The 5520A/5500A Multi-Product Calibrators Extended Specifications December 1997 General Specifications 5520A/5500A 2 Fluke Corporation Warmup Time Twice the time since last warmed up, to a maximum of 30 minutes Settling Time Less than 5 seconds for all functions and ranges except as noted Standard Interfaces IEEE-488 (GPIB), RS-232, 5725A (5500A only) Temperature Performance Operating: 0°C to 50°C Calibration (tcal): 15°C to 35°C Storage: -20°C to 70°C Temperature Coefficient Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C Relative Humidity Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C Storage: <95%, noncondensing Altitude Operating: 3,050m (10,000 ft) maximum Non-operating: 12,200m (40,000 ft) maximum Safety Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92 Analog Low Isolation 20V EMC 5520A: Designed to comply with IEC 61326-1/1997 5500A: Designed to comply with FCC Rules Part 15 Line Power Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: ± 10% about line voltage setting Power Consumption 5520A: 600 VA 5500A: 300 VA Dimensions Height:17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit Width: 43.2 cm (17 in), standard rack width Depth: 47.3 cm (18.6 in) overall Weight (without options) 5520A: 22 kg (49 lb) 5500A: 20 kg (44 lb) Absolute Uncertainty Definition 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated. Specification Confidence Interval >99% 5520A/5500A Multi-Product Calibrators Extended Specifications DC Voltage Specifications 5520A Absolute Uncertainty, tcal ± 5°C ± (ppm of output + µV) 90 Days 1 Year 24 Hours, ± 1°C ± (ppm of output + µV) 20 + 1.5 25 + 1.5 3+1 Ranges 0 to 329.9999 mV Stability Resolution Maximum Burden Current1 0.1 µV 50Ω 0 to 3.299999V 9+2 11 + 2 2 + 1.5 1 µV 10 mA 0 to 32.99999V 10 + 15 12 + 15 2 + 10 10 µV 10 mA 30 to 329.9999V 15 + 150 18 + 150 2.5 + 100 100 µV 5 mA 100 to 1000V 15 + 1500 18 + 1500 3 + 300 1000 µV 5 mA Auxiliary Output (dual output mode only)2 0 to 329.999 mV 300 + 350 400 + 350 30 + 100 1 µV 5 mA 0.33 to 3.29999V 300 + 350 400 + 350 30 + 100 10 µV 5 mA 3.3 to 7V 300 + 350 400 + 350 30 + 100 100 µV 5 mA 5+2 0.1 µV 10Ω 24 Hours, ± 1°C ± (ppm of output + µV) Resolution Maximum Burden Current1 TC Output and Measure in Linear 10 µV/°C and 1 mV/°C Modes 0 to 329.999 mV 40 + 3 50 + 3 5500A Absolute Uncertainty, tcal ± 5°C ± (ppm of output + µV) Ranges 90 Days 1 Year Stability 0 to 329.9999 mV 50 + 3 60 + 3 5+1 0.1 µV 50Ω 0 to 3.299999V 40 + 5 50 + 5 4+3 1 µV 10 mA 0 to 32.99999V 40 + 50 50 + 50 4 + 30 10 µV 10 mA 30 to 329.9999V 45 + 500 55 + 500 4.5 + 300 100 µV 5 mA 100 to 1000V 45 + 1500 55 + 1500 4.5 + 900 1000 µV 5 mA Auxiliary Output (dual output mode only)2 0 to 329.999 mV 300 + 350 400 + 350 30 + 100 1 µV 5 mA 0.33 to 3.29999V 300 + 350 400 + 350 30 + 100 10 µV 5 mA 60 + 3 5+1 0.1 µV 10Ω TC Output and Measure in Linear 10 µV/°C 0 to 329.999 mV 50 + 3 Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33V. The AUX output has an output resistance of <1Ω. TC output has an output impedance of 10Ω ± 1Ω. 2 Two channels of dc voltage output are provided. Note: Maximum voltage output is 1020V. 1 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 3 DC Current Specifications 5520A Ranges Resolution Compliance Voltage Maximum Inductive Load 90 Days 1 Year 0 to 329.999 µA 120 + 0.02 150 + 0.02 1 nA 10V 400 µH 0 to 3.29999 mA 80 + 0.03 100 + 0.03 0.01 µA 10V 400 µH 0 to 32.9999 mA 80 + 0.2 100 + 0.2 0.1 µA 7V 400 µH 0 to 329.999 mA 80 + 2 100 + 2 1 µA 7V 400 µH 160 + 40 200 + 40 10 µA 6V 400 µH 0 to 2.99999A 1 Absolute Uncertainty, tcal ± 5°C ± (ppm of output + µA) 0 to 10.9999A 380 + 330 500 + 330 100 µA 4V 400 µH 11 to 20.5A1 800 + 750 1000 + 750 100 µA 4V 400 µH Resolution Compliance Voltage Maximum Inductive Load At 20°C, full 20.5A output is available continuously for 20 minutes. 5500A Ranges Absolute Uncertainty, tcal ± 5°C ± (% of output + µA) 90 Days 1 Year 0 to 3.29999 mA 0.010 + 0.05 0.013 + 0.05 0.01 µA 4.5V 1 µH 0 to 32.9999 mA 0.008 + 0.25 0.01 + 0.25 0.1 µA 4.5V 200 µH 0 to 329.999 mA 0.008 + 3.3 0.01 + 3.3 1 µA 3.0V 200 µH 0 to 2.19999A 0.023 + 44 0.03 + 44 10 µA 3.4V 200 µH 0 to 10.9999A 0.038 + 330 0.06 + 330 100 µA 2.5V 200 µH 0.03 + 330 0.04 + 330 100 µA 4V 400 µH 5725A Amplifier 0 to 11A 4 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications Resistance Specifications 5520A Absolute Uncertainty, tcal ± 5°C ± (ppm of output + floor)2 Ranges 1 Floor Time, Temperature Since Ω Zero ppm of output 12 Hours ± 1°C Resolution Allowable Current 7 Days ± 5°C 90 Days 1 Year 0 to 10.9999Ω 35 40 0.001Ω 0. 01Ω 0.0001Ω 1 to 125 mA 11 to 32.9999Ω 25 30 0.0015Ω 0.015Ω 0.0001Ω 1 to 125 mA 33 to 109.9999Ω 22 28 0.0015Ω 0.015Ω 0.0001Ω 1 to 70 mA 110 to 329.9999Ω 22 28 0.002Ω 0.02Ω 0.0001Ω 1 to 40 mA 330Ω to 1.099999 kΩ 22 28 0.002Ω 0.02Ω 0.001Ω 250 µA to 18 mA 1.1 to 3.299999 kΩ 22 28 0.02Ω 0.2Ω 0.001Ω 250 µA to 5 mA 3.3 to 10.99999 kΩ 22 28 0.02Ω 0.1Ω 0.01Ω 25 µA to 1.8 mA 11 to 32.99999 kΩ 22 28 0.2Ω 1Ω 0.01Ω 25 µA to 0.5 mA 33 to 109.9999 kΩ 22 28 0.2Ω 1Ω 0.1Ω 2.5 µA to 0.18 mA 110 to 329.9999 kΩ 25 32 2Ω 10Ω 0.1Ω 2.5 µA to 0.05 mA 330 kΩ to 1.099999 MΩ 25 32 2Ω 10Ω 1Ω 250 nA to 0.018 mA 1.1 to 3.299999 MΩ 40 60 30Ω 150Ω 1Ω 250 nA to 5 µA 3.3 to 10.99999 MΩ 110 130 50Ω 250Ω 10Ω 25 nA to 1.8 µA 11 to 32.99999 MΩ 200 250 2500Ω 2500Ω 10Ω 25 nA to 500 nA 33 to 109.9999 MΩ 400 500 3000Ω 3000Ω 100Ω 2.5 nA to 180 nA 110 to 329.999 MΩ 2500 3000 100000Ω 100000Ω 1000Ω 2.5 nA to 50 nA 330 to 1100 MΩ 12000 15000 500000Ω 500000Ω 10000Ω 1 nA to 13 nA 1 Continuously variable from 0 to 1.1 GΩ. 2 Applies for 4-WIRE compensation only. For COMP OFF and 2-WIRE COMP, add 5 µV per amp of stimulus current to the floor specification. For example, in 2-wire mode, at 1 kΩ, the floor specification within 12 hours of an ohms zero cal for a measurement current of 250 µA is 0.002Ω + 5 µV/250 µA = 0.022Ω. 5500A Ranges 1 Absolute Uncertainty, tcal ± 5°C ± (% of output + Ω) 2 90 Days Resolution Allowable Current 1 Year 0.009 + 0.008 3 0.012 + 0.008 3 0.001Ω 1 to 125 mA 11 to 32.999Ω 0.009 + 0.0153 0.012 + 0.0153 0.001Ω 1 to 125 mA 33 to 109.999Ω 0.007 + 0.0153 0.009 + 0.0153 0.001Ω 1 to 70 mA 110 to 329.999Ω 0.007 + 0.0153 0.009 + 0.0153 0.001Ω 1 to 40 mA 330Ω to 1.09999 kΩ 0.007 + 0.06 0.009 + 0.06 0.01Ω 250 µA to 18 mA 1.1 to 3.29999 kΩ 0.007 + 0.06 0.009 + 0.06 0.01Ω 250 µA to 5 mA 3.3 to 10.9999 kΩ 0.007 + 0.6 0.009 + 0.6 0.1Ω 25 µA to 1.8 mA 11 to 32.9999 kΩ 0.007 + 0.6 0.009 + 0.6 0.1Ω 25 µA to 0.5 mA 33 to 109.999 kΩ 0.008 + 6 0.011 + 6 1Ω 2.5 µA to 0.18 mA 110 to 329.999 kΩ 0.009 + 6 0.012 + 6 1Ω 2.5 µA to 0.05 mA 330 kΩ to 1.09999 MΩ 0.011 + 55 0.015 + 55 10Ω 250 nA to 0.018 mA 1.1 to 3.29999 MΩ 0.011 + 55 0.015 + 55 3.3 to 10.9999 MΩ 0.045 + 550 11 to 32.9999 MΩ 0.075 + 550 0 to 10.999Ω 0.06 + 550 10Ω 250 nA to 5 µA 100Ω 25 nA to 1.8 µA 0.1 + 550 100Ω 25 nA to 500 nA 33 to 109.999 MΩ 0.4 + 5500 0.5 + 5500 1000Ω 2.5 nA to 180 nA 110 to 329.999 MΩ 0.4 + 16500 0.5 + 16500 1000Ω 2.5 nA to 60 nA 1 Continuously variable from 0 to 330 MΩ. 2 Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation. 3 The floor added is improved to 0.006Ω (0-10.99Ω range) and 0.010Ω (11-329.999Ω range) if the 5500A is zeroed within 8 hours and temperature is ± 1°C of zeroing ambient temperature. 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 5 AC Voltage (Sinewave) Specifications 5520A Ranges Frequency Absolute Uncertainty, tcal ± 5°C ± (ppm of output + µV) 90 Days 1.0 to 32.999 mV 10 to 45 Hz 45 Hz to 10 kHz 10 to 20 kHz 20 to 50 kHz 50 to 100 kHz 100 to 500 kHz 33 to 329.999 mV 10 to 45 Hz 45 Hz to 10 kHz 10 to 20 kHz 20 to 50 kHz 50 to 100 kHz 100 to 500 kHz 0.33 to 3.29999V 10 to 45 Hz 45 Hz to 10 kHz 10 to 20 kHz 20 to 50 kHz 50 to 100 kHz 100 to 500 kHz 3.3 to 32.9999V 10 to 45 Hz 45 Hz to 10 kHz 10 to 20 kHz 20 to 50 kHz 50 to 100 kHz 33 to 329.999V 45 Hz to 1 kHz 1 to 10 kHz 10 to 20 kHz 20 to 50 kHz 50 to 100 kHz 330 to 1000V 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz Auxiliary Output (dual output mode only)2 10 to 329.999 mV 10 to 20 Hz 20 to 45 Hz 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz 10 to 30 kHz 0.33 to 3.29999V 10 to 20 Hz 20 to 45 Hz 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz 10 to 30 kHz 3.3 to 5V 10 to 20 Hz 20 to 45 Hz 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz Resolution Maximum Burden 1 1 µV 50Ω 1 µV 50Ω 10 µV 10 mA 100 µV 10 mA 1 mV 5 mA, except 20 mA for 45 to 65 Hz 10 mV 2 mA, except 6 mA for 45 to 65 Hz 1 µV 5 mA 10 µV 5 mA 100 µV 5 mA 1 Year 600 120 160 800 3000 6000 400 110 120 300 600 1600 250 100 160 250 550 2000 250 125 220 300 750 150 160 220 600 1600 250 200 250 + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 6 6 6 6 12 50 8 8 8 8 32 70 50 25 50 50 125 600 650 200 600 600 1600 2000 6000 6000 20000 50000 10000 10000 10000 800 150 200 1000 3500 8000 500 130 150 350 800 2000 300 120 190 300 700 2400 300 150 240 350 900 190 200 250 800 2000 300 250 300 + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 6 6 6 6 12 50 8 8 8 8 32 70 50 25 50 50 125 600 650 200 600 600 1600 2000 6000 6000 20000 50000 10000 10000 10000 0.15% 0.08% 0.08% 0.15% 0.3% 0.6% 0.15% 0.08% 0.07% 0.15% 0.3% 0.6% 0.15% 0.08% 0.07% 0.15% 0.3% + + + + + + + + + + + + + + + + + 370 370 370 450 450 900 450 450 450 1400 1400 2800 450 450 450 1400 1400 0.2% 0.1% 0.1% 0.2% 0.4% 0.8% 0.2% 0.1% 0.09% 0.2% 0.4% 0.8% 0.2% 0.1% 0.09% 0.2% 0.4% + + + + + + + + + + + + + + + + + 370 370 370 450 450 900 450 450 450 1400 1400 2800 450 450 450 1400 1400 Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33V. The AUX output resistance is <1Ω. The maximum load capacitance is 500 pF, subject to the maximum burden current limits. 2 There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz for the 5520A. Note: Maximum voltage output is 1020V. 1 6 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications 5500A Ranges Frequency Absolute Uncertainty, tcal ± 5°C ± (% of output + µV) 90 Days 1.0 to 32.999 mV 33 to 329.999 mV 0.33 to 3.29999V 3.3 to 32.9999V 33 to 329.999V 330 to 1000V 5725 Amplifier 10 to 329.999 mV 10 45 Hz 10 20 50 100 10 45 Hz 10 20 50 100 10 45 Hz 10 20 50 100 10 45 Hz 10 20 50 45 Hz 1 10 45 Hz 1 5 to to to to to to to to to to to to to to to to to to to to to to to to to to to to to 45 Hz 10 kHz 20 kHz 50 kHz 100 kHz 500 kHz 45 Hz 10 kHz 20 kHz 50 kHz 100 kHz 500 kHz 45 Hz 10 kHz 20 kHz 50 kHz 100 kHz 500 kHz 45 Hz 10 kHz 20 kHz 50 kHz 100 kHz 1 kHz 10 kHz 20 kHz 1 kHz 5 kHz 10 kHz 45 Hz to 1 kHz 1 to 20 kHz 20 to 30 kHz 100 to 750V 30 to 100 kHz Auxiliary Output (dual output mode only)2 10 to 329.999 mV 10 to 20 Hz 20 to 45 Hz 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz 0.33 to 3.29999V 10 to 20 Hz 20 to 45 Hz 45 Hz to 1 kHz 1 to 5 kHz 5 to 10 kHz Resolution Maximum Burden 1 1 µV 50Ω 1 µV 50Ω 10 µV 10 mA 100 µV 10 mA 1 mV 5 mA, except 20 mA for 45 to 65 Hz 10 mV 2 mA, except 6 mA for 45 to 65 Hz 1 Year 0.26 0.11 0.15 0.19 0.26 0.75 0.19 0.04 0.08 0.12 0.17 0.53 0.11 0.02 0.06 0.1 0.17 0.38 0.11 0.03 0.06 0.14 0.17 0.04 0.06 0.07 0.04 0.15 0.15 + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 20 20 20 20 33 60 50 20 20 40 170 330 250 60 60 300 1700 3300 2500 600 2600 5000 17000 6.6 mV 15 mV 33 mV 80 mV 100 mV 500 mV 0.35 0.15 0.2 0.25 0.35 1 0.25 0.05 0.1 0.16 0.24 0.7 0.15 0.03 0.08 0.14 0.24 0.5 0.15 0.04 0.08 0.19 0.24 0.05 0.08 0.09 0.05 0.20 0.20 + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 20 20 20 20 33 60 50 20 20 40 170 330 250 60 60 300 1700 3300 2500 600 2600 5000 17000 6.6 mV 15 mV 33 mV 80 mV 100 mV 500 mV 0.04 0.06 0.08 0.38 + + + + 80 mV 100 mV 100 mV 500 mV 0.05 0.08 0.1 0.5 + + + + 80 mV 100 mV 100 mV 500 mV 0.15 0.08 0.08 0.15 0.3 0.15 0.08 0.07 0.15 0.3 + + + + + + + + + + 370 370 370 450 450 450 450 450 1400 1400 0.2 0.1 0.1 0.2 0.4 0.2 0.1 0.09 0.2 0.4 + + + + + + + + + + 370 370 370 450 450 450 450 450 1400 1400 10 µV 50 mA 70 mA 70 mA 70 mA 1 µV 5 mA 10 µV 5 mA 10 mV Remote sensing is not provided. Output resistance is <5 mΩ for outputs ≥0.33V. The AUX output resistance is <1Ω. The maximum load capacitance is 500 pF, subject to the maximum burden current limits. 2 There are two channels of voltage output. The maximum frequency of the dual output is 10 kHz for the 5500A. Note: Maximum voltage output is 1020V. 1 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 7 AC Current (Sinewave) Specifications 5520A (LCOMP Off) Ranges Absolute Uncertainty, tcal ± 5°C ± (% of output + µA) Frequency 90 Days 29.00 to 329.99 µA 10 to 20 Hz 0.16% 0.1 µA 0.2% 0.1 µA 20 to 45 Hz 0.12% 0.1 µA 0.15% 0.1 µA 0.1% 0.1 µA 0.125% 0.1 µA 1 to 5 kHz 0.25% 0.15 µA 0.3% 0.15 µA 45 Hz to 1 kHz 0.33 to 3.2999 mA 3.3 to 32.999 mA 33 to 329.99 mA 0.33 to 2.99999A 3 to 10.9999A 5 to 10 kHz 0.6% 0.2 µA 0.8% 0.2 µA 10 to 30 kHz 1.2% 0.4 µA 1.6% 0.4 µA 10 to 20 Hz 0.16% 0.15 µA 0.2% 0.15 µA 20 to 45 Hz 0.1% 0.15 µA 0.125% 0.15 µA 45 Hz to 1 kHz 0.08% 0.15 µA 0.1% 0.15 µA 1 to 5 kHz 0.16% 0.2 µA 0.2% 0.2 µA 5 to 10 kHz 0.4% 0.3 µA 0.5% 0.3 µA 10 to 30 kHz 0.8% 0.6 µA 1.0% 0.6 µA 10 to 20 Hz 0.15% 2 µA 0.18% 2 µA 20 to 45 Hz 0.075% 2 µA 0.09% 2 µA 45 Hz to 1 kHz 0.035% 2 µA 0.04% 2 µA 1 to 5 kHz 0.065% 2 µA 0.08% 2 µA 3 µA 5 to 10 kHz 0.16% 3 µA 0.2% 10 to 30 kHz 0.32% 4 µA 0.4% 4 µA 10 to 20 Hz 0.15% 20 µA 0.18% 20 µA 20 to 45 Hz 0.075% 20 µA 0.09% 20 µA 45 Hz to 1 kHz 0.035% 20 µA 0.04% 20 µA 1 to 5 kHz 0.08% 50 µA 0.10% 50 µA 5 to 10 kHz 0.16% 100 µA 0.2% 100 µA 10 to 30 kHz 0.32% 200 µA 0.4% 200 µA 10 to 45 Hz 0.15% 100 µA 0.18% 100 µA 45 Hz to 1 kHz 0.05% 100 µA 0.06% 100 µA 1 to 5 kHz 2 0.5% 1000 µA 0.6% 1000 µA 5 to 10 kHz 3 2.0% 5000 µA 2.5% 5000 µA 45 to 100 Hz 100 to 1 kHz 1 to 5 kHz 11 to 20.5A1 0.05% 2000 µA 0.06% 2000 µA 0.08% 2000 µA 0.10% 2000 µA 2.0% 2000 µA 2.5% 2000 µA 0.1% 5000 µA 0.12% 5000 µA 100 Hz to 1 kHz 0.13% 5000 µA 0.15% 5000 µA 1 to 5 kHz 2.0% 5000 µA 2.5% 5000 µA 45 to 100 Hz 1 At 20°C, full 20.5A output is available continuously for 20 minutes. 2 For compliance voltages greater than 1V, add 1 mA/V to the floor specification from 1 to 5 kHz. 3 For compliance voltages greater than 1V, add 5 mA/V to the floor specification from 5 to 10 kHz. 8 Fluke Corporation Maximum Inductive Load 1 Year 5520A/5500A Multi-Product Calibrators Extended Specifications 200 µH 200 µH 50 µH 50 µH 2.5 µH 1 µH 1 µH 5520A (LCOMP On) Ranges Absolute Uncertainty, tcal ± 5°C ± (% of output + µA) Frequency 90 Days 29.00 to 329.99 µA 10 to 100 Hz 100 Hz to 1 kHz 0.33 to 3.2999 mA 10 to 100 Hz 100 Hz to 1 kHz 3.3 to 32.999 mA 10 to 100 Hz 100 Hz to 1 kHz 33 to 329.99 mA 10 to 100 Hz 100 Hz to 1 kHz 0.33 to 2.99999A 3 to 20.5A1 Maximum Inductive Load 1 Year 0.2% 0.2 µA 0.25% 0.2 µA 0.5% 0.5 µA 0.6% 0.5 µA 0.2% 0.3 µA 0.25% 0.3 µA 0.5% 0.8 µA 0.6% 0.8 µA 0.07% 4 µA 0.08% 4 µA 0.18% 10 µA 0.2% 10 µA 0.07% 40 µA 0.08% 40 µA 0.18% 100 µA 0.2% 100 µA 10 to 100 Hz 0.1% 200 µA 0.12% 200 µA 100 to 440 Hz 0.25% 1000 µA 0.3% 1000 µA 45 to 100 Hz 0.1% 2000 µA 0.12% 2000 µA 100 to 440 Hz 0.8% 5000 µA 1.0% 5000 µA 400 µH 400 µH 400 µH 400 µH 400 µH 400 µH2 1 At 20°C, full 20.5A output is available continuously for 20 minutes. 2 Subject to compliance voltage limit. 5520A Resolution and Compliance Voltage Resolution Maximum Compliance Voltage 0.029 to 0.32999 mA 0.01 µA 7V rms 0.33 to 3.29999 mA 0.01 µA 7V rms 3.3 to 32.9999 mA 0.1 µA 5V rms 33 to 329.999 mA 1 µA 5V rms 0.33 to 2.99999A 10 µA 4V rms1 3 to 20.5A 100 µA 3V rms Ranges 1 Subject to specification adder for compliance voltages greater than 1V rms (previous page). 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 9 AC Current (Sinewave) Specifications 5500A Ranges Frequency Absolute Uncertainty, tcal ± 5°C ± (% of output + µA) 90 Days 29.00 to 329.99 µA 0.33 to 3.2999 mA 3.3 to 32.999 mA 33 to 329.99 mA 0.33 to 2.19999A 2.2 to 11A Compliance Voltage Maximum Inductive Load 0.01 µA 3V rms 1 µH 0.01 µA 3V rms 1 µH 0.1 µA 3V rms 200 µH, 10 to 500 Hz 1 Year 10 to 20 Hz 0.19 + 0.15 0.25 + 0.15 20 to 45 Hz 0.09 + 0.15 0.125 + 0.15 45 Hz to 1 kHz 0.09 + 0.25 0.125 + 0.25 1 to 5 kHz 0.30 + 0.15 0.4 + 0.15 5 to 10 kHz 0.94 + 0.15 1.25 + 0.15 10 to 20 Hz 0.15 + 0.3 0.2 + 0.3 20 to 45 Hz 0.08 + 0.3 0.1 + 0.3 45 Hz to 1 kHz 0.08 + 0.3 0.1 + 0.3 1 to 5 kHz 0.15 + 0.3 0.2 + 0.3 5 to 10 kHz 0.45 + 0.3 0.6 + 0.3 10 to 20 Hz 0.15 + 3 0.2 + 3 20 to 45 Hz 0.08 + 3 0.1 + 3 45 Hz to 1 kHz 0.07 + 3 0.09 + 3 1 to 5 kHz 0.15 + 3 0.2 + 3 5 to 10 kHz 0.45 + 3 0.6 + 3 10 to 20 Hz 0.15 + 30 0.2 + 30 20 to 45 Hz 0.08 + 30 0.1 + 30 45 Hz to 1 kHz 0.07 + 30 0.09 + 30 1 to 5 kHz 0.15 + 30 0.2 + 30 5 to 10 kHz 0.45 + 30 0.6 + 30 10 to 45 Hz 0.15 + 300 0.2 + 300 45 Hz to 1 kHz 0.08 + 300 0.1 + 300 1 to 5 kHz 0.07 + 300 0.75 + 300 45 to 65 Hz 0.05 + 2000 0.06 + 2000 65 to 500 Hz 0.08 + 2000 0.1 + 2000 0.25 + 2000 0.33 + 2000 500 Hz to 1 kHz Resolution 1 µH, 500 Hz to 10 kHz 1 µA 3V rms 200 µH, 10 to 500 Hz 5 µH, 500 Hz to 10 kHz 10 µA 3V rms 200 µH, 45 to 500 Hz 5 µH, 500 Hz to 5 kHz 100 µA 2.8V rms 200 µH, 45 to 65 Hz 1 µH, 65 Hz to 1 kHz 5725A Amplifier 1.5 to 11A 10 Fluke Corporation 45 Hz to 1 kHz 0.08 + 100 0.01 + 100 1 to 5 kHz 0.19 + 5000 0.25 + 5000 5 to 10 kHz 0.75 + 10000 1 + 10000 5520A/5500A Multi-Product Calibrators Extended Specifications 100 µA 3V rms 400 µH Capacitance Specifications 5520A Absolute Uncertainty, tcal ± 5°C ± (% of output + nF) Ranges Allowed Frequency or Charge-Discharge Rate Resolution 90 Days 1 Year To Meet Specification Typical for <0.5% Error 0.19 to 0.3999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 10 Hz to 10 kHz 20 kHz 0.4 to 1.0999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 10 Hz to 10 kHz 30 kHz 1.1 to 3.2999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 10 to 3000 Hz 30 kHz 3.3 to 10.9999 nF 0.19 + 0.01 0.25 + 0.01 0.1 pF 10 to 1000 Hz 20 kHz 11 to 32.9999 nF 0.19 + 0.1 0.25 + 0.1 0.1 pF 10 to 1000 Hz 8 kHz 33 to 109.999 nF 0.19 + 0.1 0.25 + 0.1 1 pF 10 to 1000 Hz 4 kHz 110 to 329.999 nF 0.19 + 0.3 0.25 + 0.3 1 pF 10 to 1000 Hz 2.5 kHz 0.33 to 1.09999 µF 0.19 + 1 0.25 + 1 10 pF 10 to 600 Hz 1.5 kHz 1.1 to 3.29999 µF 0.19 + 3 0.25 + 3 10 pF 10 to 300 Hz 800 Hz 3.3 to 10.9999 µF 0.19 + 10 0.25 + 10 100 pF 10 to 150 Hz 450 Hz 11 to 32.9999 µF 0.30 + 30 0.40 + 30 100 pF 10 to 120 Hz 250 Hz 33 to 109.999 µF 0.34 + 100 0.45 + 100 1 nF 10 to 80 Hz 150 Hz 110 to 329.999 µF 0.34 + 300 0.45 + 300 1 nF To 50 Hz 80 Hz 0.33 to1.09999 mF 0.34 + 1 µF 0.45 + 1 µF 10 nF To 20 Hz 45 Hz 1.1 to 3.2999 mF 0.34 + 3 µF 0.45 + 3 µF 10 nF To 6 Hz 30 Hz 3.3 to 10.9999 mF 0.34 + 10 µF 0.45 + 10 µF 100 nF To 2 Hz 15 Hz 11 to 32.9999 mF 0.7 + 30 µF 0.75 + 30 µF 100 nF To 0.6 Hz 7.5 Hz 33 to 110 mF 1.0 + 100 µF 10 µF To 0.2 Hz 3 Hz 1.1 + 100 µF Notes 1. The output is continuously variable from 190 pF to 110 mF. 2. Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. 3. The maximum peak voltage allowable is 3V. The maximum allowable peak current is 150 mA. 4. The maximum lead resistance for no additional error in 2-wire COMP mode is 10 ohms. 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 11 Capacitance Specifications 5500A Absolute Uncertainty, tcal ± 5°C ± (% of output + nF) Ranges Allowed Frequency or Charge-Discharge Rate Resolution 90 Days 1 Year Typical for <5% Error 0.33 to 0.4999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 50 to 1000 Hz 10 kHz 0.5 to 1.0999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 50 to 1000 Hz 10 kHz 1.1 to 3.2999 nF 0.38 + 0.01 0.5 + 0.01 0.1 pF 50 to 1000 Hz 10 kHz 3.3 to 10.9999 nF 0.38 + 0.01 0.5 + 0.01 1 pF 50 to 1000 Hz 10 kHz 11 to 32.9999 nF 0.19 + 0.1 0.25 + 0.1 1 pF 50 to 1000 Hz 10 kHz 33 to 109.999 nF 0.19 + 0.1 0.25 + 0.1 10 pF 50 to 1000 Hz 10 kHz 110 to 329.999 nF 0.19 + 0.3 0.25 + 0.3 10 pF 50 to 1000 Hz 10 kHz 0.33 to 1.09999 µF 0.19 + 1 0.25 + 1 100 pF 50 to 1000 Hz 5 kHz 1.1 to 3.29999 µF 0.26 + 3 0.35 + 3 100 pF 50 to 1000 Hz 2 kHz 3.3 to 10.9999 µF 0.26 + 10 0.35 + 10 1 nF 50 to 400 Hz 1.5 kHz 11 to 32.9999 µF 0.3 + 30 0.4 + 30 1 nF 50 to 400 Hz 800 Hz 33 to 109.999 µF 0.38 + 100 0.5 + 100 10 nF 50 to 200 Hz 400 Hz 110 to 329.999 µF 0.5 + 300 0.7 + 300 10 nF 50 to 100 Hz 200 Hz 1 + 300 1 + 300 100 nF 50 to 100 Hz 150 Hz 0.33 to 1.1 mF Notes 1. The output is continuously variable from 330 pF to 1.1 mF. 2. Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. 3. For all ranges, the maximum charge and discharge current is 150 mA peak. The peak voltage is 4V, except the 0.33 to 1.1 mF range is limited to 1V. 4. The maximum lead resistance for no additional error in 2-wire COMP mode is 10Ω. 12 Fluke Corporation To Meet Specification 5520A/5500A Multi-Product Calibrators Extended Specifications Temperature Calibration Specifications 5520A/5500A Thermocouple Absolute Uncertainty, tcal ± 5°C ± (°C)1 Source/Measure Range TC Type B C E J K L N R S T U 1 Minimum Maximum 90 Days 600°C 800°C 0.42°C 1 Year 0.44°C 800°C 1000°C 0.34°C 0.34°C 1000°C 1550°C 0.30°C 0.30°C 1550°C 1820°C 0.26°C 0.33°C 0°C 150°C 0.23°C 0.30°C 0.26°C 150°C 650°C 0.19°C 650°C 1000°C 0.23°C 0.31°C 1000°C 1800°C 0.38°C 0.50°C 1800°C 2316°C 0.63°C 0.84°C -250°C -100°C 0.38°C 0.50°C -100°C -25°C 0.12°C 0.16°C -25°C 350°C 0.10°C 0.14°C 350°C 650°C 0.12°C 0.16°C 650°C 1000°C 0.16°C 0.21°C -210°C -100°C 0.20°C 0.27°C -100°C -30°C 0.12°C 0.16°C -30°C 150°C 0.10°C 0.14°C 150°C 760°C 0.13°C 0.17°C 760°C 1200°C 0.18°C 0.23°C -200°C -100°C 0.25°C 0.33°C -100°C -25°C 0.14°C 0.18°C -25°C 120°C 0.12°C 0.16°C 120°C 1000°C 0.19°C 0.26°C 1000°C 1372°C 0.30°C 0.40°C -200°C -100°C 0.37°C 0.37°C -100°C 800°C 0.26°C 0.26°C 800°C 900°C 0.17°C 0.17°C -200°C -100°C 0.30°C 0.40°C 0.22°C -100°C -25°C 0.17°C -25°C 120°C 0.15°C 0.19°C 120°C 410°C 0.14°C 0.18°C 410°C 1300°C 0.21°C 0.27°C 0°C 250°C 0.48°C 0.57°C 250°C 400°C 0.28°C 0.35°C 400°C 1000°C 0.26°C 0.33°C 1000°C 1767°C 0.30°C 0.40°C 0°C 250°C 0.47°C 0.47°C 250°C 1000°C 0.30°C 0.36°C 1000°C 1400°C 0.28°C 0.37°C 1400°C 1767°C 0.34°C 0.46°C -250°C -150°C 0.48°C 0.63°C -150°C 0°C 0.18°C 0.24°C 0.16°C 0°C 120°C 0.12°C 120°C 400°C 0.10°C 0.14°C -200°C 0°C 0.56°C 0.56°C 0°C 600°C 0.27°C 0.27°C Does not include thermocouple error. Notes 1. Resolution is 0.01°C. 2. Temperature standard ITS-90 or IPTS-68 is selectable. 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 13 Temperature Calibration Specifications 5520A/5500A RTD Absolute Uncertainty, tcal ± 5°C ± (°C)1 Range RTD Type Pt 385, 100Ω Pt 3926, 100Ω Pt 3916, 100Ω Pt 385, 200Ω Pt 385, 500Ω Pt 385, 1000Ω Ni120, 120Ω Cu 427, 10Ω 2 1 2 14 Fluke Corporation Minimum Maximum 90 Days 1 Year -200°C -80°C -80°C 0.04°C 0.05°C 0°C 0.05°C 0.05°C 0°C 100°C 0.07°C 0.07°C 100°C 300°C 0.08°C 0.09°C 300°C 400°C 0.09°C 0.10°C 400°C 630°C 0.10°C 0.12°C 630°C 800°C 0.21°C 0.23°C -200°C -80°C 0.04°C 0.05°C -80°C 0°C 0.05°C 0.05°C 0°C 100°C 0.07°C 0.07°C 100°C 300°C 0.08°C 0.09°C 300°C 400°C 0.09°C 0.10°C 400°C 630°C 0.10°C 0.12°C -200°C -190°C 0.25°C 0.25°C -190°C -80°C 0.04°C 0.04°C -80°C 0°C 0.05°C 0.05°C 0.06°C 0°C 100°C 0.06°C 100°C 260°C 0.06°C 0.07°C 260°C 300°C 0.07°C 0.08°C 300°C 400°C 0.08°C 0.09°C 400°C 600°C 0.08°C 0.10°C 600°C 630°C 0.21°C 0.23°C -200°C -80°C 0.03°C 0.04°C -80°C 0°C 0.03°C 0.04°C 0°C 100°C 0.04°C 0.04°C 0.05°C 100°C 260°C 0.04°C 260°C 300°C 0.11°C 0.12°C 300°C 400°C 0.12°C 0.13°C 0.14°C 400°C 600°C 0.12°C 600°C 630°C 0.14°C 0.16°C -200°C -80°C 0.03°C 0.04°C -80°C 0°C 0.04°C 0.05°C 0°C 100°C 0.05°C 0.05°C 0.06°C 100°C 260°C 0.06°C 260°C 300°C 0.07°C 0.08°C 300°C 400°C 0.07°C 0.08°C 0.09°C 400°C 600°C 0.08°C 600°C 630°C 0.09°C 0.11°C -200°C -80°C 0.03°C 0.03°C -80°C 0°C 0.03°C 0.03°C 0°C 100°C 0.03°C 0.04°C 100°C 260°C 0.04°C 0.05°C 260°C 300°C 0.05°C 0.06°C 0.07°C 300°C 400°C 0.05°C 400°C 600°C 0.06°C 0.07°C 600°C 630°C 0.22°C 0.23°C -80°C 0°C 0.06°C 0.08°C 0°C 100°C 0.07°C 0.08°C 100°C 260°C 0.13°C 0.14°C -100°C 260°C 0.30°C 0.30°C Applies for COMP OFF (to the 5520A or 5500A front panel NORMAL terminals) and 2-wire and 4-wire compensation. Based on MINCO Application Aid No. 18. Notes 1. Resolution is 0.003°C. 2. Temperature standard ITS-90 or IPTS-68 is selectable. 5520A/5500A Multi-Product Calibrators Extended Specifications DC Power Summary Specifications 5520A Absolute Uncertainty, tcal ± 5°C ± (% of watts output) Current Range Voltage Range 0.33 to 329.99 mA 0.33 to 10.9999A 11 to 20.5A 90 days 33 mV to 1000V 0.021% 0.055% 0.10% 1 year 33 mV to 1000V 0.023% 0.07% 0.12% Note: Maximum voltage output is 1020V. 5500A Absolute Uncertainty, tcal ± 5°C ± (% of watts output) Current Range Voltage Range 3.3 to 329.99 mA 0.33 to 4.4999A 90 days 33 mV to 1000V 0.03% 0.08% 0.06% 1 year 33 mV to 1000V 0.04% 0.12% 0.09% 5725A Amplifier Current Range 1.5 to 4.4999A 4.5 to 11A 4.5 to 11A 90 days 33 mV to 1000V 0.09% 0.07% 1 year 33 mV to 1000V 0.1% 0.08% Note: Maximum voltage output is 1020V. 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 15 AC Power Summary Specifications 5520A (45 Hz to 65 Hz) PF = 1 Absolute Uncertainty, tcal ± 5°C ± (% of watts output) Current Range Voltage Range 90 days 1 year 90 days 1 year 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA 33 to 329.999 mV 0.13% 0.09% 0.13% 0.09% 330 mV to 1000V 0.11% 0.07% 0.11% 0.07% 33 to 329.999 mV 0.14% 0.10% 0.14% 0.10% 330 mV to 1000V 0.12% 0.08% 0.12% 0.08% 11 to 20.5A 0.33 to 0.8999A 0.9 to 2.9999A 3 to 10.9999A 33 to 329.999 mV 0.12% 0.10% 0.12% 0.10% 330 mV to 1000V 0.10% 0.08% 0.11% 0.09% 33 to 329.999 mV 0.13% 0.11% 0.13% 0.16% 330 mV to 1000V 0.11% 0.09% 0.12% 0.19% Note: Maximum voltage output is 1020V. 5500A (45 Hz to 65 Hz) PF = 1 Absolute Uncertainty, tcal ± 5°C ± (% of watts output) Current Range Voltage Range 3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA 0.3% 0.2% 0.25% 0.2% 90 days 33 to 329.999 mV 330 mV to 1000V 0.2% 0.12% 0.2% 0.12% 1 year 33 to 329.999 mV 0.4% 0.25% 0.35% 0.25% 330 mV to 1000V 0.25% 0.15% 0.25% 0.15% 0.12% 5725 Amplifier 90 days 100 to 1000V 0.2% 0.12% 0.2% 1 year 100 to 1000V 0.25% 0.15% 0.25% 0.15% 0.33 to 0.8999A 0.9 to 2.1999A 2.2 to 4.4999A 4.5 to 11A 5500A 90 days 1 year 33 to 329.999 mV 0.25% 0.2% 0.25% 0.2% 330 mV to 1000V 0.2% 0.12% 0.18% 0.12% 33 to 329.999 mV 0.35% 0.25% 0.35% 0.25% 330 mV to 1000V 0.25% 0.15% 0.2% 0.15% 5725 Amplifier 90 days 100 to 1000V 0.2% 0.12% 0.18% 0.12% 1 year 100 to 1000V 0.25% 0.15% 0.20% 0.15% Note: Maximum voltage output is 1020V. 5520A/5500A Phase 1 Year Absolute Uncertainty, tcal ± 5°C, (∆Φ degrees) 10 to 65 Hz 65 to 500 Hz 500 to 1 kHz 1k to 5 kHz 5k to 10 kHz 10k to 30 kHz 5520A 0.07° 0.25° 0.5° 2.5° 5° 10° 5500A 0.15° 0.9° 2.0° 6° 10° — Notes 1. The phase adjustment range for dual ac outputs is 0 to ± 179.99 degrees. 2. The phase resolution for dual ac outputs is 0.01 degrees for the 5520A and 0.02 degrees for the 5500A. 16 Fluke Corporation 5520A/5500A Multi-Product Calibrators Extended Specifications Frequency Specifications 5520A Frequency Range Resolution 0.01 to 119.99 Hz 0.01 Hz 120.0 to 1199.9 Hz 0.1 Hz 1.200 to 11.999 kHz 1.0 Hz 12.00 to 119.99 kHz 10 Hz 120.0 to 1199.9 kHz 100 Hz 1.200 to 2.000 MHz 1 kHz 1 Year Absolute Uncertainty, tcal ± 5°C 2.5 ppm, ± 5 µHz Jitter 100 nS 5500A 1 2 Frequency Range Resolution 1 Year Absolute Uncertainty, tcal ± 5°C ± (ppm of output + mHz) Jitter 2 µs 0.01 to 119.99 Hz 0.01 Hz 25 + 1 120.0 to 1199.9 Hz 0.1 Hz 25 + 1 2 µs 1.200 to 11.999 kHz 1.0 Hz 25 + 11 2 µs 2 12.00 to 119.99 kHz 10 Hz 25 + 15 140 ns 120.0 to 1199.9 kHz 100 Hz 25 + 15 140 ns 1.200 to 2.000 MHz 1 kHz 25 + 15 140 ns Uncertainty is ± 25 ppm + 15 mHz above 10 kHz. Jitter above 10 kHz is <140 ns. 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 17 Extended AC Voltage Specifications 5520A Sinewave Extended Bandwidth Ranges Frequency 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Voltage Resolution Normal Channel (Single Output Mode) 1 to 33 mV Two digits, e.g., 25 mV 34 to 330 mV 0.4 to 3.3V Three digits 0.01 to 9.99 Hz 5.0% + 0.5% Two digits 4 to 33V 10 Hz to 500 kHz 0.3 to 3.3V See “5520A AC Voltage (Sinewaves) Specifications” 500.1 kHz to 1 MHz -8 dB at 1 MHz, typical 1.001 to 2 MHz -26 dB at 2 MHz, typical Two digits Auxiliary Output (Dual Output Mode) 10 to 330 mV 0.4 to 5.0V Three digits 0.01 to 9.99 Hz 5.0% + 0.5% 10 Hz to 30 kHz Two digits See “5520A AC Voltage (Sinewaves) Specifications” 5500A Sinewave Extended Bandwidth Ranges Frequency 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Voltage Resolution Normal Channel (Single Output Mode) 1 to 33 mV Two digits, e.g., 25 mV 34 to 330 mV 0.4 to 3.3V 4 to 33V 0.3 to 3.3V Three digits 0.01 to 9.99 Hz 5.0% + 0.5% Two digits 10 Hz to 500 kHz See “5500A AC Voltage (Sinewaves) Specifications” 500.1 kHz to 1 MHz -8 dB at 1 MHz, typical 1.001 to 2 MHz -32 dB at 2 MHz, typical Two digits Auxiliary Output (Dual Output Mode) 10 to 330 mV Three digits 0.01 to 9.99 Hz 0.4 to 3.3V 18 Fluke Corporation 5.0% + 0.5% 10 Hz to 10 kHz 5520A/5500A Multi-Product Calibrators Extended Specifications Two digits See “5500A AC Voltage (Sinewaves) Specifications” AC Voltage (Non-Sinewave) Specifications 5520A/5500A Trianglewave and Truncated Sine Ranges Peak-to-Peak 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Voltage Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits on each range 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 20 kHz 0.5% + 0.25% 20 to 100 kHz 5.0% + 0.5% Frequency Normal Channel (Single Output Mode) 2.9 mV to 93V Six digits on each range 5520A Auxiliary Output (Dual Output Mode) 29 mV to 14V 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% Two digits on each range 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 5.0% + 0.5% Six digits on each range 5500A Auxiliary Output (Dual Output Mode) 29 mV to 9.3V Squarewave Ranges Peak-to-Peak 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% Two digits on each range 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 5.0% + 0.5% Frequency 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Voltage Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits on each range 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 20 kHz 0.5% + 0.25% 20 to 100 kHz 5.0% + 0.5% Six digits on each range Normal Channel (Single Output Mode) 2.9 mV to 66V Six digits on each range 5520A Auxiliary Output (Dual Output Mode) 29 mV to 14V 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% Two digits on each range 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 5.0% + 0.5% Six digits on each range 5500A Auxiliary Output (Dual Output Mode) 29 mV to 6.6V 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% Two digits on each range 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 5.0% + 0.5% Six digits on each range 5520A/5500A Squarewave Characteristics Risetime @ 1 kHz Typical <1 µs 1 Settling Time @ 1 kHz Typical Overshoot @ 1 kHz Typical Duty Cycle Range Duty Cycle Uncertainty 1 <10 µs to 1% of final value <2% 1% to 99%, <3.3V p-p, 0.01 Hz to 100 kHz ± (0.8% of period + 100 ns) For duty cycles of 10.00% to 90.00%. 5520A/5500A Trianglewave Characteristics (typical) Linearity to 1 kHz Aberrations 0.3% of p-p value, from 10% to 90% point <1% of p-p value, with amplitude >50% of range 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 19 AC Current (Non-Sinewave) Specifications 5520A (Trianglewave and Truncated Sinewave) 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Current Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits, e.g., 0.75 mA 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% Ranges Peak-to-Peak Frequency 0.047 to 0.92999 mA 0.93 to 9.29999 mA 9.3 to 92.9999 mA 93 to 929.999 mA 0.93 to 8.49999A 8.5 to 57A 20 Fluke Corporation 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.5% 1 to 10 kHz 10.0% + 2.0% 10 to 45 Hz 0.5% + 1.0% 45 Hz to 1 kHz 0.5% + 0.5% 1 to 10 kHz 10.0% + 2.0% 45 to 500 Hz 0.5% + 0.5% 500 Hz to 1 kHz 1.0% + 1.0% 5520A/5500A Multi-Product Calibrators Extended Specifications Six digits on each range Two digits on each range Six digits on each range Two digits on each range Six digits on each range Two digits on each range Six digits on each range 5520A (Squarewave) Ranges Peak-to-Peak 0.047 to 0.65999 mA 0.66 to 6.59999 mA 6.6 to 65.9999 mA 66 to 659.999 mA 0.66 to 5.99999A 6 to 41A 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Current Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits, e.g., 50 mA 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% Frequency 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 10.0% + 2.0% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.5% 1 to 10 kHz 10.0% + 2.0% 10 to 45 Hz 0.5% + 1.0% 45 Hz to 1 kHz 0.5% + 0.5% 1 to 10 kHz 10.0% + 2.0% 45 to 500 Hz 0.5% + 0.5% 500 Hz to 1 kHz 1.0% + 1.0% Six digits on each range Two digits on each range Six digits on each range Two digits on each range Six digits on each range Two digits on each range Six digits on each range 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 21 AC Current (Non-Sinewave) Specifications 5500A (Trianglewave and Truncated Sinewave) Ranges Peak-to-Peak 0.047 to 92.999 mA 93 to 929.999 mA 9.3 to 6.19999A 6.2 to 31A Frequency 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Current Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits, e.g., 0.50 mA 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% 1 to 10 kHz 5.0% + 0.5% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.5% 1 to 10 kHz 0.5% + 1.0% Six digits on each range Two digits on each range Six digits on each range 10 to 45 Hz 5.0% + 1.0% 45 Hz to 1 kHz 0.5% + 0.5% Two digits on each range 1 to 10 kHz 5.0% + 1.0% 45 to 500 Hz 2.0% + 0.5% Two digits on each range 500 to 1 kHz 5.0% + 1.0% Six digits on each range Frequency 1 Year Absolute Uncertainty, tcal ± 5°C ± (% of output + % of range) Maximum Current Resolution 0.01 to 10 Hz 5.0% + 0.5% Two digits, e.g., 50 mA 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.25% Six digits on each range 5500A (Squarewave) Ranges Peak-to-Peak 0.047 to 65.999 mA 66 to 659.999 mA 0.66 to 4.39999A 4.4 to 22A 22 Fluke Corporation 1 to 10 kHz 5.0% + 0.5% 0.01 to 10 Hz 5.0% + 0.5% 10 to 45 Hz 0.25% + 0.5% 45 Hz to 1 kHz 0.25% + 0.5% 1 to 10 kHz 0.5% + 1.0% 10 to 45 Hz 5.0% + 1.0% 45 Hz to 1 kHz 0.5% + 0.5% 1 to 10 kHz 5.0% + 1.0% Six digits on each range Two digits on each range Six digits on each range Two digits on each range Six digits on each range 45 to 500 Hz 2.0% + 0.5% Two digits on each range 500 Hz to 1 kHz 5.0% + 1.0% Six digits on each range 5520A/5500A Multi-Product Calibrators Extended Specifications Ordering Information Ordering Information 5500A Multi-Product Calibrator 5520A High Performance Multi-Product Calibrator 25 in. (63.5 cm) 21.25 in. (54.0 cm) Options 5500A-SC300 300 MHz/1 ns Oscilloscope Calibration Option 5500A-SC600 600 MHz/300 ps Oscilloscope Calibration Option 20.25 in. (51.4 cm) 9.5 in. 19 (24.1 cm) (49 .5 .5 in. cm ) 29.75 in. (75.6 cm) 11.25 in. (28.6 cm) The TC100 Instrument Cart provides three levels of work surface, each of which accommodates instruments up to 48.26 cm (19 in) wide. The top shelf can be adjusted over an angle of 25° and it has a safety lock. The cart has a capacity of 136.08 kg (300 lb). Four casters, two of which can be locked, provide excellent mobility and maneuverability. Options include a line power strip, accessories drawer and a CPU bracket. Cart Size: 51.435 cm (20.25 in) W x 53.6575 cm (21.25 in) D x 75.565 cm (29.75 in) H Drawer Size: 45.085 cm (17.75 in) W x 49.53 cm (19.5 in) D x 7.62 cm (3 in) H CPU Bracket Size: 38.1 cm (15 in) W x 13.97 cm (5.5 in) D x 43.18 cm (17 in) H Accessories 5500A/COIL 5500A/CASE 5500A/LEADS 5500A/HNDL Y5537 TC100 5725A MET/CAL® 5500/CAL 50-Turn Current Coil Roll-Aboard Transit Case Comprehensive Test Lead Kit Side Carry Handle Rack Mount Kit Test Cart Amplifier (5500A Only) Calibration Software (IEEE and RS232) Calibration Software (RS232) Pressure Modules (5520A only) FLUKE-700PCK Pressure Module Calibration Kit (Required) FLUKE-700P01 Pressure Module 0-10 IN. H20 DIFF FLUKE-700P02 Pressure Module 0-1 PSID FLUKE-700P03 Pressure Module 0-5 PSID FLUKE-700P04 Pressure Module 0-15 PSID FLUKE-700P05 Pressure Module 0-30 PSIG FLUKE-700P06 Pressure Module 0-100 PSIG FLUKE-700P07 Pressure Module 0-500 PSIG FLUKE-700P08 Pressure Module 0-1000 PSIG FLUKE-700P09 Pressure Module 1500PSIG FLUKE-700P22 Pressure Module 0-1 PSID WET FLUKE-700P23 Pressure Module 0-5 PSID WET FLUKE-700P24 Pressure Module 0-15 PSID WET FLUKE-700P29 Pressure Module 3000 PSIG WET FLUKE-700P30 Pressure Module 5000 PSIG WET FLUKE-700P31 Pressure Module 10000 PSIG WET FLUKE-700PA3 Pressure Module 0-5 PSIA FLUKE-700PA4 Pressure Module 0-15 PSIA FLUKE-700PA5 Pressure Module 0-30 PSIA FLUKE-700PA6 Pressure Module 0-100 PSIA FLUKE-700PD2 Pressure Module ± 1 PSID FLUKE-700PD3 Pressure Module ± 5 PSID FLUKE-700PD4 Pressure Module ± 15 PSID FLUKE-700PD5 Pressure Module -15+30 PSIG FLUKE-700PD6 Pressure Module -15+100 PSIG FLUKE-700PD7 Pressure Module -15+200 PSIG FLUKE-700PV3 Pressure Module -5 PSID FLUKE-700PV4 Pressure Module -15 PSID FLUKE-700PMP Pressure Pump 5520A/5500A Multi-Product Calibrators Extended Specifications Fluke Corporation 23 43.2 cm (17 in) 5500A CALIBRATOR NORMAL SCOPE AUX V, , RTD A, -SENSE, AUX V 200V PK MAX STBY OPR EARTH SCOPE HI 1000V RMS MAX 20V RMS MAX TRIG OUT LO 7 8 9 4 5 6 PREV MENU BOOST µ dBm m n V W k sec Hz SETUP ¡C NEW REF CE F MEAS TC TRIG OUT MULT DIV ¡F A 1V PK MAX 20V PK MAX 1 2 3 +/ 0 • p M TC 20V PK MAX SHIFT ENTER x 17.8 cm (7 in) EDIT FIELD RESET POWER ÷ I O Front view and dimensions. 47.0 cm (18.5 in) 6.4 cm (2.5 in) For Cable Access Side view and dimensions. Note: 5500A and 5520A dimensions are the same. YM IT ANAGEMENT SY S M TE QU AL Fluke. Keeping your world up and running. CE R TI 00 1 ISO 9001 FI E O D T O EET IS M 9 Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: U.S.A. (800) 443-5853 or Fax (425) 356-5116 Europe/M-East (31 40) 2 678 200 or Fax (31 40) 2 678 222 Canada (905) 890-7600 or Fax (905) 890-6866 Other countries (425) 356-5500 or Fax (425) 356-5116 Web access: http://www.fluke.com ©1997 Fluke Corporation. All rights reserved. Printed in U.S.A. 12/97 J0871UEN Rev A