Vector network analyzers (VNAs) are increasingly being used for impedance measurements of electronic components and circuits, especially in R&D environments where the versatility and flexibility of the VNA-based solution are preferred. This presentation introduces the impedance measurement solution of the E5061B-3L5 ENA Series LF-RF network analyzer, and discuss how to get the most of its impedance measurement capabilities for better measurements. 1 Before starting the discussion on the impedance measurement, let’s quickly look at the product overview of the E5061B-3L5 LF-RF network analyzer. 2 The E5061B-3L5 offers versatile network analysis in the broad frequency range from 5 Hz to 3 GHz. Comprehensive LF network measurement capabilities including built-in 1 Mohm inputs have been seamlessly integrated with the high-performance RF network analyzer. This slide summarizes key features of the E5061B-3L5. 3 The E5061B-3L5 has two kinds of test ports; the S-parameter test port (Port 1 and 2) and the gain-phase test port. The S-parameter test port has a 50 ohm built-in S-parameter test set that seamlessly covers a broad frequency range from 5 Hz to 3 GHz. While providing an excellent RF performance equivalent to other RF VNAs, the E5061B-3L5’s Sparameter test port offers much wider frequency coverage down to 5 Hz and better dynamic range in the low to middle frequency range below 10 MHz. 4 The gain-phase test port has reference and test receiver inputs whose input impedance is switchable to 1 Mohm or 50 ohm. The frequency range is 5 Hz to 30 MHz. The most typical application is the frequency response analysis for low-frequency devices and circuits, such as OP amps and control loop circuits of DC-DC converters. 5 And the E5061B’s function we discuss in this presentation is the E5061B005 impedance analysis function, which is a firmware option dedicated to the E5061B-3L5 LF-RF network analyzer. This impedance analysis function fully supports basic functionalities of the impedance analyzer. In addition, by supporting multiple impedance measurement methods, it covers a variety of impedance measurement applications. The E5061B-3L5/005 primarily targets at a simple and convenient VNAbased impedance measurement solution for general R&D use, in contrast to dedicated impedance analyzers which pursue an ultimate impedance measurement performance. In many applications, however, the E5061B3L5/005 allows you to perform very accurate impedance measurements comparable to the dedicated impedance analyzers by properly selecting the measurement method and optimizing measurement conditions including the calibration and fixturing. Furthermore, this impedance analysis solution offers unique capabilities such as a very wide frequency coverage and a milliohm impedance measurement. 6 Now let’s see more details about the E5061B’s impedance measurement methods. 7 The first two slides quickly review the measurement methods of impedance analyzers. Low-frequency impedance analyzers/LCR meters such as the 4294A employ the auto balance bridge method. The negative feedback loop circuit maintains the potential at the low terminal to virtual ground (zero volt), which eliminates the stray capacitance at this point and enables the voltmeters V1 and V2 to accurately measure the voltage and current at the DUT. Also, the auto ranging capability provides a very wide measurement range with an excellent linearity. Thus, the auto balance bridge method offers very high accuracy in the very wide impedance range. The frequency range is up to 110 MHz Note that the actual auto balance bridge is a more complicated feedback loop circuit than the very simplified OP-amp model shown here. 8 RF impedance analyzers/LCR meters such as the E4991A employ the RF I-V method. With the combination of the source/receiver mainframe and the test head module which senses high-frequency current and voltage at the very close location to the DUT, the RF I-V method offers higher measurement accuracy and wider measurement range than the VNA-based reflection method. However, this method is not applicable to the low frequency range because the current is sensed with a transformer. Next, let’s see the impedance measurement methods supported by the E5061B-3L5/005 network analyzer. 9 The reflection method is traditionally the most common VNA-based impedance measurement technique mainly used in the middle to high frequency range. The reflection method derives the impedance from the S11 measurement data. The impedance measurement range is a little narrower than that of the RF I-V method. The 10% accuracy range (Supplemental Performance Data) is about 1 ohm to 2 kohm. The graphs shown in the slides indicate the impedance range where each of the VNA-based measurement methods provides a good impedance measurement sensitivity. In the reflection method, the S11 measurement value dynamically varies in the magnitude (in case of resistive DUTs) or the phase (in case of reactive DUTs) when the measured impedance is about 1 ohm to 2 kohm. This is the impedance range where the reflection method provides accurate impedance measurements. The advantage of the reflection method over the RF-IV method is the lower frequency coverage due to the broadband S-parameter test set of the E5061B-3L5. We can use 7 mm type component measurement fixtures with the 16201A terminal adapter which converts the analyzer’s N-type test port to the 7 mm type. 10 The series-thru method measures the impedance by connecting the DUT in the transmission series configuration as shown in the block diagrams. This method is suitable for measuring high impedance. The 10% accuracy range is about 5 ohm to 20 kohm, which covers 1-decade higher impedance range than the reflection method. Both the gain-phase and the S-parameter test ports can be used for the series-thru method. Especially, the gain-phase test port is convenient because the 4-terminal-pair type component test fixtures can be directly connected to the gain-phase test port. The upper frequency limit is 30 MHz. Also, it is possible to perform the series-thru method up to higher frequencies by using the S-parameter test port and a user-prepared test fixture. However, it is difficult to fully eliminate the errors around the series-thru fixture as the frequency goes higher than hundreds of MHz. So the practical upper frequency limit is up to around 200 or 300 MHz. Note that the grounded DUT cannot be measured with the series-thru method. 11 The shunt-thru method measures the impedance by connecting the DUT in the transmission shunt configuration as shown in the block diagram. This method is suitable for measuring very low impedance, and commonly used for milliohm impedance measurements of power integrity applications. The 10% accuracy range is 1 mohm to 5 ohm (in the case of using the gain-phase test port), which covers lower impedance range than impedance analyzers. Both the gain phase and S-parameter test ports can be used for the shunt-thru method. For the low impedance measurement up to high frequencies over 30 MHz, the shunt-thru method using the S-parameter test port is the solution. However, for the low impedance measurement at low frequencies below 100 kHz, it is recommended to use the gain-phase test port rather than the S-parameter test port. Unlike the E5061B’s Sparameter test port and other existing VNAs, the gain-phase test port employs a semi-floating receiver architecture which eliminates the measurement error caused by the ground loop, and we can measure very low impedance in the low frequency range easily and accurately . For details about the ground loop error and the semi-floating receivers, refer to Appendix of this presentation material. 12 This slide shows an effect of the semi-floating receiver of the gain-phase test port. The DUT in this example is a resistor whose DC resistance is about 1 mohm. The screen shot in the upper left is the measurement result using the Sparameter test port whose outer shield is connected to the instrument’s chassis ground, like other conventional VNAs. The measured result in the low frequency range is incorrectly higher than the DUT’s true impedance value due to the ground loop between the source and receiver. The screen shot in the lower left is the measurement result using the Sparameter test port with a magnetic core attached to the test cable, which is a traditional way to reduce the ground loop errors. The measurement in the low frequency range is improved by increasing the outer-shield inductance with the core. But it is not so easy to fully eliminate the errors down to the very low frequency range, as shown in this screen shot. On the other hand, the screen shot on the right side is the measurement result using the gain-phase test port with no magnetic core attached to the cable. As you can see, the DUT’s milliohm impedance is correctly measured down to the very low frequency range, even without using the core. 13 This graph shows the 10 % accuracy range of three measurement methods using the S-parameter test port. 14 This graph shows the 10% accuracy range of the series-thru and shuntthru method using the gain-phase test port. 15 This table summarizes how to select the measurement methods depending on the applications. The basic recommendations are; • For general component measurements at middle to high frequency ranges, the reflection method is recommended. •For general component measurements at low to middle frequency ranges, select either of the gain-phase series-thru method or the reflection method depending on the DUT’s impedance value. • For very low impedance measurements, the shunt-thru method is recommended. On the other hand, impedance analyzers such as the 4294A and E4991A are basically recommended in the following cases; • Very high accuracy is required. • Accurately measure very high impedance. • Accurately measure very high Q (=X/R) or very low D (=R/X). 16 This slide shows recommended migration paths from legacy NA/ZA combination analyzers to the E5061B-3L5 with the option 005 impedance analysis function. 17 The next section discusses calibration techniques and other tips for improving the measurement accuracy in the actual impedance measurement applications of the E5061B-3L5/005. 18 Firstly, let’s see the basics of the calibration for impedance measurements. The series-thru and shunt-thru methods are based on the 2-port transmission measurement. If the measurement system used for these methods is the S21 measurement configuration with the 50 ohm system impedance, the VNA calibration methods such as the response thru and the 2-port full (SOLT) can be used for impedance measurements. 19 On the other hand, if we look at the impedance only, without worrying about the measurement in the S-parameter domain, we can perform the calibration by considering the measurement system as a black box and adjusting this black box with the open, short and load whose complex impedance values are known. This is the concept of the open/short/load calibration in the impedance domain. The E5061B-005 impedance analysis firmware provides this function as an Impedance Calibration function, and you can perform the open/short/load calibration not only in the reflection method but also in the series-thru and shunt-thru methods. Here it should be noted that the measurement system treated as the black box must satisfy the following conditions; • Operating in the sufficiently linear region (no gain compression and distortion). • Three standards can be clearly distinguished with measured voltages. • Load device can be measured stably. Considering these basic points about the calibration, next let discuss more details about the calibration techniques of each measurement method. 20 The typical calibration of the reflection method is the open/short/load calibration at the 7 mm coax plane, the port extension (selecting the fixture model) to compensate for the phase shift at the fixture’s coax section, plus the open/short compensation to eliminate the stray capacitance and the residual inductance around the fixture’s electrodes. For the open/short/load calibration at the 7 mm connector plane, you can use either of the Impedance Calibration function or the conventional 1port full calibration function. In the case of the Impedance Calibration function, you can perform the low-loss-capacitor calibration in addition to the open, short, and load. The low-loss capacitor is an air capacitor with approx. 3pF capacitance, and it improves the measurement accuracy of the phase, Q, or D at high frequencies over 300 MHz by reducing the phase uncertainty of the 50 ohm load termination. The low-loss capacitor is included in the 16195B 7 mm calibration kit. 21 The typical calibration of the gain-phase series-thru method is the open/short/load calibration at the 4-terminal-pair fixture. Leaded and SMD-type 50 ohm resistors are provided as the E5061B’s optional accessory. In the series-thru method up to higher frequencies using the S-parameter test port and a user-prepared test board, the most practical calibration is the SOLT calibration at the coaxial cables plus the port extension to compensate for the transmission lines on the test board. In addition, it is possible to perform the open compensation to remove the stray capacitance around the measurement terminal. 22 In the gain-phase shunt-thru method used for low impedance measurements in the low frequency range, the response thru calibration is the primarily recommended calibration method. The simple response thru calibration gives enough accuracy in most of low-frequency milliohm measurement applications, where only the impedance magnitude |Z| and the reactance Cs and Ls need to be measured. But if the impedance to be measured goes up to higher impedance ranges over 1 ohm, or if you want to measure the phase or ESR at higher frequencies over 10 MHz, it is recommended to perform the open/short/load calibration because the measurement errors in these situations cannot be fully eliminated only with the response thru calibration. In the shunt-thru method using the S-parameter test port up to hundreds of MHz or GHz ranges, the recommended calibration is the SOLT calibration at the coaxial cables plus the port extension to compensate for the fixture/probes. In case you use an RF probe station (for example, to characterize IC packages or PCBs), you can perform the SOLT calibration at the end of the probes by using the calibration standards provided by the probe vendors. 23 This table summarizes other tips for improving the impedance measurement accuracy of the E5061B. 24 This slide explains more details about the tips on how to contact to the DUT in the shunt-thru method. 25 The last section shows actual impedance measurement examples of typical DUTs using the E5061B-3L5/005. 26 The first three measurement examples are basic inductor and capacitor measurements. These examples show that the E5061B can accurately measure impedance by choosing an appropriate measurement method and taking care of the calibration and other tips discussed in the previous section. This is a measurement example of an 100 nH RF inductor using the reflection method. The test frequency range with the E5016B is from 10 kHz to 3 GHz, which cannot be covered by existing impedance analyzers. The two graphs shown on the right hand side compare the measurement results with the E5061B (at 10 kHz to 3 GHz), 4294A (at 10 kHz to 100 MHz), and the E4991A (at 1 MHz to 3 GHz). The E5061B’s measurement results show very good agreements with two impedance analyzers in both Ls and Q values, although it is partly due to the DUT’s moderate |Z| and Q values which are relatively easy to measure. 27 This is a measurement example of a high-impedance device using the gain-phase series-thru method. The DUT is a 10 nF capacitor and the test frequency range is 100 Hz to 10 MHz. As shown in the graphs on the right hand side, the E5061B’s measurement results has good agreements with the 4294A in both Cp and D. 28 This is an example of a very-low impedance measurement using the shunt-thru methods. The DUT is a low-ESR SMD electrolytic capacitor with 200 uF capacitance. The test frequency range is 1 kHz to 10 MHz. The graphs on the right hand side compare the measurement results of the gain-phase shunt-thru method (with the thru cal, or the open/short/load cal), the shunt-thru method at the S-parameter test port (with the SOLT cal), and the 4294A (with the 42941A 7 mm adapter and the 16092A 7 mm fixture). Both Cs and Rs (=ESR) measurement results of the E5061B exhibit good agreements with the 4294A, except the Rs measurement value with the thru calibration in the high frequency range. Also, it should be noted that the Rs measurement traces of the E5061B’s shunt-thru methods are more stable than the 4294A’s. This indicates an excellent low-impedance measurement sensitivity of the VNA-based shunt-thru method. 29 The next example is a resonator measurement, which is also a typical measurement application for impedance analyzers. The DUT in this example is a crystal resonator whose Q-factor is much higher than ceramic resonators. The measurement shown on the left hand side is performed in the span fully covering the DUT’s resonant frequency Fr and the anti-resonant frequency Fa. In this case, the equivalent circuit parameters are correctly derived by using the E5061B’s equivalent circuit analysis function, but the CI value (|Z| at 0 degree) read by putting the marker on the 0 degree point at Fr is incorrectly higher than the actual DUT’s CI value. To accurately measure the CI value of the high-Q resonator whose response speed is generally not so fast, we need to make the analyzer’s sweep speed slow enough and take a lot of measurement points around the Fr area. To measure CI more accurately, the measurement shown on the right hand side is performed with a narrower span just focusing on the Fr area. In this case, the CI value is correctly measured. But note that the equivalent circuit parameter C0 can be erroneous because the antiresonant area is not included in the measurement span. 30 To evaluate components under their actual operating conditions, we often want to measure their impedance by applying the DC bias. This is a measurement example of a ceramic capacitor which has a strong DC voltage dependency. The measurement performed in the channel-1 is the frequency sweep for |Z|, phase, Cs, and Rs. And the measurement performed in the channel-2 is the DC bias sweep (-10 Vdc to +10 Vdc) for Cs by using the E5061B’s internal DC source. The DC voltage dependency of the capacitor is nicely measured with the DC bias sweep. 31 To evaluate the DC current dependency of power inductors and ferrite beads, we need to apply a large DC current that cannot be provided by the E5061B’s internal DC source. This slide shows a solution example using the 16200B DC current bias adapter (1 MHz to 1 GHz, up to 5 Adc) and an external DC source. 32 Also, we sometimes want to apply large AC signals to high-power components such as ultrasound resonators, ceramic actuators, and power inductors. Similarly to the impedance analyzers of the auto balance bridge method, we can expand the E5061B’s source output level using an external power amplifier as shown in this block diagram. The measurement configuration is the series-thru connection, but we set the input impedance of the receivers to 1 Mohm and connect external homemade voltage divider circuits to prevent excessive AC voltage inputs. The large current flowing through the DUT is sensed by connecting an external high-power resistor Rc with the resistance value around 1 or 10 ohm. 33 This is a measurement example of a power inductor by applying a large AC signal using the power amplifier. The power sweep is performed from -20 to + 8 dBm, and the source output is amplified with a gain of 50. In this measurement example, about 1 Arms current is being applied to the DUT at +7 dBm source output point. 34 The last examples are in-circuit impedance measurements. This example is an in-circuit impedance measurement with hand probing. Typical target applications are in-circuit impedance measurements in MHz ranges, such as printed RFID antennas, negative impedance of oscillator circuits, and so on. The measurement method is the reflection method. The probe consists of a semi-flexible SMA cable with a nonmetal coating and the probe head of the 42941A impedance probe. The screen shot shows a measurement example of a simple LC parallel resonant circuit using this probe. The DUT’s resonant response is nicely measured. 35 The next example is an output impedance measurement of a DC-DC converter, which is a typical measurement application in the power integrity field. The gain-phase shunt-thru method is used for measuring very small impedance in the low frequency range. The receiver input impedance is 50 ohm in this configuration. But if the DUT’s output voltage is less than 5 Vdc, we can directly connect the converter without using DC blocking capacitors. The screen shot shows the measurement result. The test frequency range is 10 Hz to 10 MHz. As you can see, the output impedance in the low frequency range is suppressed to milliohm order due to the converter’s feedback loop operation. 36 The last example is also a power-integrity application. The DUT is a MPU board populated with bypass capacitors, and its PDN (Power Distribution Network) impedance is measured by using the shunt-thru method at the S-parameter test port. In this example, two SMA connectors are soldered to a thru hole from the top and bottom sides of the PCB to avoid the additive inductance caused by the inter-probe coupling (see Appendix). The screen shot shows a measurement result in the wide frequency range from 100 Hz to 1 GHz. As you can see, the PDN’s low impedance is measured up to hundreds of MHz without being affected by the additive inductance error. 37 38 39 One problem of the shunt-thru method using conventional LF VNAs is the measurement error associated with the test cable ground loop. In this block diagram, ideally the source signal flowing through the DUT should return to the source side through the outer shield of the cable as shown in the blue dotted line. But actually, in the low freq range below 100 kHz, the source current also flows into the test cable shield of receiver side, as shown in the red dotted line. The shield current flowing into the receiver side will cause the error voltage across the cable outer shield resistance, and this will cause the measurement error. It is not possible to measure the DUT’s impedance if it is smaller than the test cable’s shield resistance, which is typically around 10 mohm or more. 40 The conventional solution for this problem is to attach magnetic cores to the test cables at either of the source side or receiver side. The shield impedance caused by the magnetic core’s self inductance will reduce the source current flowing into the shield of receiver side. Also, the core attached at the source side will force the source current returns through the shield, back to the source side. However, it is actually not so easy to implement this configuration, because we need a magnetic core that has very large inductance, and it is generally not so easy to find a good core that has sufficiently large inductance down to very low freq range. Therefore, a more convenient solution has been desired. 41 Now the E5061B’s gain-phase test port offers a quite different approach to solve this problem. The receivers of gain-phase test port have a semi-floating impedance, Zg, to the chassis ground. This floating impedance is about 30 ohm in the low freq range below 100 kHz, which is sufficiently larger than cable shield resistance. So the error voltage across cable shield resistance can be ignored. As a result, the receiver can correctly monitor the DUT’s output voltage, Vo. And the DUT’s milliohm impedance can be accurately measured. With this unique receiver architecture, we can easily and accurately measure milliohm impedance of DC-DC converters and passive PDNs populated with bulk bypass capacitors at low frequencies, even without using external magnetic cores. 42 43