kV imaging dose in RT: evaluation and protocol optimization G. Bolard, Medical Physicist SSRPM FOPH Radiation Protection Day RT 09/21/2012 Imaging in RT ♦ Highly conformal treatments are becoming the standard of care – IMRT/VMAT/TOMO ♦ Conformal plans are more sensitive to patient misalignment/organ position ♦ kV or MV imaging is the main modality used in linacs for pretreatment positioning and evaluating organ movement/deformation ♦ With the introduction of adaptive radiotherapy (ART), the use of volumetric imaging (CBCT) in RT treatments is increasing FOPH Radiation Protection Day RT Bern, 09/21/12 Status ♦ Daily imaging is now the standard in RT – Planar imaging or kV/MV CBCT on linacs/TOMO ♦ Imaging dose depends on the frequency of use and tumor location ♦ Treatment machines are not equal regarding the compromise dose/image quality FOPH Radiation Protection Day RT Bern, 09/21/12 Context ♦ Volumetric imaging (MV/kV CBCT) is the most irradiating modality ♦ In some cases, daily CBCT may lead to doses of the same level as out-of-field radiation dose ♦ Dose summation with RT dose is not straightforward % – kV dose distribution differs from typical MV distribution: kV is characterized by a high surface dose and rapid attenuation in the body – Gy can be added to Gy but does it make sense from a biological point of view? ♦ Important for the RT staff (technologists and doctors) to know the level of dose delivered per imaging procedure FOPH Radiation Protection Day RT cm Bern, 09/21/12 Context ♦ ALARA principle is applicable ♦ Recommendations exist to manage additional imaging dose – AAPM TG75 2007 ♦ Technologists, doctors and physicists must all be involved in imaging dose optimization ♦ Manufacturers begin to indicate dose indices – CTDI for CBCT – Air kerma for radiography FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT ♦ Estimation of dose delivered to the patient – Derived from CTDI measurements – TLD measurements in anthropomorphic phantoms – Direct measurements – Prediction (computation) ♦ Measurements – TLDs in Rando pelvic phantom – CTDI with pencil and Farmer ionization chamber (calibrated in RQA9) – Dosimeters supplied by Institute of Radiophysics (IRA/CHUV) FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT: CTDIw Results for Varian OBI Protocol Measurement (mGy) Manufacturer Data (mGy) Standard Dose Head 4.8 3.9 Pelvis 20.0 17.7 Pelvis Spot Light 25.5 19.4 Low Dose Thorax 5.8 4.7 ♦ Keep in mind that CTDI is not patient dose – Mean dose in a slice for 32cm diameter perspex phantom – For thinner patients, the same exposure parameters (kV,mAs) result in a greater dose to the patient – AAPM TG 204 2011 (Size-Specific Dose Estimates in Pediatric and Adult Body CT Examinations) FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT: TLD results OBI: Pelvis OBI: Pelvis Spot Light FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT: Daily use scenario ♦ 39 fractions (prostate ttt, 78Gy, 2Gy/fr) Organ Absorbed dose (Gy) Pelvis spot light Pelvis Bladder 0.6 0.5 Prostate 1 0.9 Rectum 1.5 1 Left femoral head 0.6 0.7 Right femoral head 0.3 0.9 Skin 3.6 (post & lat) 1.7 Should this dose be taken into account for retrospective clinical analyses? FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT Modeling ♦ Convolution/Superposition algorithms may be used for computation on actual patient anatomy – Modeling of OBI in our TPS Philips Pinnacle3 – Monokernels with energy less than 100 keV and initial beam model supplied by Dr. Parham Alaei, University of Minnesota, Minneapolis FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT Modeling Daily use of CBCT in pelvic ttts PTV & OAR mean doses are increased by 1Gy 2-3 Gy for skin and bones Inhomogeneous dose in PTV FOPH Radiation Protection Day RT Bern, 09/21/12 Volumetric Imaging: Typical doses Imaging dose and image quality are manufacturer dependent FOPH Radiation Protection Day RT Bern, 09/21/12 Volumetric Imaging: Image quality Contrast to Noise Ratio (CNR) and resolution (lp/mm) The best CNR is not from the equipment that delivers the highest dose FOPH Radiation Protection Day RT Bern, 09/21/12 CBCT optimization ♦ Patient size-specific protocols – Adapt mAs to account for patient size – Axial collimation ♦ Consider image requirements for the specific task – Soft contrast detectability, resolution, HU accuracy – Effect on matching ♦ Also – Reconstruction parameters (filter, voxel size) – Amount of scatter (FOV) – Geometric calibration of the system ♦ → Image quality/dose relationship may be complex FOPH Radiation Protection Day RT Bern, 09/21/12 kV planar imaging ♦ Dose is mainly deposited superficially and is characterized by air kerma or entrance skin dose ♦ Dose at depth is much less than for MV images ♦ Greater possibilities to reduce dose – – – – Only bony anatomy/marker visualization needed Collimation (blades) mAs & kV Incidence ♦ kV imaging is not always necessary – Orthogonal imaging should not be default procedure – MV imaging of treatment fields where MU subtraction is possible can be better (breast, palliative) ♦ Technologists’ role is essential → training FOPH Radiation Protection Day RT Bern, 09/21/12 Measurements : Varian OBI ♦ With pancake 60cc chamber, Air kerma @ 85cm (mGy) Body site Default protocol Manufacturer Display In-house measurement AP Med 0.7 0.8 Lat Med 10.7 13.1 AP Large 1.3 1.5 Lat Large 21.6 26.8 Head AP 0.98 1.2 Head Lat 0.30 0.4 Thorax AP 0.34 0.5 Thorax Lat 4.4 6.8 Abdomen AP 2.5 3.9 Abdomen Lat 3.5 5.5 Pelvis Head Thorax Abdomen FOPH Radiation Protection Day RT Bern, 09/21/12 Planar imaging optimization Body site Protocol Original settings New settings Air kerma divided by AP Med 75 kV, 10 mAs 70 kV, 5 mAs 2.3 Lat Med 105 kV, 80 mAs 90 kV, 12.6 mAs 8.6 AP Large 80 kV, 16 mAs 90 kV, 10 mAs 1.3 Lat Large 120 kV, 126 mAs 105 kV, 16 mAs 10.3 Head AP 100 kV, 8 mAs 70 kV, 4 mAs 4.1 Head Lat 70 kV, 5 mAs 70 kV, 4 mAs 1.2 Thorax AP 75 kV, 5 mAs 70 kV, 5 mAs 1.1 Thorax Lat 95 kV, 40 mAs 85 kV, 6.4 mAs 7.8 Abdomen AP 80 kV, 32 mAs 80 kV, 5 mAs 6.4 Abdomen Lat 85 kV, 40 mAs 90 kV, 12.6 mAs 2.8 Pelvis Head Thorax Abdomen Varian OBI 1.5 FOPH Radiation Protection Day RT Bern, 09/21/12 Conclusion ♦ Important to know the level of dose delivered by imaging for the patient treatment course – For multiple fractions and daily 3D imaging, the additional dose can represent 1-2% of the prescribed dose ♦ Optimization of manufacturer protocols is possible especially for planar kV imaging ♦ Technology must progress for CBCT – Improve the image quality/dose ratio FOPH Radiation Protection Day RT Bern, 09/21/12 Thank you for your attention! FOPH Radiation Protection Day RT Bern, 09/21/12