1 LED Luminaire Lumen Maintenance – LM-80 Data APPLICABLE MODEL: MILLENIUM SEMI-RECESSED EDGE™ MR13FFR SERIES WITH 20 WATT 4000K LED (20L40K) The purpose of this document is to demonstrate that Kenall’s Millenium Edge MR13FFR follows current industry standard procedures in reporting the expected lumen maintenance of the LEDs within the product under normal operating conditions. Lumen maintenance of the Philips-Lumileds LEDs used within this product is estimated by the following sets of data: In-Situ Product Testing of a representative product’s temperature characteristics (specifically the measurement related to LED junction temperature). For this product series, this product was evaluated by a UL Accredited DAP laboratory to comply with UL8750 and UL1598 standards under an ambient temperature of 25°C (77°F). The critical measurement concerning lifetime is the LED temperature measurement point (TMPLED), which has been measured per Lumileds Application Brief AB33. LM-80 Test Results for the LED Package: Lumen maintenance evaluation of the LED source by its manufacturer under compliance with IES LM-80-08 guidelines. The purpose of LM-80-08 is to determine the lumen maintenance of an LED package, array or module after 6,000 hours of operation under specific ambient temperature and electrical operating conditions. The LEDs used within the luminaire are expected to maintain the lumen maintenance shown in the LM-80 testing if the LED temperature measurement point (TMPLED) is less than or equal to the temperature(s) specified in the LM-80 test report for the corresponding drive current and the drive current in the luminaire is less than or equal to the drive current specified in the LM-80 test report at the corresponding temperature or higher. A summary of the lumen maintenance information is shown below: LED Drive Current 700mA In-Situ LED Temperature Measurement Point (TMPLED)184.1°C Lumileds LM-80 Test Temperature285.0°C Lumen Maintenance @ 6,000 hours/10,000 hours2 94.9% / 93.7% Luminaire L70 Design Lifetime 50,000 hours 1 UL DAP Accredited laboratory test results of temperature corrected to 25°C ambient 2 Test result values from Lumileds LM-80 test results (Design Resource DR03 – 05/28/2010) Please refer to attached excerpts from Design Resource DR03: LM-80 Test Report Excerpt from Design Resource DR03: LM-80 Test Report 4. Operating cycle 4. Operating cycle LUXEON Rebel LEDs are driven with a constant direct current (DC). LUXEON Rebel LEDs are driven with a constant direct current (DC). 5. Ambient conditions including airflow, temperature, and relative humidity 5. Ambient conditions including airflow, temperature, relative humidity The thermal chamber has minimal airflow. The case temperature withinand the thermal chamber was characterized by mounting several thermocouples Thea thermal chamber stress has minimal The case temperature within thepoint, thermal chamber was characterized by mounting several thermocouples on sample reliability board airflow. at the designated thermal measurement as shown in Application Brief, LUXEON Rebel Thermal on a sample reliability stress board at the designated thermal measurement point, as shown in Application Brief, LUXEON Rebel Thermal Measurement Guidelines (AB33). In addition, several thermocouples were mounted in the air at a distance of 1.5 mm above the reliability stress Measurement Guidelines In addition, several thermocouples were mounted in the a distance of 1.5 mm above the reliability stress board. Then the reliability(AB33). stress board was mounted in the thermal chamber and driven at air theatspecified stress condition. The thermocouple board. Then the reliability stress board was mounted in the thermal chamber and driven at the specified stress condition. The thermocouple readings were monitored. After the thermocouples reached thermal equilibrium, the thermocouple readings were data-logged and averaged readings were monitored. Afterwithin the thermocouples reached thermal equilibrium, the thermocouple readings were data-logged and averaged together. The relative humidity the oven was also characterized. together. The relative humidity within the oven was also characterized. 6. Case and ambient temperatures (ambient temperature measured 1.5mm above reliability stress board) 6. Case ambient temperatures (ambient temperature above reliability The thermaland characterization results are summarized in the table below. In allmeasured cases, both TS1.5mm and TA meet or exceed the IES stress LM-80-08board) limits TheTthermal characterization results summarized in the table below. In all cases, both TS and TA meet or exceed the IES LM-80-08 limits (i.e. ] 55°C or 85°C, and that TA]are TS –5°C). S (i.e. TS ] 55°C or 85°C, and that TA] TS –5°C). Required Required LM-80 Test LM-80 Test Temperature Temperature IF IF Actual Actual TS TS Actual Actual TA TA 55°C 55°C 85°C 350 mA 350 mA 60°C 60°C 85°C 64°C 64°C 84°C 85°C * 108°C 108°C * 350 mA 350 mA 85°C 128°C 128°C 84°C 103°C 103°C LM-80 Tests LM-80 Tests Actual Actual D[T – TS] A D[TA– TS] +4°C +4°C -1°C -1°C –25°C –25°C Actual Actual Relative Relative Humidity Humidity 18% 18% 7% 7% TBD TBD Average Average lumen output atlumen 6,000output hours at 6,000 hours 0.992 0.992 0.962 0.962 0.979 0.979 Average Average lumen output atlumen 10,000output hours at 10,000 hours 0.983 0.983 0.955 0.955 0.956 0.956 * Third temperature chosen by the manufacturer, and adjusted to meet –5°C TA – TS requirement. * Third temperature chosen by the manufacturer, and adjusted to meet –5°C TA – TS requirement. Additional Additional Test Test Temperature Temperature 55°C Additional Tests Provided www. kenall.com Actual P:Additional 800-4-Kenall 847-360-1781 Tests F: Provided Actual Actual Actual IF IF Actual TS TS Actual TA TA Actual D[T – TS] A D[TA– TS] 700 mA 69°C 73°C +4°C Actual Relative Relative Humidity Humidity 18% 1020 Lakeside Drive Gurnee, Illinois 60031 Average Average Average lumen output atlumen 6,000output hours at 6,000 hours 0.978 LM80TECHDATA-020912 Average lumen output atlumen 10,000output hours at 10,000 hours 0.967 108°C * 350 mA 128°C 103°C –25°C TBD 0.979 0.956 * Third temperature chosen by the manufacturer, and adjusted to meet –5°C TA – TS requirement. 2 Additional Tests Provided Additional Test Temperature IF Actual TS Actual TA Actual D[TA– TS] Actual Relative Humidity Average lumen output at 6,000 hours Average lumen output at 10,000 hours 55°C 85°C 700 mA 700 mA 69°C 92°C 73°C 87°C +4°C –5°C 18% 7% 0.978 0.949 0.967 0.937 55°C 1000 mA 71°C 80°C +9°C 18% 0.955 0.945 7. Drive current of the LED light source during lifetime test Specified in the tables. 8. Initial Luminous flux and forward voltage at photometric measurement current See tables. 9. Lumen maintenance for data for each individual light source along with median value, standard deviation, minimum and maximum lumen maintenance value for all of the light sources See tables. Design Resource DR03 (10/05/28) Lumen Maintenance Projection for White LXM3-PWx1 LUXEON Rebel under these conditions 2 85°C, 0.7A ((Tjunction j 112°C)) Normalized to 1 at 24 hours 1.2 *Philips Lumileds recommends p extrapolations of 6x the test time Normaliz zed Light Ou utput 1.1 1.0 Extrapolated data = –6.5% at 10,000 hours 0.9 35,000 hour 25,000 hour 0.8 ENERGY STAR 25,000 and 35,000 limits after 6,000 6 000 hours of stress 6x test time* 0.7 L70 = 65,000 hours Ta within –5°C of Ts, in accordance with LM80. 0.6 0 6 1,000 10,000 100,000 1,000,000 Hours ©2010 Philips Lumileds Lighting Company. All rights reserved. www. kenall.com P: 800-4-Kenall F: 847-360-1781 1020 Lakeside Drive Gurnee, Illinois 60031