mirror generally has a
radius of curvature equal to one meter. The whole arrangement is enclosed in a
non-metallic case which is provided with a glass window in the front and leveling
screws at the base (Figure 2). A spirit level is also generally provided at the
bottom rim to ensure the vertical orientation of the coil. The torsion head is
connected to a terminal T1 of the galvanometer and the spring K is connected to
the other terminal T2, through which the galvanometer can be connected in the
circuit. A lock/unlock knob for the suspension is provided at the back of the
galvanometer. A lamp and scale arrangement (Figure 3) is used in front of the
galvanometer mirror to record the deflection of the coil of the galvanometer.
4
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Torsion-head
Mirror
Coil
T1
Binding screws
T2
}
Spirit level
} Leveling screws
Figure 2: Moving coil ballistic galvanometer
The suspended-coil type moving coil galvanometer which you are supposed to
use in the laboratory, is known as the ‘ballistic galvanometer’ (B.G.). It is a
specially designed moving coil galvanometer for measuring the total quantity of
electricity or charge that passes through it due to a transient current or voltage
of short duration, such as during charging and discharging of capacitor and in
electromagnetic induction. When such a transient current passes through the
galvanometer, its coil experiences an angular impulse and begins to perform
angular oscillations about its axis. The maximum deflection (or throw) obtained is
proportional to the total charge that passes through the galvanometer. As you
will see in the theory, it can also work as a critically damped or a dead-beat
galvanometer depending upon the circuit parameters.
© Institute of Lifelong Learning, University of Delhi
5
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
S
M
L
L-Lamp
S-Scale
M-Mirror
Figure 3: Lamp and Scale Arrangement
For the purpose of making a moving coil galvanometer ballistic, i.e., to measure
charge that passes for a short duration, its coil is required to get deflected under
the action of an impulse imparted by a momentary current. Hence, the coil has to
be light-weight.
Further, motion of the coil should not be damped. The coil is therefore wound on
a light non-metallic frame made of bamboo or ebonite, to eliminate the damping
effect of eddy currents in the metallic frame.
In order to measure the total charge, the whole transient current should have
passed through it before the coil moves appreciably from its position of rest.
Hence, the time period of oscillation of the coil should be fairly large. This is
achieved by making  (moment of inertia of the coil) large and c (torsional
constant, i.e., torsional couple per unit twist of the suspension fiber) small (as
evident from Equation (4)). For this reason, the suspension is made of thin
rectangular strip of phosphor-bronze whose torsional constant is small.
As you have gathered a brief idea about the main instrument, ballistic
galvanometer, we will now discuss in detail, the objective, working circuits and
the theory underlying this experiment.
6
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
K1
R'
BG
E
K'
Figure 4: Circuit diagram - A
Figure 5: Circuit Diagram - B
© Institute of Lifelong Learning, University of Delhi
7
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Apparatus









Ballistic Galvanometer (B.G.)
Tapping keys (K1 and K`)
Plug Key (K)
Resistance boxes with resistances upto10ΚΩ (R`, R1 and R2)
Resistance box with resistances upto 100Ω (r)
Lead accumulator (~2 volts) (E)
Lamp and scale arrangement
Stop watch
Connecting wires
Theory
With reference to Figure 1, let N be the number of turns of the coil, A, the area of
the coil and B, the magnitude of uniform magnetic induction as produced by the
cylindrical pole-pieces of the horse-shoe magnet.
The passage of charge through the galvanometer is equivalent to the passage of
a varying current for short interval of time. Let i be the current in the coil at any
instant. The coil experiences a couple due to the horizontal and oppositely
directed equal forces acting on the vertical sides. The moment of this couple or
the torque acting on the coil is given by
(1)
  NiBA .
The torque acting for an infinitesimal time dt, gives the coil an angular impulse
which is (torque × time) equal to NiBA dt . Therefore, if t be the total time for
which the current flows through the coil, the total angular impulse given to the
coil is
t
  NBAi dt  NBAq
0
t
since
 i dt  q ,
the charge that has passed through the galvanometer. This
0
impulse produces angular momentum in the coil which rotates until the restoring
couple of the suspension brings it to rest. The coil then swings back to its rest
position as the suspension unwinds, and due to its inertia, over shoots and twists
the suspension in the opposite direction. A series of back and forth angular
oscillations thus results. The amplitude of oscillation keeps decreasing with time
due to damping forces present.
Let us assume that the coil begins to oscillate after the impulse is over. Let ω be
the angular velocity at start and  its moment of inertia about the axis of
suspension. Then the angular momentum produced in the coil due to the angular
impulse is  , i.e.
  NBAq
(2)
Now, the coil possesses a kinetic energy
1 2
 at start. If damping is absent,
2
then this energy is entirely used for doing work in twisting the suspension. If c be
the restoring couple per unit twist in the suspension, then the couple for a twist θ
8
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
is cθ. The work done for an additional twist dθ is cθdθ. If the maximum twist is
θ0, then the work required is
0
  c .d 
0
1 2
c 0
2
Equating it to the kinetic energy of the coil at start, we get
1 2 1 2
  c 0
2
2
2
  c 02
or
(3)
Now, if T be the time period of oscillation of the coil, we have
T  2

or

c
(4)
T 2c
4 2
Multiplying Equation (3) and Equation (4), we get
T 2 c 2 02
4 2
Tc 0
 
2
 2 2 
or
(5)
Comparing Equation (2) and Equation (5), we get
Tc 0
2
T
c
q
.
. 0
2 NBA
q  k. 0 ,
NBAq 
or
or
where
k
constant’.
(6)
(7)
T
c
is a constant for the galvanometer, known as the ‘ballistic
.
2 NBA
It is important to rewrite Equation (7) as
Qs 
q
0
k
T
c
.
2 NBA
where Qs is known as the „charge sensitivity‟ of the ballistic galvanometer.
Thus, charge sensitivity is the same as the ballistic constant.
Hence, Equation (7) clearly shows that the first throw of a ballistic galvanometer
is a measure of the charge that has passed through it.
Equation of Motion of the moving system in Ballistic Galvanometer
The various couples acting on the moving system are as follows:
(a)
(b)
The deflecting couple,
short time τ.
NiBA due to the transient current i flowing for a
d 2
The restoring couple, 
due to the inertia of the coil, where  is the
dt 2
moment of inertia of the coil.
© Institute of Lifelong Learning, University of Delhi
9
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
(c)
The retarding couple due to damping has two components (i) due to air
resistance and (ii) due to electromagnetic damping. The component due to
air resistance is proportional to angular velocity and can be taken
b
d
,
dt
where b is a constant. The component due to electromagnetic damping
can be calculated as follows:
The e.m.f. induced in the coil,
d
d
,
G
dt
dt
where G  NBA . If R is the resistance of the galvanometer circuit, then
e  NBA
the current due to this induced e.m.f. is
I' 
e G d
 .
R R dt
Couple on the coil due to this induced current is
( NBA) I '  NBA.
G d G 2 d
.

.
R dt
R dt
The total retarding couple is then

G2
 b 
R

(d)
 d
d
.

dt
 dt
The torsional couple, cθ, due to the torsional twist of the suspension fiber.
Here, c is the torsional couple per unit twist of the suspension fiber.
The equation of motion may be obtained using the condition that under
equilibrium, the deflecting couple is equal to the sum of the restoring couple,
retarding couple and torsional couple. Hence,

d 2
d

 c  NBAi
2
dt
dt
(8)
We must consider the equation for two intervals: 0 < t < τ and τ < t < ∞. During
the first of these intervals, the coil acquires momentum due to impulse given to it
by transient current. It, however, is hardly deflected from its initial position
because of the stipulation that the transient current flows during a very small
time τ in comparison with the free period T of the instrument, i.e., τ / T << 1.
Considering the motion of the coil for t >> τ; now i = 0, so that Equation (8)
takes the form
d 2
d

 c  0
2
dt
dt
d 2
d
 2 p.
 q 2  0 ,
2
dt
dt
 1  ( NBA) 2 
2 p   b 

 
R 

or
where
and
q2 
c

(9)
(10)
(11)
Equation (9) represents a damped harmonic motion. We can solve this equation
by assuming a solution of the form
  Ae t
where the parameter α is, in general, a complex number. Substituting this
assumed solution back into the differential Equation (9), we get
 2  2 p  q 2  0
10
© Institute of Lifelong Learning, University of Delhi
(12)
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
This quadratic equation has two roots
   p  p2  q2
(13)
The complete solution for θ is, therefore
  p  p 2  q 2  t


1
2
The values of A1 and A2 can be calculated from the
 A e
Ae
  p  p 2  q 2  t


(14)
initial conditions, i.e., θ = 0
d
and
 0 at t = 0. The exact physical nature of this solution depends on the
dt
relative values of p and q. In particular, the qualitative behaviour of the system
depends on whether the quadratic Equation (12) for α has one real solution, two
real solutions or two complex conjugate solutions.
Conditions when a Moving Coil Galvanometer is Dead-beat or Ballistic
Case I: Over damping or Dead-beat
When
p 2  q 2 or p  q , i.e.,
2

1 
NBA  
b 

2 
R 
R
or
c

NBA2
4c  b
R  RCD ,
or
where, the critical damping resistance,
RCD 
In this case,
NBA2
4c  b
(15)
p 2  q 2 is less than p and, therefore, powers of exponential in
both the terms in Equation (14) will be real, say -α1 and -α2 and Equation (14)
will now take the form,
 A1e 1t  A2 e  2t
(16)

© Institute of Lifelong Learning, University of Delhi
11
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Both the terms are exponentially decreasing functions of time. Hence, the
deflection θ will go on decreasing with time and the motion will be dead-beat or
non-oscillatory (Figure 6).
A
Critical
damping -pt
Ae
Over damping
Under damping
0
time
(t)
A
Critical damping
Ae-pt
Over damping
Under damping
Figure 6: Different
cases of damping
0
t
Case II: Critical damping
A critically damped system converges to zero faster than any other, without
oscillating. This condition is achieved when p = q; so that α has only one real
value (α = -p) and the solution becomes
  Ae  pt
The damping of a galvanometer can be varied readily by altering the total
resistance R of the closed circuit in which it is used (Refer to Equation (10) where
all other parameters except R are constant for a given galvanometer).
Critical damping is obtained by choosing R so that, when the galvanometer is
deflected, the spot of light or pointer comes to rest in the minimum time without
over-swing. The motion of the coil is just non-oscillatory and the frequency of
oscillation is just zero. This is achieved when
R  RCD 
NBA2
4c  b
Since all the terms which appear on the right hand side of the above equation are
design constants of the suspension and the coil and
R  R1  RG ,
(17)
where R1 is the external circuit resistance and RG, the galvanometer resistance.
Critical damping for a particular galvanometer can be obtained by varying R1 (see
Figures 5 and 6).
12
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Case III: Under- damping or ballistic
In this case, p < q.
Thus
p 2  q 2 is complex and we can write it as jω, where
j   1 and   q 2  p 2
(18)
Then Equation (14) can be written as
  A1e  p j t  A2 e  p j t

 e  pt A1e jt  A2 e  jt

θ being a real quantity, A1 and A2 must be complex conjugates so that the
solution for θ can be written as
  Ae  pt cos(t   )
(19)
The constants A and β are determined from the initial conditions, i.e., θ = 0 and
d
  0 at t = 0. Applying these conditions to Equation (19), we get
dt
A cos   0,
 pA cos   A sin    0
 A sin   0
or
(20)
cos   0 since A  0

 3
3
or   ,
and A  0 .
, . Also, because sin β is a negative quantity,  

2 2
2
From first condition of Equation (20), we find that
Hence, Equation (19) becomes


  Ae  pt cos t 
3 

2 
 Ae  pt sin t

(21)
0  pt
e sin t

Equation (21) shows that the behaviour of the system is now damped harmonic
and the system will oscillate at the frequency ω with the exponentially decreasing
amplitude given by
Ae  pt . The time period T of these oscillations is given by
T
2


q
2
2
p
2

1
(22)
2
We have found in this special case that for the galvanometer to be
ballistic, p should be as small as possible. For this, according to Equation
(10):
(a)
The moment of inertia of the coil should be large.
(b)
The air damping should be small.
(c)
The electromagnetic damping should be small. Thus, the coil should be
wound on a non-conducting frame.
(d)
The resistance R of the circuit, including that of the coil should be large.
For this, the coil should have large number of turns.
© Institute of Lifelong Learning, University of Delhi
13
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
[Note: At this stage, it should be clear to you that for the galvanometer to be
dead-beat, just the opposite of the above mentioned conditions need to be
satisfied.]
In the ideal case, p should tend to zero. Under this condition, the second or the
middle term in Equation (9) will be absent and motion will be undamped and
simple harmonic with constant amplitude A and angular frequency ω = q (as
given by Equation (11)). In actual practice, this case can be treated as equivalent
to the working condition when the coil oscillates freely with R = ∞. This will be
done in the laboratory, under the study of oscillatory behaviour of the B.G., using
circuit diagram – A.
Correction for Damping in Ballistic Galvanometer
In deducing the relation between charge q and ballistic throw
0
(Equation (6)),
it has been assumed that the damping is entirely absent and whole of the kinetic
energy of the coil is used for twisting the suspension through an angle  0 .
Actually, it is not so. The motion of the coil is damped due to viscosity of air and
also due to opposing current induced in the coil and frame of the coil, which
rotates in the field of permanent magnet of the galvanometer. Therefore, the coil
oscillates with decreasing amplitude. Hence, the observed throw is smaller than
its true value  0 , which would have been observed if the damping were entirely
absent. A correction is therefore required.
Let
1 ,  2 ,  3 , 
be the successive throws observed at the end of first, second,
third, … swings of the coil.
and
1 ,  3 , 
are on one side of the rest position of the coil
 2 , 4 , on the other.
For Equation (21), the amplitude of motion is
amplitude  0
For
For
3T
3
t

4 2 q2  p2
t
and for t = 0, we have
 A.
T

t 
4 2 q2  p2
For
Ae  pt
5T
5

4
2 q2  p2
 p
, we have amplitude
q2  p2
1  Ae 2
3 p
 2  Ae
, we have amplitude
2 q2  p2
5 p
 3  Ae 2
, we have amplitude
q2  p2
and so on.
 p
 p
1  2  3
q p


 e
 e   e  d
2 3 4
p
T
  log e d 
 p.

2
2
Therefore,
or
14
2
© Institute of Lifelong Learning, University of Delhi
(23)
(24)
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Here, it is seen that the ratio of successive amplitudes 1 , 2 , 3 , , etc. is
constant. This constant ratio d is called decrement and
log e d  
is called
the logarithmic decrement.
We have seen that the observed throw would have been
0  A
if there were no
damping. The first throw θ1 is observed a quarter period (i.e., T/4) later and
during all this time, the damping effect has taken place, so that
 p
1  Ae
 Ae
2 q2  p2

2
Hence,
0  2
 1 
 e 1 
  
1
2 1 2  2 

2
 1
2
(approximately)
neglecting the higher powers of λ, which are very small in the case of ballistic
galvanometer.
Therefore,


 0   1 1 


2
(25)
Equation (6) can now be written as
q
cT
 
.1 1  
2 NBA 
2
(26)
Experimental determination of the logarithmic decrement
The ballistic galvanometer coil is made to oscillate by passing a discharge through
it and the successive amplitudes on the right and left are observed. Then


1  2

   9  10  d  e  .
2 3
 10 11
Multiplying all these factors, we have
1
 e10
11

or


1
2.3026
log e 1 
log 10 1
10
11
10
11
Alternatively, if throws on the same side are observed, i.e., if
observed, then
or
(27)
1 , 3 , 5 , are
1 1  2
 
 e 2  d 2
3 2 3


1
2.3026
  log e 1 
log 10 1
2
3
2
3
© Institute of Lifelong Learning, University of Delhi
15
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
As
d2 
1  3  5


 ,
3 5 7
d2 
1   3   5  
3  5  7  
we can write
Hence
      
1
  log e  1 3 5

2
 3   5   7   
(28)
Thus, all the observed values of θn, or as many observed values of θn can be used
to obtain the corresponding λ.
Q-factor
The efficiency of an oscillator is measured in terms of a quantity called Quality
factor (Q-factor). It is a dimensionless parameter that compares the time
constant for decay of the amplitude of an oscillating physical system to its
oscillation period. Equivalently, it compares the frequency at which a system
oscillates to the rate at which it dissipates its energy.
It is defined as 2 times the ratio of energy stored to the energy dissipated per
cycle, i.e.
Q  2 
Energy stored
Energy dissipated per cycle
Energy stored
Power loss  T
Energy stored

,
Power loss
(29)
 2 
(30)
where T is the time period of oscillation and ω is the angular frequency.
A higher Q indicates a lower rate of energy dissipation relative to the oscillation
frequency, so the oscillations die out more slowly.
Calculation of Q
The energy of a damped oscillator is spent in overcoming the frictional forces and
hence, decreases continuously with time. Consequently, the amplitude of
oscillation decreases. The kinetic energy of the oscillator, at any time t, is given
by
1  d 
K .E.  

2  dt 
2
(31)
where
  Ae  pt sin t
16
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Hence, the time average value of kinetic energy per cycle is given by
T
1
K .E.
T 0
 K .E. 
1 2 2 2 pt
 A e
4

(32)
where we have assumed low damping so that the time-dependent amplitude
A(t )  Ae  pt is regarded as constant over one cycle; hence taken out from
integral.
Similarly, the average value of potential energy is given by
1
 U   2 A 2 e 2 pt
4
(33)
(since, the average values of kinetic energy and potential energy for a harmonic
oscillator are the same).
Average total energy is, therefore,
 E 

1 2
 Ae  pt
2

2
(34)
The average power dissipation is
d
E
dt
  2 A2 pe 2 pt
 2p  E 
 P  
Therefore,
(35)
E
P
E

2p  E 
Q 



2p

(36)
pT
This expression can be expressed in terms of the logarithmic decrement λ using
Equation (24) as
Q
Also we may designate 2 p

2
(37)
  , i.e., resonance band width. So that
Q

2p



© Institute of Lifelong Learning, University of Delhi
(38)
17
ELECTROMAGNETIC DAMPING
PHYSICS (LAB MANUAL)
Learning Outcomes
While performing this experiment, you will be able to
1.
2.
3.
4.
5.
6.
7.
Observe the oscillatory nature of motion of the coil of the galvanometer
under minimum damping.
Determine the time period of oscillations (T) of the galvanometer coil.
Convert the oscillatory motion into dead-beat one.
Obtain the galvanometer resistance (RG).
Obtain the logarithmic decrement (λ) and Quality Factor (Q) for
various stages of damping.
Investigate the dependence of logarithmic decrement (λ) and Quality Factor
(Q) on the galvanometer circuit resistance (R).
Find out the Critical Damping Resistance (C.D.R.) of the galvanometer.
Pre-lab Assessment
Now to know whether you are ready to carry out the experiment in the
lab, answer the following questions. If you score at least 80%, you are
ready, otherwise read the preceding text again. (Answers are given at
the end of this experiment.)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
How does a ballistic galvanometer differ from a dead-beat one?
What are the different types of moving coil galvanometers?
What do you understand by logarithmic decrement?
Why and how do you determine logarithmic decrement?
What is electromagnetic damping and how can it be reduced in a B.G.?
How can the time-period of the suspended coil be increased?
How are the unwanted oscillations of the suspension of B.G. stopped?
How can you increase the sensitivity of the B.G.?
What is meant by „ballistic constant‟?
How can the couple per unit twist of the suspension fiber be kept low?
What do you understand by E.M.F. of a cell?
Why secondary cells are generally preferred over primary cells?
Choose the correct answer:
(1)
(2)
18
The coil of a suspended coil galvanometer has very high resistance. When a
momentary current is passed through the coil, it is expected that the coil
(a)
oscillates with decreasing amplitude
(b)
oscillates with the same amplitude
(c)
gets deflected and comes to rest slowly
(d)
shows steady deflection.
What does a ballistic galvanometer do?
(a) It measures current.
(b) It measures voltage.
(c) It measures charge.
(d) It measures magnetic field.
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
(3)
ELECTROMAGNETIC DAMPING
In a moving coil dead-beat galvanometer, the coil is wound on an aluminum
frame because
(a) aluminum is a heavy metal.
(b) no eddy currents are developed in aluminum.
(c)
on stopping the current in the galvanometer, the coil comes to rest
rapidly.
(d) it increases the sensitivity of the galvanometer.
(4)
To make the field radial between the pole pieces of a magnet
(a) the cylindrical (concave) pole pieces are used.
(b) the straight-cut pole pieces can be used.
(c) steel core should be used.
(d) laminated core should be used.
(5) The pointer of a dead-beat galvanometer gives a steady deflection because
(a) its magnet is very strong.
(b) its pointer is very light.
(c) eddy currents are produced in its coil .
(d) the coil is wound on an ebonite frame.
Answer in yes/no:
(1)
(2)
(3)
(4)
(5)
(6)
A moving coil galvanometer gets affected by a magnet placed close by.
A dead-beat galvanometer can be used to measure charge/transient
current.
A soft iron core is used for the coil of a B.G.
A lead accumulator can be recharged.
The coil of a dead-beat galvanometer can be wound over a non-conducting
frame.
A large value of Q (quality factor) signifies under-damped oscillations.
Procedure
1. Release the lock/unlock knob for the suspension/mirror provided at the back
of the B.G, so that the suspension fiber and hence, the mirror attached to it
are free to move.
2. Level the B.G. with the help of the three leveling screws provided at its base
and the spirit level provided on its bottom rim. (This is done to ensure that
the coil hangs freely and its suspension is vertical).
3. Arrange the lamp and scale arrangement in front of the B.G. at a distance of
1m from the mirror of B.G. so that light from the lamp falls on the mirror of
B.G. The cross-wire in the reflected spot of light is adjusted at the zero
(middle) of the scale.
Part – A
(For study of oscillatory motion of the coil)
4. Connect the circuit according to the circuit diagram - A (Figure 4). The two
terminals of the B.G. should be connected to a tapping key (K 1), through
which the other circuit connections should be made. (This key is used to
arrest the undue motion of the coil).
© Institute of Lifelong Learning, University of Delhi
19
ELECTROMAGNETIC DAMPING
PHYSICS (LAB MANUAL)
5. Take out a high resistance of the order of 10 KΩ from the resistance box R'.
(This should be done to safeguard that a large current does not flow through
the galvanometer coil that may damage it).
6. Press the tapping key K' momentarily and release. (This means that the
circuit is now open and the circuit resistance R' is infinite)
7. The coil will start oscillating about the mean position.
8. Note down n consecutive deflections, θn, on the scale, i.e. on both the sides of
zero, until the deflection reduces to zero and record the observations in Table
1.
9. Again press the tapping key K' momentarily and release. As the coil starts
oscillating, note the time taken by the spot of light to complete 20 oscillations,
using a stop-watch. (In order to time oscillations accurately, make sure that
you start your stop watch as soon as the spot crosses the zero position and
not when it is at either of the extreme positions.)
10. Repeat the above step 4-5 times and record these observations in Table 2, for
determination of time-period of the B.G.
Part – B
(To Study the effect of circuit resistance on damping)
11. Now reassemble the circuit according to Figure 5 (circuit diagram – B).
12. Keep R1 = 1 ΚΩ, R2 = 10 KΩ and adjust r to get full-scale deflection (a
deflection of about 25cm on the scale) on plugging in the key K.
13. Open the key K now, the spot would come back slowly and if it does not
overshoot the zero point on the scale, you are ready to start taking
observations. If not, i.e., if the spot overshoots the zero mark slightly, then
reduce R1 in steps of 100 Ω and again adjust r for full-scale deflection until
the spot does not cross the zero point on opening the key K. This corresponds
closely to critical damping. Record this value of R 1 as the initial observation in
Table 3.
14. Now R2 and r are to be kept constant throughout the experiment and
observations will be recorded by varying R1 only.
15. Plug in the key K. The spot of light will move to show a full-scale deflection as
already adjusted. Note this deflection as θ1 (cm). As soon as it becomes
stable, take out plug key K. Confirm that θ2 is zero. Clearly, θ3 is also zero.
Record these observations in Table 3.
16. Now, increase R1 in steps of 1 ΚΩ. Plug in the key K. The spot of light will now
show a smaller deflection (as the resistance in the circuit has been increased).
Note the first deflection θ1. Remove the plug from key K. Note the consecutive
deflections θ3, θ5 ….on the same side, till the last deflection, closest to zero.
17. Repeat step 16 by increasing R1 in steps of 1ΚΩ each time, up to about 15 ΚΩ
or 20 ΚΩ.
You have observed the cases of oscillatory motion, damped oscillation and over
damped behaviour of B.G. Now, to find the exact value of the critical damping
resistance of the B.G., you are required to proceed as follows and record your
observations in Table 4.
Part – C (For determination of critical damping resistance)
18. Start taking observations from that value of R 1 in step 17, where θ3 was very
small.
19. You should reduce R1 now in smaller steps and note θ1 and θ2 also (θ2 is the
next deflection on the other side of the zero after the first deflection θ 1).
20
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
20. Here, you are also required to note the time taken by the spot of light to
come back to zero from θ1 (i.e. start the stop watch as soon as you plug out K
and stop it as soon as the spot of light reaches the zero mark).
21. Repeat steps 19 and 20 by reducing R1in small steps.
22. You will eventually reach a step when θ2 will just be equal to zero.
23. You should stop taking further observations when θ 2 remains zero but the
time taken by the spot to come back to zero starts increasing.
24. That particular observation when θ2 is zero and the time taken by the spot of
light to come back to zero is minimum corresponds to the condition of critical
damping.
Observations
Least Count of the scale used in the lamp and scale arrangement = ……. cm
E.M.F. of the accumulator = …… V
Table 1: Oscillatory behaviour of B.G.
S. No.
1
2
3
4
5
…
…
Number of thrown
1
2
3
4
5
…
…
Throw θn (cm)
~ 20 or 30
Table 2: Time period of B.G.
S.No.
Time for 20 oscillations
(sec)
Time period (T) (sec)
1
2
3
4
5
Mean T = ………..sec.
Therefore ω = 2/T = ………..radian
© Institute of Lifelong Learning, University of Delhi
21
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Table 3: Determination of RG and effect of circuit resistance on λ and Q
Resistance R2 = ……. KΩ
Resistance r = …….. Ω
S.
No.
1
2
3
.
.
.
R1
(KΩ)
~1
2
3
.
.
.
θ3
(cm)
0.0
θ5
(cm)
θ7
(cm)
θ9
(cm)
θ11
(cm)
θ13
(cm)
…
...
_
_
_
_
_
~15
0.0
(The dashed lines indicates a pattern of the last deflection being observed closest
to zero and hence the effect of the circuit resistance on damping of the
oscillations of the coil.)
Table 4: Critical damping resistance (C.D.R.)
Resistance R2 = ……. KΩ
Resistance r = …….. Ω
S. No.
R1 (kΩ)
1
2
3
4
5
6
7
8
9
10
~ 3 to 5
Decreasing
θ 1 (cm)
θ2 (cm)
Time t (sec)
~1
Decreasing
Decreasing
0.0
0.0
0.0
Minimum
Increasing
Increasing
(The value of R1 in the highlighted row corresponding to θ2 = 0 and the minimum
time, is used in determining the CDR for the given B.G.)
22
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Calculations
Part A
Graph 1

Plot a graph between deflection
 n and the no. of deflection n as shown in
Figure 7. Take the serial number of observation (deflection), n, along x-axis
and corresponding deflection  n (on both sides of zero) along y-axis. Take the


zero of y-axis in the middle of the graph.
Calculate λ using Equation (28).
Obtain the mean value of all observations of Table 2 to find the time-period T,
of the B.G.
Part B




Calculate 1
for all observations of Table 3 and record them in Table 5 given
1
below.
Calculate λ for all observations of Table 3, using Equation (28) and record
them in Table 5 given below.
Calculate p for all observations of Table 3, using Equation (24) and record
them in Table 5 given below.
Calculate Q for all observations of Table 3, using Equation (37) and record
them also in Table 5 given below.
Table 5: For plotting various graphs
S. No.
R1 (kΩ)
1
1
(cm)-1
λ
p
2
(sec)-1
T
Q

2
1
2
3
4
5
6
7
Graph 2
Plot a graph between inverse of first throw θ1 and resistance R1, as shown in
Figure 8. The intercept on x-axis (R1 axis) gives the resistance of the
galvanometer, RG.
© Institute of Lifelong Learning, University of Delhi
23
ELECTROMAGNETIC DAMPING
PHYSICS (LAB MANUAL)
Graph 3
Plot a graph between λ and R1, as shown in Figure 9. (This represents the
variation of damping with the circuit resistance.)
Graph 4
Plot a graph between Q and R1, as shown in Figure 10.
Part C
Calculate the critical damping resistance of the B.G. using Equation (17). Here, R1
is that particular observation in Table 4 which corresponds to  2  0 and the
minimum time, and RG is obtained from Graph 2.
Nature of graphs
θn
0
n
θn
Figure 7: Graph 1
24
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
1/θ1
(cm)
RG
0
R1 (Ω)
Figure 8: Graph 2
λ
↑
0
R1 (KΩ)
Figure 9: Graph 3
© Institute of Lifelong Learning, University of Delhi
25
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Q
0
R1 (Ω)
Figure10: Graph 4
Result









The effect of electromagnetic induction on the motion of the coil of the
ballistic galvanometer was studied.
It was observed that the circuit resistance plays leading role in controlling the
behaviour of the galvanometer.
The resistance of the B.G., RG, was found to be = ….. Ω.
The Critical Damping Resistance, C.D.R., of the B.G. was found to be = … Ω
The logarithmic decrement, λ, was found to vary inversely with the circuit
resistance.
The logarithmic decrement, λ =……..for almost free oscillations of B.G.
The Q-factor was found to vary linearly with the circuit resistance.
The galvanometer behaves as ballistic for circuit resistance larger than the
C.D.R.
The galvanometer behaves as dead-beat for circuit resistance smaller than
the C.D.R.
Probable Sources of Errors
1. The accumulator may not be freshly charged. This will not send a steady
current through the circuit.
2. The B.G. may not be properly leveled. In such case, the coil will not oscillate
in a vertical plane and will lead to incorrect observations.
3. If the spot of light is not properly adjusted on the zero of the scale, all the
observations for  n as well as for the C.D.R. are liable to be incorrect.
26
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
4. The plugs of the resistance boxes may not be clean, or may be loose due to
which the actual value of the resistance introduced in the circuit may be
different from what is observed.
Glossary
Ballistic Constant: A constant for the galvanometer (k) which gives the amount
of charge that has passed through the galvanometer coil on multiplying it by the
corrected first throw ( q  k . 0 ).
Ballistic galvanometer (B.G.): A galvanometer in which an impulse is applied
to the coil of the galvanometer due to the passage of a transient current or
charge and the subsequent motion of the coil is used to determine the magnitude
of the impulse, and, hence the quantity of charge to be measured.
Charge Sensitivity: The charge in micro-coulombs which when sent through the
coil will produce a deflection of 1 mm of the spot of light on the scale placed at a
distance of 1meter from the galvanometer mirror.
Critical damping: Damping in a linear system on the threshold between
oscillatory and exponential behaviour.
Critical damping resistance (C.D.R.): The value of the total resistance in the
circuit, at which critically damped condition is achieved. When resistance in the
galvanometer circuit is larger than the C.D.R., the galvanometer becomes
oscillatory and when it is smaller than the C.D.R., the galvanometer becomes
dead-beat.
Damping: The dissipation of energy in motion of any type, especially oscillatory
motion and the consequent reduction or decay of the motion.
Damped harmonic oscillation: The linear motion of an object, subject both to
an elastic restoring force proportional to its displacement and to a frictional force
in the direction opposite to its motion and proportional to its speed.
Dead-beat galvanometer: A galvanometer which comes to rest without
vibration or oscillation.
Decrement: The ratio of the amplitudes of an under-damped harmonic motion
during two successive oscillations.
Eddy current: A vortex-like flow of current running contrary to the main current.
Electromagnetic: Pertaining to phenomena in which electricity and magnetism
are related.
Electromagnetic damping: Damping pertaining to phenomenon in which
electricity and magnetism are related.
Electromagnetic induction: The phenomenon whereby an induced E.M.F. is
produced whenever there is a change in the magnetic flux linked with the circuit.
E.M.F.: Abbreviated form of electro-motive force. It is the potential difference
between the two plates of a cell when in open circuit. It is also so called because
it is the driving force which sends the current through the cell.
Galvanometer: An instrument used to detect current or charge passing through
it.
Impulse: It is a voltage which rises to a very high value in a short interval of
time and then reduces to zero in a very short time.
Lamp and scale arrangement: An arrangement comprising of a stand with a
lamp and a scale, by which a small angular twist of the coil and hence of the
small round mirror attached to it can be measured as a much larger linear
deflection of the image of the spot of light on the horizontal scale.
Lead accumulator: It is a secondary cell or storage cell. Here, current from
some other supply is first passed through it for some time and is stored as
chemical energy. The same current is then obtained from it in the reverse
direction through an external circuit. (i) It has very low internal resistance, (ii) its
efficiency is very high and (iii) it gives a constant current.
© Institute of Lifelong Learning, University of Delhi
27
ELECTROMAGNETIC DAMPING
PHYSICS (LAB MANUAL)
Logarithmic decrement: The natural logarithm of the ratio of the amplitude of
one oscillation to that of the next, when no external forces are applied to
maintain the oscillation.
Moment of inertia: The sum of the products formed by multiplying the mass of
each element of an object by the square of its distance from a specified line. Also
known as rotational inertia.
Moving coil galvanometer: A galvanometer in which a current carrying coil is
placed in a magnetic field and is subject to a force, the direction of which is
normal both to the direction of current and the direction of magnetic field. This
leads to the motion of the coil.
Moving magnet galvanometer: A galvanometer in which a current is passed
through a fixed coil which produces a magnetic field, under which a small magnet
/ magnetic needle moves.
Oscillation: Any effect that varies periodically back and forth between two
values.
Phosphor-bronze: An alloy with Copper-92.5%, Tin-7% and Phosphorous0.5%. It has great tensile strength, its torsional couple per unit twist is small and
it is rust-resistant.
Quality factor (Q-factor): A measure of the ability of a system with periodic
behaviour to store energy equal to 2 times the average energy stored in the
system divided by the energy dissipated per cycle.
Radial magnetic field: Achieved by making the magnetic poles curved into a
concave shape. This is done in order to make the magnetic field always parallel to
the plane of the coil, whatever be the position of the coil.
Simple harmonic motion: A periodic motion about an equilibrium position for
which the displacement is a sinusoidal function of time. The acceleration of the
object is always directed towards the equilibrium position and is proportional to
the displacement from that point.
Steady current: A current which is constant with respect to time.
Time period: Time taken by a moving system to complete one vibration or
oscillation.
Torque: For a single force, the cross-product of a vector from some reference
point to the point of application of the force with the force itself. Also known as
moment of force or rotational moment. For several forces, the vector sum of the
torques associated with each of the forces gives the resultant torque.
Torsion: A twisting deformation of a solid body about an axis in which lines that
were initially parallel to the axis become helices.
Torsional rigidity: The ratio of the torque applied about the centroidal axis of a
bar at one end of the bar to the resulting torsional angle, when the other end is
held fixed.
Transient current/phenomena: A current pulse, or other temporary
phenomenon occurring in a system for a short duration.
Post-lab Assessment
Answer the following questions
(1)
(2)
(3)
(4)
(5)
(6)
28
How do you correct the observed throw for damping?
What do you understand by the sensitiveness of a galvanometer?
How do you obtain the galvanometer resistance?
Why do you multiply the expression for logarithmic decrement by 2.3026?
How do you stop the unwanted oscillations of the B.G.?
Can you use a ballistic galvanometer to measure current?
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
(7)
(8)
(9)
(10)
Why is the suspension used in the form of a rectangular strip?
Which material is used for making the B.G. coil and why?
What is the order of resistance of your B.G.?
With reference to the above question, what will be the order of the
resistance if this galvanometer was of dead-beat type?
(11) Is the value of the Ballistic Constant k, a constant for a given galvanometer
or does it depend upon the working conditions?
(12) What advantages does the lamp and scale arrangement for measuring
deflections possess over the use of mechanical pointer?
Choose the correct answer
(1)
(2)
The motion of a torsional pendulum is simple harmonic. If  is the
moment of inertia of the pendulum and c is the restoring couple per unit
angular twist, then its time period will be
a)
T  2
b)
T  2
c

c)

c
T  2 c
d)
T  2
1
.
c
The equation of motion of a moving coil galvanometer is given as
d 2
d
 0.2
 36  0 . The period of oscillation will approximately be
2
dt
dt
a)

b)

c)

d)

6
2
3
4
sec
sec
sec
sec.
(3)
In the above equation, Q-factor will be
a)
3
b)
15
c)
20
d)
30
(4)
In the above question, λ (the logarithmic decrement) will be
a)
0.05
b)
0.5
c)
0.01
d)
0.1
The sensitivity of a moving coil galvanometer is large if
a)
the magnetic field is small
b)
the area of the coil is small
c)
the torsional constant of the suspension strip is small
d)
number of turns of the coil is small.
(5)
© Institute of Lifelong Learning, University of Delhi
29
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
Answers to Pre-lab Assessment
1.
2.
3.
4.
Characteristics of Ballistic galvanometer:
(i) The moment of inertia of the coil should be large.
(ii) The air damping should be small.
(iii)The electromagnetic damping should be small. Thus, the coil should be
wound on a non-conducting frame.
(iv)The resistance R of the circuit, including that of the coil should be large.
For this, the coil should have large number of turns.
Characteristics of dead beat galvanometer:
(i) The moment of inertia of the coil should be small.
(ii) The electromagnetic damping should be large. Thus, the coil should be
wound on a conducting frame.
(iv) The resistance R of the circuit, including that of the coil should be small.
For this, the coil should have small number of turns.
There are two types of moving coil galvanometer:
(i) Pivoted-coil type, where the coil swings between two pivots and the
deflection is read by a pointer which moves on a circular scale.
(ii) Suspended-coil type, where the coil is suspended freely by a thin
phosphor bronze strip and the deflection is measured by a lamp and scale
arrangement.
On account of the resistance of the medium (air) and electromagnetic
damping, the magnitude of the amplitude of vibration of the coil decreases
exponentially with time. The ratio of the successive amplitudes on either
side of the position of rest is called „decrement‟ and the logarithm of this to
the base e is called logarithmic decrement.
It is determined to correct the observed throw for damping. The method
adopted to determine it is to observe the ratio of the first and eleventh
swing and then


2.303
log 10 1 .
10
11
5.
Electromagnetic damping is the dissipation of energy in motion of an
oscillatory system which causes decay of the motion due to the production
of electromagnetically induced e.m.f. in the moving system. It can be
reduced by winding the coil of the B.G. on a non-conducting frame made of
ivory, bamboo or ebonite.
6.
Time period of the suspended coil depends directly on the moment of inertia
of the coil and inversely on the torsional constant of the suspension. Hence,
to increase the time period, the moment of inertia of the coil should be
increased and suspension should be made of the material having low
torsional constant
7.
The oscillations are stopped by short-circuiting the coil by closing the
damping key. Eddy currents are produced in the coil in a direction which
opposes the motion of the coil.
8.
The sensitivity of the B.G. can be increased by
(i)
increasing the number of turns in the coil,
(ii) increasing the effective area of the coil,
(iii) increasing the magnetic field,
(iv) reducing the time period of the moving system,
(v) reducing the couple per unit twist of the suspension.
9.
It is a constant, which when multiplied by the throw of the galvanometer,
gives the amount of charge which has passed through it.
10. By using the suspension of phosphor-bronze in the form of a thin
rectangular strip.
30
© Institute of Lifelong Learning, University of Delhi
PHYSICS (LAB MANUAL)
ELECTROMAGNETIC DAMPING
11.
E.M.F. of a cell is the abbreviated form of electro-motive force. It is the
potential difference between the two plates of a cell when in open circuit. It
is so called because it is the driving force which sends the current through
the cell.
12. Secondary cell is preferred because (i) it gives strong current as its internal
resistance is very low, (ii) its efficiency is very high i.e. it gives back most of
the electrical energy used in charging it and (iii) it gives a constant current
due to the large area of its plates and their close proximity.
Answers (multiple choice questions)
1.
2.
3.
4.
5.
(a)
(c)
(c)
(a)
(c)
Answers (Yes/No)
1.
2.
3.
4.
5.
6.
No
No
No
Yes
No
Yes
Answers to Post-lab Assessment

1.
The observed throw is multiplied by ( 1 
2.
A galvanometer is said to be sensitive when a small current / charge
passing through it produces a large deflection in it.
A graph is plotted between 1/θ1 and R1. A straight line is obtained. On
extrapolating it backwards, it gives an intercept on the negative side of
resistance axis that gives the galvanometer resistance.
The logarithmic decrement is defined with respect to the base e. in order to
convert its expression to the base 10, it is multiplied by 2.3026.
The unwanted oscillations of the B.G. can be arrested by short circuiting the
coil, which is achieved by closing the tapping key K 1. Eddy currents are
produced in the coil in a direction so as to oppose the motion of the coil.
Yes, a B.G. can be used to measure current, but we will have to wait for a
long time for the coil to give a steady deflection.
It is so used because (i) the rectangular strip has a very small torsional
rigidity resulting in much greater twist for a given couple, and much lesser
stress on the material than with a wire of circular cross section and (ii) a
strip possesses an additional advantage of offering comparatively large
surface area for the dissipation of heat produced by the current.
Thin enamelled copper wire is used, as it has a low specific resistance.
It is of the order of 150 Ω.
[Note: The resistance is generally written on a plate attached to the case of
the B.G. by the manufacturers].
The resistance of the dead-beat galvanometer may be of the order of 25 Ω.
3.
4.
5.
6.
7.
8.
9.
10.
2
).
© Institute of Lifelong Learning, University of Delhi
31
ELECTROMAGNETIC DAMPING
11.
12.
PHYSICS (LAB MANUAL)
If k is expressed in coulombs per radian, it is independent of the working
conditions; but when it is expressed as coulomb per mm deflection at a
given distance of the scale from the mirror, it will depend on this distance.
By using a lamp and scale arrangement, a small angular twist of the coil and
hence of the mirror attached to it, can be measured as a much larger linear
deflection of the image of the spot of light on a horizontal scale.
Answers (multiple choice questions)
1.
2.
3.
4.
5.
32
b
c
d
a
c
© Institute of Lifelong Learning, University of Delhi
Download