Figure 1: The front face of the oscilloscope Cursor Channel 1 Position Power Button Cursor Run/ Movements Stop Channel 1 Menu Channel 1 VOLTS/DIV Channel 1 Input Cables Trigger Menu SEC/DIV As part of the coupled oscillator and resonance experiments, you will be using an oscilloscope. These devices automatically plot voltage versus time over and over again on a small screen. Oscilloscopes are especially useful for observing periodic behaviour, as in the case of the coupled oscillator experiment. They are also useful for plotting transient behaviour as in the resonance experiment. These instructions will tell you how to start the oscilloscopes and set them up for the measurements you are making today. You will use oscilloscopes again many times; they are very useful tools once you get to be expert with using them. 1. Attach the appropriate cables to the inputs for channel 1 on the oscilloscope, as shown in figure 1. 2. If the oscilloscope is off, turn the power on, using the power button, on the top of the oscilloscope located as shown in figure 1. When prompted, press any button, once. 3. Turn the Ch. 1 vertical position adjustment knob. A reading will appear at the bottom of the screen, telling you the offset of the Ch. 1 trace. The message at the bottom of the screen can be seen in figure 3. Continue to adjust the knob until the offset is zero. 1 Figure 2: While the vertical offset is being adjusted, it states how far offset the signal is. Adjust until this is 0 2 Figure 3: The Channel 1 menu screen. Coupling toggle BW Limit select Coarse/Fine vertical scale adjust toggle Channel 1 menu button Probe select 4. Press the ch. 1 MENU button. On the right side of the screen, a menu appears, like in figure 4. You can adjust the settings by pressing the unlabelled buttons immediately to the right of the menu items on the screen. Set them so that they match those in figure 4, (a) Coupling should be set to AC (b) BW Limit should be set to OFF (c) Volts/Div should be set to COARSE (d) Probe should be set to 1X 5. Press the TRIGGER MENU button. On the right side of the screen a menu appears, like in figure 5. Once again, you adjust the settings using the unlabelled buttons immediately to the right of the menu items on the screen. Set these so that they match those in figure 5. (a) EDGE should be highlighted 3 Figure 4: The Trigger menu screen. Edge/Video toggle Slope adjust Source select Mode select Coupling adjust (b) Slope should be set to RISING (c) Source should be set to Ch. 1 (d) Mode should be set to Auto (e) Coupling should be set to AC 6. The number of volts that each vertical grid marking represents is written in the bottom left-hand corner of the screen. Adjust the channel 1 VOLT/DIV knob (see figure 1) counter-clockwise until each division is 5 V. 7. The number of seconds each horizontal grid marking represents is written just to the right of the centre in the bottom of the screen. Adjust the SEC/DIV knob (see figure 1) clockwise until each division is 50 ms. 8. **** THIS SHOULD HAVE THE PICTURES OF “GOOD” TRACES AND “BAD” TRACES ****** 4 9. To determine frequency, make measurements of period using the cursor feature of the oscilloscope. (a) To begin, when the trace on the scope looks like what you want to be measuring, press the RUN/STOP button (see figure 1). Up near the centre, at the top of the screen, a black dot and the word STOP will appear. If the trace isn’t quite what you want, press the RUN/STOP button again, adjust things until the trace is what you want and then press the RUN/STOP button again. (b) Press the cursor button (see figure 1). Two vertical lines appear on the screen. On the right side of the screen a menu will appear. As always, you use the unlabelled buttons immediately to the right of the screen to change those menu settings. Set the first two so that they match those in figure 9. i. Type should be set to TIME ii. Source should be set to Ch. 1 iii. Delta is the difference between the two vertical lines: Cursor 1 - Cursor 2 = Delta iv. Cursor 1 reads where the first vertical line is positioned v. Cursor 2 reads where the second vertical line is positioned (c) Move cursor 1 to intersect a peak in your trace. You move cursor 1 by turning the channel 1 vertical position knob (see figure 1). (d) Move cursor 2 to intersect a neighbouring peak in your trace. You move cursor 2 by turning the channel 2 vertical position knob (see figure 1). (e) Read the difference in time between the two peaks from the menu item Delta. (f) When you have completed using the cursor, press the RUN/STOP button again to start acquiring fresh data again. 5 Figure 5: While the cursor menu is active, Delta indicates the separation between the two cursors. Type toggle Source select 6